key: cord-0879428-l78n7g8t authors: Parker, Helen E.; Sengupta, Sanghamitra; Harish, Achar V.; Soares, Ruben G.; Joensson, Haakan N.; Margulis, Walter; Russom, Aman; Laurell, Fredrik title: A Lab-in-a-Fiber optofluidic device using droplet microfluidics and laser-induced fluorescence for virus detection date: 2022-03-03 journal: Sci Rep DOI: 10.1038/s41598-022-07306-0 sha: ff2937fe5fb39b6a24d5778c3686ee631aa1af17 doc_id: 879428 cord_uid: l78n7g8t Microfluidics has emerged rapidly over the past 20 years and has been investigated for a variety of applications from life sciences to environmental monitoring. Although continuous-flow microfluidics is ubiquitous, segmented-flow or droplet microfluidics offers several attractive features. Droplets can be independently manipulated and analyzed with very high throughput. Typically, microfluidics is carried out within planar networks of microchannels, namely, microfluidic chips. We propose that fibers offer an interesting alternative format with key advantages for enhanced optical coupling. Herein, we demonstrate the generation of monodisperse droplets within a uniaxial optofluidic Lab-in-a-Fiber scheme. We combine droplet microfluidics with laser-induced fluorescence (LIF) detection achieved through the development of an optical side-coupling fiber, which we term a periscope fiber. This arrangement provides stable and compact alignment. Laser-induced fluorescence offers high sensitivity and low detection limits with a rapid response time making it an attractive detection method for in situ real-time measurements. We use the well-established fluorophore, fluorescein, to characterize the Lab-in-a-Fiber device and determine the generation of [Formula: see text] 0.9 nL droplets. We present characterization data of a range of fluorescein concentrations, establishing a limit of detection (LOD) of 10 nM fluorescein. Finally, we show that the device operates within a realistic and relevant fluorescence regime by detecting reverse-transcription loop-mediated isothermal amplification (RT-LAMP) products in the context of COVID-19 diagnostics. The device represents a step towards the development of a point-of-care droplet digital RT-LAMP platform. www.nature.com/scientificreports/ resolution temporal and spatial analysis of the sampled material in situ, in vivo, as demonstrated previously for chemical monitoring [9] [10] [11] . On the other hand, partitioning a single dilute sample into many droplets, each containing either a single target molecule or lacking target molecules enables digital detection of an amplified molecular species, significantly reducing the demands on optical detectors in e.g. droplet digital PCR 12 and digital droplet LAMP 13 . The detection and quantification of droplet-encapsulated material may involve the use of optics integrated into the chip. So called optofluidic devices offer a way to synergistically combine key microfluidics sample processing capabilities with optical functionality, opening up promising avenues for bio-and chemical-sensing 14 . Established optical techniques such as fluorescence-activated droplet sorting 15, 16 , optical microscopy 17, 18 , or laser-induced fluorescence (LIF) [19] [20] [21] , can be incorporated into microfluidic chips. Integrating microfluidic capacity directly into optical fibers and capillaries offers an alternative emerging route to benchtop optofluidics. Optical fibers possess a set of attributes which include flexibility, chemical and bio-inertness, easy light coupling, and a high aspect ratio. Moreover, their widespread use across numerous fields, from telecommunications to sensing 22, 23 and imaging in medicine [24] [25] [26] [27] have brought about high tolerance, and standardized production in kilometer lengths which can be assembled into low-cost optofluidic devices with precise channel dimensions using equipment developed for telecommunications production. In previous work, we demonstrated several all-fiber optofluidic devices for biomedical applications fabricated from an array of optical fibers and capillaries. We demonstrated devices for the detection, trapping, and collection of fluorescent microparticles 28, 29 and demonstrated the potential of these techniques by developing a low-cost all-fiber flow cytometer 30 . Most recently, we have demonstrated particle separation and counting in an uninterrupted flow through a fiber device 31 . Furthermore, structured hollow core fibers allow for the guiding of light directly in the fluid media, which has been used for Raman sensing 32 and for the tracking of diffusion of nanoscale