key: cord-0847532-f4hdyags authors: Kogoj, Rok; Korva, Misa; Knap, Nataša; Resman Rus, Katarina; Pozvek, Patricija; Avšič-Županc, Tatjana; Poljak, Mario title: Comparative Evaluation of Six SARS-CoV-2 Real-Time RT-PCR Diagnostic Approaches Shows Substantial Genomic Variant–Dependent Intra- and Inter-Test Variability, Poor Interchangeability of Cycle Threshold and Complementary Turn-Around Times date: 2022-04-12 journal: Pathogens DOI: 10.3390/pathogens11040462 sha: a2cf4873f7ef8243a06a703c25ad7ed7104853a3 doc_id: 847532 cord_uid: f4hdyags Several professional societies advise against using real-time Reverse-Transcription PCR (rtRT-PCR) cycle threshold (Ct) values to guide clinical decisions. We comparatively assessed the variability of Ct values generated by six diagnostic approaches by testing serial dilutions of well-characterized isolates of 10 clinically most relevant SARS-CoV-2 genomic variants: Alpha, Beta, Gamma, Delta, Eta, Iota, Omicron, A.27, B.1.258.17, and B.1 with D614G mutation. Comparison of three fully automated rtRT-PCR analyzers and a reference manual rtRT-PCR assay using RNA isolated with three different nucleic acid isolation instruments showed substantial inter-variant intra-test and intra-variant inter-test variability. Ct value differences were dependent on both the rtRT-PCR platform and SARS-CoV-2 genomic variant. Differences ranging from 2.0 to 8.4 Ct values were observed when testing equal concentrations of different SARS-CoV-2 variants. Results confirm that Ct values are an unreliable surrogate for viral load and should not be used as a proxy of infectivity and transmissibility, especially when different rtRT-PCR assays are used in parallel and multiple SARS-CoV-2 variants are circulating. A detailed turn-around time (TAT) comparative assessment showed substantially different TATs, but parallel use of different diagnostic approaches was beneficial and complementary, allowing release of results for more than 81% of non-priority samples within 8 h after admission. Following the initial emergence of Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) and its subsequent global pandemic spread, a range of molecular SARS-CoV-2 RNA assays, both in-house and commercial, were rapidly developed following the real-time Reverse-Transcription PCR (rtRT-PCR) protocols summary by the WHO (https: //www.who.int/docs/default-source/coronaviruse/whoinhouseassays.pdf, accessed on 22 March 2022). Consequently, a substantial number of SARS-CoV-2 RNA assays received In Vitro Diagnostics Emergency Use Authorization (EUA) approval from the US Food and Drug Administration (FDA) (https://www.fda.gov/medical-devices/coronavirus-disease-2019-covid-19-emergency-use-authorizations-medical-devices/in-vitro-diagnostics-euasmolecular-diagnostic-tests-sars-cov-2, accessed on 22 March 2022) . Due to the unprecedented need for SARS-CoV-2 testing, significant supply problems and a shortage of sample devices, instruments, reagents, and consumables soon arose and are ongoing [1, 2] . To The results of testing in triplicate of six different concentrations of 10 well-characterized samples containing different SARS-CoV-2 genomic variants using six diagnostic approaches are shown in Table 1 . None of the dilutions were SARS-CoV-2 RNA-negative for any of the target genes when tested using three fully automated rtRT-PCR analyzers. Inter-variant Ct values SDs of about one log were observed throughout all concentrations tested for all targets by all analyzers with the exception of STARLet's target gene N. For this target, intervariant Ct value SDs between 2.2 and 2.8 log were observed at respective concentrations ( Table 1) . For STARLet's target gene N, mean Ct values at all concentrations tested were also significantly higher (p < 0.001) for Beta, Eta, and Iota variants in comparison to other SARS-CoV-2 genomic variants tested (Supplemental Table S1 ). No such deviations were observed for any of the targets in all assays tested regardless of SARS-CoV-2 genomic variant included in this study. Similarly, none of the dilutions were SARS-CoV-2 RNAnegative for any of the target genes when tested by LightMix after nucleic acid isolation using three different analyzers. The SDs of LightMix's