key: cord-0876236-yav7g165 authors: Vasan, Aditya; Weekes, Reiley; Connacher, William; Sieker, Jeremy; Stambaugh, Mark; Suresh, Preetham; Lee, Daniel E.; Mazzei, William; Schlaepfer, Eric; Vallejos, Theodore; Petersen, Johan; Merritt, Sidney; Petersen, Lonnie; Friend, James title: MADVent: A low‐cost ventilator for patients with COVID‐19 date: 2020-06-05 journal: Med Devices Sens DOI: 10.1002/mds3.10106 sha: a3b871577fb3b5e10c79dae8c25377d1bca1b2db doc_id: 876236 cord_uid: yav7g165 The COVID‐19 pandemic has produced critical shortages of ventilators worldwide. There is an unmet need for rapidly deployable, emergency‐use ventilators with sufficient functionality to manage COVID‐19 patients with severe Acute Respiratory Distress Syndrome. Here we show the development and validation of a simple, portable, and low‐cost ventilator that may be rapidly manufactured with minimal susceptibility to supply chain disruptions. This single‐mode continuous, mandatory, closed‐loop, pressure‐controlled, time‐terminated emergency ventilator offers robust safety and functionality absent in existing solutions to the ventilator shortage. Validated using certified test lungs over a wide range of compliances, pressures, volumes and resistances to meet U.S. Food and Drug Administration standards of safety and efficacy, an Emergency Use Authorization is in review for this system. This emergency ventilator could eliminate controversial ventilator rationing or splitting to serve multiple patients. All design and validation information is provided to facilitate ventilator production even in resource‐limited settings. A key challenge in the battle against the disease caused by the novel coronavirus SARS-CoV-33 2, COVID-19, is a potential worldwide shortage of mechanical ventilators. The required number 34 of ventilators is projected to significantly exceed capacity, based on the number of patients ex- 35 pected to contract the disease in the United States and the percentage of these likely to require 36 assisted ventilation [1] [2] [3] [4] . Adding to this burden is the fact that COVID-19 patients who develop 37 acute respiratory distress syndrome (ARDS) often require prolonged mechanical ventilation [5- This article is protected by copyright. All rights reserved Accepted Article tion in different healthcare systems across the world; and seeking simple assembly, testing, 56 and use procedures by healthcare personnel with limited experience in ventilation and no ex-57 perience with this type of ventilator system [15] . 58 Modern ICU ventilators provide complex control and intricate feedback loops of a wide 59 variety of respiratory parameters and ventilation modalities. Their operation requires highly 60 specialized staff [16] . Regulatory requirements are understandably high, and pandemic crisis- breaths to provide the most comfortable form of ventilation in a minimally sedated patient. 78 We did not include a synchronized mode of ventilation in the design of this ventilator, recog-79 nizing that patients with COVID-19 and severe ARDS will require sedation and possibly phar-80 macologic paralysis to facilitate optimal ventilation [21, 22] . The advantages of this approach 81 include simplified ventilator settings and simplified troubleshooting with a single-mode con-82 tinuous, mandatory, closed-loop pressure-controlled time-terminated ventilator (from now on 83 referred to simply as pressure-controlled). This approach provides predictable delivery of ven-84 tilated breaths, and streamlined device production. Further design choices were based on the 85 dual goals of safe, effective ventilation and quick production as detailed in the next section. This article is protected by copyright. All rights reserved FIG. 1. The ventilator was tested on a lung simulator. All parameters were tested to their stated limits (over 200 individual experiments) and according to International Standards Organization (ISO) standards for pressure controlled ventilation. Notice that the dead space is kept to a minimum by reducing the length of tube between the bag and the lung simulator; this configuration was reproducible with a full-sized simulator manikin and a standard adjustable overbed hospital bedside table. The system shown here is an early prototype with exposed electronics, but is to be supplied with housings as depicted in Fig. 5 2. The MADVent has a novel torque conversion mechanism via a simple pulley and lanyard 112 system to convert the relatively low-torque, high speed rotation of the motor to a high-113 torque, reduced speed resuscitation bag compression mechanism. This is superior to the 114 ubiquitous geared rack-and-pinion mechanisms of other low-cost ventilators as it offers 115 greater pressure, at least doubles the maximum ventilation rate, has no backlash" and is 116 far quieter. It is also much more durable, as the nylon geared mechanisms used in other 117 systems are subject to wear and failure much faster than our approach. The overall range of parameters at which the system is capable of operating is listed in Ta- The hardware on the system allows for a volume-driven approach to ventilation in addition 156 to