key: cord-0926134-nm7oo4hy authors: Dunn, Emily F.; Akhtar, Adil; Dunn, Adam; Lacey, Scott; Pauley, Ethan; Powers, Colton; McKee, James; Petereit, Daniel title: Evaluating a UVC System for Use During SARS-CoV2 Pandemic and Personal Protective Equipment Shortage date: 2020-12-01 journal: Adv Radiat Oncol DOI: 10.1016/j.adro.2020.100636 sha: 5ef9a4e1f45c8881a72db156cda9633e12e655dc doc_id: 926134 cord_uid: nm7oo4hy BACKGROUND: The supply of N95 mask or filtering facepiece respirator (FFR) has been limited nationally due to the Covid-19 pandemic. UVC light has been suggested as a potential option for decontamination of FFRs by the Centers for Disease Control. There has been a lack of publications characterizing UVC dose distribution across FFRs. METHODS: A UVC light box and FFR rack system was assembled utilizing low pressure mercury lamps peaked at 254 nm and aluminum flashing to reduce shadowing effect. Dose was characterized with the use of UV intensity labels and a UVGI (ultraviolet germicidal irradiation) NIST (National Institute of Standards and Technology) traceable meter. Ozone production was evaluated after extended bulb run time. RESULTS: Calibration of UV intensity labels were noted to have color-change saturation at 100 mJ/cm (2). Dose measurements with the UV intensity labels on the FFR demonstrated symmetric dose to all surfaces but symmetry not supported by measurements with the UVGI meter. There was substantial dose fall off on the lateral aspects of the FFR. No ozone production was noted in the UVC system. CONCLUSIONS: UV intensity labels for characterization of dose provided a false suggestion of symmetry when compared to UVGI meter. Estimates of appropriate exposure times to reach 1000 mJ/cm(2) should be significantly increased to account for geometry of FFR and lateral dose fall off. The SARS-CoV2 (Severe Acute Respiratory Syndrome 2) pandemic has caused a patient J o u r n a l P r e -p r o o f surge in many hospitals and clinics, resulting in a nationwide shortage of personal protective equipment. One of the valuable pieces of personal protective equipment (PPE) for healthcare personnel is the N95 mask or filtering face piece respirator (FFR). The Centers for Disease Control and Prevention (CDC) has recommended several strategies to avoid crisis capacity of FFRs 1 . If decontamination is deemed necessary, ultraviolet germicidal irradiation, vaporous hydrogen peroxide (VHP) and moist heat were suggested when new respirators are not available 1 . Of these methods, VHP has been granted emergency use authorization by the Federal Drug Administration (FDA) for N95 respirator decontamination 2 . This provides a great reprieve for institutions that can utilize these services but does not yet reach the critical need in many clinics and small communities nationwide. The University of Nebraska Medical Center has been utilizing UVC light for decontamination of FFRs for several months and their protocol has been widely shared 3 . Unless identical equipment and room geometry can be utilized, it is challenging to replicate this system for use in other clinics. UVC has great potential use during the Covid-19 pandemic, but there remains a need to describe the safe implementation of any device. UVC light has a wavelength between 100-280 nm and is found in artificial sources, such as low pressure mercury lamps, which have a spectral peak at 254 nm. UVC radiation causes DNA/RNA damage by producing cyclobutane pyrimidine dimers altering DNA/RNA structure, and thus interferes with replication 4 . Some low pressure mercury J o u r n a l P r e -p r o o f lamps can generate a weak spectral line at 185 nm, which produces ozone in combination with oxygen from ambient air 5 . SARS-CoV2 has been demonstrated to be an enveloped, positive-stranded RNA virus with nucleocapsid from the family of coronaviruses 6 Multiple prior studies have evaluated the durability of FFRs with UVC exposure. UVC decontamination has been compared to microwave-generated steam, moist heat, bleach solution, ethylene oxide, and vaporized hydrogen peroxide decontamination of FFRs 9-12 . Microwave generated heat and steam have not had agreement on their safety since some studies have noted degradation of the respirator after use though others contradict this finding [11] [12] [13] . Ethylene oxide has not been recommended, as long drying times are necessary to ensure complete removal of the harmful gas 12 . Vaporized hydrogen peroxide requires gas chambers and ventilation hoods for use, though has been shown to have acceptable fit and filtration of the FFR following decontamination cycles 8,13 . J o u r n a l P r e -p r o o f Alcohol can alter the electrostatic charge or degrade the filtration of an FFR 8, 14 . Lastly, any chemical process has the potential to cause sensitivity issues on the skin of healthcare provider when reused. The integrity of the FFR when considering aerosol penetration, airflow resistance, or physical appearance of several models of FFRs