key: cord-0911052-tpw38ydy authors: Li, Tian; Pan, Rui; Wen, Yuhan; Xu, Jiaqi; Zhang, Liping; He, Suna; Liang, Gaofeng title: A Simple and Universal Nucleic Acid Assay Platform Based on Personal Glucose Meter Using SARS-CoV-2 N Gene as the Model date: 2022-04-15 journal: Biosensors (Basel) DOI: 10.3390/bios12040249 sha: 725c8264222d74e6412e27b72e93c4479e30c0a7 doc_id: 911052 cord_uid: tpw38ydy A simple, selective, and quantitative platform for point-of-care diagnostic of COVID-19 is urgently needed as a complement in areas where resources are currently relatively scarce. To meet the needs of early diagnosis and intervention, a proof-of-concept demonstration of a universal personal glucose meter-based nucleic acid assay platform (PGM-NAAP) is presented, which converts to SARS-CoV-2 detection from glucose detection. By using magnetic bead separation together with the hand-held PGM for quantitative readout, PGM-NAAP achieves the 98 pM limit of detection for a sequence related to SARS-CoV-2. The ability to discriminate target nucleic acid from genomic DNA, the satisfactory spike recoveries of saliva and serum samples, as well as the good stability all together suggest the potential of the PGM-NAAP for the screening and diagnosis of suspected patients during the outbreaks of COVID-19 in resource-limited settings without sophisticated instruments. On the basis of these findings, PGM-NAAP can be expected to provide an accurate and convenient path for diagnosis of disease-associated nucleic acid. The COVID-19 has caused more than 426 million people being infected and no less than 5.9 million deaths as of February 2022. Recent studies have shown that at least 35% of patients are asymptomatic, which greatly increases the risk of epidemic transmission [1, 2] . Since no specific medicine or treatment for COVID-19 is available, the accurate diagnosis of both symptomatic and asymptomatic cases, together with a series of prevention and control measures, including isolation and contact tracing, have become the most effective way to contain the outbreak [3] [4] [5] . Current diagnostic approaches for COVID-19 mainly include viral antigen detection and nucleic acid test [6] [7] [8] [9] . Viral antigen detection can be realized directly in untreated biological fluids without virus lysis. Recently, several electrochemical platforms and lateral flow immunoassays (LFIA) were developed for detection of SARS-CoV-2 spike protein, which are suitable for massive diagnostic testing and home self-testing [10] [11] [12] [13] [14] . Unfortunately, since proteins cannot be amplified directly, detecting trace amounts of these proteins is still a considerable challenge and the selectivity maybe not good enough to distinguish similar viruses [1] . Due to the high sensitivity and specificity, the nucleic acid test is currently the most widely used diagnostic and screening tool for COVID-19, and RT-qPCR is the gold standard in clinical practice. However, it requires expensive equipment and well-trained personnel, which have made large-scale nucleic acid tests remain a challenge. To accelerate clinical diagnostic testing for SARS-CoV-2, it is quite necessary to develop point-of-care tests (POCT) because medical systems in many parts of the world have been overloaded during the pandemic [15] [16] [17] . As some of the existing techniques cannot be applied on a large scale in many scenarios, various diagnosis and screening strategies have been established for SARS-CoV-2 [18] [19] [20] [21] [22] . Most of these sensing strategies combine nucleic acid isothermal amplification with colorimetric or lateral flow-based detection strategies in order to reduce instrument dependence [23] [24] [25] [26] , achieving only qualitative or semiquantitative results as a compromise. The personal glucose meter (PGM) is recognized as one of the most successful POCT cases due to the reliable quantitative results, simple operation, small size, and wide acceptability all over the world. Traditionally, PGM is only applied to detect glucose in blood for the diagnosis and daily monitoring of hyperglycemia based on electrochemical detection or chemical colorimetry. For the first time, Lu's group successfully used a PGM to detect non-glucose substances, which took full advantage of its simplicity and reliability [27] [28] [29] [30] . Since then, this concept has been used to detect different