key: cord-0926780-ie875yi4 authors: Nicholson, Suellen; Karapanagiotidis, Theo; Khvorov, Arseniy; Douros, Celia; Mordant, Francesca; Bond, Katherine; Druce, Julian; Williamson, Deborah A; Purcell, Damian; Lewin, Sharon R; Sullivan, Sheena; Subbarao, Kanta; Catton, Mike title: Evaluation of six commercial SARS-CoV-2 serology assays detecting different antibodies for clinical testing and serosurveillance date: 2021-05-10 journal: Open Forum Infect Dis DOI: 10.1093/ofid/ofab239 sha: 499230a06df4cab41be4f299aa3fd561a17d1b8f doc_id: 926780 cord_uid: ie875yi4 BACKGROUND: Serological testing for SARS-CoV-2 complements nucleic acid tests for patient diagnosis and enables monitoring of population susceptibility to inform the COVID-19 pandemic response. It is important to understand the reliability of assays with different antigen or antibody targets to detect humoral immunity after SARS-CoV-2 infection and to understand how antibody (Ab) binding assays compare to those detecting neutralizing antibody (nAb), particularly as we move into the era of vaccines. METHODS: We evaluated the performance of six commercially available Enzyme-linked Immunosorbent Assays (ELISAs), including a surrogate virus neutralization test (sVNT), for detection of SARS-CoV-2 immunoglobulins (IgA, IgM, IgG), total or nAb. A result subset was compared to a cell culture-based microneutralisation (MN) assay. We tested sera from patients with prior RT-PCR confirmed SARS-CoV-2 infection, pre-pandemic sera and potential cross-reactive sera from patients with other non-COVID-19 acute infections. RESULTS: For sera collected > 14 days post-symptom onset, the assay achieving the highest sensitivity was the Wantai total Ab at 100% (95% confidence interval: 94.6-100) followed by 93.1% for Euroimmun NCP-IgG, 93.1% for GenScript sVNT, 90.3% for Euroimmun S1-IgG, 88.9% for Euroimmun S1-IgA and 83.3% for Wantai IgM. Specificity for the best performing assay was 99.5% for the Wantai total Ab and for the lowest performing assay was 97.1% for sVNT (as per IFU). The Wantai Total Ab had the best agreement with MN at 98% followed by Euroimmun S1-IgA, Euro NCP-IgG and sVNT (as per IFU) with (97%, 97% and 95% respectively) and Wantai IgM having the poorest agreement at 93%. CONCLUSION: Performance characteristics of the SARS-CoV-2 serology assays detecting different antibody types are consistent with those found in previously published reports. Evaluation of the surrogate virus neutralization test in comparison to the Ab binding assays and a cell culture-based neutralization assay showed good result correlation between all assays. However correlation between the cell-based neutralization test and some assays detecting Ab’s not specifically involved in neutralization was higher than with the sVNT. This study demonstrates the reliability of different assays to detect the humoral immune response following SARS-CoV-2 infection, which can be used to optimise serological test algorithms for assessing antibody responses post SARS-CoV-2 infection or vaccination. Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) the causative agent of Coronavirus disease 2019 has infected > 131 million people and caused > 2.8 million deaths globally as of 8 April 2021 [1] . COVID-19 manifests as an acute respiratory illness, although asymptomatic infections occur [2] . Laboratory diagnosis of COVID-19 primarily relies on detection of viral RNA by RT-PCR in respiratory tract samples or antigen tests, the sensitivity of which declines as infection resolves. In contrast, antibody (Ab) is detected in most individuals 10-15 days following onset of symptoms [3] . Serological testing can detect previous infection in people who have recovered without RT-PCR testing, who are RT-PCR negative [4, 5] , or whose RT-PCR results are difficult to interpret [6] . Serology may also inform our understanding of antibody longevity and quantification of neutralising antibody (nAb) response in vaccine trials [6, 7, 8] . As the pandemic progresses estimates of prior exposure and population prevalence from serological assays will play an increasingly important role for public health decision making [9, 10] . However, there is substantial variation in assay performance depending on the antibody type detected and correlation to neutralising antibodies [11] . In this study we evaluate the sensitivity and specificity of six commercially available serological ELISA