key: cord-0848944-af51b6ji authors: Gershman, Sophia; Harreguy, Maria B.; Yatom, Shurik; Raitses, Yevgeny; Efthimion, Phillip; Haspel, Gal title: A low power flexible dielectric barrier discharge disinfects surfaces and improves the action of hydrogen peroxide date: 2021-02-25 journal: Sci Rep DOI: 10.1038/s41598-021-84086-z sha: b180a472d22d25dbb0fb619e4c0b7ee25165b22e doc_id: 848944 cord_uid: af51b6ji There is an urgent need for disinfection and sterilization devices accessible to the public that can be fulfilled by innovative strategies for using cold atmospheric pressure plasmas. Here, we demonstrate a successful novel combination of a flexible printed circuit design of a dielectric barrier discharge (flex-DBD) with an environmentally safe chemical reagent for surface decontamination from bacterial contaminants. Flex-DBD operates in ambient air, atmospheric pressure, and room temperature without any additional gas flow at a power density not exceeding 0.5 W/cm(2). The flex-DBD activation of a 3% hydrogen peroxide solution results in the reduction in the bacterial load of a surface contaminant of > 6log(10) in 90 s, about 3log(10) and 2log(10) better than hydrogen peroxide alone or the flex-DBD alone, respectively, for the same treatment time. We propose that the synergy between plasma and hydrogen peroxide is based on the combined action of plasma-generated OH(·) radicals in the hydrogen peroxide solution and the reactive nitrogen species supplied by the plasma effluent. A scavenger method verified a significant increase in OH(·) concentration due to plasma treatment. Novel in-situ FTIR absorption spectra show the presence of O(3), NO(2), N(2)O, and other nitrogen species. Ozone dissolving in the H(2)O(2) solution can effectively generate OH(·) through a peroxone process. The addition of the reactive nitrogen species increases the disinfection efficiency of the hydroxyl radicals and other oxygen species. Hence, plasma activation of a low concentration hydrogen peroxide solution, using a hand-held flexible DBD device results in a dramatic improvement in disinfection. | (2021) 11:4626 | https://doi.org/10.1038/s41598-021-84086-z www.nature.com/scientificreports/ aiding the transport of RNS/ROS into the cell. Inside the cell the ROS/RNS damage proteins, lipids, and the DNA. The combined effect of these processes is bacterial cell inactivation 8, 14 . Most of the work on medical and biological applications of DBDs has been conducted on one of three configurations, a floating electrode configuration, a plasma jet (floating electrode or two electrode), and a less common surface DBD 8, [15] [16] [17] [18] [19] [20] [21] [22] . In a floating electrode device, the high voltage electrode is encased in a dielectric material and the treated surface acts as a ground electrode 8, 15, 16 ; the treated surface is exposed to high electric fields and fluxes of charged particles. The most extensively studied is the plasma jet, which uses power from pulsed dc to microwave range and where plasma effluent is carried by a gas flow to the treated surface. The plasma effluent is suitable for medical applications but requires a compressed gas supply 15, [17] [18] [19] . Surface DBD has been primarily studied as an actuator for flow control in aeronautics applications and for large area surface modifications (ex. 20 ). Introduction of devices based on flexible printed circuit design facilitated its applications in medical and biological fields 5, [21] [22] [23] [24] . The device studied here is also based on a flexible printed circuit design. Atmospheric pressure plasmas have been shown to be effective for the decontamination of surfaces from bacteria and viruses, but the level and the rate of inactivation strongly depend on the biological species, experimental conditions, and the plasma source. For example, D-value (time for 1log10 reduction) is 225 s for the exposure to the gases produced by one DBD 25 , 150 s for another DBD 2 , 35 s for E. coli exposed to an atmospheric pressure helium/air glow discharge 3 , and 15 s for a paper-DBD 22 . The fast reduction D = 15 s was achieved by a single-use flexible DBD device using a printed patterned electrode on a paper substrate and operated at 2 kHz, 3.5 kV AC, 10 W. This is a disposable device 22 . This diversity of results and conditions makes it necessary for us to test the new device design presented here and quantify the disinfection of standard microorganisms. Another variation on plasma disinfection is the use of low-pressure plasma activated hydrogen peroxide vapor 14, 26, 27 . Systems such as the low temperature sterilization systems by Sterlis Healthcare 9 are widely accepted methods of sterilization of materials susceptible to high temperatures, humidity, and corrosion. In more recent studies, the addition of hydrogen peroxide has been explored to