key: cord-0915555-7omgkm9v authors: Uwamino, Yoshifumi; Nagata, Mika; Aoki, Wataru; Nakagawa, Terumichi; Inose, Rika; Yokota, Hiromitsu; Furusawa, Yuri; Sakai-Tagawa, Yuko; Iwatsuki-Horimoto, Kiyoko; Kawaoka, Yoshihiro; Hasegawa, Naoki; Murata, Mitsuru title: Accuracy of Rapid Antigen Detection Test for Nasopharyngeal Swab Specimens and Saliva Samples in Comparison with RT-PCR and Viral Culture for SARS-CoV-2 Detection date: 2021-04-17 journal: J Infect Chemother DOI: 10.1016/j.jiac.2021.04.010 sha: 38a5dba1b77739377b601a9dbc4966bc50a0e78e doc_id: 915555 cord_uid: 7omgkm9v Introduction Rapid antigen detection (RAD) tests are convenient tools for detecting the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) in clinics, and testing using saliva samples could decrease the risk of infection during sample collection. This study aimed to assess the accuracy of the SARS-CoV-2 RAD for testing of nasopharyngeal swab specimens and saliva samples in comparison with the RT-PCR tests and viral culture for detecting viable virus. Methods One hundred seventeen nasopharyngeal swab specimens and 73 saliva samples with positive results on RT-PCR were used. Residual samples were assayed using a commercially available RAD test immediately, and its positivity was determined at various time points during the clinical course. The concordance between 54 nasopharyngeal swab samples and saliva samples that were collected simultaneously was determined. Viral culture was performed on 117 samples and compared with the results of the RAD test. Results The positive rate of RAD test using saliva samples was low throughout the clinical course. Poor concordance was observed between nasopharyngeal swab specimens and saliva samples (75.9%, kappa coefficient 0.310). However, a substantially high concordance between the RAD test and viral culture was observed in both nasopharyngeal swab specimens (86.8%, kappa coefficient 0.680) and saliva samples (95.1%, kappa coefficient 0.643). Conclusions The sensitivity of the SARS-CoV-2 RAD test was insufficient, particularly for saliva samples. However, a substantially high concordance with viral culture suggests its potential utility as an auxiliary test for estimating SARS-CoV-2 viability. The coronavirus disease (COVID-19) pandemic caused by the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) has emerged as a serious concern worldwide. To control the spread of the SARS-CoV-2 infection, it is essential to isolate infected patients based on a laboratory-confirmed diagnosis. Reverse transcription PCR (RT-PCR) is a gold standard investigation in the clinical diagnosis of COVID-19; however, RT-PCR requires special equipment, skilled medical technologists, and considerable time (several hours) to generate test results. Rapid antigen detection (RAD) tests are a type of point-of-care testing to detect SARS-CoV-2-specific antigen proteins based on a lateral flow assay, which can be performed within 30 minutes without necessitating any special equipment or training. However, the samples for RAD tests generally comprise nasal or nasopharyngeal swab (NPS) samples and, therefore, the risk of infection among medical staff during sample collection is inevitable. Saliva, which can be used as an alternative sample in RT-PCR for SARS-CoV-2, can be self-collected by patients and thus reduce the exposure risk of medical staff. [1, 2] However, there is a lack of studies that demonstrate the accuracy of the RAD test in saliva samples. [3] Furthermore, although the viability of the virus is important for predicting its communicability, clinical studies on RAD test results and virus viability are scarce. [4] Therefore, we conducted an observational study to assess the accuracy of the SARS-CoV-2 J o u r n a l P r e -p r o o f RAD test of NPS and saliva samples in comparison with the RT-PCR tests and viral culture for detecting viable virus. In this single-centre study, we used NPS specimens and saliva samples collected for SARS-CoV-2 RT-PCR testing from July 1 to October 8, 2020 at the Keio University Hospital (Tokyo, Japan). The NPS specimens and saliva samples were collected simultaneously in some patients to assess the accuracy of RT-PCR testing of saliva samples; [5] otherwise, either an NPS specimen or a saliva sample alone were collected. NPS specimens