tests using either spike (S), nucleocapsid (NCP) or receptor binding domain (RBD) antigens for detection of specific isotypes of SARS-CoV-2 (IgA, IgM, IgG), total Ab or nAb (using a surrogate virus neutralization test (sVNT)). We also compared a subset of ELISA results to microneutralisation (MN), the current gold standard assay for SARS-CoV-2-specific antibody. A c c e p t e d M a n u s c r i p t A retrospective study evaluating the sensitivity and specificity of commercially available ELISAs for detection of Abs to SARS-CoV-2 virus on well pedigreed sera. Sera from three patient groups were assembled to assess aspects of assay performance. Euroimmun and sVNT assay sensitivity was assessed using a panel of 147 sera (Panel A) from a group of patients with prior RT-PCR confirmed SARS-CoV-2 infection. A subset of 96 sera (Panel B) were used to evaluate the Wantai assays. The Wantai total Ab assay requires 100ul of sample; 51 Panel A sera had insufficient volume for testing using the Wantai assays. While only 6 (of the 51) sera were from convalescent patients collected > 14 days post-symptom onset; the samples collected > 14 days post-symptom onset represent 49% (72/147) of Panel A and 69% (66/96) of Panel B. Specificity was calculated using 2 groups: a population group of sera representing the Victorian population collected pre-pandemic between 2011 and 2018; (N = 100 to N=312, depending on the assay tested) and a cross-reactive group representing pre-pandemic sera to assess potential crossreactivity; from patients with seasonal coronavirus (HCoV-NL63, HCoV-229E, HCoV-OC43 and HCoV-HKU1), SARS-CoV1, MERS-CoV or other non-COVID-19 acute infections (N= 30 to N=36, depending on the assay tested). See Table 1 and Table 2 for the full list. Samples were obtained from the Victorian Infectious Diseases Reference Laboratory (VIDRL) at The Peter Doherty Institute for Infection and Immunity and Royal Melbourne Hospital (RMH), Victoria, Australia. SARS-CoV-2 RNA was detected in respiratory swabs by RT-PCR and results provided by both institutions. VIDRL used an in-house RT-PCR with previously published primers [12] targeting RdRP gene with positives then tested by a second assay targeting N gene and/or S gene. RMH utilised a commercial Coronavirus Typing assay (AusDiagnostics), a 2-step, heminested multiplex tandem PCR, with 7 coronavirus RNA targets plus a proprietary artificial sequence as an internal control, [12] targeting OFR1a gene; positives were confirmed by GeneXpert, targeting N gene and E gene or tested by VIDRLs' in-house RT-PCR assay. M a n u s c r i p t ELISA ELISA testing was performed in the Serology Laboratory at VIDRL following manufacturers' Instructions for Use (IFU) with results reported semi-quantitatively as either a signal/cut-off ratio (Euroimmun and Wantai) or percentage inhibition (sVNT). Assay performance characteristics were assessed using the panels described previously. Testing was performed manually though high throughput testing could be performed using robotic platforms. Intra-assay variability was calculated by testing 10 in-house quality control sample replicates within the same microtitre plate and inter-assay variability by testing on different plates, on average, on six different days using the same kit lot number; results are reported as a coefficient of variation (CV). Euroimmun anti-SARS-CoV-2 S1-IgG, S1-IgA and anti-SARS-CoV-2 NCP-IgG (Euroimmun Medizinische Labordiagnostika, Lubeck, Germany) These are indirect ELISAs for detection of immunoglobulin (Ig) class IgG or IgA against SARS-CoV2 antigens. Wells are coated with either (i) S protein S1-domain or (ii) modified NCP. SARS-CoV-2 binding antibodies are detected using enzyme-labelled anti-human-IgG or anti-human-IgA conjugates and a colorimetric substrate and are read spectrophotometrically. (Beijing Wantai Biological Pharmacy Enterprise, Beijing China) This is a two-step incubation antigen 'sandwich' assay detecting total antibodies binding the SARSCoV-2 RBD within the S1-subunit of S protein. Patient antibody to SARS-CoV-2 that binds antigen coated on the plate, is bound by horseradish peroxidase (HRP) antigen conjugate forming an antigen-antibody-antigen-HRP complex detectable by colourimetric substrate and read spectrophotometrically. A c c e p t e d M a n u s c r i p t This is a capture ELISA for detection of IgM-class antibodies to SARS-CoV-2 virus. Anti-µ chain