enhance plasma disinfection at atmospheric pressure [28] [29] [30] [31] . Addition of H 2 O 2 droplets into a corona discharge produced 6log 10 reduction, and adding H 2 O 2 vapor to the plasma effluent produced a reduction greater than 6log 10 in the bacterial load and a significant reduction in biofilm and spores 28, 30, 31 . The dominant mechanisms responsible for the enhancement depend on the type of plasma, the state of H 2 O 2 . H 2 O 2 vapor is ionized in a plasma to form H 2 O 2 -, while droplets may be negatively charged, a water solution of H 2 O 2 is subject to the active species introduced by plasma into the solution akin to plasma activated water. Pure water is acidified by plasma enhancing the bactericidal effects, while buffered solutions such as phosphate buffer saline (PBS) maintain the pH level but are affected by the dissolved ozone, nitrates, and hydrogen peroxide radicals [32] [33] [34] . The flex-DBD described here is safe to the touch, but unlike plasma jet devices, it does not need any additional gas supply or sophisticated power sources, and unlike a paper-DBD device it is capable of long-term stable operation. We demonstrate a fast disinfection effect of over 4log 10 in less than 90 s on the standard bacteria, gram-negative Escherichia coli (E. coli). We also demonstrate the synergistic effect of the commonly available antiseptic, 3% H 2 O 2 , and the flex-DBD device, of surface decontamination > 6log 10 in 90 s. The device tested here could be used to disinfect surfaces, personal items, and protection equipment such as masks. Device assembly and electrical diagnostics. The flex-DBD is based on a printed circuit design 5, 22, 24 . It consists of a layer of copper tape (0.127 mm thick, ] 16 mm × 26 mm) serving as a high voltage electrode and covered by a layer of Kapton polyimide tape (Kapton, DuPont), 100 µm thick, ∈ rel ≈ 3.5, and a patterned ground electrode, copper (30 µm thick), electroless nickel immersion gold (ENIG) coated placed on top of the Kapton polyimide tape (Fig. 1a) . The pattern on the ground electrode consists of 200 (10 × 20) square cavities, each 0.75 × 0.75 mm in size. A regulated, 500 V-10 kV, 25-60 kHz, pulsed AC power source (PVM 500 AC, Information Unlimited) was used to generate the discharge (Fig. 1b) . The patterned electrode of the flex-DBD was connected to the powersupply ground, and the copper foil electrode to the power-supply high voltage output transformer (Fig. 1b,c) . We used a Tektronix D (2 GS/s, 250 MHz) oscilloscope to monitor the current, voltage, and charge transfer in the circuit during experiments: current to ground (Pearson Model 2877 Current Monitor, 1 V/A, 2 ns rise time); voltage at the high voltage (copper tape) electrode (Tektronix P6015 HV probe); and charge transferred was determined by measuring the voltage across a 10 nF capacitor (C M , Fig. 1b ,c) connected in series on the ground side of the flex-DBD. We operated the flex-DBD in resonance mode. The parallel connection between the DBD and the secondary coil on the high voltage output transformer has a resonance frequency, ≈ 1 2π √ LC , where L, is the inductance of the secondary (Fig. 1c) and C is the capacitance of the flex-DBD. During operation the capacitance of the flex-DBD is approximately, C ≈ C B + C BC . At the resonance frequency, the overall impedance of the L-C circuit as seen by the power source is at a maximum, therefore minimizing the current drawn and hence the power used by the device (Fig. 1c) and maximizing the voltage applied to the DBD, hence facilitating a discharge at lower power used by the power source. To start the device, we adjusted the frequency to a resonance value at a voltage amplitude below the starting voltage, then increased the applied voltage to the start the discharge (V start = 1.9 kV amplitude), and readjusted the frequency to a new resonance value for the flex-DBD with the plasma on (Fig. 1d) . The voltage was then increased until the entire surface of the flex-DBD appeared to glow to the naked eye (Fig. 1d , 2.8 kV). The resonance operating frequency was 42 ± 2 kHz. The variation in the resonance frequency is likely due to the slight differences in the hand-made devices and the operating conditions. We monitored the temperature of the grounded glowing face of the flex-DBD for several minutes prior to starting experiments to ensure that a steady-state condition was reached, and continued to monitor the temperature during experiments. Voltage Spectroscopy and imaging procedures. A PI-MAX 3 ICCD camera (Teledyne Princeton Instruments) was used for the fast imaging of the surface discharge (Supplementary material, Fig. S1a ). The camera was triggered on a high current spike with minimal delay so that the shutter was open for the next current spike. We monitored the timing of the current spikes recorded by the Pearson coil