key: cord-0923324-qhma38yq authors: Thieß, Hans-Martin; Bressem, Keno K; Adams, Lisa; Böning, Georg; Vahldiek, Janis L; Niehues, Stefan M title: Do submillisievert-chest CT protocols impact diagnostic quality in suspected COVID-19 patients? date: 2022-01-19 journal: Acta Radiol Open DOI: 10.1177/20584601211073864 sha: b9569ee393aa5839eeb2faef4972a6113812bb74 doc_id: 923324 cord_uid: qhma38yq BACKGROUND: During the ongoing global SARS-CoV-2 pandemic, there is a high demand for quick and reliable methods for early identification of infected patients. Due to its widespread availability, chest-CT is commonly used to detect early pulmonary manifestations and for follow-ups. PURPOSE: This study aims to analyze image quality and reproducibility of readings of scans using low-dose chest CT protocols in patients suspected of SARS-CoV-2 infection. MATERIALS AND METHODS: Two radiologists retrospectively analyzed 100 low-dose chest CT scans of patients suspected of SARS-CoV-2 infection using two protocols on devices from two vendors regarding image quality based on a Likert scale. After 3 weeks, quality ratings were repeated to allow for analysis of intra-reader in addition to the inter-reader agreement. Furthermore, radiation dose and presence as well as distribution of radiological features were noted. RESULTS: The exams’ effective radiation doses were in median in the submillisievert range (median of 0.53 mSv, IQR: 0.35 mSv). While most scans were rated as being of optimal quality, 38% of scans were scored as suboptimal, yet only one scan was non-diagnostic. Inter-reader and intra-reader reliability showed almost perfect agreement with Cohen’s kappa of 0.82 and 0.87. CONCLUSION: Overall, in this study, we present two protocols for submillisievert low-dose chest CT demonstrating appropriate or better image quality with almost perfect inter-reader and intra-reader agreement in patients suspected of SARS-CoV-2 infection. Since December 2019, the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), which originated from Wuhan, China, showed a rapid worldwide spread with more than 145 million infections and 3.1 million deaths as of April 2021. 1 Patients commonly present with several unspecific clinical symptoms, such as fever, dry cough, dyspnea, fatigue, and limb pain, with an unknown number of asymptomatic patients. The continuous increase in new infections and a lack of effective treatment apart from vaccinations, which are still restricted by insufficient supply, demand for fast and reliable methods for early identification of infected patients. This is aggravated by the current spread of new, possibly more infectious SARS-CoV-2-variants, 2 such as the British (B.1.1.7) and the South African variants (N501Y.V2). Currently, real-time reversetranscription polymerase chain reaction (RT-PCR) of throat swabs, deep nasal swabs, or sputum represents the gold standard for detecting SARS-CoV-2. 3 However, chest CT can play a vital role in the initial identification of infected patients for several reasons: First, due to low examination times and fast availability of results, it can bridge the time until RT-PCR results arrive, 4 thus enabling the early isolation of patients with CT findings typical of COVID-19 preventing further transmission, especially when combined with antigen rapid tests. However, these rapid tests present a widely varying sensitivity depending on which assay is used and current viral load, 5-7 the latter underlining utility of antigen rapid tests within the first week after symptom onset. 5 Second, there have been several reports of falsenegative results of initial RT-PCR [8] [9] [10] and studies stating a higher (88%-97% vs. 59%-71%, 11, 12 ) or similar sensitivity albeit low specificity of chest CT compared to RT-PCR. 3 Thus, patients with negative RT-PCR and positive CT findings may benefit from additional subsequent PCR testing, especially when providing an epidemical exposure history and/or clinical symptoms coherent with SARS-CoV-2 infection. Third, the wide availability of CT compared to remaining regional shortages of RT-PCR kits accentuates its usefulness in the early diagnosis of COVID-19. Apart from its role in the triage of possibly infected patients, CT can also be of use for follow-up to determine the course of the disease, effectiveness of treatment, and possible complications, such as acute respiratory distress syndrome 13 or abscess. Furthermore, it can help determine additional causes for rather unspecific clinical symptoms, such as lobar pneumonia, emphysema, cardiac insufficiency, and others. There are several findings in chest-CT of patients with COVID-19 considered to be typical, such as bilateral manifestation with multifocal ground-glass opacities and consolidations (predominantly in a basal and peripheral distribution), vascular thickening and a crazy-paving pattern. 