key: cord-0876248-t549rf3o authors: Noh, Ji Yun; Choi, Won Suk; Lee, Jacob; Kim, Hye Lim; Song, Joon Young; Cheong, Hee Jin; Kim, Woo Joo title: Clinical performance of the Sofia™ Influenza A + B FIA in adult patients with influenza-like illness date: 2015-06-10 journal: Diagn Microbiol Infect Dis DOI: 10.1016/j.diagmicrobio.2015.05.016 sha: e48373391a76a7985ee645fc108a2bd97543704e doc_id: 876248 cord_uid: t549rf3o The Sofia™ Influenza A + B FIA demonstrated 74.0% sensitivity and 95.4% specificity for influenza A in patients with influenza-like illness in 2012–2013 season. It yielded higher sensitivity than SD Bioline Influenza Ag A/B/A(H1N1/2009) (54.1%) for influenza A (P < 0.01). The Sofia™ Influenza A + B FIA might be useful for rapid diagnosis of influenza. A rapid diagnosis of influenza allows for early decision making and timely intervention in patients with influenza-like illness (ILI). It is also advantageous in the selection of an appropriate antiviral treatment, the reduction of antibiotics use, and the avoidance of unnecessary diagnostic examination, which, in turn, reduces medical costs (Bonner et al., 2003; Nitsch-Osuch et al., 2013) . Rapid influenza diagnostic tests (RIDTs) have been widely used in clinical practice to diagnose influenza because they are easy to use and provide results within 10-15 minutes (CDC, 2013; Cho et al., 2013) . However, the results of conventional RIDTs are limited in reliability due to various and unsatisfactory performances (CDC, 2013; Cho et al., 2013) . The sensitivity of an RIDT varies depending on the type of test and has been shown to be affected by multiple factors: study population, elapsed time from symptom onset, viral titer, sample type and status, circulating influenza virus, and epidemic size. The clinical performances of RIDTs have either been evaluated in a clinical field or by using frozen samples, with reported sensitivities ranging from 40.3% to 73.3% compared to PCR-based detection (Choi et al., 2011; Leonardi et al., 2013; Stripeli et al., 2010; Sutter et al., 2012) . The Sofia™ Influenza A + B FIA (Quidel Corporation, CA, USA) is a novel fluorescent immunoassay used to detect influenza A and B within 15 minutes using the Sofia Analyzer (Lewandrowski et al., 2013) . In this study, the clinical performance of the Sofia™ Influenza A + B FIA was prospectively evaluated and compared with the performance of the SD Bioline Influenza Ag A/B/A(H1N1/2009) (Standard Diagnostics, Yongin, South Korea) in cases of ILI in adult patients. From December 23, 2012, to April 11, 2013, a prospective study was conducted in 3 teaching hospitals in South Korea. Two nasopharyngeal swabs were obtained from adult patients (≥18 years) with ILI who visited an emergency department or outpatient clinic. ILI was defined as an acute respiratory infection with measured fever of ≥38°C and a cough that occurred within 7 days. Among 2 nasopharyngeal swabs, 1 flocked swab was randomly placed in 900 μL of viral transport medium (VTM) (BD, NJ, USA) and was used to perform promptly RIDTs at patients' bedsides. After agitating the flocked swab thoroughly in a vial, samples were tested using both Sofia™ Influenza A + B FIA and SD Bioline Influenza Ag A/B/A(H1N1/2009) simultaneously. All procedures were conducted according to the manufacturers' protocols (Quidel, n.d.; Standard Diagnostics, n.d.) . The other flocked swab was immediately placed in 3 mL of VTM (BD), and samples were kept at −70°C until use. Total RNA was extracted automatically by Nimbus (Hamilton Robotics, Reno, NV, USA) as per the manufacturer's protocol. Real-time reverse transcription polymerase chain reaction (RT-PCR) was performed using Anyplex™ II RV16 Detection (Seegene, Seoul, South Korea) to detect influenza A and B viruses. Among the samples tested, those positive for influenza A virus were selected and tested to differentiate subtype using Seeplex® Influenza A/B Onestep Typing (Seegene) and the PowerChek™ Influenza SIH1/H3/H5 Real-time RT PCR kit (Kogenebiotech, Seoul, South Korea). Sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of each RIDT were calculated using the results of RT-PCR as a gold standard. A McNemar test was performed to compare (Fig. 1) . The sensitivities of both RIDTs for influenza A virus were higher for patients who had visited the hospital within 2 days of symptom onset than for patients who visited hospital more than 2 days after symptom onset ( In a previous study, the sensitivity of the Sofia™ Influenza A + B FIA was reported to be 78.1% during the 2011-2012 influenza season, when implemented at the bedsides of infants and children (Rath et al., 2012) . Using stored samples, it displayed a sensitivity of 82.2% for influenza A, compared to real-time RT-PCR; however, the mean age of the study population was younger (21.3 years old) than ours (Lee et al., 2012) . In another prospective study, the sensitivity of Sofia™ Influenza A + B FIA for the influenza A virus, compared to real-time RT-PCR, was 85% and 69% using nasal or nasopharyngeal swabs, respectively (Lewandrowski et al., 2013) . Although more than 90% of those patients were younger than 22 years old, this study showed a lower sensitivity for detection of influenza A using nasopharyngeal swab than our results (Lewandrowski et al., 2013) . However, direct comparison of the performance of RIDTs between these studies should be limited in interpretation because the clinical performance of RIDTs can be affected by multiple factors, some of which are uncontrollable. In our study, the sensitivities of both RIDTs for influenza A virus were higher in patients who visited the hospital within 2 days of symptom onset, compared to those who visited hospital 2 days after the symptom onset. This might be due to high viral titer during the early infection phase, and this finding is consistent with a previous study (Choi et al., 2011) . However, in a previous report, the sensitivity of RIDT was low during the very early period (within 3 hours of symptom onset) and the time interval from onset to consultation was shorter in RIDT false-negative group (5.5 hours) than in true-positive group (11.5 hours) (Harada et al., 2012) . Time from symptom onset to hospital visit was recorded in the unit of day in our study; thus, analysis based on hours was not available. This study has some limitations. First, we used limited commercial RT-PCR kits as a gold standard. The sensitivities to detect influenza viruses vary with each commercial kit. Second, samples were used promptly for Sofia™ Influenza A + B FIA after specimen collection. However, samples for PCR were eluted in 3 mL of VTM and were frozen at −70°C and used after thawing. There is a chance that influenza virus RNA was diluted in VTM. Freezing and thawing can reduce viral titer in clinical samples. These limitations can produce false negatives in PCR, causing false positives of the Sofia™ Influenza A + B FIA in this study. Also, there is a chance that uneven distribution of influenza virus between 2 nasopharyngeal swabs could affect the results. Only adult patients were included in this study and relatively lower viral titer in adult than children could affect low sensitivity of RIDTs. However, viral titer was not determined. During the study period, influenza B virus rarely circulated; thus, we did not evaluate the performance of RIDTs for influenza B. In addition, direct comparison of sensitivity against influenza A(H1N1)pdm09 between Sofia™ Influenza A + B FIA and SD Bioline Influenza Ag A/B/A(H1N1/2009) is limited in interpretation. Because the Sofia™ Influenza A + B FIA distinguishes influenza A and B, whereas the SD Bioline Influenza Ag A/B/A(H1N1/2009) has 3 lines to differentiate influenza A, B, and influenza A(H1N1)pdm09. Despite limitations, this study is valuable as a prospective study on the performance of the Sofia™ Influenza A + B FIA in the clinical field during the 2012-2013 influenza season in Northern Hemisphere. Due to its ability to provide relatively high sensitivity in the detection of the influenza A virus, it will be one of viable tools for the rapid diagnosis of influenza in clinical practice. Impact of the rapid diagnosis of influenza on physician decision-making and patient management in the pediatric emergency department: results of a randomized, prospective, controlled trial Evaluation of 11 commercially available rapid influenza diagnostic tests-United States Evaluation of five rapid diagnostic kits for influenza A/B virus The clinical usefulness of the SD Bioline Influenza Antigen Test(R) for detecting the Reliability of a rapid test for the clinical diagnosis of influenza A/H1N1 2009 Evaluation of Sofia fluorescent immunoassay analyzer for influenza A/B virus Comparison of conventional lateral-flow assays and a new fluorescent immunoassay to detect influenza viruses Detection of influenza a and B viruses with the sofia analyzer: a novel, rapid immunofluorescence-based in vitro diagnostic device Influence of rapid influenza test on clinical management of children younger than five with febrile respiratory tract infections Early detection of influenza A and B infection in infants and children using conventional and fluorescencebased rapid testing Standard Diagnostics. SD BIOLINE Influenza Ag A/B/A(H1N1) pandemic. Available at Performance of rapid influenza testing in hospitalized children Performance of five FDA-approved rapid antigen tests in the detection of 2009 H1N1 influenza A virus Test sensitivities were compared between the Sofia™ Influenza A + B FIA and the SD Bioline Influenza Ag A/B/A(H1N1/2009) (A) influenza A virus (B) A/H3N2 influenza virus This work was supported by a grant of the TEPIK (Transgovernmental Enterprise for Pandemic Influenza in Korea), which is part of the Korea Healthcare Technology R&D Project by Ministry of Health & Welfare, Republic of Korea (grant number A103001) and Dow Biomedica (Seoul, South Korea). The funders had no role in study design, data collection or analysis, decision to publish, or preparation of the manuscript. This study was supported in part by Dow Biomedica (Seoul, Korea).