key: cord-0915034-89rvfavn authors: Biswas, Gokul Chandra; Suzuki, Hiroaki title: Simple manual roller pump-driven valve-free microfluidic solution exchange system for urgent bioassay date: 2022-01-21 journal: RSC advances DOI: 10.1039/d1ra08052k sha: 7672d267688fbbcc61035b6a02c3bfc1a2b0b0e8 doc_id: 915034 cord_uid: 89rvfavn We introduce a simple-to-use manual roller pump (MRP)-driven and valve-free microfluidic system for sequential solution exchange, followed by a bioassay to detect protein. The polydimethylsiloxane (PDMS)/glass-based disposable device comprises a reaction chamber, multiple micro-flow channels (μFCs), and air vents. The practical solution exchange was realized by sequential injection and withdrawal of several solutions into and from the reaction chamber through constricted μFCs by utilizing changing air pressure of an MRP when a small cylindrical roller was pressed and rolled over a soft silicone tube using a finger. Furthermore, we investigated the effect of surface hydrophobicity on solution exchange. A sandwich fluorescence-based immunoassay to detect human interleukin 2 (IL-2) was performed using this simple microfluidic scheme to demonstrate its suitability for analytical bioassays. The system allowed quick IL-2 detection in 20 min in a pre-functionalized device with a detection limit of 80 pg mL(−1) and a range of 125 pg mL(−1) to 2.0 ng mL(−1). We have thus developed a microfluidic scheme that non-experts can efficiently perform and that can be the fundamental module for low-cost bioassays necessary for emergencies and situations where resources are constrained. Infectious diseases are highly prevalent in developing countries. [1] [2] [3] Contagious diseases can become global pandemics. They can exert catastrophic impacts on countries regardless of their developmental status, as recently exposed by the millions of infections and deaths caused by the COVID-19 pandemic. 4 To deal with the consequences of the pandemic, the World Health Organization (WHO) urged for an increase in the quantity of diagnostic tests by promoting their "test, test, test," 5 and ranked point-of-care testing (POCT) as the rst of eight urgent research actions identied at the COVID-19 pandemic epoch. 6 In this respect, Micro Total Analysis Systems (mTAS) 7, 8 can act as a agship that enables diagnosis seekers to receive low-cost, rapid diagnostic outputs from non-experts at nearby clinics, at home, or in resource-constrained settings. If realized, this method should augment personalized therapy or treatment against critical diseases in contrast to expensive, laborious, laboratory-oriented conventional diagnostic approaches. [1] [2] [3] 9, 10 Alongside nucleic acid testing, protein detection can aid indirect diagnosis and prediction of immune status against particular infectious diseases. 11 ELISA is commonly performed to detect target proteins or biomarkers; 12 even so, traditional ELISA requires skilled operators, occupies more space and assay times, and consumes a bigger quantity of costly reagents or samples, obstructing its great chance of extensive applications for domestic or clinical diagnosis. As a problem-solving approach, microuidics enables immunoassays on miniaturized chips or devices. 11, 13, 14 The small assay chambers in microsystems maintain the high surface-to-volume ratios to support prompt and high-throughput analysis. 13, 14 Micro-ow systems carry the enormous potential to facilitate emergency medical care with the simple analytical tool in remote corners of resource-inadequate settings. Practically, despite its prospects, the application of these systems has not been successfully expanded so far except for the lateral ow immunoassay (LFIA). 15 This is because the simplication-allied challenges in the sequential exchange of several solutions essential for consecutive reactions and washings in microuidic affinity biosensing is yet to be addressed. Solution exchange for microuidic bioassays is usually governed by microuidic components such as pumps or valves. Among the reported pumps, pressure-driven pumps are straightforward for transporting solutions into and from the sensing/assay/reaction chamber through micro-ow channels (mFCs). 13, 16, 17 However, these pumping systems rely on bulky peripheral instruments such as electric syringe pumps or pneumatic apparatus. Moreover, comparatively wide mFCs and high pressure are required to eliminate uidic resistance. 