key: cord-0891509-0hhrx0lu authors: Ohira, Shingo; Kanayama, Naoyuki; Komiyama, Riho; Ikawa, Toshiki; Toratani, Masayasu; Ueda, Yoshihiro; Washio, Hayate; Miyazaki, Masayoshi; Koizumi, Masahiko; Teshima, Teruki title: Intra-fractional patient setup error during fractionated intracranial stereotactic irradiation treatment of patients wearing medical masks: comparison with and without bite block during COVID-19 pandemic date: 2020-11-12 journal: J Radiat Res DOI: 10.1093/jrr/rraa101 sha: 79ccab5ddc703e45c598b6767f73c7d4ff00a679 doc_id: 891509 cord_uid: 0hhrx0lu The immobilization of patients with a bite block (BB) carries the risk of interpersonal infection, particularly in the context of pandemics such as COVID-19. Here, we compared the intra-fractional patient setup error (intra-SE) with and without a BB during fractionated intracranial stereotactic irradiation (STI). Fifteen patients with brain metastases were immobilized using a BB without a medical mask, while 15 patients were immobilized without using a BB and with a medical mask. The intra-SEs in six directions (anterior–posterior (AP), superior–inferior (SI), left–right (LR), pitch, roll, and yaw) were calculated by using cone-beam computed tomography images acquired before and after the treatments. We analyzed a total of 53 and 67 treatment sessions for the with- and without-BB groups, respectively. A comparable absolute mean translational and rotational intra-SE was observed (P > 0.05) in the AP (0.19 vs 0.23 mm with- and without-BB, respectively), SI (0.30 vs 0.29 mm), LR (0.20 vs 0.29 mm), pitch (0.18 vs 0.27°), roll (0.23 vs 0.23°) and yaw (0.27 vs 22°) directions. The resultant planning target volume (PTV) margin to compensate for intra-SE was <1 mm. No statistically significant correlation was observed between the intra-SE and treatment times. A PTV margin of <1 mm was achieved even when patients were immobilized without a BB during STI dose delivery. The a ermath of the outbreak of the coronavirus disease 2019 in Wuhan, China, has witnessed the spread of infections from Asia to Europe and the USA. Consequently, the World Health Organization (WHO) has declared COVID-19 a global pandemic. According to the WHO scienti c brief, interpersonal transmission of COVID-19 occurs through droplet transmission from coughing, sneezing or talking between people during close contact [1] . Because infective respiratory droplets can transmit the virus through exposed mucosae or conjunctiva, the wearing of a medical mask is one of the primary prevention measures for minimizing the spread of COVID-19 [2] . In this context, Liang et al. have reported that cancer patients are more susceptible to infection by COVID-19 than individuals without cancer owing to their immunosuppressive condition caused by malignancy and cancer treatments [3] . As regards radiotherapy treatments, which generally cover several weeks, medical sta need to protect susceptible cancer patients from infection (from sta to patients) during the treatment period. Similarly, medical sta need to be protected from interpersonal transmission (from patients to sta ). Additionally, the normally used closed environment for radiotherapy treatments (including shielded treatment rooms with inadequate ventilation in underground locations) may increase the risk of infection. In this regard, a er the outbreak, Wei et al. studied the patient setup accuracy of cranial and head-and-neck patients immobilized with thermoplastic masks as well as medical masks, and they found that the setup accuracy with the medical mask-thermoplastic mask combination was comparable with that of thermoplastic masks alone [4] . In the context of radiation treatment, we note that patients with brain metastases have historically been treated with conventional fractionated whole-brain radiotherapy (WBRT). Meanwhile, the recently developed stereotactic irradiation (STI) technique allows the accurate delivery of high radiation doses in a few fractions with the application of sophisticated radiation planning, accurate patient immobilization and volumetric image guidance irradiation techniques [5] [6] [7] . Thus, STI treatment results in reduced radiation-induced toxicity without compromise on the treatment outcomes [8, 9] . On the other hand, because a relatively longer treatment time is required for STI, accurate patient immobilization becomes imperative for the success of the treatment. Although the initial patient setup error (SE) can be corrected by using a volumetric image and a six degrees of freedom (6-DoF) couch, the patient position during treatment delivery should remain unchanged. In this regard, several studies