and camera shutter on the oscilloscope and made the appropriate adjustments in the lengths of the connecting cables so that the images correlated to the current spikes. Optical emission spectra (OES) of the light emitted by the flex-DBD were recorded in the range of 300-650 nm (Supplementary material, Fig. S1b ) with an SpectraPro HRS-750 spectrometer coupled with PI-MAX 3 ICCD camera (both of Teledyne Princeton Instruments). Fourier transform infrared absorption spectroscopy (FTIR AS) was used to assess the presence of more stable gaseous species in the plasma effluent from the flex-DBD. We placed the flex-DBD horizontally inside a test gas cell and turned it on (Supplementary material, Fig. S1c ). The spectra were taken in the range of 700-3500 cm −1 , at a resolution of 2 cm −1 , using a FTIR-6800 with the MCT detector (Jasco). Each measurement of 64 scans took about 1 min. We then turned off the flex-DBD, flushed the instrument chamber and the gas cell with air, and recorded a new background spectrum prior to each measurement. To demonstrate the disinfection ability of the flex-DBD, we tested the effectiveness of the device in reducing the bacterial load of E. coli 10-beta (New England Biolabs) and the standard E. coli AMS 198 (ATCC-11229). Bacteria were cultured following vendor's instructions at 37 °C in Luria broth or Luria broth agar (both from Research Products International). For the surface test experiments, we used a suspension of the E. coli strain (OP-50-GFP, Caenorhabditis Genetics Center) that expresses cytoplasmic green fluorescent protein and forms a uniform bacterial lawn rather than discrete colonies. The flex-DBD was attached to a holder, or in other experiments, to the lid of a 60 mm petri dish (Fig. 2a) . The experiments included the treatment of bacteria seeded in petri dishes, on a disposable textile type material, www.nature.com/scientificreports/ on metal (aluminum), and on glass (microscope cover slips). For treatment of bacterial plates, we spread 50 µl of a fresh bacteria culture on LB-agar plates and treated the plate surface with the flex-DBD attached to a lid, placed over the petri dish ( Fig. 2a) , for different amounts of time. We then incubated the plates overnight at 37 °C and visually compared them to untreated control and examined for areas that were clear of bacteria. To test the disinfection of the textile-type surfaces, 100 µl of E. coli OP-50 was spread on textile-like polyethylene material (Tyvek, DuPont). The petri dish cover with the flex-DBD attached was placed over the inoculated area. The petri dish cover used in this experiment was cut to maintain a 1-2 mm distance between the treated surface and the face of the flex-DBD. Each region was treated for a set amount of time and at the end of the treatment time, immediately stamped with an LB contact plate (Carolina Biological Supply Company). The untreated area was stamped as the control. The contact plates were incubated for 24 h at 37 °C. Qualitative results were assessed visually by observing GFP expression using a gel imaging station (FastGene Blue/Green LED GelPic Box, Nippon Genetics). To quantify disinfection, bacteria (E. coli 10-beta, Standard E. coli AMS 198 or OP50-GFP) were inoculated on glass coverslips (25 mm diameter). Four droplets of 5 µl each (20 µl) of the bacterial culture (starting concentration 10 8 CFU/ml) were placed onto each coverslip and allowed to dry for approximately 40 min. The slides with dry bacterial culture were then placed onto the flex-DBD with the inoculated side directly in contact with the discharge (Fig. 2b) . The coverslips were treated for 10, 30, 90, and 270 s. At the end of the treatment time, the treated coverslip was placed in a centrifuge tube with 7.5 ml LB, enough to cover the coverslip. The tubes were vortexed on a medium setting for 20 s to recover the bacteria from the treated surface but not damage the cell membrane. The resulting bacterial suspension was spread on LB-agar plates and incubated for 24 h. Cultures were then counted and the number used to calculate the logarithmic reductions of bacterial concentration. All disinfection experiments were conducted with the flex-DBD operating at 3 kV, 20% duty cycle, and 40-50 °C. We tested the efficiency of the flex-DBD disinfection in conjunction with a commonly available 3% solution of H 2 O 2 . The discharge in the flex-DBD is suppressed by water so we used a semipermeable polyethylene material (Tyvek, DuPont) to keep the H 2 O 2 solution from the surface of the flex-DBD (Fig. 2c) . Pieces of material were disinfected by soaking for 5 min in a 70% solution of isopropyl alcohol then dried thoroughly for at least 30 min. For each trial, we placed a piece of the sterile material on top of the operating flex-DBD and five 2 µl droplets of 3% H 2 O 2 solution on top of the polyethylene material. We