key: cord-0893200-0fd6n1h5 authors: Koçak, Berna; İpek, Yeliz title: Electrochemical Detection of Ivermectin Used for the Treatment of COVID-19 with Glutardialdehyde-Modified Glassy Carbon Electrode date: 2022-01-05 journal: Electrocatalysis (N Y) DOI: 10.1007/s12678-021-00704-4 sha: dcaed98a67e452553f0ff6bfce4c0f8f256f2b5d doc_id: 893200 cord_uid: 0fd6n1h5 In this study, the surface of the glassy carbon electrode was modified with glutardialdehyde. The modified glassy carbon electrode showed electrocatalytic activity against ivermectin. The glassy carbon electrode modified with glutardialdehyde showed high sensitivity, selectivity, and stability in the determination of ivermectin. The peak current of glutardialdehyde oxidation obtained by differential pulse voltammetry decreased inversely with the ivermectin concentration. Ivermectin inhibited the oxidation reaction of glutardialdehyde and caused a decrease in current. This change made the analysis of ivermectin electrochemically possible. In order to demonstrate the applicability of the developed method in real samples, recovery studies were carried out in tap water and urine. The highest sensitivity (0.45 µA/((µmol·L(−1))(cm(2)))) was achieved with urine sample and the lowest detection limit as 2.66 × 10(−6) mol·L(−1) was obtained with BRT solution sample. [Image: see text] The COVID-19 outbreak first appeared in the city of Wuhan, in Hubei Province of China, in early December 2019, according to the World Health Organization (WHO). By spreading to Europe, the USA, and other continents, it has deeply affected international political and economic relations as well as the world's health system [1, 2] . In order to prevent this disease, various vaccines have been developed and vaccination processes have started rapidly. In addition, various studies have been carried out on treatment methods for patients. COVID-19 is caused by the betacoronavirus that causes pneumonia by affecting the lower respiratory tract, called SARS-CoV-2 [3] . SARS-CoV-2 is considered as a relative of Middle East respiratory syndrome (MERS) and severe acute respiratory syndrome (SARS). The betacoronavirus is a positive single-stranded RNA virus [4] . Ivermectin (IVM) is a FDA-approved broad-spectrum anti-parasitic agent, and recent studies have demonstrated antiviral effects on many RNA and DNA viruses, including SARS-CoV-2 [2, [5] [6] [7] [8] [9] . Based on drug safety sheet data for ivermectin (New Drug Application Identifier: 50-742/S-022), possible side effects are as follows: increase in alanine aminotransferase and aspartate aminotransferase (2%), headache (6%), dysmenorrhea (5.5%), leukocyte count decrease (3%), dizziness (3%), itching (3%), peripheral edema (3%), tachycardia (3%), nausea (2%), upper respiratory tract infection symptoms (1.8%), and diarrhea (1.8%) [10] . In addition, ivermectin, which is frequently used by cattle breeders due to its effectiveness against parasites, can accumulate in fat and muscle tissue over time due to its lipophilicity [11] . When meat obtained from cattle is consumed by humans, it is important to analyze the accumulation of ivermectin in its structure. If the ivermectin concentration in beef is higher than the maximum allowable residue limit (100 μg·kg −1 ), the beef is considered as contaminated and may cause embryotoxicity [12] . Various studies have been conducted on residue analyzes of ivermectin in blood plasma [13, 14] and urine [15] . For the determination of IVM, liquid chromatography-tandem mass spectrometry [14, [16] [17] [18] [19] [20] [21] [22] , high-performance liquid chromatography [23] [24] [25] [26] , and amperometric methods [27] were used. Difficulties in implementation of the chromatographic methods, such as need for specialists, need for expensive devices, high solvent consumption, and the complex preprocessing treatments, made electrochemical sensor applications more attractive. In this study, the electrochemical analysis methods which are more cost-effective, have low sensitivity, and have less solvent consumption were applied for detection of IVM residual in BRT buffer solution, in tap water, and in urine sample. Glutaraldehyde