key: cord-0866149-0fo42uxh authors: Price, Courtney; Ben-Yakov, Maxim; Choi, Joseph; Orchanian-Cheff, Ani; Tawadrous, Davy title: Barrier enclosure use during aerosol-generating medical procedures: A scoping review date: 2020-11-06 journal: Am J Emerg Med DOI: 10.1016/j.ajem.2020.10.071 sha: 977d3617690703525918d18fa566f31d5358944f doc_id: 866149 cord_uid: 0fo42uxh INTRODUCTION: Barrier enclosure devices were introduced to protect against infectious disease transmission during aerosol generating medical procedures (AGMP). Recent discussion in the medical community has led to new designs and adoption despite limited evidence. A scoping review was conducted to characterize devices being used and their performance. METHODS: We conducted a scoping review of formal databases (MEDLINE, Embase, Cochrane Database of Systematic Reviews, CENTRAL, Scopus), grey literature, and hand-searched relevant journals. Forward and reverse citation searching was completed on included articles. Article/full-text screening and data extraction was performed by two independent reviewers. Studies were categorized by publication type, device category, intended medical use, and outcomes (efficacy – ability to contain particles; efficiency – time to complete AGMP; and usability – user experience). RESULTS: Searches identified 6489 studies and 123 met criteria for inclusion (k = 0.81 title/abstract, k = 0.77 full-text). Most articles were published in 2020 (98%, n = 120) as letters/commentaries (58%, n = 71). Box systems represented 42% (n = 52) of systems described, while plastic sheet systems accounted for 54% (n = 66). The majority were used for airway management (67%, n = 83). Only half of articles described outcome measures (54%, n = 67); 82% (n = 55) reporting efficacy, 39% (n = 26) on usability, and 15% (n = 10) on efficiency. Efficacy of devices in containing aerosols was limited and frequently dependent on use of suction devices. CONCLUSIONS: While use of various barrier enclosure devices has become widespread during this pandemic, objective data of efficacy, efficiency, and usability is limited. Further controlled studies are required before adoption into routine clinical practice. During the COVID-19 pandemic, the threat of diminishing supplies of personal protective equipment sparked an interest in alternative means to protect healthcare providers. One such means included barrier enclosure devices, which are generally described as a plastic sheet over a structural frame or a transparent plastic four-sided box that are used as a potential method of protecting healthcare providers from SARS-CoV-2 during aerosol generating medical procedures (AGMP) (e.g. intubation or extubation). This device is typically placed in between a patient and airway operator during an AGMP as a means of physically limiting the transmission of aerosols and/or droplets to healthcare providers. Currently there is limited evidence to support the use of barrier enclosure devices and important questions remain regarding their efficacy in reducing contamination, efficiency of use, and usability within various healthcare settings. In May 2020, the United States Food and Drug Administration issued a temporary emergency medical device license for the use of protective barrier enclosures [1] . While healthcare institutions continue to test, modify, and adopt these barriers into practice, we sought to collate and characterize the published literature on devices that are being used in various settings, as well as elucidate any performance outcomes (i.e., efficacy, efficiency, usability) associated with each system. A scoping review was selected given the heterogeneity of the literature on this topic. A protocol of our methodology was published a priori and followed PRISMA-ScR guidelines [2, 3] . A search to identify barrier enclosure devices was executed by an academic information specialist in bibliographic databases including Ovid MEDLINE, Ovid Embase, Cochrane Database of Systematic Reviews, Cochrane Central Register of Controlled Trials and Scopus (2000-01-01 to 2020-06-24) for the main concepts of AGMPs and barrier enclosure devices (Appendix A) across all languages. The year 2000 was chosen to capture barrier devices potentially used during previous pandemics (e.g., SARS-CoV-1). We excluded nonhuman studies, conference and book materials. Additionally, a grey literature search of Google Scholar, clinical trials registries (ClinicalTrials. gov, WHO Clinical Trials), pre-print repositories (OSF, MedRvix), disseminated reports ( were manually searched on 2020-07-03 reviewing articles published in March to July 2020 issues and those available as early release. Forward and reverse citation searching was completed on all included articles (2020-07- 19) . All