objects 33, 34 including virus particles 35 . Structured optical waveguides, such as light cages, have been developed for absorption spectroscopy with possible applications in bioanalytics 36 . These seamlessly integrated Lab-in-a-Fiber optofluidic devices indicate potential for extending the use of Lab-in-a-Fiber technology to molecular diagnostics. Starting in late 2019, the SARS-CoV-2, or COVID19, pandemic represented a major global health challenge and continues to have far reaching societal and economic consequences. The pandemic has highlighted the need for globally deployable, fast and scalable biomolecular testing 37 for high as well as low-income countries. The tests developed for ongoing SARS-CoV-2 infection include rapid antigen tests, which detect the presence of key viral proteins in nasopharyngeal swab specimens, and nucleic acid tests which amplify and detect viral genetic material in nasal, throat or mouth swabs. Of these, nucleic acid amplification tests are expected to provide higher sensitivity and potentially higher specificity than rapid antigen tests 38 . In this work, motivated by fortifying and broadening the range of test formats for sample-to-answer medical diagnostics, we have developed a droplet microfluidic Lab-in-a-Fiber device to carry out two key bioassay functions for a rapid nucleic acid amplification test based on digital LAMP. These two steps are: droplet encapsulation of sample material, and in situ detection of optical signatures through LIF. Features of our Lab-in-a-Fiber device are: droplets are generated in a negative pressure regime (a requisite for in situ sampling); LIF excitation and detection is carried out on board within a novel compact fiber arrangement; and operation of the device is carried out via hardware connected at only one end, simplifying the experimental layout and freeing up the length of the Lab-in-a-Fiber device. We describe the fabrication of this Lab-in-a-Fiber device and its characterization using a dilution series of fluorescein solutions from 100 µ M to 10 nM. As a proof-of-concept demonstration, we present results from a pre-amplified RT-LAMP-based nucleic acid amplification SARS-CoV-2 assay. Thus, we demonstrate that the integrated Lab-in-a-Fiber device operates within a diagnostically relevant fluorescence detection regime for such nucleic acid amplification tests. This section describes: the fabrication of the Lab-in-a-Fiber device; the optical detection setup; and the experimental methods used to characterize the complete setup using fluorescein and a pre-amplified RT-LAMP assay. Fabrication of the Lab-in-a-Fiber device. The Lab-in-a-Fiber device was made from a catalogue of fibers and capillaries drawn in-house and assembled using a Vytran Automated Glass Processor Workstation (GPX3000, Thorlabs). The device schematic is shown in Fig. 1 and comprises two modules with the following roles: droplet generation via flow-focusing; and optical detection. A detailed description of the fabrication process of each module is given below. The Lab-in-a-Fiber device operates with proximal-end light coupling and application of negative pressure via a syringe pump, allowing for distal-end sample loading. Droplet generation module. A compound capillary was made by first aligning a 90 µm/125 µ m inner diameter/ outer diameter (ID/OD) capillary to the central hole of a five-hole capillary (Fig. 2a) . Each of the holes of the five-hole capillary had an ID of 30 µ m and were arranged like the dots on the five-side of six-faced dice (Fig. 2a (inset) ). The 90 µm/125 µ m ID/OD capillary was then fusion spliced to the central hole (Fig. 2b) . The splice process slightly collapsed the 90 µm/125 µ m ID/OD capillary so that the four outer holes were not blocked but the 90 µm/125 µ m ID/OD capillary remained open. The five-hole capillary was then cleaved to a length of 200 µ m from the splice position (Fig. 2c ). This compound capillary section formed the droplet generation element and was inserted from the left side into a 5 cm long section of a 255 µm/333 µ m ID/OD housing capillary (Fig. 2d) . A 5 cm long section of 125 µm/255 µ m ID/OD capillary was inserted from the right side of the housing capillary and its end face was brought to a distance of 100 µ m from the cleaved end of the five-hole capillary (Fig. 2e ). This distance influences the produced droplet size 39, 40 and was determined empirically. Following this, a second 90 µ m/125 µ m ID/OD capillary was inserted a short way into the left hand entrance of the housing capillary to provide oil supply for the outer four holes of the