then placed an inoculated glass coverslip on top of the solution with the bacteria in contact with the solution (Fig. 2c) . At the end of each treatment, we www.nature.com/scientificreports/ dropped both the coverslip and the cloth into a centrifuge tube with 7.5 ml of LB solution. The same recovery and plating procedure was used in all experiments. The controls for this experiment were inoculated but untreated coverslips, as well as the same procedure with H 2 O 2 but with the flex-DBD remaining turned off. We also compared the disinfection efficiency of H 2 O 2 aided only by the UV light produced by the flex-DBD. To block all the output from the plasma except light, we placed a thin film filter transparent down to 190 nm between the DBD and H 2 O 2 . Finally, to eliminate the operation temperature as a factor contributing to the disinfection process, we placed inoculated glass coverslips on a heating block at 47 °C and repeated the same disinfection procedures to determine the reduction in the bacterial load. We did not observe any reduction in the number of CFU/ml. To quantify bacterial load reduction, we imaged the treated plates after 24 h of growth and counted the number of bacterial colonies, interpreted as colony forming units (CFUs) in the plateseeding solution. When possible, we used ImageJ (1.53c) software 35 to count CFUs, otherwise we counted visually. The bacterial load in CFU/ml was calculated by multiplying the CFU count by 9375 (follows from 40 µl spread on each plate from coverslips washing volume of 7.5 ml LB broth, and original inoculation volume of 20 µl), and multiplying to account for any serial dilutions. The logarithmic reduction in the bacterial load was calculated as log 10 Chemical tests of the treated solutions. We determined the pH of the LB broth, Phosphate Buffer Saline solution (Sigma-Aldrich), and the 3% hydrogen peroxide solutions before and after the application of the flex-DBD, using the a pH meter (Aspera Instruments, model SX823-B, ± 0.01 pH) and pH test strips (Esee, ± 0.5) for amounts too small for the pH sensor. We We used a scavenger method to assess qualitatively the production of the OH · in solutions during plasma treatment. We used coumarin (> 99%, Sigma-Aldrich) as a scavenger because it reacts with OH · in solution to produce 7-hydroxycoumarin that fluoresces at 460 nm when excited at 390 nm 36 . Coumarin itself does not fluoresce in this spectral range. We prepared a 5 mM solution of coumarin in a 20 mM PBS by dissolving crystalline coumarin in the PBS solution at a pH9 and readjusting the pH of the resulting solution back to 7.4 with HCl 36 . We used three reference solutions: the coumarin stock solution, a solution prepared by adding equal amounts of coumarin solution and PBS, and a solution prepared by adding equal amounts of the coumarin stock solution and 3% hydrogen peroxide. The fluorescence of each solution was recorded at room temperature using an Ocean Optics Fluorescence/Absorption spectrometer. Coumarin/PBS solution and coumarin/hydrogen peroxide solutions were treated with the DBD discharge for 5 min, and the fluorescence of was recorded immediately following the treatment. Surface dielectric barrier discharge in a flex-DBD device. A 40 kHz sinusoidal voltage, amplitude of 1.9-3 kV, was applied to the high voltage electrode for ~ 200 µs (20% duty cycle at 1 kHz repetition rate). A surface dielectric barrier discharge ignites inside the cavities and around the perimeter of the flex-DBD (Fig. 1d) . The discharge propagates along the surface of the dielectric and eventually erodes the substrate of the ground electrode. The erosion pattern observed on post-run devices indicates that the discharge occurs in the center portion of each cavity leaving the angles intact. This is due to the electric field topology in the cavity with the maximum electric field E = V/r ~ 80 kV/cm, at the center of the circular cavity with a given radius (r). The erosion of the substrate of the patterned electrode happens over months of operation, possibly because only a few cells are lit at any time (apparent only at high temporal resolution imaging synchronized with the current spikes, Fig. 3b ). During the discharge, the maximum current can be up to 100 mA for several tens of ns. Water can form a conductive film that prevents charge accumulation on the electrode, preventing breakdown conditions 32, [37] [38] [39] . This effect depends on the voltage rise time and breakdown voltage may be reached at subnanosecond rise times. However, in the flex-DBD, the voltage rise time is ≈7 µs, too slow to prevent the charge leakage from the conductors. The parts of the device that become moist will not light until dry. If that occurs when the power is on, heating due to the resistive and dielectric losses will self-dry the device and it will restart once