have reported that the use of a bite block (BB) in conjunction with the normal immobilization devices improves the accuracy of patient setup [10] [11] [12] . However, the use of the BB carries the risk of buccal secretion adhering to the immobilization system, which can lead to indirect contact transmission between patients and medical sta . Moreover, droplet transmission may also occur when a patient detaches the medical mask during the setup procedure. Against this backdrop, in this study, we conducted treatment simulations and procedures with patients wearing medical masks. The intrafractional patient SE (intra-SE) with and without the BB during STI treatment was compared for patients with brain metastasis. This retrospective study, approved by the ethics committee of our institution, included 30 patients with brain metastases who underwent fractionated STI; Table 1 lists the patient details. In the simulations, all patients were immobilized by means of the QFix Encompass SRS immobilization system (Avondale, PA, USA), which consists of a couch insert and a clam-shell-style thermoplastic mask [13] . The Encompass mask consists of anterior and posterior portions, and it can exactly conform to the shape of each patient's head region to minimize patient motion during the treatment. For a randomly selected 15 patients, who received fractionated STI until the end of 2019, the medical mask was detached during simulation, and a BB was used (Figure 1a , le panel); these patients were designated as belonging to the 'with-BB' group. To maintain the patient position during treatment, the thermoplastic mask was extended to shape the nose region as well. For a randomly selected 15 patients, who received fractionated STI from the beginning of 2020, the thermoplastic mask was formed without a BB ('without-BB' group), and patients wore medical masks (Figure 1a , right panel) to avoid the risk of interpersonal infection; the nose region was not covered by the thermoplastic mask. In accordance with the interim guidelines recommended by the WHO, the medical mask used was either a surgical or at/pleated procedure mask [2] . Computed tomography (CT) scanning was performed by means of a dual-energy CT system (Revolution HD; GE Medical Systems, Milwaukee, WI) with the following scanning parameters: tube voltage of 140/80 kVp; tube current of 550-600 mA; helical pitch of 0.531:1. Images were reconstructed with a slice thickness of 1 mm and a eldof-view (FOV) of 320 mm. First, the acquired CT images were transferred to a treatment planning system (Eclipse; Varian Medical Systems, Palo Alto, CA). A planning target volume (PTV) was generated by the addition of an isotropic 1mm margin to the gross tumor volume. The prescription dose of 30-35 Gy was delivered to the PTV in 3-5 fractions (Table 1) . Automated noncoplanar volumetric modulated arc therapy (named HyperArc) plans [5] were generated. One full or half coplanar arc with a couch rotation of 0 • and three half non-coplanar arcs with couch rotations of 315, 45 and 90 • (or 270 • ) were used in the treatment. TrueBeam STx and Edge treatment units with a 6-DoF couch were used for dose delivery (Varian Medical Systems). Figure 1b shows the work ow of the STI dose delivery applied in this study. For treatment, patients were immobilized in the same manner as in the case of the CT simulation, and cone-beam (CB) CT images (pre-CBCT) were acquired before treatment (slice thickness of 1 mm). The patient position was corrected by using the CBCT and the corresponding planning CT images in six directions (three translational directions: anterior-posterior (AP), superior-inferior (SI) and le -right (LR), and three rotational directions: pitch, roll, and yaw), and doses were delivered with the use of a coplanar arc (couch rotation of 0 • ). Subsequently, the megavoltage (MV) linac graphy (AP or PA direction) was acquired for each couch angle, and bony registration was performed by using the MV images and the corresponding digitally reconstructed radiographs (DRRs) before dose delivery. A er the treatments, the acquisition of CBCT images (post-CBCT) were performed to assess the intra-SE. The intra-SE was de ned as the di erence in the patient's position between the pre-and post-CBCT in the six directions, and thus, the intra-SE included the patient's position correction done using the MV images during treatment. Subsequently, the intra-SE without MV correction was simulated by adding the amount of shi done using each MV image. Because the MV images were acquired by using the AP/PA beams, the intra-SE without MV correction was simulated along the SI, LR and yaw directions. The 