key: cord-0919419-r39s8rw1 authors: Xun, H.; Shallal, C.; Unger, J.; Tao, R.; Torres, A.; Vladimirov, M.; Frye, J.; Singhala, M.; Horne, B.; Yesantharao, P.; Kim, B. S.; Burke, B.; Montana, M.; Talcott, M.; Winters, B.; Frisella, M.; Kushner, B.; Sacks, J.; Guest, J.; Kang, S. H.; Caffrey, J. title: Vent-Lock: A 3D Printed Ventilator Multiplexer to Enhance the Capacity of Treating Patients with COVID-19 date: 2020-09-22 journal: nan DOI: 10.1101/2020.09.16.20195230 sha: 20ff0dfaf549d0983aa05201bbc0589950e8bcaa doc_id: 919419 cord_uid: r39s8rw1 Mechanical ventilators are essential to patients who become critically ill from acute respiratory distress syndrome (ARDS), and shortages have been reported due to the novel severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). We utilized cost-effective, on-demand 3D printing (3DP) technology to produce critical components for a novel ventilator multiplexer system, Vent-Lock, to split one ventilator or anesthesia gas machine between two patients. FloRest, a novel 3DP flow restrictor, provides clinicians control of tidal volumes and positive end expiratory pressure (PEEP), using the 3DP manometer adaptor to monitor pressures. We tested the ventilator splitter circuit in simulation centers between artificial lungs and used an anesthesia gas machine to successfully ventilate two swines. As one of the first studies to demonstrate splitting one anesthesia gas machine between two swines, we present proof-of-concept of a de novo, closed, multiplexing system, with flow restriction for individualized patient therapy. Our studies underscore that while possible, ventilator multiplexing is a complicated synergy between machine settings, circuit modification, and patient monitoring. Consequently, ventilator multiplexing is reserved only as a last emergency resource, by trained clinicians and respiratory therapists with ventilator operative experience. The novel severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) has led to a 55 global pandemic resulting in rapid depletion of resources necessary to care for critically ill 56 patients, such as mechanical ventilators and their associated parts. Mechanical ventilators are 57 critical for the treatment of approximately 5-10% of patients with coronavirus disease (COVID-58 without tooling and expensive machines 21 . 3D printing enables faster design and manufacturing 123 processes 22 , so that it can be utilized in emergency situations to fill gaps in the supply chain 23 . 124 Specifically, in reference to COVID-19, there have been multiple shortages in ventilator 125 associated equipment and valves, leading healthcare providers to look into different avenues of 126 manufacturing to address the gap in the supply 20 . Ventilator splitter products which primarily 127 use commercial medical equipment may have limited or unreliable supply 17 in urgent situations. 128 We address urgent medical needs arising from unstable supply chains by using 3D printing to 129 rapidly and cost-effectively prototype and test components of our ventilator splitter circuit using 130 biocompatible and sterilizable materials. 131 control versus pressure control) on multiplexing, and additional challenges. The novel, de novo, 146 3DP Vent-Lock circuit and FloRest is well positioned to rapidly increase capacity of mechanical 147 ventilators to provisionally meet ventilator shortages due to the COVID-19 pandemic and future 148 pandemics and disasters. 149 150 RESULTS 151 We validated a 1+n(1) system which can split one ventilator between one standard patient 153 and one or potentially more variable patients (Fig.1) . The standard patient ideally has the lowest 154 lung compliance and has minimal components in the circuit to establish low resistance allowing 155 the ventilator to maintain standard function. The standard patient will be ventilated at pressure 156 settings unaltered from that delivered by the ventilator. Additional patients (n) added to the 157 circuit are considered variable patients and can have their tidal volumes and PEEP altered by 158 circuit components. This paper demonstrates use of a ventilator splitter adjusting for 1 control 159 and n=1 variable patients. The 1+n(1) split contains Vent-Lock 3DP parts and commercial parts 160 ( Fig. 1) . We 3D printed the splitters and the flow restrictors (needle valves). The other parts 161 including the one-way check valves, the filters and the PEEP valves were all commercial parts 162 ( fig. S1 ). The Vent-Lock circuit was designed to be closed circuit and leak-free to minimize risk 163 of aerosolizing viral particles into the surrounding environment. 