key: cord-0864139-j4c4qx7b authors: Patout, Maxime; Fresnel, Emeline; Lujan, Manuel; Rabec, Claudio; Carlucci, Annalisa; Razakamanantsoa, Léa; Kerfourn, Adrien; Nunes, Hilario; Tandjaoui-Lambiotte, Yacine; Cuvelier, Antoine; Muir, Jean-François; Lalmoda, Cristina; Langevin, Bruno; Sayas, Javier; Gonzalez-Bermejo, Jesus; Janssens, Jean-Paul title: Recommended approaches to minimize aerosol dispersion of SARS-CoV2 during noninvasive ventilatory support can deteriorate ventilator performances: a benchmark comparative study date: 2021-03-02 journal: Chest DOI: 10.1016/j.chest.2021.02.047 sha: 2ae68433f87bbe0d88d8c48378f7ebd4cb3e11be doc_id: 864139 cord_uid: j4c4qx7b Background SARS-CoV-2 aerosolization during noninvasive positive pressure ventilation may endanger healthcare professionals. Various circuit setups have been described in order to reduce virus aerosolization. However, these setups may alter ventilator performances. Research question What are the consequences of the different suggested circuit setups on ventilator’s efficacy during continuous positive airway pressure (CPAP) and noninvasive ventilation (NIV)? Study Design and Method Eight circuit setups were evaluated on a bench made of a 3-D printed head and an artificial lung. Setups were a dual-limb circuit with an oro-nasal mask, a dual-limb circuit with a helmet interface, a single-limb circuit with a passive exhalation valve, three single-limb circuits with custom-made additional leaks and two single-limb circuits with active exhalation valves. All setups were evaluated during NIV and CPAP. The following variables were recorded: the inspiratory flow preceding trigger of the ventilator, the inspiratory effort required to trigger the ventilator, the triggering delay, the maximal inspiratory pressure delivered by the ventilator, the tidal volume (Vt) generated to the artificial lung, the total work of breathing (WOB) and the pressure time product to trigger the ventilator (PTPt). Results With NIV, the type of circuit setup had a significant impact on inspiratory flow preceding the trigger of the ventilator (p<0.0001), the inspiratory effort required to trigger the ventilator (p<0.0001), the triggering delay (p<0.0001); the maximal inspiratory pressure (p<0.0001), the Vt (p:0.0008), the WOB (p<0.0001), the PTPt (p<0.0001). Similar differences and consequences were seen with CPAP as well as with the addition of bacterial filters. Best performance was achieved using a dual limb circuit with an oro-nasal mask. Worst performance was achieved using a dual-limb circuit with a helmet interface. Interpretation Ventilator performance is significantly impacted by the circuit setup. The use of dual-limb circuit with oro-nasal masks should be used preferentially. expiratory pressure (PEEP) as described 13 . We used three indicators to characterize pressurization. The indicators were: delivered inspiratory pressure (cmH 2 O) defined as the peak pressure reached 168 during the inspiratory pressurization phase, tidal volume (ml) defined as the difference between the For each setup, 135 respiratory cycles were analyzed with NIV, and 150 with CPAP. Each experiment 212 was conducted with 3 different filter configurations: no filter, low pressure filter or HME filter. In total, 213 2,430 respiratory cycles were analyzed: 810 (33%) with a normal compliance and resistance profile, 214 810 (33%) with a low compliance and normal resistance profile and 810 (33%) with a normal 215 compliance and increased resistance profile. With NIV, the addition of a low pressure and a HME filter had a significant impact on ventilator 218 performance. The addition of a filter (low pressure or HME) was associated with an increase in flow pressure and HME filters except for PTPt which was higher using an HME filter (p=0.0136) (Table 1 224 and Figure 4 ). Similar results were seen with CPAP: albeit for inspiratory flow preceding trigger, all 225 parameters were significantly influenced by the addition of filters. Except for tidal volume, HME filters 226 performed significantly worse than low pressure filters. (Table 1 and Figure 5 ). With NIV, the type of circuit setup had a significant impact on in flow preceding triggering (p<0.0001), inspiratory effort to trigger the ventilator (p<0.0001), triggering delay (p<0.0001); maximal inspiratory 230 pressure (p<0.0001), V t (p:0.0008), WOB (p<0.0001), PTPt (p<0.0001) and sPVA (p<0.0001) ( Table 2 231 and Figure 6 ). Type of sPVA varied significantly between circuit setups (p<0.0001) (e- With the use of single-limb circuits, we did not assess CO 2 re-breathing. However, CO 2 re-breathing is proportional to the dead-space volume between the patient and the exhalation port. Therefore, the 279 exhalation port is