key: cord-0859863-l039gw49 authors: Yuen, John G.; Marshilok, Amy C.; Benziger, Peter Todd; Yan, Shan; Cello, Jeronimo; Stackhouse, Chavis A.; Kisslinger, Kim; Bock, David C.; Takeuchi, Esther S.; Takeuchi, Kenneth J.; Wang, Lei; Babu, Sruthi; Itzkowitz, Glen; Thanassi, David; Knopf, Daniel A.; Shroyer, Kenneth R. title: Dry heat sterilization as a method to recycle N95 respirator masks: the importance of fit date: 2021-09-15 journal: bioRxiv DOI: 10.1101/2021.09.15.460526 sha: 1f3c779b1154b3341132d7fe599e07bc51f067e7 doc_id: 859863 cord_uid: l039gw49 In times of crisis, including the current COVID-19 pandemic, the supply chain of filtering facepiece respirators, such as N95 respirators, are disrupted. To combat shortages of N95 respirators, many institutions were forced to decontaminate and reuse respirators. While several reports have evaluated the impact on filtration as a measurement of preservation of respirator function after decontamination, the equally important fact of maintaining proper fit to the users’ face has been understudied. In the current study, we demonstrate the complete inactivation of SARS-CoV-2 and preservation of fit test performance of N95 respirators following treatment with dry heat. We apply scanning electron microscopy with energy dispersive X-ray spectroscopy (SEM/EDS), X-ray diffraction (XRD) measurements, Raman spectroscopy, and contact angle measurements to analyze filter material changes as a consequence of different decontamination treatments. We further compared the integrity of the respirator after autoclaving versus dry heat treatment via quantitative fit testing and found that autoclaving, but not dry heat, causes the fit of the respirator onto the users face to fail, thereby rendering the decontaminated respirator unusable. Our findings highlight the importance to account for both efficacy of disinfection and mask fit when reprocessing respirators to for clinical redeployment. 93 spectroscopy, and contact angle measurements to analyze filter material changes as a consequence of 94 different decontamination treatments. We further compared the integrity of the respirator after autoclaving 95 versus dry heat treatment via quantitative fit testing and found that autoclaving, but not dry heat, causes the fit 96 of the respirator onto the users face to fail, thereby rendering the decontaminated respirator unusable. Our 97 findings highlight the importance to account for both efficacy of disinfection and mask fit when reprocessing 98 respirators to for clinical redeployment. N95 respirators were placed in a paper bag and sealed with a piece of heat-stable tape (Fig 1) . The 12 bags were placed onto a metal rack which was loaded into a TPS Gruenberg truck-in oven. Unless otherwise 13 specified, all N95 respirators were treated for four cycles at either 80˚C for 60 minutes or 100˚C for 30 minutes, 14 with at least 10 minutes of cooling time to room temperature in between. Autoclaved respirators were 15 subjected to a single cycle of 121˚C at 15-25 psi for 30 minutes. The SARS-CoV-2 isolate USA-WA1/2020 was obtained from BEI Resources and used for the 22 experiments in this study. VeroE6 cells were obtained from ATCC and used to titer and passage the SARS-23 COV-2 virus. VeroE6 cells were routinely cultured in DMEM containing Glutagro (Corning) and 8% Fetal 24 Bovine Serum (FBS) at 37ºC with 5% CO 2 . All growth and manipulations of the SARS-CoV-2 virus were 25 performed under BSL3 containment conditions. Approximately 5 x 10 5 PFU of SARS-CoV-2 virus in DMEM was spotted in triplicate onto a N95 mask 27 for each condition and the masks were left to dry within the biosafety cabinet for 2 hours at room temperature. 28 N95 masks containing SARS-COV-2 virus were either left at room temperature or treated with dry-heat using a 29 TPS/Tenney T2 series (Tenney Environmental) small dry-heat sterilizer for the indicated time and temperature. 30 As a negative control, N95 masks were treated similarly with DMEM media alone and left at room temperature 31 during heat treatment. 