18, 19 Otherwise, valves, such as pneumatic or check valves, are formed in mFCs to control solution delivery. 13, 16, 17, [19] [20] [21] To simplify pressure-driven pumping, approaches such as by a nger, 22,23 screw, 24 or chemicals 25 have been reported, eliminating off-chip control for pumping. However, they involve extra fabrication steps and additional cost of a thin layer PDMS membrane, rotary electric motor, and chemical modication in the pump unit. Alternatively, pumping utilizing the passive driving forces of capillarity, 26-28 evaporation 29 or PDMS degassing 30 have become attractive for bioassays; but their respective dependence on tiny capillary patterns and preset solution volumes, perpetual temperature and humidity, and more prolonged vacuum exposure generated scope of thinking more simplication regarding applications under resource constraints or in emergencies. Moreover, these micropumps may face difficulty in controlling solution ow without valve components. 27, 31 Solution exchange systems that require stopped-ow in the bioassay process may be unrealistic for these pumping approaches. The issue can either be addressed by narrowing the mFCs in the direction of solution ow or by altering plug or column ow, either in capillary-triggered 28, 31 or syringe-driven settings. [32] [33] [34] However, maintaining simplicity in pump operation remains problematic. Earlier, we incorporated a narrow mFC pattern on electrowetting-based valve areas and a ushing port in the reaction chamber for serial injection and removal of solutions, respectively. 35 Later, we eased the exchange of solutions and applied wide mFCs by integrating electrochemically switchable hydrophobic valves 36,37 and a superabsorbent pump 27 against capillary-owed solutions. Moreover, the solutions can be transported automatically, 37, 38 allowing this system to be operated on the spot without an external power or pressure source. However, extra fabrication steps and material costs are involved with the construction of these components. Overall, microuidic solution exchange without electric power and valve component remains a hurdle for on-the-spot bioassays. As a result, interest in solution pumping in micro-uidic devices that does not require electricity is growing. [39] [40] [41] Thus, it is imperative to offer structural and functional simplicity in pumping strategies and microuidic devices to realize the practical exchange of multiple solutions and facilitate instant microuidic immunoassays. In this article, to address the above-stated limitations, we present a competent scheme of multiple solution exchange based on a manual roller pump (MRP)-driven valve-free simple microuidic device. We employed an ordinary so silicone tube with a small cylindrical roller as a pump (Fig. 1) . Solution loaded in a tube or reservoir is injected and removed into and from the central reaction chamber through the connected, constricted mFCs by simple rolling of the roller with a nger, thus realizing a simple solution exchange without expensive instruments ( Fig. 1A and B ). Our strategy minimizes the cost and complexity of device fabrication and operation. Additionally, we demonstrated the effectiveness of the scheme by performing on-device immunoassay to detect human interleukin 2 (IL-2), which is a potential biomarker for testing/diagnosis of infections, immune response, cytotoxicity, and lymphocyte transformation. 33, [42] [43] [44] We envision this simple microuidic approach as a valuable tool for enabling many easy-to-perform instantaneous bioassays or rapid POC tests during emergencies or in situations where access to resources is constrained. Materials and reagents utilized for device fabrication and experiments were purchased from the following commercial sources: Glass wafers (TEMPAX Float®; diameter ¼ 3 inches, thickness ¼ 500 mm) from Schott Japan (Tokyo, Japan); a thick-lm photoresist SU-8 25 from MicroChem (Newton, MA, USA); a negative photoresist (OMR-83) from Tokyo Ohka Kogyo (Kawasaki, Japan); a prepolymer solution of poly(dimethylsiloxane) (PDMS; KE-1300T) and (3-aminopropyl) triethoxysilane (APTES) from Shin-Etsu Chemical (Tokyo, Japan); phosphate-buffered saline (PBS, pH 7.4), bovine serum albumin (BSA), uorescein, and a dye, sunset yellow FCF (C 16 -H 10 N 2 Na 2 O 7 S 2 ), from Wako Pure Chemical Industries (Osaka, Japan). Orange G (C 16 H 10 N 2 Na 2 O 7 S 2 ) and methylviolet (C 24 H 28 N 3 Cl) were obtained from Merck (Darmstadt, Germany). Food dyes were obtained from Kyoritsu Foods (Tokyo, Japan). An FITC labeling kit (lot. PG201955) was procured from Thermo Scientic (Rockford, IL, USA). The standard ELISA kit for human IL-2 (Human IL-2 ELISA MAX Deluxe) was obtained from BioLegend (San Diego, CA, USA). Information regarding the concentrations of the reagents in the ELISA kit was proprietary. A PBS solution containing 0.5% BSA was used to dilute the protein solutions. A diluent in the kit was used to prepare the 200-fold diluted antibody (capture and detection antibody) solutions. All the other reagents not stated above were acquired from Wako Pure Chemical Industries (Osaka, Japan). Milli-Q pure water of 18 MU cm À1 resistivity was utilized to make all the necessary solutions. The devices were fabricated with a PDMS substrate with microuidic patterns and a glass substrate. A thick-lm photoresist (SU-8 25) was used to form a template for the PDMS structures via the replica molding technique, as detailed in the ESI. † Fig. 2A shows one of the devices which was used in the nal stage. Other congurations are discussed in later sections and shown in ESI (Fig. S1 †) . The devices contained ve connecting mFCs, a reaction chamber, and four air vents in the PDMS substrate. The height of the ow channels and reaction chamber was 80 mm for all devices. The width of each of the ow channels was 50 mm. The diameter of the reaction chamber was 1.5 mm (volume: $140 nL). To smoothen the transport of the solution by applying air pressure, air vents of 30 mm width were added to the reaction chamber near the outlets. For injecting and removing the solution, the inlet and outlet ports were formed by making through-holes in the PDMS layer using a disposable 1 mm biopsy punch (ref. BP-10F, Lot no. 16B25; KAI Industries Co. Ltd., Seki, Japan). The outside ends of the air vents contained through-holes in the PDMS stratum. The inlets and outlet were connected to the MRP through broken needles inserted into silicone tubes (Fig. 2B) . To complete the device, the patterned PDMS and glass substrate were exposed to oxygen plasma for 20 s at 20 W and 30 Pa of oxygen pressure to make strong bonding. The activated glass and PDMS substrates were aerward aligned and bonded by applying gentle pressure. The MRP consists of a so silicone tube (inner and outer diameters of 1 mm and 2 mm, respectively) with a connected piece of a small needle (1 mm outer diameter) inserted into the port of the inlet reservoir to load solutions and pump the device. With the assistance of a small cylindrical rigid roller, such as a glass bottle, rod, or dry cell battery, the pump is ready to use (Fig. S2 †) . The operation of the MRP is very simple and transport of a solution can be started immediately by placing, pressing, and rolling a roller on a tube (Fig. 3A) . Keeping the solution plug at the front, a small roller is gently pressed using a nger so that the top and bottom surfaces of the hollow tube become fused, creating an airlock before the solution plug in the tube. Aerward, positive air pressure is generated by displacing the air towards the solution by rolling the roller over the tube, which pushes the solution plug forward to realize forward pumping, like in a pneumatic pump (Fig. 3A [i] and B[i and ii]). In another way, applying the same motion to the free end of the tube will provide a suction pump action that creates a negative air pressure in the tube, resulting in the solution moving forward by suction pumping (Fig. 3A [ii] and B[iii and iv]). We call this process roller pumping, which we applied to realize solution exchange in the microuidic device. Likewise, solution exchange can be performed by MRP-based air pumping with solutions in reservoirs formed in the mFC structures. In that case, the solutions are rst loaded using a micropipette, and only air is pumped as described above (Fig. S3 †) . The volume of solutions used for analysis is determined by the volume of the reaction chamber. Therefore, as long as only the reaction chamber is lled with the solutions precisely without leakage into the other ow channels, the same volume of solutions is guaranteed and no additional procedure for the measurement of solution volumes is necessary. Then again, if the quantitative loading or delivery of a solution is