key: cord-0913034-ewg9d4ei authors: Kitagawa, Hiroki; Nomura, Toshihito; Nazmul, Tanuza; Keitaro, Omori; Shigemoto, Norifumi; Sakaguchi, Takemasa; Ohge, Hiroki title: Effectiveness of 222-nm ultraviolet light on disinfecting SARS-CoV-2 surface contamination date: 2020-09-04 journal: Am J Infect Control DOI: 10.1016/j.ajic.2020.08.022 sha: a8a2fb0a94a8337e207beadd9e70e4975bd4b3ec doc_id: 913034 cord_uid: ewg9d4ei BACKGROUND: Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), which causes coronavirus disease 2019 (COVID-19), has emerged as a serious threat to human health worldwide. Efficient disinfection of surfaces contaminated with SARS-CoV-2 may help prevent its spread. This study aimed to investigate the in vitro efficacy of 222-nm far-ultraviolet light (UVC) on the disinfection of SARS-CoV-2 surface contamination. METHODS: We investigated the number of viable SARS-CoV-2 after UV irradiation (0.1 mW/cm(2)) at 222 nm for 10–300 s using the 50% tissue culture infectious dose (TCID(50)). In addition, we used quantitative reverse transcription polymerase chain reaction to quantify SARS-CoV-2 RNA under the same conditions. RESULTS: One and 3 mJ/cm(2) of 222 nm UVC irradiation (0.1 mW/cm(2) for 10 and 30 s) resulted in 88.5 and 99.7% reduction of viable SARS-CoV-2 based on the TCID50 assay, respectively. In contrast, the copy number of SARS-CoV-2 RNA did not change after UVC irradiation even after a 5-min irradiation. CONCLUSION: This study shows the efficacy of 222-nm UVC irradiation against SARS-CoV-2 contamination in an in vitro experiment. Further evaluation of the safety and efficacy of 222-nm UVC irradiation in reducing the contamination of real-world surfaces and the potential transmission of SARS-CoV-2 is needed. spread. This study aimed to investigate the in vitro efficacy of 222-nm far-ultraviolet light (UVC) on the disinfection of SARS-CoV-2 surface contamination. We investigated the number of viable SARS-CoV-2 after UV irradiation (0.1 mW/cm 2 ) at 222 nm for 10-300 s using the 50% tissue culture infectious dose (TCID 50 ). In addition, we used quantitative reverse transcription polymerase chain reaction to quantify SARS-CoV-2 RNA under the same conditions. One and 3 mJ/cm 2 of 222 nm UVC irradiation (0.1 mW/cm 2 for 10 and 30 s) resulted in 88.5 and 99.7% reduction of viable SARS-CoV-2 based on the TCID50 assay, respectively. In contrast, the copy number of SARS-CoV-2 RNA did not change after UVC irradiation even after a 5-min irradiation. This study shows the efficacy of 222-nm UVC irradiation against SARS-CoV-2 contamination in an in vitro experiment. Further evaluation of the safety and efficacy of 222-nm UVC irradiation in reducing the contamination of real-world surfaces and the potential transmission of SARS-CoV-2 is needed. Keywords: SARS-CoV-2; COVID-19; Environmental contamination; Disinfection; Ultraviolet light; Far-UVC Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), which causes coronavirus disease 2019 (COVID- 19) , has emerged as a serious threat to human health worldwide. 1, 2 SARS-CoV-2 is mainly transmitted through infected respiratory droplets and close contact with infected people. Recent studies showed that SARS-CoV-2 can remain viable for days on surfaces under controlled experimental conditions. 3, 4 Furthermore, surfaces in hospitals treating patients with COVID-19 were found to be contaminated by SARS-CoV-2, suggesting that the hospital environment could be a potential medium of transmission. [5] [6] [7] Therefore, if environmental disinfection is not effective, SARS-CoV-2 may spread widely and cause nosocomial infections. Recently, ultraviolet (UV) light disinfection systems have been increasingly used in healthcare settings in an attempt to decrease the transmission of nosocomial pathogens and prevent healthcare-associated infections. 