key: cord-0881979-m99g2opd authors: de Ree, Roy; Willemsen, Jorn; te Grotenhuis, Gilbert; de Ree, Rick; Kolkert, Joé; Peppelman, Malou title: Continuous monitoring in COVID-19 care; a retrospective study in time of crisis date: 2021-04-10 journal: JAMIA Open DOI: 10.1093/jamiaopen/ooab030 sha: cb8909a881e306561d096dac667721362b52d458 doc_id: 881979 cord_uid: m99g2opd BACKGROUND: A new monitoring system was implemented to support nursing staff and physicians on the COVID-19 ward. This system was designed to remotely monitor vital signs, to calculate an automated Early Warning Score (aEWS) and to help identify patients at risk of deterioration. METHODS: Hospitalized patients who tested positive for SARS-CoV-2 were connected to two wireless sensors measuring vital signs. Patients were divided into two groups based on the occurrence of adverse events during hospitalization. Heart and respiratory rate were monitored continuously and an automated EWS was calculated every 5 minutes. Data were compared between groups. RESULTS: Prior to the occurrence of adverse events, significantly higher median heart and respiration rate and significantly lower median SPO2 values were observed. Mean and median automated EWS were significantly higher in patients with an adverse event. CONCLUSION: Continuous monitoring systems might help to detect clinical deterioration in COVID-19 patients at an earlier stage. LAY SUMMARY: A new monitoring system was implemented to support nursing staff and physicians on the COVID-19 ward. This system was designed to remotely monitor vital signs, like respiratory rate, heart rate and the oxygen level in the blood. These parameters were used to calculate an automated early warning score which helps to identify patients at risk of deterioration. Hospitalized patients who tested positive for SARS-CoV-2 were connected to two wireless sensors. Heart and respiratory rate were monitored continuously and an automated EWS was calculated every 5 minutes. Data were compared between patients at the COVID-19 ward and patients who were transported to the ICU or died. COVID patients at the ICU or those who died had significantly higher median heart and respiration rate and significantly lower median oxygen levels. These findings showed that continuous monitoring systems might help to detect clinical deterioration in COVID-19 patients at an earlier stage. Monitoring vital signs is essential in detecting clinical deterioration in hospitalized patients. The Early Warning Score (EWS) has proven to be a very useful tool in identifying patients at risk of deterioration. [1] This bedside assessment of six physiological parameters (respiratory rate, oxygen saturation, temperature, blood pressure, pulse rate, level of consciousness) provides the nursing staff a tool to evaluate the clinical condition of patients and notify rapid response teams quickly in case of imminent clinical instability. This might shorten the delay to intervention and improve patient outcome. [2, 3] However, manual monitoring is time consuming thus imposing a substantial burden on the nursing staff. Innovative sensor technology allows more frequent assessment of physiological parameters and, in theory, detects clinical deterioration in an earlier stage. [4] Several recent papers have shown that current wireless monitoring systems are capable of detecting changes in vital signs in patients who develop adverse events. These measurements correlate well with gold standard measurements. [4] [5] [6] However, due to the heterogeneity of systems and sensors used in different studies there is of yet lacking evidence supporting universal adoption of continuous monitoring. [7, 8] Different patient populations might need their own specific systems. We adjusted a monitoring system, originally designed to detect clinical deterioration in electively admitted patients to support nursing staff and physicians on the COVID-19 ward. This system was designed to remotely monitor vital signs (pulse rate, respiratory rate and oxygen saturation), calculate an automatic Early Warning Score (aEWS) and used to try to identify COVID-19 patients at risk of deterioration. In this overview we describe the system used, the lessons learned and the challenges yet remaining. https://mc.manuscriptcentral.com/jamiao All patients who had been tested positive for SARS-CoV-2 and admitted to our hospital were considered at risk for rapid clinical deterioration and were connected to two wireless sensors (figure 1). Vital parameters obtained by these wireless systems were 24/7 monitored by a trained operator at distance. Also, heart rate, respiratory rate, temperature, blood pressure, level of consciousness, urine production and oxygen saturation were measured at bedside at least three times a day. The collected data were retrospectively analyzed after formal approval for this study had been given by the local medical ethical committee. All patients admitted to the COVID-19 ward of our hospital between April 1 st and May 15 th 2020 were considered for inclusion. Impaired cognition and/or psychiatric illness were defined as exclusion criteria. Monitoring was discontinued if patients had to be transferred to the intensive care for further treatment or in case palliative care was started. Patients were excluded from further analyses if subsequent PCR did not test positive for SARS-CoV-2. Heart rate and respiratory rate were recorded every second using the Everion sensor (Biovotion AG, Zürich, Switzerland). Saturation was recorded every fifteen minutes using the iHealth Labs Pulse Oximeter (iHealth Labs Inc. Mountain View, United States). Patients were allowed to ambulate while wearing the devices. However, most patients remained in bed due to physical impairment. Data from these devices were transmitted to a mobile device via Bluetooth. Wi-Fi facilitated data transmission from the mobile device to a secured multisensor platform (Digistat) developed by Ascom ( figure 1) . A link between the Digistat platform and the electronic medical https://mc.manuscriptcentral.com/jamiao record (EMR) and a mobile notification device allowed real-time insight in the patients vital signs. All data registered by the Everion and the Pulse Oximeter were eventually retrieved from the Ascom's Digistat platform (Ascom Holding AG, Baar, Switzerland) for further analysis. Other patient data were manually collected from the hospital's EMR. An automated Early Warning Score (aEWS) was calculated every 5 minutes using heart-and respiratory rate (table 1) . aEWS values of less than three were considered to be normal. To correct for outliers, the aEWS was calculated using the median values of both heart-and respiratory rate within a five minute window. Oxygen saturation values were not included in the aEWS because they were only measured every 15 minutes. The aEWS is based on the Early Warning Score used in our hospital's Rapid Response Team (RRT) activation protocol. Trained