key: cord-0919064-wmzpoxv8 authors: Hoffman, Henry T.; Miller, Robert M.; Walsh, Jarrett E.; Stegall, Helen R.; Diekema, Daniel J. title: Negative pressure face shield for flexible laryngoscopy in the COVID‐19 era date: 2020-07-29 journal: Laryngoscope Investig Otolaryngol DOI: 10.1002/lio2.437 sha: bbcb0f2de22c397c92f7b6008dd369dc16d60daf doc_id: 919064 cord_uid: wmzpoxv8 OBJECTIVE: Introduce novel methods and materials to limit microdroplet spread when performing transnasal aerosol generating procedures in the COVID‐19 era. METHODS: Prototypes of a negative pressure face shield (NPFS) were tested then used clinically to create a suction‐clearing negative pressure microenvironment with controlled access to the nose and mouth. Air pressure measurements within prototypes were followed by prospective evaluation of 30 consecutive patients treated with the device assessed through questionnaires and monitoring oximetry. RESULTS: The NPFS is a transparent acrylic barrier with two anterior instrumentation ports and a side port to which continuous suction is applied. It is positioned on a stand and employs a disposable antimicrobial wrap to secure an enclosure around the head. This assembly was successfully used to complete transnasal laryngoscopy in all 30 patients studied. Tolerance of the design was excellent, with postprocedure questionnaire identifying no shortness of breath (27/30), no claustrophobia (27/30), no pain (29/30), and no significant changes in pulse oximetry. CONCLUSION: Diagnostic laryngoscopy was successfully performed in a negative pressure microenvironment created to limit dispersion of aerosols. Further application of the NPFS device is targeted for use with transnasal laryngeal laser and biopsy procedures to be followed by additional modification to enable intranasal and intraoral procedures in a similar protected environment. LEVEL OF EVIDENCE: Level 2b (Cohort Study). The coronavirus disease 2019 (COVID-19) pandemic has radically altered the approach to medical care to impact the management of disorders of the upper aerodigestive tract. Risk of viral spread during the aerosol generating procedure (AGP) of transnasal laryngoscopy was addressed through a consensus development webinar involving 300 participants from the American laryngology community on March 24, 2020. 1 This consensus was presented early in the pandemic and recommended screening patients to radically limit use of this procedure. The recommendation at that time concluded that "flexible laryngoscopy should only be performed in critical cases" and "alternatives to laryngoscopy should be considered." Additional recommendations were made for practitioner to use protective gear in a negative pressure or designated isolation room, and to attend to appropriate disinfection of equipment and the examination room. Through evolving knowledge regarding protective measures, restrictions have been eased over the past months. Guidelines for safe return to practice to resume the practice of flexible laryngoscopy were disseminated online by the American Academy of Otolaryngology-Head and Neck Surgery (AAO-HNS) on May 8. 2 These guidelines identify that flexible nasal laryngoscopy "may potentially increase the likelihood of cough, gag, and sneeze, with possible subsequent aerosolization, and therefore appropriate precautions should be considered based on individual clinical circumstances." The AAO-HNS recommendations focused on use of personal protective equipment (PPE) and also discussed concern of the variability in COVID testing-emphasizing a high false negative rate. Screening patients with nasopharyngeal or oropharyngeal swabs to test for SARS-CoV-2 infection with the reverse transcription polymerase chain reaction (RT-PCR) assay has become common practice at some institutions before performing transnasal laryngoscopy but is still limited at others. A recent publication from the Johns Hopkins School for Public Health acknowledges a high false negative rate early in the course of infection and suggest that "care must be taken" when using these results as a basis for removing precautions. 3 In addition to use of PPE, other methods have been improvised to help separate the patient from the health care worker in addressing the concept that all patients should be managed as if they harbor transmissible virus. One approach has been to outfit the patient with a mask modified to permit passage of the flexible scope into the nose. 4, 5 More elaborate barriers have been constructed to shield the health care workers during endotracheal intubation. 