key: cord-0881569-4f4wrwjl authors: Ho, Johannes; Stange, Claudia; Suhrborg, Rabea; Wurzbacher, Christian; Drewes, Jorg E.; Tiehm, Andreas title: SARS-CoV-2 wastewater surveillance in Germany: long-term RT-digital droplet PCR monitoring, suitability of primer/probe combinations and biomarker stability date: 2021-12-19 journal: Water Res DOI: 10.1016/j.watres.2021.117977 sha: 8e873d7e759bf9a188f4da9e8bd6e35a5146824f doc_id: 881569 cord_uid: 4f4wrwjl In recent months, wastewater-based epidemiology (WBE) has been shown to be an important tool for early detection of SARS-CoV-2 circulation in the population. In this study, a detection methodology for SARS-CoV-2 RNA (wildtype and variants of concern) in wastewater was developed based on the detection of different target genes (E and ORF1ab) by polyethylene glycol (PEG) precipitation and digital droplet PCR. This methodology was used to determine the SARS-CoV-2 concentration and the proportion of N501Y mutation in raw sewage of the wastewater treatment plant of the city of Karlsruhe in south-western Germany over a period of 1 year (June 2020 to July 2021). Comparison of SARS-CoV-2 concentrations with reported COVID-19 cases in the catchment area showed a significant correlation. As the clinical SARS-CoV-2 official case report chain takes time, viral RNA titre trends appeared more than 12 days earlier than clinical data, demonstrating the potential of wastewater-based epidemiology as an early warning system. Parallel PCR analysis using seven primer and probe systems revealed similar gene copy numbers with E, ORF, RdRP2 and NSP9 assays. RdPP1 and NSP3 generally resulted in lower copy numbers, and in particular for N1 there was low correlation with the other assays. The occurrence of the N501Y mutation in the wastewater of Karlsruhe was consistent with the occurrence of the alpha-variant (B.1.1.7) in the corresponding individual clinical tests. In batch experiments SARS-CoV-2 RNA was stable for several days under anaerobic conditions, but the copy numbers decreased rapidly in the presence of dissolved oxygen. Overall, this study shows that wastewater-based epidemiology is a sensitive and robust approach to detect trends in the spread of SARS-CoV-2 at an early stage, contributing to successful pandemic management. In the current pandemic situation, wastewater-based epidemiology (WBE) is considered an important tool to estimate SARS-CoV-2 prevalence, genetic diversity and geographical distribution (European Commission, 2021; Kitajima et al., 2020) . Sewage systems provide a feasible approach to survey faecal viruses across an entire region (Hart and Halden, 2020; Hill et al., 2021) , even if asymptomatic courses of infection occur (Qi et al., 2018) and at low frequency of clinical diagnostic testing (Lodder and de Roda Husman, 2020) . Therefore, wastewater surveillance can provide an alternative method to detect the spread of infections in different areas (Ahmed et al., 2020a; Lodder and de Roda Husman, 2020; Medema et al., 2020; Nemudryi et al., 2020; Wu et al., 2020) -especially for regions with limited diagnostic capacity and without a functioning reporting system, such as developing countries . In addition, wastewater surveillance can help detect variations in circulating strains, allowing comparisons among regions and an assessment of the evolution of the viral genome over time (Bisseux et al., 2018) . Moreover, wastewater-based surveillance can serve as an early warning system (Ahmed et al., 2020a; Chavarria-Miró et al., 2021; Hill et al., 2021; Kitajima et al., 2020; Kumar et al. 2020; La Rosa et al., 2020; Medema et al, 2020) . In light of the current pandemic, this approach could determine whether new SARS-CoV-2 infections have occurred in a community or whether the number of infected people decreases after measures in the affected population have been taken (e.g., lockdown or social distancing). Recent studies have shown the overall benefit of WBE for SARS-CoV-2 monitoring (Agrawal et al., 2021a; Medema et al., 2020; Rossmann et al., 2021b; Rossmann et al., 2021a; Westhaus et al., 2021) . For WBE purposes, SARS-CoV-2 is detected in wastewater using PCR-based methods. Multiple sets of PCR primers and probes have been published, targeting different locations of the genome such as