3D translational intra-SE was calculated as the square root of the sum of squares of the three translational intra-SEs, and the 3D rotational intra-SE was the sum of the three rotational angles. The treatment time (TT) was de ned as the amount of time taken from the acquisitions of pre-CBCT to post-CBCT. For the translational intra-SE, the systematic and random error was de ned as the mean and standard deviation (SD) of intra-SE, respectively, for each patient. The statistical values of Σ and σ were calculated as the SD of the systematic error and the root mean square of random error, respectively. The PTV margin (M) to compensate for the patient's positional uncertainty during STI treatment was calculated by using the following margin recipe [14] :M = 2.5 + β σ 2 + σ 2 p − βσ p ,whereσ p (3.2 mm) is the width of the penumbra modelled by a cumulative Gaussian, and β (1.64) is the value of the inverse cumulative standard normal distribution at the prescribed PTV minimum dose level. All statistical analyses were performed by using SPSS so ware (version 24; IBM, Armonk, NY). The Mann-Whitney U test was performed for the statistical measurement of the di erence in the absolute intra-SE values between the with-and without-BB cases. Subsequently, the paired Wilcoxon signed-rank test was performed to compare the absolute intra-SE corrected using MV image and that without MV correction. To statistically measure the strength of the linear relationship between the 3D translational/rotational intra-SE and TT, a Spearman rank correlation coe cients analysis was performed. The absolute value of Spearman's r s was considered very weak for r s ≤ 0.2, weak for 0.2 < r s ≤ 0.4, moderate for 0.4 < r s ≤ 0.7, and strong for 0.7 < r s . A two-sided P value of < 0.05 was considered to indicate statistical signi cance. A total of 53 and 67 sessions corresponding to the with-and without-BB groups, respectively, were analyzed to assess the intra-SE during STI dose delivery. Figure 2 shows the histograms of the translational intra-SE. A comparable absolute mean (SD) intra-SE is observed in the AP direction (0.19 (0.16) mm vs 0.23 (0.23) mm for the withand without-BB cases, respectively, P = 0.90), in the SI direction (0.30 There is no statistically signi cant di erence in the rotational intra-SE between the with-and without-BB groups (P > 0.10). Figure 4 illustrates the translational (SI and LR) and rotational (yaw) intra-SEs when the MV images were not used during dose delivery. A relatively wide variation in intra-SE was observed in the SI direction for the without BB group, and the intra-SE was >1 mm for 18 out of 67 sessions, whereas only 5 sessions exceeded an intra-SE of 1 mm in the SI direction for the with BB group. A statistically signicant di erence in the absolute intra-SE was not observed (P > 0.05) between the with and without BB groups (0.43 (0.47) vs 0.6 (0.51) mm, 0.38 (0.28) vs 0.41 (0.29) mm, and 0.34 (0.31) vs 0.31 (0.26) mm in the SI, LR and yaw directions, respectively). In contrast, a signi cant di erence was observed in the SI direction when the intra-SE with MV correction for the with-BB group was compared with that without MV correction (P < 0.001). For the without-BB group, the intra-SEs without MV correction were signi cantly larger in all directions than those with MV correction (P < 0.01). Table 2 lists the systematic and random error for each patient, and the PTV margins calculated to compensate for translational intra-SEs during the STI treatment. For both with and without BB groups, systematic and random errors were <1 mm for all patients. The resultant PTV margins were <1 mm for the with (0.3, 0.4 and 0.5 mm, in the AP, SI and LR directions, respectively) and without (0.5, 0.5 and 0.8 mm, in the AP, SI and LR directions, respectively) BB groups. Figure 5 shows the relationships between the translational and rotational 3D intra-SEs and the treatment time for each group. In the study, because a beam energy of 6 MV with a maximum dose rate of 600 monitor unit/min was used for 8 patients, a longer treatment time (P < 0.001) was required for patients of the BB group (mean (SD), 15.4 (3.3) min) than for their without-BB counterparts (mean (SD), 13.0 (1.5) min). There are only 'very weak' correlations (|r s | ≤ 0.2), and no statistically signi cant correlation is observed between the translational/rotational 3D intra-SE and treatment time (P > 0.18). In this study, we compared the intra-SE for patients immobilized with and without BBs during fractionated STI dose delivery in the context of protecting patients/medical sta from the interpersonal transmission of COVID-19. In this