kinds of targets, such as metal ion [31, 32] , toxins [33] , viruses [34] , bacteria [35] , and disease biomarkers [36] [37] [38] [39] [40] . Recently, several groups have realized the detection of COVID-19 related biomarkers using the widespread, commercially available PGMs, with high sensitivity and specificity, which show good performance in clinical samples [6, 20, 41, 42] . However, these experimental designs can hardly avoid the interference of endogenous glucose in patient samples. Recently, some electrochemical biosensors combining magnetic beads, sandwich hybridization assays, and enzymatic reaction have realized fast and sensitive nucleic acid detection [43, 44] . Herein, in this work, we present a universal and quantitative PGM-based nucleic acid assay platform (PGM-NAAP). The nucleocapsid phosphoprotein gene (N gene) of SARS-CoV-2 was tested as the model, which can be recognized by the capture probes modified with invertase using sandwich-binding design. Then, the sandwich structure conjugates were separated and the invertase, which is directly related to the target concentration, could hydrolyze sucrose into glucose, thus the target could be determined by monitoring the glucose signal on the PGM. This reliable, simple, low-cost, and widely available PGM-NAAP makes it possible to carry out large-scale screening and diagnosis in communities and resource limited areas, effectively avoiding the overload of medical resources. Sucrose, Na 2 HPO 4 , NaCl, and NaH 2 PO 4 were sourced from Yongda Chemical Reagent Co., Ltd. (Tianjin, China). Tris (2-carboxyethyl)phosphine (TCEP) and diethylpyrocarbonate (DEPC)-treated water were from Sangon Biotech Co., Ltd. (Shanghai, China). Streptavidin-modified magnetic beads (10 mg/mL) were sourced from Thermo Fisher Scientific. Salmon sperm DNA and 4-(N-Maleimidomethyl)cyclohexane-1-carboxylic acid 3-sulfo-N-hydroxysuccinimide ester sodium salt (sulfo-SMCC) (>98%) were from Macklin Biochemical Co., Ltd. (Shanghai, China). Invertase (≥100 units/mg) was obtained from Yuanye Biochemical Co., Ltd. (Shanghai, China). Ultrafiltration tubes with molecular weight cut-off of 3, 30, and 50 kDa were purchased from Millipore Inc. Oligonucleotides used in this article were purchased from Sangon Biotech Co., Ltd. (Shanghai, China) and the corresponding sequences were as follows [45, 46] The PGM (ACCU-CHEK Active) was from Roche (Mannheim, Germany). The working principle of the PGM was provided in supporting information. The Probe1−invertase conjugates were synthesized as described in the previous reports [29, 38] . Thirty microliters of 100 µM sulfhydryl modified Probe1, 1 µL of 10 mM TCEP, and 2 µL of 1 M sodium phosphate buffer (pH = 5.3) was mixed and reacted at room temperature for 1 h. Then, the activated Probe1 was purified by ultrafiltration centrifuge tube (3 kDa) for eight times using buffer A (0.1 M sodium phosphate buffer, pH = 7.3). Meanwhile, 50 µL of 20 mg/mL invertase was activated with 1 mg sulfo-SMCC in buffer A at room temperature. One hour later, the insoluble sulfo-SMCC was first removed by centrifugation and the supernatant was further purified by an ultrafiltration centrifuge tube (50 kDa) for eight times using buffer A. Subsequently, the activated Probe1 and invertase were mixed and reacted at room temperature for 48 h. The product was then purified by an ultrafiltration centrifuge tube (50 kDa) eight times. Finally, the synthesized Probe1−invertase conjugates compound was diluted to 100 µL with buffer A and stored at 4 • C. The detection of target nucleic acid was first carried out in buffer A. The assay mixture consisted of 1 µL of Probe1−invertase (10 mg/mL, the concentration of invertase represent the concentration of Probe1-invertase), 100 µL of target nucleic acid with concentration as indicated, and 1 µL of Probe2. Final concentrations of Probe1−invertase, Probe2, and target DNAs were 0.1 mg/mL, 100 nM, and 100 pM-20 nM, respectively. The mixture was cooled from 37 to 25 • C in 20 min to allow the DNAs to hybridize and form the sandwich structure (Probe1−invertase and target and Probe2). Then, 8 µL of streptavidin-conjugated magnetic beads (streptavidin-MBs) were added, and the mixture was kept at 25 • C for 30 min to allow the streptavidin-MBs to capture the sandwich structure. The assemblies, along with unattached MBs, were magnetically separated and washed