key: cord-0909768-xfpeo0yl authors: Wang, Ying; Wang, Geng; Zhang, Lulu; Ren, Lili; Wang, Jianwei title: The sampling efficiencies of commercial nasopharyngeal swabs date: 2022-03-11 journal: Biosaf Health DOI: 10.1016/j.bsheal.2022.03.007 sha: a1da9a46f519eb905a7cf8004e1b6e4d82246e70 doc_id: 909768 cord_uid: xfpeo0yl Upper respiratory tract samples are the most commonly used samples for coronavirus disease 2019 (COVID-19) diagnosis. The samples collected from the nasopharynx are preferred for viral nucleic acids detection. Commercial nasopharyngeal swabs are the major factor that influences the sampling quality. We here evaluated the acceptability and efficiency of nasopharyngeal swabs from five manufacturers by examining the concentration of glyceraldehyde-3-phosphate dehydrogenase gene (GAPDH) retrieved from the swabs using the RT-PCR method. Significant different concentrations of GAPDH were detected, ranged from 4.36×10(8)copise/mL to 6.98×10(10)copise/mL among the five swabs (p<0.05). The designation of the swab head, with or without tip expansion, had limited influence on the collection efficiency. The discrepancy among the nasopharyngeal swabs emphasized the improvement of the swab head material. The outbreak of coronavirus disease 2019 (COVID-19) caused by severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) epidemic has ravaged the world, compelling the World Health Organization (WHO) to declare it as a global pandemic on 11 March 2020, caused more than 263 million confirmed cases and 5.23 million deaths of 3 December 2021 [1] . The laboratory-based viral nucleic acid testing is the principal evidence for early confirmation of SARS-CoV-2-infected cases. The sampling procedures are the main factors influencing the sensitivities of nucleic acid testing, including anatomic sampling site, transportation, transient storage conditions, etc. Respiratory tract samples for viral diagnostic tests could be taken from the upper respiratory tract, such as nasopharyngeal swabs (NPS), oropharyngeal swabs and saliva, or lower respiratory tract, including those of the lower respiratory tract sputum and bronchoalveolar lavage. The lower respiratory tract specimens are commonly collected in symptomatic or severe COVID-19 cases [2] [3] [4] . For patients with mild symptoms or population nucleic acids screening, upper respiratory samples are preferred to be collected. NPS is recommended for SARS-CoV-2 screening to acquire more reliable results than saliva, nasal and oropharyngeal swabs [2, [5] [6] [7] [8] . The currently available commercial consumables of NPSs are significant factors influencing the sampling quality. The designation of the swab is invariance on the shape, diameter, length of the swab head. Flocking material is commonly used on swab heads. As the nasopharynx is very sensitive, the unsuitable sampling process may cause tissue damage, bleeding, or irritation. The comfortability of NPSs sampling would also be considered. This study evaluated the retention efficiency of five widely used commercial NPSs; one was COPAN flocked swab, the other four swabs were manufactured in China. The designation, individual comfortability, and cost-effectiveness are also compared. Five brands of NPSs, with the code numbers 1, 2, 3, 4, 5, were selected. Code 1 4 was COPAN flocked swab (Cat:503CS01, Brescia, Italy). Code 2 to 5 were manufactured in China, with catalog numbers MSF-96000BQ, BZ0302-1, ST001-2, CY-96000 in order. All the heads of the swabs were flocked. The characteristics of NPSs were recorded, including the shape, diameter, and length of the swab head, breakpoint from the head, and expansion of the swab head. A total of 33 volunteers were randomly recruited, and they had no known respiratory tract symptoms or nasal and pharyngeal illness. The median age of the participants was 26 (21-48) years old (interquartile range [IQR]); 18 (55%) were female. The sampling procedure was performed by the same person trained well on samples collection. The swab was inserted randomly into either right or left naris, then gently rotated for several seconds at the site of the nasopharynx, then kept in the viral transport medium (VTM). Five kinds of nasopharyngeal swabs for each volunteer were tested, alternating the two nostrils at one-hour intervals. To comfortability of each swab was evaluated as a score, which included pain (usually-, slightly+, very painful++), soreness (usually-, slightly+, very painful++), and causing tear (no-, less+, tears++). Volunteers selected the most comfortable and uncomfortable swabs then summarized the frequency for each swab. The efficiency of sampling with different NPSs was evaluated by testing the concentration of the glyceraldehyde-3-phosphate dehydrogenase gene (GAPDH) retrieved from the swabs by using the PCR method. A total of 400µl VTM extracted nucleic acid using the chemagic™ 360 instrument (PerkinElmer, USA). The primers and probe for detecting GAPDH were used as follows: the forward primer GAPDH copy numbers were measured as surrogates of nucleic acid recovery and were quantified using the GAPDH RNA standard curve. Comparisons between groups were made by using t-tests. Two-sided p <0.05 was considered statistically significant. All statistical analyses were conducted by using GraphPad Prism software version 9.0. The commercial NPSs consisted of flocked swab heads plastic rods with a breakpoint. A significant red line was marked on the breakpoint of Code 1 plastic rod, and the others had no marker on the breakpoint (Figure 1 ). The parameters, as well as the price for each swab, were summarized in Table 1 . The NPSs showed different appearances on total length, length of the breakpoint from the head, head length, head diameter, and expansion of the swab head. Code 1 was the easiest to break at the breakpoint, followed by Code 5 and Code 3, Code 2, and Code 4. The softness of the plastic rod was tested, and Code 1 swab was the best, followed by Code 5, Code 2, Code 4, and Code 3. The volunteers decided