citations were managed in Covidence (covidence. org) screening software. Two reviewers (CP, DT) independently evaluated the eligibility of studies on the basis of title and/or abstract using pre-established inclusion and exclusion criteria (Table 1) . A sample of 100 articles were screened to ensure consistency among reviewers and fidelity of established criteria. Reviewers independently evaluated the eligibility of all articles; disagreements were resolved by re-evaluation, discussion, and when necessary, in consultation with a third reviewer. Full-text articles were retrieved if reviewers considered a citation potentially relevant. Reviewer agreement for study eligibility was assessed using the unweighted Cohen's kappa coefficient. Data abstraction was completed independently by one reviewer (CP) using a standardized form (Appendix B) and verified by a second reviewer (DT, JC, MBY). We abstracted publication details (author, title, publication date, country of origin, publication status, publication type), setting, device design details, intended medical use, methods and outcomes. A list and definition of variables collected can be found in Appendix B. Devices were categorized as either a box, plastic sheet (with frame), plastic sheet (without frame), or other system. Outcomes, both qualitative and quantitative, were categorized as either efficacy (i.e., related to the device's ability to protect the intubator and contain particles), efficiency (i.e., time taken to perform an AGMP) or usability metrics (i.e., feedback on experience of the use of the system). A total of 6336 articles were identified through formal database search strategies, and 153 articles through grey literature and citation screening (Fig. 1) . After duplicate removal, 4509 unique articles were screened, and 169 full-text articles were assessed for eligibility. We identified 123 articles for inclusion. Our kappa coefficient was good for title/abstract screening (kappa = 0.81) and full-text review (kappa = 0.77). Most articles were published between March 2020 -July 2020 (n = 120), with three articles published prior to 2020 [4] [5] [6] . Over half of articles were published as letters/commentaries (58%, n = 71), 29% as original research studies (n = 36), and 13% (n = 16) as brief/ short reports. Publications originated from 27 unique countries with the top 3 countries including the United States (34%, n = 42), India (10%, n = 12), and Canada (10%, n = 12). Commonly reported barrier enclosure devices include box (42%, n = 52) and plastic sheet (54%, n = 66) systems. Over half (59%, n = 39/66) of plastic sheet systems utilized a supportive frame and 41% (n = 27/66) had no supportive structure (Fig. 2) . Box designs were often similar to the original design in Canelli et al. [7] , which is a transparent 4-sided structure with two open faces: an inferior face bound by the stretcher and a caudal face pointed towards the foot of the bed. Common modifications included change in the number or size of ports (i.e., for operators and/or tools) [8] [9] [10] [11] [12] , increased device size for improved operator ergonomics and/or patient body habitus [13] [14] [15] , built-in gloves and/or port coverings [11, 12, 16, 17] , addition of a plastic drape or covering on the caudal face [18] [19] [20] , a sloped top panel for improved visibility [15, 18, 21] and the use of a negative suction system [22] [23] [24] . Conversely, plastic sheet systems with frames were constructed using polyvinyl chloride tubing [25, 26] , operating room equipment (e.g., anesthetic screens) [24, 27] , and Mayo stands [28, 29] . Six articles introduced a plastic canopy system, which is semicircular in shape enclosing the patient's upper or full body [4, [30] [31] [32] [33] [34] . Alternatively, plastic sheet systems without a frame were often akin to surgical draping, where a clear sheet drapes over the patient's head, neck and/or entire body and the physician works beneath the drape or cuts an opening into the plastic sheet [5, [35] [36] [37] . There were a number of other unique designs. Three articles introduced large, non-mobile plastic chamber units for COVID-19 testing [38, 39] and outpatient ENT procedures [40] in which the patient entered the closed system and the procedure was performed through Table 1 Inclusion and exclusion criteria Inclusion criteria Exclusion criteria 1) Descriptions, design, and/or protocol for barrier enclosure use in AGMPs a 2) All article types (e.g. original research, reviews) 3) Any publication status (e.g., pre-print, online) 4) Time frame: 2000-2020-06-24 5) Studies in: humans, experimental, simulation 6) Any language 1) Conference abstracts, posters or proceedings, registered trials, online website material 2) Critique or opinion on