the S, N and E gene Corman et al., 2020; US CDC, 2020) . For diagnostic testing evaluations of primer and probe sets used in SARS-CoV-2 RT-qPCR detection assays were already performed (Vogels et al., 2020) . While one of these clinical studies showed a high similarity in the analytical sensitivities for the different assays (Vogels et al., 2020) , other studies identified primer probe sets that showed higher sensitivity than others (Jung et al, 2020; Nalla et al., 2020) . However, there is a lack of data with respect to the reliability of primers/probes and genes at different positions of the genome, thus hampering the quantitative comparison of results, in particular for wastewater surveillance. Recently, predominantly RT-qPCR methods have been published. However, as an alternative to quantitative real-time PCR, digital droplet PCR (ddPCR) can be used for wastewater monitoring (Cervantes-Avilés et al., 2021) , which is a quantitative PCR method allowing an absolute quantification of DNA or RNA. By using ddPCR, a higher sensitivity can be achieved compared to qPCR since PCR inhibitors and the competition of background DNA and target molecules play a negligible role (Rački et al., 2014) . The genome of the SARS-CoV-2 virus is subject to a high mutation rate. Mutations are distributed across the genome and either have no effect on the phenotype (silent mutation) or can lead to altered infectivity and pathogenicity (Mercatelli and Giorgi, 2020 These SARS-CoV-2 variants were classified as variants of concern (VoC). All three VoCs have the mutation A23063T, also named N501Y, in common, which is involved in the receptorbinding mechanism and may have clinical impacts (Makowski et al., 2021) . For the single base mutation, a primer/probe set that allows the specific detection of the mutation has been described Korukluoglu et al., 2021) . The detection of mutations of SARS-CoV-2 in wastewater additionally reveals a more detailed view of the infection process and the local spread of virus strains. The SARS-CoV-2 RNA is excreted with the faeces of infected patients (Bogler et al., 2020) . This viral RNA reaches the wastewater treatment plant (WWTP) via the sewer system, where composite samples for wastewater monitoring are usually taken from the influent. However, relatively little is known about the stability of SARS-CoV-2 RNA in sewer systems (Hart and Halden, 2020) . In general, stability is strongly influenced by various environmental factors such as temperature, pH, biological activity or solid content (Foladori et al., 2020) . In laboratory experiments, significantly higher reductions were determined at higher temperatures (Ahmed et al., 2020b; Gundy et al., 2009) . Coronaviruses are enveloped viruses, and an intact envelope is important for virus pathogenicity (Mandala et al., 2020) . Overall, there are indications that enveloped viruses are less resistant to environmental conditions and inactivation compared to non-enveloped viruses such as adeno-or noroviruses (Gundy et al., 2009; Ye et al., 2016) . The virus envelope is highly sensitive to chemical and physical conditions (e.g., pH, lipid solvents, disinfectants) (Mohan et al., 2021; Scheller et al., 2020) . In contrast, the capsid protein is probably less subjected to lipid solvents, temperature and pH changes, although so far, little information is available on its stability. In this context, the aims of this study were (i) to establish a robust wastewater monitoring procedure including the evaluation of different target genes and the use of digital droplet PCR, (ii) to apply the WBE approach in southern Germany, (iii) to determine the distribution of SARS-CoV-2 variants and (iv) to investigate the stability of its genome in wastewater samples. This study was designed to contribute to the assessment of the sensitivity and robustness of SARS-CoV-2 wastewater surveillance, and therefore to pandemic and post-pandemic management. Material and Methods Long-term monitoring was performed between June 2020 and July 2021 at the WWTP of the city of Karlsruhe. Karlsruhe is the second-largest city of the German federal state Baden-Württemberg and located in south-western Germany. The WWTP of Karlsruhe has a capacity of 850,000 population equivalents and treats the sewage of about 370,000 