antibodies on the plate capture patient IgM antibodies; detection is by recombinant SARS-CoV-2antigen-HRP-conjugate followed by a colourimetric substrate read spectrophotometrically. (GenScript USA, Inc. New Jersey, USA) This is a species and isotype independent blocking ELISA which mimics virus neutralization, detecting circulating neutralizing SARS-CoV-2 antibodies that block the interaction between the viral spike RBD and host angiotensin converting enzyme 2 (ACE2) cell surface receptor. HRP-conjugated recombinant SARS-CoV-2 RBD fragment binds to any circulating nAb to RBD preventing capture by the hACE2 protein on the well which is subsequently removed in the following wash step. Substrate reaction incubation time is determined by temperature; ideal reaction temperature and time in the IFU are 25°C for 15 minutes. For temperatures lower than 25°C, the time can be extended. At 15 minutes our control values did not meet the assay validity criteria, but at 20 minutes they fell within the acceptable ranges. Colour intensity is inversely dependent on the titer of anti-SARS-CoV-2 nAbs. We report three types of alternative estimates: (i) 20% cut-off without retesting (as per the IFU), (ii) 20% cut-off with retesting for equivocal results (18-22%) and (iii) 25% cut-off without retesting. In a subset of sera, a comparison was performed using an in-house MN assay and the commercial ELISAs. To compare MN sensitivity and specificity to the Euroimmun and sVNT assays, eighty-five (Panel A) samples from RT-PCR positive patients were assessed. Seventy samples (Panel B) were assessed for comparison of MN to the Wantai assays, which require a higher volume of sera than the other assays. The MN assay was performed using an in-house assay as described previously [13] . Briefly, the ability of serial 2-fold dilutions of sera to neutralize the infectivity of 100 median tissue culture infectious doses of SARS-CoV-2 was assessed by inhibition of viral cytopathic effect in Vero cells. The nAb titre was calculated using the Reed/Muench method [14, 15] . All statistical analyses were conducted using R version 4.0.2 [16] . Responses were assessed in each of the 3 patient groups (Infected, population, cross-reactive). Sensitivity was estimated using the infected group separately for each of three categories from time of symptom onset: (<7 days, 7-14 days, >14 days). Each observation was treated as independent within each subgroup. Specificity was A c c e p t e d M a n u s c r i p t estimated separately for the population group and the cross-reactive group. The 95% confidence intervals were generated using the exact binomial Clopper-Pearson method with PropCIs R package [17] . Sensitivity estimates (point estimates and CI bounds) were averaged across the onset categories to give the "averaged" sensitivity estimate. Point estimates and interval bounds for the sensitivity and the population group specificity were used to calculate the corresponding estimates and bounds for PPV and NPV at different theoretical levels of population prevalence (0.1%, 0.5%, 1%, 5%, 10% and 20%). Project ethical approval for RMH specimens was obtained from Melbourne Health Human Research Ethics Committee (RMH HREC QA2020052), this included patient's written consent. The in-house panel consisted of anonymised excess diagnostic specimens sent to VIDRL for COVID -19 testing and the 'VIDRL Serum Reference Collection'. Performance characteristics of the six commercial assays are shown in (Table 3) . Increased sensitivity as well as increased antibody level, reported as index values or % inhibition, was observed across all timeframes for all assays (Figure 1 ). For comparative assay sensitivity and specificity see ( Figure 2 ). Euroimmun S1 IgA, S1 IgG and NCP IgG Sensitivity was assessed using sera from subjects with RT-PCR proven COVID-19 disease (Table 3) . When sera collected > 14 days post-symptom onset were considered, sensitivity was 88.9% (79.395.1) for S1-IgA, 90.3% (81.0-96.0) for S1-IgG and 93.1% (84.5-97.7) for NCP-IgG. The averaged sensitivity across all timeframes was 58.0% (44.4-71.1) for S1-IgA, 46.7% (35.4-59.6) for S1-IgG and 51.4% (39.4-64.2) for NCP IgG. Specificity was assessed using the population and cross-reactive groups (Table 3) . Specificity when testing the pre-pandemic population group was lowest at 95.0% (90.7-97.7) using the S1-IgA, 97.2% (93.6-99.1) using S1-IgG and highest 