key: cord-0890126-d4qlz1ob authors: Zhao, Bin; Che, Congnyu; Wang, Weijing; Li, Nantao; Cunningham, Brian T. title: Single-step, wash-free digital immunoassay for rapid quantitative analysis of serological antibody against SARS-CoV-2 by photonic resonator absorption microscopy date: 2021-04-01 journal: Talanta DOI: 10.1016/j.talanta.2020.122004 sha: f302a9b7fc4f1216b2d72970b7c136e9b2ff3a04 doc_id: 890126 cord_uid: d4qlz1ob Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2), the cause of Coronavirus Disease 2019 (COVID-19), poses extraordinary threats and complex challenges to global public health. Quantitative measurement of SARS-CoV-2 antibody titer plays an important role in understanding the patient-to-patient variability of immune response, assessing the efficacy of vaccines, and identifying donors for blood transfusion therapy. There is an urgent and ever-increasing demand for serological COVID-19 antibody tests that are highly sensitive, quantitative, rapid, simple, minimally invasive, and inexpensive. In this work, we developed a single-step, wash-free immunoassay for rapid and highly sensitive quantitative analysis of serological human IgG against SARS-CoV-2 which requires only a single droplet of serum. By simply incubating 4 μL human serum samples with antibody-functionalized gold nanoparticles, a photonic crystal optical biosensor coated with the recombinant spike protein serves as a sensing platform for the formation of sandwich immunocomplex through specific antigen–antibody interactions, upon which the detected IgG molecules can be counted with digital precision. We demonstrated a single-step 15-min assay capable of detecting as low as 100 pg mL(−1) human COVID-19 IgG in serum samples. The calculated limit of detecting (LOD) and limit of quantification (LOQ) is 26.7 ± 7.7 and 32.0 ± 8.9 pg mL(−1), respectively. This work represents the first utilization of the Activate Capture + Digital Counting (AC + DC)-based immunoassay for rapid and quantitative analysis of serological COVID-19 antibody, demonstrating a route toward point-of-care testing, using a portable detection instrument. On the basis of the sandwich immunoassay principle, the biosensing platform can be extended for the multiplexed detection of antigens, additional IgGs, cytokines, and other protein biomarkers. The ongoing Coronavirus Disease 2019 (COVID-19) pandemic is caused by a novel coronavirus, now known as Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) [1] . The rapid spread of the SARS-CoV-2 virus worldwide has led to >23.5 million documented cases and over 800 000 deaths at the time of this writing (August 2020) (https://coronavirus.jhu.edu/map.html). While detection of SARS-CoV-2 specific nucleic acid sequences by reverse transcriptase-polymerase chain reaction (RT-PCR) and (to a lesser extent) serological tests for the presence of bloodborne antigen are the predominant diagnostic assays for COVID-19 infection [2] [3] [4] [5] [6] [7] , these tests do not reveal the presence or intensity of a patient's immune response, or enable identification of a patient who experienced a COVID- 19 infection, but was not tested while displaying a measurable viral load. Serological antibody testing is an important diagnostic tool for combating the COVID-19 pandemic [8] . Studies have shown that measurement of SARS-CoV-2 specific antibodies may be helpful for the diagnosis of suspected patients with negative RT-PCR results and for the identification of asymptomatic infections [9] . More importantly, antibody testing can detect both recent and prior infections, while playing vital roles in epidemiology studies [10] . Measuring the immune response against SARS-CoV-2 by antibody testing is an important tool for assessing the outcomes of patients and understanding global prevalence. Quantitative assessment of SARS-CoV-2 antibody titer is especially important as clinicians and researchers more fully understand the patient-to-patient variability of immune response, in terms of the onset time for post-infection antibody production, and the