key: cord-0889431-0bu2ugl8 authors: Farnsworth, Christopher W.; Case, James B.; Hock, Karl; Chen, Rita E.; O'Halloran, Jane A.; Presti, Rachel; Goss, Charles W.; Rauseo, Adriana M.; Ellebedy, Ali; Theel, Elitza S.; Diamond, Michael S.; Henderson, Jeffrey P. title: Assessment of serological assays for identifying high titer convalescent plasma date: 2021-07-06 journal: Transfusion DOI: 10.1111/trf.16580 sha: a84f53ba014775a2ca9cad7b414c9260e14a398b doc_id: 889431 cord_uid: 0bu2ugl8 BACKGROUND: The COVID‐19 pandemic has been accompanied by the largest mobilization of therapeutic convalescent plasma (CCP) in over a century. Initial identification of high titer units was based on dose–response data using the Ortho VITROS IgG assay. The proliferation of severe acute respiratory syndrome coronavirus 2 serological assays and non‐uniform application has led to uncertainty about their interrelationships. The purpose of this study was to establish correlations and analogous cutoffs between multiple serological assays. METHODS: We compared the Ortho, Abbott, Roche, an anti‐spike (S) ELISA, and a virus neutralization assay. Relationships relative to FDA‐approved cutoffs under the CCP emergency use authorization were identified in convalescent plasma from a cohort of 79 donors from April 2020. RESULTS: Relative to the neutralization assay, the spearman r value of the Ortho Clinical, Abbott, Roche, anti‐S ELISA assays was 0.65, 0.59, 0.45, and 0.76, respectively. The best correlative index for establishing high‐titer units was 3.87 signal‐to‐cutoff (S/C) for the Abbott, 13.82 cutoff index for the Roche, 1:1412 for the anti‐S ELISA, 1:219 by the neutralization assay, and 15.9 S/C by the Ortho Clinical assay. The overall agreement using derived cutoffs compared to a neutralizing titer of 1:250 was 78.5% for Abbott, 74.7% for Roche, 83.5% for the anti‐S ELISA, and 78.5% for Ortho Clinical. DISCUSSION: Assays based on antibodies against the nucleoprotein were positively associated with neutralizing titers and the Ortho assay, although their ability to distinguish FDA high‐titer specimens was imperfect. The resulting relationships help reconcile results from the large body of serological data generated during the COVID‐19 pandemic. K E Y W O R D S convalescent plasma, COVID-19, SARS-CoV-2, serology 1 | INTRODUCTION COVID-19 convalescent plasma (CCP) has been one of the primary therapies deployed in the COVID-19 pandemic. In this current iteration of a classic therapy, serological assays to quantify antibodies to the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) spike (S) protein play a critical role in characterizing human immune responses and identifying CCP donors. Commercial SARS-CoV-2 serological assays have accordingly emerged at a rapid pace. Within the first year of the pandemic, more serological assays were available for SARS-CoV-2 than for any other infectious disease, with over 65 emergency use authorizations (EUA) granted for serological testing alone. 1 The CDC and Infectious Diseases Society of America have both defined relatively narrow and limited clinical applications for SARS-CoV-2 serology to include CCP donor identification, infection diagnosis in patients more than 14 days from symptom onset, and seroprevalence determinations. [2] [3] [4] Nevertheless, the clinical utility of these assays has been questioned, 5, 6 in part, due to the challenge of reconciling results from serological assays with clinical outcomes [7] [8] [9] and poor agreement between commercial serological and virus neutralization assays. [10] [11] [12] Identification of CCP with antibody content sufficient for therapeutic CCP use has emerged as a key quantitative application for SARS-CoV-2 serological assays. 2, 5 Anti-S IgG responses in particular were identified early as key correlates of SARS-CoV-2 immunity. Early in the pandemic, the absence of an FDA-approved serological assay was a major obstacle to deploying CCP units with sufficient antibody content. A highly sensitive and specific laboratory developed S-based ELISA was quickly developed 13 and used to identify CCP donors with antibodies to SARS-CoV-2 following RT-PCR confirmed infection. 14 The initial FDA recommendation was to use a minimum specific antibody titer of 1:160, with an ideal titer ≥1:320, as a criterion for CCP donation. 