pressure-controlled ventilation with continuous feedback. Tests were conducted to charac-FIG. 2. The MADVent Mark V has alarms for high and low volume that may be set between 200 and 1000 mL. In this example, the system was run at a rate of 13 breaths per minute (ventilation rate), a PEEP value of 15 cm H 2 O and the compliance on the lung simulator was initially set to 0.03`/cmH 2 O. A) The high-volume alarm threshold was set to 500 mL for the first case. PEEP was decreased from 15 cm H 2 Oto 5 cm H 2 O in order to increase the tidal volume delivered to the lung simulator. A high-volume alarm was triggered when the calculated tidal volume exceed the limit set by the healthcare provider. A relevant clinical scenario for this alarm would be a leak in the inspiratory circuit leading to an increase in volume delivered without the target pressure being reached. B) The low-volume alarm is triggered once the calculated volume drops below the lower limit set by the healthcare provider. This was simulated by increasing the PEEP up to 17 cm H 2 O. A relevant clinical scenario for this alarm would be the inspiratory line being kinked. C) The high-pressure scenario was simulated by interrupting the expansion of the lung simulator during inspiration to simulate a patient coughing. The high-pressure alarm was triggered when the pressure exceeded the set value of 30 cm H 2 O. Other scenarios are provided in the Supplementary . Furthermore, a post-hoc quadratic curve fit (3.47£10°4¡ 2 +0.322¡°52.5 with R 2 = 0.953) is provided showing a slightly improved fit, indicating that a quadratic function can adequately represent the tidal volume as a function of the angle ¡. In B, the volume corresponding to a given motor rotation is seen to increase with compliance-accounting for the spread in the data along with experimental error. In C, the difference between peak pressure and PEEP is seen to increase along the model, as expected due to the ideal gas law. and increasing reliance on an intact supply chain. However, we did test the system as a volume- was programmed to trigger upon two consecutive high pressure events, after which the system 168 will release the bag compression arm and commence a new respiration cycle at lower tidal vol-169 umes but increased rate in order to meet the minute ventilation set by the healthcare provider. In the event of an accidental disconnection of the endotracheal tube or other significant leak 171 in the system, a low-pressure alarm will be triggered as illustrated in Fig. 4B A high-pressure condition was simulated by decreasing patient lung compliance. The system triggered an alarm once the pressure went above 42 cm H 2 O. B) The low-pressure alarm is triggered once the in-line pressure value drops below the lower limit. A low-pressure situation was simulated by disconnecting the endotracheal tube to trigger an alarm which results in the system immediately stopping. C) In the event that the tubing is kinked or there is a blockage in the endotracheal tube, the pressure begins to rise until the upper threshold is reached. This triggers a high-pressure alarm and causes the system to resume ventilation at a lower volume, but at an increased rate according to the set minute ventilation. This article is protected by copyright. All rights reserved Coventor ventilators are all less expensive and simpler to manufacture than the PB560. The following information on the MIT E-Vent is representative of the publicly available in- This article is protected by copyright. All rights reserved The bag is mounted into a frame under a lever arm that is subsequently used to compress 379 the bag, as shown in Fig. 5 . The entire ventilator structure, including the bag mounting frame 380 and arm, can be rapidly laser cut from polyoxymethylene (acetal) in 15 min, and assembled us-381 ing readily available hardware. An alternate material choice is polycarbonate, which has supe-382 rior resistance to commonly used hospital disinfectants such as sodium hypochlorite (bleach). Complete design files are provided for the reader (see Supplementary Information) . Two con- sound. This condition has been programmed to occur in three cases: when the low-volume 483 alarm threshold is higher than the high-volume alarm threshold, when the set peak pressure is 484 higher than the high-pressure alarm threshold and when the user set inspiratory time is more 485 than 75% of the inspiratory time calculated from the user set respiratory rate. After the parameters have been set, the system waits for activation via a toggle switch before 487 initiating ventilation. During inspiration, the motor rotates an amount proportional to the dif-488 ference between the intended pressure and the current measured pressure at each time-step. The intended pressure at each time-step is determined by a monotonically increasing function 490 between p(t = 0) = 0 and p(t = t i ) = p p , where p is pressure, t is time, p p is the peak pressure 491 set by the provider, t i is the inspiratory time set by the provider. Once the peak pressure or the 492 inspiratory time has been reached, the motor reverses direction at a set speed until it reaches 493 the