key: cord-0923186-8bghg3ox authors: Chan, Shannon Melissa; Ma, Tsz Wah; Ka-Chun Chong, Marc; Chan, Daniel Leonard; Ng, Enders Kwok Wai; Chiu, Philip Wai Yan title: A proof of concept study: Esophagogastroduodenoscopy is an aerosol-generating procedure and continuous oral suction during the procedure reduces the amount of aerosol generated date: 2020-07-07 journal: Gastroenterology DOI: 10.1053/j.gastro.2020.07.002 sha: 8a696bcccd944275346dff8a759d7545d2057acb doc_id: 923186 cord_uid: 8bghg3ox nan Email: philipchiu@surgery.cuhk.edu.hk This study was not supported by any research grant. All authors do not have any conflicts of interest to disclose. Aerosol-generating procedures pose a potential threat to healthcare workers, especially during this coronavirus 2019 (COVID-19) pandemic. Esophagogastroduodenoscopy (EGD) was assumed to be an aerosol-generating procedure and recommendations, therefore reflect evidence generated from non-gastrointestinal (GI) aerosol-generating procedures such as bronchoscopy. However, there is no scientific evidence to support this claim. This study aims to provide scientific evidence on whether EGD is an aerosol-generating procedure and to examine ways of decreasing the amount of aerosol generated. This study was a prospective observational trial to examine the aerosol generation of EGD applying a quantitative approach (see additional text for details). All patients undergoing EGD at the endoscopy center of the Prince of Wales Hospital from 7 th May 2020 to 1 st Jun 2020 were included. Procedures were performed with the patient in the left lateral position with a mouthguard, using a 9.9 mm flexible video gastrointestinal scope (GIF-H290) (Olympus Hong Kong and China Limited). Measurements were taken using the portable GT-526S Handheld Particle Counter (Met One Instruments, Inc., United States). The six-channel particle sizes were programmed at 0.3μm, 0.5μm, 0.7μm, 1μm, 5μm, and 10μm. The particle counter was placed within 10cm from the mouth of the patient once the patient entered the room and measured for at least one minute before the start of the procedure. The measurement was continued during the procedure until after the patient left the endoscopy suite. Multilevel modeling was used to test all the hypotheses in the study 1 . The particle counts per cubic feet of each particle size are presented in dCF. We tested the association between the uses of sedation, dental sucker, and Log(dCF). The interaction terms between sedation and procedure, and between dental sucker and procedure were added to the multilevel models. Age, sex, endoscopist seniority, procedure length, diagnostic or therapeutic procedure, use of biopsy were included as confounders. A random effect of repeated measures was accounted for in the models to capture the intra-subject variability for the changes in dCF over time. From 7 th May 2020 to 1 st Jun 2020, a total of 93 patients were recruited into the study. There were 59 (63.4%) unsedated patients and 34 (36.6%) sedated patients. 30 (32.3%) patients had the use of the dental sucker. Majority of the procedures were diagnostic procedures (n=85; 91.4%) with biopsy taken (n=68, 73.1%). A multivariate analysis was also performed with the results summarized in Table 1 . During the EGD procedure, the level of dCF of all sizes were significantly higher than the baseline period (ps <0.001 to 0.02). The use of dental sucker significantly reduced the amount of particles with sizes 0.3 μm, 0.5 μm, 0.7 μm, 1 μm, 5 μm and 10 μm expelled during the procedure as compared to baseline (p<0.001, p<0.001, p<0.001, p=0.02, p<0.01, p=0.046 respectively). Simple slope tests revealed that when compared with baseline, the number of dCF_0.3, dCF_0.5, dCF _0.7, dCF_1 during EGD were significantly increased among procedures performed without dental sucker (p<0.01, p<0.01, p<0.01, p=0.02 respectively) while non-significantly decreased among participants when dental sucker was used. For larger particle sizes of 5um and 10um, the use of dental sucker was associated with a significantly smaller magnitude of increase in particles during the procedure (p<0.01, p<0.01 respectively) (Appendix 1). There were no significant differences in the level of all dCF outcomes between the individuals with or without sedation (ps = .13 to .96). Sedation did not affect the association between procedure time and all dCF outcomes. The above results are summarized in Table 1 . One per-oral endoscopic myotomy (POEM) was recorded. The procedure was performed under general anesthesia (GA). The data showed that there was a surge in all particle sizes during the initial endoscope intubation and diagnostic EGD (Appendix 2). Currently, the term droplet is often taken to refer to droplets >5 microns (μm) in diameter while the term aerosol refers to particles ≤5 μm in diameter that can remain suspended in air for a significant time, allowing them to be transmitted over longer distances at >1 meter. The increase of particle size at 0.3 μm, 0.5 μm, 0.7 μm, and 1 μm supports EGD as an aerosolgenerating procedure. Another important finding of this study was that the use of dental sucker in the oral cavity for continuous suction during the procedure significantly decreased the particle counts of all sizes detected during EGD, suggesting the use of the dental sucker may be recommended during EGD. The usual practice in our center was to provide intermittent oral suction via suction catheter. The postulation was continuous oral cavity suction reduced pooling of saliva and thus may reduce aspiration and patients' coughing. Another reason may be the dental sucker also partially suctioned out the particles that were generated during the procedure. The smaller particles may be subject to the negative pressure effect generated by the continuous suction of the dental sucker, thus causing a paradoxical drop in the particle counts during the procedure. In this study, we found that conscious sedation was not able to reduce the amount of aerosol and droplets generated during the procedure. This may be due to small sample size and the presence of confounding factors. We also found that there was still a surge of all particle sizes during GA suggesting that GA may not eliminate the aerosol generated. However, the increase in aerosol does not equate to an increase in the infectivity of respiratory viruses during EGD. Despite numerous studies on the transmission routes of respiratory viruses over the past, the results remained inconclusive 