or glutardialdehyde is used as a cross-linker for electrode modification in electrochemical sensor studies [28] [29] [30] and allows the sensor material to be immobilized to the electrode surface. İpek [31] used glutaraldehyde as a sensing element for the analysis of carbaryl pesticide. Glassy carbon electrode (GCE) was modified with glutardialdehyde and GA 2 /GCE electrode was achieved for IVM detection. Electrochemical studies were performed with the Ivium pocketstat potentiostat/galvanostat impedance analyzer. During the experimental studies, ultrapure water from the Millipore direct-3QV brand device was used. pH measurements were carried out with the Ohaus starter 3000 Bench pH meter. In voltammetric studies, glassy carbon electrode (GCE, CHI104) and GCE modified with glutardialdehyde electrode (GA 2 /GCE) were used as working electrodes in separate measurements. Ag/AgCI (3 mol·L −1 NaCI) electrode was utilized as reference electrode (MF-2052, BASI) and as counter electrode Pt wire was used. Ivermectin (IVM, CAS no: 70288-86-7) determined during the study was obtained from Sigma-Aldrich company. The stock solution of 1 × 10 −3 mol·L −1 IVM was prepared in methanol and stored at + 4 °C. Glutardialdehyde (GA 2 , 25%) which was used in the modification of the electrode surface was obtained from Merck company. In addition, all chemicals (sodium hydroxide (NaOH), hydrochloric acid (HCl), boric acid (H 3 BO 3 ), acetic acid (CH 3 COOH), and phosphoric acid (H 3 PO 4 )) which were used for the Britton-Robinson buffer (BRT) solution preparation were of analytical purity and were obtained from Merck company. The surface of the GCE was cleaned and polished using aluminum oxide (alumina, Al 2 O 3 ) with 0.05 µm particle size. Glassy carbon electrode was sonicated in an ultrasonic bath in deionized water for 10 min in order to clean the alumina residues remaining on its surface. GA 2 /GCE electrode was obtained as a result of dropping 5.0 µL of 5% glutardialdehyde (GA 2 ) solution onto the cleaned electrode surface and drying it at room temperature and conditions for 2 h. Glassy carbon electrode modified with GA 2 (GA 2 /GCE) was used as the working electrode. In order to increase the solubility of ivermectin, methanol was added to the solution medium at 25% of the volume of the buffer solution. Before adding ivermectin to the phosphate buffer, differential pulse voltammograms of the buffer were recorded. Then, voltammetric measurements were made by adding ivermectin with known concentration to the electrolytic buffer solution. The optimized operating conditions for differential pulse voltammetry (DPV) measurements of ivermectin are as follows: pH value of 12.02 for phosphate buffer as supporting electrolyte, pulse time: 10 ms, pulse amplitude: 200 mV, scan rate: 50 mV s −1 , and Estep: 2 mV. Modified electrode surface was investigated with a stereo microscope equipped with Sony IMX485 sensor, coaxial lamp, and Nikon monocular C-Mount lens. The achieved optical images were given in Fig. 1 . A homogenous, uniform thin film with a thickness about 40 µm was observed with the microscope. After preparation of modified electrode, electrochemical measurements were applied with differential pulse voltammetry (DPV) technique. DPV images of GCE and modified GA 2 /GCE electrode were exhibited in Fig. 2 . After a smooth and clean DPV voltammogram had been achieved with GCE electrode, it was modified with GA 2 . GA 2 /GCE electrode had two oxidation peaks at 0.9 V and 1.45 V as O1 and O2 peaks, respectively. Anodic behavior of 2.85 × 10 −4 mol·L −1 ivermectin in BRT solution at different pH values (pH 1.99, pH 5.01, pH 7.02, pH 9.01, pH 11.02, pH 12.02) was investigated with GCE and DPV technique. The highest anodic peak current was obtained at pH 12.02, as shown in Fig. 3 . Thus, pH 12.02 BRT buffer solution was used throughout the electrochemical detection studies. Figure 3 (a) shows the change in the current value of the ivermectin peak detected using GCE with the change in the hydrogen concentration of the medium. Accordingly, pH 9 was a critical value, and an increasing trend in the peak current was observed after pH 9. The