key: cord-0924967-spc5euhh authors: Pérez-Cataluña, Alba; Cuevas-Ferrando, Enric; Randazzo, Walter; Falcó, Irene; Allende, Ana; Sánchez, Gloria title: Comparing analytical methods to detect SARS-CoV-2 in wastewater date: 2020-12-08 journal: Sci Total Environ DOI: 10.1016/j.scitotenv.2020.143870 sha: 745549838df81b6da9e65937c9d9f6a423432cc6 doc_id: 924967 cord_uid: spc5euhh Wastewater based epidemiology (WBE) has emerged as a reliable strategy to assess the coronavirus disease 2019 (COVID-19) pandemic. Recent publications suggest that SARS-CoV-2 detection in wastrewater is technically feasible; however, many different protocols are available and most of the methods applied have not been properly validated. To this end, different procedures to concentrate and extract inactivated SARS-CoV-2 and surrogates were initially evaluated. Urban wastewater seeded with gamma-irradiated SARS-CoV-2, porcine epidemic diarrhea virus (PEDV), and mengovirus (MgV) was used to test the concentration efficiency of an aluminum hydroxide adsorption-precipitation method and a polyethylene glycol (PEG) precipitation protocol. Moreover, two different RNA extraction methods were compared in this study: a commercial manual spin columns centrifugation kit and an automated protocol based on magnetic silica beads. Overall, the evaluated concentration methods did not impact the recovery of gamma-irradiated SARS-CoV-2 nor MgV, while extraction methods showed significant differences for PEDV. Mean recovery rates of 42.9 ± 9.5%, 27.5 ± 14.3% and 9.0 ± 2.2% were obtained for gamma-irradiated SARS-CoV-2, PEDV and MgV, respectively. Limits of detection (LoD95%) for five genomic SARS-CoV-2 targets (N1, N2, gene E, IP2 and IP4) ranged from 1.56 log genome equivalents (ge)/mL (N1) to 2.22 log ge/mL (IP4) when automated system was used; while values ranging between 2.08 (N1) and 2.34 (E) log ge/mL were observed when using column-based extraction method. Different targets were also evaluated in naturally contaminated wastewater samples with 91.2%, 85.3%, 70.6%, 79.4% and 73.5% positivity, for N1, N2, E, IP2 and IP4, respectively. Our benchmarked comparison study suggests that the aluminum precipitation method coupled with the automated nucleic extraction represents a method of acceptable sensitivity to provide readily results of interest for SARS-CoV-2 WBE surveillance. The use of wastewater as a tool for epidemiology tracking, known as wastewater-based epidemiology (WBE), has a long history of use in public health, particularly for human enteric viruses (Asghar et al., 2014; Cuevas-Ferrando et al., 2020; Hellmér et al., 2014; Miura et al., 2016; Prevost et al., 2015; Santiso-Bellón et al., 2020) . In the midst of the ongoing COVID-19 pandemic, WBE is being implemented globally for the detection of SARS-CoV-2 RNA shed into wastewater, sewers, and sludge (Ahmed et al., 2020a; Bivins et al., 2020; Guerrero-Latorre et al., 2020; Haramoto et al., 2020; (Ahmed et al., 2020b; Polo et al., 2020; Rusiñol et al., 2020; WHO, 2020) . In an attempt to coordinate current knowledge and data gaps, the European Commission (EC) created a Pan-European Umbrella Study to better understand the limitations and challenges of this approach including the development of a roadmap for a systemic rollout of complementing ongoing national and regional surveillances in a unique approach (EC, 2020) . One of the problems highlighted by these collaborative studies is the need of standardized procedures, spanning from sampling to data analysis. In this sense, viral concentration and nucleic acid extraction methods are two critical steps for the analysis of viruses in wastewater and quality controls must be accurately defined. To our knowledge, three studies have compared different concentration methods using SARS-CoV-2 surrogates (Ahmed et al., 2020c; Jafferali et al., 2020; Rusiñol et al., 2020) . However, the analytical performances of SARS-CoV-2 concentration, extraction, and detection procedures tested alongside are not yet characterized for wastewater samples. Thus, the aim of this work was to evaluate different concentration methods, nucleic acid extraction procedures, and quantitative RT-qPCR assays to efficiently detect SARS-CoV-2 in wastewater