three times with buffer A. Subsequently, 100 µL 1 M sucrose was added, and the mixture was kept at 37 • C for 120 min. The glucose signal was finally determined using the PGM and the sample-to-result turnaround time was roughly 180 min. The test was carried out in a large amount of salmon sperm DNA. The concentration of target nucleic acid was 10 nM and the mass ratio of salmon sperm DNA to target increased from 0 to 10 4 . The detection procedures were identical as mentioned above. In order to explore the feasibility of PGM-NAAP in clinical application, we tested it in fetal bovine serum and saliva. Healthy saliva samples (~500 µL) were collected in sterile tubes from volunteers. It is suggested to avoid eating 1 h before saliva collection, and rinse the mouth with water before sample collection [13] . Fetal bovine serum and saliva, diluted 10 times with buffer A, was filtered through an ultrafiltration centrifuge tube (30 kDa) and the filtrate were spiked with different concentrations of target nucleic acid separately. The detection process was the same as mentioned above. A statistical data analysis was performed using Statistical Program for Social Sciences (SPSS) ver. 22.0 software packages. Comparisons between the dependent variables were determined by using analysis of variance (ANOVA), and Fisher's least significant difference (LSD). Results are expressed as mean value ± standard deviation (SD) of three determinations. The principle of PGM-NAAP is illustrated in Scheme 1. With the existence of target nucleic acid, a complex of "Probe1−invertase & target & Probe2" was formed through sandwich type DNA hybridization. The amount of invertase captured by magnetic beads was directly related to the target concentration. After magnetic separation and washing away of nonspecific adsorption, the bound DNA-invertase could catalyze the hydrolysis of sucrose, and the product, glucose, could be quantified easily by any untrained users with the help of a hand-held PGM. When compared with the existing SARS-CoV-2 detection methods based on PGMs [6, 20, 42] , PGM-NAAP could effectively avoid the interference of endogenous glucose in the sample, since the sandwich structure conjugates with invertase were separated from the detection system and hydrolyzed sucrose after the washing step. That is to say, the invertase on the sandwich structure conjugates, which was directly related to the target concentration, hydrolyzed sucrose into glucose in a completely new and independent system, which has nothing to do with the original sample and the interferents in it. In general, the whole detection procedures of PGM-NAAP did not need any professional instruments or well-trained operators. As a consequence, the PGM-NAAP was expected to provide a simple and useful tool for POCT of COVID-19. During the current spread of SARS-CoV-2, research shows that the RdRP gene, E gene, and N gene of SARS-CoV-2 have conserved sequences. The N gene, which has 1260 nucleotide bases, was tested for the specific detection of SARS-CoV-2. To obtain better sensing performance, five different regions of N gene (40 mers) were selected as recognition sites and their capture probes were rationally designed separately. The capture probes and the target sequence could form a sandwich structure and the binding energy was predicted by NUPACK software for high detection efficiency (Table S1 ). Finally, the fragment (601-640) which had the lowest binding energy with its capture probes was chosen as a target model. In order to obtain an optimal sensing performance, the reaction time, the amount of Probe1−invertase conjugation, the dosage of streptavidin-MBs, and incubation time were optimized. The reaction time for invertase to hydrolyze sucrose has a very important impact on the performance of PGM-NAAP. The glucose concentration in the test tube was quantified by a PGM at intervals. For 20 nM target nucleic acid, the best signal-to-background ratio (SBR) was achieved with a reaction time of 120 min ( Figure S1 ). Thus, 120 min was chosen in the following research. The amount of Probe1-invertase conjugates is a crucial factor affecting the analytical performance of PGM-NAAP. On the one hand, enough Probe1-invertase conjugates are needed to capture the target nucleic acid. On the other hand, background signal may increase at high invertase levels due to nonspecific binding between