necessary in the differently congured devices, a scale can be marked/printed on the outer surface of the silicone tube. In real situations, reagent solutions used for the analysis may be loaded beforehand in the tube or solution reservoirs on the chip using any tools such as syringes or micropipettes. Sample solutions can also be loaded into the tube in the same manner. However, for this purpose, MRP can also be used without the assistance of any other tools. Aer loading in the tube by suction pumping, the solution plug can be moved to the solution reservoirs and reaction chamber by forward pumping. The detection IL-2 antibodies (anti-IL-2) were labeled following the instructions of Thermo Fisher Scientic. In short, 40 mL of a borate buffer solution (0.67 M) was added to 500 mL of the detection antibody (dAb) diluted to 2 mg mL À1 in PBS. The FITC reagent was dissolved in 500 mL of the antibody solution by pipetting the solution up and down 10 times with brief vortexing. The mixture was incubated for 60 min at 4 C and then added to the spin column lled with a purication resin and mixed by brief vortexing to purify the labeled antibodies. The puried antibodies were collected by centrifugation of the spin column at 1000Âg for 30-45 s. Aer which, the FITC-to-antibody ratio was estimated as stated by the guidelines of Thermo Fisher Scientic. Approximately four FITC uorophores were calculated for each FITC-labeled antibody. Before labeling, all the reagents were stored at room temperature. We adopted a sandwich uorescence immunoassay scheme to detect human IL-2 (Fig. 4) . Unlike the traditional lengthy ELISA process, a rapid APTES functionalization-based colorimetric ELISA protocol 33 was applied with a slight modication for uorescence immunoassay. APTES forms a contact network for antibodies through its self-polymerization ability in an aqueous solution. 45 Moreover, the partially polymerized APTES network offers immobilization sites for cAb via electrostatic interactions between the amino and carboxylic groups of APTES and cAb, respectively. 45, 46 The one-step reaction chamber functionalization and cAb immobilization started with the injection of 1 mL of instantly mixed cAb solution, APTES (1%, v/v), and coating buffer into the assay chamber, and continuing the incubation for 25 min at 4 C. To block the non-specic binding of proteins on the surface, 1 mL of PBS containing 0.5% BSA and 0.01% Tween 20 was introduced and incubated for 30 min at room temperature aer removing the previous solution and subsequent washing process. Before IL-2 detection, we formulated a standard series of IL-2 solutions by diluting the IL-2 stock solutions with PBS. To enable the rapid assay step in our system, we introduced an immediately prepared mixture of IL-2 and FITC-labeled dAb (at a 1 : 1 ratio) in the assay chamber in a single step instead of delivering the target protein and dAb solution in separate steps, as practiced in a typical immunoassay. The mixture formed a protein complex (cAb-protein [IL-2]-FITC labeled dAb) with cAb while incubated for 20 min at 4 C. Later, the assay chamber was washed 2-3 times using PBS containing Tween 20, and the uorescence signal from the protein-antibody conjugate was measured using a uorescence microscope (IX-73; Olympus, Tokyo, Japan) with a lter unit (Cy5-4040C, Olympus) and CMOS camera (ORCA-Flash 4.0; Hamamatsu Photonics, Hamamatsu, Japan). Exposure to UV light with 495 nm wavelength was provided during measurement. The IL-2 concentration was determined from the uorescence signal. Device design, pumping, and solution exchange The solution exchange for immunoassay necessitates introducing the desired solutions into the assay chamber without entering the other mFCs, and successive removal through the exit port. To this end, we rst constructed a type 1 microuidic device with seven 200 mm wide inlet ow channels and an outlet ow channel that surrounded a reaction chamber of 2 mm diameter (Fig. S1A †) . Each channel was constricted with a 50 mm-wide narrow structure at the entry to the reaction chamber, and each contained an inlet port and a solution reservoir. Narrowing the PDMS mFC to $50 mm can provide a hydrophobic barrier; 27,32 therefore, we can prevent the undesired delivery of multiple solutions in the reaction chamber and through other mFCs (Fig. S4 †) . Using the MRP, we successfully transported solutions to the reaction