using gamma-irradiated SARS-CoV-2, porcine epidemic diarrhea virus (PEDV) as coronavirus model, and mengovirus as non-enveloped counterpart. Importantly, limits of detection were established using gamma-irradiated SARS-CoV-2 particles and five different RT-qPCR assays targeting various genetic fragments. We finally validate the selected RT-qPCR assays in wastewater samples collected during the COVID-19 pandemic in different regions of Spain. enveloped member of the Picornaviridae designated in the ISO 15216-1: 2017 standard method as process control. The PEDV cytopathogenic CV777 strain (Friedrich-Loeffler-Insitut, Greifswald, Germany) and MgV vMC0 were propagated in Vero and HeLa cell monolayers, respectively (Puente et al., 2020) . Two hundred milliliters of seeded wastewater samples (n=4) were concentrated through AlCl 3 adsorption-precipitation (Randazzo et al., 2019 (Randazzo et al., , 2020a (Randazzo et al., , 2020b . A final concentrate was then formed by centrifugation at 1,900 × g for 30 min and the resulting pellet was resuspended in 1 mL of PBS, pH 7.4. Alternatively, 200 mL of seeded wastewater samples (n=4) were concentrated through precipitation with 20% polyethylene glycol (PEG) 8000 (PanReac, Spain) and resuspended in 1 mL of PBS, pH 7.4. Briefly, 25 mL of Tris Glycine-Beef Extract buffer (TGEB) pH 9.5 were added to each sample and incubated in agitation at 300 rpm for 2 hours at 4°C. After incubation, samples were centrifuged at 2,500 x g for 10 minutes at 4°C. Supernatant was adjusted to pH 7.0-7.2. PEG and NaCl were added to a final concentration of 20% and 0.3M, respectively, and mixed gently. Sample was incubated in agitation overnight at 4°C and then centrifuged at 3,500 x g for 30 min at 4°C. Pellet was resuspended in 1 mL of PBS and concentrated samples stored at -80°C for further analysis. For both concentration methods, recovery controls were prepared by spiking each virus at the concentration detailed above in 1 mL of PBS. For each sample, the percentage recovery was calculated dividing the viral titer of concentrated sample by the titer of the recovery control. original provider's protocol were established based on preliminary laboratory results during a method optimization step (data not shown). Finally, 400 µL of cetyltrimethyl ammonium bromide (CTAB) and 40 µL of proteinase K solution (both provided with the kit) were added to 300 µL of concentrated water samples, the mix was then incubated at 60°C for 10 min and centrifuged for 10 min at 16000 x g. Then, the supernatant was transferred to the loading cartridge along with 300 µL of lysis buffer. The cartridge was loaded in the Maxwell® RSC Instrument and the extraction performed by selecting the "Maxwell RSC Viral total Nucleic Acid" running program in the instrument software. For both manual and automated extractions, RNA was finally eluted in 100 µL nuclease-free water. Negative controls constituted by nuclease-free water instead of concentrated sample were included in both extraction methods. Viral detection of SARS-CoV-2, PEDV, and MgV was performed by RT-qPCR using One Step Tables S1, S2 and S3, respectively. All RT-qPCR assays were performed in duplicate on a LightCycler 480 instrument (Roche Diagnostics, Germany). Positive (genomic RNA) and negative (nuclease-free water) controls were always included. Standard curves for PEDV, MgV and SARS-CoV-2 quantifications were performed using the genomic RNA of each virus with serial 10-fold dilutions in triplicate. Differences between methods were statistically analyzed using Saphiro test for normal distribution and T-student for mean comparison (p < 0.05). Influence of concentration and extraction methods was analyzed using multifactorial ANOVA for each virus (p < 0.05). The limits of detection at 95% and 50% confidence intervals (LoD 95% and LoD 50% , respectively) were obtained by detecting gamma-irradiated SARS-CoV-2 (Bei Resources, NR-52287) ten-fold serially J o u r n a l P r e -p r o o f Journal Pre-proof diluted from 1.7 x 10 3 to 1.7 ge/mL and seeded in 200 mL of wastewater samples tested negative for SARS-CoV-2. Samples were also spiked with PEDV (10 7 gc/mL) as process control. Viral particles were concentrated by the aluminum