usually placed as close as possible to the mask. Given the SARS-CoV-2 pandemic 280 and risk of droplet aerosolization, it has been suggested to connect the filter directly to the mask. Such 281 a strategy increases the dead-space volume. In setups 4 and 6, the filter was placed after the leak. This may therefore limit CO 2 re-breathing with limited droplet aerosolization. Given the lack of available ventilators during the critical phase of the pandemic, the use of a Boussignac CPAP has been suggested as an alternative. In our study, the Boussignac CPAP 286 achieved lower pressures, and a lower tidal volume for a higher patient inspiratory effort than 287 ventilator-based CPAP. Because of technical limitations, we were unable to increase the flow above 288 30L/min: a higher flow may have helped to achieve a similar pressure than ventilator-based CPAP. The level of intentional leakage of each circuit setup may be different: this could have had an impact 291 on ventilator performance. However, in setups 2 and 3, the level of leaks was identical but, on NIV, 292 setup 3 performed better than setup 2. Hence, we hypothesize that the resistance added on the circuit 293 by the second filter is one of the main drivers of the differences seen. Our results suggest that the use of low-pressure filters had a less deleterious impact than that of HME 296 filters. However, these results need to be interpreted with caution as we ran our tests for a limited 297 period of time and without the impact of humidification coming from air exhaled by the patient. Therefore, in a non-simulated environment, humidity may increase more rapidly the resistance of low-pressure filters than that of HME filters. This may lead to an increase in the work of breathing as well 300 as a decrease in the delivered pressure. There are a few limitations of our study. Firtsly, we did not assess aerosol dispersion. This J o u r n a l P r e -p r o o f The onset of pressure support allows to measure the Triggering delay, the Flow to trigger and the The authors would like to thank Asten Santé for providing the ventilator, the filters, the masks and the 432 circuits used in this trial. The authors would also like to thank Alexandre Nicol and Phoenix effect for prototyping the 3D model 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 254 255 With the use of single-limb circuits, we did not assess CO 2 re-breathing. However, CO 2 re-breathing is proportional 256 to the dead-space volume between the patient and the exhalation port. Therefore, the exhalation port is usually 257 placed as close as possible to the mask. Given the SARS-CoV-2 pandemic and risk of droplet aerosolization, it 258 has been suggested to connect the filter directly to the mask. Such a strategy increases the dead-space volume. In 259 setups 4 and 6, the filter was placed after the leak. This may therefore limit CO 2 re-breathing with limited droplet 260 aerosolization. 282 Therefore, in addition to the use of dedicated circuit setups, clinicians should carefully choose their CPAP/NIV 283 interface whilst initiating patients with SARS-CoV2 infection on a ventilator. Secondly, this is a bench-model 284 study. We identified significant differences between setups, but we were not able to assess their clinical relevance. 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 J o u r n a l P r e -p r o o f 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 ScholarOne -http://mchelp.manuscriptcentral.com/gethelpnow/index.html -(434) 964-4100 CHEST 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 The dead space between the mask and the exhalation valve was*: - 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 Proposal for a systematic analysis of 365 polygraphy or polysomnography for identifying and scoring abnormal events occurring during 366 non-invasive ventilation Helmet with specific settings versus 142 rise time: 100ms, trigger sensitivity: high, cycling: 50% of the peak inspiratory flow. With CPAP, the expiratory 143 pressure was set at 8 cmH 2 O. These settings were chosen according to the clinical experience of the authors in the 144 management of patients with severe SARS-CoV-2 infection Measurements were performed using the flow and pressure curves provided by the artificial lung The indicators were: triggering delay (ms) which measured the time lag between 150 the beginning of simulated effort and the onset of pressure support, flow to trigger (L/min) defined as the value of 151 patient flow measured at the onset of support pressure, pressure to trigger (cmH 2 O) defined as the value of 152 muscular pressure P mus measured at the onset of support pressure and inspiratory pressure-time product (PTPt) 153 (cmH 2 O.s) defined as