chamber without them entering other channels (Fig. S5A, B †) . Nonetheless, we experienced problems with solution removal due to the imbalance of suction pressure and positions of the inlet channels, particularly those situated near the outlets. The solutions were withdrawn improperly, leaving undesired residues in the reaction chamber (Fig. S5C †) and the inlet mFCs (Fig. S5C, D †) . This could contaminate other solutions and was not expected for solution exchange. To solve the problems, we fabricated the type 2 device (Fig. 5A, B ; S1B †). The width of mFCs was narrowed to 100 mm with the 50 mm constriction at the front. The inlet and outlet mFCs were placed at opposite sides of the reaction chamber (2.5 mm diameter), and the solution reservoirs in the middle of the inlet mFCs of the type 1 device were removed. Air vents (of 20 mm width) were additionally formed near the outlet to facilitate the proper injection and lling of solutions without air trapping. With the type 2 device, we succeeded in delivering and maintaining the solutions in the reaction chamber without penetrating the other mFCs, facilitating practical sequential solution exchange. Fig. 5C demonstrates the stepwise process of sequential injection, lling, and subsequent removal of seven dye solutions when employing the device. None of the solutions moved to other mFCs or fragmented during the exchange process. No residue was le in the reaction chamber while smoothly pumping out the solutions using suction. This simple, new microuidic scheme restrained the solutions in the reaction compartment for over 30 min. In immunoassay systems, frequent washing of the residual solution from the assay area is essential for producing an accurate signal. In this simple device, a single mFC can be effectively engaged to perform multiple washes without any interruption (Fig. 5D) . Pumping in the long so tube is subjected to extra back pressure because of the higher tube wall friction. Furthermore, the so tube assists in uidic pulsation absorption. 19 The MRP also carries those features in its operation. As a consequence, MRP delivered a smooth pumping experience while exchanging solutions in the reaction chamber. The solutions were transported and withdrawn to and from the reaction chamber slowly (at a ow rate of $6 mL min À1 ) using the MRP. We calculated the ow rate based on the ratio of solution volume transported and the time required to ll and empty the reaction chamber by the MRP action. With the MRP, we tested and succeeded in transporting solution volume over 30 mL in the tube. As the pump is manually operated, it was not easy to achieve a constant ow rate. However, the MRP can be used whenever an operator needs it. Interestingly, MRP can enable bidirectional pumping. The bidirectional motion of the roller from the same side of the solution plug correspondingly can initiate the forward and backward ow of a solution by utilizing the positive and negative pressure of the MRP (Fig. 6A) . Consequently, solution exchange without outlet mFCs can be accomplished in a micro-uidic device using the bidirectional MRP (Fig. 6B) . While performing immunoassays, the surface of the substrate can be hydrophobic, leading to becoming air trapped inside the chamber by non-specic or physical adsorption of the antibody, antigen, or other biomolecules. The presence of air in the assay chamber may produce a misleading signal by altering the antibody immobilization and binding process because air can deteriorate the quality of proteins and cause light scattering. 33 Therefore, we examined the effect of hydrophobicity on solution exchange using the devices formed with an intentionally formed hydrophobic layer of negative photoresists (i.e., SU-8-25 and OMR-83) or PDMS beneath the reaction chamber. The contact angle of the water (1 mL) for PDMS-stamped, SU-8-25, and OMR-83 surfaces were 59 , 73.8 , and 90 , respectively (Fig. S6A †) , and no problems with solution transport, lling, and withdrawal was observed due to the increase of hydrophobicity (Fig. S6B †) . Finally, we constructed the type 3 device ( Fig. 2A ; S1C †), with slight modications to the design of the type 2 device, to make it simpler and reduce the reaction volume of the assay. The number of inlet mFCs was reduced to ve, and the width of all inlet mFCs was made 50 mm along their entire length instead of forming denitive constriction, and the