key: cord-0929131-7sndlov9 authors: Agaoglu, Nihat Bugra; Yıldız, Jale; Dogan, Ozlem Akgun; Alkurt, Gizem; Kose, Betsi; Demirkol, Yasemin Kendir; Irvem, Arzu; Doğanay, Levent; Doganay, Gizem Dinler title: COVID-19 PCR Test Performance for Samples Stored at Ambient Temperature date: 2020-06-16 journal: bioRxiv DOI: 10.1101/2020.06.15.153882 sha: 417abf992fc6944393ee2edcfb24bdd4635d4fc8 doc_id: 929131 cord_uid: 7sndlov9 Background The new type of Coronavirus infection had become a pandemic in a very short period since it was first seen in Wuhan. The outbreak had a negative impact on all health care systems throughout the world and overwhelmed the diagnostic laboratories as well. During the pandemic, handling patient specimens in accordance with the universal guidelines was troublesome as WHO, CDC and ECDC required cold chain compliance during transporting and storing the swap samples. Materials and methods In this study, we tested diagnostic performance of RT-PCR on 30 swab samples stored at ambient temperature and compared them with the samples stored at +4°C. Results Our results revealed that all the samples stored at ambient temperature remain PCR positive for at least five days. We did not see any false negativity. Conclusion In conclusion, we report that transferring and storing of nasopharyngeal/oropharyngeal samples at ambient temperature could be possible in the resource-limited conditions like pandemic. Coronavirus disease 19 (COVID-19), which stems from a new type of coronavirus, severe 42 acute respiratory syndrome coronavirus 2 (SARS-CoV-2), has become a universal pandemic 43 since it first appeared in Wuhan, China in November 2019. Until the 10th of June, there were 44 7,127,753 confirmed cases, 407,159 deaths, and 108,918 new cases (WHO COVID-19 Dash-45 board) 1 . With increasing number of infected people in a such short period of time, the pan-46 demic has been an enormous burden to healthcare system including diagnostic laboratories. 47 Real-time polymerase chain reaction (RT-PCR) is considered as the gold-standard confirma-48 tory diagnostic laboratory test for CoVID-19 2 . Samples for this method are obtained from the 49 upper and lower respiratory tract (oropharyngeal, nasopharyngeal swabs, sputum, lower res-50 piratory tract aspirates, bronchoalveolar lavage, and nasopharyngeal wash/aspirate or nasal 51 aspirate) 3 and put into a transport medium. The World Health Organization (WHO), Centers 52 for Disease Control and Prevention (CDC), European Centre for Disease Prevention and Con-53 trol (ECDC) and several national health authorities announced several RT-PCR protocols and 54 sample storage guidelines and all emphasized that accuracy of the RT-PCR tests mostly relies 55 on proper specimen collection and storage 4-6 . All guidelines agreed that samples should be 56 transferred and kept in conditions maintaining cold chain. However, processing outrageous 57 number of samples during the outbreak hampers keeping transport and storing conditions in 58 line with the guidelines. Here, we aimed to test the RT-PCR diagnostic performance on naso-59 pharyngeal/oropharyngeal swab samples stored at ambient temperature. 60 samples. 2 mL of VTM solution was added to the samples. Then these tubes were vortexes 67 and aliquoted into two separate tubes (2 mL of solution for each tube) to be stored at refriger-68 ator (+4°C) and room temperature (20-24°C), respectively. Additionally, five RT-PCR nega-69 tive samples were stored under the same conditions as a control group. The room temperature 70 and refrigerator temperature were monitored continuously and recorded using data loggers 71 ( Fig. 1) . 72 All samples (+4°C, room temperature, and controls) were tested every 24 hours for 9 days by 74 RT-PCR with the SARS-CoV-2 detection kit (Coyote Bioscience Co., Ltd) according to the 75 protocols provided by the manufacturer. 76 The primers of the commercial kit designed to target ORF1ab and N gene of SARS-CoV-2. 77 All tests were done with Biorad CFX 96 platform. The cut off of Ct value was taken below 29 78 for the detection of positive results according to the suggestions of the manufacturer. 79 A positive result is reported if both the ORFlab and N gene were positive (Fig. 2) . If both tar-81 gets were negative, the result is negative. When only one target region is positive then the test 82 was repeated. 83 84 5 A total of 30 SARS-CoV-2 positive samples and 5 negative samples (all stored in +4°C and 86 ambient temperature) were studied. All positive samples were run in PCR every day, and neg-87 ative samples were studied on day 5 and day 9. 