post-recovery time that antibodies continue to be present [11] . As vaccines become available, it will be urgent for patients to know the extent to which their immune response has been stimulated. Furthermore, quantitative measurement of SARS-CoV-2 antibody titer in donated blood is important for blood transfusion therapy, which has shown promising results for treating patients with severe symptoms [12] . Therefore, there will be an urgent and increasing demand for serological COVID-19 antibody tests that are highly sensitive, quantitative, rapid, simple, minimally invasive, and inexpensive. Currently, there are three methods that represent the most widely adopted serological COVID-19 antibody tests. The traditional enzymelinked immunosorbent assay (ELISA) performed in microplates can quantify the different isotypes and subclasses of antibodies, and has been used for the detection of COVID-19 antibodies against SARS-Cov-2 spike and nucleocapsid proteins [13] [14] [15] . Chemiluminescent immunoassays (CLIA) combine the immunoassay with photon-generating chemiluminescence reporters by using a luminescent molecule-labeled antibody [16] . CLIA systems have been developed for verifying the performance of commercial COVID-19 IgG and IgM antibodies testing kits [17] . Although ELISA and CLIA offer high sensitivity and specificity, both ELSIA and CLIA-based workflows are time-consuming due to the multiple sample reagent handling and washing steps, while they also require a relatively large sample volume. The lateral flow immunoassay (LFIA) is a paper-based method that has emerged as a rapid diagnostic tool for point-of-care non-quantitative detection that offers benefits, such as low cost and ease of use [18] . However, while a number of commercial LFIAs have been rapidly developed for measuring COVID-19 IgG and IgM qualitatively [19, 20] , the sensitivity of LFIA is relatively low (ng mL − 1 level) and it lacks the capability for quantitative analysis. Due to the limitations of ELISA, CLIA, and LFIA, there is an important unaddressed gap in the currently available technologies for quantitative and simple SARS-CoV-2 antibody testing. Ideally, the diagnostic workflow would require a single step, an inexpensive/portable detection instrument, an inexpensive/disposable assay cartridge, and only a fingerstick quantity of serum. To achieve the single molecule biosensing, numerous sensing technologies have been intensely explored recently, such as electrochemical biosensors (droplet digital PCR) [21] , impedance-based biosensors (nanopores) [22] , interferometric biosensors (interferometric detection of scattering (iSCAT)) [23] , and photonic crystal (PC)-based biosensors [24, 25] . Among these digital biosensing technologies, PC-based biosensors provide a combination of low cost, simple operation, short assay time, and high sensitivity [24] . PC-based biosensors have been widely applied to the detection of multiple biological targets, such as cells, proteins, metabolites, glucose, DNAs, and RNAs [25] [26] [27] [28] [29] [30] . Additionally, a diverse range of PC structures have been incorporated with other emerging technologies, such as microfluidics [31] , flexible materials [32] , and smartphones [33] , to offer portable platforms in biosensing applications in point-of-care (POC) testing. For a recent review of digital-resolution biomolecular detection technologies, we refer the reader to Refs. [34] . In recent studies, we reported the development of a highly sensitive, single-step, enzyme-free assay with digital biomolecular precision for quantifying the presence of serum biomarkers called "Activate Capture + Digital Counting (AC + DC)" [25, 31] . AC + DC uses a PC biosensor and a new form of microscopy called photonic resonator absorption microscopy (PRAM). AC + DC was initially utilized for ultrasensitive (100 aM limits of detection) detection of cancer-specific microRNA (miRNA) sequence with digital-resolution and single-mismatch discrimination without target amplification or wash steps [25] , while using only a 30 μL sample