people. We collected 24-h-composite samples (from 0 to 12 pm) twice a week using a flow-proportional auto sampler. The collected samples were homogenised using the dispersing and homogenizing tool MiniBatch D-9 (MICCRA GmbH) and transported to the laboratory at 4°C. Upon arrival, the samples were concentrated and analysed according to the method described below. In total, 89 samples were taken in the study area as duplicates. Wastewater samples from other sites in southern Germany (Leonberg, Baden-Wuerttemberg and Berchtesgadener Land, Bavaria) were also used for the evaluation of different target genes (802 samples in total, Table 1 ). For sample concentration, a volume of 45 mL of wastewater sample was centrifuged (30 min; 5,000 g) to settle larger particles. The supernatant was than concentrated using PEG/NaCl precipitation as described by Wu et al. (2020) , with slight modifications. Briefly, 10% (w/v) PEG 8000 (Carl Roth) and 2.25% NaCl (w/v) (Carl Roth) were added to the sample and mixed. The mixture was agitated on a shaking incubator at 120 rpm for 2 h on ice, and subsequently, samples were centrifuged at 20,000 g for 2 h at 4°C. The supernatant was discarded, and the pellet was resuspended in 600 μL PCR-grade water (ThermoFisher Scientific). For all samples two independent replicates were analysed. Recovery of viruses by the PEG precipitation was determined using phi6 bacteriophage as a surrogate for enveloped viruses. Using spiked samples, the phi6 recoveries were approx. 26% using PEG precipitation. In the first phase of the study, SARS-CoV-2-RNA in particle-bound and the dissolved portion were processed in parallel. For the solid phase, the pellet of 45 mL wastewater was resuspended in 600 µL of PCR-grade water and nucleic acid extraction was performed according to (section 2.2.2). However, since the measured SARS-CoV-2 concentration in the solid phase was at least 0.7 log levels lower than in the liquid phase, the solid phase was not taken into account for trend analysis. Nucleic acids were extracted directly after concentration using the innuPrep Virus DNA/RNA Kit -IPC16 (Analytik Jena GmbH) as described by the manufacturer (protocol for isolation from 600 µL cell-free body fluid). Extracted nucleic acids were eluted in a final volume of 100 µL and analysed immediately or stored at -80°C. In case of the four main assays (N1, E, ORF, RdRP1), all ddPCR reactions were performed within a week. Analysis using further primer and probes set were performed later with a maximum storage time of six month. Parallel to these analyses, the E and ORF assays were repeated. A comparison of the results for E and ORF analysed at the two different time points showed variations within the standard deviation of the ddPCR, and no degradation of the RNA over the storage period. The SARS-CoV-2-specific sequences were quantified using one-step reverse transcription (RT)-ddPCR. For the PCR analyses, seven primers/probe sets were compared (Table 1 ). All wastewater samples were analysed using two duplex PCR assays with the primers and probe sets 1-4. Additional investigations of selected samples for the evaluation of alternative primers and probes (set 5-7) and the detection of the N501Y mutation and wildtype (sets 8 and 9) were carried out. All primers and probes were obtained from ThermoFisher Scientific. The ddPCR was performed on the QX200 Droplet Digital PCR System (BioRad) using the One- QuantaSoft Analysis Pro (BioRad) were used to manually threshold and export the data. The number of accepted droplets was 13,000 on average and similar for all gene targets. Analyses with a droplet number below 8,000 were repeated with an additional tenfold dilution in accordance with the digital MIQE guidelines (dMIQE Group and Huggett, 2020). Due to this procedure, more than 8,000 droplets could always be generated in the repetition. Tests for possible inhibition in ddPCR were performed with ten-and hundredfold diluted samples in PCRgrade water showing no inhibitory effects (diluted samples had an average of 5 % deviation from the undiluted samples). Concentrations per reaction were converted to copies per mL of wastewater. All detections were performed in the same lab and with the same methods. Previous reports showed an actual detection limit of 