prior published work, with no introduction of a new device 3) Non-infectious risk exposure (e.g. chemotherapy, radiation) a Defined as any enclosure which surrounds the patient and aims to prevent droplet spread and aerosol dispersion into the environment during an intervention. The most commonly reported use was for airway management (67%, n = 83) ( Table 2 ). Within these, 84% reported use for intubation or extubation (n = 70/83), 7% (n = 6/83) for tracheostomies [23, [47] [48] [49] [50] [51] , and 6% (n = 5/83) for non-invasive respiratory support (e.g. highflow nasal cannula) [26, 30, 34, 52, 53] . Two studies (2%, n = 2/83) used a device in pediatric laryngoscopy and bronchoscopy [54, 55] . Nine studies (7%) discussed these devices for general AGMPs [22, 25, 32, 46, [56] [57] [58] [59] [60] and 9 (7%) for endoscopic procedures. [10, 11, 43, 45, [61] [62] [63] [64] A small proportion of studies (11%, n = 14) used an enclosure during surgical procedures, mainly for otolaryngology procedures (57%, n = 8/ 14) (e.g. mastoidectomy, endo-nasal/endo-oral procedures) [40, [65] [66] [67] [68] [69] [70] [71] as well as in other types of surgery (43%, n = 6/14) (e.g. craniotomy, oral maxillofacial, colorectal surgery) [5, 6, [72] [73] [74] [75] . Other uses included dental procedures [41, 76] , dermatological procedures [29] , and regional anesthesia [9, 77] (4%, n = 5/123). Over half of articles included an evaluation component (54%, n = 67), the majority of which only included qualitative outcomes (54%, n = 36/67). Among these, 70% (n = 47) of studies reported on the use of enclosures in simulation settings, 19% (n = 13) reported their use in real patients, and 10% (n = 7) reported on use in both environments. Efficacy was the most frequently reported outcome among articles (82%, n = 55/67) followed by usability (39%, n = 26/67) and efficiency (15%, n = 10/67). The most common method to assess the device's ability to contain particles or prevent contamination was through the visual assessment of droplets or smoke (71%, n = 39/55), primarily with box systems (51%, n = 20/39). Four studies used the ability to smell [78, 79] or taste a bitter solution [60, 80] as a proxy for aerosols escaping into the environment. Using these qualitative methods, studies concluded that the use of a barrier device was effective at either preventing or reducing the number of particles escaping the system. Only 40% (n = 22/55) studies reported quantitative results. Three of these studies used pre-established grids to quantify exposure outside of the enclosure with fluorescent dye or gross droplets and reported success in reducing contamination [13, 49, 81] . Two studies reported no SARS-CoV-2 infection rates of physicians after using the system [59, 82] , while another four studies detected the presence of molecules contained within the enclosure and/or a decrease in particles outside the enclosure as a proxy for its effectiveness [5, 6, 34, 74] . The majority of barriers (77%, n = 17/22) with objective findings used suction to generate negative pressure and reported particle counts or aerosol clearance rates (59%, n = 13/22) ( Table 3 ). In contrast to the visual contamination studies showing effectiveness, evidence from quantitative data was often less favourable and contingent on the use of suction devices. For example, Simpson et al. [24] evaluated the efficacy of four different designsa box, sealed box (caudal end closed), and two plastic sheet barrier systems and found that only when suction was applied, particle counts decreased. Similar results were seen in in Lyaker et al. with increased particle detection outside the chamber without the use of suction [83] . Usability was assessed primarily by self-reported qualitative feedback from physicians using these devices (81%, n = 21/26). Generally, authors reported success carrying out procedures using the device with no major issues, [33, 44, 54, 61, 75, 77, 84] however four studies using the box system reported additional workflow complexities [85] and challenges while performing intubation. [12, 86, 87] Six articles included a quantitative assessment of usability for intubation (23%, n = 6/26, 4, 12, 88-91], mainly in the box (83%, n = 5/6) [12, [88] [89] [90] [91] and one in a plastic sheet system (canopy) (17%, n = 1/6) [4] . Seger et al. reported a limited increase in time required for device maneuverability: removal and disposal within 10 s, and completion of a position change within the enclosure in less than 2 s. [91] However, Clariot et al. [88] , Begley et al. [12] , and Hamal et al. [89] reported worsening laryngoscopic views when using a box system. Similarly, Serdinšek et al. [90] and Plazikowski et al. [4] both reported more difficulty with airway management when