14 A subsequent study demonstrated that infusion of high-titer CCP, defined as a signal of ≥18.45 on the Ortho-Clinical Diagnostics VITROS IgG assay, was associated with lower risk of mortality than low-titer CCP infusion in a large retrospective analysis of patients treated through an FDA expanded access program. 15 A subsequent analysis of this cohort through the Biomedical Advanced Research and Development Authority (BARDA) found that patients receiving CCP with a neutralizing antibody titer >1:250 experienced lower mortality than those receiving units with titers <1:250. 16, 17 Neutralizing antibody assays, however, are highly laborious and require biosafety level 3 facilities if using live SARS-CoV-2, limiting their use primarily to research laboratories. As a result, neutralizing assays were correlated with the Ortho Clinical IgG assay, with a minimum signal of 12.0 distinguishing units with high neutralizing titers. 18 In February 2021, the FDA reissued a letter of authorization for CCP with several revisions to the previous EUA. 19 Importantly, this included a decision to release only high-titer CCP units for patient use. Cutoffs were provided so that multiple serological assays could be used to define high-titer CCP and previously established titers approved by the FDA were modified. The titer to establish high-titer units with the Ortho Clinical assay was lowered from 12.0 to ≥9.5 signal-to-cutoff (S/C), and the original anti-S ELISA threshold was raised from 1:320 to ≥1:2880 in an ELISA performed at Mt. Sinai Hospital. The revised EUA also established cutoffs for distinguishing high-titer units using several other commercial serological assays. For example, the Abbott SARS-CoV-2 IgG assay and the Roche Elecsys anti-SARS-CoV-2 assay were approved for qualifying high-titer units with results ≥4.5 S/C and ≥109 cutoff index (COI), respectively. Little published literature is available to correlate neutralizing antibody titers, commercial serological assays, and anti-S ELISAs. Several studies have assessed the positive percent agreement and negative percent agreement (PPA and NPA) between assays. 10, 11, 20 However, the signal from other commercial, serological assays that best correlates to anti-S ELISA titers of 1:320, neutralizing titers of 1:250, and the Ortho Clinical S/C of 12.0, has not been determined. The purpose of our study was to establish correlations and analogous cutoffs between widely used commercial serological assays, anti-S ELISA, and neutralizing assays with authentic SARS-CoV-2. The resulting relationships will help reconcile results from the large body of serological studies and CCP trials results that continue to emerge during the COVID-19 pandemic. This study was approved by the Washington University Institutional Review Board. Serum specimens were drawn on patients with RT-PCR-confirmed SARS-CoV-2 infection at least 14 days after infection and prior to donation of convalescent plasma. Patient reported demographic information including age, sex, race, comorbidities, and duration of symptoms was collected by survey on each patient. After collection, specimens were immediately frozen in 100 μl aliquots and stored at À80°C until further analysis. Specimens were thawed at room temperature and analyzed within 3 days. Three commercial serological assays and an anti-S ELISA granted EUA at Mt. Sinai Hospital, but used on a research basis for this study, were used to directly measure antibody levels in serum specimens. These assays detected antibodies to SARS-CoV-2 S or nucleocapsid proteins. The anti-S ELISA was performed as previously described. 13 In short, plasma specimens were diluted to 1:30 in PBS, then serially diluted to 1:65,610 in a 96-well plate (Corning, NY). Wells were washed, incubated with a secondary anti-human IgG, followed by another wash step. Wells were then incubated with o-Phenylenediamine dihydrochloride (Sigma-Aldrich, St. Louis, MO) followed by a stop solution (3 M hydrochloric acid). The optical density was then measured at 490 nm and the cutoff for a positive result was determined as an optical density that was three standard deviations above the mean signal from a negative control specimen run with each plate. This signal was extrapolated from the generated curves to quantify the titer. 21 An authentic SARS-CoV-2 neutralization assay was used to measure neutralizing antibody titers. Focus reduction neutralization assays were performed as previously described. 