key: cord-0917797-3ejeoirs authors: Zamzami, Mazin A.; Rabbani, Gulam; Ahmad, Abrar; Basalah, Ahmad A.; Al-Sabban, Wesam H.; Nate Ahn, Saeyoung; Choudhry, Hani title: Carbon nanotube field-effect transistor (CNT-FET)-based biosensor for rapid detection of SARS-CoV-2 (COVID-19) surface spike protein S1 date: 2021-10-15 journal: Bioelectrochemistry DOI: 10.1016/j.bioelechem.2021.107982 sha: 8cdfe518217019cadb2cb0cf9573f87812f275bf doc_id: 917797 cord_uid: 3ejeoirs The large-scale diagnosis of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) is important for traceability and treatment during pandemic outbreaks. We developed a fast (2-3 min), easy-to-use, low-cost, and quantitative electrochemical biosensor based on carbon nanotube field-effect transistor (CNT-FET) that allows digital detection of the SARS-CoV-2 S1 in fortifited saliva samples for quick and accurate detection of SARS-CoV-2 S1 antigens. The biosensor was developed on a Si/SiO(2) surface by CNT printing with the immobilization of a anti-SARS-CoV-2 S1. SARS-CoV-2 S1 antibody was immobilized on the CNT surface between the S-D channel area using a linker 1-pyrenebutanoic acid succinimidyl ester (PBASE) through non-covalent interaction. A commercial SARS-CoV-2 S1 antigen was used to characterize the electrical output of the CNT-FET biosensor. The SARS-CoV-2 S1 antigen in the 10 mM AA buffer pH 6.0 was effectively detected by the CNT-FET biosensor at concentrations from 0.1 fg/mL to 5.0 pg/mL. The limit of detection (LOD) of the developed CNT-FET biosensor was 4.12 fg/mL. The selectivity test was performed by using target SARS-CoV-2 S1 and non-target SARS-CoV-1 S1 and MERS-CoV S1 antigens in the 10 mM AA buffer pH 6.0. The biosensor showed high selectivity (no response to SARS-CoV-1 S1 or MERS-CoV S1 antigen) with SARS-CoV-2 S1 antigen detection in the 10 mM AA buffer pH 6.0. The biosensor is highly sensitive, saves time, and could be a helpful platform for rapid detection of SARS-CoV-2 S1 antigen from the patients saliva. The World Health Organization (WHO) announced on 11 March 2020 that the SARS-CoV-2 epidemic was a pandemic and a global health emergency [1] . It is well known that there have been three human pathogenic coronavirus outbreaks in the 21 st century that triggered transmission and posed immense challenges to global public health and economic growth [2] . The three virus outbreaks were SARS-CoV in 2003 [3] , MERS-CoV in 2012 [4] , and the new SARS-CoV-2 in December 2019 [5] . Phylogenetic data analysis suggests that SARS-CoV-2 is further away from the SARS-CoV (79%) and MERS-CoV (50%) strains [6] . SARS-CoV-2 has a large positive-stranded RNA genome size of 29.8 kb that gives signals for encoding accessory and four structural proteins: spike glycoprotein (S), envelope (E), matrix (M), and nucleocapsid (NC) [7] . Among those, only spike (S) glycoprotein faces the outer structure on the virion body, which is highly immunogenic for the host and the best reliable biomarker for the diagnosis of SARS-CoV-2 infection [8] [9] [10] . S-protein monomer is folded and glycosylated by sialic acid at the distal end [11] . S-protein is cleaved by host proteases into two subunits, S1 and S2; the S1 subunit functions in host cell recognition, cell invasion, and antibody neutralization [12] . SARS-CoV-2 infection is mediated by the binding of the S1 subunit of spike (S) protein to the host cell receptor through the receptor-binding domain (RBD) present on the angiotensin-converting enzyme 2 (ACE2) [13, 14] . SARS-CoV-2 inhaled through aerosol droplets binds to host cellular receptor ACE2 and targets activating proteases (for example, TMPRSS2, furin, and cathepsin L) for membrane fusion and reaches the host respiratory cells [15, 16] . The virus easily spreads by touch, droplets, air contamination, the oral-fecal route, or blood, and from mothers to newborns and animals to humans [17] [18] [19] . As a result the world increasingly needs diagnostic tools with the capability of identifying the infected individuals e.g. the source of infection. Electrochemical biosensors are advantageous for sensing biomolecules because of their ability to detect biomarkers with accuracy, specificity and high sensitivity [20] . Previously electrochemical biosensors have been successfully used in medical diagnostics for the detection of viruses such as MERS-CoV [21] , hepatitis C virus (HCV) [22] , human influenza A virus H9N2 [23] and avian influenza virus (AIV) H5N1 [24] . Recently, Shao et al. developed a SARS-CoV-2 S antigen detection device based on a carbon nanotube