key: cord-0924680-ro10b9rk authors: Lin, Hsin-Yung; Lai, Pei-Chen; Chen, Wei-Lin title: A narrative review of hydrogen-oxygen mixture for medical purpose and the inhaler thereof date: 2020-09-27 journal: Med Gas Res DOI: 10.4103/2045-9912.295226 sha: 4523ca178d574dd09de203115f1645363f862624 doc_id: 924680 cord_uid: ro10b9rk Recent development regarding mixture of H(2) (concentration of ~66%) with O(2) (concentration of ~34%) for medical purpose, such as treatment of coronavirus disease-19 (COVID-19) patients, is introduced. Furthermore, the design principles of a hydrogen inhaler which generates mixture of hydrogen (~66%) with oxygen (~34%) for medical purpose are proposed. With the installation of the liquid blocking module and flame arresters, the air pathway of the hydrogen inhaler is divided by multiple isolation zones to prevent any unexpected explosion propagating from one zone to the other. An integrated filtering/cycling module is utilized to purify the impurity, and cool down the temperature of the electrolytic module to reduce the risk of the explosion. Moreover, a nebulizer is provided to selectively atomize the water into vapor which is then mixed with the filtered hydrogen-oxygen mix gas, such that the static electricity of a substance hardly occurs to reduce the risk of the explosion. Furthermore, hydrogen concentration detector is installed to reduce the risk of hydrogen leakage. Result shows that the hydrogen inhaler implementing the aforesaid design rules could effectively inhibit the explosion, even ignition at the outset of the hydrogen inhaler which outputs hydrogen-oxygen gas (approximately 66% hydrogen: 34% oxygen). The currently available hydrogen inhalers can be classified as ion exchange membrane type or electrolytic solution type. Figure 1 shows an ion exchange membrane-based hydrogen inhaler which conventionally uses a platinum-coated ion exchange membrane to separate the electrolysis tank into an anode chamber and a cathode chamber. When the anode and cathode electrodes of the electrolysis tank are supplied with electrical power, oxygen is generated in the anode chamber, and the extra protons (H + ) are transported from the anode chamber to the cathode chamber through the ion exchange membrane and combine with electrons on the cathode to form hydrogen therein. Since the ion exchange membrane is structured to separate the electrolytic tank, theoretically almost 100% hydrogen gas could be obtained from the cathode chamber. However, 100% hydrogen gas is not suitable for direct inhalation of human body and shall be mixed with air. 57 It is considered as oxygen-deficient gas when the O 2 concentration of the inhaled gas is slower than 19.5%, and it is dangerous for human begins when the O 2 concentration of the inhaled gas is slower than 16%. 58 Mixing 100% hydrogen gas from the ion exchange membrane-based hydrogen inhaler with air for inhaling will inevitably dilute the original O 2 concentration (~21%) in the air and likely poses negative impact to subject users. For example, a human body consumes approximately air of 6 L/min. Supposedly the ion exchange membrane-based hydrogen inhaler outputs 100% hydrogen gas of 1 L/min and only half of them (500 mL) is inhaled and mixed with the 5.5 L of air for human body, the oxygen concentration of the mixed gas (air of 5.5 L + 100% hydrogen gas of 500 mL) is approximately reduced to 19.25%. Therefore, it is not recommended to inhale 100% hydrogen gas for a long period of time. Moreover, deionized water is required for ion exchange membrane-based hydrogen inhaler to maintain the normal operation; otherwise impurities in regular water would block ion channels of the ion exchange membrane and then the function thereof fails. Additionally, the lifetime of such hydrogen inhaler is shorter due to the inevitable decay of the platinum catalyst coated on the ion exchange membrane. Furthermore, ion exchange membrane is a polymer membrane (such as perfluorosulfonic acid proton exchange membrane) mainly used in fuel cells and may not be suitable for medical purpose. 59 To be worse, fluoride could be likely released from the perfluorosulfonic acid proton exchange membrane during operation. 