33 Virus recovery and quantification 34 SARS-CoV-2 virus was recovered by cutting each virus-treated spot, including all three mask layers, 35 from each N95 mask and placing each spot in an Eppendorf screw-cap tube containing 1 mL DMEM plus 10 36 units/ml penicillin, 10 µg/ml streptomycin, and 1 µg/ml amphotericin B. Samples were submerged and 37 incubated for 5 minutes at room temperature then rocked gently by hand for 5 minutes to recover virus. Plaque 38 assays were performed to quantify the amount of virus recovered by performing serial dilutions of recovered 39 virus and infecting VeroE6 cells seeded at 4.5x10 5 cells/well in a 6-well tissue culture treated plate for one 40 hour. Cells were then overlaid with DMEM containing 0.8% tragacanth gum, 2.5% FBS, 10 units/ml penicillin, 41 10 µg/ml streptomycin, and 1 µg/ml amphotericin B and incubated at 37ºC with 5% CO 2 for 48 hours. To 42 quantify the amount of virus recovered, the overlay was removed, and the plaques were visualized by staining 43 VeroE6 cells with 0.5% crystal violet and 0.8% glutaraldehyde in 50% methanol for 10 minutes followed by 44 several washes with distilled water. Total PFU/mL for each condition was calculated by averaging the mean 45 PFU/mL recovered for each biological replicate (n=3). A PortaCount Pro 8048 (TSI, Shoreview, MN) was used to obtain OSHA compliant quantitative fit 49 factors at Stony Brook University Hospital, in the Department of Occupational Health and Safety (17). The 50 most penetrating particle size (MPPS) for most N95 FFRs is around 300 nm (18-22). For this reason, the 51 quantitative fit testing protocols evaluate leakages at particle size ranges outside of the MPPS to be more 52 sensitive to leakages across the sealant between mask and face. The principle of operation is to choose a 53 particle size, typically about 40 nm, that is filtered with great efficiency due to electrostatic interactions and to 54 compare the number concentration of those particles outside the mask (i.e. ambient air), the particle 55 concentration inside the mask (23, 24). Detection of those particles is achieved by a condensation particle 56 counters (CPC) that grows these small particles via condensation of vapor to size detectable via light 57 scattering. Quantitative fit testing was performed on the same operator for all mask types and for all conditions: 58 dry heat treated (n = 3), untreated (n = 1), and autoclaved (n = 1) respirators. Quantitative fit testing procedures for N95 respirators were performed according to Occupational Safety 60 and Health Administration (OSHA) guidelines found in Appendix A to §1910.134. N95 respirators were fitted 61 with a mask sampling adapter that allows for the measurement of particles inside the respirator while donned 62 by an individual. Four exercises were performed (Table 1) 78 layers using forceps, where layer 1 is the outside layer (farthest from the mask wear), layer 2 is the middle 79 meltblown layer and layer 3 is the inside layer (closest to the mask wearer). Scanning electron microscopy (SEM) images were collected using a high-resolution SEM (JEOL 81 7600F) instrument. SEM images were acquired at an accelerating voltage of 5 kV. A thin layer of silver (Ag) 10 82 nm in thickness was applied to the mask materials prior to SEM imaging to reduce sample charging. 83 Raman. Raman spectra of the pristine and treated mask materials were recorded on a Horiba Scientific XploRA 85 instrument with a 532 nm laser at 10% intensity using a 50X objective and a grating of 1,200 lines/mm. The 86 spectra were calibrated with a Si standard. The test subject will recite the Rainbow Passage loud enough to be heard by another person in the room. 30 second mask sample. 3) Head Side-to-Side Turn head from side to side for and inhale 2 times at each extreme. 30 second mask sample. Slowly move head up and down for inhale 2 times at each extreme. 30 second mask sample, followed by a 9 second ambient sample. 93 the volume change as function of time was determined using the droplet profile and Kyowa FAMAS software. 