the moisture evaporates. We used a water resistant material in the experiments with H 2 O 2 solutions, to prevent the flex-DBD becoming wet while allowing active species from the plasma to reach the treated surface. This sensitivity to water is important for bio-related applications. A typical current trace is comprised of a displacement current sinusoidal component of (42 ± 2) kHz and the superimposed sharp spikes 10 -50 ns in duration corresponding to the discharges (Fig. 3a,c) . The displacement current was subtracted from the total measured current to obtain the discharge current (Fig. 3c) . The number of discharges, their overall duration, and their amplitude increase with increasing voltage (Fig. 3c) . Although at 3 kV, the flex-DBD appeared completely lit, fast imaging triggered on a current spike demonstrates that during each current spike only a few bright regions are observed (Fig. 3b) . Individual current spikes appear to correspond to isolated discharge events that appear randomly on the surface of the DBD. The number and the amplitude of the current spikes is not symmetrical in each half cycle of the AC current/voltage with a greater www.nature.com/scientificreports/ number of spikes occurring during the time when the mesh electrode acts as the anode, the voltage applied to the copper tape electrode is negative. In case of a positive mesh electrode, the electrons are able to flow into the anode and the current grows, but if the mesh electrode is negative, the electrons accumulate on the dielectric and the current stops resulting in lower current spikes. This asymmetry has been observed in plasma actuators that are a single edge surface DBD similar to the flex-DBD 20, 35 . The number of individual discharges or current peaks varies depending on the maximum applied voltage (overvoltage). For example, the number of current peaks (over 10 mA) is 15 ± 8 at 2 kV and increases to 45 ± 8 for 3 kV (Fig. 3c) . The greater number of current spikes results in a greater amount of charge transferred in the circuit as evident from the Lissajous plots (Fig. 3d) . The Lissajous plot has a two-slope shape with a slight asymmetry due to a greater number of more intense discharges for the negative voltage (positive patterned electrode). The energy dissipated in the circuit per one cycle can be calculated as the area of the Lissajous plot, and the power is then determined using the frequency, f , the duty cycle, ν, www.nature.com/scientificreports/ where Q is the charge measured by the capacitor probe and dV is the voltage obtained by the high voltage probe. For example, for the peak voltage of 1.9 kV the energy per cycle was 0.04 mJ/cycle. For the frequency of 41 kHz and a 20% duty cycle this gives the power of 0.3 W. For the max voltage of 2.9 kV the energy per cycle was 0.14 mJ/cycle, and the power, 1.1 W. The corresponding power density for the ~ 2 cm 2 device is 0.15-0.5 W/ cm 2 . The applied max AC voltage was varied from 1.6 kV to about 3 kV while keeping the frequency and the duty cycle constant. The resulting power varied linearly (Fig. 3e) with the applied voltage, which can be used as a calibration curve to set the desired power for a given device. Increasing the operating voltage increases the discharge power and corresponds to an increase in the number of individual discharges and the production of plasma. Increasing the duty cycle increases the overall power consumption by the device, but does not change the number of individual discharges per cycle. To evaluate the effectiveness of the flex-DBD device in decontamination of surfaces from biological contaminants, we conducted qualitative and quantitative experiments. The qualitative experiments included the treatment of E. coli in petri dishes, the decontamination of inoculated aluminum and fabric surfaces (Fig. 4a,b) ; the quantitative bacterial load reduction was determined by treating bacterial culture dried onto glass coverslips (Fig. 4c) . We placed droplets of the bacterial culture on an aluminum surface, treated by exposure to the flex-DBD, and then stamped with contact plates. The flex-DBD effectively reduced the bacterial load when the DBD was placed 1 mm from the surface and operated at 3 kV and 44 kHz, duty cycle of 20%. The flex-DBD device was also effective at disinfecting textile-type textile-like polyethylene material. To assess the spatial extent of disinfection we uniformly inoculated the fabric with a transgenic E. coli strain that expresses green fluorescent protein (OP-50-GFP) and treated a 10 × 20 mm area with the same operating parameters remained. Only viable bacteria contain GFP and fluoresce when excited with blue light. Indeed, there was a marked reduction in GFP positive colonies around the treated area (Fig. 4b) . The distance of the DBD from the surface is also important because reducing the distance