60 The traditional electrolytic solution-based hydrogen inhaler is illustrated as follows. As shown in Figure 2 , two electrodes of the electrolytic tank undergo an electrochemical reaction against water to produce oxygen and hydrogen which are then mixed within the electrolytic tank. The H 2 concentration in the hydrogen-oxygen mixed gas is approximately equal to 66%. Unlike the ion exchange membrane-based hydrogen inhaler, deionized water is not necessary for the operation of electrolytic solution-based hydrogen inhaler, but it is recommended to use regular pure water or distilled water. Although the electrolytic solution-based hydrogen inhaler is a simple structure, electrolyte is required in the electrolytic solution to enhance electrolysis efficiency, and alkali mist generated during high temperature may irritate the human body if such alkali mist is not properly filtered and isolated. Moreover, heat released during the electrolysis process unavoidably raises the temperature of the electrolytic tank, and it is necessary to intermittently shut down the traditional electrolytic solutionbased hydrogen inhaler for safety purpose. Furthermore, there exists risk of detonation due to the hydrogen (~66%)-oxygen (~34%) mixed gas in the event no suitable anti-explosion devices are designed in the solution-based hydrogen inhaler. Medical-grade hydrogen inhaler is the prerequisite for the development of hydrogen biomedicine. It should be safe, durable, and can stably generate higher flow rate (e.g., over 2 L/min) and higher hydrogen concentration (e.g., over 50%) over a long period (e.g., at least over 8 hours without any interrupt of operation). The lifetime of the hydrogen inhaler shall be more than 5000 operating hours, and no additional electrolyte shall be supplemented during the above lifetime of the solution-based hydrogen inhaler. Furthermore, all materials of the hydrogen inhaler used in contact with gas/liquid should be medical-grade and the hydrogen inhaler itself should be made in compliance with all international standards of medical device. Additionally, all impurity generated during the electrolysis process shall be filtered and isolated, ensuring that the outputted hydrogenoxygen mixture or hydrogen meets the specifications of medical respiratory gas. The examples of available standards are shown in Table 1 . The following are the proposed design rules of a safe and Figure 3 , the electrolytic module is the kernel for the hydrogen inhaler, it will either generates almost 100% hydrogen gas (ion exchange membrane type) or hydrogenoxygen mixed gas (H 2 ~66%: O 2 ~34%, electrolytic solution type) which is then combined with air for the breath of human body. It is important to reduce the space of the gas chamber in the electrolytic module, such that only few residuals of hydrogen gas or hydrogen-oxygen mixed gas would be kept in the electrolytic module to reduce the risk or impact of the unexpected explosion. Furthermore, filtering module shall be installed to purify the hydrogen gas or hydrogen/oxygen mixed gas in order to comply with available medical gas standards. Additionally, double or triple purification processes are recommended to filter various potential impurities, such as fluoride or electrolyte. During electrolytic process, heat will be generated and eventually raises the temperature of the electrolytic module. Such continuously raised temperature could shorten the lifetime of the electrolytic module, breaks sealing of the electrolytic module and then causes the leakage of the liquid/gas therein, and likely triggers the explosion. Therefore, a cycling module to circulate and cool down the liquid in the electrolytic module is critical to maintain the lifetime and safety, as shown in Figure 3 . A temperature sensor coupled to the electrolytic module can monitor the temperature of the electrolytic module and activates the cycling module whenever necessary. Moreover, when the cycling module is utilized in the electrolytic solution type hydrogen inhaler, it could also circulate water and flush residual electrolyte in the air pathway of the hydrogen inhaler back to the electrolytic module. Flushing the residual electrolyte back to the electrolytic module could recycle the electrolyte and extend the operation hours of the hydrogen inhaler. A liquid blocking module in the hydrogen inhaler is an anti-explosion device. It will isolate the electrolytic module from the output of the liquid blocking module; therefore, to block the subject user from any impact of unexpected explosion happened in the electrolytic module. On the other hand, the liquid blocking module prohibits any accidental explosion propagating from the output of the liquid blocking module into the electrolytic module, and therefore to prevent any damage to the electrolytic module caused by any ignition happening outside the liquid blocking module. Thus, a first