five-hole capillary (Fig. 2f) Fig. 1a . We used basic hydrodynamic resistance calculations to select suitable lengths of each inlet such that the droplets were generated with sufficient separation from one another. This was 1:3.5 (continuous phase inlet length:dispersed phase inlet length)), which, in our device, was 7 cm and 25 cm, respectively. The entire arrangement resulted in monodisperse sets of droplets of 120 µ m in diameter (Fig. 1b) . A 40 cm long piece of 90 µ m/125 µ m ID/OD capillary was inserted and secured with UV-cured glue in the open end of the 125 µm/255 µ m ID/OD capillary. The droplet detection module was then built onto this length of capillary. As a final step in the fabrication of the droplet generation module, a siliconizing reagent, Sigmacote ® (SL2, Sigma-Aldrich), was flushed steadily through the device for 45 min. This ensured that the interior walls of the flow channels were appropriately hydrophobic and fluorosilicic to prevent unwanted interaction between them and the disperse aqueous liquid. Droplet detection module. Optical interrogation was performed within the Lab-in-a-Fiber device itself using a periscope fiber as illustrated in Fig. 1c . A double-clad fiber (DCF), drawn in-house, consisting of an 8 µ m core (numerical aperture (NA) of 0.14), a light-guiding 50 µ m 1st cladding region (NA of 0.25), and a 125 µ m 2nd cladding region was flat-cleaved ( Fig. 3a (i)) and the tip was masked using vacuum-safe paint ( Fig. 3a(ii) ). The masked tip was then cut at an angle of 45 • normal to the fiber propagation axis using a dicing saw (DAD-2H/6M, Disco) ( Fig. 3a(iii) ). The DCF was then loaded into a thermal evaporator (Auto 306/FL400, Edwards) and the tip was coated with aluminum until a 70 nm layer of aluminum built up on the angled cut surface (Fig. 3a (iv)). The DCF was then immersed in acetone and sonicated for 20 min to remove the vacuum-safe paint, leaving behind a mirrored angled tip ( Fig. 3a(v) ). The periscope fiber was then secured within a housing capillary in parallel Even without aluminium deposition on the angled cut surface of the periscope fiber, sideways propagation of light from the tip was observed. However, the result was two local regions of light: a primary illumination region at 90 • to the propagation axis, and a secondary illumination region at an oblique angle which extended ahead of the fiber tip. This resulted in droplet peaks consisting of three sub-peaks as the droplets moved through the two illumination regions, shown in Fig. 3d (red line), representing: the leading end of the droplet coincident with the primary illumination region (Fig. 3d(i) ); the body of the droplet coincident with both the primary and the secondary illumination region (Fig. 3d(ii) ); and the trailing end of the droplet coincident with the secondary illumination region (Fig. 3d(iii) ). Aluminum deposition ensured that light was better reflected to the primary region and this resulted in peaks closer to a top hat shape, see Fig. 3d (blue line). Optical setup. The optical setup was built from off-the-shelf components and is shown in Fig. 4 . Excitation light from a 20 mW 488 nm single-mode CW fiber laser (LP488-SF20, Thorlabs) was collimated via an achromatic fiber port ( L 1 ) (PAF2-A4A, Thorlabs). The light was passed through an excitation filter (Ex. filter) (MDF-GFP2 Excitation, Thorlabs) and reflected from a dichroic mirror (DM) (MDF-GFP2 Dichroic, Thorlabs). A second achromatic fiber port ( L 2 ) (PAF2-A4A, Thorlabs) coupled the light into the central core of the periscope fiber. Since the mode field diameter of the fibered laser was 3 µ m we slightly defocused ( L 2 ) in order to not underfill the 8 µ m core of the periscope fiber while also taking care not to guide excitation light through the 1st cladding region. This was achieved by pre-aligning a dummy DCF fiber in the system and checking central core coupling using a camera at the distal end. Once good pre-alignment was achieved, the dummy DCF was replaced by the device. Avoiding cladding coupling minimized high order mode illumination of the droplets which would have resulted in an unstable incident intensity profile across the droplet cross-section. Microscope images showed a localized laser spot incident on the droplet itself. Thus, as droplets passed, their crosssections were scanned along the capillary axis. Any misalignment of the laser light into the 1st cladding region could be seen by a broadening of the laser spot incident on the droplet, coupled with an unstable fluorescence measurement, so could easily be identified and realigned if