key: cord-0849850-8a702clp authors: Cai, C.; Floyd, E. L.; Aithinne, K. A.; Oni, T. title: Is the Current N95 Respirator Filtration Efficiency Test Sufficient for Evaluating Protection Against Submicrometer Particles Containing SARS-CoV-2? date: 2020-05-18 journal: nan DOI: 10.1101/2020.05.14.20102327 sha: b4bb61d4e687f80423c51634116bec58c44712e6 doc_id: 849850 cord_uid: 8a702clp Previous researchers measured the Overall Filtration Efficiency using sodium chloride testing aerosol with aerosol photometer to analyze the reuse of respirators after sterilization. However, this study showed that the Overall Filtration Efficiency would significantly overestimate the respirators' ability to protect against submicrometer particles containing SARS-CoV-2. For example, after three cycle H2O2 vaporous sterilizations, the most penetrating particle size for the N95 level respirators was from 250 nm to 500 nm, which carried the peak concentration of the SARS-CoV-2 in hospitals. The Size Specific Filtration Efficiency (250 nm to 500 nm) dropped to 56%{+/-}4%, however, the Overall Filtration Efficiency was still 86%{+/-}5%. In order to protect health care personnel adequately, we recommend measuring the Size Specific Filtration Efficiency to evaluate the reuse of respirators. Filtration Efficiency to evaluate the reuse of respirators. CC-BY-NC 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted May 18, 2020. . https://doi.org/10.1101/2020.05.14.20102327 doi: medRxiv preprint Introduction 28 Recent studies indicates that aerosol transmission of the severe acute respiratory 29 syndrome coronavirus 2 (SARS-CoV-2) is plausible since the virus can remain viable and 30 infectious in aerosol form for hours (1, 2) . The actual SARS-CoV-2 size ranges from 60 nm -31 150 nm (3,4,5). The peak concentration of the SARS-CoV-2 in hospitals was found in 250 32 nm -500 nm particles (6). Due to the COVID-19 pandemic and shortage of mask supplies 33 (7), some studies analyzed the reuse of N95/KN95 masks after sterilization by measuring the 34 filtration efficiency using sodium chloride (NaCl) test aerosol (8, 9) . The N95 grade mask 35 should have a minimum filtration efficiency of 95% for 300 nm NaCl aerosols. The light-36 scattering instruments (e.g., aerosol photometer), which are commonly used for filtration 37 efficiency test, are usually limited to measuring aerosol size from 100 nm to 10 µm. However, the count median diameter of the NaCl test aerosol is 75±20 nm with the geometric from 100 nm -250 nm, which is not in the particle size range of containing the peak 43 concentration of the SARS-CoV-2 in hospitals. Therefore, a mask efficient for particles ≤ 44 250 nm may not capture most SARS-CoV-2, and the size specific filtration efficiency is a 45 . CC-BY-NC 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) The copyright holder for this preprint this version posted May 18, 2020. . https://doi.org/10.1101/2020.05.14.20102327 doi: medRxiv preprint 3 better assessment of the mask's protection ability rather than the overall filtration efficiency 46 to prevent the SARS-CoV-2 transmit via airborne particles. The goal of this demonstration 47 study was to test the effects of multiple sterilization cycles on the size specific filtration 48 efficiency compared to the overall filtration efficiency of KN95 and N95 masks. The effects of multi-cycle sterilization on the filtration efficiency by aerosol size are 66 . CC-BY-NC 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted May 18, 2020. . https://doi.org/10.1101/2020.05.14.20102327 doi: medRxiv preprint 4 summarized in the Figure 1 . After three H2O2 sterilization cycles, the filtration efficiency for 67 250 nm -500 nm particles dropped from 84%±2% to 68%±3% for KN95, and from 97%±2% 68 to 56%±4% for N95. The differences between two and three cycles were significant for both 69 KN95 (p-value = 0.028±0.037) and N95 (p-value = 0.0009±0.0008). After UV sterilization, CC-BY-NC 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted May 18, 2020. is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted May 18, 2020. This study proposed that, in addition to considering the overall filtration efficiency, 91 the filtration efficiencies for particle sizes similar to infectious particles should be considered. This study also found that the two different multi-cycle sterilization processes have unique is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted May 18, 2020. . https://doi.org/10.1101/2020.05.14.20102327 doi: medRxiv preprint 7 the American College of Surgeons for H2O2 sterilization. The filter material might degrade 99 further with more cycles, which should be investigated for UV treatment. Response pilot grant, and by the student volunteers that assisted with data collection. 107 108 . CC-BY-NC 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted May 18, 2020. . https://doi.org/10.1101/2020.05.14.20102327 doi: medRxiv preprint Table 1 . Comparison of overall filtration efficiency and size specific (250 nm -500 nm) 143 filtration efficiency after multi-cycle H2O2 treatment. . CC-BY-NC 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review) The copyright holder for this preprint this version posted May 18, 2020. . https://doi.org/10.1101/2020.05.14.20102327 doi: medRxiv preprint Aerosol and surface stability of SARS-CoV-2 as compared with SARS-111 The coronavirus pandemic and 113 aerosols: Does COVID-19 transmit via expiratory particles? Aerosol Science and 114 Technology Coronavirus Isolated from a Patient in Korea with COVID-19. Osong Public Health 117 and Research Perspectives Features, evaluation 119 and treatment coronavirus (COVID-19) Enhanced 122 isolation of SARS-CoV-2 by TMPRSS2-expressing cells Aerodynamic analysis of 125 SARS-CoV-2 in two Wuhan hospitals Critical Supply Shortages -The Need for 128 Ventilators and Personal Protective Equipment during the Covid-19 Pandemic It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. (which was not certified by peer review)The copyright holder for this preprint this version posted May 18, 2020. CC-BY-NC 4.0 International license It is made available under a is the author/funder, who has granted medRxiv a license to display the preprint in perpetuity. The copyright holder for this preprint this version posted May 18, 2020. . https://doi.org/10.1101/2020.05.14.20102327 doi: medRxiv preprint