6 have not yet been reported to improve the safety of transnasal laryngoscopy. We present development followed by implementation of a negative pressure face shield (NPFS) designed to establish a negative pressure microenvironment about the patient's face during transnasal and transoral procedures. The NPFS is a reusable acrylic apparatus implemented with the common clinical materials of wall suction, disposable tubing, tape, and draping material. The H100 sterilization wrap is a "a disposable infection control product" made of water-repellant breathable "low lint spunbond meltblown spunbond (SMS) fabric" as is also used in production of face masks and lab jackets. 7 An electronic micromanometer (AirdataMultimeter ADM-870C, Shortridge Instruments, Inc, Scottsdale, Arizona) was used to measure air pressure in the NPFS during use. Evaluation of the prototype involved assessment of the vacuum generated within the face shield. The probe was placed through one of the anterior ports with H100 sterilization wrap in place around the shield. Pressure readings were obtained both with the second anterior port opened and closed with tape ( Figure 4 ). Initial testing in human volunteers included assessment of fluctuations in the negative pressure generated within the shield during inspiration and ventilation. Additional pressure testing was done employing a single vs double layer of the sterilization wrap as well as comparison of use of the wall suction on maximum vs portable suction with a HEPA filter. The prepared NPFS with sterilization wrap applied is rotated into position with the patient leaning slightly forward and adjusting the exam chair to a level optimal for the exam. The most optimal of the two anterior access ports is chosen for use during the procedure ( Figure 5A ,B). The unused port is sealed with tape. A twill tape is drawn around the sterilization wrap and the patient to improve the seal and optimize negative pressure within the device ( Figure 5C If there is perceived need to further decrease the risk of aerosol spread, a tighter seal may be formed around the laryngoscope access port. Clear tape is placed over the access port, through which a narrow slit is created. The laryngoscope is then placed through the taped port, forming a tight barrier around the scope Pressure measurements with the model in place were obtained at rest, during inspiration, and during expiration. Further measurements were obtained during sequential modifications to the size and shape of the shield as well as methods to occlude the region around the patient's face (combinations of towels, blankets, and positioning). Depending on the cycle of respiration, employing a single barrier wrap and wall suction at 300 mmHg, the pressure varied between −0.002 (expiration) to −0.020 (inspiration) inches of water. Consistently greater vacuum was achieved with a double layer of the barrier wrap and higher suction achievable (530 mmHg) with the Neptune portable suction system. The duration of the procedure was measured from the initial placement of shield and barrier wrap around the patient to its removal and included the nasal spray, a purposeful delay to allow time for the topical application to work, and then performing the laryngoscopy. This duration ranged from 2 minutes and 10 seconds to 6 minutes and 5 seconds. The longest procedure included interaction with our speech pathologist involved to elicit additional vocalization maneuvers with the flexible laryngoscope in place. Among the 30 patients, 27 reported no shortness of breath during the procedure, with the remainder only reporting "slight" shortness of breath. Twenty-nine reported no pain in the process. The one F I G U R E 2 A stable camera stand is employed to secure the negative pressure face shield in front of the patient's face patient reporting "slight" pain during the procedure attributed the discomfort to manipulation of the flexible laryngoscope and not the face shield. Twenty-seven patients reported no claustrophobia, with the remainder reporting only "slight" claustrophobia. Among the comments from the three patients who reported "slight" shortness of breath was "shortness of breath wasn't a big deal-just a little nervous." Evaluation of scores in each category were significantly less than the compared, arbitrarily agreed-upon, acceptable tolerance score of 2, as determined by the one sample sign test (P < .0001) (Figure 11 ). Additional qualitative assessments about the process were solicited from the patients following the procedure and included the following. "I think this was a very good idea with COVID virus." "I thought it was absolutely ridiculous-because this is much ado about nothing a certain amount of people are going to croak anyway society is just a bunch of lemmings." "I think this is going to be a very good instrument to use in the future"; "anything you can do to prevent the spread of COVID 19 is very good." "Was worried I'd be claustrophobic but with the suction on and a slight breeze through the hood was calming and didn't feel claustrophobic." "I was more worried about being claustrophobic before the procedure than during" and "very professional and fast, in and out in a hurry." "Everything felt fine; didn't