10 SARS-CoV-2 strain n-CoV/USA_WA1/2020 was obtained from the Centers for Disease Control. Virus was propagated in Vero E6 cells in Dulbecco's modified Eagle medium (Corning) that was supplemented with 10% FBS, glucose, L-glutamine, and sodium pyruvate. Patient sera were diluted and incubated with 1 Â 10 2 focus forming units of SARS-CoV-2 for 1 h at 37°C. The plasma/ virus complex was then added to Vero E6 monolayers at 37°C for 1 h. After overlaying with methylcellulose, cells were harvested at 30 h, methylcellulose was removed, and fixed with 4% paraformaldehyde for 20 min. Plates were washed and incubated with 1 μg/ml anti-S antibody (CR3022) and HRP conjugated goat anti-human IgG. Infected cells were visualized with TrueBlue peroxidase substrate (KPL) and quantified using an ImmunoSpot microanalyzer (Cellular Technologies, Cleveland, OH). A minimum of eight dilutions was performed for each specimen, a standard curve generated, and the 1/log10 plasma dilution (EC50) determined as the dilution at which 50% of the cells were infected. All specimens were analyzed on three commercially available serological assays. The Ortho Clinical VITROS SARS-CoV-2 IgG (Ortho Clinical Diagnostics, Raritan, NJ) assay was performed on an Ortho Clinical VITROS 5600 Immunodiagnostic System and targets antibodies to the S protein. The Abbott SARS-CoV-2 IgG assay (Abbott Diagnostics, Abbott Park, IL) was performed on an Abbott Architect i2000 and detects antibodies to the nucleocapsid protein. The Roche Elecsys anti-SARS-CoV-2 assay (Roche, Indianapolis, IN) was performed on a Cobas e601 and identifies antibodies to the nucleocapsid protein. All commercial assays have FDA EUA as qualitative methods and were performed and interpreted according to the manufacturer's instructions. The positive cutoffs for each assay are 1.0 (S/C), 1.4 (S/C), and 1.0 (COI) for the Ortho Clinical, the Abbott, and the Roche assays, respectively. All three assays report a numeric signal to cut-off that is the amount of signal generated by the sample for each assay relative to the signal from a single calibrator. Cutoffs used to assess each assay were identified previously by either literature review or the FDA [14] [15] [16] [17] [18] [19] and can be found in Table S1 . Normality of the distribution of the reported signal from each assay was assessed using the D'Agostino and Pearson test. Correlation between the various assays was assessed using nonparametric spearman correlation. Association between assays was compared with least squares regression to calculate intercept, slope, and r 2 . Receiver operator curves (ROC) were plotted to assess the ideal cutoffs using Youden Index to establish cutoff with maximum positive and negative percent agreement relative to the Ortho Clinical assay at a cutoff of 12.0 S/C and a neutralizing antibody titer of 1:250. Final cutoffs for distinguishing high and low-titers units by each assay were established by averaging the ideal cutoffs established by Youden's index for the Ortho Clinical and neutralizing assays. 6.6-61.5 COI) ( Figure 1E ). These results are consistent with substantial variability in neutralizing antibody responses to SARS-CoV-2 proteins among recovered adults. 12 Linear relationships between each commercial assay were defined relative to the neutralizing antibody titer, anti-S ELISA IgG titer, and Ortho Clinical assay ( Figure 2 ). Each assay was positively correlated with neutralizing antibody titers. Relative to the neutralizing assay, the assay with the highest correlation was the anti-S ELISA (r = 0.76, 95%; confidence interval [CI] 0.64-0.84) and the lowest was the Roche assay (r = 0.45, 95% CI; 0.25-0.62, Figure 2A ). Relative to the anti-S ELISA, the Ortho Clinical assay had the highest correlation (r = 0.90, 95% CI; 0.84-0.93) followed by the Abbott assay (r = 0.76, 95% CI; 0.65-0.84) and the Roche assay (r = 0.61, 95% CI; 0.45-0.74, Figure 2B ). Of the commercial assays, there was comparable correlation between the Ortho Clinical and the Abbott assay (r = 0.76, 95% CI; 0.65-0.84) and the Ortho Clinical and Roche assay Table 3 . Dashed gray lines represent the FDA-approved cutoffs from EUA. EUA, emergency use authorization; ROC, receiver operator curve; SARS-CoV-2, severe acute respiratory syndrome coronavirus 2 (r = 0.64, 95% CI; 0.48-0.76, Figure 2C ). The slopes and intercepts, r 2 , and interpolated cutoff for each commercial assay relative to the neutralization assay, the anti-S ELISA, and the Ortho Clinical assay are found in Table S2 . ROC curves