operators were situated in a control room, where they remotely monitored the vital signs of all patients and noticed elevated aEWS alarms. The protocol for alerting nursing staff was based on the existing COVID-19 literature at that time and the expertise of our pulmonologists. Nursing staff were alerted if a patient exhibited two or more elevated aEWS's in a time span of half an hour or if oxygen saturation dropped below 90% or dropped more than 4%. Nursing staff were then instructed to check on the patient and if deemed necessary perform extra measurements or alert the attending physician. All vital signs and aEWS scores were also accessible for the nursing staff itself via the hospital's EMR. The aEWS is calculated based on respiration rate and heartrate. Both paramaters have an associated score ranging from 0 to 3. The total score is the sum of both components. Patients were divided into two groups depending on whether adverse events had occurred. To provide a confidence interval to our results, we report a 95% confidence interval of the Mann-Whitney parameter as described in "Confidence Intervals of the Mann-Whitney Parameter that are compatible with the Wilcoxon-Mann-Whitney Test" [9] . The Mann-Whitney parameter is consistent with the Mann-Whitney U test. This parameter can be understood as the probability that a random individual from the adverse event group has a greater value (e.g. mean automated EWS) than a random individual from the other group. We compared multiple derived parameters between groups. For every patient mean automatic early warning scores (EWS), median automatic EWS, median SPO2, median heart rate (HR), median respiration rate (RR) and the average number of automatic EWS ≥ 3 per day were calculated. in COVID-19 care. We believe our main finding is the significance of the difference in vital signs measurements between patient with and without severe adverse events. The numerical difference were less relevant due to variability between sensors. COVID-19 patients transferred to intensive care or those who died showed changes in heart rate, respiratory rate and oxygen saturation prior to the occurrence of adverse events. Although differences between groups were found, the number of aEWS is quite substantial for both patients with or without severe adverse events. This may be the effect of the severity of COVID-19 causing more fluctuation of vital signs resulting in more aEWS alarms compared to other conditions. Evaluation of the sensitivity of the aEWS alerting therefore requires further evaluation in another patient population. To our best of knowledge, no data has been published about continuous monitoring of COVID-19 patients. Previous studies in other patient populations frequently tested only one device and did not have an automated alarm system incorporated. [3] [4] [5] We believe future medical care will be increasingly dependent on automatic measurements and it will be vital to have adequate systems in place to handle the data efficiently. ]. The implementation of this new system presented some challenges.. Although possible, most pulse oximeters are not intended to be used for continuous monitoring. Wireless pulse oximeter sensors showed to be prone to failure after long periods of use. With technical and practical solutions it is possible to minimize the downtime of these sensors. Other causes of missing data from both the saturation finger sensor and the arm sensor were attributed to low batteries and connectivity (WiFi and Bluetooth) problems. Due to the non-normality of our data, we used the Mann-Whitney U test to assess difference in mean ranks between groups. We used the techniques described in "Confidence Intervals of the Mann-Whitney Parameter that are compatible with the Wilcoxon-Mann-Whitney Test" [9] to provide confidence intervals to our results. The paper introduces the notion of the Mann-Whitney parameter. To illustrate how the Mann-Whitney parameter should be interpreted we shall take a closer look at the parameter estimate for mean automated EWS. The Mann-Whitney parameter estimate is 0.84 with a corresponding 95% confidence interval ranging from 0.63-0.94. This means that based on our data, 95 out of 100 times the true probability that a random patient from the adverse event group has a greater mean automated EWS than another random patient from the non-adverse event group, is contained within the range 0.63 to 0.94. Validation of a modified Early Warning Score in medical admissions The impact of the use of the Early Warning Score (EWS) on patient outcomes: A systematic review Do either early warning systems or emergency response teams improve hospital patient survival? A systematic review Are current wireless monitoring systems capable of detecting adverse events in high-risk surgical patients? A descriptive study Reliability of a wearable wireless patch for continuous remote monitoring of vital signs in patients recovering from major surgery: A clinical validation study from the TRaCINg trial Continuous Monitoring of Vital Signs Using Effectiveness of continuous or intermittent vital signs monitoring in preventing adverse events on general wards: A systematic review and meta-analysis The value of vital sign trends in predicting and monitoring clinical deterioration: A systematic review Confidence intervals of the Mann-Whitney parameter that are compatible with the Wilcoxon-Mann-Whitney test A call to alarms: Current state and future directions in the battle against alarm fatigue Alarm fatigue: impacts on patient safety The implementation of this new system took a combined effort of a large project group. First of all we thank Koen van Dulmen as project manager. We thank Ascom's and Slingeland's technical division for their continued support. We also thank all our control room operators and the members of the project team. We thank all medical professionals who worked with us for their efforts. The system we used in this study is a first iteration, it will be improved and evaluated for neurological and vascular patients in our hospital in a prospective study. Future research should focus on additional vital parameters, different sensors and advanced pattern recognitions. While trained operators were able to spot trends, we were unable to quantify these changes in this study. Furthermore, once a robust system is available, the focus of research should shift to improving clinical outcomes with the implementation of that system. In COVID-19 care this system has proven to be a valuable risk screening tool in addition to standard care.  This research received no specific grant from any funding agency in the public, 11 commercial or not-for-profit sectors  The data underlying this article cannot be shared publicly due to the privacy of 18 individuals involved in this study. The data will be shared on reasonable request to the All authors report no conflicts of interest related to the topic of the article.