needed. Fluorescence induced in the droplets was coupled back through the periscope fiber via the core and the 1st cladding region, which ensured maximized collection efficiency. The fluorescence was passed through the dichroic mirror (DM) (MDF-GFP2 Dichroic, Thorlabs) and filtered using an edge pass filter (Em. filter) (FELH0500, Thorlabs). An aspheric condenser lens ( L 3 ) (ACL2520U-A, Thorlabs) was used to loosely focus the collected fluorescence onto a silicon photomultiplier (SiPM) (PM3325-WB, Ketek). The signal was first passed through a pre-amplifier (SR560, Stanford Research Optics and hardware for operating the Lab-in-a-Fiber device. Excitation from a 488 nm fibercoupled laser was collimated using a fiber port ( L 1 ). The light passed through an excitation filter (Ex. filter) and was reflected from a dichroic mirror (DM) to a second fiber port ( L 2 ) which was used to couple the light into the Lab-in-a-Fiber device. Collected fluorescence passed through the dichroic mirror (DM) and an emission filter (Em. filter). The normalized transmission spectra for the filter set are shown on the right. An achromatic condenser lens ( L 3 ) loosely focused the light onto a silicon photomultiplier (SiPM). The SiPM signal was coupled to an oscilloscope via a low noise pre-amplifier which was used as a 1 kHz low pass filter. www.nature.com/scientificreports/ Systems), which we used as a 1 kHz low pass filter, and onto an oscilloscope (MSO54 5-BW-1000 Mixed Signal Oscilloscope, Tektronix) for readout. Experimental procedure. The full experimental procedure consisted of three steps: a continuous phase flushing step, a sample loading step, and a second continuous phase flushing step. For the first flushing step, both the dispersed phase inlet and the continuous phase inlet at the distal end of the Lab-in-a-Fiber device (left end of Fig. 1 ) were submerged in separate Eppendorf tubes each containing oil. At the proximal end (right end of Fig. 1 ) the droplet flow capillary was secured to a 2.5 mL syringe and loaded onto a syringe pump (AL-1000, World Precision Instruments). The pump was then set to a 12 µL min −1 withdrawal rate and was kept constant throughout the experiment. This rate was determined empirically to produce monodisperse droplets (of ∼ 0.9 nL from 1 µ M fluorescein solution) at a generation rate of 41 ± 0.5 Hz which was sustained for the entire experimental duration. As a first step, the Lab-in-a-Fiber device was allowed to run for twenty minutes before sample loading, aspirating fluorinated oil through both inlets. For the sample loading step, the dispersed phase inlet was quickly lifted out of its oil Eppendorf tube and lowered into an Eppendorf tube containing the sample: in our case either fluorescein (for characterization experiments) or pre-amplified RT-LAMP SARS-CoV-2. The continuous phase inlet remained unmoved. The total sample volume was withdrawn before the final flushing step. For the final flushing step, the dispersed phase inlet was quickly lifted out of the sample Eppendorf tube and returned to its original oil Eppendorf tube and allowed to flush for twenty minutes. In the case of pre-amplified RT-LAMP SARS-CoV-2 experiments, we carried out the experiment both at room temperature and 65 • C. The latter was achieved by securing the Lab-in-a-Fiber device to a homemade hot plate with a thermocouple and PID controller. This thermocouple provided live feedback of the temperature and the PID controller actively maintained 65 • C. We measured the hot plate to have a stability of 65 • C ± 1 • C throughout the experiment. We did not determine a minimal time for fluorinated oil flushing before or after sample loading. Data analysis. Analysis of droplet data was performed offline using relevant Python packages according to the following method: background fluoresence (which was determined to originate primarily from the periscope fiber itself) was captured during each experiment and was subtracted from the raw data. Then a Savitzky-Golay filter was applied to the raw data and local maxima were found using a find peaks function 41 . The mean and standard deviation of the peak maxima 42 Fluorescein. Fluorescein dye (Sigma-Aldrich) was used to prepare solutions of different concentrations. 