field-effect transistor (CNT-FET) functionalized with anti-SARS-CoV-2 spike antibody with a detection range of 5.5 fg/mL to 5.5 pg/mL [25] . The real-time polymerase chain reaction (RT-PCR) is rigorously used for the diagnosis of SARS-CoV-2. RT-PCR is excellent in terms of sensitivity and selectivity, but can only be performed in well-equipped laboratory-based hospitals, requires costly chemicals and enzymes and spacious machinery, and requires up to 1-2 days. In order to detect SARS-CoV-2 quickly with high accuracy, a simple, sensitive, and low-cost biosensor is needed. This research paper describes the development of a new CNT-FET-based biosensor for reliable, fast detection of SARS-CoV-2 S1 antigen with high selectivity and sensitivity. Here, we present a highly sensitive electrical immunosensor for the detection of SARS-CoV-2 S1 antigen using single wall carbon nanotubes (SWCNTs) deposited on the surface of SiO 2 between the S-D channel. The developed biosensor is based on the field-effect transistor (FET) technology, and its fabrication involves the immobilization of antibodies between the S-D sensing area. The newly developed CNT-FET immunosensor can detect SARS-CoV-2 S1 antigen, based on the change in source-drain current (I DS ), and the antigen-binding effect is characterized by a change in electrical signal (Scheme 1). The basic advantages of the FET-based biosensor are easy fabrication, fast sensing response, and ease of use because the dynamic real-time response can be acquired with low-cost digital readers that can be calibrated for different applications. Certainly, at central hospitals and testing centers, advanced and traditional equipment can analyze and diagnose SARS-CoV-2 infection with high accuracy and precision. In pandemic circumstances, CNT-FET-based biosensors could be helpful as potential detection devices for tracing SARS-CoV-2 infection at the mass level. Single-wall CNTs with a length of 5 to 30 µm and diameter of 1 to 2 nm were purchased from NanoIntegris, Canada. SARS-CoV-2 spike/S1 subunit-His tag (Cat# 40591-V08H), SARS-CoV-1 spike/S1 subunit-His tag (Cat# 40150-V08B1) and MERS-CoV spike/S1 subunit His tag recombinant protein (Cat# 40069-V08H) were purchased from Sino Biological Inc., Beijing, China. Monoclonal rabbit anti-spike S1 antibody (40150-R007) was purchased from Sino Biological Inc, Beijing, China (antibody specificity was not validated with corresponding SARS-CoV-2 real positive virus saliva/serum sample). HRP-conjugated secondary antibody goat pAb to rabbit IgG was purchased from Abcam (ab6721). Source and drain (S-D) electrodes of Ti 10 nm/Au 30 nm were patterned by a conventional photolithography method and lift-off technique. After leaving it for 1 h, the Si wafer was soaked in dichlorobenzene, acetone, and methanol for 30 s for each co-solvent, and finally, DI water was used for rinsing and nitrogen (N 2 ) gas blowdown. A uniform CNT ink (separated semiconducting SWNTs at 0.1 mg/mL in dichlorobenzene) was printed on Si/SiO 2 substrate by inkjet printer between the active areas of patterned electrodes. To fix the printed CNT pattern on the base material, it was hard baked on a hot plate at 120 °C in the vacuum oven for 10 min. To avoid direct contact with the S-D electrode metal surface and minimize the gate current leakage due to flowing antigen test solution an additional photolithography step was employed to make SU-8 passivation layer on the S-D electrode. In order to investigate the semiconductor electrical properties and sensing efficiency of the CNT-FET device, measurements were performed by a Keithley 3 probe station at room temperature. The semiconductor property of the CNT-FET biosensor was validated by acquiring the source-drain current (I DS ) versus gate voltage (Vg, -1.5 to 0.5 V) with a stable source-drain voltage (Vds = 0.5 V). X-ray photoelectron spectroscopy (XPS) spectra were acquired on a K-Alpha spectrometer (Thermo Fisher Scientific, USA). The machine is equipped with a micro-focused AlKα X-ray source (1486.6 eV) using a 10 mA emission current and an applied anode potential of 12 kV (120 W). The spectra were collected in constant analyzer energy mode with a pass energy of 200 eV for the survey. For reliable digital acquisition and data processing, Thermo Scientific Advantage software was used. Final CNT fabricated FET chips were soaked in 2 mM PBASE solution prepared in methanol for 1 h at room temperature. Subsequently, the sensing area of the CNT-FET chip was successively washed 3 times with pure methanol and DI water to remove the unbound excess