using box and plastic canopy systems. The most frequent efficiency metric reported was time to intubation or related metrics to securing an airway (e.g., first-pass success) (70%, n = 7/10, 4, 12, [88] [89] [90] [91] [92] primarily in the box system (86%, n = 6/7) [12, [88] [89] [90] [91] [92] . Those assessing intubation times (40%, n = 4/10) noted [13, 20, 21, 24, 87] . In Brown et al. [87] , Ibrahim et al. [20] ., and Gore et al. [21] , particles escaped through the open caudal end of box systems with increased contamination of the operator and/or environment relative to the plastic systems during airway management. Laosuwan et al. also assessed droplet contamination on a standardized grid in an extubation simulation and similarly reported increased contamination with box systems relative to plastic sheet systems [13] . Simpson et al. found that there was no significant difference in particle exposure outside the enclosure when comparing plastic sheet systems to no system during intubation, whereas the use of the box system concentrated particles without limiting dispersion [24] . Only one study compared usability between a box and plastic sheet system and reported that physicians favoured the plastic sheet system due to ease of mobility and the ability to accommodate an airway assistant [87] . Barrier enclosures are described as innovative systems which protect healthcare workers from infectious disease transmission. We identified 123 articles from 27 countries, the majority of which were published following the original aerosol box design released in April 2020 [7] . Across these studies, three general device types were identified: box, plastic sheet with frame, and plastic sheet without frame systems for use in airway management (intubation, extubation, tracheostomies or respiratory support) or general aerosolizing medical procedures. To date, there is a lack of strong evidence to support the use of barrier systems in clinical settings. Our review demonstrated a reliance on short letters/commentaries to validate various devices' medical use and safety and limited rigorous trials. Currently, evidence to support the reduction of aerosol and droplet contamination is based primarily on visual assessments of aerosol and droplet spread. While these results are generally positive, emerging quantitative studies have reported less favourable results that frequently depend on concurrent use of a suction device [24, 83] . Often discussed as a lowcost, pragmatic means of protecting physicians, use of the box systems in some instances demonstrated a delay in time to intubation [12, 88] AGMPs. Particle generator and a particle counter inside enclosure. Hospital suction / commercially available suction / none Aerosol clearance was significantly hastened with suction vs. passive clearance and the commercial suction device was better than in-hospital system. AGMPs. Simulated cough with normal breath measured through a particle generator. Suction position /device openings/suction flow rate / device vs. no device Device prevented particle escape. Optimal condition was when the suction was vertically next to the patient's head. Le, US 95 Airway Management. Atomizer used to simulate aerosol production with counter inside/outside enclosure. None listed Containment of greater than 90% of sub-micrometer particles across all particle sizes. Lang, US 57 AGMPs. Humidifier generated particles with particle counters placed inside/outside enclosure. Device vs. no device / Suction vs. no suction Particle count detected outside hood significantly decreased with the system. Suction system reduced particle count inside the enclosure. Shaw, US 52 Respiratory Support. Humidifier generated particles with particle counters placed inside/outside enclosure. Nasal cannula used inside hood. Airway Management. Flow rate testing completed with two suction sources. Theoretical times for airborne contaminant at 99% and 99.9% would be faster with the in-hospital suction system due to higher flow rates. Adir, Israel 30 Respiratory Support. Face velocity and smoke direction of air measured perpendicular to hood. Photometry used for particle leakage. The average air flow velocity was 4.4 m·s − 1 with the smoke flowing into the back side of the canopy. Filtration efficiency was reported at 0.0006%. Box & Plastic Sheet. Airway Management. Nebulized saline through a simulated cough with particle counter outside enclosure. Box / vertical drape / horizontal drape / sealed box with suction / sealed box without suction The sealed intubation box with suction resulted in a decrease in almost all particle sizes across all time periods. The box had an increase in particle exposure. No difference using the plastic drapes. and worsening laryngoscopic views [12, 88, 89] , which has important clinical implications in physiologically difficult intubations. In fact, while simplistic, plastic sheet systems appear to outperform box systems in efficacy and usability characteristics, with less environmental contamination [13, 24] and better ergonomics [87] . These variable characteristics are important to consider in light of the evolving SARS-CoV-2 pandemic. In May 2020, the United States FDA granted emergency approval for barrier enclosure device manufacturing, distribution and use during AGMPs without guidance on the design or intended medical uses for these devices [1] . Subsequently, a plethora of various devices were heavily promoted through social media, press and in many medical journals translating to a large uptake of these systems [93] despite limited scientific evidence on efficacy, efficiency and usability. Recognizing this risk, the FDA has since revoked its emergency license for barrier enclosures devices in August 2020 [1] , and now recommends the use of enclosures with suction devices in keeping with emerging objective evidence. [12, 24] The pandemic has highlighted the delicate balance of thorough evaluation with the need for immediate solutions. Commercial medical devices undergo rigorous testing in order to prove efficacy and safety for the patient and physician and requires strict reporting of adverse events through a centralized system to make decisions regarding continued use [94] . This is an opportunity for regulatory bodies to reexamine how emergency approvals are granted, and to set up infrastructure to encourage local innovation while providing a platform to register and monitor its effects, similar to how trials are registered. In light of the established characteristics and performance outcomes, researchers and innovators looking to further develop and optimize barrier enclosures should focus on quantitative assessments of efficacy, efficiency, and usability in real clinical environments. Other opportunities for further exploration include focusing on patient-centered outcomes, such as frequency of desaturations and peri-intubation cardiac arrest, as well as the economics associated with implementation, widespread adoption, and maintenance (e.g., sterilization) of these devices. Our review focused on the published literature related to the use of barrier enclosure devices and did not include designs that were published on websites, social media or design sites. While devices published in non-academic mediums may have been missed in our scoping review, we believe this further highlights the need for a central platform to catalog and regulate the use of barrier enclosures. We also performed the last formal search on 2020-06-24 and were unable to obtain the full text of one study. As a rapidly growing field of research, other studies published since that time were not included in this review. However, forward and reverse citation screening on included articles was completed. Many of these enclosure systems were devised in the early stages of the pandemic when things were rapidly evolving with many unknowns. As a consequence of that, the studies largely included qualitative and simulation-derived data on process measures. It will be important to perform quantitative studies analyzing real-world outcome (e.g. infectivity rates) in order to make any conclusions on the efficacy of these devices. The use of barrier systems in clinical care was introduced to protect physicians during AGMPs. However, the efficacy of barrier enclosures in protecting physicians is limited. Overall, clinical use of these devices in the absence of thorough medical device testing is concerning and contrary to regulatory legislation intended to safeguard patient and physician safety. This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors. None. -Device Description or Modificationpresenting a new device, including the modification of an existing device. -Evaluationarticle includes measurement of any outcomes (quantitative or qualitative). -Other Design details of all the device (e.g. # of drapes, different size / shapes, coverage provided) and any unique features included. Options: -Plastic Boxsimilar in design to the 4-sided aerosol box design -Plastic Sheet -Rigid Frame Actual use or intended medical use of the device. If describing a device for "general use", but no procedure specified, list "AGMPs". Options: intubation, extubation, tracheostomy, endoscopy, bronchoscopy, NIPPV, ENT surgeries (e.g. endonasal / endo-oral), dental procedures, other (free-text) Study design type, if applicable and available. Description of the patient population in the study. Study participants, sample size and method of measurements. Listed outcomes for efficacy, efficiency, usability, and