assay chamber was downsized to 1.5 mm. Air vents were widened to 30 mm and placed closer to an outlet with 200 mm width and 50 mm constriction. The reaction chamber effectively enclosed the solution, preventing penetration to any mFC (Fig. 2B) . The device was employed for exchanging assay reagents in the immunoassay mentioned later. BSA is a popular blocking reagent applied in immunoassays to prevent the undesired adsorptions of biomaterials; 41, 47 nonetheless, the BSA concentration should be adjusted to enable smooth transport and lling of solutions in the assay chamber without generating air bubbles. We tested the feasibility of exchanging protein solutions in the type 3 device with BSA solutions of 0.1-1% concentration. The device presented no issues of bubble trapping during the solution exchange operation for BSA solutions up to 0.5%. In contrast, concentrations above 0.5% sometimes resulted in the air becoming trapped inside the reaction chamber (Fig. S7 †) . Accordingly, a 0.5% BSA solution was chosen for blocking. To show the practicality of the MRP and valve-free device, we performed a sandwich uorescence immunoassay for human IL-2 aer the compatibility testing of MRP-driven exchange of assay reagents in type 3 device (ESI and Fig. S8 †) . For the immunoassay purpose, inlet mFC 3 was used to transport the coating buffer mixture, APTES (1%, v/v), and cAb. mFCs 2 and 4 were used to deliver the blocking reagent (0.5% BSA + 0.01% Tween 20) and the mixture of FITC-labeled dAb and antigen (IL-2), respectively. mFCs 1 and 5 were used for washing. For ELISA, either mFC 1 or 5 can be used to transport either the substrate or washing solution. The uorescence signal derived from the formed protein complexes (i.e., cAb-Ag-FITC-labeled dAb) was measured only from the reaction chamber. Fig. 7 shows that uorescence intensity depends on IL-2 concentration. The uorescence image taken from the reaction chamber indicated the variation in intensity due to the change of IL-2 concentration, corresponding to the amount of IL-2 trapped by the immobilized cAb. Some faint uorescence was also observed for the blank sample without IL-2, which indicated the non-specic adsorption of FITC-labeled dAb onto the surfaces of the assay chamber. The intensity increased with the increment of IL-2 concentration in the range of 125 pg mL À1 to 2.0 ng mL À1 and inclined to saturate aer that range. The platform showed the limit of detection (LOD) for IL-2 as 80 pg mL À1 , the average response value for blank and three times its standard deviation. The LOD of our device lies within the normal range of IL-2 concentrations (i.e., 10 pg mL À1 to 1 ng mL À1 ) in serum. 42 Recent research has explored the elevated levels of IL-2 found in mild or asymptomatic COVID-19 cases and highlighted the latent use of IL-2 among cytokines as prognostic biomarkers for pandemic SARS-CoV-2 infection and degree of immunity aer vaccination. 43, 44 Though our current LOD does not satisfy the required level (below 10 pg mL À1 ) of IL-2 detection to diagnose COVID-19 infection; an improved suitable nanomaterial-assisted assay 15 could meet that diagnostic demand. We also investigated the usefulness of APTES-based cAb immobilization. This was followed up by two methods of immobilization for the assay chamber-the rst one was by delivering the freshly mixed solution of coating buffer, APTES (1% v/v) and cAb; the second one was functionalization with APTES solution (for $25 min) followed by the injection of a mixed solution of cAb and coating buffer. For both approaches, they were incubated for 25 min because APTES-based cAb immobilization over this period can result in a maximum signal response in the assay process, which can be stabilized for up to 7 days. 33 Both the methods were found effective for cAb immobilization based on the similar uorescence responses from the assay chamber (Fig. 8) . However, the former decreased the number of operational steps; therefore, it was adopted for our study. Later, we studied the surface blocking effect of BSA to prevent non-specic adsorption of FITC-labeled dAb (Details in ESI †). The 0.5% BSA by itself and together with a surfactant (0.01% Tween 20) can reduce the noises of non-specic protein adsorptions by 35% and 55%, respectively, when compared to an unblocked surface (Fig. S9 †) . The considerable reduction in non-specic protein adsorption by adding the surfactant with BSA indicates the addition of a hydrophilic property (by providing a hydroxyl group) to the hydrophobic wall of PDMS lessens the hydrophobic interactions that cause undesirable binding of proteins with the surfaces. 