Ct values in samples processed by three different nucleic acid isolation systems were also about 1 log across all serial dilutions created, which is comparable to the data obtained by fully automated integrated rtRT-PCR analyzers ( Table 1) . Table 1 . Inter-variant mean Ct values with standard deviations (SD), coefficient of variation (CV), and individual variant lowest mean/highest mean Ct at respective concentration points generated by specific PCR targets in all rtRT-PCR diagnostic approaches assessed (Alinity, cobas, STARLet, and LightMix) when testing in triplicate six different concentrations of 10 well-characterized isolates containing different SARS-CoV-2 genomic variants. 11.0 ± 0.6 Different numbers of rtRT-PCR targets, target gene regions, or even a combination of targets are used in the diagnostic approaches evaluated. Therefore, rtRT-PCR efficiency was chosen rather than Ct values for comparison of results obtained by six diagnostic approaches related to respective SARS-CoV-2 genomic variants. As presented in Table 2 , the comparative assessment showed that, regardless of the SARS-CoV-2 variant, all six diagnostic approaches performed with rtRT-PCR efficiencies between 80.1% and 106.8%. However, rtRT-PCR efficiencies for the same genomic variant were not consistent between assays. Regardless of the SARS-CoV-2 variant tested, cobas showed the most consistent rtRT-PCR efficiency, ranging from 98.8% to 105.0% and from 95.8% to 106.8% for Target 1 (ORF1ab) and Target 2 (E gene), respectively (Table 2 ). Overall, the targets in the respective assays were amplified with a mean inter-test rtRT-PCR efficiency of at least 91% for each genomic variant tested, with the exception of the Gamma (P.1) variant, with a mean inter-test rtRT-PCR efficiency of 88.5% (Table 2) . Cobas had the best rtRT-PCR efficiency (100.1% Target 1; 96.8% Target 2) for detection of the Gamma variant, whereas in all other diagnostic approaches rtRT-PCR efficiency for this genomic variant did not reach the 90% threshold, the lowest being for STARLet gene N (80.6%) ( Table 2 ). Direct comparison of Ct values between different diagnostic approaches is hampered because each assay/analyzer used different targets, chemistry, and design, but the RdRP gene was represented as a target in all six approaches chosen. Thus, detailed head-to-head Ct comparison between six diagnostic approaches was performed for the RdRP gene only (Table 3 and Figure 1 ). The results show substantial diagnostic approachdependent differences in mean Ct values across all dilutions tested, looking at genomic variants separately. At respective concentrations, the differences between the platform with the maximum mean Ct value and the platform with the minimum mean Ct value span from 2.4 to 7.1. If all genomic variants are considered together, these differences are even greater: between 5.7 and 7.6 ( Table 3 ). Figure 1 shows the distribution of RdRP mean Ct values generated by six diagnostic approaches evaluated for 10 SARS-CoV-2 genomic variants tested. It reveals that, when testing the same concentrations of either genomic variant, the resulting Ct range varies greatly and is dependent on the diagnostic approach and genomic variant. Assessment of systemic error, robustness of the systems used, and pipetting accuracy are presented in Supplemental Figure S1 . Table 3 . Inter-test comparison of RdRP gene mean Ct values with standard deviations (SD) and coefficient of variation (CV) generated by all diagnostic approaches tested and difference between platform with highest (hi) and lowest (lo) individual test mean Ct value (∆ hi-lo) for respective concentration of 10 clinically relevant SARS-CoV-2 genomic variants altogether and separately. TATs were comparatively assessed during the period when all six diagnostic approaches were used in parallel. From March 2020 to the end of May 2021, a total of 504,173 samples for SARS-CoV-2 RNA testing were received at IMI. Of these, 96.1% (484,336/504,173) were processed as regular samples. The remaining 3.9% (19,837/504,173) were ordered as high-priority samples (STAT) and were processed either by XpertXpress SARS-CoV-2 (Cepheid, Sunnyvale, CA, USA) or cobas LIAT (Roche). However, due