8 Most UV disinfection systems use germicidal lamps emitting UV radiation C (UVC) around 254 nm. However, it is well-known that 254-nm UVC is harmful to the skin and eyes. Previous reports demonstrated that 222-nm UVC light, belonging to far-UVC (207-222 nm), has the same highly effective germicidal properties as 254-nm UVC; however, it is less harmful to the skin and eyes than 254-nm UVC because far-UVC light has a very limited penetration depth in the skin or eye. [9] [10] [11] [12] [13] In addition, previous studies have reported that low dose of 222-nm UVC inactivated aerosolized H1N1 influenza virus and human coronaviruses alpha HCoV-229E and beta HCoV-OC43. 14, 15 Although there are a few reports about the effectiveness of UV disinfection on SARS-CoV-2 16, 17 , the effect of 222-nm UVC on SARS-CoV-2 is poorly understood. The aim of this study was to investigate the in vitro efficacy of 222-nm UV disinfection of SARS-CoV-2 contamination. method and expressed as TCID 50 /mL, as described previously. 19 Viral suspensions (100 µL) containing ca. 5 × 10 6 TCID 50 were loaded onto a sterile polystyrene plate (8.8-cm diameter) and spread into a circle with a diameter of approximately 7 cm. The inoculum was allowed to dry under sterile laminar flow in a biosafety cabinet at room temperature. A total of five plates were prepared in triplicate for a non-irradiated control and for 222-nm UV disinfection for 10, 30, 60, and 300 s. All experiments with SARS-CoV-2 virus were performed in a biosafety level 3 (BSL3) containment facility at Hiroshima University. All experimental results are reported as means across three replicates. The Care222 TM (Ushio Inc., Tokyo, Japan) UVC emitting device used in this study was a 222-nm krypton-chloride (Kr-Cl) excimer lamp module. The lamp contains an optical filter that removes essentially all but the dominant 222 nm emission wavelength. [9] [10] [11] [12] [13] [14] [15] For 222-nm UVC irradiation, the Care222 TM was placed 24 cm above the surface of the plates, and the radiation intensity at the surface of the plates was 0.1 mW/cm 2 , as measured with a recently calibrated Unimeter SNK005 meter (Ushio Inc., Tokyo, Japan). After irradiation for each treatment time, the control and treated plates were washed with 1 mL of phosphate-buffered saline (PBS) for 1 min and the virus-containing PBS was collected in a test-tube for analysis. We used the standard TCID 50 method to determine virus titer for the collected viral samples from each plate as previously described. 19 Log10 TCID 50 /mL reductions were calculated by comparing the log10 TCID 50 /mL values recovered from plates after 222-nm UVC irradiation with those from control (non-irradiated) plates. SARS-CoV-2 RNA was extracted from the collected viral samples from each plate using a The results of the TCID 50 assay and RT-qPCR are shown in Table 1 . Three mJ/cm 2 of 222 nm UVC light (0.1 mW/cm 2 for 30 s) reduced viable SARS-CoV-2 (99.7% reduction) to undetectable levels based on the TCID50 assay. In contrast, the copy number of SARS-CoV-2 RNA did not change after UV irradiation compared to that of the non-irradiation control, even after the maximum 300 s exposure (cycle quantification (Cq), 17.02-18.32). 3.03 ± 1.73  10 7 * SARS-CoV-2, Severe acute respiratory syndrome coronavirus 2; TCID50, 50% tissue culture infectious dose; SD, standard deviation † SARS-CoV-2 RNA was measured by quantitative reverse transcription PCR The results of this study demonstrated that 222-nm UVC irradiation at 3 mJ/cm 2 (0.1 mW/cm 2 for 30 s) resulted in a 99.7% reduction of viable SARS-CoV-2 based on the TCID 50 assay. Previous reports showed that a low dose of 222-nm UV inactivated several viruses, although the treated viruses were aerosolized; 2 mJ/cm 2 of 222-nm UV inactivated >95% of H1N1 aerosolized influenza virus. 14 In addition, 1.7 and 1.2 mJ/cm 2 of 222-nm UV inactivated 99.9% of aerosolized human coronaviruses alpha HCoV-229E and beta HCoV-OC43, respectively. 