the conjugates and magnetic beads, which in turn affects the sensitivity. For 10 nM target nucleic acid, the best SBR was obtained with 0.1 mg/mL Probe1−invertase conjugation ( Figure S2) . Hence, the optimal amount of Probe1−invertase conjugates was 0.1 mg/mL. The dosage of streptavidin-MBs should be sufficient to capture and separate the sandwich structure (Probe1−invertase and target and Probe2). Five different dosages of streptavidin-MBs (0.1, 0.3, 0.5, 0.8, and 1 mg/mL) were used. As shown in Figure S3 , the largest SBR was obtained at the dosage of 0.8 mg/mL and decreased with further increases in the amount of streptavidin-MBs. Thus, the dosage of streptavidin-MBs in the following study was 0.8 mg/mL. The incubation time between the sandwich structure and streptavidin-MBs also influences the sensitivity of the PGM-NAAP. To study its effect, we detected 10 nM target nucleic acid with different incubation times, and found that the SBR initially increased from 15 to 30 min and then decreased ( Figure S4 ). Therefore, incubation time was selected as 30 min in the following study. The sensitivity of PGM-NAAP was first evaluated by measuring target nucleic acid at different concentrations under the optimized experimental conditions. The PGM signal change (∆PGM) is defined as P 0 -P 1 , where P 0 and P 1 indicate the PGM signal in the absence and presence of target sequence, respectively. The background signal (P 0 ) was (1.63 ± 0.048) mM (n = 10). As shown in Figure 1A , the ∆PGM increased with increasing concentration of target nucleic acid, and a good linear relationship (R 2 = 0.9922) existed in the range from 0.1 to 20 nM, with a detection limit of 98 pM (5.9 × 10 7 copy/µL, based on 3 σ/k, n = 10). In the selectivity test, two random interference sequences and the sequences related to severe acute respiratory syndromes (SARS), Middle East respiratory syndrome (MERS), as well as the influenza viruses such as H1N1 and H7N9 were applied to the PGM-NAAP. A high PGM signal was obtained from the sample containing target N gene. In addition, as presented in Figure 1B , the observed PGM signal in the presence of target at 2 nM (1.2 × 10 9 copy/µL) was statistically higher than those of the non-target sequence at 20 nM (p ≤ 0.013; LSD), confirming the high selectivity of PGM-NAAP. We further used salmon sperm DNA to simulate the complexity of the genome to verify the selectivity of PGM-NAAP. With the existence of 1 mg/mL salmon sperm DNA, a good linear relationship still existed between ∆PGM and the target nucleic acid concentration, with a detection limit of 0.2 nM (1.2 × 10 8 copy/µL) (Figure 2A) . Besides, the PGM signal did not fluctuate significantly as the mass ratio of genomic DNA to target increased from 0 to 10 4 . These results indicated that genomic DNA has little effect on target detection, which further confirmed the selectivity of PGM-NAAP. The ∆PGM signal was obtained with the existence of different amounts of target sequence: 0.5 nM (3 × 10 8 copy/µL), 1 nM (6 × 10 8 copy/µL), 2 nM (1.2 × 10 9 copy/µL), 5 nM (3 × 10 9 copy/µL), 10 nM (6 × 10 9 copy/µL), and 20 nM (1.2 × 10 10 copy/µL).; (B) the signal value of the blood glucose meter changes with the quality of salmon sperm DNA in the system. The mass ratio of target sequence (N-gene) to salmon sperm DNA is 1:0 to 1:10 4 . The blue bar represents the concentration of N-gene at 10 nM (6 × 10 9 copy/µL). The red bar is blank, and the dotted line is the average value of all blank signals. Error bars represent the standard deviation (n = 3). For glucose detection, 1 mM equals 18 mg/dL. The stability of an assay is very important for clinical application and diagnosis. The prepared Probe1−invertase conjugates were kept at 4 • C, and five tests were conducted in the following 20 days for the detection of 10 nM (6 × 10 9 copy/µL) target nucleic acid. No significant fluctuation was observed and the relative standard deviation of the five tests was calculated as 3.3%, indicating the good stability of both Probe1−invertase conjugates and PGM-NAAP (Figure 3) . To explore the potential of PGM-NAAP in practical application, spike and recovery assays were carried out in fetal bovine serum, which is a complex matrix containing a variety of peptides, proteins, fats, carbohydrates, growth factors, hormones, inorganic substances, and so on. The diluted serum sample was spiked