PBASE (linkers) followed by N 2 gas blowdown. For functionalization, the monoclonal anti-SARS-CoV-2 S1 (1 µL of 100 µg/mL) in 0.01× PBS solution was dropped on the channel area and left overnight at 4 °C in a humid chamber for binding with the PBASE-modified CNT surface. The unbound anti-SARS-CoV-2 S1 remains were removed by washing with 0.01× PBST and DI water. The device was further exposed to 1% bovine serum albumin (BSA) at room temperature in humid conditions for 20 min to block and inactivate nonspecific sites of the remaining nonfunctionalized CNTs and exposed SiO 2 substrate to ensure the binding of the specific target molecule. To eliminate excess BSA remaining after blocking, the device was thoroughly washed with 0.01× PBST and DI water and dried by N 2 gas blowdown for 5 minutes. After complete drying, the CNT-FET device was connected to source, drain, and gate electrodes with a probe station for measurement of its post-treatment semiconductor behavior. For SARS-CoV-2 S1 antigen (50 µg/mL) was slowly added to the colloidal AuNPs suspension drop by drop with gentle stirring on a stir plate. Then, the AuNPs-SARS-CoV-2 S1 antigen mixture was allowed to react for 60 min at room temperature with gentle stirring. For blocking, 50 µL of freshly prepared 10% BSA in DI water was added in colloidal AuNPs-SARS-CoV-2 S1 suspension and vortexed for 60 min at room temperature on a stir plate.After further incubation, the colloidal suspension was spun at 9660 × g for 30 min at 4 °C. The obtained supernatant was discarded and only the pelleted AuNPs-conjugated SARS-CoV-2 S1 antigen was resuspended and recovered by mixing with 2 mM borate buffer (pH 8.0, total volume was 100 µL). It was estimated that the final AuNPs-conjugated SARS-CoV-2 S1 antigen carried 6.37×10 11 particles/mL (supplementary information figure S1: by absorbance measurement at 520 nm). The AuNPs-conjugated SARS-CoV-2 S1 antigen was stored for further experimental use at 4 °C in an air-tight vial to prevent evaporation. CNT-FET chips with good S-D current versus gate voltage (I DS -V GS ) signal properties were selected for SEM imaging experiments. Step 1: The upper side of the CNT-FET chips (containing S-D and conducting CNT channel) was soaked in acetone solution for 10 min and air-dried at room temperature. For rigorous surface contamination removal, the acetone washing step was repeated 3 times. Step 2: Acetone-washed CNT-FET chips were soaked in 2 mM PBASE (linker) and left for 1 h at room temperature in air tight glass container. Step 3: After 1 h, CNT-FET chips were removed from the PBASE solution and washed 3 times with methanol and DI water followed by N 2 gas blowdown. Step 4: 1 µL of anti-SARS-CoV-2 S1 (100 µg/mL) in 0.01× PBS solution was immobilized on the PBASE-modified CNT-FET biosensor and incubated for 4 h at room temperature in a humid chamber. Step 5: Washing with 0.01× PBST was performed 5 times to remove unbound anti-SARS-CoV-2 S1 followed by air drying at room temperature. Step 6: To block and inactivate nonspecific sites, CNT-FET devices were further exposed to 1% BSA and incubated at room temperature for 20 min in a humid chamber. Step 7: Washing with 0.01× PBST was performed 5 times to remove unbound BSA traces, followed by air drying at room temperature. Step 8: 1 µL SARS-CoV-2 S1 antigen with conjugated AuNPs (diluted in 10 mM AA buffer, pH 6.0) was immobilized and incubated for 1 h at room temperature in a humid chamber. Step 9: Three times washing with 0.01× PBST and DI water was performed to remove unbound AuNPs-conjugated SARS-CoV-2 S1 antigens, followed by a N 2 gas blow dry at room temperature. Step 10: The morphology and microstructure of CNT were characterized by field emission scanning electron microscope JEOL JSM-7610F (JEOL Ltd, Akishima, Japan). Western blot experiment was utilized to examine the specificity of anti-SARS-CoV-2 S1 and antigen binding. SARS-CoV-2 S1, SARS-CoV-1 S1, and MERS-CoV S1 with an antigen concentration of 1 µg/mL (lanes 1, 2, and 3) were separated under reducing conditions of 12% polyacrylamide gel. After electrophoresis, protein bands were transferred from the SDS-PAGE gel to the PVDF membrane. The membrane was blocked in 1× phosphate-buffered saline containing 0.05% Tween-20 (PBST) and 5% (w/v) skim milk and left overnight at 4 °C. After 3 washes with 1× PBST, PVDF membrane was stained with SARS-CoV-2 (2019-nCoV) spike protein S1 primary antibodies at 1:5000 dilution (200 ng/mL) for 2 h in 5% skim milk solution in 1× PBST. After 3 washes with 1× PBST, the membrane was further incubated