3 copies per reaction for the ddPCR (Alteri et al., 2020) . In this case, the detection limit also represents the limit of quantification. Based on the initial volume of wastewater and the volume of RNA extract used in our study, a detection limit of 2.5 gene copies per mL of wastewater (without taking into account the losses during concentration and extraction) was calculated. Epidemiological data on COVID-19 cases in the studied area were retrieved from the daily report on COVID-19 infections in the city and districts of Karlsruhe Basic mathematical calculations (mean, median, sliding average, regression and p-tests) were performed in Microsoft Excel (14.0.7212.5). To compare trends for the results of different gene targets (e.g. N1 and E), a linear regression statistic was performed with and without regression models forced through zero. Batch experiments were established in 1-L glass bottles using untreated wastewater from the WWTP of Karlsruhe. The 24-hr composite sample of the previous day (4 L) was transported to the laboratory and directly used for the experiment. Overall, six test bottles containing 600 mL of wastewater were set up. Three bottles with wastewater were purged with ambient air using aquarium pumps, and the other three bottles were purged with nitrogen gas for ten minutes and then sealed immediately. The bottles were incubated at room temperature. Samples were taken to determine dissolved oxygen (DO) and pH using a WTW meter (36020 IDS) with optical sensor (FDO 925) and pH electrode (SenTrix) from Xylem Analytics. After each sampling the wastewater was purged again with ambient air or nitrogen for five minutes. The PCR samples (40 ml) were processed using the Vac-Man Laboratory Vacuum Manifold (Promega) and the Maxwell ® RSC Enviro Wastewater TNA Kit (Promega). The ddPCR data for a total of 802 field samples, including results for four primer and probe sets targeting N, E, ORF and RdRP genes (2019-nCoV-N1, E_Sarbeco, HKU-ORF1b-nsp14, RdRP_SARS; hereinafter referred to as N1, E, ORF and RdRP1), were compared (Fig. 1) . The coefficient of variation of independent replicates was 26%. Using N1, 94% of the samples gave a positive result. For other primer and probe combinations the proportion of positive detections were considerably lower (62% for E, 54% for ORF and 20% for RdRP1). A similar trend was observed for average gene copy numbers, with highest values for N1 and lowest one for RdRP1. Median over all values were 27.5, 11.8, 11.5 and 8.5 gene copies per mL for N1, E, ORF and RdRP1, respectively. In particular, for N1, we observed several results much higher than those measured in samples days before and after. These peaks did not appear in the E, ORF and RdRP assays. In some cases, the N1 values were more than 2 log 10 levels higher than the values obtained by the other assays. Linear regression between the data (Fig. 1) Gonzalez et al. (2020) and Westhaus et al. (2021) ; both studies excluded N1 for wastewater monitoring. Considering the observed effects in our study, initial comparison of detection methods and the parallel detection of at least two SARS-CoV-2-specific genes is recommended to assure reliable WBE. For the investigated WWTP, 89 samples were taken during the 1-year survey from June 2020 to Germany. For each sample, two independent replicates were analysed using E and ORF assays. An average SARS-CoV-2 RNA concentration was calculated from the results of the two replicates and the gene copy numbers for E and ORF. Based on the correlation and a methodological detection limit of 2.5 genomic copies per mL sewage by ddPCR, the theoretical limit of detection for the wastewater monitoring is approx. 