zero position, which is defined by the compressor arm photointerrupter switch and con-494 firmed by the motor encoder. ance and circuit integrity. In addition to alarms for pressure, the system is equipped with temperature sensors that The alarm system of the MADVent Mark V ventilator was tested by simulating the same 526 alarm conditions that would normally be detected by a commercial ventilator. Excessively high 527 and low volume conditions were simulated by changing the PEEP values as shown in Fig. 2(a,b) ; 528 each of these conditions triggered the respective alarms on our ventilator. Likewise, high pres- This article is protected by copyright. All rights reserved sure events that could be due to a patient coughing or a kink in the ventilation tube, blocking 530 air flow to the patient, produce an alarm in Fig. 2(c) , but only after repeated coughing-as de-531 sired. Triggering alarms after a single cough might inappropriately encourage the healthcare 532 provider to find a way to defeat the alarm. The admissible range of operating pressure, PEEP, 533 time, and breaths per minute were determined for our system from the lung compliance and 534 the peak and PEEP pressure values as shown in Fig. 3 . Once the alarms were confirmed to operate according to expectations by our anesthesiolo- Clinical characteristics of 138 hospitalized patients with 2019 novel coronavirus-568 infected pneumonia in wuhan Locally informed simulation to predict hospital capacity needs during the 571 COVID-19 pandemic Critical supply shortages -the need for ventila-573 tors and personal protective equipment during the COVID-19 pandemic COVID-19 -navigating the uncharted Features, evaluation and treat-579 ment coronavirus (COVID-19) Intensive care management of coronavirus disease 2019 (COVID-19): challenges and 581 recommendations COVID-19 in critically ill patients in the seattle region-case series Clinical course and outcomes of critically ill patients with SARS-CoV-2 pneumo-585 china: a single-centered, retrospective, observational study. The Lancet Respiratory Stockpiling ventilators for influenza pandemics The world is scrambling to buy ventilators in the covid-19 pandemic. one coun-597 try has only four of them -for 12 million people High contagiousness and rapid spread of severe acute respiratory syndrome coro-600 navirus 2 The need for ventilators in the developing world: 602 An opportunity to improve care and save lives Ford and GM are making tens of thousands of ventilators. it may al-604 ready be too late Low tidal volume ventilation use in acute respiratory distress syndrome. Critical 607 Care Medicine Driving pressure and survival in the acute respiratory distress syndrome Medicine/Society of Critical Care Medicine clinical practice guideline: Mechanical ventilation in 612 adult patients with acute respiratory distress syndrome Ventilation with lower tidal volumes as compared with traditional tidal volumes 615 for acute lung injury and the acute respiratory distress syndrome Sedation and neuromuscular blocking agents in acute respiratory distress syn-618 drome Prevention & of Acute Lung Injury (PETAL) network, E. T. Early neuromuscular blockade in the 620 acute respiratory distress syndrome Emergency use authorization for ventilators (en) lung ventilators 624 and related equipment -vocabulary and semantics Intubation and ventilation amid the covid-19 outbreakwuhan's experience. Anesthe-627 siology (en) medical electri-629 cal equipment -part 2-80: Particular requirements for basic safety and essential performance of 630 ventilatory support equipment for ventilatory insufficiency Critical care surge response strategies for the 2020 COVID-19 outbreak in the united 633 states The design and evaluation of a novel 637 low-cost portable ventilator Design and prototyping of a low-cost portable mechanical ventilator. Trans-639 actions of the The Other Option Is Death York starts sharing of ventilators MIT emergency ventilator (E-Vent) project COVID-19 ventilator Surge capacity mechanical 648 ventilation Health Systems Research Inc. Bioterrorism and other public health emergencies: Altered standards 650 of care in mass casualty events The effects of gas humidification with high-flow nasal cannula on cultured hu-653 man airway epithelial cells Incidence, risk factors and outcome of barotrauma in mechanically ventilated 655 patients Coronavirus disease 2019 (COVID-19): Critical care 657 issues International Standards Organization ISO/TC 210. ISO 14971:2019 medical devices -application 659 of risk management to medical devices The realities of risk-cost-benefit analysis Trends in mechanical ventilation: are we ventilating our pa-663 tients in the best possible way? Breathe Flow sensor Nonlinear flow sensor calibration with an accurate 669 syringe Do new anesthesia ventilators deliver small 671 tidal volumes accurately during volume-controlled ventilation? Accuracy of small tidal volume measurement comparing 673 two ventilator airway sensors Pressure-volume curves of the respiratory system Bench test evaluation of volume delivered by modern icu ventilators during 676 volume-controlled ventilation Medicine (2020).