regard, we note that a number of e orts (personal protective equipment, accelerated irradiation schedule, area zoning, terminal disinfection, etc.) aimed at protecting patients and medical sta have been reported speci cally for radiation oncology [4, [15] [16] [17] [18] . Zhang et al. suggested that patients should wear medical masks during treatment to avoid cross infection [16] . They otherwise suggested the use of an additional face shield with the thermoplastic mask. Wei et al. attempted a careful balancing of the comfort level of mask wearing against the immobilization accuracy, and they used thermoplastic masks for cranial and head-and-neck patients [4] . They demonstrated that the mean translational and rotational SEs for patients with medical masks were ≤1 mm and ≤ 0.8 • , respectively, and no signi cant di erence was observed in relation to patients without medical masks. Intra-SE during STI may induce non-negligible dosimetric e ects, which can induce radiation injury or underdosage for the target volume. In this regard, Roper et al. showed that small targets are susceptible to rotational SE during the application of the single-isocentric irradiation technique for multiple targets [19] . Another study by Sagawa et al. demonstrated that a 3D rotational SE of ≥2 • signi cantly reduced the target coverage, and this e ect was more prominent for tumors located far from the isocenter [20] . Furthermore, for the addition of tight PTV margins (1-2 mm) to the gross tumor volume (GTV) for STI treatment [21] [22] [23] , the patient SE should be minimized to ensure successful treatment. In this regard, the 6-DoF setup overcomes limitations of the 4-DoF setup in correcting rotational SE, which signi cantly a ects the STI treatment in the case of small targets, small PTV margins, high doses and steep dose gradients [24] . However, maintaining the patient's position during STI treatment is still challenging in modern radiotherapy. Previous investigators have reported on the intra-SE during STI dose delivery [12, 25] . In particular, using the thermoplastic mask, Carminucci et al. demonstrated that the mean translational intra-SEs during STI were 1.67, 0.73 and 0.75 mm in the AP, SI and LR directions, respectively, whereas the rotational intra-SEs were 0.73, 1.44 and 0.76 • in the pitch, roll and yaw directions, respectively [25] . Babic et al. compared the intra-SEs among various immobilization systems and found that the thermoplastic mask in conjunction with the BB could stringently limit the intra-SE over the thermoplastic mask alone Fig. 4 . Histograms of translational and rotational intra-SE during STI with and without BB when megavoltage images were not used for patient correction. [12] . In this study, with the aim of protecting patients and medical sta from COVID-19 infections, we rstly compared the intra-SE between the 'with' and 'without-BB' patients during STI. The mean absolute translational and rotational intra-SE was small (<0.3 mm and <0.3 • , respectively) when the BB was not used for patient immobilization, and the intra-SE was comparable with that of patients immobilized using BB (Figs 2 and 3) . The resultant PTV margin to compensate for the intra-SE was <1 mm (Table 2 ). In the STI treatment, tight PTV margins (1-2 mm) are widely adopted to minimize radiation injury to the surrounding normal tissue [26, 27] . Thus, the patient's position seems to be maintained during the STI treatment, even though the BB was not used for patient immobilization. However, a signi cantly larger intra-SE was observed when the patient's position was not corrected using the MV images for both with BB (SI direction) and without BB (SI, LR and yaw directions) groups (Fig. 4) . Thus, we need to be aware of the importance of patient correction using MV images especially for the without BB group. Here, we also mention the limitations of our study. First, the number of patients was limited, and patients were not randomized for either the with-or without-BB categories, which may have resulted in selection bias. In particular, 'un-attened' [ attening lter free (FFF)] beams (6X-FFF and 10X-FFF) can generate a higher dose rate than attened beams (6X), which can result in a shorter treatment time in the former cases. Second, mechanical uncertainties (such as the kV imager and couch position accuracies) were not considered in the intra-SE obtained using the CBCT images in this study. Third, corrections of the patient's position during dose delivery (with the use of MV images) were limited to only three directions (SI, LR and yaw); current technologies allow real-time patient motion monitoring in six directions