key: cord-0863604-awsuugw7 authors: Lowe, Christopher F.; Matic, Nancy; Ritchie, Gordon; Lawson, Tanya; Stefanovic, Aleksandra; Champagne, Sylvie; Leung, Victor; Romney, Marc G. title: Detection of low levels of SARS-CoV-2 RNA from nasopharyngeal swabs using three commercial molecular assays date: 2020-04-28 journal: J Clin Virol DOI: 10.1016/j.jcv.2020.104387 sha: e0b43c0c111fe3a673cb582f9b18044641283235 doc_id: 863604 cord_uid: awsuugw7 Abstract In response to the COVID-19 pandemic, commercial molecular assays for SARS-CoV-2 testing have been rapidly developed and broadly deployed in laboratories worldwide. Although these assays have been reported to correlate well, we sought to compare the Xpert® Xpress SARS-CoV-2 to the cobas® SARS-CoV-2 or the Lightmix® Modular SARS and Wuhan CoV E-gene assay for nasopharyngeal (NP) swabs with low levels of SARS-CoV-2 RNA. Thirty-seven NP swabs were studied, including 10 samples with a moderate cycle threshold (Ct) between 30-33.9, and 22 with Ct≥34, and 5 negative for SARS-CoV-2. Overall concordance on initial comparison was 86.5% (32/37), which was 100% concordance for samples with Ct values ranging between 30-33.9. Discordance amongst samples showing a Ct ≥34 was 22.7% (5/22). Endpoint value analysis on the Xpress SARS-CoV-2 within the discordant samples noted two with an endpoint value >5, which were detected by the cobas® or Lightmix®. Testing of SARS-CoV-2 on the three commercial assays was comparable for NP swabs with moderate Ct values, while high Ct values were less concordant. Importantly, analysis of Xpert® endpoint values improved interpretation of discrepant results. The global COVID-19 pandemic caused by SARS-CoV-2 has challenged healthcare systems throughout the world. Timely diagnosis of COVID-19 cases and subsequent infection control and public health containment measures are essential for healthcare facilities and the community. One of the World Health Organization's key tenets to control the pandemic is to build and maintain public health capacity, including detection of symptomatic cases, testing of suspect cases within 24 hours of specimen collection, self-isolation of confirmed cases and contact tracing for confirmed COVID-19 cases [1] . Laboratory testing for SARS-CoV-2 is a crucial component of this process, and as an emerging virus, there has been an ongoing evolution of diagnostic molecular testing including an array of laboratory developed tests and commercial high-volume and automated throughput, as well as the ability to process specimens other than nasopharyngeal (NP) swabs using the Lightmix® Kit. On average, the expected laboratory testing time on these platforms is approximately 3.5 hours. A rapid molecular assay, which provides a result in ~45 minutes, for SARS-CoV-2, Xpert® Xpress SARS-CoV-2 (Cepheid, Sunnyvale, CA) also recently received FDA EUA. Comparison of the cobas® and Xpert® showed high overall concordance (>99%) [3] . In a public health crisis, rapid molecular assays can have a considerable impact on the ability to make immediate decisions regarding, for example, the isolation of non-adherent patients or the assessment of infection risk to healthcare workers performing invasive procedures on critically ill patients. As the ability of molecular assays to detect SARS-CoV-2 infection can be limited by Of the thirty-seven samples, ten samples had a Ct value between 30-33.9, twenty-two with Ct values ≥34, and five samples were negative for SARS-CoV-2. Overall concordance on initial comparison was 86.5% (32/37). There was 100% concordance for samples with Ct values between 30-33.9. Results comparing the Xpert® to the cobas® and Lightmix® are shown in Table 1 and 2, respectively. Among the samples with a Ct value ≥34, thirteen were initially detected by the Lightmix® assay and nine by the cobas® assay. Discordance within this subgroup was 23% (5/22). For the five discordant results, two samples had endpoint values >5, with one positive by the alternate assay (Lightmix®). Of note, sample 1 (Table 1 ) was a clinical follow up SARS-CoV-2 test. This J o u r n a l P r e -p r o o f patient was positive for SARS-CoV-2 twenty-two days prior using the Lightmix® assay (Ct value for E=28.82), suggesting that the follow-up test represented viral clearance. There