isolation zone is formed between the electrolytic module and the liquid blocking module, and any unexpected explosion will be restrained therein (Figure 3) . Besides, the liquid blocking module could lower the temperature of the hydrogen gas or hydrogen-oxygen mixed gas generated from the hot zone of the electrolytic module, and further dissolve potential impurity of the generated gas in the liquid as well. In the conventional hydrogen inhaler, a carbon dioxide absorbent is common used to partially remove the vapor in the hydrogen gas or hydrogen-oxygen mix gas. As a result, the mucus layer of the respiratory organ could not be kept in a normal state, and the respiratory airway could not be protected by the mucus layer. Therefore, it is necessary to provide appropriate humidity in the hydrogen gas or hydrogen-oxygen mixed gas for the breath of subject users. The nebulizer is provided to selectively atomize the water into vapor at the nebulizing chamber and the vapor will be mixed with the hydrogen gas or hydrogen-oxygen mixed gas in the mixing chamber, as shown in Figure 3 . Furthermore, the nebulizer could atomize suitable medicine for breath as well to assist treatment of human body. Moreover, when the humidity increases, the electrical resistance of a substance (gas or the subject user) decreases and static electricity hardly occurs. The disappearing time during which static electricity is eliminated also decreases exponentially as the absolute humidity increases. 61 Therefore, the higher the relative humidity, the less likely static electricity causes explosion. With the help of vapor combined with the hydrogen gas or hydrogen/oxygen mix gas, the static electricity Biological evaluation of medical devices-Part 10: Tests for irritation and skin sensitization www.medgasres.com of a substance (gas or the subject user) hardly occurs, and the explosion incurred by the static electricity could be prevented. It is known that flame will be extinguished if the dimension of the air pathway is 0.20 mm or less at any hydrogen gas concentration. 62 Based on the 0.20 mm air pathway criteria, a flame arrester made of stainless steel meshes which are randomly compressed and overlapped together can set up pathway the dimension of which is less than 0.2 mm. Therefore, the flame arrester is also an anti-explosion device to extinguish flame and to prevent explosion from propagating. The flame arrester could be installed at the input of the nebulizer such that a second isolation zone is formed between the nebulizer and the liquid blocking module, and any unexpected explosion in the second isolation zone will be confined therein. For the convenience of breath of the subject user, a comfortable nasal tube is necessary. An additional flame arrester could be installed in the nasal tube such that a third isolation zone is formed between the nasal tube and the nebulizer, and again any unexpected explosion in the third isolation zone will be confined therein. Moreover, such flame arrester is preferred to be installed closed to the output of the nasal tube to reduce the space between such flame arrester and the outlet of the nasal tube (e.g., within the range of 1-2 cm 3 ). In the event any static electricity generated from the subject user or intentional ignition happens at the outlet of the nasal tube, the small space of 1-2 cm 3 hardly causes explosion, and therefore prevent damage to the subject user. It is well discussed that the H 2 concentration of the mixed gas higher than 4% may cause explosion. Nevertheless, even the hydrogen inhaler outputs the hydrogen-oxygen mixed gas and the H 2 concentration of which is around 66%, H 2 concentration around the subject user breathing the hydrogen-oxygen mixed gas is less than 1% (as discussed later). Thus, during normal operation it is almost unlikely to incur explosion even there is unexpected flame or ignition close to the subject user. However, in the event there is any abnormal situation happened during the operation of the hydrogen inhaler, such as leakage in the hydrogen inhaler, H 2 concentration around the subject user may be higher than 4% and there still exits risk of detonation. Therefore, a detector could be installed to monitor the H 2 concentration and the hydrogen inhaler should be immediately shut down if the detected H 2 concentration is approaching 4%. Pursuant to the aforesaid design principles, AMS-H-03 hydrogen inhaler is developed and built. It provides inhaled gas with H 2 concentration of ~66% and O 2 concentration of ~34%, at output rate