with two concentration levels of target and the spike recovery was (109.4 ± 12.4)% for the 100 pM target and (107.1 ± 6.7)% for the 20 nM target (n = 3). Previous research has shown that SARS-CoV-2 virus load is high in saliva [6] , which is one of the most readily available and ideal human specimens for daily COVID-19 screening. So, we finally challenged the PGM-NAAP with the healthy saliva specimens spiked with different concentrations of target nucleic acid to prove the feasibility of the assay in a more real scenario. A calibration plot was generated with increasing target concentration from 0.1 to 20 nM (Figure 4 ). The LOD was calculated (3σ/slope) to be 89 pM (5.4 × 10 7 copy/µL) in such a complex and viscous saliva sample. These preliminary results show that the PGM-NAAP is feasible for application in biological fluids. Figure 4 . Application of the PGM-NAAP in 10% saliva. The ∆PGM signal was obtained with the existence of different amounts of target sequence: 0.1 nM (6 × 10 7 copy/µL), 0.5 nM (3 × 10 8 copy/µL), 2 nM (1.2 × 10 9 copy/µL), 10 nM (6 × 10 9 copy/µL), and 20 nM (1.2 × 10 10 copy/µL). Error bars represent the standard deviation (n = 3). For glucose detection, 1 mM equals 18 mg/dL. In summary, we have successfully developed a simple and universal PGM-based nucleic acid assay platform with a detection limit of 98 pM for a sequence related to SARS-CoV-2. Since no sophisticated equipment is required, it provides the possibility for screening and early diagnosis of diseases in resource-limited areas such as healthcare centers and community hospitals. The ability to discriminate target nucleic acid from genomic DNA, and the satisfactory spike recoveries in saliva and serum samples, together with the good stability, all suggest the potential of the PGM-NAAP for monitoring suspected patients during COVID-19 outbreaks. However, some limitations still exist for fully translating the current PGM-NAAP to a widely available TOC platform. The LOD is far from that obtained with PCR and there is no obvious advantage from the perspective of detection speed. Since the sandwich-binding design is versatile and can be easily expanded, pre-amplification should be adopted to achieve a higher sensitivity. A smartphone app, which can directly convert the PGM readout into the concentration of nucleic acid, should be developed to enhance the data progressing capability. In addition, the proposed PGM-NAAP contained several steps, which would be better to integrate into a one-pot reaction in potential by automation, minimal instrumentation, or a simplified process [47, 48] . The above-mentioned limitations could be further investigated and solved by the future work. In conclusion, considering the universal experimental design and simplicity of PGMs, the PGM-NAAP has great potential to be used as a simple, quantitative, and portable sensing platform for the detection of other disease-related nucleic acids. Supplementary Materials: The following supporting information can be downloaded at: https:// www.mdpi.com/article/10.3390/bios12040249/s1, Table S1 : Binding energy of different recognition site with their capture probes; Figure S1 : Optimization of reaction time for invertase catalysis; Figure S2 : Optimization of the amount of DNA-invertase conjugates; Figure S3 : Optimization of the amount of magnetic beads; Figure S4 : Optimization of the incubation time. Diagnostics for SARS-CoV-2 detection: A comprehensive review of the FDA-EUA COVID-19 testing landscape Occurrence and transmission potential of asymptomatic and presymptomatic SARS-CoV-2 infections: A living systematic review and meta-analysis SARS-CoV-2 asymptomatic and symptomatic patients and risk for transfusion transmission The COVID-19 Pandemic: A Comprehensive Review of Taxonomy, Genetics, Epidemiology, Diagnosis, Treatment, and Control The impact of the hypoxia-VEGF-vascular permeability on COVID-19-infected patients Hitting the diagnostic sweet spot: Point-of-care SARS-CoV-2 salivary antigen testing with an off-the-shelf glucometer COVID-19 diagnosis -A review of current methods Diagnosing COVID-19: The Disease and Tools for Detection Highly Sensitive and Specific SARS-CoV-2 Serological Assay Using a Magnetic Modulation Biosensing System Coronavirus Nucleocapsid Antigen-Detecting Half-Strip Lateral Flow Assay Toward the Development of Point of Care Tests Using Commercially