20 infections per 100,000 inhabitants. The results of another German study indicated that RT-qPCR can detect 50 acute infected persons per 100,000 inhabitants in dry-weather periods (Westhaus et al., 2021) . In our study one genomic copy per mL is equivalent to 7 active cases per 100,000 inhabitants. However, the case/gene copy number factor is hard to compare with literature, as parameters often differ (e.g., cumulative cases (Medema et al., 2020) or gene copies per day (Agrawal et al., 2021a) ). Furthermore, these values are highly specific for a sewer system and depend on numerous boundary conditions, such as characteristics of the sewer system. Therefore, SARS-CoV-2 copy number/number of infections factors cannot be transferred directly to other catchment areas. The results for the primers/probes N1, E, ORF and RdRP1 exhibited the highest gene copy numbers for N1 and the lowest one for RdRP1. Based on these results, it could be hypothesised that the SARS-CoV-2 genome degradation might preferably start from the 5' end. In general, viral RNA can be degraded enzymatically via endonucleases, 3'-to-5' exonucleases and 5'-to-3' exonucleases (Houseley and Tollervey, 2009 ). However, viruses have developed specific mechanisms to counteract this degradation. These include various 5' end modifications, as well as processes of mimicking cellular RNAs via 'cap snatching' or even direct blocking of receptor proteins, thus preventing recognition (Dickson and Wilusz, 2011; Markiewicz et al., 2021 ) . The genome of coronaviruses has a 5′-terminal cap structure and a poly(A) sequence at the 3′-end . To our knowledge, information about the specific enzymatic degradation processes of SARS-CoV-2 RNA in wastewater is not available yet. To test the hypothesis of the preferential 5' end degradation, two additional genes located more closely to the 5' end, namely NSP3 and NSP9, and an alternative primer for the RdRP gene (hereinafter referred to as RdRP2) were investigated with a set of 64 samples (Fig. 4) . For this set, samples with high, medium and low copy numbers were selected. High N1 outliers were excluded. Generally, the observed median gene copy numbers differed between 17.5 (RdRP1) and 84.6 copies per mL (N1). The medians followed the order N1, RdRP2, NSP9, E, ORF, NSP3, RdRP1. The E, ORF, RdRP1, RdRP2, NSP3 and NSP9 primer/probe sets showed a good correlation with coefficients up to 1.00 (p < 0.05) (Fig. S3 in Supplementary Information) . All correlation coefficients with the N1 assay were generally lower and ranged from -0.24 (RdRP2) to 0.74 (ORF). Comparing the slopes of the linear regression, nearly all significant correlations, except those with RdRP1, ranged from 0.5x to 2.0x (n = 46-58, p < 0.05). Generally, RdRP1 provided considerably lower values than the other genes (0.23x-0.46x, n = 50-47, p < 0.05). Based on this analysis, we can conclude the following: (i) due to the high gene copy numbers (outliers) observed with N1 and the low gene copy numbers using RdRP1 and NSP3, these primer/probe sets are not recommended for wastewater surveillance, (ii) E, ORF, RdRP2 and NSP9 worked well for monitoring using ddPCR and provided comparable gene copy numbers (average of deviations: 14%) and (iii) the hypothesis of the 5'-directed degradation of the SARS-CoV-2 genome was rejected. The observed effects of deviation between the different PCR-assays might be explained with variations in PCR efficiency as well as general primer design or secondary structures of target regions in the viruses. Primers and probes did not show any mismatches in comparison with the corresponding target sequences (e.g., SARS-CoV-2 reference sequence NC_045512.2), except for the RdRP1 reverse primer with a single nucleotide mismatch in the middle. This mismatch could lead to lower copy numbers (Vogels et al., 2020) ; however, single mismatches not in close 3´proximity are not that critical (Lefever et al., 2013) . Overall, we strongly recommend the application of different primers and probes. Such an approach will assure reliable and robust results and prevent underestimation of results due to newly occurring primer binding site mutations. Additionally, multiple targets will ensure comparability with earlier results in the case a primer has to be exchanged. A good example for this is the N3 forward primer published by the CDC, with a mutated nucleotide position in about 4% of observed sequences (Vogels et al., 2020) . We also quantified VoC of sewage samples in the sewershed of the city of Karlsruhe. The proportion of N501Y to wildtype sequences was compared with publicly available data from the diagnostic tests of infected persons. The proportion of the N501Y mutation in wastewater follows the same course as the trend seen for infected persons (Fig. 5) . The earliest detection of N501Y was made in the wastewater sample of December 28, 2020. The proportion of