the testing order of the swabs in a single-blinded way. After all the swabs were finished, the volunteer decided on comfortability. All the volunteers (39.39%) complained code 2 was the most uncomfortable one, while code 5 was the most comfortable. The comfortability of the NPSs varied considerably ( Table 2) . The shape and tip expansion of the nasopharyngeal swab head had a limited impact on the comfortability of different codes ( Table 1 ). Nucleic acid was extracted from the swabs, and the concentrations of GAPDH (copies/mL) were tested using RT-PCR to evaluate the sampling efficiency of the 6 NPSs. As shown in Figure 2 , the concentration of GAPDH for Code 1 and Code 2 swabs were significantly higher than Code 5 (p=0.025 and p=0.036), indicating retrieving efficiency of Code 1 and Code 2 swabs. The fluctuation range of the GAPDH concentrations displayed a stable collection efficiency of the Code 1 swab. For the Code 2 swab, a significant GAPDH concentration fluctuation was observed, indicating an unstable collection efficiency. Our findings suggested that the head material of the swabs should be the most influential factor in sampling efficiency. During the control and prevention of COVID-19, swabs are the most commonly used materials to collect samples and screen the pathogens. Current knowledge considers that the nasopharyngeal swab's sensitivity is higher than that of the oropharyngeal swab in the early identification of the COVID-19 case [2, 5, 6, 10] . The nasopharyngeal swab contains flocked swabs and traditional fiber swabs, and the flocked swabs are better than conventional fiber swabs on sampling efficiency [11, 12] . In this study, we compared the comfortability, and sampling efficiency of the commercial flocked NPSs. The sampling procedure was performed by one person who had been trained well in case of the sampling bias to eliminate human factors during sampling. The volunteers without symptoms of respiratory tract infections were randomly recruited. Each volunteer was informed only swabs code number, and they decided the sequence of the swabs and the inserted naris when sampling. The results showed that four of the NPSs had similar retrieving efficiency according to the concentration of GAPDH. In contrast, one domestic manufacturer showed lower and unstable sampling efficiency than others. During the outbreak of COVID-19, swabs were the most commonly used consumables. Three dimensions printed NPSs have been commercially available to support the prevention and control of COVID-19 [13, 14] . The preclinical evaluation of NPSs includes swab design, collection sufficiency, and PCR compatibility [14] . Our findings showed that the shape tip expansion of the swab head had a limited impact on the collection efficiency based on the concentration of GAPDH retrieved from the swabs. However, the optimization of swab head material should be considered. 7 NPS sampling is a critical step for viral nucleic acids detection. Non-qualified sampling may make diagnostic errors, especially when patients have a very low viral load [15] . When performing nasopharyngeal sampling, rotation of the swab upon contact with the nasopharynx has no significant improvement for sampling quality; rotation is also less tolerable for participants [16] . If the insertion depth is not enough, mid-nasal or mid-turbinate samples but not NPS would be collected, potentially reducing the sensitivity of detections [17] . The sampling procedure using NPS was easy to handle, and the comfortability was acceptable. However, the inconsistent sampling efficiency of NPSs from five manufacturers indicated an urgency to standardize the evaluation and manufacture procedure. The discrepancy among the nasopharyngeal swabs emphasized the improvement of the swab head material. The main limitation of our study was the small sample size and limited kinds of swabs. Therefore, further evaluation on the quality of the consumable of NPSs should be performed based on a larger scale sample size. The Ethical Review Committee approved this study at the Institute of Pathogen World Health Organization, WHO Coronavirus (COVID-19) Dashboard Detection of SARS-CoV-2 in different types of clinical specimens Positive rate of RT-PCR detection of SARS-CoV-2 infection in 4 Negative nasopharyngeal and oropharyngeal swab does not rule out COVID-19 Comparison of nasopharyngeal and oropharyngeal swabs for SARS-CoV-2 detection in 353 patients received tests with both specimens simultaneously SARS-CoV-2 viral load in upper respiratory specimens of infected patients Effect of throat washings on detection of Sensitivity of nasopharyngeal swabs and saliva for the detection of Severe Acute Respiratory Syndrome Coronavirus 2 Simultaneous typing and HA/NA subtyping of inflfluenza A and B viruses including the pandemic influenza A/H1N1 2009 by multiplex real-time RT-PCR Laboratory diagnosis and monitoring the viral shedding of SARS-CoV-2 infection Direct PCR amplification of DNA from human bloodstains, saliva, and touch samples collected with microFLOQ® swabs Comparison of nasopharyngeal nylon flocked swabs with universal transport medium and rayon-bud swabs with a sponge reservoir of viral transport medium in the diagnosis of paediatric influenza Prospective clinical validation of 3D printed nasopharyngeal swabs for diagnosis of COVID-19 Open development and clinical validation of multiple 3D-printed Nasopharyngeal collection swabs: rapid resolution of a critical COVID-19 testing bottleneck SARS-CoV-2 RNA identification in nasopharyngeal swabs: issues in pre-analytics Evaluation of nasopharyngeal swab collection techniques for nucleic acid recovery and participant experience: recommendations for COVID-19 diagnostics Variability in instructions for performance of nasopharyngeal swabs across Canada in the era of COVID-19 -what type of swab is actually being performed? We want to thank the volunteers who contributed to the sample collection. This The authors declare that there are no conflicts of interest.