were generated for each serological assay using the neutralizing cutoff of 1:250 ( Figure 3A ) and the Ortho Clinical cutoff of 12.0 S/C ( Figure 3B ). The assay with the greatest correlation to the neutralizing assay was the anti-S IgG ELISA, with an area under the curve (AUC) of 0.865 (95% CI; 0.783-0.947). Relative to the Ortho Clinical assay, all assay's AUC values were >0.8, with the anti-S IgG ELISA having the best correlation (AUC = 0.927). The cutoff that maximized the PPA and NPA for each assay relative to the neutralizing assay cutoff of 1:250 was averaged with the cutoff that maximized the PPA and NPA relative to the Ortho Clinical assay cutoff of 12.0 to establish ideal cutoffs ( Table 2) . With this approach, the average cutoff for distinguishing high-and low-titer units by the Abbot assay was 3.87 S/C, the Roche assay was 13.82 COI, the anti-S IgG ELISA was 1:1412 1:219 for the neutralization assay and 15.34 S/C for the Ortho-Clinical assay ( Table 2) . ROC curves were also generated relative to the low and high neutralizing titers of 1:150 and 1:500, respectively ( Figure S1 ), and for the low and high Ortho-Clinical cutoffs of 4.62 S/C and 18.45 S/C, respectively ( Figure S2 ). Specimens were segregated as low-or high-titer using the neutralizing assay cutoff of 1:250 ( Figure 4A ) and the Ortho Clinical cutoff of 9.5 S/C or 12 S/C and scatterplots generated ( Figure 4B,C) . Using the cutoffs established in Table 2 (dotted black lines), all four assays (Ortho Clinical, ELISA, Abbott, and Roche) demonstrated comparable performance relative to the neutralizing cutoff of 1:250 for identifying patients with high and low antibody titers. The overall agreement using the derived cutoffs with the neutralization assay cutoff of 1:250 was 78.5% for Abbott, 74.7% for Roche, 83.5% for the anti-S ELISA, and 78.5% for the neutralization assay (Table S3) . Specimens also were segregated as low-or high-titer using the Ortho Clinical assay cutoff of 12.0 S/C and 9.5 S/C with similar results (Figure 4B,C) . Decreasing the signal for identifying high-titer plasma on the Ortho Clinical assay led to improved NPA and PPA with the Abbott and Roche assay and an improved NPA with a modest decrease in PPA with the anti-S IgG ELISA and the neutralization titer. Relative to the FDA Abbott cutoff of 4.5 S/C (dashed gray line) for identifying high-titer units, 5 additional specimens would have been labeled as lowtiter by the Abbott but high-titer by the Ortho Clinical assay. Using the FDA cutoff of ≥109 COI for the Roche assay, all 79 specimens would have been qualified as lowtiter units. The tiered neutralizing and Ortho Clinical cutoffs used to identify patients with low medium and high titers are found in Table S4 and Figures S3, S4. Despite accumulating evidence of associations between commercial serological assay values and neutralizing antibody titers with human immunity and CCP efficacy, few published studies permit correlation between the assay formats in use. Here, we tested three widely used commercial serological assays, an EUA anti-S IgG ELISA, and neutralizing antibodies and correlated each assay with the ideal cutoffs for establishing high-titer plasma. An important finding from this study is that commercial assays and the anti-S ELISA performed similarly for identifying specimens with high neutralization titers. Our approach using ROC curves that established maximal PPA and NPA-identified cutoffs made these assays largely interchangeable for identifying high-titer CCP. The antigenic target of the assay did not change the PPA and NPA, with assays measuring antibodies to the viral S protein performing similarly to those measuring antibodies to the nucleocapsid protein. This finding is similar to other studies from acutely infected patients with severe symptoms and patients with mild symptoms. 10, 22 It is notable that FDA's reissued CCP authorization letter incorporated multiple EUA serological assays, several of which are included in this report. 