has been minimal after UVC exposure 10, 12 . Material degradation has been demonstrated after extremely high doses of UVC (710 and 950 J/cm 2 ) to FFRs 10 . The breaking strength of the respirator straps also demonstrated a dose dependent decline with a 14-28% decrease in strength at 1180 J/cm 2 and 20-51% decrease at 2360 J/cm 2 10 . A 2018 study examined UVC deactivation of H1N1 influenza on fifteen NIOSHapproved N95 FFRs of varying shapes and materials 15 . Statistically significant reduction in influenza viability was demonstrated for all FFRs exposed to UVC. Specific design characteristics such as geometry that increase shadowing (folds) and hydrophilic material were found to make sterilization less amenable to UVC. Even FFRs with these characteristics demonstrated statistically significant reductions in influenza viability following UVC exposure. Additional research from 2019 supports FFR integrity and fit after several UVC decontamination cycles 16 . This study specifically evaluated 15 types of FFRs to UVC exposure. Data presented in the above study also provided dose and virus susceptibility specific to the family of coronaviruses and UVC on FFR material. MERS-CoV (Middle Eastern Respiratory Syndrome Coronavirus) and SARS-CoV1 were soiled on numerous FFRs and exposed to 1000 mJ/cm 2 . Following exposure to UVC at this dose, no detectable virus was recovered 16 . Notably, this evaluation was completed on flat J o u r n a l P r e -p r o o f coupons and did not take into account utilizing UVC for clinical need, which requires evaluation of geometry and shadowing of FFRs. 1000 mJ/cm 2 should serve as optimal minimum dose to the FFR based upon this data. Total exposure times should be determined by the time required for the lowest dose-receiving location on the FFR to receive at least 1000 mJ/cm 2 The following outlines the way a UVC system was constructed and evaluated for symmetry of exposure across FFRs. These measurements were utilized to establish exposure times ensuring the surfaces of the FFR received at least 1000 mJ/cm 2 . Materials utilized in the production of the UVC system are listed: 12 Ushio G30T8 germicidal low pressure mercury arc lamps, ILT2400 Meter (calibrated on March 26, 2020), XSD140T254 detector (silicon sensor), N010-004 UVC Intensity labels, 3 garment stands, 6 Lithonia 36" light fixtures, 2 extension cords, pack of 10 conduit connectors, 16-guage wire for securing fixtures and hanging FFRs, 1 roll of electrical tape, aluminum reflective insulation, Aeroqual series 200 Monitor and Ozone sensor Head 0-0.5 ppm (calibrated January, 2020), and tools for assembly (drill, screw driver, hammer, pliers, wire cutters). The complete structure consists of three racks, one FFR mask rack sandwiched between two source racks ( To ensure a constant irradiance, the lamps were initially turned on for 20 minutes prior to measurements. Prior to each measurement, the lamps were turned off to reposition the detector. At initiation, and two minutes of exposure, the output irradiance was found to be on average 85% and 95%, respectively, of the maximum output at 5 minutes. Thus, all individual measurements were made at approximately 5 minutes from the turning on of the lamp. A calibration curve of UVC intensity labels was created by exposing labels to a known dose. The label was affixed to the UVGI sensor, just inferior to the detector to avoid shadowing as demonstrated in Figure S1 . The dose was determined by utilizing the ILT2400 & XSD140T254 meter and detector system in integration mode while simultaneously exposing the label (Figure S1 ). A FFR was placed in the upper corner of the mask rack for UVC exposure evaluation (location 3, 131.5 cm from base as noted in Figure 1 ). Nine UVC Intensity Labels were adhered on the convex and concave aspects with clear plastic tape as noted in Figure S2 . The symmetry of UVC dose was tested by exposing the FFR to 20 mJ/cm 2 , which was confirmed by the silicon sensor that was placed next to the FFR. For further evaluation of symmetry of dose across the respirator, the UVGI meter was adhered to the FFR at 5 different locations on the convex and concave surface (Figure 3) and exposed to UVC to evaluate the irradiance. This was repeated at the four corner locations (rack height 68.5 cm and 162 cm in both columns 1 and 3), which are felt to be the most divergent from the light sources. Ozone generation was evaluated with a NIST traceable meter. Both light sources were run for 40 minutes at room temperature to assess if any residual ozone was being created during the run cycle. Table 1 demonstrates the irradiance and output at the 12 different locations along the rack to each light source when placed en face to the bulb rack. A goal dose of 1000 mJ/cm 2 was reached by running the light sources for 1320 seconds in the absence of aluminum flashing. The UVC label calibration curve showed discernable color change up to 100 mJ/cm 2 , with the darkest color recorded at 125 mJ/cm 2 . Color