was complete concordance of the Xpert® Xpress SARS-CoV-2 with the cobas® SARS-CoV-2 or the Lightmix® assay for samples with a Ct value <34 (100%). Compared to previously published data indicating >99% concordance, overall agreement was lower in our study as we aimed to assess challenging diagnostic samples with low levels of viral RNA [3, 4] . Interestingly, on review of the Xpert® results for the discordant signals, endpoint values >5 samples were detected in 40% (2/5) of the samples. Based on our previous study on endpoint values with Xpert® Flu/RSV, our laboratory would test negative samples with endpoint values >5 with a second assay due to the concern for a low level viral load [5] . Including those two samples with endpoint values >5 (Sample 1 and 2, Table 1 ), concordance between the assays would have been 92%. Importantly, in the absence of endpoint value review and further testing, a discordance rate of 23% may result in laboratories that rely exclusively on the Xpert® assay to report negative results in samples harboring low levels of viral RNA. With expansion of testing to identify pre-symptomatic patients, the number of specimens falling into this low level viral RNA range may be expected to increase, and repeat testing of negative Xpert® results may be indicated with a secondary method. There are clinical, infection prevention and control, and public health implications for the detection of samples with low levels of SARS-CoV-2 viral RNA. For patients presenting early, studies on viral dynamics suggest that they may have a low burden of disease [6] . As a potentially missed diagnosis, these patients can serve as ongoing reservoirs for transmission within the community or healthcare facility [7] . For diagnosed COVID-19 patients, in situations where a test-based strategy is preferred for discontinuation of infection control precautions, identification of these patients with low viral loads is essential to avoid premature discontinuation [8] . Although detection of viral RNA by PCR may not correlate with live transmissible virus [6] , a conservative approach to discontinuing transmission-based precautions may be indicated, given this strategy is primarily used for high-risk populations such as inpatients, severely immunocompromised patients, or long-term care facility residents. There are limitations to this study, including the small sample size which was due to the intended aim of low positive SARS-CoV-2 samples. Inherent in testing these low positives are discordant results, such as in Table 1 , which could be the result of sampling bias, specimen storage or assay design/sensitivity. In addition, we were unable to test all samples on the three assays, using the cobas® or Lightmix® assays as the reference Ct in which to test the Xpert®. However, a prospective comparison of 502 nasopharyngeal swabs between the cobas® and Lightmix® reported 99.6% overall agreement and good correlation (r² = 0.96) between Ct [4] . This was for pragmatic reasons, as those two assays were already implemented in our laboratory, and due to limited sample volumes and the high demand for patient testing, our laboratory needed to balance the kits used for verification versus clinical testing. We report on the test characteristics of three commercial assays on specimens with low levels of viral RNA. As laboratories integrate different testing platforms to optimize patient care, such as the cobas® for high-volume testing and the Xpert® for rapid testing, they should consider The authors have no relevant conflicts of interest to disclose. J o u r n a l P r e -p r o o f World Health Organization. COVID-19 strategy update 14 Laboratory diagnosis of emerging human coronavirus infections-the state of the art The Detection of SARS-CoV-2 using the Cepheid Xpert Xpress SARS-CoV-2 and Roche cobas SARS-CoV-2 Clinical evaluation of the cobas SARS-CoV-2 test and a diagnostic platform switch during 48 hours in the midst of the COVID-19 pandemic Relevance of reviewing endpoint analysis for negative results on the Xpert® Xpress Flu/RSV Virological assessment of hospitalized patients with COVID-2019 The authors would like to acknowledge the staff at the SPH Microbiology and Virology Laboratory for their commitment to providing quality testing for our patients.