of 2-3 L/min. Since O 2 concentration of the inhaled gas is higher than 21%, any combination of the inhaled gas and air for the breath of human body will not be considered as oxygen-deficient gas. During the operation, the electrolytic module of the AMS-H-03 hydrogen inhaler generates hydrogen-oxygen mixed gas (H 2 ~66%: O 2 ~34%, electrolytic solution type) which then passes through the filtering module and the liquid blocking module for purification. One additional filter is installed after the liquid blocking module to further percolate the hydrogenoxygen mixed gas. Such triple purification (including the filtering module, the liquid blocking module, and the other filter) will assure that the outputted hydrogen-oxygen mixed gas is complied with available medical gas standards. When the temperature of the electrolytic module raises due to the heat generated by the electrolytic process, one cycling module of AMS-H-03 then activated to circulate and cool down the liquid in the electrolytic module. Furthermore, in the event the water in the electrolytic module is consumed or exhausted due to the electrolytic process, another cycling module of AMS-H-03 will supply water into the electrolytic module through the filtering module, and the residual electrolyte originally blocked by the filter module will then be flushed The gas pathway of AMS-H-03, beginning from the water surface of the electrolytic module to the outlet of the nasal tube, is designed as short as possible. Moreover, such gas pathway is further divided into three isolation zones, one is between the electrolytic module and the liquid blocking module, another is between the liquid blocking module and a flame arrester coupled to the nebulizer, and the other is between the flame arrester coupled to the nebulizer and anther flame arrester close to the outlet of the nasal tube worn by a user. With the installation of the liquid blocking module and flame arresters, any unexpected explosion generated in one isolation zone will be prevented from propagating to the other isolation zone. Moreover, the nebulizer is provided to selectively atomize the water into vapor which is then mixed with the filtered hydrogen/oxygen mix gas, not only to maintain appropriate humidity in the hydrogen gas or hydrogen/oxygen mix gas for the respiratory system of subject users, but also to reduce the risk of explosion incurred by the static electricity. A multicenter randomized clinical trial that verifies the efficacy and safety regarding inhalation of H 2 (~66%) and O 2 (~34%) in patients with COVID-19 was reported based on utility of hundreds of AMS-H-03 hydrogen inhalers. It is found that inhalation of H 2 (~66%) and O 2 (~34%) significantly improves disease severity of patients with COVID-19, such as improvement of chest distress, chest pain, resting oxygen saturation, and dyspnea scale, etc. 63 As previously mentioned, the inhaled gas with higher H 2 concentration has the greater diffusion speed and lower pressure. Thus, mixture of H 2 (~66%) and O 2 (~34%) could decrease the inspiratory efforts due to the significantly lower resistance when passing through the respiratory tract, 63 and could be an option for relieving respiratory symptoms in patients with COVID-19. The inhaled gas of the AMS-H-01/AMS-H-03 is subject to the tests set forth in GB8982-2009 (CN) regarding standards of medical gas and aviation gas. With the help of filtering module and cycling module, the inhaled gas of the AMS-H-01/AMS-H-03 passes the section 5.4 of GB8982-2009 (CN) regarding the residual amount of acid material and alkaline material. Therefore, the alkali mist is totally removed from the inhaled gas of the AMS-H-01/AMS-H-03. As shown in Table 2 , during the normal operation, even H 2 concentration at the output of the nebulizer and nasal tube is around 64-67%, the H 2 concentration around the subject user is less than 1% and the H 2 concentration of exhaled breath of the subject user is less than 4%, both of which are less than the critical explosive 10% H 2 concentration in the mixture of hydrogen and air (or oxygen). 