volume. We further developed the AC + DC assay approach by integrating the PC biosensor into a microfluidic cartridge to demonstrate a simple quantitative viral load assay for HIV-1 p24 antigen in human serum, which required only 40 μL sample volume and provided a result in ~30 min [31] . In this work, we adapted the AC + DC assay method to demonstrate a single-step, wash-free immunoassay for rapid quantitative analysis of serological human IgG against SARS-CoV-2. Simply incubating the human COVID-19 IgG with the test sample and secondary antibodyfunctionalized gold nanoparticles (2 o Ab-AuNPs) the recombinant COVID-19 spike protein-coated PC biosensor serves as a surface for formation of sandwich immunocomplex through specific antigen-antibody interactions, enabling highly sensitive and quantitative digitalresolution detection of human IgG antibody within 15 min. Of note, the term "single-step" refers to the one-pot reaction of human COVID-19 IgG as target, 2 o Ab-AuNPs as detection probe and spike protein coated PC sensing platform. Without any pre-incubation step after mixing the sample and 2 o Ab-AuNPs in a vial, the mixture is immediately applied on the spike protein-coated PC biosensing platform for the subsequent PRAM imaging. Also, no washing steps are involved in the assay. This work represents the first utilization of the AC + DC-based immunoassay platform for rapid quantitative analysis of serological COVID-19 antibody with high sensitivity and selectivity. The calculated limit of detection (LOD) is 26.7 ± 7.7 pg mL − 1 and the calculated dynamic detection ranges from 93.3 pg mL − 1 to 25.7 ng mL − 1 . The estimated limit of quantification (LOQ) is 32.0 ± 8.9 pg mL − 1 . The results support the potential for POC testing of serological antibodies against SARS-CoV-2 using a portable detection instrument. On the basis of the sandwich immunoassay principle, this PC digital biosensing platform can be extended to become a universal analytical platform for the multiplex detection of antigens, additional IgGs, cytokines, and other protein biomarkers [35, 36] . The PC biosensors were fabricated on 8-inch diameter glass substrates (Moxtek, Orem, UT) based on our design specifications, and cut into 10 × 12 mm 2 pieces that were used in all experiments. The PC periodic structure, fabrication process, and resonant optical properties have been described in prior publications [25, 31] . Briefly, the PC has a linear grating period of 380 nm and a grating depth of 97 nm etched into a glass substrate by reactive ion etching. The grating is coated by TiO 2 thin film with a high refractive index of 2.25 and a thickness of 98.5 nm, which yields a surface structure with high-efficiency resonant reflection at a wavelength of 625 nm when the PC surface is covered in aqueous media (Fig. S1 ). PCs are used once and are discarded after an assay. The PC used here is a sub-wavelength periodic grating structure acting as a highly efficient narrow bandwidth resonant reflector [26, 31, 37] . The digital-resolution immunoassay in this work is based on the PRAM system we reported previously [25] , where the light absorption of single bound plasmonic nanoparticles is dramatically enhanced when the surface plasmon-resonant wavelength of the nanoparticle matches the PC resonance wavelength. As a result, the local reflected resonant intensity from the PC is substantially reduced by the presence of individual nanoparticles (Fig. S1 ). Our PRAM system therefore enables the observation of single particles and digital-resolution counting of detected target molecules. A schematic diagram of the PRAM instrument used in this work is in Fig. S2 . Peak intensity value (PIV) images are obtained to visualize the captured single AuNPs under PRAM, where the value of the reflection intensity at the resonant wavelength are used as the contrast modality to provide high signal-to-noise microscopy, while using low intensity LED illumination and an ordinary uncooled image sensor. AC + DC detection is highly quantitative and robust, as it does not utilize