key: cord-0844514-6s7k9319 authors: LeConte, Bailey; Low, Garren M.I.; Citardi, Martin J.; Yao, William C.; Eguia, Arturo A.; Luong, Amber U. title: Aerosol generation with common rhinologic devices: Cadaveric study conducted in a surgical suite date: 2020-08-07 journal: Int Forum Allergy Rhinol DOI: 10.1002/alr.22679 sha: e251e6f44a6ad3ad9c34b1b0e1bcfb098a242cbd doc_id: 844514 cord_uid: 6s7k9319 nan Adding to the understanding of aerosol generation of rhinologic instruments in recently published data, we evaluated additional instruments typically used for in-office procedures and analyzed generation of submicron aerosols. We found that in addition to the bipolar cautery and highspeed drill, the bipolar radiofrequency (RF) ablation (coblation) and cryotherapy were aerosol generating procedures (AGPs), and other than the high-speed drill, these AGPs generated aerosols that skewed towards the submicron size. An aerosol suction device placed in the nasopharynx reduced aerosol escape from the nasal cavity by >99.8% with all AGPs. The World Health Organization along with hundreds of medical experts recognize that aerosol transmission of SARS-CoV-2, the virus associated with the COVID-19 pandemic, can lead to infection 1 . Overall, particles under 100 µm can be infectious with those greater than 20 µm subject to gravity 2 . However, particles <10 µm follow airflow streamlines and can deposit in the upper airway, and those 5-10 µm or smaller are capable of penetrating the lower airway to the alveoli 3 . This cadaveric study simulated real-world conditions by modeling intubated patients in the operating room and awake patients during in-office procedures to evaluate submicron and greater aerosol production of rhinologic instruments. Set in an operating room within an ambulatory surgery center (2058 cubic feet), a thawed cadaver head was placed supine on the operating room table with a laser diode-based particle counter (Lasair III 310C, Particle Measuring Systems [Boulder, CO]) flow intake positioned at the tip of the nose above the nostril. The intake flow rate measuring total particle counts (0.3 m to 25.0 m) as well as counts by particle size was constant at 1 ft³/min (28.3 L/min). To simulate an in-office awake patient, a plastic suction tubing (6 foot tube with 3/8 inch inner diameter, Medi-Vac® non-conductive suction tubing, Cardinal Health [Dublin,OH]) was placed in the nasopharynx via the oral cavity, and the oral cavity was sealed with saline soaked gauze using the same cadaver head as described above. A simulated exhalation was produced by one of the authors (BL) exhaling through the suction tubing into the nasopharynx at approximately 12 breaths/minute. To determine the number of aerosols generated by each instrument procedure, the calculated mean background particle count/ft 3 for each instrument was then subtracted from the mean total particle counts/ft³ for each instrument. Of the 6 instruments evaluated, 4 were noted to generate aerosols over background: bipolar cautery, powered high-speed drill, bipolar RF ablation (coblation) and cryotherapy (Fig 1) . When evaluating the size distribution of the aerosols generated, the high-speed drill was the only instrument that generated a significant amount of 1 µm or greater sized aerosols (Fig 2) . The other AGPs generated more aerosols less than 1 µm, than 1 µm or greater. No significant aerosols 10 µm or greater were generated from any instruments. In the presence of a novel aerosol evacuation device placed in the contralateral nasopharynx, the hands-free device captured >99.8% of generated aerosols of all sizes between 0.3 to 25+ µm. Similar to Workman et al, we confirmed that the high-speed drill and bipolar cautery both were significant aerosol producers while the microdebrider with in-line suction was not 4, 5 . We found that the bipolar RF ablation with in-line suction (coblation) also created aerosols, but to a lesser extent than the bipolar cautery. Despite the in-line suctions associated with the bipolar RF ablation and high-speed drill, the suction produced was insufficient to prevent the generated aerosols from escaping out of the nasal cavity. Of the two instruments tested with the exhalation model, the cryotherapy device generated aerosols, mainly in the 0.3 -1.0 m range, but overall produced noticeably less aerosols than the bipolar cautery or high-speed drill (Fig 1 and 2) . Even without tissue contact, the cooling of the device balloon during activation generated a visible plume which would be detected as aerosols by the particle counter. With the exception of the high-speed drill, the other AGPs generally produced submicron aerosols. The distribution of aerosol size with the high-speed drill skewed to 1 µm or greater, however, submicron aerosols were also produced. This is an important observation given aerosols < 10 µm can be suspended in a gas and inspired 2,6 . Aerosols < 5 µm, the majority of the aerosols generated in this study, are able to penetrate to the alveolar spaces with smaller aerosols penetrating deeper 6 . The SARS-CoV-2 spike protein receptor used for viral attachment is highly expressed on type II airway pneumocytes 7 making exposure to these cells an important pathway for infection. It is important to note that this study only demonstrates that aerosols were generated with instrumentation. It is not possible to make a firm determination on the make-up of these particles or whether they were able to carry a virus such as SARS-CoV-2. We would like to thank Aerin Medical, Hemostasis, Medtronic ENT, and Smith-Nephew for providing in-kind disposables to this study. We would also like to thank Michael and Brian Spearman of Hemostasis for providing in-kind the cadaver head, CoronaVac aerosol evacuation device, and use of the particle counter. Finally, we thank Matt Brokaw and Andrew Frazier for sharing their internal studies and data that inspired our studies. cautery, microdebrider, high-speed drill, and bipolar radiofrequency ablation were tested in the intubated patient simulation. Temperature-controlled RF ablation and cryotherapy were tested with the in-office awake patient simulation in which exhalation was modeled. * denotes a p-value <0.05 and ** denotes a p-value <0.005. Figure 2 . Size distribution of aerosol generating producers. Mean particle counts by micron per cubic feet of those procedures generating particles over the background reading with error bars representing the standard deviation. * denotes a p-value <0.05 and ** denotes a p-value <0.005. Pneumonia of unknown cause -China Recognition of aerosol transmission of infectious agents: a commentary Aerosol or droplet: critical definitions in the COVID-19 era. Int Forum Allergy Rhinol Endonasal instrumentation and aerosolization risk in the era of COVID-19: simulation, literature review, and proposed mitigation strategies Airborne Aerosol Generation During Endonasal Procedures in the Era of COVID-19: Risks and Recommendations Otolaryngol Head Neck Surg Recognition of aerosol transmission of infectious agents: a commentary Tissue distribution of ACE2 protein, the functional receptor for SARS coronavirus. A first step in understanding SARS pathogenesis