were collected by trained medical staff using a FLOQ SWAB and a Copan UTM container (Copan, Brescia, Italy), and saliva samples were self-collected by patients themselves by spitting into sterile containers (ASIAKIZAI, Tokyo, Japan) after 1 min of salivation. SARS-CoV-2 real-time RT-PCR was performed using the LightCycler96 (Roche, Basel, Switzerland) and the 2019 Novel Coronavirus Detection Kit (Shimadzu, Kyoto, Japan) by using N1 and N2 primers and probes in accordance with the manufacturer's instructions. Furthermore, to assess the viability of the virus in the sample when the patients' consent was obtained for sample use, the refrigerated residual samples were sent to biosafety level 3 facilities, and viral culture was performed to detect infective virus using VeroE6/TMPRSS2 cells in Dulbecco's Modified Eagle Medium containing 10% fetal calf serum (FCS), 1 mg/mL G418, 100 units/mL penicillin, 100 µg/mL streptomycin, and 5 µg/mL Plasmocin Prophylactic (InvivoGen, San Diego, CA, USA). A 24-well plate containing VeroE6/TMPRSS2 cell culture monolayer at a low crowding density (70-95% confluent) was prepared. After the medium was discarded, 100 µL of sample was added to the cells and incubated for 1 hr at 37°C under 5% CO 2 . Then, 0.5 mL of growth medium was J o u r n a l P r e -p r o o f added and incubated at 37°C under 5% CO 2 for 1 week until a cytopathogenic effect was observed. (supplemental file. Fig. S1 ). [7] Information about the date of sample collection, patient age, sex, and date of symptom onset were obtained from the electronic hospital medical records. The date of symptom onset was defined as the date when fever, respiratory symptoms, olfactory disorder, or taste disorder were first observed in symptomatic patients, and as the date that the first RT-PCR-positive sample was collected in presymptomatic or asymptomatic patients. The duration after symptom onset was calculated as the interval (days) between sample collection and symptom onset. The relationship between the RAD test results and the days after symptom onset and the Ct values of RT-PCR were assessed. The Ct values of the N1 prime probe sets were used for the assessment because the elevation of the amplification curve for the N2 primer probe sets was not observed in some samples according to the difference in the amplification efficacy among the two primer probe sets. For the NPS specimens and saliva samples that were collected simultaneously, the concordance between the RAD tests between the two samples was calculated. Additionally, among the viral culture samples, the concordance between the RAD test results and viral culture results was assessed. A total of 190 RT-PCR positive samples obtained from 72 patients were used. One hundred and seventeen samples were NPS specimens and 73 were saliva samples. Patient age, sex ratio, and number of days from onset were equivalent for the two sample types. The Ct values of the saliva samples were slightly lower than those of the NPS specimens although the difference was not statistically significant (Table 1) . Seventy-two percent of NPS specimens that were collected within 4 days of symptom onset were positive for the RAD test. However, the RAD test positivity rate of NPS specimens collected 5 days after symptom onset was less than 30%, and those of saliva samples were lower than 30% in each time period (Fig. 1) . 2) . Table 3 ). SARS-CoV-2 RAD tests are convenient and can be performed easily even at the patient's bedside, although their low sensitivity is of considerable concern for making differential diagnosis of suspected acute respiratory infection cases in COVID-19 pandemic. Scohy et al. reported that the RAD test detected only 42.1% of RT-PCR-positive samples [8] ; similarly, this study revealed the poor sensitivity of RAD tests. Therefore, it might be difficult to substitute the RAD test for RT-PCR testing in COVID-19 diagnosis. In particular, a poor concordance of the RAD test results between NPS specimens and saliva samples was demonstrated in this study, and this is compatible with previous reports of the poor sensitivity of the RAD test using saliva samples. [3] Additionally, Diao et al. mentioned the high accuracy of RAD testing of NPS specimens collected immediately after the onset of symptoms [9] , and this study