164 165 The Vent-Lock 3DP flow restrictor (Vent-Lock FloRest) is a flow restrictor based on a 167 needle control valve design optimized for low flow rates to offer clinicians robust control over a 168 . CC-BY-NC-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The goal of FloRest ( Fig. 2A) was to allow the clinician to modify the flow rate of air 180 being delivered to the patient, thus providing ranges of clinical tidal volumes and PEEPs with 181 control sensitivity and a reliable linear relationship between closure and flow rate, tidal volume, 182 ed . CC-BY-NC-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. CC-BY-NC-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint We tested the use of Vent-Lock FloRest in the Johns Hopkins Medicine Simulation 239 Center (JHMSC) to confirm the following: 1) Allowing volumes to be distributed evenly 240 between patients, 2) variable patient control of PEEP, 3) variable patient tidal volume control 241 and 4) changes in the variable patient breathing settings does not alter breathing dynamics to the 242 standard patient. 243 We tested the Vent-Lock multiplexing system using a 1+1 split patient circuit (Fig. 1) . 244 We used one ventilator (Puritan Bennett 840 Ventilator System, Avante Health Solutions) to 245 ventilate two patients with different lung compliances of 20 cmH 2 O and 50 cmH 2 O. We first 246 tested using a pressure control mode, with inspiratory pressures set at 25 cmH 2 O (additional 247 . CC-BY-NC-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The copyright holder for this this version posted September 22, 2020. . https://doi.org/10.1101/2020.09.16.20195230 doi: medRxiv preprint ventilator settings available in Table S2 ). The Vent-Lock FloRest allowed adjustment of tidal 248 volumes delivered to patients between 7-ଷ ସ turns to 9-଼ turns (fully closed), corresponding to 249 21.5% of tidal volume control range and 2-⅛ turning range control (Fig. 3A) . It allowed control 250 of 85.7% of the total range of delivered tidal volumes (compared to initial variable patient tidal 251 volume) with negligible change in tidal volume delivery to the standard patient (range: 99.86% 252 and 103.2% initial standard patient tidal volume, mean: 102.1% ± 0.98%). We note that the total 253 expiratory volume reported by the ventilator trends with tidal volume delivered to the variable 254 patient (Fig. 3A ) and the peak inspiratory pressures (PIP) of the variable patient and ventilator 255 peak inspiratory volumes also correspondingly decrease with decreases in tidal volume ( Fig. 256 3B,C), while peak inspiratory volumes remain stable for the standard patient and peak end 257 expiratory volumes remain stable for both patients during these changes. 258 We repeated Vent-Lock 1+1 multiplexing patient circuit with the ventilator on volume 259 control mode to deliver a total of 2L of volume, corresponding to approximately 600 mL of tidal 260 volume per patient (additional ventilator settings available in table S1). We note that turning of 261 . CC-BY-NC-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. Two domestic swine were anesthetized and ventilated using the Vent-Lock system with constant 356 volume delivery. Swine were successfully ventilated for approximately four hours using a single 357 ventilator. Initial calculated dynamic lung compliances were 50.3 and 48.1 ml/cm H 2 O for the 358 standard and variable swine, respectively. Throughout the experiment, minimum and maximum 359 dynamic compliance ranged from 50.3 to 244.5 and 37.9 to 87.1 ml/cm H 2 O for the standard and 360 variable swine, respectively, reflecting the differences in tidal volumes those swine received 361 during the flow restriction trial. Serial ABGs were monitored (Fig. 6 ) and initial shared ventilator 362 settings were determined to be too high as both swine developed a respiratory alkalosis. At 363 approximately two hours this was corrected and pH and paCO 2 were allowed to normalize for 364 one hour. Vent settings were not changed following this