0.66 mg of fluorescein dye was measured using a weighing balance (ABS-120-4N, KERN) and was dissolved in 20 mL of PBS buffer solution (pH 7.2) (Alfa Aesar) to prepare a 100 µ M stock solution. The solution was homogenized by ultrasonication for one minute using an ultrasonication bath (Selecta Ultrasonic). Using the serial dilution method six lower concentrations of fluorescein solutions (10 µ M, 1 µ M, 500 nM, 100 nM, 10 nM, 1 nM) were prepared from the 100 µ M stock solution. Each solution was homogenized using ultrasonication for one minute before dilution. Amplification of RNA was performed similarly to that described in Soares et al. 43 . The RT-LAMP was performed in an Eppendorf nexus gradient Mastercycler with a total of 100 µ L per reaction mixture. The reaction mixture containing 10 µ L template RNA (103 copies µL −1 ) or DI-water was prepared with final concentrations of 1x Isothermal amplification buffer II (New England Biolabs), 6 mM dNTPs (RP65, Blirt), 6 mM MgSO 4 (New England Biolabs), 1x primer mixture (six primers described in detail below), 1x EvaGreen ® (Biotium, 20x solution in water), 3. Characterization using fluorescein. In order to characterize the detection of droplets in the Lab-in-a-Fiber device, we selected fluorescein as an appropriate fluorophore surrogate for EvaGreen ® . From our experience of the fluorescence of RT-LAMP products and fluorescein, whereby negative SARS-CoV-2 RNA RT-LAMP, positive SARS-CoV-2 RNA RT-LAMP, and 500 nM fluorescein displayed RFUs of 6600, 8700, and 7800, respectively, we determined that the most appropriate set of fluorescein dilutions for our characterization should be: 100 µ M, 10 µ M, 1 µ M, 500 nM, 100 nM, 10 nM, 1 nM. These dilutions were run through the Lab-in-a-Fiber device and data were collected for sets of approximately 20,000 droplets per dilution. The amplitude and standard deviation were calculated for each of the droplet sets for each concentration, see Fig. 5a . From our characterization, we determined that it was possible to distinguish between each droplet set and we established a limit of detection (LOD) of 10 nM fluorescein (Fig. 5b) . Our LOD was defined according to 3.29 × the standard deviation of the background signal, in our case fluorinated oil. It has been reported that droplet velocity can result in significant variation in fluorescence intensity of fluorescein droplets, particularly at slow velocities ( ≤ 2.5 cm s −1 ) 45 . Therefore, as a post-acquisition quality check, peak widths were calculated throughout all experiments, see Fig. 5c . Peak widths provide a representation of droplet size and/or speed. These measurements ensured we could attribute observed differences in droplet fluorescence intensity to fluorophore concentration and not changes in droplet velocity in this slow droplet regime. that our Lab-in-a-Fiber device was operational within a relevant biological regime, pre-amplified positive and negative SARS-CoV-2 RNA RT-LAMP samples were tested. A clear difference was observed in the average fluorescence value of the negative (9.7 mV ± 0.2) and positive (14.6 mV ± 0.7) RT-LAMP samples at room temperature, see Fig. 5d (left), with a signal-to-noise ratio (SNR), here defined as the ratio of the fluorescence value of positive droplets to that of negative droplets, of 1.5. When the detection of the same sample was done at 65 • C, an increased difference between the average amplitude of negative (4.0 mV ± 0.2) and positive (14.5 mV ± 0.5) RT-LAMP samples was observed, see Fig. 5d (right), with an SNR of 3.6. At room temperature, primer dimer generation is more significant due to lower hybridization stringency, so the observed increase in population resolution is due to a decreased background fluorescence in the negative sample at 65 • C. The aim of this work was to demonstrate that well-established microfluidics techniques could be translated into an all-fiber optofluidic arrangement. Hence, we focused on developing two fundamental capabilities that could apply to a plethora of potential use cases: (i) monodisperse droplet generation and (ii) sensitive LIF detection. These capabilities were further developed with certain criteria in mind: (i) that the design remains uniaxial and compact, and (ii) that the device operates under negative pressure. We achieved droplet generation using a flowfocusing approach and detection using a periscope fiber without compromising on device footprint. The diameter of the Lab-in-a-Fiber device was no greater than 333 µ m at any point along its length (at the generation and detection modules), with the rest of the device being 125 µ m in diameter. The sample inlet capillary was 25 cm in length in 125 µ m in diameter, which makes it flexible and easy to maneuver by hand for simple sample loading. Our device exhibits the conventional benefits of optical fibers and capillaries, namely, biological and chemical inertness, ready integration with conventional