57,64-66 Therefore, during normal operation, in the event there is unexpected flame or ignition close to the subject user, it is almost unlikely to incur explosion or detonation due to the fact that the H 2 concentration surrounding the subject user is less than 4%. The ignition experiments at flow rate of 3 L/min, 2.5 L/min, and 2 L/min of hydrogen-oxygen gas (approximately 66% H 2 : 34% O 2 ) generated from the AMS-H-03 are performed. The AMS-H-03 is ignited at output of nasal tube, as shown in Figure 4 . As shown in the following Table 3 , it is found that only clean ignition noises could be detected for three different flow rates. The results are reasonable since, as previously discussed, the volume between the flame arrester in the nasal tube and the outlet of the nasal tube is quite small, almost no explosive sound is detected when ignition is made at the output of the nasal tube. Therefore, during the operation of AMS-H-03, it is likely that the user who wears the nasal tube for hydrogen inhaling would not be subject to damage even there is ignition caused by static electricity of human body. conclusion Hydrogen-oxygen mixture (approximately 66% H 2 : 34% O 2 ) for medical purpose is an oncoming and developing research topic. A safe and durable hydrogen inhaler for generating such hydrogen-oxygen mixture is proposed based on reliable design principles. Experiment and measurement result shows that the hydrogen inhaler implementing the proposed design rules could effectively reduce the explosion, and that the hydrogen-oxygen mixture generated from the hydrogen inhaler is suitable for medical purpose. This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms. additional file Additional file 1: Additional file 1. Note: Furthermore, H 2 concentration is measured by New Cosmos Electric: XP-3140 (Osaka, Japan). First, when the AMS-H-01 is tuned on, the hydrogen concentration at: (1) ambient environment which is one meter away from the AMS-H-01 ("VA"); (2) output of the nebulizer ("OA"); and (3) output of special nasal tube or nasal mask connected to the nebulizer ("ONT") are measured separately. Secondly, when each subject user is equipped with the nasal tube, H 2 concentration around the subject user is detected at the location of 1 cm away from: (1) right ear ("A"); (2) left ear ("B"); (3) top of head ("C"); and (4) back of head ("D") of the subject user after 5-minute, 60-minute, and 120-minute operation of the AMS-H-01 respectively. Last, the exhaled breath of the subject was collected using a gas sample collection bag immediately after the nasal tube was removed from the subject user, and then the hydrogen concentration of the exhaled breath was measured. Hydrogen Molecular Medicine Inhalation of hydrogen gas reduces infarct size in the rat model of myocardial ischemia-reperfusion injury Hyperbaric hydrogen therapy: a possible treatment for cancer Response of five established solid transplantable mouse tumors and one mouse leukemia to hyperbaric hydrogen Hydrogen acts as a therapeutic antioxidant by selectively reducing cytotoxic oxygen radicals Hydrogen therapy may be an effective and specific novel treatment for acute radiation syndrome Hydrogen mediates suppression of colon inflammation induced by dextran sodium sulfate Anti-inflammatory effect of hydrogen-rich saline in a rat model of regional myocardial ischemia and reperfusion Hydrogen inhalation ameliorates oxidative stress in transplantation induced intestinal graft injury Hydrogen therapy reduces apoptosis in neonatal hypoxia-ischemia rat model Molecular hydrogen: a preventive and therapeutic medical gas for various diseases Molecular hydrogen alleviates nephrotoxicity induced by an anticancer drug cisplatin without compromising anti-tumor activity in mice Hydrogen-rich saline attenuates cardiac and hepatic injury in doxorubicin rat model by inhibiting inflammation and apoptosis Hydrogen gas inhibits lung cancer progression through targeting SMC3 Consumption of hydrogen water prevents atherosclerosis in apolipoprotein E knockout mice Hydrogen-rich pure water prevents superoxide formation in brain slices of vitamin C-depleted SMP30/GNL knockout mice Oral hydrogen water prevents chronic allograft nephropathy in rats Drinking hydrogen water ameliorated cognitive impairment in senescence-accelerated mice Physical Chemistry An Introduction to Fluid Dynamics Protective effects of hydrogen inhalation on rat model of chronic obstructive pulmonary disease Hydrogen inhalation attenuated bleomycin-induced pulmonary fibrosis by inhibiting transforming growth factor-β1 and relevant oxidative stress and epithelial-to-mesenchymal transition Inhalation of hydrogen gas ameliorates glyoxylate-induced calcium oxalate deposition and renal oxidative stress in mice Inhalation of water electrolysis-derived hydrogen ameliorates cerebral ischemia-reperfusion injury in rats -A possible new hydrogen resource for clinical use Protective effect of hydrogen on the lung of sanitation workers exposed