16 The cutoff for the Abbott assay established here (≥3.87 S/C) is similar to the cutoff of ≥4.5 S/C established by the FDA. This is despite the lowered Ortho Clinical assay S/C cutoff (from 12 to 9.5) in the reissued CCP EUA. 16 Nonetheless, the Abbott assay cutoff of 3.87 S/C had better NPA with the Ortho Clinical cutoff of 9.5 S/C than the cutoff of 12.0 S/C, without sacrificing PPA. In contrast, the FDA-approved cutoff of 109 COI for the Roche assay would have disqualified all units as low-titer, with a signal approximately 10-fold higher than the ideal cutoff identified in this study. The derived cutoff from this study with the anti-S IgG ELISA (1:1412) also was lower than that established by the FDA (1:2880) . This resulted in a considerable difference in PPA, with far more convalescent donor units being excluded under the FDA cutoffs for the ELISA than for the Ortho Clinical assay. The cutoff identified in our study that best distinguishes neutralizing titers ≥1:250 with the Ortho Clinical assay was 15.9 S/C, higher than the original cutoff of 12.0 S/C from the FDA. An S/C of 9.0 on the Ortho Clinical assay correlated to a neutralizing titer of $1:100. This is notably similar to the neutralizing titer of 1:104 we found to be sufficient to reduce weight loss in mice, a common outcome of enhanced disease severity in animal models. 23 Nevertheless, the neutralizing assay used in this study cannot be assumed to perform similarly to the assay used in the BARDA study 14, 15 due to nonstandardization of SARS-CoV-2 strains, cell lines, and reagents/procedures. These differences underlie the difficulties in harmonizing SARS-CoV-2 serological assay results. These discrepancies should be considered when attempting to use any serological assay as a proxy for measuring neutralizing antibodies. Several studies demonstrate survival benefits with early, high-titer CCP administration and in patients with hematological malignancies, implying a continued need to identify CCP or immune globulin donors. 15, 24, 25 This study attempts to harmonize several commercially available assays that have been extensively studied and published. While numerous studies have addressed the analytical performance of available serological assays, little correlative information is available in the published literature to relate multiple serological assay results. The data presented here suggest that multiple commercial assays could be used to identify CCP donors with high levels of neutralizing antibodies. However, this study highlights the relatively moderate to weak correlation between commercially available serological assays and neutralizing antibody titers. This directly impacts the PPA and NPA of the various assays for determining donors with high-titer units. These results highlight the potential shortcomings of a single cutoff for identifying high-titers units. There are several limitations associated with this study. Among the greatest limitations is the lack of standardization between assays, even among the same manufacturers. This was previously noted with the neutralization assay, though the same is true among commercial assays. Since several of the assays have been designated as qualitative (i.e. the Roche, Abbott, and Ortho Clinical assays), there is limited evidence that semi-quantitative results are comparable between different instruments by the same manufacturer above the cutoff. For example, since there is no material to verify linearity at higher concentrations, a result of 15 S/C at one institution using the Ortho Clinical assay may vary from the Ortho Clinical assay at another institution. This may underlie the differences between the established cutoff and FDA cutoff for the Roche assay. In general, this problem will continue to plague the field until quantitative assays are universally adopted and standardized to SARS-CoV-2 antibody reference material, such as that recently released by the World Health Organization. 26 This is further complicated by unclear direction as to how to report a qualitative assay result as quantitative under an EUA, which does not permit modification of the manufacturer's Instructions for Use. Another limitation of the current study is that a limited number of assays were evaluated, limiting the generalizability of results. It is also important to note that these specimens were obtained early during the course of the pandemic, and that continued viral evolution (which may lead to extensive antigenic changes in the S protein) means that the quantitative relationships in this manuscript could become outdated. Ongoing studies are required to confirm the relationships established here in patients infected with SARS-CoV-2 variants and using neutralizing assays that utilize SARS-CoV-2 variants. Furthermore, cutoffs were derived using previously established cutoffs on the Ortho Clinical assay (12.0 S/C) and