key: cord-0882528-2d2tde0n authors: Jafferali, Mohammed Hakim; Khatami, Kasra; Atasoy, Merve; Birgersson, Madeleine; Williams, Cecilia; Cetecioglu, Zeynep title: Benchmarking virus concentration methods for quantification of SARS-CoV-2 in raw wastewater date: 2020-10-14 journal: Sci Total Environ DOI: 10.1016/j.scitotenv.2020.142939 sha: 7c398578fb9d59e8fe85bc14aad318dfc57548ad doc_id: 882528 cord_uid: 2d2tde0n Wastewater-based epidemiology offers a cost-effective alternative to testing large populations for SARS-CoV-2 virus, and may potentially be used as an early warning system for SARS-CoV-2 pandemic spread. However, optimal viruses are highly diluted in wastewater, and a validated method to be concentrated for further processing and suitable reference viruses are the main needs to be established for reliable SARS-CoV-2 municipal wastewater detection. For this purpose, we collected wastewater from two European cities during the Covid-19 pandemic and evaluated the sensitivity of RT-qPCR detection of viral RNA after four concentration methods (two variants of ultrafiltration-based method and two adsorption and extraction-based methods). Further, we evaluated one external (Bovine corona virus) and one internal (pepper mild mottle virus) reference virus. We found a consistently higher recovery of spiked virus using the modified ultrafiltration-based method. This method also had a significantly higher efficiency (p-value <0.01) for wastewater SARS-CoV-2 detection. The ultracentrifugation method was the only method that detected SARS-CoV-2 in the wastewater of both cities. The pepper mild mottle virus was found to function as a potentially suitable internal reference standard. The first cases of the current global pandemic of severe acute respiratory syndrome corona virus 2 (SARS-CoV-2) infections were reported in December 2019, in China (WHO, 2020). Survival of coronaviruses in water and wastewater has been previously confirmed (Gundy et al., 2009) , making wastewater based epidemiology (WBE) a robust tool in developing an early warning or surveillance system for infections or rises of the SARS-CoV-2 virus. WBE has been previously used as a successful approach to grasp the severity and prevalence of pathogenic outbreaks in Sweden (Hellmér et al., 2014) and Israel (Kopel et al., 2014) . Therefore, several studies have focused on identification of SARS-CoV-2 in wastewater during the current pandemic (La Rosa et al., 2020b; Medema et al., 2020; Randazzo et al., 2020) . One of the hurdles is the recovery methods, which are primarily developed for nonenveloped viruses. However, the novel SARS-CoV-2 belongs to the coronaviridae family (Gorbalenya et al., 2020) of enveloped viruses with single-stranded RNA. Different functional groups on the outer layer of enveloped and nonenveloped viruses impact wastewater recovery methods (Ye et al., 2016) . Ahmed et al., (2020) have recently compared the efficiency of seven different concentrations methods for the recovery of SARS-CoV-2. Nevertheless; recovery J o u r n a l P r e -p r o o f Journal Pre-proof methods of enveloped viruses, their efficiencies, and internal and external surrogates (reference viruses) require further research (La Rosa et al., 2020a) as the existing information indicate different recovery efficiencies for each method (Ahmed et al., 2020b ). In the current study, the sensitivity of four virus concentration methods were assessed in the perspective of external and internal reference viruses for wastewater samples from two countries: Sweden and Italy. The regions of Stockholm and North of Italy were chosen, as they represented regions with high case numbers of SARS-CoV-2 infection. The four different virus concentration techniques examined in this study were; 1) ultrafiltration 1.A) modified ultrafiltration 2) adsorption-vacuum filtration and 2.A) centrifugation combined with adsorption-vacuum filtration. Bovine coronavirus (BCoV), from betacoronavirus genus, the same genus as SARS-CoV-2 is endemic in cattle. This single stranded positive sense enveloped RNA was used as an external reference virus. Pepper mild mottle virus (PMMoV), a non-enveloped single stranded RNA virus, was assessed as internal reference virus. PMMoV, from the tobamovirus family, is an indicator of fecal contamination in wastewater as it is found abundantly in various aquatic environments (Kitajima et al., 2018) . Following recovery of virus, RNA isolation and one-step reverse transcriptase quantitative polymerase chain reaction (RT-qPCR) were conducted to determine the efficiency of each virus concentration technique. Twenty l of BCoV (stock prepared in human colorectal tumor cell line HRT-18G, ATCC CRL-11663 (Christensen and Myrmel, 2018) ) were spiked to 50 ml of all wastewater samples as an external reference. In this study, four approaches were tested. Method 1-Ultrafiltration: wastewater samples were centrifuged at 4600xg for 30 min at 4C in order to remove large and coarse particles, and the supernatant (approx. 