88 On day 6, 2 samples (in positive group) failed to run due to insufficient material. On day 7, 12 89 samples, on day 8, 17 and on the 9th day 28 samples had insufficient material for the PCR 90 assay. Study is terminated at day 9 (Table 1) . 91 Five PCR negative samples were aliquoted and stored in refrigerator and ambient temperature 92 as control samples. These controls tested on day 5 and day 9, all remained negative. The cycle 93 values obtained are indicated in the tables (Table S1, Table S2, Table S3, Table S4 ). 94 PCR results of daily tested study samples are given in Table 1 . On day 1, 3 samples in +4°C 95 group and 4 samples in RT group resulted as "test repeat". Ultimate PCR results for all these 96 samples were positive. On day 2, 3 and 5, only one sample in +4°C group tested negative, in-97 terestingly this sample showed a positive result on day 4. After day 7 the available test mate-98 rial in some tubes remained insufficient for the PCR assay. Till day 7 test sensitivity in ambi-99 ent temperature was 100%. We did not observe false negative in room temperature group 100 even for the sample showing variation at +4°C. 101 Samples stored at +4°C showed instability in the test results by day 9 and also remaining 102 samples in tubes were not sufficient for continuing PCR testing. On the other hand, samples 103 kept at room temperature revealed higher consistency without showing any false negativity. tively. Also, they all suggest storing at -70°C for samples that needs to be stored for more 116 than 5 days or whose transfer will be delayed. However, this restrictive suggestion causes lo-117 gistic and cost problems related to the transportation and storage of samples especially in 118 pandemic periods when massive amounts of samples have to be analyzed. In addition, many 119 laboratories, following the CDC and WHO recommendations, reject samples that cannot be 120 delivered to them under the recommended transport conditions. This situation requires re-121 sampling, where both creates discomfort for the patient and causes delays in diagnosis. 122 It is obvious that keeping the number of RT-PCR tests as high as possible is the most crucial 123 step in preventing the spread of infection. For this purpose, it is also very important to use ex-124 isting infrastructure in an efficient way. In light of the data we obtained from our study, we 125 suggest that in a resource-limited setting like pandemics, transferring and storing of nasopha-126 ryngeal/oropharyngeal samples at ambient temperature (20-24℃) should not be considered as 127 inadequate, and these samples should not be rejected, can be analyzed and reported. Such a 128 change in daily practice will result in considerable time and cost savings as well as a reduced 7 number of sample rejection. When sample transportation or storage at cold chain conditions 130 becomes a limiting factor for the pandemic laboratory, keeping samples at ambient tempera-131 ture will enable much more testing. 132 The limitation of our study is the considered range of temperatures, between 20°C and 24°C. 133 Temperatures higher than 24°C may be common in some tropical research and/or storage set-134 tings so the time period that samples remain positive may be variable. Also, the limited num-135 ber of samples can be counted as another limitation of our study. Epi-151 demiology, Diagnosis, Treatment, and Control Laboratory diagnosis of emerging human coro-156 navirus infections -the state of the art Methodology for estimating point prevalence of SARS-CoV-2 infection by pooled RT-PCR 158 estimating-point-prevalence%20-SARS-CoV-2-infection-pooled-RT-PCR-testing.pdf 160 5. Laboratory testing for coronavirus disease (COVID-19) in suspected human cases Interim Guidelines for Collecting, Handling, and Testing Clinical Specimens from Persons 164 for Coronavirus Disease Responding to COVID-19 in Istanbul: Perspective from genomic laborato-168 ry Possi-170 ble environmental effects on the spread of COVID-19 in China. The Science of the to-171 tal environment 731 Association between ambient temperature and COVID-19 in-173 fection in 122 cities from China. The Science of the total environment The role of environmental factors to transmission of 175 SARS-CoV-2 (COVID-19) Consistent viral DNA quantification after prolonged storage at ambient 178 temperature Evaluation of swabs, 180 transport media, and specimen transport conditions for optimal detection of viruses by 181 PCR