to haze Postconditioning with inhaled hydrogen promotes survival of retinal ganglion cells in a rat model of retinal ischemia/reperfusion injury Effects of hydrogen gas inhalation on endometriosis in rats Hydrogen coadministration slows the development of COPD-like lung disease in a cigarette smoke-induced rat model High-concentration hydrogen protects mouse heart against ischemia/reperfusion injury through activation of thePI3K/Akt1 pathway Molecular hydrogen increases resilience to stress in mice Inhalation of high concentrations of hydrogen ameliorates liver ischemia/reperfusion injury through A(2A) receptor mediated PI3K-Akt pathway Effect and mechanism of hydrogen suppressing non-small cell lung cancer Hydrogen gas inhalation protects against cigarette smoke-induced COPD development in mice Inhalation of hydrogen of different concentrations ameliorates spinal cord injury in mice by protecting spinal cord neurons from apoptosis, oxidative injury and mitochondrial structure damages Hydrogen gas reduces chronic intermittent hypoxia-induced hypertension by inhibiting sympathetic nerve activity and increasing vasodilator responses via the antioxidation Hydrogen ameliorates chronic intermittent hypoxia-induced neurocognitive impairment via inhibiting oxidative stress Improvement of hydrogen on liver oxidative stress injury in chronic intermittent hypoxia rats Breathing hydrogen-oxygen mixture decreases inspiratory effort in patients with tracheal stenosis Effects of post-treatment hydrogen gas inhalation on uveitis induced by endotoxin in rats Effect of hydrogen gas inhalation on postoperative delirium in elderly patients receiving hip fracture sur-gery Hydrogen gas inhalation protects against cutaneous ischaemia/reperfusion injury in a mouse model of pressure ulcer Therapeutic potential of molecular hydrogen in ovarian cancer An animal model and protection of hidden hearing loss Beneficial effects of hydrogen gas inhalation on a murine model of allergic rhinitis Hydrogen postconditioning promotes survival of rat retinal ganglion cells against ischemia/reperfusion injury through the PI3K/Akt pathway Hydrogen and oxygen mixture to improve cardiac dysfunction and myocardial pathological changes induced by intermittent hypoxia in rats Hydrogen gas alleviates chronic intermittent hypoxia-induced renal injury through reducing iron overload Hydrogen protects against chronic intermittent hypoxia induced renal dysfunction by promoting autophagy and alleviating apoptosis Hydrogen gas inhalation enhances alveolar macrophage phagocytosis in an ovalbumin-induced asthma model Protective effects of hydrogen gas in a rat model of branch retinal vein occlusion via decreasing VEGF-α expression Molecular hydrogen suppresses glioblastoma growth via inducing the glioma stem-like cell differentiation Hydrogen gas therapy induced shrinkage of metastatic gallbladder cancer: A case report Real world survey" of hydrogencontrolled cancer: a follow-up report of 82 advanced cancer patients Inhalation of hydrogen gas attenuates airway inflammation and oxidative stress in allergic asthmatic mice Notice of the 7th version of guidelines for the diagnosis and management of Covid-19 Preventing explosions of hydrogen gas inhalers Oxygen Deficiency, Honeywell International Perfluorosulfonic Acid Membranes for Fuel Cell and Electrolyser Applications Static Electricity. What is the Relationship with Humidity? Osaka Prefectural Industrial Technology Research Industrial Safety Technology Tree Association No.248 p Hydrogen/oxygen mixed gas inhalation improves disease severity and dyspnea in patients with Coronavirus disease 2019 in a recent multicenter, open-label clinical trial Flame acceleration and the development of detonation in fuel-oxygen mixtures at elevated temperatures and pressures Study on the explosion risk of hydrogen (1 st Report). Effect of pressure on the explosive limit of hydrogen. Research Report of National Institute for Industrial Safety Explosion pressure of hydrogen air mixture. Research Report of National Institute for Industrial Safety Date of decision Date of acceptance Date of web publication The references cited in this article are mainly based on search of the published literature in Internet and electronic databases, such as PubMed, Science Direct, and Wiley Online Library, etc. Related papers or publications were selected and reviewed based on the search terms, including but not limited to "hydrogen", "medical gas", "cancer", "respiratory", "inhaler", "explosion", and "Asclepius Meditec", etc.