darkening was difficult to discern beyond 100 mJ/cm 2 . The most contrasting color change to the labels occurred between 0 mJ/cm 2 and 25 mJ/cm 2 ( Figure S1) . Therefore, the symmetry of FFR exposure was evaluated with labels and a dose of 20 mJ/cm 2 . Based upon findings noted in Figure S2 , the FFR appeared to have symmetric exposure to both the convex and concave surfaces when exposed to 20 mJ/cm 2 by UV intensity labels. The color change of the labels was noted to be within 20-25 mJ/cm 2 as per comparison to the previous calibration curve. Additional quantitative evaluation of dose via UVGI meter at 5 different locations on the FFR is demonstrated in Table 2 . This confirmed a dose gradient across the FFR noting substantial dose fall off at the lateral aspect of the respirator. Of the four locations measured for mask contouring effect, rack height 68.5 cm, column 3 showed the largest exposure dose gradient with a minimum of 34% and a maximum of 126% compared to the central convex position 3A. The inner mask exposure range was between 66% and 107% compared to the central convex position 3A. The least exposed location on the rack was 68.5 cm from the base, column 3, mask position 4A. There was a large gradient in the measured dose compared to the perceived dose based on UV intensity labels. In order to reach a minimum of 1000 mJ/cm 2 to the lateral aspects of the respirator, the exposure times required an increase to 1950 seconds. Ozone was measured at baseline to be 0.000 parts per million and 0.000 parts per million when read immediately after bulb sources were run for 40 minutes. Minimum dose exposure to the FFR was evaluated by both UVC labels and a UVGI meter. The meter is likely to under represent the total dose that is received by the FFR as angular response of the meter was not determined. The UVC labels may provide a general guideline for light exposure and to avoid shadowing. UVC labels utilized here are useful for UVC exposure, but are also reactive to UVA and UVB spectrum of light. Caution should be used as these can potentially overestimate UVC exposure. Furthermore, overexposure has few downsides in comparison to underexposure when considering FFR decontamination. Thus, the UVGI meter is favored for determining minimum exposure times in any given UVC decontamination system. FFRs with exhalation valves, folds, or coating in hydrophilic material should not be considered for UVC decontamination. Folds and expiratory valves create unavoidable shadowing. FFRs made of hydrophilic material are also felt to be incompatible due to viral particle migration away from the surface of the respirator 15 . Hard-bodied FFRs without the above features are considered most optimal for UVC decontamination. It is important to note that maximum doses should also be evaluated when utilizing any UVC system. We have recommended using a conservative maximum dose of 120 J/cm 2 based on prior studies 10 While a maximum number of decontamination cycles could theoretically be calculated, the user fit will be altered after several wears. User self-seal-check is required to confirm the respirator remains effective for reuse. Any alteration in the seal renders the FFR highly ineffective and should be discarded 17 . It is necessary that the FFR is reused only by the original wearer after UVC exposure and should be labeled for this purpose due to the molding of the FFR to the face. The chain-of-command within any clinic or hospital setting is highly important to avoid any potential cross contamination between FFRs or to the staff processing the respirators. The wearer of FFR should avoid any cosmetics, sunblock, or other materials that could transfer to the mask since this will impede the ability of UVC to reach the surface of the respirator. Any defects in the FFR strap or mask integrity should not be sent for decontamination. The respirator should be thoroughly examined prior to donning to ensure appropriate structure remains in tact. The FFR should not be considered sterile after UVC decontamination since 1000 mJ/cm 2 has not been evaluated for all potential pathogens. A processed FFR with UVC should be donned with the use of gloves and followed by hand hygiene. The use of UVC for FFR decontamination should only be considered in dire situations where new FFRs are not available. Low pressure mercury lamps utilized in our process are ozone-free and peaked at 254 nm. It is necessary to evaluate ozone production within each potential setup of UVC lights as exposure is carcinogenic. UVC lights should be used in a well-ventilated space for extra caution. ALARA (as low as reasonably achievable) should be employed when J o u r n a l P r e -p r o o f using UVC as with any other radiation source. The lights can be carcinogenic and cataractogenic when exposed to bare skin and eyes. Those that are looking to use UVC light should be aware that we were unable to independently assess the decontamination of any pathogens. The above utilized physical