to the frequent supply shortage of cartridges/assay tubes needed for two ultra-fast rtRT-PCRs, TATs were comparatively assessed during the period when all six diagnostic approaches were used in parallel. From March 2020 to the end of May 2021, a total of 504,173 samples for SARS-CoV-2 RNA testing were received at IMI. Of these, 96.1% (484,336/504,173) were processed as regular samples. The remaining 3.9% (19,837/504,173) were ordered as high-priority samples (STAT) and were processed either by XpertXpress SARS-CoV-2 (Cepheid, Sunnyvale, CA, USA) or cobas LIAT (Roche). However, due to the frequent supply shortage of cartridges/assay tubes needed for two ultra-fast rtRT-PCRs, 29.9% (5933/19,837) of STAT samples were processed with LightMix or Alinity, amounting to a total of 490,269 samples included in the TAT analysis. As shown in Figure 2 In the past 2 years, molecular laboratories around the world were forced to unprecedentedly increase their capacity for rtRT-PCR testing to meet extraordinary demands for SARS-CoV-2 RNA testing from authorities and healthcare facilities. In addition to highvolume testing, laboratories were challenged daily by several interpretation issues, one of the most frequent being applying rtRT-PCR Ct values for better management of COVID-19 patients. Predicting and determining the duration of infectivity and transmissibility is indeed a very important aspect in management of COVID-19 patients as well as exposed individuals to allow timely hospital discharge, and as short a quarantine and isolation duration as possible and their appropriate termination. Although the quantification ability of rtRT-PCR is indispensable in management of several important infectious diseases, the use of Ct values in management of COVID-19 is still controversial [20] . Namely, due to numerous analytical and clinical factors known to impact Ct values in general, the nonstandardized nature of sample collection using a nasopharyngeal swab, and the fact that patients that have recovered from COVID-19 can remain rtRT-PCR-positive for a prolonged period of time [12, 21] , there is a consensus that the mere presence of SARS-CoV-2 RNA does not necessarily reflect the infectivity and transmissibility of a patient [22, 23] . Despite all these uncertainties and the almost uniform recommendation of professional societies and some authorities and findings by researchers [24] against the routine use of Ct values to guide clinical decisions, a general concept that classifies patients with a Ct value of ≥30 as non-infectious unfortunately seems to have prevailed in peer-reviewed literature [25, 26] . To the best of our knowledge, no comprehensive comparison of Ct values and rtRT-PCR efficiencies generated by several rtRT-PCR platforms and various nucleic acid isolation methods across a range of the clinically most relevant SARS-CoV-2 genomic variants has been performed yet. For this study, all variants of concern (VOC) and variants of interest (VOI) detected in the central European region until the end of 2021 were selected for analysis. In addition, variant B.1.258.17 as the main lineage during the second wave, In the past 2 years, molecular laboratories around the world were forced to unprecedentedly increase their capacity for rtRT-PCR testing to meet extraordinary demands for SARS-CoV-2 RNA testing from authorities and healthcare facilities. In addition to highvolume testing, laboratories were challenged daily by several interpretation issues, one of the most frequent being applying rtRT-PCR Ct values for better management of COVID-19 patients. Predicting and determining the duration of infectivity and transmissibility is indeed a very important aspect in management of COVID-19 patients as well as exposed individuals to allow timely hospital discharge, and as short a quarantine and isolation duration as possible and their appropriate termination. Although the quantification ability of rtRT-PCR is indispensable in management of several important infectious diseases, the use of Ct values in management of COVID-19 is still controversial [20] . Namely, due to numerous analytical and clinical factors known to impact Ct values in general, the nonstandardized nature of sample collection