48 The total time needed for our immunoassay was approximately 80 min including the time for the immobilization of capture antibodies and blocking with BSA, which is much shorter than the traditional ELISA. 12 If the reaction chamber is already functionalized with cAb and blocked with BSA, the assay time is shortened further to approximately 20 min, which will be acceptable in emergencies or resource-decient regions. The MRP is favorable for integrating microuidic systems where stopped-ow solutions, 20,35 a liquid plug, 25, 32, 34 or column 33 are necessary for bioassays. Furthermore, the MRP can process multiple liquid plugs (Fig. S10 †) , which can suit the system that requires the transport of a series of liquid plugs. 32, 34 The bidirectional MRP is well suited for platforms where the backward ow of the solutions is necessary. 25, 34 The ESI and Fig. S11 † shows a conceivable microuidic solution exchange scheme that utilizes the bidirectional MRP for bioassays. In this case, inlet chambers or tubes will act as waste reservoirs where forward pumped solutions can be retrieved with MRP-driven backow from the assay site to the inlet reservoirs aer solution processing (i.e., immobilization/functionalization, assay reactions, and washings followed by incubation). MRP can also actuate membrane displacement-based valves usually driven by the air pressure of pneumatic or peristaltic pumps, 20,21 which will further expand the possibility of novel microuidic devices with high performance. We demonstrated an easy-to-apply microuidic system for the sequential exchange of multiple solutions, which can be performed by untrained staff to detect proteins or biomarkers through a bioassay process. By adopting an uncomplicated roller pump and simplifying the design of a microuidic reaction chamber by connecting it with narrow ow channels and air vents, we facilitated sequential solution exchange without any crosscontamination of solutions. The manual roller pump, which is simply a hollow silicone tube with a small cylindrical rigid roller, can be a cheap alternative pressure source to pressure-driven pneumatic or syringe pumps that rely on expensive and bulky external devices; thus, this method alleviates cost and portability issues. With this system, we performed a stepwise uorescence immunoassay for detecting IL-2 in 20 min with a LOD of $80 pg mL À1 . Considering the weeklong stability of the APTES-based cAbimmobilized surface, the prefabricated and functionalized device can be applied for on-the-spot POCT. Although we used a microscope and benchtop analyzer for assay signal readouts, integration of a smartphone-based analyzer 49 can make this approach simpler and more suitable to on-the-spot or urgent biosensing applications. Moreover, colorimetric or electrochemical assays can also be planned with this easy-to-operate solution exchange platform, though further investigation is needed. Controlled solution processing is the prime issue for realizing stepwise assays in disposable devices, which can be addressed by the MRP and valvefree micro-ow structures. Thus, our streamlined microuidic scheme holds the potential to be applied for instantaneous bioassays or POCTs in emergencies such as pandemics or in resource-inadequate settings where many tests are necessary and trained manpower or materials are unavailable. There are no conicts of interest to declare. World Health Organization, WHO Director-General's opening remarks at the media brieng on COVID-19 -16 COVID-19 Public Health Emergency of International Concern (PHEIC) Global research and innovation forumj12 Lab-on-a-Chip Devices and Micro-Total Analysis Systems Transducers -2015 18th International Conference on Solid-State Sensors, Actuators and Microsystems This study was partly supported by a promotional research grant (Project code: LS/2018/1/02) of the SUST Research Centre, Shahjalal University of Science and Technology, Bangladesh and Grants-in-Aid for Scientic Research (No. 26560365) from the Japan Society for the Promotion of Science (JSPS), Japan. We thank Professor Masatoshi Yokokawa and Mr Shishir Kanti Pramanik of University of Tsukuba for their valued discussions.