in 5% skim milk containing secondary antibody goat pAb to rabbit IgG conjugated to HRP (1:5000 dilution) for 60 min at room temperature on a rotating shaker plate. The membrane was washed 5 times with 1× PBST to remove unbound remains. ECL solution was mixed in a 1:1 ratio following the proportions of solutions A and B provided by the manufacturer. ECL is a chemiluminescent substrate that detects horseradish peroxidase (HRP) conjugated secondary antibodies where light emission is proportional to the protein concentration. ECL consists of 2 reagents, reagent A (luminol) and reagent B (an enhancer). The PVDF membrane was placed on plastic film, freshly mixed ECL reagents A and B (1 mL each, total volume 2 mL) at a 1:1 ratio were added, and the membrane was incubated up to 3 min to complete the reaction. The PVDF membrane was lifted from the plastic film and transferred to the membrane cassette. The signal was exposed in the darkroom with an exposure time of 40 s, and the X-ray film was developed by immersing in the appropriate developing and fixing solutions for 5 min each. SARS-CoV-2 S1, SARS-CoV-1 S1 and MERS-CoV S1 antigen stocks were prepared in 10 mM ammonium acetate (AA) buffer at pH 6.0 and, to prevent denaturation, stored at 4 °C. The SARS-CoV-2 S1 antigen was diluted in 10 mM AA buffer at pH 6.0 with increasing concentrations (0.1, 1, 10, 100, and 5000 fg/mL). The CNT-FET device was connected to 3 probe station, and 0.8 µL of 10 mM AA buffer at pH 6.0 was dropped to equilibrate the I DS and achieve the stable baseline. After achieving a stable baseline of the biosensor, increasing concentrations of SARS-CoV-2 S1 antigen were sequentially dropped onto the S-D channel regions. The sample volume was 0.8 µL in each drop. A selectivity test experiment was performed to determine the biosensor's ability to select target molecules. Target SARS-CoV-2 S1 antigen and non-target SARS-CoV-1 S1 and MERS-CoV S1 antigens were tested using the anti-SARS-CoV-2 S1 immobilized on CNT-FET biosensor. For this experiment, 1 μL SARS-CoV-2 S1 antigen was added to 99 μL healthy human saliva sample to mimic a SARS-CoV-2-positive saliva sample and incubated for 30 min at 37 °C. In the fortified human saliva final concentration of SARS-CoV-2 S1 antigen was (0.1, 1, 10, 100, 1000, 2000, 3000, 4000 and 5000 fg/mL). These fortified saliva solution used as a test sample instead of a clinical SARS-CoV-2 positive sample. Then 1.0 µL SARS-CoV-2positive simulated saliva sample was added to the anti-SARS-CoV-2 S1 functionalized CNT-FET biosensor and the electrical signal was read up to 60-120 s. Reproducibility tests of each biosensor were characterized by the various parameters for practical applications. For this, three batches (1, 2, and 3 batch) of fabricated CNT-FET biosensors were taken, each batch was replicated with saliva and fortified saliva and a total of 24 chips were selected and semiconductor property were validated by I-V measurements. The reproducibility was determined by the SARS-CoV-2 S1 fortified saliva and only saliva dropping experiment. The used concentration of SARS-CoV-2 S1 antigen was 100 fg/mL in 10 mM AA buffer of pH 6.0. For the reusability experiment, different chips from the new batch were used after the antigen test was completed. The tested SARS-CoV-2 antigen and immobilized anti-SARS-CoV-2 S1 were removed from the biosensor sensing area by 5 times 0.01× PBST washing. To fix the fabrication structure on the base material, it was hard baked on a hot plate at 120 °C in the vacuum oven for 10 min. After it was completely dry, S-D current versus gate voltage (I DS -V GS ) measurement was again performed to see the semiconductor properties either retained or destroyed in the washing process. Further on the dried CNT-FET biosensor PBASE linker and anti-SARS-CoV-2 S1 were immobilized (the same steps are repeated as described in section 3.4). SARS-CoV-2 S1, SARS-CoV-1 S1 and MERS-CoV S1 antigen test measurements were again conducted S-D current versus time (I DS -T) and acquired data of the CNT-FET biosensor was compared with the used and reused CNT-FET biosensor. The basic structure of the field-effect transistor (FET) is shown in Figure 1A . The complete fabricated CNT-FET device after a single die is shown in Figure 1B . Optical microscopy (OM) was used to identify the locations of nanotubes bridging source-drain (S-D) electrode pairs in order to determine the fabrication process ( Figure 1C ). Inkjet printing was used to deposite CNTs, random network on the SiO 2 surface between the S-D of the device. The schematic