optics, and the potential to be developed towards in vivo use cases. With the capabilities of fibers well demonstrated across the fields of telecommunications and biomedicine, it is exciting to consider the impact that Lab-in-a-Fiber technology could have in the field of optofluidics, which has so far been mostly explored within Lab-on-a-Chip architectures. Although Lab-in-a-Fiber research is in its nascent stage, one could imagine multiplexed functional fiber devices that incorporate, for example, microfluidics, sensing, and imaging to address a range of diagnostic challenges. In order to fully realize the potential of Lab-in-a-Fiber optofluidics, certain challenges, which we discuss below, will need to be addressed. Thermal drawing techniques have led to the mass manufacturability of fibers and capillaries. This has already been exploited for biomedical device development. For example, imaging fibers can be drawn using relatively inexpensive materials 46, 47 , enabling fiber-based endoscopy. It may be possible to address the bottleneck in our device fabrication process, namely, manual assembly of the constituent fibers and capillaries, with more sophisticated manufacturing techniques. For example, photolithography and etching could be used to make connective parts to facilitate the assembly process, ensuring tighter device-to-device repeatability while keeping down cost. We determined a LOD of 10 nM fluorescein of our Lab-in-a-Fiber device. The main limiting factor of our LOD was the collection efficiency of our periscope fiber. The NA of the 1 st cladding region of the DCF was 0.25 and this resulted in a collection efficiency of < 5 % of the total fluorescence emission from the droplet. Greater collection efficiency could be achieved through the use of a DCF with a larger light-guiding cladding region, or with higher NA. Capillary flow channels with a high aspect ratio, i.e., long and narrow, enable simple and easy sample loading but introduce a significant resistance which limits throughput. As described in the introduction, a key possibility of microfluidics is high-throughput analysis, with droplet generation possible at rapid rates (> kHz). In the results presented, our Lab-in-a-Fiber device produced droplets at a rate of 41 ± 0.5 Hz. Increasing the droplet rate could be achieved by shortening the device and increasing the inner diameter of the capillaries while still maintaining ease-of-use. Although outside the scope of this work, it is also relevant to discuss how onboard amplification of the nucleic acid material could be achieved. The amplification would need to occur after droplet generation, either through a continuous flow isothermal temperature zone at 65 • C or stationary in a large diameter capillary that provides sufficient volume to house and hold the sample plus continuous phase, in our case 10 µ L and 30 µ L, respectively. This capillary could be readily incorporated into our pre-existing design and then be heated from an external thermal source. Due to the wider research interests of our group, we chose RT-LAMP 43 as an appropriate first demonstration of LIF detection. In this work we describe the development of a novel Lab-in-a-Fiber device and its potential application for droplet microfluidics. The fiber device is constructed from a variety of capillaries and fibers which are widely available, and we demonstrate that the device is functional for both the generation of monodisperse droplets ( ∼ 0.9 nL) and their optical detection via laser-induced fluorescence. The Lab-in-a-Fiber device generates droplets using a uniaxial capillary flow-focusing arrangement through the application of negative pressure. Optical interrogation is performed on board through a double-clad fiber with an angled mirrored tip, which operates like a periscope to allow light coupling perpendicular to the fiber axis, keeping the device compact. The lowest detectable concentration that was tested was 10 nM fluorescein which is on par with many traditional Lab-on-a-Chip microfluidic platforms. Beyond our current work, we believe that further development could involve incorporating a modular temperature control unit and incubation chamber for onboard nucleic acid amplification. A review of state-of-the-art microfluidic technologies for environmental applications: Detection and remediation. Global Chall. 