99.0% (96.3-99.9) using NCP-IgG. In the cross-reactive assessment group, the lowest specificity was 83.3% (65.3-94.4) using NCP-IgG, 86.7% (69.3-96.2) using S1-IgA and highest using S1-IgG at 90% (73.5-97.9). Initial testing of the S1-IgA kit gave poorer specificity than reported here; however, with the introduction of a new buffer by Euroimmun, retrospective testing of pre-pandemic samples showed an 8.5% specificity improvement. Cross-reactivity was observed with sera containing anti-SARS-CoV-1 antibody (all tests), anti-MERS antibody (only NCP IgG, 1/9 samples), and sera positive for mycoplasma and CMV (all tests) and M a n u s c r i p t dengue (only S1-IgA, 1/5 samples) antibodies. No cross-reactivity was observed to sera positive for parvovirus-B19 and seasonal coronavirus antibodies (Table 2) . For the COVID-19 RT-PCR positive group, sensitivity of the sVNT was 93.1% (84.5-97.7) when sera collected > 14 days from symptom onset were considered and was 54.4% (42.3-66.7) across all time periods ( Table 2 ). The assay specificity was 97.1% (94.6-98.7) when the pre-pandemic population group were tested. All other assays required re-testing of samples with results within 10-20% of the assay cut-off (the 'equivocal zone'), as described in each manufacturer's IFU. Because the sVNT had similar intra-assay and inter-assay variability to the other assays, we wanted to see if specificity improved with retesting of samples within the 10% equivocal zone. When the pre-pandemic population group sera were tested with repeat testing of equivocal zones, specificity increased to 99.4% (97.7-99.9) with 2 of the 312 samples false positive (both MN negative). Both samples gave results close to the 20% inhibition cut-off (22.5%, 21%). Because the IFU suggests setting a population specific cut-off, we considered a range of thresholds (Supplementary Figure 1a , 1b, 1c & Table 1a , 1b). Using a cut-off value of 25% increased specificity and PPV without changing the sensitivity and NPV (Table 3 ). In the cross-reactivity assessment group cross-reactivity was observed with sera containing antibody to SARS-CoV-1 (Table 2) . The averaged sensitivity for the Wantai Total Ab was 71.6% (52.8-87.2) and for the IgM assay 61.6% (41.8-80.8) which improved to 100.0% (94.6-100.0) and 83.3% (72.1-91.4), respectively, when restricted to sera collected > 14 days post-symptom onset. (Table 3 ). The specificity for the Wantai Total Ab was 99.5% (97.4-100.0) and for IgM was 99.0% (94.6-100.0) in the population group and for the cross-reactive group these were 97.2% (85.8-99.9) and 96.7% (82.8-99.9), respectively. Cross-reactivity was observed with sera containing antibody to SARS-CoV1 (Table 2) . PPV and NPV were calculated across a range of population prevalence estimates:0.1%, 0.5%, 1%, 10% and 20% were calculated for each assay (Supplementary Table 2 and Supplementary Figure 2) . In a low prevalence setting like ours, assuming population prevalence of 0.32% for Victoria and 0.11% for Australia [18] PPV for the best performing assay, Wantai Total Ab was 40.2% and 18.7%, respectively while the NPV was 100% for sera collected >14 days post-symptom onset. M a n u s c r i p t In the RT-PCR positive group there were 41 discordant results, with discordance increasing with time since disease onset from 8/41 at < 7 days post-onset to 19/41 at > 14 days post-onset. In the population and cross-reactive assessment groups there were 17 and 6 discordant results, respectively. In part this appeared attributable to the type of antibody being detected, the antigen targeted in the assay and the assay format. For samples collected > 14 days post-onset, 11 were Wantai IgM negative but Total Ab positive. All 11 samples were collected 21 to > 30 days post onset, suggesting discordance attributable to IgM loss [20] . We retrospectively tested these samples for IgA (results not shown). Six of eleven (55%) were IgA positive, suggesting waning IgM antibody and to a lesser extent IgA (Supplementary Figure 3-Heat-map) . Intra-assay variability was low for all assays, below 10% and ranged from CV=3.1% for the Euroimmun S1-IgA to CV=8.8% for the Euroimmun S1-IgG. Inter-assay variability ranged from Euroimmun NCP-IgG CV=3.1% to Wantai IgM CV=14.9% (Supplementary Table 3a , 3b). Eighty-five samples from RT-PCR positive Panel A were tested by MN (Table 4a ). Seventy three percent (62/85) were MN positive. Euroimmun S1-IgA had equal