using a nasopharyngeal swab, and the fact that patients that have recovered from COVID-19 can remain rtRT-PCR-positive for a prolonged period of time [12, 21] , there is a consensus that the mere presence of SARS-CoV-2 RNA does not necessarily reflect the infectivity and transmissibility of a patient [22, 23] . Despite all these uncertainties and the almost uniform recommendation of professional societies and some authorities and findings by researchers [24] against the routine use of Ct values to guide clinical decisions, a general concept that classifies patients with a Ct value of ≥30 as non-infectious unfortunately seems to have prevailed in peer-reviewed literature [25, 26] . To the best of our knowledge, no comprehensive comparison of Ct values and rtRT-PCR efficiencies generated by several rtRT-PCR platforms and various nucleic acid isolation methods across a range of the clinically most relevant SARS-CoV-2 genomic variants has been performed yet. For this study, all variants of concern (VOC) and variants of interest (VOI) detected in the central European region until the end of 2021 were selected for analysis. In addition, variant B.1.258.17 as the main lineage during the second wave, variant A.27 as a rare variant, with mutation N501Y, and variant B.1 (with D614G) as the first main lineage, which caused a substantial part of the first wave of infections in Europe, were included. All 10 genomic variants were tested in triplicate using six selected diagnostic approaches at concentrations ranging from 1 × 10 4 PFU/mL to 1 × 10 −1 PFU/mL. This study clearly showed substantial differences in Ct values generated by six selected diagnostic approaches with respect to the SARS-CoV-2 genomic variant, even without taking into account sample quality variability. Differences ranging from 2.0 to 8.4 Ct values were observed for the same concentration of different SARS-CoV-2 genomic variants, clearly showing a relevant influence of genomic variant on the detection capability of rtRT-PCRs. Although larger Ct variability was initially expected for manual diagnostic approaches over fully automated analyzers, it turned out that SD values for respective concentrations were quite similar between manual and automated approaches, except when targeting STARLet's N gene. In addition, for STARLet's N gene, significantly higher Ct values (p < 0.001) for the Beta (B.1.351), Eta (B.1.525), and Iota (B.1.526) genomic variants were found at all concentrations tested compared to other genomic variants included in the study. Similar deviations for the SARS-CoV-2 Alpha genomic variant were previously observed with the SARS-CoV-2/FluA/FluB/RSV assay [27] [28] [29] . There is reason to believe that these genomic variants most likely have mutation(s) in the N gene at primers and/or probe binding site(s) that affect the amplification efficacy and/or intensity of the probe signal, resulting in higher Ct values. In this study, such N-gene Ct value deviations were not observed for the Alpha genomic variant using the SARS-CoV-2 specific Allplex assay. With other diagnostic approaches included in this study such deviations were not observed for any of the rtRT-PCR targets regardless of SARS-CoV-2 genomic variant tested. These results clearly indicate that Ct values as a surrogate for viral load are unreliable and should not be used as an indicator of infectivity and transmissibility, especially when different rtRT-PCR assays are used in the laboratory in parallel and multiple SARS-CoV-2 genomic variants are circulating in a population. This observation is supported by the findings of Arons et al. [30] and Scola et al. [14] , who have shown that 25% and even 50% of clinical samples with Ct values above 30 can be cultured and could be potentially infectious. In addition, the results of recent external quality assessment showed that Ct values reported by laboratories can vary by more than ±4 (up to 18) depending on the nucleic acid extraction protocol and rtRT-PCR platform used [31] . Furthermore, the results of this study showed that some SARS-CoV-2 genomic variants are not amplified with equal rtRT-PCR efficiency using different rtRT-PCR assays. Although all targets in the diagnostic approaches