key: cord-0839569-d3kpwgsq authors: Nold, Julian; Metzger, Marc C.; Schwarz, Steffen; Wesemann, Christian; Wemken, Gregor; Pieralli, Stefano; Kernen, Florian; Weingart, Julia; Schirmeister, Carl G.; Schumann, Stefan; Schlager, Stefan; Spies, Benedikt C. title: Air seal performance of personalized and statistically shaped 3D-printed face masks compared with market-available surgical and FFP2 masks date: 2021-09-29 journal: Sci Rep DOI: 10.1038/s41598-021-98963-0 sha: 2ce818f8d5a49bdfe559c123315419abedbae388 doc_id: 839569 cord_uid: d3kpwgsq The ongoing COVID-19 pandemic has revealed alarming shortages of personal protective equipment for frontline healthcare professionals and the general public. Therefore, a 3D-printable mask frame was developed, and its air seal performance was evaluated and compared. Personalized masks (PM) based on individual face scans (n = 8) and a statistically shaped mask (SSM) based on a standardized facial soft tissue shape computed from 190 face scans were designed. Subsequently, the masks were additively manufactured, and in a second step, the PM and SSM were compared to surgical masks (SM) and FFP2 masks (FFP2) in terms of air seal performance. 3D-printed face models allowed for air leakage evaluation by measuring the pressure inside the mask in sealed and unsealed conditions during a breathing simulation. The PM demonstrated the lowest leak flow (p < 0.01) of inspired or expired unfiltered air of approximately 10.4 ± 16.4%, whereas the SM showed the highest (p < 0.01) leakage with 84.9 ± 7.7%. The FFP2 and SSM had similar values of 34.9 ± 18.5% leakage (p > 0.68). The developed framework allows for the time- and resource-efficient, on-demand, and in-house production of masks. For the best seal performance, an individually personalized mask design might be recommended. (EU) for N95/FFP2 masks. Only N95 respirators (corresponding to the FFP2 standard) and N99 respirators (corresponding to the FFP3 standard) meet the standards required for protection against aerosol and droplet transmission 14 . The air seal performance of standard masks has already been evaluated and compared in various studies, showing the importance of a proper fit 15, 16 . A universal one-piece facemask with a perfect fit seems improbable. This has been confirmed in in vivo investigations examining the fit of commercially available facemasks 8, 17 . However, further testing is needed to evaluate the seal performance of standard masks versus new mask shape designs. 3D printing has been widely applied in healthcare and medical research, spanning from bioprinting bone scaffolds for enhanced bone regeneration 18 to printing medical models for surgeons 19 . It offers the unique possibility of a cost-effective and individual manufacturing method for specific applications 20 . The aim of this study was to develop a workflow to be shared with the community to design and manufacture personalized 3D-printed masks based on individual face scans. Second, a mask design was created that allowed for large-scale production based on 190 face scans by means of statistical shape modeling. Third, the seal performance of the developed mask designs was compared with that of commonly available surgical masks and FFP2 respirators. Fourth, we evaluated the feasibility of the rapid 3D printing of face mask frameworks. The null hypothesis assumed that none of the masks used in the study would show differences in seal performance. Computer-aided design and computer-aided manufacturing of test faces. To allow for comparison ( Fig. 1) with individually designed masks, the faces of 12 employees were digitalized (FaceHunter, Zirkonzahn, South Tyrol, Italy), representing a wide variety of face shapes (Fig. 2) . In cooperation with 3D-LABS (St. Georgen, Baden-Württemberg, Germany), the surface files of the digitized faces were loaded into 3D modeling software (Geomagic Freeform, 3D Systems, Rock Hill, South Carolina, United States) to create the test faces for seal performance testing. This included adding an inward offset of 2 mm to create a rigid and printable design while retaining the dimensional accuracy of the original surface. Additionally, perforations in the nose (diameter: 7 mm) and mouth (width: 