other. Main results of each study outcome(s). Main study limitations as listed by the author(s) Protective Barrier Enclosures Without Negative Pressure Used During the COVID-19 Pandemic May Increase Risk to Patients and Health Care Providers -Letter to Health Care Providers Barrier enclosure use during aerosol-generating medical procedures: a scoping review protocol PRISMA extension for scoping reviews (PRISMA-ScR): checklist and explanation Emergency airway Management in a Simulation of highly contagious isolated patients: both isolation strategy and device type matter The extent of environmental and body contamination through aerosols by hydro-surgical debridement in the lumbar spine Aerosols produced by high-speed cutters in cervical spine surgery: extent of environmental contamination Barrier enclosure during endotracheal intubation Aerosol box for protection during airway manipulation in covid-19 patients Aerosol prevention box for regional anaesthesia for eye surgery in COVID times Development and evaluation of a novel protective device for upper gastrointestinal endoscopy in the COVID-19 pandemic: the EBOX The C-Cube: an endoscopic solution in the time of COVID-19 The aerosol box for intubation in coronavirus disease 2019 patients: an in-situ simulation crossover study Acrylic Box Versus Plastic Sheet Covering on Droplet Dispersal During Extubation in COVID-19 Patients Maximising application of the aerosol box in protecting healthcare workers during the COVID-19 pandemic Novel Coronavirus Disease 2019 (COVID-19) Aerosolization Box: Design Modifications for Patient Safety. J Cardiothoracic Vascular Anesthesia Low-cost double protective barrier for intubating patients amid COVID-19 crisis Protective device during Airway Management in Patients with coronavirus disease 2019 (COVID-19). Anesthesiology Barrier enclosure for airway management in COVID-19 pandemic Rapid prototype feasibility testing with simulation: improvements and updates to the Taiwanese "aerosol box Comparison of the effectiveness of different barrier enclosure techniques in protection of healthcare workers during tracheal intubation and Extubation Intubation Containment System for Improved Protection from Aerosolized Particles during Airway Management Minimising exposure to droplet and aerosolised pathogens: a computational fluid dynamics study. medRxiv St George's COVID shield for use by ENT surgeons performing tracheostomies Measurement of airborne particle exposure during simulated tracheal intubation using various proposed aerosol containment devices during the COVID-19 pandemic. Anaesthesia (Lond) A multipurpose portable negative air flow isolation chamber for aerosol-generating procedures during the COVID-19 pandemic Aerosol Barrier Hood for Use in the Management of Critically Ill Adults With COVID-19 Use of a plastic barrier curtain to minimize droplet transmission during tracheal extubation in patients with COVID-19 Barrier enclosure using a Mayo stand and plastic sheet during cardiopulmonary resuscitation in patients and its effect on reducing visible aerosol dispersion on healthcare workers Thinking outside the box in COVID-19 era COVID-19: minimising risk to healthcare workers during aerosol-producing respiratory therapy using an innovative constant flow canopy Developing a Simply Fabricated Barrier for Aerosol Generating Procedures Rapid development of a portable negative pressure procedural tent Introducing the "corona Curtain": An innovative technique to prevent airborne COVID-19 exposure during emergent intubations Environmental safety evaluation of the protection and isolation system for patients with covid-19 Adaptation to the Plastic Barrier Sheet to Facilitate Intubation during the COVID-19 Pandemic. Anesthesia analgesia Barrier system for airway management of COVID-19 patients Safer intubation and extubation of patients with COVID-19 Protecting health care workers in the front line: Innovation in COVID-19 pandemic Cost-effective innovative personal protective equipment for the management of COVID-19 patients A Closed Chamber ENT Examination Unit for Aerosol-Generating Endoscopic Examinations of COVID-19 Patients Development of a Device to Reduce Oropharyngeal Aerosol Transmission Development of a protective device for RT-PCR testing of COVID-19 Simple barrier device to minimize facial exposure of endoscopists during COVID-19 pandemic Modified Wake Forest Type Protective Shield for an Asymptomatic, COVID-19 Non-Confirmed Patient for Intubation Undergoing Urgent Surgery Patient-Worn Enhanced Protection Face Shield for Flexible Endoscopy. Otolaryngology Head and Neck Surgery Universal mobile protection system for aerosol-generating medical interventions in COVID-19 patients Barrier device prototype for