hydroxide adsorption-precipitation method and RNA extracted using both RNA extraction protocols as described above. Experiments were performed in triplicate by concentrating three independent samples for each inoculation level. LoD 95% and LoD 50% were calculated according to Wilrich & Wilrich (2009) . To determine SARS-CoV-2 detection limits, five different targets were used: N1 and N2 regions of the nucleocapsid gene, the envelope gene (E), and regions IP2 and IP4 of the RNA-dependent RNA polymerase gene (RdRp). The amplification of the N1 region was conducted as previously described. Region N2 detection was fulfilled using primers and probes available at the diagnostic panel assays 2019-nCoV RUO Kit from the US CDC (CDC, 2020). Detection of gene E was performed using primers and probes described by Corman et al. (2020) (Table S1 and S2). To amplify and quantify IP2 target, One Step PrimeScript™ III RT-PCR kit (Takara Bio) was used. A total of 34 influent wastewater samples collected in different regions of Spain were analyzed for the detection and quantification of SARS-CoV-2. Samples were concentrated using the aluminum hydroxide adsorption-precipitation method as described before. Detection of SARS-CoV-2 was conducted through the analysis of five aforementioned SARS-CoV-2 genome targets (Tables S1, S2, and S3). Each reaction was performed in duplicate. Genomic RNA of SARS-CoV-2 (ATCC VR-1986D) and nuclease free water were used as positive and negative controls, respectively. Viral quantifications were calculated by using two different standard curves for N1, N2 and E genes. The Figure 1 shows the viral recovery rates of eight wastewater samples tested negative for SARS-CoV-2 spiked with gamma-irradiated SARS-CoV-2, PEDV and MgV, subjected to two different concentration and two nucleic acid extraction methods (Figure 1 ). To determine the presence of potential inhibitors, a 10-fold dilution of each sample was also analyzed by RT-qPCR for the three targeted viruses. RT-qPCR results showed that no significant inhibitions occurred (data not shown). into account that these protocols are intended to be used for early SARS-CoV-2 monitoring, in which the readily availability of results is crucial to set a timely public health response up, the choice of a suitable analytical method should be based on the bench work time needed for each procedure, alongside its sensitivity. Since the PEG protocol includes an overnight incubation step, unlike the aluminum hydroxide adsorption-precipitation method (total time less than 2 hours), we select the aluminum protocol for further comparisons. Validation of the two extractions methods was performed in wastewater samples naturally contaminated with SARS-CoV-2 using N1 target. As occurred with spiked samples, slightly better results were obtained when automated system was used (Supplementary Table S4 ). Regarding the (2020), applying the aluminum concentration method combined with the MN extraction, reported LoD 95% of 2.46 log gc/mL for HEV in wastewater. Naturally contaminated wastewater samples were analyzed by targeting five different SARS-CoV-2 genomic fragments (N1, N2, E, IP2 and IP4) to evaluate the sensitivity of each RT-qPCR assay. As shown in Figure For viral quantification of N1, N2 and gene E, standard curves were built using complete genomic RNA of SARS-CoV-2 (ATCC VR-1986D) and synthetic plasmids developed for genes N and E (Supp. Figure S1 ). Figure 3 shows the difference in the logarithm of gc/L obtained with each standard curve for each gene. Mean values of these differences were 1.27±0.01, 0.27±0.03, and 1.61±0.01 log (gc/L) for N1, N2, and E, respectively. This overestimation was produced by the plasmid quantification as described previously (Lin et al., 2011) . Quantification bias was observed depending on the reference material used, which is very important when comparing quantification data from different studies. Thus, the use of genomic RNA of SARS-CoV-2 as standard material should be recommended. Moreover, given the final WBE aim, that is the estimation of the number of infected The introduction of SARS-CoV-2 and its spread to the pandemic status have put all countries on alert and WBE has been readily implemented