key: cord-0905241-btnz5jru authors: Ha, Yeonjeong; Kim, Yoonsub; Song, Eugene; Yoo, Hyun Jung; Kwon, Jung‐Hwan title: Development of a personal passive air sampler for estimating exposure toeffective chlorine while usingchlorine‐based disinfectants date: 2020-09-26 journal: Indoor Air DOI: 10.1111/ina.12747 sha: a22c535b0e732e9936930dbcf61a941abec59ec1 doc_id: 905241 cord_uid: btnz5jru With an increasing use of indoor disinfectants such as chlorine (Cl(2)) and hypochlorous acid (HOCl), a convenient sampler for estimating exposure to oxidants, such as effective chlorine, is necessary. Here, we developed a personal passive air sampler (PPAS) composed of a redox dye,o‐dianisidine, in a polydimethylsiloxane(PDMS) sheet. o‐Dianisidine readily reacts with gaseous oxidants generated by bleach usage, and its colorchanges as the reaction progresses; hence, personal exposure to effective chlorinecould be easily detected by the naked eye,while cumulative exposure could be determined by measuring concentrations of o‐dianisidine reacting with it. The PPAS was calibrated,and asamplingrate of 0.00253 m(3)h(−1)was obtained using a small testchamber. The PPAS was tested with the help of ten volunteers whosepersonal exposure toCl(2)‐equivalent gas was estimated after bathrooms were cleaned using spray and liquid‐type household disinfection products, and the accumulated exposure‐gas concentrations ranged from 69 to 408 ppbv and 148 to 435 ppbv, respectively. These PPAS‐derived exposure concentrations were approximatelytwoorders lower than those estimated using ConsExpo(TM), suggesting a significant overestimation by prevailing screeningmodels,possibly due to the ignorance of transformation reactions. Bleach,a hypochlorite (OCl  ) solution, is the most common chlorinated disinfectant used in homes and workplacesglobally. As the hypochlorite solution (hereafter termed bleach) is a good oxidizing agent and effectively inactivates a broad spectrum of microorganisms, 1 it has been extensively used as an antimicrobial agent for cleaning surfaces, bleaching laundry, disinfecting food, and sanitizing drinking-water and swimming pools. However its usehas raised concerns associated with adverse respiratory health effects due to the generation of reactive chlorine and their gaseous byproducts (e.g., chloroform and carbon tetrachloride) in indoor air. [2] [3] [4] [5] Two important effective chlorine species from bleach arechlorine (Cl 2 ) and hypochlorous acid (HOCl); recent studies have reported that concentrations of both thesegases in indoor air significantly increased while using commercial bleach, with concentrationsof Cl 2 and HOCl reaching 20-25 and 50-250 of ppbv, respectively. 6, 7 Acute inhalation of these gasesgenerally causehealth problems such as reactive airways dysfunction syndrome (RADS)owing to Cl 2 gas generation from bleach use, especially when bleach mixes with acidic solutions such as hydrochloric acid (HCl) and vinegar. [8] [9] [10] Even at low exposures (1-15 ppmv) , Cl 2 gas could cause mucous membrane irritation.At 30 ppmv of Cl 2 exposure, cough, substernal chest pain, and shortness of breath could occur. 11 Inhalation toxicity of HOCl gas has not yet been reported; however, being a good oxidant,it could damage mammalian-cells by oxidizing glutathione (GSH) and protein thiols. 12 In addition, once Cl 2 and HOCl gas are liberated, OH and Cl radicalscould be generated by photolysis, thus increasing the oxidizing capacity of indoor air. 6 The coronavirus disease is an ongoing global pandemic, caused by the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). 13 This article is protected by copyright. All rights reserved centers received a spike in calls (45,550) related to cleaners and disinfectants during January-March, 2020. 15 Bleaches indicated the largest percentage increase (62.1%) among all disinfectants from 2019 to 2020, while inhalation accounted for the largest percentage increase (108.8%) for all disinfectants among all exposureroutes. 