feel anxious at all." "The fact the walls of the box is clear makes claustrophobia unlikely." Airflow made it comfortable-better than wearing mask he came in with. F I G U R E 3 The H100 "one step sterilization wrap" is unfolded and taped circumferentially around the back of the negative pressure face shield (NPFS) (black arrows) and then pulled forward to permit the patient to place their face in the device (white arrows) F I G U R E 4 Testing negative pressure with full wall suction applied (here shown with second port opened). An average of 18 individual readings was −0.0131 in. of water F I G U R E 5 The negative pressure face shield with sterilization wrap is rotated in place and secured around the neck F I G U R E 6 Instillation of nasal decongestant with anesthesia through anterior access port. Note the full protective equipment as depicted in examining this patient early in the study was modified with further experience to employ only a face shield, mask, and gloves without perceived need for gown or N95 Mentioned the scope in nose was irritating but process not painful. Assembly of instrumentation and preparation of the face shield by assistants before the patient enters the room is facilitated by use of a checklist (see Supporting Information). It is helpful to optimize patient positioning prior to starting the procedure to minimize need for repositioning during the procedure. Positioning is initiated by adjusting the height of the exam chair, the angle of the stand, and the posture of the patient with the face shield in front of the patient before placing the antimicrobial wrap over the head. The full suction as applied to the NPFS is maintained through the end of the procedure to perform a vacuuming maneuver as the assembly is rotated from the patient's face with the blue draping material collapsing into the face shield under suction. F I G U R E 8 The end of a cotton tipped applicator is used to create a small opening providing a tight fit through which the flexible laryngoscope is placed. Note the procedure may be done with the patient on oxygen as this man required 24 hours/day at 3 L per minute F I G U R E 9 Successful experience in use of the negative pressure face shield permitted safely performing the flexible transnasal laryngoscopy without addition of gown or N95 F I G U R E 1 0 Pre-, intra-, and postprocedure pulse oximetry measurements identified saturation SpO 2 consistently above 96%. NPFS, negative pressure face shield F I G U R E 1 1 Postprocedure assessment by the patients identified no more than "slight" ratings for claustrophobia, shortness of breath and pain during the procedure. NPFS, negative pressure face shield A high level of patient tolerance was identified in this effort to limit dispersion of bioaerosols created during transnasal laryngoscopy. As a feasibility study, these findings offer support for an approach that-as is true for many infection control strategies-is guided more by basic principles and common sense than evidence-based medicine. 10 Other barriers to aerosolization have been developed to mitigate dispersion of aerosol during the COVID-19 era and include the "aerosol box" or intubation box designed to provide protection to anesthesiologists during direct laryngoscopy. [11] [12] [13] Canelli et al reported trial of a transparent plastic cube placed over a mannequin's face with "crude simulation of a cough" found to be help contain the spread fluorescein droplets in a manner felt to provide "a modicum of additional protection" as "an adjunct to standard PPE." 6 Other investigators identified shortcomings to this device including that it "introduces another contaminated device that must be properly handled during use and disinfected between use to prevent cross contamination." 14 The NPFS addresses this concern by its simple design with smooth surfaces that permit rapid disinfection employing antimicrobial wipes followed by air-drying in a 3-minute period before reuse. Additional concerns about enclosures restricting access to the patient 15 during intubation were addressed by the original investigators who reiterated in their reply that "operators should be ready to abandon use of the box should airway management prove difficult." 