N501Y gradually increased to 97% on March 11, 2021, and then settled at around 86%. In December 2020, the state of Baden-Württemberg reported the detection of the alpha variant (B.1.1.7) for the first time, corresponding to the first detection of N501Y in wastewater influent from Karlsruhe. After the first detections of variants containing the N501Y mutation, the increase followed a trend that is comparable with the increase of N501Y-containing variants in patients in Baden-Wuerttemberg. Analogous to a Dutch study , the results of this study show the applicability of RT-ddPCR to detect the spread of relevant mutations of SARS-CoV-2 in the population. In order to obtain more information on the stability of SARS-CoV-2 RNA in sewer systems, studies into the effect of oxygen were conducted. Depending on flow velocity, temperature, dissolved organic carbon, and other boundary conditions, the DO concentration in the wastewater of the sewer system can vary. For this reason, the decay of SARS-CoV-2 RNA was investigated under aerobic and anaerobic conditions. At the beginning of the experiment, DO contents of 2.8 mg/L and SARS-CoV-2 RNA concentrations between 95 and 112 gene copies per mL were determined (Fig. 6 ). In the bottles aerated with atmospheric oxygen, the DO increased up to 8.5 mg/L while the concentration of SARS-CoV-2 RNA decreased by 1 log 10 level within 24 h. After 4 days the copy numbers were below the detection limit. In the nitrogen treated bottle, the DO concentration decreased below 0.1 mg/L and viral RNA decreased by about 0.3 log 10 levels (52%) within one day. Subsequently The virus stability studies published so far have mostly used other human or animal pathogenic coronaviruses, showing an influence of temperature, pH and matrix on the decay of coronaviruses in several water types (Foladori et al., 2020) . At room temperature, coronaviruses suspended in tap water, primary WWTP effluent or untreated and autoclaved wastewater showed > 99.9% reduction within 2-5 days (Ahmed et al., 2020b; Gundy et al., 2009) . In contrast, animal coronaviruses such as transmissible gastroenteritis virus (TGEV) and mouse hepatitis virus (MHV) showed a higher persistence in pasteurized settled sewage at room temperature (Casanova et al., 2009) ; times for a 99% reduction were 7 days for MHV and 9 days for TGEV (Casanova et al., 2009) . However, oxygen concentration has not been measured in these studies. It has been known for a long time that the oxygen concentration can be an important factor influencing virus survival. Scheuerman et al. (1991) have studied the effects of temperature and DO on the persistence of enteric viruses in sludge (polio 1, coxsackie B3 and echo 1, and rotavirus SA-11) under laboratory conditions. The inactivation rates under aerobic conditions were significantly higher than those under anaerobic conditions (-0.77 log 10 /day vss -0.33 log 10 /day) (Scheuerman et al., 1991) . Also in groundwater studies, the presence of oxygen accelerated enteric viruses decay (Gordon and Toze, 2003; Sidhu and Toze, 2012) . Within this study, a robust method for the detection of SARS-CoV-2 RNA in wastewater was established. This detection methodology is based on PEG precipitation followed by automated nucleic acid extraction and the detection of different target genes by RT-ddPCR. The following findings were obtained: -Comparison of ddPCR results for seven different primer and probe assays showed that E, ORF, RdRP2 and NSP9 were suitable for wastewater monitoring and yielded similar results. The N1, RdRP1 and NSP3 assays clearly deviated from the average of the other primer and probe combinations. Based on these results and the high mutation potential of coronaviruses, we recommend the parallel detection of at least two, preferably three or more, target sequences to increase the robustness of the method. -Comparison of the wastewater monitoring results for the city of Karlsruhe with reported COVID-19-infection numbers confirmed the potential of WBE for the early detection of trends in the incidence of infection within a catchment area. -With PCR-based wastewater monitoring also the spread of VoCs in a catchment area can be monitored, making it a fast and cost-effective alternative to NGS-based approaches. -SARS-CoV-2 RNA in