assessed were amplified with rtRT-PCR efficiencies between 80 and 120%, which is generally considered acceptable, rtRT-PCR efficiencies for some SARS-CoV-2 genomic variants showed a 14.8 to 24.1% difference between diagnostic approaches with potential patient management consequences. Such differences are best recognizable when mean Ct values for the RdRP gene, which is a common target for all six diagnostic approaches, are compared head-to-head. The RdRP gene mean Ct values varied from 2.4 to 7.1 across diagnostic approaches tested when the genomic variants are considered individually, and when the results are calculated independent of genomic variant, the differences for given concentrations ranged from 5.7 to 7.6 at respective concentrations across the diagnostic approaches tested. This clearly indicates that the use of the Ct value 30 as a simple cutoff for infectivity and transmissibility is inappropriate and that the Ct values measured are dependent on the diagnostic platform and genomic variant. However, quantification precision can be significantly improved by using standardized quantitative rtRT-PCR or droplet digital PCR [32] . Several strategies have been employed by laboratories worldwide to meet the high COVID-19 testing demand, including pooling of samples [33] [34] [35] , multiplexing [36] [37] [38] , and omission or simplification of nucleic acid isolation [39, 40] , all with expected tradeoffs in overall analytical sensitivity. In many laboratories, including our laboratory, requested diagnostic scaling up could only be achieved by using several molecular platforms in parallel, ranging from fully automated integrated sample-to-result analyzers such as cobas or Alinity to different semi-automated or manual approaches. Although such a strategy was proven to be beneficial timewise, the use of diverse diagnostic approaches in parallel undeniably led to potentially clinically relevant intra-laboratory assay-to-assay variations and poor result interchangeability, as shown in this study. However, on the positive side, having diverse diagnostic approaches running in parallel allowed us to meticulously measure and comparatively evaluate TATs of different diagnostic approaches in a real-life laboratory setting. Overall, during three SARS-CoV-2 waves, a combination of six automated and manual diagnostic approaches made possible the release of final results for more than 81% of routine non-priority samples within 8 h after admission to the laboratory. A detailed TAT analysis showed that most results were released in two peak time periods, the major one being within the first 8 h after admission (Figure 2 ). This first peak time period was composed of several sub-peaks (Figure 3) , the first containing samples processed by LightMix and Alinity, followed by cobas and finally STARLet. Both Alinity and LightMix were found to be comparable in terms of TAT and the fastest of the diagnostic approaches comparatively assessed, with TAT in 3 to 4 h. This observation is not surprising because Alinity takes 140 min to process a batch of the first 12 samples, adding another 12 samples every 16 min. Similarly, LightMix, in conjunction with the slowest of the automated nucleic acid isolation approaches assessed (MagNA), requires up to 160 min for 96 samples. Cobas follows with mean TAT of 4 to 6 h, processing samples in batches of up to 94 in 180 min. The slowest of the diagnostic approaches evaluated was STARLet, which took 6 to 7 h to process most of the samples because it had the longest processing time for batches of 94 samples (250 min). Although having substantially different individual TATs, a parallel use of six diagnostic approaches was found to be quite complementary in a real-life laboratory setting, allowing day-today flexibility, testing of up to 8000 samples in 24 h without significant prolongation of TAT, and providing a valuable back-up solution in cases of technical issues, failures, and temporary reagent shortages. Other possibilities to shorten TAT are available such as ultrafast PCR [41, 42] and isothermal assays [43, 44] however, implementation of such methods in parallel to rtRT-PCR may bring even further complexity and complications when a request for Ct