3, 1800060 Microfluidics in drug discovery Microfluidic cell culture systems for drug research The origins and the future of microfluidics Smartphone-based rapid quantification of viable bacteria by single-cell microdroplet turbidity imaging Living droplets Application of droplet digital PCR to detect the pathogens of infectious diseases Recent advances in droplet-based microfluidic technologies for biochemistry and molecular biology In vivo chemical monitoring at high spatiotemporal resolution using microfabricated sampling probes and droplet-based microfluidics coupled to mass spectrometry Push-pull perfusion sampling with segmented flow for high temporal and spatial resolution in vivo chemical monitoring Improved temporal resolution for in vivo microdialysis by using segmented flow High-throughput droplet digital PCR system for absolute quantitation of DNA copy number Digital droplet lamp as a microfluidic app on standard laboratory devices Integrated optofluidics: A new river of light Fluorescence-activated droplet sorting for single-cell directed evolution Directed evolution of custom polymerases using droplet microfluidics Droplet-based microfluidic platform for high-throughput screening of streptomyces Microfluidic based optical microscopes on chip Laser-induced fluorescence detection system for microfluidic chips based on an orthogonal optical arrangement Customizing droplet contents and dynamic ranges via integrated programmable picodroplet assembler Recent advances on sorting methods of high-throughput droplet-based microfluidics in enzyme directed evolution Fiber-based electrochemical biosensors for monitoring pH and transient neurometabolic lactate In vivo photopharmacology enabled by multifunctional fibers Design and evaluation of a device for fast multispectral time-resolved fluorescence spectroscopy and imaging FLImBrush: Dynamic visualization of intraoperative free-hand fiber-based fluorescence lifetime imaging Human in vivo fluorescence microimaging of the alveolar ducts and sacs during bronchoscopy Molecular detection of gram-positive bacteria in the human lung through an optical fiber-based endoscope Fluidic trapping and optical detection of microparticles with a functional optical fiber A fiber optic system for detection and collection of micrometer-size particles High performance micro-flow cytometer based on optical fibres Optofluidic fiber component for separation and counting of micron-sized particles Highly sensitive broadband Raman sensing of antibiotics in step-index hollow-core photonic crystal fibers Tracking and analyzing the Brownian motion of nano-objects inside hollow core fibers Three-dimensional spatiotemporal tracking of nano-objects diffusing in water-filled optofluidic microstructured fiber Optofluidic fiber-based nanoparticle tracking analysis: Tool to characterize diffusing nanoscale specimen such as SARS-CoV-2 The optofluidic light cage-On-chip integrated spectroscopy using an antiresonance hollow core waveguide The Covid-19 xprize and the need for scalable, fast, and widespread testing Rapid, point-of-care antigen and molecular-based tests for diagnosis of SARS-CoV-2 infection Modeling of droplet generation in a microfluidic flow-focusing junction for droplet size control Passive and active droplet generation with microfluidics: A review SciPy 1.0: Fundamental algorithms for scientific computing in Python Array programming with NumPy Sample-to-answer COVID-19 nucleic acid testing using a low-cost centrifugal microfluidic platform with bead-based signal enhancement and smartphone read-out Rapid detection of COVID-19 coronavirus using a reverse transcriptional loop-mediated isothermal amplification (RT-LAMP) diagnostic platform Effect of velocity on microdroplet fluorescence quantified by laserinduced fluorescence Low index contrast imaging fibers Core crosstalk in ordered imaging fiber bundles This work was funded by a Knut and Alice Wallenberg Foundation Grant (2016.0104). W.M. acknowledges support by the Office of Naval Research Global (Award N62909-20-1-2033). H.P. acknowledges support from KTH Life Sciences Platform and the Lars Hiertas Memorial Foundation. We would like to thank Tharagan Kumar for helpful discussions and assistance with sourcing experimental supplies, Frans Forsberg and João Manoel Barbosa Pereira for technical support at RISE, Laura Barrett for carrying out aluminum deposition, and Cristine C. Kores and Andrius Zukauskas for assistance with dicing fibers. The authors declare no competing interests. The online version contains supplementary material available at https:// doi. org/ 10. 1038/ s41598-022-07306-0.Correspondence and requests for materials should be addressed to H.E.P. or S.S. Publisher's note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http:// creat iveco mmons. org/ licen ses/ by/4. 0/.