16 The NPFS is deployed in a manner that permits it to be immediately rotated away from the patient. However, the excellent patient tolerance identified in the course of our study makes this process less of a concern. The limited access to the nose and oral cavity through use of the NPFS was not an issue for diagnostic imaging but is problematic if other manipulations such as nasal packing or cautery are needed. Modifications to the NPFS, including the strength of suction, the positioning of the suction, as well as the size, shape and covering of the aperture are under evaluation to address these issues. Management options other than placing a barrier around the patient have been proposed to limit viral spread. These options include pretreatment testing of patients for COVID-19. Greater access to COVID-19 testing has permitted us to apply this option to patients who are not candidates for use of the NPFS-including those with tracheostomies and those requiring transoral procedures. Current accessibility to testing remains limited and is still unavailable for use prior to routine laryngoscopy at our institution (as of June 5, 2020). Shortcomings to testing all patients include the need for patients to make a separate trip for testing before the laryngoscopy to allow time for processing. Additionally, concerns about reliability of testing have been raised and have identified false negative results to occur. Other strategies to decrease viral spread include use of topical preparations to reduce the viral load in the upper aerodigestive tract. Parhar et al advocate application of a dilute povidone-iodine solution to the nasal and oral cavities prior to instrumentation of the upper airway as safe and potentially helpful in decreasing the viral load. 17 Studies addressing influenza have identified that both infectivity and the severity of an infection are dependent on the viral load to which a patient is exposed. 18, 19 While this dose-dependence has not been specifically identified for SARS-CoV-2, eliminating or minimizing the inoculum of exposure is a reasonable goal for any aerosolized virus that has the potential for infection. Although there is currently no single strategy available to completely eradicate the risk of coronavirus transmission, methods to decrease dissemination such as the NPFS are likely to be beneficial. As identified by Rameau et al "Flexible laryngoscopy in the age of COVID-19 requires adaptation." 1 The NPFS provides an adaptation designed to decrease dissemination of virus and has been successfully used clinically to examine the larynx. It is under active study for use in transnasal surgical procedures including laser treatments and biopsies. Shortcomings to this study of the NPFS include the lack of a comparison of its efficacy in diminishing aerosol distribution when compared to other methods. Sophisticated techniques to identify bioaerosol dissemination are available and have been used by a member of the research team to assess generation of bioaerosols with toilet flushing in the hospital setting. 20 Evaluation is planned to employ these techniques to quantify and compare the degree to which the bioaerosols dispersion is mitigated with the NPFS. A new device is presented with accompanying methods and materials designed to limit exposure to aerosols and to enhance the safety of transnasal AGPs. We have incorporated the NPFS into routine clinical practice due to its ease of use, rapid disinfection, and high level of patient tolerance. The NPFS has the potential to extend its application beyond diagnostics to include procedures at even higher risk for generating contagious aerosols, including transnasal laser and biopsy procedures. Flexible Laryngoscopy and COVID-19 COVID-19 RESOURCES Guidance for return to practice for Otolaryngology-Head and Neck Surgery Variation in False-Negative Rate of Reverse Transcriptase Polymerase Chain Reaction-Based SARS-CoV-2 Tests by Time Since Exposure Trans-nasal laryngoscopy in COVID-19 era-method for additional protection with disposable mask Application of a Modified Endoscopy Face Mask for Flexible Laryngoscopy During the COVID-19 Pandemic Barrier enclosure during endotracheal intubation 510(k) Summary for the Kimberly-Clark Corporation Kimguard Sterilization Wrap Research electronic data capture (REDCap)-a metadata-driven methodology and workflow process for providing translational research informatics support The REDCap consortium: building an international community of software partners Infection prevention strategies for procedures performed outside operating rooms: a conceptual integrated model Airborne Aerosol Generation During Endonasal Procedures in the Era of COVID-19: Risks and Recommendations Endonasal instrumentation and aerosolization risk in the era of COVID-19: simulation, literature review, and proposed mitigation strategies Application of the aerosol box in protecting healthcare workers during the COVID-19 pandemic More on barrier enclosure during endotracheal intubation More on barrier enclosure during endotracheal intubation. Reply Topical preparations to reduce SARS-CoV-2 aerosolization in head and neck mucosal surgery A dose-finding study of a wild-type influenza a(H3N2) virus in a healthy volunteer human challenge model Influenza infectious dose may explain the high mortality of the second and third wave of 1918-1919 influenza pandemic Bioaerosol concentrations generated from toilet flushing in a hospital-based patient care setting Negative pressure face shield for flexible laryngoscopy in the COVID-19 era