14 These may correlate with different times in the course of the infection, severity of the course of the disease, and necessity of ICU admission. 15, 16 The major drawback of CT is radiation exposure, which can be reduced by using techniques, such as iterative reconstruction (e.g., SAFIRE/ ASIR/AIDR) in combination with specific low-dose protocols. 17 Due to the high examination count of chest-CT during the coronavirus pandemic, we introduced new, dedicated low-dose protocols for COVID-19 diagnostic, which this study aims to analyze regarding image quality, radiation dose, and reproducibility. For this retrospective study, we analyzed 100 noncontrast low-dose CT-examinations between April of 2020 and February of 2021. To avoid bias, we randomly selected scans from 50 patients with and 50 patients without confirmation of SARS-CoV-2 infection via RT-PCR from deep nasal or throat swabs. Patients included presented in the emergency department with symptoms suspicious of SARS-CoV-2 infection. This study was priorly approved by the institutional review board (EA4/ 140/17). Image acquisition was conducted using either a Canon Aquilion Prime or GE Lightspeed VCT. Exams included in the study were performed only of the thorax, and no additional scans using different parameters were performed. All studies were non-contrast helical scans in supine position, from the diaphragm upwards andwhen possible-in deep inspiration, using examination parameters and reconstruction methods as seen in Table 1 . The selected parameters were based on our experience with these scanners utilizing CT protocols routinely used for interstitial lung disease imaging focusing on a middle ground between image noise and the lung as a high-contrast organ. All images were obtained on either an 80-row detector (Aquilion Prime, Canon Medical Systems) or a 64-row detector (GE Lightspeed VCT, GE Healthcare Systems). All parameters used are summed up in Table 1 . Individual CT dose index (CTDI vol ) and dose length product (DLP) of each exam was extracted from the DICOM-header. Effective radiation dose was estimated by multiplying DLP by a conversion factor for chest imaging at 100 kV: 0.0144 mSv × mGy À1 × cm À1 . 18 Two radiological residents with 5 and 2 years of experience in CT imaging (KKB and HMT) independently rated all exams on a Workstation using MERLIN Diagnostic Workcenter by Phoenix-PACS regarding image quality on a Likert scale from 0 to 3 (0: best quality, no limitations in diagnostic value; 1: good quality, only slight impairments; 2: moderate quality, noticeable decrease in diagnostic value; and 3: worst quality, non-diagnostic scan). The evaluation was based on subjective ratings for the parameters image sharpness (defined as delineation of margins, such as bronchial walls or interlobar fissures in not breath dependent lung sections), image noise in lung parenchyma and soft tissue, respectively, as well as ring and other artifacts, for example, those caused by foreign materials. Three weeks after the initial rating, reevaluation of all included scans was performed, allowing for analysis of intra-reader reliability. To allow for additional assessment of organizational/fibrotic changes of lung parenchyma, including inter-reader reliability, both readers furthermore scored scans for the extent of reticulation, honeycombing, and emphysema to the nearest 5% in the following three zones: The upper zone being defined as at or above the aortic arch, the middle zone as between the aortic arch and the confluence of the pulmonary veins, and the lower zone as at or below the pulmonary veins. 19 The extent of lung fibrosis was then calculated as the mean of the mean extent of reticulation and honeycombing for each zone. Readers only used axial images for ratings yet were allowed to change window settings. Additionally, based on the study by Wang et al., 20 presence and extent of pathological findings of the lung, such as consolidations, ground-glass opacities, distribution, and pleural effusion, were noted; presence of crazy-paving patterns was not included in accordance with Wang et al.'s results. Statistical analysis was performed using the R statistical language Version 4.0.0 and the "tidyverse" library. 21 Quantitative parameters (e.g., patient characteristics, presence/extent of radiological features) were expressed as median value and interquartile ranges. Cohen's kappa was used for analysis of inter-reader and intra-reader reliability with κ-values characterized as 0-0.20 = poor, 0.21-0.40 = fair, 0.41-0.6 = moderate, 0.61-0.80 = substantial, and 0.81-1.00 = (almost) perfect agreement. 