structure of the FET circuit diagram along with the CNT semiconductive network situated between S-D electrodes on the substrate is shown in Figure 1D . The deposited CNT network characterized by SEM imaging is shown in Figure 1E . The basic mechanism of PBASE functionalization with CNT surface is shown in Figure 1F . To confirm the CNT-FET biosensor surface modification and the presence of PBASE on the surface of the device, X-ray photoelectron spectroscopy (XPS) was employed. The peak height of XPS difference spectra can be used to confirm the absence or presence of functionalized PBASE on the CNT surface by analyzing the N1s peak at 400.2 eV. As seen in Figure 1G , the bare surface did not show any peak increase at 400.2 eV, and after PBASE modification of the device, the N1s peak appeared at 400.2 eV. The binding energy values after N substitution were assigned in the range of 400.2 to 401.8 eV. These XPS results clearly suggest successful functionalization since the only source of nitrogen atoms in functionalized CNT is the PBASE molecule. The PBASE was immobilized on the surface of the carbon nanostructure by π-π non-covalent stacking between the six-membered ring of the CNT sidewall and the pyrene group of PBASE [26] [27] [28] , while the succinimidyl ester group of the PBASE facing upward on the surface of CNT can react with an amine group of the target antibody and forms an amide bond [29] . The morphology of CNT from scanning electron microscopy (SEM) is shown in Figure 2 . Bare SiO 2 shows a clean surface (Figure 2A ) surface of anti-SARS-CoV-2 S1 immobilized area, AuNPs-SARS-CoV-2 S1 conjugated antigen solution was applied, and AuNPs-conjugated SARS-CoV-2 S1 antigen was bound with the immobilized anti-SARS-CoV-2 S1 between the S-D channel area. Figure 2C shows the sphere-like structure that proves the presence of AuNPs-conjugated SARS-CoV-2 S1 antigen. SEM images in Figure 2D show the AuNP sizes in the range of 19.3 to 23 nm. The increased size of AuNPs was due to the binding of the highly glycosylated SARS-CoV-2 S1 antigen. These findings suggest that the SARS-CoV-2 S1 antibody was successfully immobilized on the biosensing area, and the formation of the antibody-antigen complex localized by AuNPs is illustrated in Figure 2C . To confirm the specificity of anti-SARS-CoV-2 S1 with target SARS-CoV-2 S1 antigen, a Western blotting experiment was performed. Western blot data of SARS-CoV-2 S1, SARS-CoV-1 S1 and MERS-CoV S1 antigens at a concentration of 1 µg/mL (lanes 1, 2, and 3 ) are shown in Figure 3 . A single distinct band recognizes the 77 kDa SARS-CoV-2 surface spike glycoprotein S1 antigen (lane 1). The increased molecular size was observed due to the high glycosylation of the S1 protein. The nonspecific SARS-CoV-1 S1 and MERS-CoV S1 antigens did not show binding with the anti-SARS-CoV-2 S1(lanes 2 and 3). These results indicate that the used anti-SARS-CoV-2 S1 was specific for the SARS-CoV-2 S1antigen. Before any treatment, the transfer property of the CNT-FET device was characterized as the best fit for a p-type CNT-FET device, acting as a conductor with semiconductor properties SARS-CoV-2 S1 (1 μL of 100 μg/mL) antibody diluted in 0.01× PBS was then exposed to the PBASE-modified CNT surface. After anti-SARS-CoV-2 S1 immobilization, surface was washed and dried in N 2 gas blow and the I DS of the CNT-FET biosensor was measured, and depleted drain current indicated that the antibody was successfully immobilized on the biosensor surface ( Fig. 4A light blue line). Antibody immobilization leads to a 10-50% decrease in ON-state current because of enhanced electron dispersion from crowded and more complex chemical environments along the CNT-FET source-drain channel area [32] . In each of these steps, the mobility of the CNT-FET device generally decreases by a small amount, as there are additional sources of dispersion of the charge carrier [33] . Garcia-Aljaro et al. reported that the decrease in I DS of the biosensor, that is, the increase in resistance, is due to the accumulation of negative charge from the immobilized antibodies on the biosensor [34] . Based on the same modification and attachment chemistry, changes in I DS versus V GS induced by the functionalization process were consistent with previously reported findings [35] . The mechanism of source-drain current (I DS ) decrease in carbon nanotube transistors may arise from the electrostatic gating effect and the Schottky barrier effect [36] . On the dry surface of the CNT-FET biosensor 1 µL of SARS CoV-2 S1 antigen was droped on the