dose and knowledge of the efficacy on UVC from other coronavirus strains (SARS-CoV1 and MERS-CoV) 15 . Exposure dose should be verified independently to deliver at least 1000 mJ/cm 2 to all surfaces of the FFR and minimize shadowing. If UVC labels are used, determination of the exposure dose range is recommended through a calibration curve. UVC labels are considered less accurate since the evaluation is based on visual acuity and labels can change in color when exposed to UVA or UVB light., NIST traceable radiometry equipment should be used to accurately measure the exposure dose within the desired UVC spectral range. Substantial dose falloff at the lateral aspects of the FFRs was noted within our system. Therefore, minimum exposure to en face measurements of the light source will not provide adequate dose to FFRs. Exposure fall-off from the light source can be approximated by 1/r 2 if assumed to be a point source, where r is the distance from the source. If the source is light is spread out, as it is with multiple UV lamps, the exposure fall-off will be more gradual than 1/r 2 . Given the geometrical complexity of real world light sources it is essential that exposure be calibrated by measurement rather than calculation. We strongly recommend review of The Ultraviolet Germicidal Irradiation Handbook for further resources on mathematical modeling, safety, and the potential use of UVGI in clinical settings 18 . An example of a system implemented at a second community hospital is shown in Figure S3 where UVC towers were already in use to sterilize rooms for C. Difficile It is imperative that any use of UVC light be completed with a team that has knowledge of its capabilities and limitations and the input of a medical physicist is strongly recommended. UVC light has been described as a possible tool for FFR decontamination by the CDC 1 . Many institutions may have UVC lights already in place for C. Difficile cleansing of rooms or are available at lighting distributors but will require additional engineering to be used during the pandemic. The above describes the creation, geometry, and dose characterization of a UVC system for FFRs. The total time of UVC exposure was substantially longer once dose fall off of the lateral aspect of the mask was taken into account via UVGI meter. The use of UV intensity labels did not provide accurate analysis of dose distribution and may lead to J o u r n a l P r e -p r o o f false conclusions of symmetric exposure. The minimum dose utilized herein was extrapolated from other studies and virucidal effect was not validated. UVC has several limitations in its use due to the susceptibility of shadowing, carcinogenic if exposed to humans, and possible generation of ozone. Accurate characterization of the spectrum of light, and geometry of the UVC system is imperative for success. J o u r n a l P r e -p r o o f Table S1 : Quantitative measurement of exposure dose in mW/cm 2 at 5 locations on convex (A) and concave (B) surface of the FFR as demonstrated in Figure 3 . Measurements obtained via ultraviolet germicidal irradiation meter. Locations selected are the most divergent from light sources and provide the minimum exposure time to achieve at least 1000 mJ/cm 2 to all positions on FFR within the rack. The least exposed position is 68.5 cm from the base, column 3, mask position 4A. COVID-19) N95 Filtering Facepiece Respirator Ultraviolet Germicidal Irradiation (UVGI) Process for Decontamination and Reuse UV inactivation of pathogenic and indicator microorganisms UV-Lamps for Disinfection and Advanced Oxidation -Lamp Types Coronaviridae Study Group of the International Committee on Taxonomy of Viruses. The species Severe acute respiratory syndrome-related coronavirus: classifying 2019-nCoV and naming it SARS-CoV-2 Predicted Inactivation of Viruses of Relevance to Biodefense by Solar Radiation Effectiveness of N95 Respirator Decontamination and Reuse against SARS-CoV-2 Virus Effectiveness of three decontamination treatments against influenza virus applied to filtering facepiece respirators Effects of Ultraviolet Germicidal Irradiation (UVGI) on N95 Respirator Filtration Performance and Structural Integrity Impact of Three Cycles of Decontamination Treatments on Filtering Facepiece Respirator Fit Evaluation of Five Decontamination Methods for Filtering Facepiece Respirators A scalable method of applying heat and humidity for decontamination of N95 respirators during the COVID-19 crisis Electrostatic Respirator Filter Media: Filter Efficiency and Most Penetrating Particle Size Effects Ultraviolet germicidal irradiation of influenza-contaminated N95 filtering facepiece respirators Research to Mitigate a Shortage of Respiratory Protection Devices During Public Health Emergencies Performance of an N95 Filtering Facepiece Particulate Respirator and a Surgical Mask During Human Breathing: Two Pathways for Particle Penetration Ultraviolet Germicidal Irradiation Handbook: UVGI for Air and Surface Disinfection