open tracheotomy during COVID-19 pandemic Health workers' safety during tracheostomy in COVID-19 patients: Homemade protective screen. Head Neck Tracheostomy during COVID-19 pandemic-Novel approach. Head Neck Negative-Pressure Aerosol Cover for COVID-19 Tracheostomy Open Tracheostomy for Covid19 Positive Patients: A Method to Minimize Aerosolization and Reduce Risk of Exposure Intensive care unit isolation hood decreases risk of aerosolization during noninvasive ventilation with COVID-19 Negative Pressure Tent to Reduce Exposure of Health Care Workers to SARS CoV-2 During Aerosol Generating Respiratory Therapies Pediatric Microlaryngoscopy and Bronchoscopy in the COVID-19 Era Pediatric laryngoscopy and bronchoscopy during the COVID-19 pandemic: A four-center collaborative protocol to improve safety with perioperative management strategies and creation of a surgical tent with disposable drapes Barrier Shields: Not Just for Intubations in Today's COVID-19 World? Effectiveness of a negative-pressure patient isolation hood shown using particle count Rapid clearing of aerosol in an intubation box by vacuum filtration The vacuum assisted negative pressure isolation Hood (VANISH) system: novel application of the Stryker Neptune TM suction machine to create COVID-19 negative pressure isolation environments Enhanced draping for airway procedures during the COVID-19 pandemic Endoprotector: protective box for safe endoscopy use during COVID-19 outbreak New protection barrier for endoscopic procedures in the era of pandemic COVID-19 Shielding for patients using a single-use vinyl-box under continuous aerosol suction to minimize SARS-CoV-2 transmission during emergency endoscopy Cover Page Dual Purpose Easily Assembled Aerosol Chamber Designed for Safe Endoscopy and Intubation during COVID Pandemic A Simple Technique for Droplet Control During Mastoid Surgery Aerosol dispersion during Mastoidectomy and custom mitigation strategies for Otologic surgery in the COVID-19 era. Otolaryngol Head Neck Surg New device and technique to protect intubation operators against COVID-19. Intensive Care Medicine Disposable customized aerosol containment chamber for oral cancer biopsy: A novel technique during COVID-19 pandemic The nasal tent: an adjuvant for performing endoscopic endonasal surgery in the Covid era and beyond Endoscopic skull base and transoral surgery during COVID-19 pandemic: minimizing droplet spread with negative-pressure otolaryngology viral isolation drape Exoscope and personal protective equipment use for Otologic surgery in the era of COVID-19 COVID-19: impact on colorectal surgery Surgeon's protection during ophthalmic surgery in the Covid-19 era: a novel fitted drape for ophthalmic operating microscopes The craniotomy box: an innovative method of containing hazardous aerosols generated during skull saw use in autopsy on a COVID-19 body Reducing transmission of COVID-19 using a continuous negative pressure operative field barrier during oral maxillofacial surgery Aerosol box for dentistry Novel method of performing brachial plexus block using an aerosol box during COVID-19 pandemic Barrier Devices for Reducing Aerosol and Droplet Transmission in Coronavirus Disease 2019 Patients: Advantages, Disadvantages, and Alternative Solutions Endotracheal intubation with barrier protection A Rapidly Deployable Negative Pressure Enclosure for Aerosol-Generating Medical Procedures. medRxiv Demonstration and mitigation of aerosol and particle dispersion during mastoidectomy relevant to the COVID-19 era Easy and Accessible Protection against Aerosol Contagion during Airway Management Looking beyond tracheal intubation: addition of negative airflow to a physical barrier prevents the spread of airborne particles Airtight, flexible, disposable barrier for extubation Evaluating intubation boxes for airway management Protecting staff and patients during airway management in the COVID-19 pandemic: are intubation boxes safe? Thinking outside the Box: A Low-cost and Pragmatic Alternative to Aerosol Boxes for Endotracheal Intubation of COVID-19 Patients. Anesthesiology Minimising COVID-19 exposure during tracheal intubation by using a transparent plastic box: a randomised prospective simulation study. Anaesth Crit Care Pain Med An affordable videolaryngoscope for use during the COVID-19 pandemic Evaluation of a foldable barrier enclosure for intubation and extubation procedures adaptable for patients with COVID-19: a mannequin study A Novel Negative Pressure Isolation Device for Aerosol Transmissible COVID-19. Anesthesia Analgesia Opening the discussion on a closed intubation box How a simple plastic box could protect health-care workers across Canada from COVID-19 The MacGyver bias and attraction of homemade devices in healthcare