22 Differences in quality ratings were analyzed depending on RT-PCR results, presence of CT findings typical of COVID-19, and patients' sex using Pearson's chi-squared test. A p-value of < 0.05 was considered statistically significant. One hundred patients were included in this study, 38 females and 62 males, with a mean age of 71 years (IQR: 22.0). Of those subjects, 17 female and 33 male patients were tested positive for SARS-CoV-2 via RT-PCR on the same day equaling 44.7% and 53.2%, respectively. Data about patients' BMI was available for 40 patients with a mean BMI of 27.2 kg/m 2 (IQR: 6.3 kg/m 2 ). Median DLP was 29.6 mGy*cm (IQR: 29.5 mGy*cm) for females and 40.7 mGy*cm (IQR: 19.7 mGy*cm) for males, which results in a median effective dose of 0.43 mSv (IQR: 0.42 mSv) for females, 0.59 mSv (IQR: 0.28 mSv) for males, and 0.53 mSv (IQR: 0.35 mSv) overall using a tissue conversion factor of 0.0144 mSv/(mGy*cm) for chest CT. 18 Median DLP was 36.3 mGy*cm (IQR: 30.6 mGy*cm) using the Aquilion Prime and 39.6 mGy*cm (IQR 14.8 mGy*cm) using the Lightspeed VCT. This results in a median effective dose of 0.52 mSv and 0.57 mSv. The radiological findings in patients suspected of COVID-19 pneumonia based on positive CT results at presentation are summarized in Table 2 . Among these, most scans (51.9%) with ground-glass opacities and/or consolidations showed a peripherally accentuated pattern with basal (31.5%) or dorsal (24.1%) distribution (Figure 1 ). Pleural effusion was present in only 10 patients (18.5%), and cavitation was seen in one patient. Presence of crazy-paving patterns was not detected in this study, and extent of reticulation was only scored for the purpose of calculating a fibrosis score. The mean extent of fibrosis across the three predefined zones was 2.4%, and the mean extent of emphysema was 0.4%. The majority of all scans were rated unrestricted positive (Likert score = 0, n = 62). 32 patients had slight limitations (Likert value = 1). For those rated impaired (n = 6, Likert score = 2 + 3), limitations in patient positioning or motion artifacts accounted for the rating (see and only in three cases due to foreign bodies (internal or external). These artifacts primarily impaired image quality of soft tissues while only mildly affecting the evaluation of lung parenchyma (see Figure 3 ). The second most common reasons for decreased ratings were a lack of sharpness (e.g., blurriness of the interlobar fissure in the upper thorax) or high noise (grainy appearance of mediastinal/subcutaneous adipose tissue or air-filled structures), the latter too predominantly affecting the evaluation of soft tissues. In a few cases, distinguishing pleural effusions from atelectasis was hardly possible, and one scan showed movement artifacts throughout the entire chest, leading to reduced ratings. Only one scan was evaluated as non-diagnostic (score = 3) by both raters due to excessive artifacts throughout the complete scan caused by lowered arms (Figure 4) . Table 3 sums up the reasons for reduction of image quality ratings. There was no significant difference in image quality ratings depending on the result of RT-PCR (p = .09). We were neither able to show a significant difference between quality ratings depending on the patient's sex (p = .44) nor findings in chest CT (p = .50). Inter-reader reliability for the assessment of image quality was almost perfect with a κ value of 0.82 (p < .001) for all scans and substantial for scans of PCRpositive patients (κ = 0.79, p < .001). Analysis of intrareader reliability demonstrated similar results with a κ value of 0.91 (p < .001) for patients with confirmed SARS-CoV-2 infection and 0.82 (p < .001) for patients without confirmed disease. Thus, the overall intra-reader reliability of all scans was almost perfect too, with a κ value of 0.87 (p < .001). Subgroup analysis, based on which scanner was used, showed substantial agreement (κ = 0.76, p < .001) for inter-and almost perfect agreement (κ = 0.91, p < .001) for intra-reader reliability for the Canon Aquilion Prime compared to almost perfect agreement (κ = 0.81, p < .001) for inter-and substantial agreement (κ = 0.71, p < .001) for intra-reader reliability for the GE Lightspeed VCT. Assessment of the extent of lung fibrosis and emphysema showed almost perfect and perfect inter-reader reliability with a κ of 0.995 and 1, respectively (p < .001). During the COVID-19 pandemic, reliable and readily available diagnostic methods to identify infected patients are in high demand. Chest-CT, being one of those methods due to its frequent use in patients with suspected COVID-19 pneumonia, makes it imperative to reduce radiation dose during scans as much as reasonably achievable, in accordance with the ALARA principle. In this study, we analyzed two low-dose chest-CT protocols using scanners from two different manufacturers, focusing on possible reductions