15 Disinfection could become increasingly important for preventing infections caused by various viruses and bacteria,while bleach has been reported as the most likely cause of accidental inhalation poisoning; hence, estimating personal exposure to oxidant gases generated from bleaches isan urgent requirement. To address this requirement, a personal passive air sampler (PPAS) was developed for estimating personal exposure levelsto reactive chlorine gasesresulting from disinfection using bleaches. The PPAS has been considered as a powerful tool for sensitive personal monitoring and estimating timeweighted average concentrations of individual chemical exposure to various gaseous compounds. [16] [17] [18] [19] The PPAS developed in this studyconsists of two components: (i) polydimethylsiloxane (PDMS) as a permeable medium and (ii) a redox dye, o-dianisidine, contained with in it. PDMS has been successfully used as a main component of passive air samplers in numerous studies as gas molecules easily permeatethrough theirpores. [20] [21] [22] [23] [24] Previous studies have assessed the exposure to volatile organic compounds (VOCs) or semi-volatile organic compounds (SVOCs) such as n-alkanes, aromatic hydrocarbons and phthalates;however, studies related to passive air samplers that could estimate exposure levels of reactive gases, especially chlorine gasesgenerated by bleach use,have not been reported. o-Dianisidine was chosen as a dye that changes color rapidly and stably onoxidation. The objectives of this study are (i) to develop and calibrate a PPAS for estimating personal Cl 2 -equivalent gasexposure from bleach use and (ii) to conduct a panel study to estimate the personal exposure levels of chlorinated gases when consumers clean residential bathrooms using chlorinated disinfectants. In addition, we compared the exposure levels estimated by the PPAS to that calculated using a screening model, ConExpo TM . 25 This article is protected by copyright. All rights reserved This article is protected by copyright. All rights reserved Responses of PTR/SRI-MS toCl 2 gas were calibrated using standard chlorine gas (4.7 ppmv) purchased from RIGAS (Daejeon, Republic of Korea). This standard Cl 2 gas was injected continuously at aflow-rate of 5.0Lmin 1 into a custom-made 125 L acrylic chamber at 23-24 C. Signals for Cl 2 gas were detected at six intervals (1, 5, 9, 12, 16 (h).The sampling rate (R s ) of the passive sampler was obtained from the slope of a plot, EV vs.t. To estimate the personal exposure levels of chlorine gas during household disinfection using bleach, we recruited a panel of ten volunteers (five fromGyeongsan-si,Gyeongsangbuk-do, Republic of Korea, and five fromGimhae,Gyeongsangnam-do, Republic of Korea), who were asked to wear the PPAS This article is protected by copyright. All rights reserved patch while they cleaned bathrooms using disinfectants. This panelstudy was approved by the institutional review board (IRB) of Korea University (Approval number: KUIRB-2019-0175-03). The panelstudy was conducted in February 2020. The package per participant included two PPAS patches[ Figure 1 The panel experiments were performed over two days: on day one, each panelmember cleaned their bathroom(s) using one type of bleach (e.g., spray-type); on day two, they cleaned their bathroom(s) similarlybut using the other type of bleach (e.g., liquid-type) wearing a PPAS patch [ Figure 1 After completing the process, each panel-member was asked to record the type of disinfectant used, cleaning-time, bathroom-area, and whether they turned on the air ventilation systems. Pictures were taken by each panelmember after the process to check for color changes of the PPAS; each passivesampler was stored in the glassvialcontaining 10 mL isopropyl alcohol. All the used bleach products and PPAS containing glassvials were sent back to the laboratory for measuringthe amounts of disinfectants used and for an instrumental analysis (which were conducted after methods described in section 2.1.2) of passive-sampler extraction, respectively. To estimate the Cl 2 equivalent exposure, we first calculated the equivalent mass of reacted Cl 2 gas with o-dianisidine in the passivesamplers (M sampled ) that each panelmember used, using equation (2): This article is protected by copyright. All rights reserved where,C i and C f are initial and final o-diansidine concentrations in 10mL extraction solvents (isopropyl alcohol), respectively (gL 1 ); V E is the volume of extraction solvent (0.01 L); and M