fluorescent dyes that are subject to the effects of photobleaching. The PC surface was cleaned by acetone, isopropyl alcohol (IPA), and Milli-Q water within a series of separate glass jars and sonicated for 2 min, after which the PC was dried by N 2 and baked in an 80 • C oven for 10 min to allow the water on the PC surface evaporate. The PC surface was further cleaned and activated by a 10-min oxygen plasma treatment (Pico plasma system, diener electronic, power = 100%). Then, each PC was coated with a layer of (3-Glycidyloxypropyl) trimethoxysilane (GPTMS, Sigma Aldrich) in an 80 • C vacuum for 6 h. Redundant silane was removed by washing with toluene, methanol, and Milli-Q water sequentially. A PDMS gasket with six circular openings (diameter = 3 mm) was placed onto the PC surface to form six independent liquid compartments. For the human COVID-19 IgG capture assay, the PC was 2 o Ab-AuNPs were prepared via EDC/NHS chemistry and heterobifunctional polyethylene glycol (PEG) linkers following the protocol we previously reported [31] . In detail, aqueous solutions of 36.5 μL of 4 μM heterobifunctional polyethylene glycol (SH-PEG-COOH, JenKem Technology), 146 μL of 1 μM 1-ethyl-3-(3-dimethylaminopropyl) carbodiimide (EDC, Thermo Scientific), 58 μL of 2.5 μM N-hydroxy succinimide (NHS, Thermo Scientific) were mixed with 174.5 μL of Milli-Q water. After pH adjustment to 7.4 by adding 50 μL of 10 × concentrated PBS and 25 μL of 0.01 M NaOH, the solution was incubated at room temperature for 1 h with gentle shaking. Then, 7 μL of 1.4 μM goat anti-Human IgG Fc highly cross-adsorbed secondary antibodies (2 o Ab, A18831, Thermo Scientific) were added into the solution and incubated at room temperature for 2 h. After centrifuging at 4 • C for 10 min at 14, 000 g using a 50-kDa centrifuge filter, the mixture was further washed with PBS buffer twice under the same centrifuge conditions to remove any byproduct. The as-prepared SH-PEG-2 o Ab conjugate (12 μL) was mixed with 200 μL of 80 nm diameter gold nanourchin stock solution (13 pM, Cytodiagnostics) and incubated at room temperature for 1 h with gentle shaking. After overnight incubation at 4 • C, 2 μL of Super-Block™ blocking buffer was added to block the bare AuNP surface to eliminate non-specific binding. The 2 o Ab-AuNPs were gently shaken at room temperature for 1 h and washed by centrifuge at 4 • C for 30 min at 800 g. The pellet was re-dispersed with PBS and stored at 4 • C for later use. The 2 o Ab-AuNPs were characterized by a NanoDrop® ND-1000 Spectrophotometer (pathlength = 1 mm) and Dynamic Light Scattering Analyzer to confirm the conjugation of antibodies (Fig. S3 ). In a typical immunoassay in buffer, 9 μL of 2 o Ab-AuNPs conjugate (Abs = 0.070 ± 0.005, diluted with 0.1 × PBS) was mixed with 1 μL of human COVID-19 IgG in PBS at variable concentration (0, 1, 10, 100, 1000 and 5000 ng mL − 1 ) in a 0.6 mL microcentrifuge tube. The final concentration of COVID-19 IgG is 0, 0.1, 1, 10, 100 and 500 ng mL − 1 . For the human serum sample analysis, 6 μL of 2 o Ab-AuNPs (Abs = 0.155 ± 0.005) were added into a 0.6-mL microcentrifuge tube. Serial dilutions of 100% human serum samples containing COVID-19 IgG (4 μL for each) were added to achieve a final concentration of 0.1, 1, 10, 100 and 500 ng mL − 1 . After thoroughly mixing, the mixture was immediately added into a PDMS well attached on the spike protein-coated PC for PRAM imaging. All the assays were repeated three times independently. We selected native IgG, IgM and Human Serum Albumin (HSA) isolated from complete human serum (ab98981, ab91117, ab205808, Abcam) as control proteins. 