immobilized anti-SARS CoV-2 S1 and I DS of the CNT-FET biosensor was measured. As shown in figure 4A (green line), depleted drain current indicated that the SARS CoV-2 S1 antigen solution was strongly binds with immobilized anti-SARS CoV-2 S1 on the biosensor surface. The CNT-FET biosensor response time, limit of detection, and detection range were characterized by loading of target antigen (SARS-CoV-2 S1) at serially increased concentrations. The anti-SARS-CoV-2 S1 immobilized CNT-FET biosensor was used in systematic electrical signal measurements. First, at 10 sec, 0.8 µL of 10 mM (AA) buffer solution at pH 6.0 was dropped on the sensor to equilibrate the source-drain current (I DS ). After achieving a stabile I DS signal, continuously increasing SARS-CoV-2 S1 antigen concentrations (0.8 µL each drop) were loaded. The real-time response with increasing concentration of SARS-CoV-2 S1 antigen added every 50 s was recorded after achieving saturation of the observed I DS . Each concentration of SARS-CoV-2 S1 antigen generated a characteristic spike and saturation, seen in all tested antigen concentrations ( Figure 4B ). A calibration curve was plotted as a function of SARS-CoV-2 S1 antigen concentration against normalized sensor response (∆I/I o ), shown in Figure 4C . These results indicate that the CNT-FET biosensor detection lies in the range of 0.1 to 5000 fg/mL. The calibration curve indicates that the sensing response starts becoming saturated from 100 fg/mL of SARS-CoV-2 S1 antigen concentration; however, little increase in I DS was observed above a 100 fg/mL concentration. SARS-CoV-2 spike S1 protein using monoclonal antibody immobilized, which had a LOD of 1 fg/mL [37] . The variation in LOD values of graphene and CNT-FET was due to the varying amount of semiconducting graphene and CNT in the device fabrication. Table 1 presents the data on SARS-CoV-2 detection by various biosensors reported in recent years. The summarized parameters collected from the relevant publications as shown in Table 1 include the analyzed biomarker target, LOD, measurement range, time required for final result generation, and compared with the results of this study on detection of SARS-CoV-2 S1. Our biosensor is highly sensitive comparing with the other reported biosensors. The sensitivity and time response of the biosensor varies with the number of immobilized detecting antibodies on the semiconducting CNT layer [38] . The dimension and width of the S-D may be other parameters that are responsible for the sensitivity of the biosensor [39] . The SARS-CoV-2 S1 antigen is positively charged (pI = 8.27) [40] , and the SARS-CoV-2 S antigen is negatively charged (pI = 6.24) at neutral pH [25] . Table 2 summarizes the physicochemical properties of the SARS-CoV-2, SARS-CoV-1, and MERS-CoV spike proteins. The SARS-CoV-2 S1, SARS-CoV-1 S1, and MERS-CoV S1 antigens carry a net charge of +5, -8, and -7, respectively, at pH 7.0 ( Table 2) . Based on these data, it was concluded that the SARS-CoV-2 S1 antigen carries more positive charge than the other two antigens at neutral pH. The SARS-CoV-2 S1 antigen was prepared in 10 mM AA buffer at pH 6.0. At this acidic pH, the antigen was more protonated and carried a net charge of +13. This increased positive charge (+13) on the antigen strongly binds with the SARS-CoV-2 S1 antibody and generates a strong I DS response. Practically, a biosensor should be not only sensitive but also specific to a specific antigen. Selectivity is an important parameter in immunosensor applications because of the complex composition of biomolecules. Therefore, to confirm the selectivity of the CNT-FET biosensor, a control set of specific (SARS-CoV-2 S1) and non-specific (SARS-CoV-1 S1 and MERS-CoV S1) antigens were tested under identical experimental conditions. Figure 5A shows the normalized biosensor response of the tested antigens prepared in 10 mM AA buffer at pH 6.0. First, after running the test, at time 10 s, the 10 mM AA buffer was dropped to stabilize the I DS of the CNT-FET biosensor. After I DS stabilization up to 50 s, the MERS-CoV S1 antigen was loaded on the biosensor. After saturation of the I DS signal, the SARS-CoV-1 S1 antigen was again dropped ( Figure 5A ). These negative control experiments provide strong evidence that the no signal response of the CNT-FET biosensor reflects a highly selective response against non-specific antigens that reavels a non-specific binding. Immobilized anti-SARS-CoV-2 S1 recognized the target antigen, not the two tested nonspecific antigens ( Figure 5B ). SARS-CoV-2 S1 