the area under the pressure-time curve between the onset of inspiratory effort and the return 154 to the set positive end-expiratory pressure (PEEP) as described 13 . We used three indicators to characterize 155 pressurization. The indicators were: delivered inspiratory pressure (cmH 2 O) defined as the peak pressure reached 156 during the inspiratory pressurization phase, tidal volume (ml) defined as the difference between the maximal 157 volume delivered within the current cycle to the mechanical lung and the residual volume and total work of 158 breathing of the system (WOB) (mJ) defined as the sum of the patient WOB (integral of muscular pressure x flow 159 product preceding the onset of ventilatory support) and ventilator WOB We distinguished 164 rate asynchronies from intracycle asynchronies. Rate asynchronies were defined as a mismatch between ventilator 165 and patients' rates. We identified: ineffective efforts when an inspiratory effort was not assisted by the ventilator 166 (ie. a drop of airway pressure associated with an increase or decrease of airflow occurring during expiratory or 167 inspiratory phase respectively), double triggering when two mechanical cycles were triggered by the patient, 168 separated by a very short expiratory time (< 30% of mean inspiratory time) and auto-triggering when mechanical 169 cycles were unrelated to patient's spontaneous breathing. Rate intracycle asynchronies were defined as a distortion 170 of the flow and pressure curves during inspiration and/or expiration. We identified premature cycling when the 193 Results 194 For each setup, 135 respiratory cycles were analyzed with NIV, and 150 with CPAP. Each experiment was 195 conducted with 3 different filter configurations: no filter, low pressure filter or HME filter With NIV, the addition of a low pressure and a HME filter had a significant impact on ventilator performance The addition of a filter (low pressure or HME) was associated with an increase in flow preceding triggering 202 (p=0.0423), inspiratory effort to trigger the ventilator (p<0.0001), triggering delay (p<0.0001), WOB (p<0.0001) 203 and PTPt (p<0.001) and a decrease in maximal inspiratory pressure (p<0.0001), and Vt (p<0.0001). Percentage of 204 sPVA was the only parameter for which adding a filter did not cause a significant difference Similar results were seen with CPAP: albeit for inspiratory flow preceding 207 trigger, all parameters were significantly influenced by the addition of filters With NIV, the type of circuit setup had a significant impact on in flow preceding triggering (p<0.0001) Setup 5 using dual limb circuit was the best 214 setup as flow preceding triggering, inspiratory effort to trigger the ventilator, triggering delay, and PTPt were 215 lower than in other setups with similar V t delivered pressure and sPVA. Setup 8 using a helmet interface had the 216 poorest performance regarding triggering delay, PTPt, inspiratory effort to trigger the ventilator and sPVA. Similar 217 results were obtained with CPAP (Table 2 and Figure 7). With CPAP, the use of setup 8 (helmet) was associated 218 to a significant delay in pressurization and the use of setup 9 (Boussignac CPAP) was associated with higher PTPt 219 values Expiratory Limb Protection: An Experimental Bench Study Covid-19 Does Not Lead to a 332 Performance characteristics of 10 334 home mechanical ventilators in pressure-support mode: a comparative bench study Proposal for a systematic analysis of polygraphy or 337 polysomnography for identifying and scoring abnormal events occurring during non-invasive ventilation Framework for patient-ventilator asynchrony during 340 long-term non-invasive ventilation Helmet with specific settings versus facemask 342 for noninvasive ventilation Oronasal mask versus helmet in acute hypercapnic respiratory 344 failure Prone and Lateral Positioning in Spontaneously Breathing Patients The authors would also like to thank Alexandre Nicol and Phoenix effect for prototyping the 3D model used in EF: conception, acquisition, analysis, interpretation, drafting the work, revising critically 408 LR: acquisition, interpretation ML: conception, interpretation 410 and Competing Interests statement: 413 MP reports personal fees from Resmed, Philips Respironics, grants and non-financial support from In the last 2 years JGB received fees from Philips, Resmed, Breas, Lowenstein and Air Liquide for 418 expertise, and a grant from BREAS for a trial, outside the submitted work JS reports advisory and teaching payments from ResMed, Philips, Chiesi and Menarini outside the 420 submitted work In the last two years, CR received fees from Philips, ResMed, Lowenstein and Fischer & Paykel for 422 expertise, outside the submitted work