from 1 mm above the surface to a direct contact with the ground electrode, increased the rate of inactivation of bacteria. We obtained a similar spatial pattern by treating E. coli bacterial culture dried onto the surface of a glass coverslip, following 30 s treatment with the flex-DBD. (1) P = f ν QdV , www.nature.com/scientificreports/ To quantify the bactericidal effect of the flex-DBD we inoculated and dried glass coverslips and measured the surviving bacterial load in colony forming units per milliliter (CFU/ml). Treatment with the flex-DBD device reduced viable bacteria log 10 (N o /N) = 4.1 after 90 s (Fig. 4c) . The flex-DBD was operated at a voltage of 3 kV and discharge power of 0.5 W/cm 2 , and the temperature of the grounded surface was below 50 °C. We repeated the inactivation of E. coli using the Standard E. coli strain AMC 198 (ATCC 11,229) (Fig. 4c) . Two experiments were conducted, one using a lower voltage, 2 kV peak voltage and the temperature of the grounded surface T < 40 °C, and 3 kV peak voltage and T < 50 °C. The higher applied voltage resulted in faster (p = 0.003, ANOVA) inactivation of E. coli; log 10 (N o /N) = 5.8 after 180 s treatment (Fig. 4c ) as compared to log 10 (N o /N) = 2.6 after 180 s, demonstrating a dependence on the flex-DBD peak voltage. We calculated the 1log 10 reduction (D value) for E. coli AMS 198 because the data is less variable than that of 10-beta, probably due to a greater control of the strain characteristics. At the start of the plasma treatment, the plasma affects the most susceptible bacteria that is located the closest to the plasma and hence is subjected to shorter-lived reactive plasma species. Hence the inactivation rate is the highest for the short treatment times. A linear fit to the treatment times of 10 s to 270 s give the times for 1log 10 reduction, D = 74 ± 2 s with the correlation coefficient, R = 0.996. We tested whether it has been the device operating temperature (40 °C and 50 °C) that caused disinfection of E. coli 10 Beta and E. coli AMC 198 strains. Instead of treatment with flex-DBD we incubated contaminated coverslips at 50 °C. We found no reduction in the bacterial load even at the longest exposure times of 180 s and 270 s. Therefore, the improvement in the disinfection at higher voltage may be attributed to plasma related effects such as the increase in the concentration of reactive species. Spectral analysis of the flex-DBD confirms the production of the OH · (Fig. 5a) and a wide range of ROS and RNS (Fig. 5b) by the discharge. Hydroxyl radical is the highest oxidizer and it reacts with lipids in the cell membrane and oxidizes proteins and nucleic acids inside the cells. Because of its reactivity, it is very short-lived and needs to be produced at the site of action. Since the inoculated coverslips were in contact with the patterned side of the flex-DBD during treatment, we can speculate that the reactions of OH · and other short-lived ROS are responsible for the fast initial rate of bacterial inactivation. Ozone and nitrogen oxides evident in the IR absorption spectrum can diffuse into the substrate and continue the disinfection process at a slower rate limited by the rate of diffusion and cellular processes 10-12 . Combined effect of plasma and hydrogen peroxide. To augment the disinfection we applied the flex-DBD treatment in combination with 3% H 2 O 2 solution commonly available and, widely used for oral, skin, and wound disinfection. Applying the DBD together with H 2 O 2 results in a 3.5 log 10 reduction in the bacterial population in just 10 s and in > 5log 10 reduction in 90 s (Fig. 4c) . The combined disinfection effect of hydrogen peroxide and plasma is faster than either DBD alone (p = 0.04, repeated measures ANOVA test) or H 2 O 2 alone (p = 0.03). The antibacterial mechanism of H 2 O 2 solutions alone is based on the production of the highly reactive hydroxyl radicals, but there is no appreciable equilibrium concentration of OH . in the solution itself. We used a chemical scavenger method to detect the presence of the OH . radical in 3% H 2 O 2 solution and found that the concentration is of the same order as coumarin/PBS solutions that contain no hydrogen peroxide (Fig. 5c) . The hydroxyl radical can be produced inside a cell by the Fenton reaction: The production of OH . leads to oxidation and eventually to cell death. This disinfection by H 2 O 2 alone depends on the concentration of the solution and can be concentration limited, slowing down as the hydrogen peroxide is used up 40 . Our results show that a bacterial load reduction due to the H 2 O 2 alone (Fig. 4c ) slows down faster than the corresponding DBD treatment (the times for 1log 10 reduction, D = 139 ± 5 s, R = 0.97 for E. coli 10-beta; and D = 135 ± 5 s, R = 0.947 for E. coli AMC 198). The sustained rate of reduction increases