1 μL of 100 ng mL − 1 IgG, IgM or HSA in PBS was added to 9 μL of 2 o Ab-AuNPs (Abs = 0.070 ± 0.005). For the human serum blank control (HS blank), 4 μL of complete human serum was mixed with 6 μL of 2 o Ab-AuNPs (Abs = 0.155 ± 0.005). After mixing well, the mixture was immediately introduced into the PDMS well on the PC surface for PRAM imaging. The PRAM system is built upon the body of a bright field microscope (Carl Zeiss Axio Observer Z1) (Fig. S2) . A 20 × objective lens was used to acquire the resonant reflection spectrum of each pixel within the fieldof-view at a line-scanning mode. A 512 × 512 pixel image was generated by collecting the reflected light using a charge-coupled device camera (Photometrics Cascade). As shown in Fig. S1 , the Peak Intensity Value (PIV) represents the highest intensity of the resonantly reflected wavelength from the PC surface, and a localized reduction in PIV occurs when an AuNP attaches to the PC surface. By gathering a baseline PIV image of the PC biosensor surface before introduction of the test sample, reductions in PIV are registered by comparison of the baseline image to subsequent PIV images gathered after AuNP attachment. A watershed algorithm is applied to the resulting PIV "shift" image to identify clusters of pixels with reduced PIV, and to thus count the number of attached AuNPs in the field of view. The design of a PRAM-based AC + DC immunoassay for human COVID-19 IgG detection is schematically illustrated in Fig. 1 . The immunoassay employs a typical sandwich configuration, which is based on the principle of AC + DC we previously reported [25, 31] . We prepare 2 o Ab-AuNPs by covalently conjugating urchin-shaped nanoparticles (nanourchins) with anti-human IgG Fc secondary antibody (Fig. 1A) . The human COVID-19 IgG sample and 2 o Ab-AuNPs are mixed in a tube and the mixture is immediately applied on the spike protein-coated PC biosensing platform for PRAM-based single-step immunoassay (Fig. 1B) . COVID-19 IgGs will activate the system by binding either functionalized AuNPs or the spike protein-coated PC through specific antigen-antibody interaction, allowing for the subsequent capture of COVID-19 IgG-AuNP complexes or AuNPs to form sandwich immunocomplex on the PC. PRAM imaging is then conducted for quantitative detection of human COVID-19 IgG antibody by digitally counting the number of bound AuNPs on the PC (Fig. 1C) , where each AuNP in the PRAM image represents one IgG target molecule. We have previously demonstrated that AC + DC-based immunoassays can effectively reduce the time requirement for a diffusion-limited assay [31] . In this work, because no pre-incubation and washing steps are required, our AC + DC immunoassay method allows for a 15-min rapid test of human COVID-19 IgG in human serum in the clinically relevant concentration range. We prepared 2 o Ab-AuNPs by conjugating goat anti-human IgG Fc highly cross-adsorbed secondary antibodies with 80 nm diameter gold nanourchins via EDC/NHS chemistry and heterobifunctional polyethylene glycol (PEG) linkers. Gold nanourchins with a protruding tip morphology (Fig. S3A) were chosen as the label for PRAM-based imaging due to their enhanced light harvesting across the particle surface, and because their surface plasmon resonance wavelength closely matches the PC resonant reflection wavelength [38] . The as-prepared 2 o Ab-AuNPs were characterized by UV-Vis spectroscopy to confirm the conjugation of the secondary antibody (Fig. S3B) . A slight red-shift (~5 nm) was observed after the antibody modification and the maximum absorption wavelength is approximately 625 nm, which perfectly matches the resonant reflection wavelength of the PC. As a gauge of conjugation performance, dynamic lighting scattering (DLS) measurements showed that 2 o Ab-AuNPs displayed an distinct increase of ~25.6 nm in average diameter compared to non-functionalized bare AuNPs (103.4 nm vs 77.8 nm, Fig. S3C ). As a proof-of-concept demonstration, we first tested the human COVID-19 IgG against SARS-Cov-2 spike protein in buffer using the developed single-step immunoassay. The PC coated with recombinant COVID-19 spike protein was first blocked with SuperBlock™ blocking buffer to eliminate non-specific binding during the assay. A series of human COVID-19 IgG samples in phosphate buffered saline (PBS) at variable concentrations were simply mixed with 2 o Ab-AuNPs, followed by the direct introduction into the polydimethylsiloxane (PDMS) container on the PC surface for PRAM imaging. As