during dose delivery. In this regard, we note that Wiant et al. demonstrated that surface imaging o ers a non-ionizing, near real-time alternative to radiographic imaging for patient localization during STI [28] . Finally, the PTV margin calculated in the present study could not account for the uncertainties in the mechanical accuracy of the linear accelerator (such as the accuracy of the isocenter, image center and treatment couch movement) [29] . Meanwhile, despite the abovementioned limitations, our quantitative data provide useful information toward protecting against the spread of infections between patients and medical sta during STI treatment. In conclusion, we found that a PTV margin of <1 mm could be achieved even when patients were immobilized without the BB. That is, the required accuracy for STI can be maintained even Modes of transmission of virus causing COVID-19: Implications for IPC precaution recommendations Advice on the use of masks in the context of COVID-19 Cancer patients in SARS-CoV-2 infection: A nationwide analysis in China Radiotherapy work ow and protection procedures during the coronavirus disease 2019 (COVID-19) outbreak: Experience of the Hubei cancer Hospital in Wuhan, China HyperArc VMAT planning for single and multiple brain metastases stereotactic radiosurgery: A new treatment planning approach Five-fraction CyberKnife radiotherapy for large brain metastases in critical areas: Impact on the surrounding brain volumes circumscribed with a single dose equivalent of 14 Gy (V14) to avoid radiation necrosis Stereotactic radiosurgery plus whole brain radiotherapy versus radiotherapy alone for patients with multiple brain metastases Stereotactic radiosurgery versus whole brain radiation therapy: A propensity score analysis and predictors of Care for Patients with brain metastases from breast cancer Stereotactic radiosurgery for patients with multiple brain metastases ( JLGK0901): A multi-institutional prospective observational study Improved setup and positioning accuracy using a three-point customized cushion/mask/biteblock immobilization system for stereotactic reirradiation of head and neck cancer Cone-beam computed tomography in hypofractionated stereotactic radiotherapy for brain metastases To frame or not to frame? Conebeam CT-based analysis of head immobilization devices speci c to linac-based stereotactic radiosurgery and radiotherapy Evaluation and veri cation of the QFix encompass(TM) couch insert for intracranial stereotactic radiosurgery The probability of correct target dosage: Dose-population histograms for deriving treatment margins in radiotherapy The impact of coronavirus (COVID-19) on head and neck cancer patients' care Prevention and control measure to avoid cross infection during radiotherapy in coronavirus disease 2019 (COVID-19) epidemic in Wuhan, China Navigating the challenges of the COVID-19 outbreak: Perspectives from the radiation oncology service in Singapore Preparation for the COVID-19 pandemic in the department of radiation oncology in the National Cancer Center Hospital in Tokyo Single-Isocenter multipletarget stereotactic radiosurgery: Risk of compromised coverage Dosimetric e ect of rotational setup errors in stereotactic radiosurgery with HyperArc for single and multiple brain metastases Inhomogeneous tumor dose distribution provides better local control than homogeneous distribution in stereotactic radiotherapy for brain metastases Repeated courses of radiosurgery for new brain metastases to defer whole brain radiotherapy: Feasibility and outcome with validation of the new prognostic metric brain metastasis velocity Present clinical practices of stereotactic irradiation for metastatic brain tumors in Japan: Results of questionnaire survey of the Japanese radiation oncology study group ( JROSG) working subgroup for neurological tumors The use of six degrees of freedom couch is only clinically bene cial in stereotactic radio surgery Assessment of motion error for frame-based and noninvasive mask-based xation using the Leksell gamma knife icon radiosurgery system Dosimetric factors related to radiation necrosis a er 5-fraction radiosurgery for patients with resected brain metastases LINAC based stereotactic radiosurgery for multiple brain metastases: Guidance for clinical implementation Direct comparison between surface imaging and orthogonal radiographic imaging for SRS localization in phantom Evaluation of the truebeam machine performance check (MPC) geometric checks for daily IGRT geometric accuracy quality assurance None declared. This study was supported by JSPS KAKENHI Grant (Grant-in-Aid for Young Scientists 19 K17285).