sensitivity with 73% (62/85) samples positive, 67% (57/85) sVNT positive, 67% (56/84) Euroimmun NCP IgG positive and 62% (53/85) Euroimmun S1 IgG positive. Hence MN and Euroimmun S1-IgA had the highest sensitivity. At > 14 days post symptom onset Euroimmun S1 IgA and Euroimmun NCP-IgG had 100% (38/38) sensitivity, followed by MN and Euroimmun S1 IgG at 97% (37/38). sVNT was 95% (36/38) sensitive. (Table 4a ). The MN assay specificity for the population and cross-reactive assessment sera groups was 100%. Reliable COVID-19 serosurveillance is important for guiding the pandemic response [9, 10] . At the individual level, serology can provide a tool for resolving the diagnosis for patients with infections not confirmed by RT-PCR. At the population level, it provides policymakers with an assessment of overall impact of the pandemic and vaccination efficacy. Assay sensitivities ranged from 46.7% to 71.6%; and specificities from 95.0% to 99.5% (pre-pandemic population group), and from 83.3% to 97.2% (cross-reactivity assessment group). M a n u s c r i p t Sensitivity for sera collected >14 days post-symptom onset ranged from 83.3% to 100%. Consistent with the biology of an immune response, and as reported by others, when results from all timepoints were considered, sensitivity was low, but increased substantially when only sera collected >14 days post-onset were assessed [19] , underlining the need to wait a sufficiently long period before confirming infection status by serology. The Euroimmun NCP-IgG and sVNT had good sensitivity, however the highest sensitivity was achieved by Wantai Total Ab at 100%. The Euroimmun NCP-IgG and Wantai IgM had better pre-pandemic population group specificity at 99%, than sVNT at 97.1%. Wantai Total Ab achieved highest specificity in the cross-reactivity assessment group at 97.2%. We observed improved sVNT specificity when we varied from the IFU to use a +/-10% repeat equivocal zone however optimal specificity was achieved when we adjusted the cut-off to 25%, without reducing the sensitivity and NPV. (Table 3) Cross-reactivity was observed in all assays with sera containing anti-SARS-CoV-1 antibody. Crossreactivity was more commonly seen in the Euroimmun assays, and depending on the assay, in sera positive for MERS, mycoplasma, CMV and dengue antibodies, consistent with other reports [19] . Assays challenged with seasonal coronavirus antibody positive sera were not reactive, and no crossreactivity to sera containing parvovirus B-19 antibodies was seen (Table 2) The Wantai Total Ab and sVNT (using a modified population specific cut-off of 25%) demonstrated the highest specificity. This may be because both assays detect Ab to the RBD within the S1-subunit, therefore increasing specificity by exclusion of cross-reacting epitopes outside this domain [19] . sVNT is the only immunoassay correlating with nAb [21, 22] although studies have shown samples with Wantai Total Ab positive results with an index ratio > 10 had detectable levels of nAb by VNT [11] . Assay formats which utilise a single detector antibody, for example Wantai Total Ab, may show greater specificity than those using two antibodies in an antigen-antibody-antibody format, such as the Euroimmun assays [23]. Except for the Wantai Total Ab, manufacturers for all assays reported higher sensitivities than we demonstrated. However, direct comparison of results is limited by differences in available sample cohorts and sampling timeframes. Our results are consistent with prior reports of sensitivity for Wantai Total Ab of 100% for samples collected > 14 days post-onset [11] ; Euroimmun S1-IgG sensitivity post-onset, 43% (7 to 13 days), 67% (14 to 20 days) and 78% (≥ 21 days) and the specificity was 96.0% [19] ; Euroimmun NCP-IgG 88.89% (>14 days) [22] ; and 95.2% (14-17 days), and specificity 94.7% [24] . While other groups have reported slightly higher sVNT sensitivity (95-100%) and specificity 100% [22] than we observed, sensitivity in our assessment was 93.1% (95% CI 84.5-97.7) for sera collected > 14 days post-symptom onset, with specificity 97.1% (pre-pandemic population group). This improved to 100% with use of a modified 25% inhibition cut-off, and 96.7% for the cross-reactive group. We compared agreement between the ELISAs and an in-house MN assay, which is the current gold standard to assess protective immunity against SARS-CoV-2 [11] . The Wantai IgM assay had the highest