w is the molecular weight of o-dianisidine (244.30 gmol 1 ).A factor of 0.5 was applied to this equation, since a molar ratio of o-dianisidine to Cl 2 was 1:2 in the oxidation reaction shown in Figure 2 . Then, the time-weighted average concentration equivalent to Cl 2 gas (C v , mol Cl 2 m 3 ) was calculated (1)], andt is cleaningtime of each panel (h).Unit conversion of C v frommol Cl 2 m 3 to ppbvwas conducted applying the ideal gas equation. This article is protected by copyright. All rights reserved To optimize the PPAS, we comparedpassivesamplers using two differentPDMS-thicknesses (0.22 mm and 0.55 mm) and varying o-dianisidine doping concentrations in tolueneranging from 0.12 gL 1 to 2.0 gL 1 . Figure S1 shows the sampling-rate of the passive air sampler was determined as 0.00253 m 3 h 1 using equation (1). Since thepassive-sampler patches were prepared carefully to prevent any oxidation of o-dianisidine before use,the concentrations of o-dianisidine in the passivesamplers was confirmedto not have changedfortwelve days( Figure S4 ). The usage amount, bathroomarea, and time for cleaning bathrooms using a spray-type/liquid-type bleach product are summarized in Table 1 that the median estimated exposure concentrations using spray-type and liquid-type products were almost similar( Figure 5 ). However, variations of Cl 2 exposure between the ten panelmembers were higher whileusing spray-type products,rather than liquid-typeproducts, suggesting that when panelmembers used spray-type bleaches,gasexposure depended more on personal cleaning habits and bathroom conditions. Even though the usage amounts of both spray-type and liquid-type bleaches variedamongthe panelmembers, personal Cl 2 -equivalent gasexposure was almost independentof the bleach usage This article is protected by copyright. All rights reserved amounts ( Figure 6 ). Hence, other factors such as the gas removal rate via chemical reactions, air ventilation, and drainage during cleaning could be more important than the usage amount. Here, we compared the personal exposure levels of gases (C V ) obtained through the panelstudy with Cl 2 gas concentrations from the chamber experiment for calibrating a passivesampler (C air = 1,508 ppbv) and estimated C v values usingthe ConsExpo TM model ( Figure 6 ). To run ConsExpo TM , we used an exposure to vapor model and an increasing release-area mode; all other parameters haddefault values, except that for the amount of solution used (average values of bathroomarea and usage times of the ten panelmembers listed in Table 1 were used to derivethe room volume andduration of exposure, respectively; a ventilationrate of 10 h 1 was used). The exposure-gas concentrations estimated using ConsExpo TM increased with increasing amounts of bleaches used ( Figure 6) ; the values obtained were approximately100 times higher than the estimated C v values obtained using PPAS, even 1.5 to 30 times higher than the C air values obtained from the chamber study. ConsExpo TM is a higher-tier model for estimatinginhalation exposure from consumer products. The reported differences between ConsExpo TM -derived air concentrationsand experimental measurements werewithin a factor of ten for some consumer products, including transfluthrin 34 and silver nanoparticles(AgNP) 25 . However, this model highly overestimatesthe exposure level,especially for reactive gas compounds such as chlorine,as various removal mechanisms such as gas phase reactions and reactive uptake on the surfaces are not considered. In addition, the predicted exposure levels for consumer applications far exceed the NRC standard of 500 ppbv. Thus, the PPAS developed in this studycould greatly improve the assessment of exposure conducted by thesehigher-tier exposure models. We developed a cost-effective and user-friendly PPASfor estimating time-weighted personal exposure levelsto chlorine gas compounds from the use of household bleaches.After determining the samplingrate ofthe PPAS, a panelstudy was conducted to assess Cl 2 -equivalent gas-exposure by cleaning residential bathrooms using spray-and liquid-type bleaches.These time-weighted average exposure levels could be compared to the exposure limits of occupational and consumer products, providing more specific