wastewater showed low decay under anaerobic conditions, and fast decay under aerobic conditions. These findings are relevant for a better understanding of biomarker stability in sewer systems, and storage of wastewater samples. The data obtained from the wastewater monitoring were made available to the Covid-19 Task Force of the city of Karlsruhe and used as an additional tool for pandemic management. The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper. 5 . Concentration of N501Y mutation divided by the total gene copy numbers of the S-gene (ratio of wildtype and N501Y mutation) in wastewater, and ratios of variants of concern (VoC) in new COVID-19 Fig. 6 . Decay of SARS-CoV-2 biomarkers (as means of ORF, E, RdRP2 and NSP3 copy numbers) and oxygen concentration over time in wastewater samples purged with air (left) or nitrogen (right). Errror bars represent standard deviations for the three independent test bottles (n = 3). Long-term monitoring of SARS-CoV-2 RNA in wastewater of the Frankfurt metropolitan area in Southern Germany Metatranscriptomic analysis reveals SARS-CoV-2 mutations in wastewater of the Frankfurt metropolitan area in southern First confirmed detection of SARS-CoV-2 in untreated wastewater in Australia: a proof of concept for the wastewater surveillance of COVID-19 in the community Decay of SARS-CoV-2 and surrogate murine hepatitis virus RNA in untreated wastewater to inform application in wastewater-based epidemiology Detection and quantification of SARS-CoV-2 by droplet digital PCR in real-time PCR negative nasopharyngeal swabs from suspected COVID-19 patients Epidemiological surveillance of SARS-CoV-2 by genome quantification in wastewater applied to a city in the northeast of France: Comparison of ultrafiltration-and protein precipitation-based methods Monitoring human enteric viruses in wastewater and relevance to infections encountered in the clinical setting: a one-year experiment in central France Rethinking wastewater risks and monitoring in light of the COVID-19 pandemic Survival of surrogate coronaviruses in water Approaches applied to detect SARS-CoV-2 in wastewater and perspectives post-COVID-19 Sentinel surveillance of SARS-CoV-2 in wastewater anticipates the occurrence of COVID-19 cases Molecular diagnosis of a novel coronavirus (2019-nCoV) causing an outbreak of pneumonia Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR Genome sequencing of sewage detects regionally prevalent SARS-CoV-2 variants Quantitative analysis of SARS-CoV-2 RNA from wastewater solids in communities with low COVID-19 incidence and prevalence The digital MIQE guidelines update: minimum information for publication of quantitative digital PCR experiments for 2020 Commission Recommendation of 17.3.2021 on a common approach to establish a systematic surveillance of SARS-CoV-2 and its variants in wastewaters in the EU Wastewater monitoring outperforms case numbers as a tool to track COVID-19 incidence dynamics when test positivity rates are high SARS-CoV-2 from faeces to wastewater treatment: what do we know?: A review Early-pandemic wastewater surveillance of SARS-CoV-2 in Southern Nevada: Methodology, occurrence, and incidence/prevalence considerations COVID-19 surveillance in Southeastern Virginia using wastewater-based epidemiology Influence of groundwater characteristics on the survival of enteric viruses Survival of coronaviruses in water and wastewater Wastewater surveillance for SARS-CoV-2: Lessons learnt from recent studies to define future applications Computational analysis of SARS-CoV-2/COVID-19 surveillance by wastewater-based epidemiology locally and globally: Feasibility, economy, opportunities and challenges Droplet digital RT-PCR to detect SARS-CoV-2 variants of concern in wastewater SARS-CoV-2 known and unknowns, implications for the water sector and wastewater-based epidemiology to support national responses worldwide: Early review of global experiences with the COVID-19 pandemic Monitoring SARS-CoV-2 in municipal wastewater to evaluate the success of lockdown measures for controlling COVID-19 in the UK The many pathways of RNA degradation Real-time RT-PCR assays for the detection of SARS-CoV-2 Monitoring SARS-CoV-2 circulation and diversity through community wastewater Sequencing, the Netherlands