agreement with MN at 91%, followed by Wantai Total Ab at 86% and the lowest was with the Euroimmun S1-IgG at 79%. The sVNT performed according to the IFU gave the best concordance with MN at 84% compared to 83% (when an equivocal repeat zone was used) and 81% (when a 25% cut-off was used). At > 14 days post symptom onset Wantai Total Ab had the best agreement with MN at 98% followed by Euroimmun S1-IgA, Euro NCP-IgG and sVNT (as per IFU) with (97%, 97% and M a n u s c r i p t 95% respectively) and Wantai IgM having the poorest agreement at 93%. MN and ELISA result concordance for sera from the pre-pandemic population group ranged from 55-100% and from 0-80% for the cross-reactive assessment group; MN was the only assay not to show reactivity to the SARS-CoV-1 antibody positive sample. Interestingly, some of the Ab binding assays showed better agreement with the MN than the sVNT. The sVNT was not as sensitive as MN in our comparison. However, it is simpler to perform than a VNT and does not require a level 3 Biocontainment laboratory, a highly skilled operator nor 5-day test turnaround time, and it has high specificity (if a population adjusted cut-off is employed), and generally good correlation with VNT [25]. Further comparison is required to determine its suitability as an alternative to VNT, in certain settings, particularly as we move into the post-vaccination phase. In our study, as reported by others, IgA and IgM were detected earlier than IgG, and IgA specificity was lower than IgM or IgG [20] . For clinical testing, the Wantai total antibody assay demonstrated the best overall performance with highest sensitivity, specificity, PPV and NPV. Detecting RBD-Ab, the strong linear correlation between S-and RBD-Ab and ACE-2 receptor binding, suggests nAb is being detected [11, 25] . It is also suitable for population screening where high specificity is required and high sensitivity is desirable. Other assays such as Euroimmun NCP-IgG have high sensitivity but lower specificity, giving it a low PPV. Further research on the utility of IgA and IgM detection is required, especially as indicators of recent infection. The high specificity sVNT assay was further improved to 100% with cut-off adjustment to 25% for our population. Although sensitivity of the sVNT was lower than the Wantai Total Ab, its specificity makes it a reliable supplemental test to use with screening assays. The assays evaluated here detect antibodies to different viral targets: S, NCP or RBD and emerging humoral response dynamics studies suggest these antibodies have different half -lives [26], an important consideration for serosurveillance. An additional important consideration for assay selection is understanding whether the assay detects non-nAb or nAb (particularly for testing post-vaccination); however further research is required. The Wantai SARS-CoV-2 total Ab assay was alone among those tested in our hands in achieving sensitivity and specificity very close to that, reported by the manufacturer (94.5%; and 100% respectively); one limitation of this assay is the requirement for 100ul of serum versus 10ul for the other assays. In low prevalence populations and with low test PPV's, serological testing algorithms should use highly sensitive assays for screening such as Wantai Total Ab, followed by supplemental or confirmatory testing with highly specific assays such as sVNT or MN to ensure reliable results. An important component of this study is the comparison of sensitivities of the ELISAs and MN, which should be considered in the context of discordant screening and supplemental results in testing algorithms. M a n u s c r i p t Limitations of our study include the use of two different RT-PCR POS sample panels however there was sample overlap between the panels with all of Panel B included within Panel A. Panel B also included more convalescent patient samples (69%) collected > 14 days post symptom onset compared to Panel A which contained 49%. A rise in index value/inhibition for all assays was observed over time ( Figure 1 ) however these measurements are semi-quantitative, indicating antibody amount present for comparison between assays, not antibody titre which can only be obtained by sample titration to 'end-point.' In conclusion, our study provides a detailed comparison of six commercially available serology assays with performance consistent with what others have reported. The