40-50 ml) was filtered through 10 kDa cut off centrifugal ultrafilters (Sartorius) at 1500xg for 15 min (Megastar 1.6R benchtop centrifuge) (Medema et al., 2020) . Method 1.A-Double Ultrafiltration (Method 1 modified): The obtained concentrate from Method 1 was centrifuged a second time (10kDa cut off Sartorius centrifugal ultrafilters, 1500xg, 15 min, 4C). The obtained concentrate of method 1 and 1.A varied between 3-5 ml and 0.5 to 1.5 ml, respectively. Method 2-Adsorption-Extraction: MgCl 2 (final concentration 25 mM) was added to the wastewater samples, followed by filtration through 0.45-m pore size electronegative membranes (Supor 450, plain) (Ahmed et al., 2020b) . Method 2.A-Centrifugation combined with adsorption-extraction (Method 2 modified): Wastewater samples were centrifuged at 4600xg for 30 min at 4C in order to remove the large and coarse particles before addition of MgCl 2 (final concentration 25 mM). The obtained concentrate was then passed through 0.45-m pore size electronegative membranes (Supor 450, plain). All concentrated samples were stored at -80 C until further analysis. RNA from Method 1 and 1.A processed municipal wastewater was extracted by adding three volumes of Trizol LS reagent for liquid samples (Thermofisher Scientific) to one volume of J o u r n a l P r e -p r o o f Journal Pre-proof concentrated wastewater. For each ml of Trizol-wastewater mixture 0.2 ml of chloroform (Sigma-Aldrich) was added, and the aquous phase purified by miRNeasy Mini Kit (Qiagen, Chatsworth, CA). RNA from filter papers obtained from Method 2 and 2.A were isolated using RNeasy PowerMicrobiome Kit (Qiagen, Chatsworth, CA). The RNA was eluated in 50 µl, and stored at −80 °C. Primers targeting the nucleocapsid (N) gene were used to detect the SARS-COV-2 gene. The specificity of N-gene primer set against human corona viruses and other viruses (respiratory) has been previously reported by (Medema et al., 2020) . For the internal municipal wastewater virus reference, primers targeting PMMoV were used, and for the external (spike) reference virus, primers targeting BCoV were used. All primers are listed in Table 1 . Preliminary experiments of waste water samples showed that inhibition of the RT-qPCR reaction was reduced by addition of Bovine Serum Albumin (BSA) to the reaction mixture. Therefore, in all reaction 2 µl of 4 mg/ml BSA (Sigma-Aldrich) was added. For each reaction either 8 µl (N gene detection) or 2 µl (PMMoV and BCoV detection) of RNA template was used. This corresponds to 8 ml initial wastewater volume per SARS-CoV-2 RT-qPCR reaction, and 2 ml for PMMoV detection. Since the same initial volume of wastewater sample were used for all samples and methods (for the same virus), the results are directly comparable. SYBR Green chemistry was used to detect the expression of genes, and RNA from inactivated cultured human SARS-CoV-2 (gift from the Public Health Agency of Sweden) and BCoV were used as positive controls. Negative controls were included in each qPCR run. The reaction was performed according to the manufacturer's recommendations using iTaq universal SYBR Green one-step kit (Bio-Rad) and a final reaction volume of 20 µl (SARS-CoV-2) or 10 µl (PMMoV, BCoV). Thermal cycling (50°C 10 min, 95°C J o u r n a l P r e -p r o o f Journal Pre-proof (Bio-Rad) machine were performed. Melting curve detection (65 °C to 95 °C with increment of 0.5 °C for 5 sec) were analyzed for all included genes and compared to positive controls, to ensure specific amplification. Reactions were considered positive if the cycle threshold (Ct) was below 40 cycles with a single melting peak at correct temperature. RNA was extracted from 200 µl of cultured SARS-CoV-2 at 6 x 105 plaque-forming unit (PFU)/ml, and 80 µl of BCoV at 4.5 x 105 50% tissue-culture-infective dose (TCID50)/ml. Ten-fold serial dilutions of the RNAs were prepared and RT-qPCR performed as described above. Standard curves were generated from the log-linear regression of Ct values of replicates, and the amplification efficiencies for SARS-CoV-2 and BCoV were calculated (Nolan et al., 2013) . The lowest number of diluted standards detected in duplicate assays was considered limit of detection (LOD) for the RT-qPCR assay. The presence of qPCR inhibitor in the concentrated municipal wastewater RNA sample was subsequently assessed using the PMMoV qPCR assay. RNA templates were added in series of 1 µl, 2 µl and 4 µl. The qPCR reaction was set up as described (section 2.4). The expression of PMMoV gene was analyzed alongside non-template controls and corresponding amplification efficacy calculated (Nolan et al 2013) and compared to that of RNA from cultured samples. First, we investigated if SARS-CoV-2 were present in the collected samples. Four out of five samples tested positive for SARS-CoV-2 with N-gene primers ( Italy 2) it was only detected in one of two technical replicates. This suggests border limit detection in the assay, possibly due to low occurrence of SARS-CoV-2 in the municipal wastewater during the sample collection week or indicative of varying presence of inhibitors. To be noted, in quantitative