15 Previous report showed that the viral titer of SARS-CoV-2 was reduced on a plastic plate over time. 3 The decrease of the viral titer of the control plate by dry To protect healthcare workers and to help stop the spread of SARS-CoV-2, more thorough prevention and control practices are needed. Since COVID-19 has become a pandemic and there are asymptomatic patients, 20 attention is being focused on environmental disinfection in public spaces beyond healthcare settings in which many people come into contact with various surfaces. Although the most common UV disinfection system in use is a germicidal lamp emitting around 254-nm UVC, 8 it is well-known that this wavelength is harmful to the skin and eyes. Thus, the germicidal lamp UV disinfection systems can mainly only be used in unoccupied spaces. In contrast, recent studies have shown that 222-nm UVC is less harmful than 254-nm UVC because far-UVC light has a very limited penetration depth in the skin or eyes and it is also an efficient anti-microbial technology. 9-13 Therefore, a 222-nm UVC disinfection system could be used in occupied public spaces including a hospital room, which is occupied by patients and frequented by healthcare workers and possibly, visitors. However, further study about the safety and efficacy of 222-nm UVC irradiation in occupied space is needed. One limitation of our study is that we only investigated the in vitro efficacy of 222-nm UV irradiation on SARS-CoV-2 contamination. We did not evaluate this technology in a real-world setting, such as a surface counter-top. In conclusion, our study demonstrated the effectiveness of 222-nm UVC irradiation on viable SARS-CoV-2. Our results suggest that this technology could be used for infection prevention and control against COVID-19, not only in unoccupied spaces but also occupied spaces. Further evaluation of the safety and efficacy of 222-nm UVC irradiation in reducing the contamination of real-world surfaces and the transmission of SARS-CoV-2 is needed. Early transmission dynamics in Wuhan, China, of novel coronavirus-infected pneumonia Clinical features of patients infected with 2019 novel coronavirus in Wuhan, China Aerosol and surface stability of SARS-CoV-2 as compared with SARS-CoV-1 Persistence of coronaviruses on inanimate surfaces and their inactivation with biocidal agents Environmental contamination by SARS-CoV-2 in a designated hospital for coronavirus disease 2019 Air, surface environmental, and personal protective equipment contamination by severe acute respiratory syndrome cronavirus 2 (SARS-CoV-2) from a symptomatic patient Environmental contamination of SARS-CoV-2 in healthcare premises No-touch disinfection methods to decrease multidrug-resistant organism infections: A systematic review and meta-analysis Germicidal efficacy and mammalian skin safety of 222-nm UV light Disinfection and healing effects of 222-nm UVC light on methicillin-resistant Staphylococcus aureus infection in mouse wounds Chronic irradiation with 222-nm UVC light induces neither DNA damage nor epidermal lesions in mouse skin, even at high doses Long-term effects of 222-nm ultraviolet radiation C sterilizing lamps on mice susceptible to ultraviolet radiation Ultraviolet C light with wavelength of 222 nm inactivates a wide spectrum of microbial pathogens Far-UVC light: A new tool to control the spread of airborne-mediated microbial diseases Far-UVC light (222 nm) efficiently and safely inactivates airborne human coronaviruses Deactivation of SARS-CoV-2 with pulsed xenon ultraviolet: Implications for environmental COVID-19 control Susceptibility of SARS-CoV-2 to UV irradiation Enhanced isolation of SARS-CoV-2 by TMPRSS2-expressing cells Inactivation of pathogenic viruses by plant-derived tannins: strong effects of extracts from persimmon (Diospyros kaki) on a broad range of viruses Estimating the asymptomatic proportion of coronavirus disease 2019 (COVID-19) cases on board the Diamond Princess cruise ship This work was supported by the Japan Agency for Medical Research and Development (AMED) under grant number JP20he0922003. We would like to thank Editage (www.editage.com) for English language editing. There are no conflicts of interest to declare.