shown in Fig. 2A , the surface-attached AuNPs that form through the sandwich immunocomplexes were clearly observed at single-particle digital resolution after 30 min. The count of AuNPs increased with the increased human COVID-19 IgG concentration from 0 to 500 ng mL − 1 . Fig. 2B show the counting of AuNPs using an optimized algorithm for PIV images. A customized watershed algorithm for the bound AuNPs is utilized to ensure the accuracy of the particle counting [25] . Fig. 2C presents the dose-response relationship obtained by correlating the AuNPs counts with COVID-19 IgG concentration. Detecting the target IgG in buffer, the PRAM-based single-step immunoassay has a wide dynamic range and good linear relationship from 0.1 ng mL − 1 to 100 ng mL − 1 (Fig. 2C and S4 ). The R square (R 2 ) is 0.986 (Table S1 ). The calculated LOD is determined by extrapolating the COVID-19 IgG concentration to a signal equal to a blank signal (all conditions the same, but no IgG present) plus 3 standard deviations (blank+3σ) [39] , which is 62.3 ± 19.3 pg mL − 1 for the assay in buffer. We also calculated the LOQ to be 565.2 ± 101.0 pg mL − 1 , which is defined as the COVID-19 IgG concentration at a signal equal to a blank signal plus 10 standard deviations. We next evaluated the selectivity of the method and its response to interfering proteins in human serum. Native human IgG and IgM were selected as control proteins due to their similarity to the target IgG and their abundance in serum [40] . We also tested human serum albumin (HSA) because it is the most abundant protein in human blood plasma [41] . As depicted in Fig. 3 , all control proteins (10 ng mL − 1 ) show a small number of particles (HSA: 20 ± 3, IgM: 12 ± 2, IgG: 15 ± 4) and no significant difference to the blank control (18 ± 5) in PBS after 30 min. In contrast, the COVID-19 IgG at the same concentration results in an elevated number of AuNPs (144 ± 38) bound to the PC through specific antigen-antibody binding, which is 7 times higher than that of control proteins. We further examined complete human serum (100% human serum) which contains different types of interfering substances, including proteins, electrolytes, antigens and hormones. A small number of particles (34 ± 8) were observed, indicating that the 2 o Ab-AuNPs can tolerate the complicated components in human serum. Taken together, the results demonstrate our PRAM-based immunoassay has good specificity towards human COVID-19 IgG and is not affected by a variety of protein interferences in human serum. Having demonstrated AC + DC detection of human COVID-19 IgG in buffer with high sensitivity and selectivity, we next challenged the method with human serum samples spiked with human COVID-19 IgG. Interestingly, we found that the signal-to-noise ratio (SNR) of the assay relies on the volume mixing ratio of 2 o Ab-AuNPs and serum samples. A relatively high SNR of 13.9 was achieved when 2 o Ab-AuNPs and the human serum blank sample were mixed at a volume ratio of 6:4 compared to 5:5, which yielded a SNR of 5.1 (Fig. S5 ). Considering the natural IgG proteins in human serum may react with 2 o Ab-AuNPs and therefore reduce the efficacy of the assay, we further increased the concentration of 2 o Ab-AuNPs (2 × [AuNPs] 0 , Abs 625nm = 0.155 ± 0.005) to obtain a SNR of 18.9 for the detection of 400 ng mL − 1 COVID-19 IgG (Fig. S5) . The AuNP counts were determined to be 643 ± 34. In contrast, for the same concentration of COVID-19 IgG in serum, 2 o Ab-AuNPs with the original concentration (1 × [AuNPs] 0 , Abs 625nm = 0.070 ± 0.005) only yielded an AuNP count of 132 ± 16. This number is much lower than that of assays performed in PBS for 100 ng mL − 1 COVID-19 IgG (432 ± 111) using 1 × [AuNPs] 0 . The optimal volume IgG can be differentiated from the blank control at concentration as low as 0.1 ng mL − 1 . We incubated the assay for an additional 15 min, and repeated the imaging procedure to explore whether increased assay time would reduce the detection limit. Interestingly, the PIV images and quantitative analysis