to D = 40 s (R = 0.86) when the treatment is carried out by both the DBD and the H 2 O 2 solution (Fig. 4c) and reaches a > 6log10 reduction in just 90 s. In humid environments, air plasma produces hydroxyl radicals generally through the interactions of electrons and excited nitrogen with water vapor. But the hydroxyl radical reacts, oxidizes or recombines to form hydrogen peroxide on a microsecond scale 32 To test whether the UV radiation from the flex-DBD was sufficient to explain the improved E. coli inactivation with H 2 O 2 we blocked all plasma products except for the UV radiation with a UV filter. UV radiation improved the inactivation of bacteria in the first 30 s compared with H 2 O 2 alone, but it did not achieve any additional reduction with increasing treatment time. After the first 30 s, the survival curve flattens (D > 250 s). This effect of UV radiation is insufficient to explain the improvement in the inactivation with H 2 O 2 achieved by the addition of the DBD plasma. Hence the plasma, not UV radiation alone, improves the action of H 2 O 2 . The remarkable synergy between plasma and H 2 O 2 can be explained by the combination of the peroxone process with the RNS produced by the flex-DBD. The IR absorption spectrum is dominated by ozone and RNS. Peroxone is an advanced oxidation process that has been known for over 100 years. It involves the reactions of ozone and hydrogen peroxide that promote the production of OH · , which is a much more effective oxidizer www.nature.com/scientificreports/ than ozone alone 42 . Plasma-generated superoxide can also aid in the process of generating OH · in solution. The scavenger method shows that plasma generates not only a stable hydrogen peroxide in a liquid solution but a measurable OH · concentration (Fig. 5c) . Plasma treatment of a H 2 O 2 solution results in a significant increase in the concentration of OH · in solution, well above plasma treatment alone. OH · oxidation can lead to an easier penetration of the membrane by RNS that damage the proteins inside the cell and improve the overall disinfection process [10] [11] [12] 43, 44 . Hence the combination of plasma treatment with hydrogen peroxide is a powerful tool for disinfection of bacterial contaminants. Although many years of investigation continue to demonstrate the potential of plasma in disinfection of biological and non-biological surfaces, the devices that have been accepted alongside the steam sterilization in medical and pharmaceutical industries are low pressure plasma-H 2 O 2 vapor chambers 9,29,30 . Our results demonstrate Fluorescence of the 7-hydroxycoumarin indicating the presence of OH · in solution. Solution of coumarin in PBS, and coumarin with H 2 O 2 are not different from the coumarin in water that is used to produce the baseline spectra. Fluorescence intensity increased at 460 nm following plasma treatment (DBD) and further increased for plasma treated H 2 O 2 solution (DBD + H 2 O 2 ). Spectra were taken immediately after treating 1.5 ml liquid samples for 5 min at the same DBD setting. Each curve is an average of three trials. www.nature.com/scientificreports/ that plasma disinfection could fill a technological gap not as a replacement for the standard bulk sterilization methods but in special niche applications and in personal consumer use 14 . Here we demonstrate the effectiveness of the first hand-held flex-DBD device that is suitable for personal use by untrained personnel. This flex-DBD device achieves fast disinfection, 4log 10 reduction, in under 90 s and reaches > 5log 10 in 270 s. Augmented with H 2 O 2 flex-DBD it achieves high bacterial load reduction two times faster than alone: > 6log 10 in 90 s with D = 2.5 s for the first 10 s and D = 30 s for 10 s -180 s treatment times. These results are faster than UV and chemicals alone and faster than atmospheric pressure glow discharge, mesh DBD, and plasma activated water. The synergy of plasma and hydrogen peroxide is due to plasma activation of H 2 O 2 . Our experimental results indicate that the mechanism of this activation is due to two factors. A peroxone process produces OH · in H 2 O 2 solution and the combined action of OH · and RNS leads to enhanced disinfection. Novel in-situ FTIR AS measurements of the flex-DBD show the presence of ozone and RNS in the plasma effluent. The scavenger method demonstrates a significant increase in the OH · concentration in the H 2 O 2 solution. These measurements support the proposed mechanism. The flex-DBD device can make effective disinfection accessible to the untrained public. It is safe in operation since the user facing components are grounded. The electrical measurements described above show that the flex-DBD operates consistently over prolonged periods, at least 15 min at a time for a total of more than 100 h of operation. The synergistic action