equilibration. Over the next hour the 365 Vent-Lock system was adjusted from fully open to fully closed, where air was still allowed to 366 pass even when Vent-Lock is closed to prevent unintentional hypoventilation. Respiratory 367 characteristics including tidal volume ratios, percent of total set tidal volume delivered, 368 inspiratory pressure and tidal volume are presented in Fig. 7 . While the tidal volume delivered to 369 the variable swine decreased marginally, a substantial increase in tidal volume was noted to the 370 standard swine (Fig. 7D) , similar to what is seen in simulation center testing with ventilator on 371 volume control mode. Arterial blood gas measurements demonstrated hyperoxia in both swine 372 (Fig. 7C) . A hypercarbic respiratory acidosis occurred in the variable swine ( Figure 6A We used 3D-printing to rapidly prototype components of a ventilator splitter circuit that 396 allows one ventilator or anesthesia gas machine to ventilate two patients. The circuit addresses 397 aforementioned challenges and allows the clinician to control individual patient's pressures and 398 volumes, in a closed system with bacterial filters to reduce viral contamination. 3D printing was 399 selected as the prototyping and production modality due to rapid iterative production for research 400 and development and on-demand manufacturing to meet urgent needs in context of 401 biocompatible and sterilizable 3D printing materials. 3D printed components of the splitter 402 circuit were designed to include ventilator splitters, manometer adaptors for continuous pressure 403 monitoring and a flow restrictor to control tidal volumes and pressures. 404 The novel flow restrictor valve, Vent-Lock FloRest, is able to regulate tidal volumes to 405 one patient while maintaining flow to another patient (Fig. 3) , and control PEEP when placed on 406 the expiratory limb with the manometer adaptor (Fig. 5) . Design, research, and development was 407 driven by creating an airtight valve that allows precise control by the clinician. We demonstrated 408 that FloRest is compatible with multiple modes of ventilation strategies and anesthesia gas 409 machines (Fig. 4D) . Vent-Lock FloRest has demonstrated volume and pressure control by 410 different physicians, using different mechanical ventilators or anesthesia gas machines, thus 411 demonstrating potential compatibility with different ventilators and anesthesia gas machines. 412 While the reproducibility of the Vent-Lock FloRest must be further investigated, this has 413 important global scalability implications as we demonstrate potential compatibility and 414 reproducibility not limited to make or type of ventilator or anesthesia gas machines, which may 415 be more commonly available in developing countries. 416 . CC-BY-NC-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. Our studies reveal fundamental differences in tidal volume patterns with flow restriction 417 with the ventilator in volume versus pressure mode. In pressure control, Vent-Lock FloRest 418 allowed for reduction in delivered tidal volumes to the variable patient with stable volumes 419 delivered to the standard patient (Fig. 3A) . However, in volume control, reduction in delivered 420 tidal volume to the variable patient resulted in a concomitant increase in tidal volume delivery to 421 the standard patient. This pattern is expected due to the continuous delivered volume maintained 422 by the machine; therefore, the Vent-Lock FloRest allowed regulation of the ratio, the 423 standard/variable patient tidal volume ratio (Fig. 4B ). This ratio pattern of control is seen in both 424 ventilator volume control settings in the simulation center and replicated in vivo swine studies 425 using anesthesia gas machines (Fig. 5) . The variation in both patients may be difficult for 426 clinicians to manage simultaneously. However, this ratio-based control delivered by Vent-Lock 427 FloRest can be critical for splitting legacy ventilators or anesthesia gas machines that may only 428 have volume control settings. One of the biggest challenges of splitting patients on ventilators is 429 that air will preferentially travel to the patient with the highest baseline lung compliance 430 resulting in unequal ventilation between the two patients. However, if Vent-Lock FloRest is 431 placed on the patient with the highest baseline lung compliance (variable patient), air flow can be 432 decreased, while simultaneously increasing air flow to the standard patient, until tidal volumes 433 are equilibrated between the two. Consequently, while we demonstrate that flow restriction 434 provides tidal volume control on ventilator pressure or volume control settings through different 435 mechanisms. We emphasize that especially during emergency use settings, providers appreciate 436 these differences in tidal volume control mechanisms, and select the setting most appropriate for 437 patients and practice settings. 