14 in failed quantitative fit testing (fit factor < 100) (Fig 3) . In contrast, dry heat incubation yielded passing fit test 15 scores (≥100) for all the respirators that passed quantitative fit testing prior to any treatment. None of the 16 Bacou Willson 801 respirators passed quantitative fit testing, presumably due to poor fit on the user. It is 17 notable, however, that masks which were autoclaved yielded a lower fit factor than either the untreated, 100˚C, 18 or the 80˚C dry heat treatment groups (Fig 3c) . The mesoscale morphologies of 3M 1860 N95 were characterized by SEM before and after heat and 26 autoclave treatment. The cross-section view was taken and application of EDS indicated that only carbon 27 signal was detected at layers-1 and -2, while both carbon and oxygen were detected at layer-3 (Fig 4a and Fig 28 4b) . Layer-1 is ~300 µm in thickness, with millimeter scale patterning, comprised of microfibers with a diameter 29~20 µm (Fig 4c) . Layer-2 is ~300 µm in thickness, comprised of microfibers with a diameter in the range of 1-30 10 μm (Fig 4d) . Layer-3 is ~1 mm in thickness, comprised of microfibers with a diameter ~30 µm, and some 31 defects were also observed (Fig 4e) . The morphologies of the 100˚C dry heat treatment (Fig 4f-h) and 32 autoclave treatment (Fig 4i-k) of 3M 1860 N95 did not show obvious differences from SEM images. Layer 1 showed significant fluorescence, as indicated by the broad peak features, possibly due to the 46 dye used in this layer. Despite the strong fluorescence, a decrease in the ratio of two bands at 810 and 840 cm −1 49 at 1220 cm -1 ascribed to the helical chain of 14 monomeric units of polypropylene suggested shorting of the 50 helical chain length. Layer 2 (middle layer) and layer 3 (inner layer) (Fig 5b and 5c ) also contain polypropylene 51 fibers that are lower in crystallinity and narrower in thickness. X-ray diffraction (XRD) analysis was also performed 52 to identify the composition and crystallinity of each layer before and after dry heat and autoclave treatment (S1 07 Similarly, dry heat sterilization after 4 cycles did not affect their fit as measured by quantitative fit testing. In 08 contrast, autoclaved N95 respirators appeared to have some physical damages in the overall structure and 09 shape of the respirators and failed quantitative fit testing in all respirator types tested. These results after 10 autoclaving are consistent with other reports on most N95 respirator types that show degradation of the 11 respirators (7, 13, 32 27 failed the mask fit test and suggested that the change in filtration efficiency is up to 7.73%. However, the mask 28 material analysis does not corroborate significant changes in the filtration material that could lead to these 29 decreases in filtration efficiency. This points to the fact, that most likely, autoclaving the respirators led to changes 30 in the mask fit by either altering the face mask mold, sealant, and/or straps. Taken together, these data suggest 31 that autoclaving indeed has an impact on the fit of the respirator material. In combination with previous studies 32 on the effects of autoclaving N95 respirators (13, 32) and their deleterious impact on both fit and filtration, the 33 results confirm that autoclaving is not a consistently viable method for the decontamination of N95 respirators 34 for their reuse. Although we assessed the function of decontaminated masks by quantitative fit testing and material 36 characterization, our study does not directly distinguish whether failed fit testing is due to the impairment of the 37 filtration efficiency, including any impact on the electret properties of N95 respirators (16), or due to the failure of 38 fit or some combination of both fit and filtration, although our data suggests that impact on fit as the most likely 39 cause. In summary, dry heat sterilization is a potentially scalable, accessible, and effective method of Aerodynamic analysis of SARS-CoV-2 in two Wuhan 49 hospitals Droplets and Aerosols in the Transmission of SARS-CoV-2 Critical supply shortages-the need for ventilators and personal 53 protective equipment during the Covid-19 pandemic Contributing factors to personal protective equipment shortages 55 during the COVID-19 pandemic Control CfD, Prevention. Implementing filtering facepiece respirator (FFR) reuse, including reuse after 57 decontamination, when there are known shortages of N95 respirators Low risk of SARS-CoV-2 transmission by 59 fomites in real-life conditions. The Lancet Infectious Diseases Decontamination of N95 masks for re-use 61 employing 7 widely available sterilization methods A method to determine the available UV-C dose for the decontamination of 66 filtering facepiece respirators the United States -Considerations Can N95 respirators be reused after disinfection? 