gene expression analysis, SYBR Green and TaqMan are the two commonly used methods. The one used here, SYBR Green, is cheaper and does not require additional probes (Valasek and Repa, 2005) . These are benefits during the pandemic, as there was less shortage of these reagents. This method can be less accurate, but by including melting curve analysis and positive controls, accuracy can be ensured. Next, we compared the detection sensitivity taken the different concentration methods into account. We evaluated three viruses, PMMoV, the spiked BoCoV and SARS-CoV-2. As presented in Table 3 and Figure Table 3 ). Again, the adsorption and extraction methods 2 (33.5±2) and 2.A (34.0±2) presented lower detection efficiencies. The recovery rate of the spiked virus was further calculated, by comparing the RT-qPCR detection of RNA extracted from equivalent amount of spiked virus. The recovery rate was substantially higher for Method 1 and 1.A, compared to method 2 and 2.A ( Figure 1B) . Of note is that the recovery was low (less than 10% in most samples) also by Method 1 Figure 1C ). Thus, the centrifugal ultrafilter methods, Method 1 and 1.A, enabled higher recovery rate and virus detection ability. As we concluded that Method 1 and 1.A were more sensitive in regards to detection of all three viruses, but still did not detect SARS-CoV-2 in all replicates, we were interested in whether inhibitors affected the detection. We had already added BSA to counter inhibition, but in order to investigate remaining inhibition we calculated qPCR amplification efficiencies for the SARS-CoV-2 and BCoV virus primers used in this set up. We first calculated amplification efficiency using RNA purified from cultured viruses. With a theoretical optimal doubling of DNA molecules for each replication cycle, the amplification efficiency should be 100%. Desired amplification efficiencies range from 90% to 110%. Our calculations showed that amplification of pure SARS-CoV-2 and BCoV virus RNA (not from waste water) resulted in the desired range of 90% and 99.6%, respectively ( Figure 2A) . Further, the limit J o u r n a l P r e -p r o o f Journal Pre-proof of detection (LOD) for SARS-CoV2 was found to be0.12 PFU/reaction and for BCoV to 0.045 TCID 50 /reaction. It has to be noted that PFU represent the number of infectious virus particle capable of lysing the host cell and TCID 50 represent the dose that infects 50% of the cells (Dulbecco and Vogt, 1954; Khatib et al., 1980) , and they do not directly indicate virus copy number. Next, we calculated amplification efficiencies for PMMoV in two municipal wastewater RNA samples (Stockholm 3 and North of Italy 1following three virus concentration methods (Method 1, 1.A and 2) . This yielded amplification efficiencies of 67% to 116% (Figure 2 ). An efficiency below 90% is indicative of non-optimal conditions, such as presence of inhibitors. Method 1 (single ultracentrifugation) or Method 1.A (double centrifugation) generated samples with the highest qPCR efficiencies, whereas the filter paper concentration (Method 2) had the lowest efficiency for both the Stockholm (67.3%) and Italian (85.9%) samples. We conclude that inhibitors appear to impact the amplification using the adsorption and extraction method, which likely contributes to the markedly higher Ct values and resulting lower sensitivity. BCoV, which belongs to the Coronavirus genus in the Coronaviridae family, order Nidovirales, has similar genetic and serological properties as well as host range with other mammalian coronaviruses (Valarcher and Hägglund, 2014) . Therefore, BCoV is classified in the same group of other mammalian coronaviruses, such as rat coronavirus, human enteric coronavirus and human coronaviruses (Valarcher and Hägglund, 2014 ). In the current study, BCoV was selected as surrogates to calculate the recovery rate during sample processing and filtration, based on the similar properties. By adding a known amount of BCoV before filtration, we could estimate loss during filtration, RNA extraction and RT-qPCR analysis. Furthermore, PMMoV is an indicator for fecal contamination in the water sources owing to it J o u r n a l P r e -p r o o f Journal Pre-proof is globally distributed and present in various water sources in greater abundance than human pathogenic viruses, without substantial seasonal fluctuations (Kitajima et al., 2018) . PMMoV has been used for the detection of pathogenic enteric viruses because of increased concentrations of PMMoV tend to be correlated with increased fecal contamination (Kitajima et al., 2018) . In the current study, PMMoV was selected for normalization of external factors such as flow rate of the wastewater, which is changed based on wet and dry season/periods. Such fluctuation would affect the concentration of both PMMoV and SARS-CoV-2 in the samples. After concluding that the methods exhibited varying recovery rates for the spiked BCoV, and that they varied significantly in their detection of PMMoV and SARS-CoV-2 detection, we explored whether the internal or external reference viruses were