Available Reagents Rapid and Efficient Detection of the SARS-CoV-2 Spike Protein Using an Electrochemical Aptamer-Based Sensor Nanozyme chemiluminescence paper test for rapid and sensitive detection of SARS-CoV-2 antigen SARS-CoV-2 RapidPlex: A Graphene-Based Multiplexed Telemedicine Platform for Rapid and Low-Cost COVID-19 Diagnosis and Monitoring Paper-Based Test for Rapid On-Site Screening of SARS-CoV-2 in Clinical Samples Diagnostic methods and potential portable biosensors for coronavirus disease 2019 Detection of COVID-19: A review of the current literature and future perspectives Ultrasensitive supersandwichtype electrochemical sensor for SARS-CoV-2 from the infected COVID-19 patients using a smartphone CRISPR-Cas12-based detection of SARS-CoV-2 Ultrasensitive and visual detection of SARS-CoV-2 using all-in-one dual CRISPR-Cas12a assay A CRISPR-Cas12a-derived biosensor enabling portable personal glucose meter readout for quantitative detection of SARS-CoV-2 Selective Naked-Eye Detection of SARS-CoV-2 Mediated by N Gene Targeted Antisense Oligonucleotide Capped Plasmonic Nanoparticles Botulinum toxin as an ultrasensitive reporter for bacterial and SARS-CoV-2 nucleic acid diagnostics Rapid molecular diagnostics of COVID-19 by RT-LAMP in a centrifugal polystyrene-toner based microdevice with end-point visual detection Colorimetric reverse transcription loop-mediated isothermal amplification (RT-LAMP) as a visual diagnostic platform for the detection of the emerging coronavirus SARS-CoV-2 One-pot visual RT-LAMP-CRISPR platform for SARS-cov-2 detection Simultaneous Dual-Gene Diagnosis of SARS-CoV-2 Based on CRISPR/Cas9-Mediated Lateral Flow Assay Portable Detection of Melamine in Milk Using a Personal Glucose Meter Based on an In Vitro Selected Structure-Switching Aptamer Using the widely available blood glucose meter to monitor insulin and HbA1c Using personal glucose meters and functional DNA sensors to quantify a variety of analytical targets An invasive DNA approach toward a general method for portable quantification of metal ions using a personal glucose meter Enhanced 3D paper-based devices with a personal glucose meter for highly sensitive and portable biosensing of silver ion A portable and quantitative biosensor for cadmium detection using glucometer as the point-of-use device Portable Aptasensor of Aflatoxin B1 in Bread Based on a Personal Glucose Meter and DNA Walking Machine Detection of Enzymes, Viruses, and Bacteria Using Glucose Meters Portable and quantitative point-of-care monitoring of Escherichia coli O157:H7 using a personal glucose meter based on immunochromatographic assay Interfacing a Personal Glucose Meter with Cell-Free Protein Synthesis for Rapid Analysis of Amino Acids Enzyme-Encapsulated Liposome-Linked Immunosorbent Assay Enabling Sensitive Personal Glucose Meter Readout for Portable Detection of Disease Biomarkers Bioinspired DNA-Inorganic Hybrid Nanoflowers Combined with a Personal Glucose Meter for Onsite Detection of miRNA Liposome-encoded magnetic beads initiated by padlock exponential rolling circle amplification for portable and accurate quantification of microRNAs Biocomputing for Portable, Resettable, and Quantitative Point-of-Care Diagnostics: Making the Glucose Meter a Logic-Gate Responsive Device for Measuring Many Clinically Relevant Targets A glucose meter interface for point-of-care gene circuit-based diagnostics Translating daily COVID-19 screening into a simple glucose test: A proof of concept study Detection of Ostreopsis cf. ovata in environmental samples using an electrochemical DNA-based biosensor Electrochemical biosensor for the dual detection of Gambierdiscus australes and Gambierdiscus excentricus in field samples. First report of G. excentricus in the Balearic Islands Rapid Differential Diagnosis of Seven Human Respiratory Coronaviruses Based on Centrifugal Microfluidic Nucleic Acid Assay Detection of SARS-CoV-2 and Its Mutated Variants via CRISPR-Cas13-Based Transcription Amplification Isothermal Amplification and Ambient Visualization in a Single Tube for the Detection of SARS-CoV-2 Using Loop-Mediated Amplification and CRISPR Technology A rapid and ultrasensitive dual detection platform based on Cas12a for simultaneous detection of virulence and resistance genes of drug-resistant Salmonella The authors declare no conflict of interest.Biosensors 2022, 12, 249