andrelevant criteria for guidelines for chlorinated gas exposure from bleach This article is protected by copyright. All rights reserved use. In addition to estimation ofexposure levels, the PPAS changes its color onexposure to chlorinated gases, and the colorimetric responses variedwith an increase in exposure concentrations. Thus, this PPAS could find applicationsas a versatile sensor in preventing accidental inhalation poisoning from bleach usage. Occupational exposure levels to chlorine gases arising from disinfection of public places such as hospitals,schools, and restaurantswhich is a priority for preventing infections could also beeasily assessed using the PPAS without expensive instruments. This article is protected by copyright. All rights reserved Table 1 was used here. Boxplots show the 10th, 25th, median, 75th, and90th percentiles. Table 1 was used here. Box plots show the 10th, 25th, median, 75th, and 47 90th percentiles. This article is protected by copyright. All rights reserved Accepted Article Uses of inorganic hypochlorite (bleach) in health-care facilities Asthma, chronic bronchitis, and exposure to irritant agents in occupational domestic cleaning: a nested case-control study Short-term respiratory effects of cleaning exposures in female domestic cleaners Domestic use of hypochlorite bleach, atopic sensitization, and respiratory symptoms in adults Halogenated volatile organic compounds in chlorine-bleach-containing household products and implications for their use Observations and impacts of bleach washing on indoor chlorine chemistry Multiphase chemistry controls inorganic chlorinated and nitrogenated compounds in indoor air during bleach cleaning Household bleaches based on sodium hypochlorite: review of acute toxicology and poison control center experience Severe asthma and ARDS triggered by acute shortterm exposure to commonly used cleaning detergents Reactive airways dysfunction syndrome in housewives due to a bleach-hydrochloric acid mixture Chlorine gas inhalation: human clinical evidence of toxicity and experience in animal models Living with a killer: the effects of hypochlorous acid on mammalian cells -2019)-and-the-virus-that-causes-it. 2020. 14. World Health Organization (WHO). Considerations for quarantine of individuals in the context of containment for coronavirus disease (COVID-19): interim guidance Cleaning and Disinfectant Chemical Exposures and Temporal Associations with COVID-19-National Poison Data System Development of a passive sampler to measure personal exposure to gaseous PAHs in community settings Silicone wristbands as personal passive samplers Preparation and performance features of wristband samplers and considerations for chemical exposure assessment Field evaluation of a passive personal air sampler for screening of PAH exposure in workplaces Polydimethylsiloxane (silicone rubber) brooch as a personal passive air sampler for semi-volatile organic compounds Passive air sampling of flame retardants and plasticizers in Canadian homes using PDMS, XAD-coated PDMS and PUF samplers Calibration of polydimethylsiloxane and polyurethane foam passive air samplers for measuring semi volatile organic compounds using a novel exposure chamber design Polydimethylsiloxane-based permeation passive air sampler. Part I: Calibration constants and their relation to retention indices of the analytes Polydimethylsiloxane-based permeation passive air sampler. Part II: Effect of temperature and humidity on the calibration constants Comparison of modeled estimates of inhalation exposure to aerosols during use of consumer spray products Data available on request from the authors This article is protected by copyright. All rights reserved panelstudy, the redline represents measured chlorine gas concentrations (C air ) from the chamber study described in section 2.2.2, and the blue-line indicates estimated C v values using ConsExpo TM . The errorbar indicates standard deviations of three replicates. Table1.Summary of the panelstudy using(a) spray-type and (b) liquid-type chlorinated disinfectant.(a) Spray-type disinfectant † Passive air sampler (PAS) color changes after cleaning bathrooms. (O: color changes vividly, X: color does not change,asverifiedby the naked eye). Figure S6