identified the high sensitivity of RAD testing J o u r n a l P r e -p r o o f in NPS specimens collected within 4 days from symptom onset. However, even in the early phase, the sensitivity of the RAD test of saliva samples was low, which demonstrates that the clinical utility of RAD testing of saliva samples is limited in our study setting. It is interesting that a substantially high concordance was observed between RAD test results and viral culture in both NPS specimens and saliva samples. In most cases, the samples with positive viral culture showed Ct values of less than 25. [10] The Ct value threshold of RAD test positivity was approximately 25 in saliva samples and 30 in NPS samples. Therefore, it was simple for viral culture-positive samples to also become RAD positive. Despite this confounding relationship, this fact implies RAD test can be a potential auxiliary diagnostic tool for estimating the infectivity of the COVID-19 patients. Patients with the viable virus are thought to be highly contagious, and strict isolation is required. RT-PCR, which can detect not only viable virus but also fragmented RNA, is known to continue to test positive for as long as 1 month from symptom onset. [11] Therefore, negative conversion of RT-PCR is an unsatisfactory criterion for ending isolation, as it can result in unnecessarily prolonged isolation. Although the sensitivity of viral culture for detecting viable virus is uncertain, only viral culture can detect viable virus. Therefore, negative conversion of viral culture is an ideal criterion for ending the isolation of COVID-19 patients; however, viral culture requires special equipment and experienced staff trained in biosafety protocols. Although the sample size is insufficient to draw a firm conclusion, the negative J o u r n a l P r e -p r o o f predictive value of the RAD test for predicting the viral culture result suggests that one or two negative RAD test results is potentially useful as a criterion for assessing the viability of SARS-CoV-2. This study has several limitations. First, RAD test samples of NPS specimens were not collected directly from the patient but were obtained from residual UTM medium. Therefore, the viral load of the samples might be lower than that of the directly collected samples. Second, the sensitivity of viral culture is difficult to determine. Therefore, a negative viral culture does not always mean the absence of communicability, although viral culture is the only available means for assessing the infectivity of the sample at this time. Third, although the false positivity of RAD test is problematic, [12] this study could not evaluate the specificity of RAD tests compared to RT-PCR because RT-PCR-negative samples were not included in this study. Finally, the variance of the accuracy among commercial kits was not studied. Therefore, to determine the external validity of the study results, further studies including a variety of kits and large sample sizes are essential to optimise the usage of RAD kits in the clinic management of COVID-19. In conclusion, although the sensitivity of the SARS-CoV-2 RAD test was insufficient for ruling out COVID-19 diagnosis, particularly when using saliva samples, the substantially high concordance of the RAD test and viral culture suggests its potential utility in estimating SARS-CoV-2 viability. J o u r n a l P r e -p r o o f J o u r n a l P r e -p r o o f Comparison of sars-cov-2 detection in nasopharyngeal swab and saliva Saliva as a non-invasive specimen for detection of sars-cov-2 Clinical evaluation of self-collected saliva by quantitative reverse transcription-pcr (rt-qpcr), direct rt-qpcr, reverse transcription-loop-mediated isothermal amplification, and a rapid antigen test to diagnose covid-19 Evaluation of rapid antigen test for detection of sars-cov-2 virus Accuracy and stability of sample for reverse transcription pcr detection of sars-cov-2 A cdc 2019-novel coronavirus (2019-ncov) real-time rt-pcr diagnostic panel. CDC Low performance of rapid antigen detection test as frontline testing for covid-19 diagnosis Accuracy of a nucleocapsid protein antigen rapid test in the diagnosis of sars-cov-2 infection Prolonged persistence of sars-cov-2 rna in body fluids Another false-positive problem for a sars-cov-2 antigen test in japan