with H 2 O 2 reaches high levels of disinfection and opens exciting new possibilities for decontamination and treatment. This device can be used to disinfect personal protection equipment, hands, and frequently touched surfaces, as well as for wound treatment and other medical applications. The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request. Received: 10 October 2020; Accepted: 9 February 2021 Gas plasma technology-an asset to healthcare during viral pandemics such as the COVID crisis Nonthermal decontamination of biological media by atmospheric-pressure plasmas: Review, analysis, and prospects Low temperature plasma-based sterilization: Overview and state-of-the-art Nonthermal dielectric-barrier discharge plasma-induced inactivation involves oxidative DNA damage and membrane lipid peroxidation in Escherichia coli A new flexible DBD device for treating infected wounds: In vitro and ex vivo evaluation and comparison with a RF argon plasma jet Cold argon-oxygen plasma species oxidize and disintegrate capsid protein of feline calicivirus Mechanism of virus inactivation by cold atmospheric-pressure plasma and plasma-activated water Disinfection and sterilization using plasma technology: Fundamentals and future perspectives for biological applications American Sterilizer Co, assignee. Hydrogen peroxide sterilization method Mechanisms of selective antitumor action of cold atmospheric plasma-derived reactive oxygen and nitrogen species Reactive species in non-equilibrium atmospheric-pressure plasmas: Generation, transport, and biological effects Synergistic effect of H 2 O 2 and NO 2 in cell death induced by cold atmospheric He plasma Transport and accumulation of plasma generated species in aqueous solution White paper on plasma for medicine and hygiene: Future in plasma health sciences Dielectric barrier discharges: Progress on plasma sources and on the understanding of regimes and single filaments Blood coagulation and living tissue sterilization by floating-electrode dielectric barrier discharge in air Arc-free atmospheric pressure cold plasma jets: A review Atmospheric pressure plasma jets: An overview of devices and new directions Applications of the COST plasma jet: More than a reference standard Asymmetric surface dielectric barrier discharge in air at atmospheric pressure: Electrical properties and induced airflow characteristics Low-temperature sterilization of wrapped materials using flexible sheet-type dielectric barrier discharge Paper-based plasma sanitizers Investigation of the roles of plasma species generated by surface dielectric barrier discharge Flexible electronics technologies for the fabrication of surface dielectric barrier discharge devices Quantification of air plasma chemistry for surface disinfection Gas plasma sterilization-application of space-age technology Gas-plasma sterilization: Relative efficacy of the hydrogen peroxide phase compared with that of the plasma phase Sterilization by H 2 O 2 droplets under corona discharge Hydrogen-peroxide-enhanced nonthermal plasma effluent for biomedical applications Decontamination with vaporized hydrogen peroxide is effective against Mycobacterium tuberculosis Combined effect of cold atmospheric plasma and hydrogen peroxide treatment on mature listeria monocytogenes and salmonella typhimurium biofilms Plasma-liquid interactions: A review and roadmap Non-thermal plasmas in and in contact with liquids Long-term antibacterial efficacy of air plasma-activated water NIH Image to ImageJ: 25 years of image analysis Coumarin-3-Carboxylic Acid as a Detector for Hydroxyl Radicals Generated Chemically and by Gamma Radiation Characterization of a flexible surface DBD device Electrical and mechanical characteristics of surface AC dielectric barrier discharge plasma actuators applied to airflow control Temporal evolution of a surface dielectric barrier discharge for different groups of plasma microdischarges The development of dielectric barrier discharges in gas gaps and on surfaces Inactivation of Escherichia coli cells in aqueous solution by atmospheric-pressure N 2 , He, Air, and O 2 microplasmas Disinfection of Model Indicator Organisms in a Drinking Water Pilot Plant by Using PEROXONE Measurement of reactive species generated by dielectric barrier discharge in direct contact with water in different atmospheres Aqueous-phase chemistry and bactericidal effects from an air discharge plasma in contact with water: Evidence for the formation of peroxynitrite through a pseudo-second-order post-discharge reaction of H 2 O 2 and HNO 2 The authors declare no competing interests. The online version contains supplementary material available at https ://doi. org/10.1038/s4159 8-021-84086 -z.Correspondence and requests for materials should be addressed to S.G. Publisher's note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.