70 How many times? ACS nano Stability of SARS-CoV-2 in different environmental 72 conditions Effectiveness of 74 N95 respirator decontamination and reuse against SARS-CoV-2 virus. Emerging infectious diseases Decontamination of SARS-CoV-2 and other 77 RNA viruses from N95 level meltblown polypropylene fabric using heat under different humidities Recommendations for Sponsors Requesting EUAs for Decontamination and Bioburden Reduction Systems for Surgical Masks and Respirators During the Coronavirus Disease KN95 and N95 Respirators Retain Filtration 83 Efficiency despite a Loss of Dipole Charge during Decontamination Appendix A to § 1910.134: fit testing procedures (mandatory) Aerosol penetration through filtering facepieces and respirator 87 cartridges Effect of particle size on the performance of an N95 filtering Electrostatic respirator filter media: filter efficiency and most penetrating 92 particle size effects Do N95 respirators provide 94 95% protection level against airborne viruses, and how adequate are surgical masks? American journal of 95 infection control Performance of N95 respirators: filtration 97 efficiency for airborne microbial and inert particles Fit testing using size-selected aerosol Estimation of crystallinity of isotactic polypropylene using Raman 04 spectroscopy Isotactic polypropylene morphology-Raman spectra correlations Estimation of crystallinity of isotactic polypropylene using Raman 08 spectroscopy Investigation of the Molecular Mechanisms of Melting and Crystallization 10 of Isotactic Polypropylene by in Situ Raman Spectroscopy Contact Angle and Wetting Properties Artificial Saliva: Challenges and 13 Future Perspectives for the Treatment of Xerostomia Personal protective equipment shortages during COVID-19-15 supply chain-related causes and mitigation strategies Autoclave sterilization and ethanol treatment of re-used 18 surgical masks and N95 respirators during COVID-19: impact on their performance and integrity In-Line Monitoring the Degradation of Polypropylene under Compared with the 28 XRD patterns of the pristine masks (black), after dry air treatment (red) and autoclave/steam treatment (blue), 29 the 3M 1860 N95 mask materials showed no compositional changes and insignificant crystallite sizes changes 30 after both types of thermal treatment. Specifically, the respective crystallite sizes of the pristine, dry air treated 31 and steam treated layer are 15, 17 and 14 nm for layer 1 (Figure S1a) This indicated that dry air treatment slightly increased the crystallize size at layer 1 with no significant 33 change at layer 2. Interestingly, steam treatment decreased crystallite size for both layers 1 and 2 for the 3M in comparison to that of the pristine, which evince an initial contact angle of 103.9°±7.7°, 105.4°±6.2°, and 41 96.0°±15.2°, respectively. The treated samples' contact angle values remain consistent over time, whereas the for the three samples. This observation suggests that the wettability of the pristine sample increases over 44 time, but this behavior is ameliorated by the dry heat and steam treatments and after dry heat and steam heat treatment. The XRD patterns of layer 1 63 and layer 2 indicated a semicrystalline character, and as marked therein, the major diffraction patterns were 64 indexed to reflections from Compared with the XRD 70 patterns of pristine (untreated) Bacou Willson 801 N95 (black curve), the XRD patterns after dry air treatment 71 (red) and steam treatment (blue) indicate higher crystallinity for layers 1 and 2 (S5-b). Specifically, the crystallite 72 sizes of the respective pristine, dry heat treated, and steam treated 16, 18 and 19 nm for layer 1 (S5a) and 4, 8 layer 2, and the dry air treatment and steam treatment has more After the dry air and steam 82 treatment, the ratio of the two bands at 810 and 840 cm -1 layer 1 decreased, along with the decreasing intensity 83 of the 972 cm -1 peak (S6a), as highlighted in the yellow dashed regions, suggesting a shorting of the helical 84 chain conformation of polypropylene after heat treatment (28) (34) as indicated by the strong C=C stretching 90 band (ring deformation) at 1615 cm -1 and C=O stretching band at 1730 cm -1 . Similarly, no significant differences dry heat and steam treatment Whilst there are no significant initial differences between 05 the pristine and heat-treated samples, 106.9°±6.8° and 111.9°±15.1°, respectively, both samples show a 06 decreasing trend over time. The most rapid decrease is observed within the pristine sample