show that the AuNP counts after 30 min remain the same as those obtained after only 15 min ( Fig. 4A and B) . The results imply that the kinetics of antigen-antibody interaction in human serum are fast enough that assembly of the AuNP tagged immunocomplex is not limited by the rate of diffusion to the PC biosensor surface. As the 15 min sample incubation time is sufficient to obtain detection limits in the clinically relevant range, and the particle counts have reached a stable threshold, we conclude that the assay volume, sample size, and fluid compartment dimensions provide a robust environment for obtaining easily reproduced results. Fig. 4C depicts the dose response relationship between COVID-19 IgG concentration and counts of bound AuNPs on the PC. The calculated dynamic detection range is from 93.3 pg mL − 1 to 25.7 ng mL − 1 . The immunoassay has a good linear range from 0.1 to 100 ng mL − 1 (Fig. S6) . The R 2 is determined to be 0.995 and all the linear correlation coefficients are listed in Table S2 . The calculated LOD and LOQ is 26.7 ± 7.7 pg mL − 1 and 32.0 ± 8.9 pg mL-1, respectively. It is thus demonstrated that our PRAM-based single-step digital-resolution immunoassay is capable of quantitatively detecting serological antibody against SARS-CoV-2 with high sensitivity in 15 min. It is noteworthy that the high sensitivity of our method rivals that of ELISA or CLIA which typically range 1-100 pg mL − 1 [42] . It is > 50 times more sensitive than commercial pGOLD High Accuracy IgG/IgM Assay Kit (https://www.nirmidas.com/pgold-covid-19-igg-igm-assaykit), which is based on NIR fluorescence and offers a detection limit of ~1.6 ng mL − 1 for COVID-19 IgG [43] . In terms of assay time, our 15-min assay method is comparable to LIFA and substantially more rapid than ELISA and CLIA (>2 h). More importantly, with calibration against concentration standards, it can achieve quantitative detection of COVID-19 antibody which is vital for assessment of the level of immune response and identifying donors for blood transfusion therapy [12, 44] . In summary, we have developed a single-step and washing-free immunoassay for rapid quantitative detection of serological human COVID-19 IgG with digital resolution, high sensitivity, low sample volume, and high selectivity. Taking advantage of the AC + DC assay principle, COVID-19 IgG proteins induce the activation of the functionalized AuNPs in solution and the specific binding of activated AuNPs to the PC, followed by the digital imaging by PRAM. We first demonstrated the proof-of-concept assay for highly sensitive and selective detection of human COVID-19 IgG in buffer and in human serum. We achieved a detection limit of 100 pg mL − 1 with a 15 min assay time. Our method possesses many attractive features, such as easy operation, a simple assay cartridge, and the need of a small volume of serum sample (~4 μL). Although the benchtop PRAM instrument was used in this work, we are developing portable and inexpensive instrument by utilizing LED illumination and a webcam-quality 2-dimensional image sensor that eliminates the spectrometer and requirement for line-scanning. We will describe the design and operation of the portable PRAM in a future publication as we consider how to perform serological assays in point of care settings. Given the broad set of assays that are performed using the sandwich assay approach used here, our method holds great promise for the development of a universal diagnostic platform for multiplexed detection of a variety of analytes, including antigens, proteins additional IgGs, cytokines, and other protein biomarkers. The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper. Detection of SARS-CoV-2 in different types of clinical specimens Improved molecular diagnosis of COVID-19 by the novel, highly sensitive and specific COVID-19-RdRp/hel real-time reverse transcription-PCR assay validated in vitro and with clinical specimens Diagnosing COVID-19: the Disease and tools for detection Development and clinical application of a rapid IgM-IgG combined