value is made. There are some important limitations of this study. The input volume of samples differed slightly across the diagnostic approaches evaluated, both automated and manual, but it was dictated by manufacturers' instructions. Unfortunately, the elution volumes during RNA extraction and the volumes used for subsequent rtRT-PCR in the automated analyzers are not disclosed by the manufacturers. The sample input volumes as well as elution volumes for the three automated nucleic acid isolation platforms were also slightly different. In addition, virus quantification by PFU is associated with some degree of inaccuracy due to the method design, which cannot be fully avoided [45] . Finally, only the clinically most relevant VOCs, VOIs, and major lineages circulating in central Europe until the end of 2021 were tested. Although the factors listed may introduce some biases and variability in the Ct values generated, we remain confident that the differences in Ct values and rtRT-PCR efficiencies identified in this comparative study are mainly driven by differences in the assays' design and are genomic variant-dependent. In summary, testing serial dilutions of well-characterized SARS-CoV-2 isolates of a range of the clinically most relevant SARS-CoV-2 genomic variants using six previously thoroughly evaluated diagnostic approaches showed substantial inter-variant intra-test as well as intra-variant inter-test variability. The differences obtained in Ct values were dependent on both the rtRT-PCR platform and SARS-CoV-2 genomic variant. This study further supports recommendations against the routine use of Ct values to guide clinical decisions and reconfirms the need for close monitoring of new emerging SARS-CoV-2 genomic variants and their potential impact on the performance of rtRT-PCR assays. A detailed TAT analysis showed that, despite substantially different individual TATs in a real-life laboratory setting, the parallel use of several diagnostic approaches was beneficial and complementary and made possible the release of final results for more than 81% of routine non-priority samples within 8 h after admission to the laboratory during three SARS-CoV-2 waves. Since March 2020, complete SARS-CoV-2 genomes have been routinely sequenced at IMI with next-generation sequencing (NGS) using residual nasopharyngeal swab samples of COVID-19-positive patients to actively monitor the epidemiology, circulation, and emergence of SARS-CoV-2 variants in the country. All fully characterized samples have been stored at −70 • C and whole genome sequences uploaded to the GISAID database (https: //www.gisaid.org, (accessed on 22 March 2022)). Selected samples with Ct values below 25 have been additionally cultured on Vero E6 cells and quantified by TCID50 as described previously [46] . The following 10 SARS-CoV-2 variants were selected for this study: Alpha Six different diagnostic approaches were comparatively evaluated: three fully automated and integrated rtRT-PCR analyzers and a reference manual rtRT-PCR assay using RNA isolated with three different instruments for automated nucleic acid isolation. All testing was performed strictly following the respective manufacturers' instructions. Briefly, logarithmic dilutions of nine samples containing the SARS-CoV-2 genomic variants listed above were prepared in fresh RPMI medium (Sigma Aldrich, St. Louis, MO) to obtain 1 × 10 4 /1 × 10 3 /1 × 10 2 /1 × 10 1 /1 × 10 0 /1 × 10 −1 PFU/mL. The dilutions were distributed in aliquots of appropriate volume and tested in triplicate with cobas SARS-CoV-2 Test (Roche) on a cobas 6800 analyzer (cobas), Allplex SARS-CoV-2 Assay (Seegene) on a STARLet analyzer (STARLet), and SARS-CoV-2 Kit (Abbott) on an Alinity_m analyzer (Alinity). Similarly, for manual LightMix Wuhan CoV kits (LightMix), nucleic acids were isolated from triplicates of appropriate sample volume for each automated nucleic acid isolation instrument: 190 µL for NX-48, 200 µL for MagNA Pure 96, and 300 µL for Maelstrom 9600. Nucleic acids were eluted in volumes as recommended by the manufacturers. Isolated RNA and reagents for rtRT-PCRs were automatically pipetted into 384-well PCR plates (Applied Biosystems, Foster City, CA, USA) using a