2 . SARS-CoV2 RNA, however, has been found in fecal samples of infected patients for a prolonged period of time 3 . The role of the gastrointestinal tract in the transmission of SARS-CoV2, remains uncertain. All in all, EGD is an aerosol-generating procedure and the use of dental sucker may decrease the amount of aerosol generated, and thus decreasing the risk to healthcare workers. These findings may inform current guidelines in infection control and the most appropriate PPE during digestive endoscopy. Chiu Table 1 The association between sedation and log (dCF_0.3, dCF_0.5, dCF_0.7, dCF_1, dCF_5, dCF_10). Appendix 1 (a) Line graphs of particle counts size 0.3, 0.5, 0.7, 1, 5, 10 µm of an unsedated diagnostic procedure without the use of dental sucker. (b) Line graphs of particle counts size 0.3, 0.5, 0.7, 1, 5, 10 µm of an unsedated diagnostic procedure with the use of dental sucker. (c) The change in in LN_dCF during the procedure as compared to baseline with or without sucker. The use of dental sucker reduced the association between timing of procedure (before vs. during) and dCF_0. 3 Confounders were controlled at the multilevel modeling. The line graph showing the change in dCF during a per-oral endoscopic myotomy (POEM) procedure, done under general anesthesia with CO2 insufflation. There was a surge seen in all particle sizes during the initial intubation of the endoscope and diagnostic EGD. There was also a significant increase in all particles generated once the energy cutting devices were used. These findings suggest that general anesthesia may not have a protective effect on the amount of aerosol generated and the use of energy cutting devices generates a significant amount of aerosols. In this study, we included adult patients (≥ 18 years of age) who received an EGD in an elective or emergency setting, for either a diagnostic or therapeutic indication. Exclusion criteria were patients undergoing endoscopy other than EGD, patients undergoing EGD outside of the All procedures (except for POEM) were performed with air insufflation with high flow setting was used in all patients. All EGD were conducted in a designated procedure room (including the POEM), measuring 37m 2 , with a total supply of 6 air changes per hour with internal recirculation, at 23°C ± 1°C and humidity ≤ 60%. Reflecting on our institutional practice, endoscopy was performed by both surgeons and physicians, and conscious sedation was provided by the endoscopists. During these times of COVID-19 pandemic, the procedures were performed either by a specialist or by trainees with at least 4 years of experience in EGD. All patients were given 8 puffs of 10% topical xylocaine before the procedure. The decision for conscious sedation was up to the endoscopists' discretion usually depending on the patients' condition, the anticipated difficulty, or duration of the procedure. Conscious sedation was provided by either intravenous midazolam with or without intravenous pethidine. The endoscopy team included the chief endoscopist, endoscopy nurse, and airway nurse. Extra personnel were strictly forbidden once the patient enters the room to minimize the disturbance to the room's airflow and particle flow. All staff wore surgical N95 face masks and enhanced personal protective equipment, per the recent Asian Pacific Society for Digestive Endoscopy position statements for the practice of endoscopy during COVID-19 pandemic 4 . A dental sucker (DIS-110 Saliva Ejector, Svenska Dentorama AB, Sweden) was also placed in the oral cavity in the latter part of the study to provide continuous suction during the procedure (Figure 1a & b) . The suction pressure was set at 30kPa. Patient demographics, seniority of endoscopists, use of conscious sedation, use of dental sucker, patient flow and endoscopy timings, and air particle count (particles/cubic feet, dCF) were recorded. Timings were marked for patients entering the room, scope in, scope out, and when patients left the room. Measurements were taken using the portable GT-526S Handheld Particle Counter (Met One Instruments, Inc., United States) and extracted using Comet Deploy v2.0.4 software (Met One Instruments, Inc., United States). The six-channel particle sizes were programmed at 0.3μm, 0.5μm, 0.7μm, 1μm, 5μm, and 10μm. The 0.3-10μm range selected represents the minimum and maximum particle sizes detectable by this device. Samples were collected over periods of 5 seconds with continuous monitoring, and there were no intervals in between measurements. The particle counter was calibrated at the beginning of each endoscopy session, at least 30 minutes before the first procedure to ensure a stable baseline reading. The particle counter was placed within 10cm from the mouth of the patient once the patient enters the room and measured for at least one minute before the start of the procedure. The measurement was continued during the procedure until after the patient left the endoscopy suite. The endoscopy team was blinded to the measurements of the particle counter. The methodology used in this study is to use laser particle counters to detect a change in the concentration of particles before and during EGD. Laser particle counters are used to count the size of particles suspended in both air and liquids. They are predominantly used as a tool for characterizing cleanrooms and other contamination-controlled areas. To count particles laser light is used to reflect light from the particles as they pass through the laser beam. Airborne The room used for this observational cohort measures 37m 2 with 6 air changes per hour. The particle counter was placed directly in front of the patient's mouth before and during the procedure. All personnel in the room were restricted to the endoscopist, the airway nurse, and the endoscopy nurse, and the doors were closed during the procedure. They were all equipped with enhanced PPE with N95 masks according to international guidelines. It also served to minimize the change of particles in the air introduced by talking and breathing of the team. The number of particles of different sizes in the room were relatively stable just before the procedure, which was regarded as the baseline. Therefore, any increase in particle counts during the procedure should be generated by the procedure itself. Hierarchical Linear Models: Applications and Data Analysis Methods Transmission routes of respiratory viruses among humans Prolonged presence of SARS-CoV-2 viral RNA in faecal samples