neutralizing assay (1:250) which may have caused a selection bias. However, current methods for identifying the FDA-approved cutoffs use results from a single neutralizing assay at a titer of 1:250. Finally, while these results provide evidence of that the cutoffs identified by the FDA are helpful for identifying high-titer CCP units, there were several specimens by each assay that were not in agreement. These specimens demonstrate the requirement for further study before the cutoffs proposed by the FDA are modified. In conclusion, we demonstrate that assays based on nucleoprotein antibodies (Roche, Abbott) and neutralization were positively associated with Ortho assay results (anti-S), though their ability to distinguish FDA high-titer specimens was marginal. Association with a traditional ELISA serologic test was high. All assays were positively associated with neutralization titers, though associations were strongest with S-based assays. Diagnostics. JPH-Consulting: Immunome, Inc. EST-Consulting/Advisory Board: Roche Diagnostics, Accelerate Diagnostics, Serimmune Inc. JBC, KH, REC, JAO, RP, CWG, AMR, AE, and MSD have disclosed no conflicts of interest. Christopher W. Farnsworth https://orcid.org/0000-0001-7169-3850 Consulting: Roche REFERENCES 1. FDA. 2020 Infectious Diseases Society of America. 2020. IDSA COVID-19 antibody testing primer Interim guidelines for COVID-19 antibody testing. Centers for Disease Control and Prevention The Infectious Diseases Society of America guidelines on the diagnosis of COVID-19: molecular diagnostic testing SARS-CoV-2 serology: much hype, little data The role of antibody testing for SARS-CoV-2: is there one? Clinical performance of two SARS-CoV-2 serologic assays Clinical performance of the Roche SARS-CoV-2 serologic assay Performance characteristics of four high-throughput immunoassays for detection of IgG antibodies against SARS-CoV-2 Association between SARS-CoV-2 neutralizing antibodies and commercial serological assays Evaluation of high-throughput SARS-CoV-2 serological assays in a longitudinal cohort of patients with mild COVID-19: clinical sensitivity, specificity and association with virus neutralization test Sex, age, and hospitalization drive antibody responses in a COVID-19 convalescent plasma donor population A serological assay to detect SARS-CoV-2 seroconversion in humans Recommendations for investigational COVID-19 convalescent plasma. FDA Convalescent plasma antibody levels and the risk of death from COVID-19 COVID convalescent plasma EUA COVID-19 convalescent plasma EUA decision FDA issues emergency use authorization for convalescent plasma as potential promising COVID-19 treatment, another achievement in administration's fight against pandemic Revised COVID convalescent plasma EUA Commercial serology assays predict neutralization activity against SARS-CoV-2 Sinai Laboratory COVID-19 ELISA IgG antibody test EUA summary Antibody responses to SARS-CoV-2 in patients with COVID-19 Human neutralizing antibodies against SARS-CoV-2 require intact Fc effector functions for optimal therapeutic protection Effectiveness of convalescent plasma therapy in severe COVID-19 patients Convalescent plasma and improved survival in patients with hematologic malignancies and COVID-19 WHO. 2021. First WHO International Standard anti-SARS-CoV-2 immunoglobulin (human Assessment of serological assays for identifying high titer convalescent plasma This study utilized samples obtained from the Washington University School of Medicine's COVID-19 biorepository, which is supported by: the Barnes-Jewish Hospital Foundation; the Siteman Cancer Center grant P30 CA091842 from the National Cancer Institute of the National Institutes of Health; and the Washington University Institute of Clinical and Translational Sciences grant UL1TR002345 from the National Center for Advancing Translational Sciences of the National Institutes of Health. Jeffrey P. Henderson acknowledges support from National Institute of Diabetes and Digestive and Kidney Diseases grant RO1DK111930 and the Longer Life Foundation. The content is solely the responsibility of the authors and does not necessarily represent the view of the NIH. Specimen analysis was funded by the Washington University Department of Pathology and Immunology and the Mayo Clinic. No industry sponsors were involved with the collection, analysis, and presentation of data, drafting of the manuscript, or the decision to publish.