Janus G3 pipetting robot (Perkin Elmer, Waltham, MA, USA) dispensing 5 µL of RNA eluate and 7.5 µL of rtRT-PCR master mix per well. Two separate master mixes were prepared from 1 × TaqMan FastVirus 1-Step Master Mix (Applied Biosystems) and each Wuhan SARS-CoV-2 E gene and Wuhan SARS-CoV-2 RdRP gene kits. Real-time RT-PCR amplification and detection were performed using a QuantStudio7Pro instrument (Applied Biosystems). Ct values generated by six diagnostic approaches were collected and, after computing mean Ct values for each concentration tested, used for calculation of respective rtRT-PCR efficiencies obtained from the slopes of linear regression lines across dilutions tested using the following equation: Efficiency (%) = ((10e (−1/slope)) − 1) × 100). PCR efficiency of 100% means that in each PCR cycle the amount of target is doubled. Statistical calculations and graph plotting were performed using Excel 2016 version 16.0.5188.1000 (Microsoft, Redmond, MA, USA), Prism7 version 7.04 (GraphPad, San Diego, CA, USA), and R software version 4.1.2 (The R Foundation for Statistical computing, Vienna, Austria). Data for laboratory admission time and time-to-result reporting were analyzed from March 2020 until the end of May 2021 and used to calculate the total TAT for each of the six diagnostic approaches. Only routinely processed samples tested with cobas, STARLet, Alinity, and LightMix were included in the TAT analysis, and fast-track prioritized samples were not considered. Supplementary Materials: The following supporting information can be downloaded at: https: //www.mdpi.com/article/10.3390/pathogens11040462/s1. Table S1 : STARLet's mean Ct values obtained by testing serial dilutions of 10 SARS-CoV-2 genomic variants; Figure S1 : Assessment of systemic error, robustness of the systems used, and pipetting accuracy. The study protocol conformed to the Declaration of Helsinki, Oviedo Convention on Human Rights and Biomedicine, and Slovenian Code on Medical Deontology, and it was approved by the Medical Ethics Committee of the Republic of Slovenia (consent number 0120-211/2020/7). All data used were linked to randomized numerical codes only. Data Availability Statement: Additional data presented in this study are available on request from the corresponding author. An alternative workflow for molecular detection of SARS-CoV-2-escape from the NA extraction kit-shortage Automated SARS-CoV-2 RNA extraction from patient nasopharyngeal samples using a modified DNA extraction kit for high throughput testing SARS-CoV-2 Infection: New Molecular, Phylogenetic, and Pathogenetic Insights. Efficacy of Current Vaccines and the Potential Risk of Variants COVID-19 Diagnostic Methods and Detection Techniques: A Review Diagnostics for SARS-CoV-2 detection: A comprehensive review of the FDA-EUA COVID-19 testing landscape Laboratory diagnosis of emerging human coronavirus infections-the state of the art Summary of the Available Molecular Methods for Detection of SARS-CoV-2 during the Ongoing Pandemic Viral Load Difference between Symptomatic and Asymptomatic COVID-19 Patients: Systematic Review and Meta-Analysis COVID-19), viral load and clinical outcomes; lessons learned one year into the pandemic: A systematic review Viral Dynamics and Real-Time RT-PCR Ct Values Correlation with Disease Severity in COVID-19 SARS-CoV-2, SARS-CoV, and MERS-CoV viral load dynamics, duration of viral shedding, and infectiousness: A systematic review and meta-analysis Duration of infectiousness and correlation with RT-PCR cycle threshold values in cases of COVID-19 Residual SARS-CoV-2 RNA in nasal swabs of convalescent COVID-19 patients: Is prolonged quarantine always justified? Viral RNA load as determined by cell culture as a management tool for discharge of SARS-CoV-2 patients from infectious disease wards Utility of Ct values in differentiating COVID-19 reinfection versus prolonged viral shedding in an immunocompromised patient High proportion of low cycle threshold value as an early indicator of COVID-19 surge Real-Life Head-to-Head Comparison of Performance of Two High-Throughput Automated Assays for the Detection of SARS-CoV-2 RNA in Nasopharyngeal Swabs: The Alinity m and cobas 6800 SARS-CoV-2 Assays