as an early warning tool for outbreaks. In most of the countries, wastewater surveillance for monitoring COVID-19 started very hastily, even before the scientific community could have robust data on the optimal methodologies. In fact, procedures used for viral detection have been little studied and standardization is still needed. This study benchmarked two concentration methods and two nucleic acid extraction methods widely used in environmental virology. Results obtained in this study reveal the variability obtained depending on the surrogate used as process control to validate the analyses, the extraction method, and the molecular target used for SARS-CoV-2 detection. These are critical decision which will affect the sensitivity of the analyses. On the other hand, despite the difference on sample processing time, both aluminum and PEG concentration methods can be indiscriminately used, as they did not show significant differences. However, the reduced time needed for the concentration of the samples using the aluminum hydroxide adsorption-precipitation method, makes it the preferred method for this step. In contrast, a different sensitivity of the RT-qPCR assay has been observed suggesting that the selection of the molecular target for detection is crucial. The findings of this study expand the knowledge on the analytical procedures and its efficiencies for SARS-CoV-2 detection in wastewater constituting a step forward for the global implementation of COVID-19 WBE. A recurrent mutation at position 26,340 of SARS-CoV-2 is associated with failure of the E-gene qRT-PCR utilized in a commercial dual-target diagnostic assay Environmental surveillance for polioviruses in the global polio eradication initiative Wastewater-Based Epidemiology: Global Collaborative to Maximize Contributions in the Fight against COVID-19 Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR HEV Occurrence in Waste and Drinking Water Treatment Plants Feasibility assessment for an EU-wide Wastewater Monitoring System for SARS-CoV-2 Surveillance COVID-19 surveillance in Southeastern Virginia using wastewater-based epidemiology SARS-CoV-2 in river water: Implications in low sanitation countries First environmental surveillance for the presence of SARS-CoV-2 RNA in wastewater and river water in Japan ISO 15216-1:2017-Microbiology of the Food Chain-Horizontal Method for Determination of Hepatitis A Virus and Norovirus Using Real-Time RT-PCR-Part 1: Method for Quantification Benchmarking virus concentration methods for quantification of SARS-CoV-2 in raw wastewater First proof of the capability of wastewater surveillance for COVID-19 in India through detection of genetic material of SARS-CoV-2 First detection of SARS-CoV-2 in untreated wastewaters in Italy Quantification Bias Caused by Plasmid DNA Conformation in Quantitative Real-Time PCR Assay SARS-CoV-2 in wastewater: potential health risk, but also data source Interlaboratory Comparative Study to Detect Potentially Infectious Human Enteric Viruses in Influent and Effluent Waters SARS-CoV-2 RNA in wastewater anticipated COVID-19 occurrence in a low prevalence area Metropolitan wastewater analysis for COVID-19 epidemiological surveillance Presence and infectivity of SARS-CoV-2 virus in wastewaters and rivers Concentration methods for the quantification of coronavirus and other potentially pandemic enveloped virus from wastewater First detection of SARS-CoV-2 RNA in wastewater in North America: A study in Louisiana SARS-CoV-2 RNA detection of hospital isolation wards hygiene monitoring during the Coronavirus Disease 2019 outbreak in a Chinese hospital Status of environmental surveillance for SARS-CoV-2 virus Estimation of the pod function and the LOD of a qualitative microbiological measurement method SARS-CoV-2 Titers in Wastewater Are Higher than Expected from Clinically Confirmed Cases Comparison and evaluation of conventional RT-PCR, SYBR green I and TaqMan real-time RT-PCR assays for the detection of porcine epidemic diarrhea virus The authors thank Agustin Garrido Fernández, Azahara Reolid, and Andrea Lopez de Mota for their technical support.We acknowledge Global Omnium S.L., NILSA, Aguas de Malaga, ESAMUR and FACSA for managing water sampling.