antibodies binding with specific antigens produced far greater changes in the I DS signal than the non-specific antigens. For particular antigen-antibody pairs, the CNT surface showed high electron transport properties compared to the non-specific pairs. These findings show a reduced normalized biosensor response by 8 to 9.5% compared with specific antigen-antibody pairs. Saliva samples with and without SARS-CoV-2 S1 antigen were tested. There was no significant change in I DS after dropping of non-fortified saliva on the biosensor ( Figure 5C ). Saliva without SARS-CoV-2 S1 antigen showed a negligible response and did not show binding with immobilized anti-SARS-CoV-2 S1. However, saliva with SARS-CoV-2 S1 antigen showed a strong binding response with immobilized anti-SARS-CoV-2 S1. The concentration-dependent SARS-CoV-2 S1 fortified saliva was tested to see the biosensor response. As shown in figure 5D the sensor response was dependent on the SARS-CoV-2 S1 concentration present in the fortified saliva. The sensor response was almost saturated at 4000-5000 fg/mL SARS-CoV-2 S1 fortified saliva. These results indicate that the sensor response data is very similar as discussed in section 4.6. Furthermore, the reproducibility tests of each biosensor are needed to characterize the parameters for practical applications. Because reproducibility is one of the key factors that affecting the practical application of electrochemical immunosensors. Therefore, the reproducibility of fabricated CNT-FET biosensors from three batches (1, 2, and 3) was tested under the identical experimental condition by the SARS-CoV-2 S1 antigen dropping experiment ( Figure 6A ). The obtained results demonstrate that all tested CNT-FET biosensors fabricated in all three batches exhibited very similar SARS-CoV-2 S1 antigen responses. The relative standard deviation (RSD) was estimated to be 2.65% based on the generated electrical signal of the twelve replicates of SARS-CoV-2 S1 fortified saliva ( Figure 6A ). Thus, these results indicate that fabricated CNT-FET biosensors have stable and good reproducibility for antigen recognition. For the reusability experiment of CNT-FET-biosensor, SARS-CoV-2 S1 antigen tested biosensor was washed 5 times with 0.01× PBST to clean the bound SARS-CoV-2 S1 antigen with anti-SARS-CoV-2 S1 and dried with N 2 gas blow. The MERS-CoV S1, SARS-CoV-1 S1 and SARS-CoV-2 S1 antigens were tested and the observed sensor response showing that after one use still fabricated material was stable and not damaged in the washing process. The normalized sensor response drop of SARS-CoV-2 S1 was 0.068, however in the reused biosensor the normalized sensor response was 0.062 ( Figure 5A and 6B). Based on the comparison of normalized sensor response ( Figure 5A and 6B), it was concluded that the sensitivity of the biosensor was decreased by 8.82% ( Figure 6B ). With less than 10% of sensor response decrease, the regeneration step was working quite well for the developed system. The biosensor reusability response depends on the retained CNT-FET fabricated materials and the biosensor offers 91.18% retained or higher sensing response. We designed and successfully developed a very sensitive, selective, and quantifiable CNT- were immobilized on the PBASE layer. The biosensing responses were rapid (minutes) and produced reliable, reproducible results. Antigen-specific concentration-dependent sensing response over a broad range of concentrations (0.1 to 5000 fg/mL) was shown in the calibration experiments. The CNT-FET biosensor is highly sensitive and able to discriminate SARS-CoV-2 S1, SARS-CoV-1 S1, and MERS-CoV S1 antigens, indicating good selectivity to SARS-CoV-2 S1. The experimental data indicate that the biosensor can be a good, lowcost biosensor for rapidly testing people for SARS-CoV-2 infection that is easy to handle. Thus the developed biosensor is inexpensive, simple, versatile, and easy to use, can be a prototype for SARS-CoV-2 S1 antigen detection and diagnosis of SARS-CoV-2 infection. The authors would like to acknowledge King Abdulaziz City for Science and Technology CoV-2 S1-mAb antibodies. Lanes 1, 2, 3 represent loaded recombinant SARS-CoV-2 S1, SARS-CoV-1 S1, and MERS-CoV S1 antigens. For each antigen concentration, 1 µg/mL was loaded in SDS-PAGE. SARS-CoV-2 S1 antibody immobilized CNT-FET surface (light blue) and SARS-CoV-2 S1 antigen drop (green line). The 100 fg/mL SARS-CoV-2 S1 was prepared in10 mM AA buffer of pH 6.0. All the transfer characteristics are measured under ambient conditions. (B) Real time senror response