Clinical Evaluation of the cobas SARS-CoV-2 Test and a Diagnostic Platform Switch during 48 Hours in the Midst of the COVID-19 Pandemic AccuPower SARS-CoV-2 Multiplex RT-PCR kit (Bioneer, South Korea)" and "Allplex 2019-nCoV Assay (Seegene, South Korea)" for SARS-CoV-2 RT-PCR Diagnosis: Korean CDC EUA as a Quality Control Proxy for Developing Countries Caution in interpretation of SARS-CoV-2 quantification based on RT-PCR cycle threshold value Prolonged Detection of Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) RNA in an Obstetric Patient with Antibody Seroconversion Culture-Based Virus Isolation To Evaluate Potential Infectivity of Clinical Specimens Tested for COVID-19 SARS-CoV-2 shedding and infectivity CT-Values, and Infectivity-Conclusions to Be Drawn from Side Observations. Viruses Can the cycle threshold (Ct) value of RT-PCR test for SARS CoV2 predict infectivity among close contacts? Cycle threshold values in RT-PCR to determine dynamics of SARS-CoV-2 viral load: An approach to reduce the isolation period for COVID-19 patients Rapid screening for SARS-CoV-2 VOC-Alpha (202012/01, B.1.1.7) using the Allplex™ SARS-CoV-2/FluA/FluB/RSV Assay SARS-CoV-2 N gene dropout and N gene Ct value shift as indicator for the presence of B.1.1.7 lineage in a commercial multiplex PCR assay Diagnostic pre-screening method based on N-gene dropout or delay to increase feasibility of SARS-CoV-2 VOC B.1.1.7 detection Presymptomatic SARS-CoV-2 Infections and Transmission in a Skilled Nursing Facility Variability of cycle threshold values in an external quality assessment scheme for detection of the SARS-CoV-2 virus genome by RT-PCR Digital droplet PCR accurately quantifies SARS-CoV-2 viral load from crude lysate without nucleic acid purification Assessment of Sample Pooling for Clinical SARS-CoV-2 Testing Selfcollection and pooling of samples as resources-saving strategies for RT-PCR-based SARS-CoV-2 surveillance, the example of travelers in French Polynesia Pooled RT-qPCR testing for SARS-CoV-2 surveillance in schools-a cluster randomised trial Detection of SARS-CoV-2 RNA by multiplex RT-qPCR Triplex Real-Time RT-PCR for Severe Acute Respiratory Syndrome Coronavirus 2 Enhanced throughput of the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) real-time RT-PCR panel by assay multiplexing and specimen pooling Clinical usefulness of extraction-free PCR assay to detect SARS-CoV-2 Evaluation of three extraction-free SARS-CoV-2 RT-PCR assays: A feasible alternative approach with low technical requirements Fast detection of SARS-CoV-2 RNA via the integration of plasmonic thermocycling and fluorescence detection in a portable device Ultrafast Photonic PCR Based on Photothermal Nanomaterials COVID-19 Infection Diagnosis: Potential Impact of Isothermal Amplification Technology to Reduce Community Transmission of SARS-CoV-2. Diagnostics Rapid Detection of COVID-19 Coronavirus Using a Reverse Transcriptional Loop-Mediated Isothermal Amplification (RT-LAMP) Diagnostic Platform Comparison of the plaque assay and 50% tissue culture infectious dose assay as methods for measuring filovirus infectivity Performance of the rapid high-throughput automated electrochemiluminescence immunoassay targeting total antibodies to the SARS-CoV-2 spike protein receptor binding domain in comparison to the neutralization assay We express our gratitude to all members of the COVID-19 diagnostic team, who have been working diligently for the past 20 months to ensure timely SARS-CoV-2 results, and to Alen Suljič for help with graphic artwork. We also thank the Office of Defense Cooperation, U.S. Embassy in Slovenia, for the donation of a Janus G3 pipetting robot. The Institute of Microbiology and Immunology, Faculty of Medicine, University of Ljubljana, Slovenia, received research grants from Roche, Seegene, and Abbott Molecular in the past 5 years, but not for this work. All authors declare no competing interest. The manufacturers of the evaluated fully automated integrated rtRT-PCR analyzers and processors for automated nucleic acid isolation had no role in the study design, data collection, analysis and interpretation of results, writing of the manuscript, or decision to submit the work for publication.