maximum peak current was obtained when the pH value of the BRT solution used was 12.02. In Fig. 3(b) , the changes in the peak potential simultaneously with the peak current changes are shown. The breaking point at the peak potential was observed at pH 7.02. While the peak potential decreased up to pH 7.02, the potential value increased after pH 7.02. Since the high peak current will allow the sensor to make more sensitive measurements, the pH was chosen as 12.02 for the electrochemical ivermectin sensor studies. Electrochemical behaviors of 2.85 × 10 −4 mol·L −1 ivermectin were investigated using GCE and GA 2 /GCE electrodes separately using DPV technique. The resulting DPV voltammograms are shown in Fig. 4 . The peak current obtained with the GA 2 /GCE electrode was sharper and more pronounced than the peak current obtained with the GCE electrode, and its intensity was 1.46 times higher. This result showed that the GA 2 /GCE-modified electrode was more sensitive to IVM. DPV is preferred more than CV because of its sensitivity and low background current [32] . Quantitative determination of IVM was applied using the DPV method and GA 2 /GCE modified electrode in BRT solution at pH 12.02. The DPV voltammograms and calibration curve was given in Fig. 5 . DPV measurements started with 0.99 µmol·L −1 concentration of IVM and gradually IVM concentration increased until reaching 444 µmol·L −1 concentration. Ten repetitive measurements were applied for each concentration of IVM in BRT solution at pH 12.02. The standard deviation of the results were calculated with these repetitive measurements. The peak potential was 1.286 V and the current intensity decreased with addition of IVM into solution. The oxidation peak was thought to be corresponding with glutardialdehyde and addition of IVM inhibited the oxidation activity of glutardialdehyde. The inhibition leads the decrease in the peak current intensity as given in Fig. 5a , b. To see this decrease and calculate the sensor properties, peak current intensities plotted against each concentration of IVM and exhibited in the inner graph as Fig. 5(b) . There was two linear concentration ranges. In the linear concentration, ranges were between 9.99 × 10 −7 mol·L −1 -1. 1 × 10 −4 mol·L −1 and 1.1 × 10 −4 mol·L −1 -4.41 × 10 −4 mol·L − 1 , respectively. In both ranges, the peak current intensity of the IVM decreased linearly. Linear regression equation for the 1st operating range was I (µA) = −0.755 C (µmol·L −1 ) with a regression coefficient of R 2 = 0.985. Linear regression equation for 2nd working range was I (µA) = −0.121 C (µmol·L −1 ) + 1 × 10 −4 having the regression coefficient of R 2 = 0.985. The limit of detection (LOD) value of the modified electrode can be calculated using Eq. 1 [33] : In this equation, S is standard deviation of the measurements and m is the slope of the calibration line in the linear range. LOD was calculated to be 2.66 × 10 −6 mol·L −1 for the first linear range and 3.89 × 10 −5 mol·L −1 for the second linear range. Limit of quantitation (LOQ) value can be calculated using Eq. 2 [33] : 8 .87 × 10 −6 1.30 × 10 −4 Repeatability of peak potential, RSD a % (intra-day) 0.94 Repeatability of peak potential, RSD a % (intra-day) 0.50 Repeatability of peak current, RSD a % (intra-day) 0.82 Repeatability of peak current, RSD a % (intra-day) 1.12 LOQ was found to be 8.87 × 10 −6 mol·L −1 for the first linear range and 1.30 × 10 −4 mol·L −1 for the second linear range. The sensor parameters of GA 2 /GCE electrode for IVM detection were given in Table 1 . This proposed method for the determination of IVM was carried out in a shorter time than the previous chromatographic and electrochemical methods and at an affordable cost without any synthesis or pre-separation process [15, 23, 27, 34] . In order to determine the measurement repeatability of GA 2 /GCE, 3 separate measurements of 0.285 mM IVM were taken during the day with the same electrode. The relative standard deviation (RSD%) of this electrode was calculated as 0.82. To estimate the reproducibility of GA 2 /GCE, the relative standard deviations (RSD%) of the three independently prepared electrodes using solution