in image quality and inter-/intra-reader reliability of those ratings. Using our low-dose protocols, we achieved submillisievert effective doses in both female and male patients. Steuwe et al. reported in their study of 105 patients from a similar, middle-European demographic a mean DLP of 89.3 ± 27.7 mGy*cm. They calculated an effective dose of 1.3 ± 0.4 mSv, 23 which is substantially higher than our findings due to an increased reference mAs of 60 at 100 kV. Dangis et al. achieved similarly low radiation doses in their study, including 192 middle-European patients, with a mean DLP of 39.9 ± 17.8 mGy*cm resulting in a mean effective dose of 0.56 ± 0.25 mSv. 24 An even lower radiation dose was reported by Agostini et al., 25 who compared nonlow-dose to low-dose chest CT. They stated a median DLP of 19.5 mGy*cm (IQR: 17.5-29.02) and a median effective dose of 0.28 mSv (IQR: 0.25-0.42) for low-dose CT as opposed to 226.2 mGy*cm (IQR: 176.0-322.0) and 3.28 mSv (IQR: 2.55-4.67) for nonlow-dose scans. This resulted in a median dose reduction of 90.6% using long-pitch, dual-source acquisition at 100 kV and an increased reference mAs of 180 with spectral shaping. Spectral shaping using a dedicated tin filter reduces radiation dose by removing a portion of the low-energy photons from the X-ray beam, which usually would not reach the detector and thus does not contribute to image acquisition. This effect was demonstrated by Hau Despite the very low radiation doses, our radiologists rated most scans as optimal or slightly suboptimal in image quality, and only a few scans as suffering from limitations in diagnostic value, primarily due to artifacts caused by arms or foreign material within the scan range. Only one scan was rated as being non-diagnostic due to significant artifacts caused by lowered arms. Therefore, in patients, who cannot raise their arms, especially if findings within the soft tissue are of interest to the radiologist, regular chest CT should be considered. The fact that only 62% of scans were given optimal ratings and only a few scans were rated as below average indicates an adequate reduction in radiation dose without loss of diagnostic value. The presented examination technique shows results comparable to Agostini et al. but omits the requirement of spectral shaping and can thus be performed with hardware of other vendors. 25 However, Agostini et al. did not find significant reductions in diagnostic reliability and evaluation of pathological findings using their low-dose protocol. Moreover, Agostini et al. also did not report poor or nondiagnostic image quality scans. In their analysis of signalto-noise ratio (SNR) and contrast-to-noise ratio (CNR), values were overall higher in HD-DECT; however, no significant difference was found when comparing these values for the lung parenchyma. Steuwe et al. presented similar results, stating 13% of low-dose scans as having above average image noise ratings and none as having unacceptable image noise, based on subjective ratings on a five-point Likert scale. 23 Neither Steuwe et al. nor Agostini et al. analyzed their results concerning inter-reader or intrareader reliability, the latter due to the low number (10) of included patients. Inter-reader reliability showed almost perfect agreement in assessing image quality. Three weeks after the first rating, reevaluation too showed almost perfect agreement for patients with and without confirmed SARS-CoV-2 infection, demonstrating the reproducibility of scan results and further underlining reliability of chest-CT in a clinical setting. Inter-reader reliability for assessment of the extent of lung fibrosis and emphysema showed almost perfect and perfect agreement, respectively. Validity, however, is limited due to the patient population not having been selected based on the presence of such chronic changes and thus only few scans demonstrating fibrosis/emphysema. Furthermore, COVID-19-related changes of lung parenchyma, such as ground-glass opacities and consolidations, may mask fibrosis and emphysema. Radiological findings in our patients are mostly in line with those considered typical of COVID-19, as described by several studies, such as the study by Wang et al. They documented ground-glass opacities, consolidations, or both in all patients undergoing chest-CT, usually in a peripheral (43.6% of subjects) or combined peripheral and central distribution pattern (56.4%). 20, 28 Similar results were reported by Pan et al., who also defined four stages of lung involvement from the early stage (days zero to four) to an absorption stage (≥ 14 days after symptom onset) with an increasing extent of lung opacities until 9-13 days after onset of symptoms and steady decrease thereafter over a time span of in some cases more than 26 days total. 