suitable as calibrators in order to compare virus levels between samples. First, we plotted the level of PMMoV detection in relation to BCoV recovery ( Figure 3A ). This showed that the normalizing to the recovery rate of the spike can adjust the comparison to some extent, but not completely. A perfect callibration would result in equal levels of PMMoV in relation to the spike, but this was not achieved. Next, we normalized SARS-CoV-2 detection to PMMoV ( Figure 3B ) or BCoV recovery rate ( Figure 3C ). Both normalizations indicate that Stockholm 3 had higher levels of SARS-CoV-2 than Stockholm 1 and 2. However, the two normalization strategies rendered different relative values for the Italian sample (higher level in North of Italy 2 than the Stockholm 3 in relation to PMMoV, but lower when normalizing to spike recovery only). Since March 2020, there are many ongoing studies relating to the detection of SARS-CoV-2 in municipal wastewater (Ahmed et al., 2020a; Daughton, 2020; Nghiem et al., 2020) . However, there is no standardized method. The most important need in wastewater-based epidemiology is to develop or improve and evaluate a standardized and sensitive method for the detection (Kitajima et al., 2020) . Ahmed et al. (2020) compared seven concentration methods by using murine hepatitis virus as external reference (surrogate) and found that the best recovery was obtained from MgCl 2 adsorption and vacuum filtration (Method 2 in the current study), in wastewater samples from Australia. This approach is cheaper and easier. However municipal wastewater characteristics might change across different countries, regions, etc. (Pons et al., 2004) . The methods may also perform differently depending on type of virus (Lu et al., 2020) . In our study, municipal wastewater samples and SARS-CoV-2 from two different countries (Sweden and Italy) were tested using four different approaches in order to find out the best applicable and sensitive virus concentration method for these conditions. We also tested two reference viruses: 1) PMMoV which naturally exist in municipal wastewater, and 2) spiked animal pathogen BCoV which belongs to same genus as SARS-CoV-2. In the light of our findings: 1-Ultrafiltration and modified ultrafiltration (Method 1 and 1.A) had higher recovery efficiencies than adsorption and extraction methods (Method 2 and 2.A). The latter two appeared to accumulate more PCR inhibitors. Thus, in opposition to the results obtained by Ahmed et al. (2020) , the current study showed a better performance of the ultracentrifuge-based methods in terms of recovery efficiency and capacity for viral detection. 3-Internal reference virus (PMMoV) was found to be a sensitive and representative standard, and adding external surrogate (BCoV) did not add meaningful additional information. We conclude that PMMoV internal standard is sufficient to inform of relative recovery and to normalize between samples. SARS-CoV-2 N gene detection normalized to input and recovery rate of spiked BCoV. Student's t-test were used to calculate statistical significance (*p<0.05, **p>0.01, ***p<0.001). J o u r n a l P r e -p r o o f ☒ The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper. ☐The authors declare the following financial interests/personal relationships which may be considered as potential competing interests: J o u r n a l P r e -p r o o f Comparison of virus concentration methods for the RT-qPCR-based recovery of murine hepatitis virus, a surrogate for SARS-CoV-2 from untreated wastewater Making waves: Coronavirus detection, presence and persistence in the water environment: State of the art and knowledge needs for public health Coagulant residues' influence on virus enumeration as shown in a study on virus removal using aluminium, zirconium and chitosan Wastewater surveillance for population-wide Covid-19: The present and future Plaque formation and isolation of pure lines with poliomyelitis viruses The species Severe acute respiratory syndrome-related coronavirus: classifying 2019-nCoV and naming it SARS-CoV-2 Age-dependent pathogenicity of group B coxsackieviruses in Swiss-Webster mice: Infectivity for myocardium and pancreas SARS-CoV-2 in wastewater: State of the knowledge and research needs Pepper mild mottle virus as a water quality indicator. npj Clean Water Lessons from a public health emergency -Importation of wild poliovirus to Israel Coronavirus in water environments: Occurrence, persistence and concentration methods -A scoping review First detection of SARS-CoV-2 in untreated wastewaters in Italy Primary concentration -The critical step in implementing the wastewater based epidemiology for the COVID-19 pandemic The COVID-19 pandemic: Considerations for the waste and wastewater services sector Long. Case Stud Good Practice Guide for the Application of Quantitative PCR (qPCR) Wastewater characteristics in Europe -A survey SARS-CoV-2 RNA in wastewater anticipated COVID-19 occurrence in a low prevalence area Bovine coronavirus, in: Infectious and Parasitic Diseases of Livestock The power of real-time PCR