strength of our study is comparing these assays detecting different antibody types with a surrogate virus neutralization test and a gold standard virus neutralization assay. As we move into the post-vaccination phase consideration should be given to the use of GenScript sVNT in certain settings, as an alternative to VNT. ELISA detection of anti-SARS_CoV-2 specific antibody provides evidence of prior SARS-CoV-2 infection or vaccination, however further research is required to determine correlates of protection, antibody longevity and the role played by the adaptive immune system. M a n u s c r i p t M a n u s c r i p t Lin M a n u s c r i p t M a n u s c r i p t M a n u s c r i p t M a n u s c r i p t False positives in reverse. COVID-19 situation update worldwide, as of 27 Longitudinal evaluation and decline of antibody responses in SARS-CoV-2 infection Molecular and serological investigation of 2019-nCoV infected patients: implication of multiple shedding routes SARS-CoV-2 use cases white paper, Mickey Urdea, Halters Associates. Rockefeller Foundation Connecting cluster of COVID-19: an epidemiological and serological investigation Clinical and epidemiological features of 36 children with coronavirus disease 2019 (COVID-19) in Zhejiang, China: an observational cohort study Oxford University Hospitals NHS Foundation Trust. Evaluation of sensitivity and specificity of four commercially available SARS-CoV-2 antibody immunoassays Use cases for SARS-CoV-2 assays 2020 The COVID-19 serology studies workshop: recommendations and challenges 2019-nCoV) by real-time RT-PCR. Euro surveillance: bulletin Europeen sur les maladies transmissibles = European communicable disease bulletin Evaluation of Serological Tests for SARS-CoV-2: Implications for Serology Testing in a Low-Prevalence Setting Prophylaxis with a Middle East respiratory syndrome coronavirus (MERS-CoV)-specific human monoclonal antibody pretects rabbits from MERS-CoV infection Prior infection and passive transfer of neutralizing R: A language and environment for statistical computing. R Foundation for Statistical Computing PropCIs: Various Confidence Interval Methods for Proportions Mucosal versus systemic antibody responses to SARS-CoV-2 antigens in COVID-19 patients For all calculations, equivocal results were treated as positive. 95% CI, 95% confidence interval; Best-worst; best and worstcase intervals -intervals obtained by setting the corresponding sensitivity and specificity estimates to their upper/lower interval bound) Euro Averaged; average sensitivity across the symptom onset categories.PPV; Positive Predictive Value. NPV; Negative Predictive Value, c/o; cut-off, ^ *20% c/o; s-vnt ≥ 20% cut-off, no repeats+ s-vnt ≥ 20% cut-off + repeats for equivocals (18%-22%)+, # *25% c/o; ≥ 25% cut-off Group A -49% of sera (> 14 days post symptom onset) Group B -69% of sera (>14 days post symptom onset) Abbreviations: Euro GenScript surrogate virus neutralization test, CI; Confidence Interval, pos; positive, c/o; cut-off, equ; equivocal Financial support. This work was supported by the Victorian Department of Health and Human Services and by the Australian Government Department of Health. A c c e p t e d M a n u s c r i p t A c c e p t e d M a n u s c r i p t A c c e p t e d M a n u s c r i p t A c c e p t e d M a n u s c r i p t Figure 1 -Boxplots of data distribution, as signal/cut-off value for each assay for the RT-PCR Positive and control sera by days post symptom onset; <7 days, 7-14 days and >14 days. Boxes represent median values and interquartile range, and whiskers represent largest and smallest values. Grey horizontal lines represent the cut-off value, and the shaded grey indicates the equivocal/borderline zone. Measure is index value (sample optical density (OD) value/ cut-off OD) for all assays except sVNT. sVNT measure is %inhibition, calculated as per IFU. Comparative analysis of assay specificity in cross-reactive assessment and population groups Abbreviations: E-S1-IgA, Euroimmun S1-IgA; E-S1-IgG, Euroimmun S1-IgG; E-NCP-IgG, Euroimmun NCP-IgG; sVNT, GenScript surrogate virus neutralization test 20% cut-off with repeat testing for equivocal results (18) (19) (20) (21) (22) ; sVNT-20 with 20% cut-off with no repeat testing; sVNT-25 with 25% inhibition cut-off and no repeat testing; W-IgM, Wantai IgM, W-T, Wantai Total Ab.A c c e p t e d M a n u s c r i p t A c c e p t e d M a n u s c r i p t