15 Table 3 . Reasons for reduced image quality ratings (multiple answers possible per scan). Reason for a reduction in image quality rating In future research, low-dose scans will be using more recent, AI-assisted reconstruction methods (i.e., AiCE/ DLIR). These could additionally be checked for the extent of dose reduction and impact on image quality. In conclusion, in this study, we demonstrated two protocols for chest-CT with a median ED of 0.53 mSv suitable for examining patients with suspected SARS-CoV-2 infection, which provide reliable scans without significant sacrifices in perceived image quality or diagnostic value. The major limitations of our study are the retrospective design, including patients from two centers and the limited sample size. Although the data was acquired on devices from two different manufacturers, these results cannot be guaranteed to be transferred to other manufacturers or other devices with different operating or reconstruction software or design. As reviewers were not blinded to PCR results, this may have caused a bias in image quality ratings. Due to the retrospective design of the study and missing documentation of objective criteria (i.e., scoring) for the clinical severity of COVID-19 in infected patients, we were unable to analyze correlation between clinical severity and image quality ratings or inter-/intra-rater reliability. COVID-19 situation update worldwide, as of week 16, updated 29 Assessing transmissibility of SARS-CoV-2 lineage B. 1.1. 7 in England Systematic review with meta-analysis of the accuracy of diagnostic tests for COVID-19 Lowdose chest CT for the diagnosis of COVID-19 Recommendations for use of antigenic tests in the diagnosis of acute SARS-CoV-2 infection in the second pandemic wave: attitude in different clinical settings Limits and opportunities of SARS-CoV-2 antigen rapid tests: an experienced-based perspective WHO. Antigen-detection in the diagnosis of SARS-CoV-2 infection using rapid immunoassays: Interim guidance Chest CT for typical coronavirus disease 2019 (COVID-19) pneumonia: relationship to negative RT-PCR testing Positive rate of RT-PCR detection of SARS-CoV-2 infection in 4880 cases from one hospital in A case report of COVID-19 with false negative RT-PCR test: necessity of chest CT Sensitivity of chest CT for COVID-19: comparison to RT-PCR Correlation of chest CT and RT-PCR testing for coronavirus disease 2019 (COVID-19) in China: a report of 1014 cases Acute respiratory failure in COVID-19: is it "typical Chest CT manifestations of new coronavirus disease 2019 (COVID-19): a pictorial review Time course of lung changes at chest CT during recovery from coronavirus disease 2019 (COVID-19) Diagnostic value of initial chest CT findings for the need of ICU treatment/ intubation in patients with COVID-19 Submillisievert chest CT in patients with COVID-19 -experiences of a German Level-I center Multisection CT protocols: sex-and age-specific conversion factors used to determine effective dose from dose-length product Idiopathic pulmonary fibrosis: physiologic tests, quantitative CT indexes, and CT visual scores as predictors of mortality Imaging manifestations and diagnostic value of chest CT of coronavirus disease 2019 (COVID-19) in the Xiaogan area Tidyverse: easily install and load the 'tidyverse. R Package Version The measurement of observer agreement for categorical data Dose-optimised chest computed tomography for diagnosis of Coronavirus Disease 2019 (COVID-19) -evaluation of image quality and diagnostic impact Accuracy and reproducibility of low-dose submillisievert chest CT for the diagnosis of COVID-19 Proposal of a lowdose, long-pitch, dual-source chest CT protocol on thirdgeneration dual-source CT using a tin filter for spectral shaping at 100 kVp for CoronaVirus Disease 2019 (COVID-19) patients: a feasibility study Unenhanced third-generation dual-source chest CT using a tin filter for spectral shaping at 100 kVp The efficacy of tinfiltration for computed tomography in diagnosing urolithiasis Imaging and clinical features of patients with 2019 novel coronavirus SARS-CoV-2 This work is part of the doctoral thesis of Hans-Martin Thieβ at the Charité -University Medicine Berlin. The author(s) declared the following potential conflicts of interest with respect to the research, authorship, and/or publication of this article: Stefan Markus Niehues declares relationships with the following companies: Vital Images, CanonMedical Systems, Guerbet, Bracco Imaging, Teleflex/Vidacare. The author(s) received no financial support for the research, authorship, and/or publication of this article. Hans-Martin Thieß  https://orcid.org/0000-0002-2381-6075 Georg Böning  https://orcid.org/0000-0002-8819-6209 Janis L Vahldiek  https://orcid.org/0000-0002-1537-1424