characteristics after dropping increasing concentration (0.1, 1, 10, 100, and 5000 fg/mL) of SARS-CoV-2 S1 antigen. (C) Calibration curve after normalizing biosensor response versus applied concentration data of target SARS-CoV-2 S1 antigen. SARS-CoV-2 S1 and non-target SARS-CoV-1 S1 and MERS-CoV S1 antigens were tested. Each antigens 100 fg/mL were prepared in10 mM AA buffer of pH 6.0. [45] . SARS-CoV-2 S1 SARS-CoV-2 S1 non-fortified saliva SARS-CoV-2 S1 fortified saliva SARS-CoV-1 S1 WHO Director-General's opening remarks at the media briefing on COVID-19 -11 Origin and evolution of pathogenic coronaviruses SARS-beginning to understand a new virus Middle East respiratory syndrome A novel coronavirus from patients with Pneumonia in China SARS-CoV-2: Camazotz's Curse The genetic sequence, origin, and diagnosis of SARS-CoV-2 Molecular architecture of the SARS-CoV-2 virus Virus-receptor interactions of glycosylated SARS-CoV-2 spike and human ACE2 Receptor Characterization of spike glycoprotein of SARS-CoV-2 on virus entry and its immune cross-reactivity with SARS-CoV Spike glycoprotein-mediated entry of SARS coronaviruses Neutralizing antibodies targeting SARS-CoV-2 spike protein Structural and functional properties of SARS-CoV-2 spike protein: potential antivirus drug development for COVID-19 Structure, function, and antigenicity of the SARS-CoV-2 spike glycoprotein Immune-mediated approaches against COVID-19 Animal and translational models of SARS-CoV-2 infection and COVID-19 SARS-CoV2 in different body fluids, risks of transmission, and preventing COVID-19: A comprehensive evidence-based review Transmission of SARS-CoV-2: A review of viral, host, and environmental factors On the whereabouts of SARS-CoV-2 in the human body: A systematic review Electrochemical biosensors for pathogen detection An electrochemical immunosensor for the corona virus associated with the Middle East respiratory syndrome using an array of gold nanoparticle-modified carbon electrodes Electrical detection of hepatitis C virus RNA on single wall carbon nanotube-field effect transistors Electrochemical detection of influenza virus H9N2 based on both immunomagnetic extraction and gold catalysis using an immobilizationfree screen printed carbon microelectrode An impedance immunosensor based on low-cost microelectrodes and specific monoclonal antibodies for rapid detection of avian influenza virus H5N1 in chicken swabs Rapid detection of SARS-CoV-2 antigens using high-purity semiconducting single-walled carbon nanotube-based fieldeffect transistors Ultrasensitive label-free sensing of IL-6 based on PASE functionalized carbon nanotube micro-arrays with RNA-aptamers as molecular recognition elements Electrical detection of specific versus nonspecific binding events in breast cancer cells Labelfree capture of breast cancer cells spiked in buffy coats using carbon nanotube antibody micro-arrays Development of single-walled carbon nanotubebased biosensor for the detection of Staphylococcus aureus Extreme oxygen sensitivity of electronic properties of carbon nanotubes Local-gated single-walled carbon nanotube field effect transistors assembled by AC dielectrophoresis A carbon nanotube immunosensor for Salmonella Carbon nanotube transistors for biosensing applications Carbon nanotubes-based chemiresistive biosensors for detection of microorganisms Enhancement of sensitivity and specificity by surface modification of carbon nanotubes in diagnosis of prostate cancer based on carbon nanotube field effect transistors Identifying the mechanism of biosensing with carbon nanotube transistors Characterization of the receptor-binding domain (RBD) of 2019 novel coronavirus: implication for development of RBD protein as a viral attachment inhibitor and vaccine Rapid detection of COVID-19 causative virus (SARS-CoV-2) in human nasopharyngeal swab specimens using fieldeffect transistor-based biosensor Isoelectric point determination by imaged CIEF of commercially available SARS-CoV-2 proteins and the hACE2 receptor Dual-functional plasmonic photothermal biosensors for highly accurate severe acute respiratory syndrome coronavirus 2 detection Rapid SARS-CoV-2 spike protein detection by carbon nanotube-based near-infrared nanosensors Minute-scale detection of SARS-CoV-2 using a low-cost biosensor composed of pencil graphite electrodes Two-Dimensional-Material-Based Field-Effect Transistor Biosensor for Detecting COVID-19 Virus (SARS-CoV-2) ExPASy: The proteomics server for in-depth protein knowledge and analysis The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.