438 . CC-BY-NC-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The copyright holder for this this version posted September 22, 2020. . https://doi.org/10.1101/2020.09.16.20195230 doi: medRxiv preprint access to 3D print for instances where manufacturing infrastructure is unavailable to produce this 507 device. Consequently, we believe that open sourced 3D printing methodology of Vent-Lock 508 production is appropriate for scalability in the face of urgent demand, such as the COVID-19 509 pandemic, but can also be transitioned into traditional manufacturing. 510 In this study, we developed Vent-Lock, a ventilator or anesthesia gas machine splitter 511 system with a flow restrictor (FloRest) that can modify flow rates per patient for patient tailored 512 therapies. We provide proof-of-concept that two swine can be safely split using one anesthesia 513 with cross-ventilation restricted by using one-way check valves. Ventilation was maintained with 587 volume control. One swine was not flow-regulated and thus considered the standard patient, 588 while the other had a Vent-Lock 3DP 4.0 connected in the inspiratory limb and thus considered 589 the variable patient. Flow was measured at each expiatory limb with a SS11LB airflow 590 transducer (Biopac; Goleta, CA). Flow data were collected at 2kHz using an MP36 data 591 acquisition unit and BSL 4.1.3 software (Biopac; Goleta, CA). The spirometry data was then 592 smoothed with a 0.25 sec wide moving median filter after removal of instrument noise below 593 0.08 L/sec (determined by histogram inspection). The smoothed data was then numerically the spirometry results were then aligned using common timestamps. All breaths spontaneously 598 initiated by the swine (identified by respiratory rates more than 30% away from the ventilator set 599 point) were removed from analysis. The mean and standard deviation for each anesthesia record 600 entry were calculated for respiratory rate, tidal volume, minute ventilation, and lung compliance. 601 All of the described analysis was performed using a custom MATLAB script (MATLAB 2019b, 602 The MathWorks, Inc., Natick, MA)]. Arterial and venous blood gas data were collected 15 603 minutes following any changes to the Vent-Lock 3DP device. Following the procedure, swine 604 were euthanized with an overdose (~150mg/kg) of supersaturated potassium chloride IV while 605 under anesthesia. Necropsy was performed to assess for any gross lung pathology. Table S1 . The biodurability and sterilization conditions. 775 Table S2 . Ventilator settings on 840 Ventilator System, Nellcor Puritan Bennett. 776 Mov S1. Leaky bubble test demonstrates Vent-Lock FloRest is airtight 777 Data S1. STL print files of Vent-Lock splitter, FloRest, and manometer adaptor 778 Data S2. MATLAB code used to analyze swine data 779 780 . CC-BY-NC-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. CC-BY-NC-ND 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) preprint The copyright holder for this this version posted September 22, 2020. . https://doi.org/10.1101/2020.09.16.20195230 doi: medRxiv preprint The Potential Health Care Costs And Resource Use Associated With COVID-19 In The United States. Health Aff Critical supply shortages -The need for 627 ventilators and personal protective equipment during the Covid-19 pandemic Beware of the second wave of COVID-19 SCCM Rapid Cycle 3 Response Summary Health Care Supply Chains: COVID-19 Challenges and Pressing Actions Multiplex Ventilation: A Simulation-647 based Study of Ventilating Two Patients with One Ventilator Consensus Statement on Multiple Patients Per Ventilator. The Society of Critical Care 650 Medicine Individualized System for Augmenting Ventilator Efficacy (iSAVE)