antibody test for SARS-CoV-2 infection diagnosis Real-time PCR in virology Real-time reverse transcriptionpolymerase chain reaction assay for SARS-associated coronavirus Assay techniques and test development for COVID-19 diagnosis Highly sensitive and specific multiplex antibody assays to quantify immunoglobulins M, A and G against SARS-CoV-2 antigens The trinity of COVID-19: immunity, inflammation and intervention Effectiveness of convalescent plasma therapy in severe COVID-19 patients Evaluation of nucleocapsid and spike protein-based enzymelinked immunosorbent assays for detecting antibodies against SARS-CoV-2 Evaluation of the EUROIMMUN anti-SARS-CoV-2 ELISA assay for detection of IgA and IgG antibodies A serological assay to detect SARS-CoV-2 seroconversion in humans Chemiluminescent immunoassay technology: what does it change in autoantibody detection? Performance verification of detecting COVID-19 specific antibody by using four chemiluminescence immunoassay systems Lateral flow (immuno)assay: its strengths, weaknesses, opportunities and threats. A literature survey Validation and performance comparison of three SARS-CoV-2 antibody assays Fast, portable tests come online to curb coronavirus pandemic ddPCR: a more accurate tool for SARS-CoV-2 detection in low viral load specimens SERS detection of nucleobases in single silver plasmonic nanopores Direct optical sensing of single unlabelled proteins and super-resolution imaging of their binding sites Photonic crystals: emerging biosensors and their promise for point-of-care applications Digital-resolution detection of microRNA with single-base selectivity by photonic resonator absorption microscopy Quantitative analysis of focal adhesion dynamics using photonic resonator outcoupler microscopy (PROM) Photonic crystal surface mode imaging for multiplexed and high-throughput label-free biosensing Development of molecularly imprinted 2D photonic crystal hydrogel sensor for detection of Lkynurenine in human serum 2D photonic crystal hydrogel sensor for tear glucose monitoring Efficient fluorescence resonance energy transfer between quantum dots and gold nanoparticles based on porous silicon photonic crystal for DNA detection Activate capture and digital counting (AC+ DC) assay for protein biomarker detection integrated with a self-powered microfluidic cartridge Flexible sensors for biomedical technology Multimode smartphone biosensing: the transmission, reflection, and intensity spectral (TRI)-analyzer Critical Review: digital resolution biomolecular sensing for diagnostics and life science research DNA nanostructure-based universal microarray platform for high-efficiency multiplex bioanalysis in biofluids Gold nanostructures encoded by non-fluorescent small molecules in polyA-mediated nanogaps as universal SERS nanotags for recognizing various bioactive molecules Microcavity-Mediated spectrally tunable amplification of absorption in plasmonic nanoantennas Plasmon resonances of a gold nanostar Single-molecule enzyme-linked immunosorbent assay detects serum proteins at subfemtomolar concentrations Serum levels of immunoglobulins (IgG, IgA, IgM) in a general adult population and their relationship with alcohol consumption, smoking and common metabolic abnormalities Human serum albumin: from bench to bedside Predicting detection limits of enzyme-linked immunosorbent assay (ELISA) and bioanalytical techniques in general Diagnosis of Zika virus infection on a nanotechnology platform Anti-SARS-CoV-2 virus antibody levels in convalescent plasma of six donors who have recovered from COVID-19 This work is supported by the National Institutes of Health (NIH) R21 AI130562 and R01 AI20683. B. Z. is supported by a Carl R. Woese Institute for Genomic Biology (IGB) fellowship in the Center for Genomic Diagnostics. The authors gratefully acknowledge the members of the Nanosensors Group (NSG) and staff in the Nick Holonyak Jr. Micro and Nanotechnology Laboratory for their support. Supplementary data to this article can be found online at https://doi. org/10.1016/j.talanta.2020.122004.