containing 0.285 mmol·L −1 IVM were calculated as 1.12. These results showed that the measurement repeatability and reproducibility of the developed modified electrode were at desirable level. Increasing amount of 1 × 10 −3 mol·L −1 IVM solution was added into the electrolytic BRT solution and tap water. Tap water was 25% by volume of the total solution. The electrochemical activity against the GA 2 /GCE electrode was measured and the DPV data were recorded (Fig. 6(a) ). A linear calibration plot was drawn using peak current intensities of the voltammograms (Fig. 6(b) ). An excellent linearity was found between 3.98 × 10 −6 mol·L −1 and 2.85 × 10 −4 mol·L −1 . The linear regression equation was calculated as I (µA) = − 0.445 C (µmol·L −1 ) and the regression coefficient was R 2 = 0.990. LOD and LOQ values were calculated as 1.54 × 10 −5 mol·L −1 and 5.13 × 10 −5 mol·L −1 , respectively. In order to determine the feasibility of this method, recovery studies were carried out with tap water sample. Achieved data were shown in Table 2 . Canga et al. [5] reported in their review article that only 1% of ivermectin and its metabolites were removed via urine from the body. The urine sample taken from a healthy individual was added to the BRT (pH 12.02) solution, constituting 25% of the total electrolytic solution by volume, and the stock IVM solution (1 × 10 −3 mol·L −1 ) was gradually added to the resulting electrolyte solution and the DPV data were recorded ( Fig. 7(a) ). A linear calibration plot was drawn using the recorded voltammograms ( Fig. 7(b) ) and an excellent linearity was found between 3.98 × 10 − Recovery calculations were made to show the applicability of this method. Recovery rates were shown in Table 3 . As a summary, LOD, LOQ, linear detection ranges, and sensitivity parameters of GA 2 /GCE-modified electrode were given in Table 4 . While the highest sensitivity (0.45 µA/ ((µmol·L −1 )(cm 2 ))) was achieved with urine sample, the lowest detection limit as 2.66 × 10 −6 mol·L −1 was obtained with BRT solution sample. Since it has a high sensitivity in urine sample that consists of different materials having interference effect, it can be concluded that the selectivity of the electrode was in desired level. Excellent selectivity and sensitivity properties of GA 2 /GCE-modified electrode make it a simple and low-cost sensor of IVM for water and urine samples. In this study, GA 2 /GCE-modified electrode was used for the determination of IVM. Compared with GCE and GA 2 /GCE, the higher peak current of IVM in the modified electrode showed that the modified electrode was more suitable for detection. The value of this study is that the electrode surface is modified without the need for any synthesis process and it allows rapid, sensitive, and selective determination of IVM at low concentrations. The modified electrode showed both good repeatability and stability. With the proposed method, recovery studies were carried out on real samples (tap water and urine) using the direct calibration graph method. The highest sensitivity (0.45 µA/((µmol·L −1 )(cm 2 ))) was achieved with urine sample and the lowest detection limit as 2.66 × 10 −6 mol·L −1 was obtained with BRT solution sample. Although the GA 2 /GCE electrode sensitivity was high with urine sample, its recovery % average was 61.54. The interference effect of the other metabolites consisted in the urine prevented detection with 100% recovery ratio. Thus, it can be said that the selectivity of the electrode was lower for urine sample. In further studies, ivermectin analysis in milk sample can be studied. In a study by Ogbuokiri et al. [35] , when a healthy mother used ivermectin at a dosage of 150 μg/kg, ivermectin concentration in breast milk was measured as 14.1 ng·mL −1 6.5 h later. Therefore, it will help mothers who have to use the ivermectin for medical purposes to measure the ivermectin residue in breast milk rapidly, sensitively, and selectively to choose the appropriate time to feed their babies. Detection Methods and Novel Sensor Studies for Corona Virus Physical Testing and Chemical Analysis (Part B: Chemical Analysis) Materials Today: Proceedings