and Belgium Comparative analysis of primer-probe sets for RT-qPCR of COVID-19 causative virus (SARS-CoV-2) The architecture of SARS-CoV-2 transcriptome SARS-CoV-2 in wastewater: State of the knowledge and research needs 40 minutes RT-qPCR assay for screening spike N501Y and HV69-70del mutations First proof of the capability of wastewater surveillance for COVID-19 in India through detection of genetic material of SARS-CoV-2 First detection of SARS-CoV-2 in untreated wastewaters in Italy Single-nucleotide polymorphisms and other mismatches reduce performance of quantitative PCR assays SARS-CoV-2 in wastewater: Potential health risk, but also data source Genomic characterisation and epidemiology of 2019 novel coronavirus: Implications for virus origins and receptor binding Biological and clinical consequences of integrin binding via a rogue RGD motif in the SARS CoV-2 spike protein Structure and drug binding of the SARS-CoV-2 envelope protein transmembrane domain in lipid bilayers Tricks and threats of RNA virusestowards understanding the fate of viral RNA Presence of SARScoronavirus-2 RNA in sewage and correlation with reported COVID-19 prevalence in the early stage of the epidemic in the Netherlands Geographic and genomic distribution of SARS-CoV-2 mutations SARS-CoV-2 in environmental perspective: Occurrence, persistence, surveillance, inactivation and challenges Occurrence, persistence, surveillance, inactivation and challenges Multicentre comparison of quantitative PCR-based assays to detect SARS-CoV-2 Comparative performance of SARS-CoV-2 detection assays using seven different primer-probe sets and one assay kit Temporal detection and phylogenetic assessment of SARS-CoV-2 in municipal wastewater Measurement of SARS-CoV-2 RNA in wastewater tracks community infection dynamics Comparing analytical methods to detect SARS-CoV-2 in wastewater Global prevalence of asymptomatic norovirus infection: A meta-analysis One-step RT-droplet digital PCR: A breakthrough in the quantification of waterborne RNA viruses SARS-CoV-2 RNA in wastewater anticipated COVID-19 occurrence in a low prevalence area Ahead of the second wave: Early warning for COVID-19 by wastewater surveillance in Hungary SARS-CoV-2 Krisenmanagement mit Hilfe eines Abwasserfrühwarnsystems am Beispiel Berchtesgadener Land SARS-CoV-2. Crises Management using Wastewater Surveillance in the County Berchtesgadener Land Innovatives SARS-CoV-2-Krisenmanagement im öffentlichen Gesundheitswesen: Corona-Dashboard und Abwasserfrühwarnsystem am Beispiel Berchtesgadener Land. Innovative SARS-CoV-2-2 crises management in the pubic health sector corona dashboard and wastewater surveillance in the county Berchtesgadener Land A novel single nucleotide polymorphism assay for the detection of N501Y SARS-CoV-2 variants Physicochemical properties of SARS-CoV-2 for drug targeting, virus inactivation and attenuation, vaccine formulation and quality control Laboratory studies of virus survival during aerobic and anaerobic digestion of sewage sludge Assessment of pathogen survival potential during managed aquifer recharge with diffusion chambers US CDC (US Center for Disease Control and Prevention), 2020. 2019-novel coronavirus (2019-nCoV) real-time rRT-PCR panel primers and probes Analytical sensitivity and efficiency comparisons of SARS-CoV-2 RT-qPCR primer-probe sets Detection of SARS-CoV-2 in raw and treated wastewater in Germany: Suitability for COVID-19 surveillance and potential transmission risks SARS-CoV-2 titers in wastewater are higher than expected from clinically confirmed cases Evaluation of lockdown impact on SARS-CoV-2 dynamics through viral genome quantification in Paris wastewaters Survivability, partitioning, and recovery of enveloped viruses in untreated municipal wastewater This study was financially supported by the German Federal Ministry of Education and Research as part of the funding program Sustainable Water Management (NaWaM-RiSKWa) (Biomarker, grant number 02WRS1557). We thank the City Civil Engineering Office of the city of Karlsruhe (Tiefbauamt der Stadt Karlsruhe), especially Martin Kissel, Albrecht Dörr, Dr. Gaby Morlock und Stephen Kemper. We also thank Dr. Marion Woermann und Carmen Kraffert for their assistance with wastewater monitoring and experimental work.