22 mm, height: 7 mm) were integrated. Tubes with an inner diameter of 7 mm for the nose holes and a tube with an inner diameter of 14 mm for the mouth with connectors at the inward end were incorporated. Furthermore, an adapter was designed to join the nose and the mouth tube connections to a single connection. This allowed for easy access of all tubes, which was needed for insertion of the pressure sensor tube (polyurethane tube, 4 × 2.5 mm, Sang-A Pneumatic Co., Daegu, Korea) through one side of the nose while retaining the ability to use both the nose and mouth for airflow testing. For the realistic fit evaluation of earloop-worn surgical masks, the area behind the ear was carved out to allow for a reliable retention spot (Fig. 3) . After completing the test face designs, all 12 face models were additively manufactured using a MultiJet Printer (HP 4200, HP, Palo Alto, California, U.S.). Statistical shape models for estimating the midfacial shape. For surface registration, statistical analysis and generation of the respective surface shapes, the mathematical/statistical platform R (R Core Team, 2020, Vienna, Austria) and the Rpackages Morpho, Rvcg (Schlager, 2017, Germany), mesheR (Schlager, 2015a, After a Procrustes alignment based on the registered meshes' vertices, the average shape was computed for both sexes. The averaging of these estimations yielded a sex-neutral face. For scaling to the appropriate size, the sex-neutral shape was then scaled to the average centroid size of each sex, resulting in two sizes, small and large. The calculated facial surfaces were imported into Rapidform software (Inus Technology, Seoul, South Korea) (Fig. 4) , and a forward offset of 1 mm was generated to create a volume out of the surface, which was finally saved as an STL file. Computer-aided design of the statistically shaped mask. For the computer-aided design (CAD) of the statistically shaped mask (SSM) frame, the STL data of the reference faces were imported into browser-based 3D CAD software (Tinkercad, Autodesk, California, USA) and aligned to the horizontal plane. A two-piece mask design was chosen, combining a mask frame with a horizontal tensioner that allows for easy mounting of the filter cloth. To create the frame part of the mask, an 8 mm thick vertical slice ranging from the glabella to the menton was cropped from the volume of the reference faces and served as a precursor (Fig. 5) . Rectangles (9 × 11 × 3.5 mm) were attached to the proximal side of the frame at the level of the labial angle for retention of the horizontal tensioner. To attach the filter cloth (Freudenberg, Weinheim, Germany) and buttonhole elastics (12 mm; Prym, Stolberg, Germany), 4 attachments consisting of grouped rectangles (3 × 8 × 10 mm and 3 × 7 × 10 mm) were connected to the outer surface of the frame. For the CAD of the tensioner, a rectangle (150 × 20 × 1.5 mm) was built and reinforced by another rectangle measuring 0.5 mm in height and 23 mm in length at both ends. Two rectangular holes (8 × 15 mm) attached to the referring retention on the frame were cut into the reinforced area. Finally, the data were exported as a STL file and imported to open-source slicing software (Cura 4.5, Ultimaker). For improved bed adhesion, a brim of 8 outlines was used, and the GCODE was generated for 3D printing (Fig. 7) . for creating the models used for measurement were selected for designing personalized face masks (PM). Eight personalized mask frames were designed following the same approach used for creating the statistically shaped masks. The same tensioner design was used with the personalized masks. Manufacturing and assembly of 3D-printed masks. For 3D printing of the mask frames, a polypropylene (PP) copolymer (Purell EP274P, LyondellBasell, Rotterdam, Netherlands, MFR = 15 g (10 min) −1 , T m = 165 °C, Vicat softening temperature (VST A50) = 142 °C, Young's modulus (3D printed) = 1000 MPa) was used, allowing a wide temperature range for the application. Due to the certified biocompatibility (ISO 10993 & class VI according to USP 88) and the requirements on heavy metal content and additives in close collaboration with pharmaceutical associations, irritation in case of direct contact of the PP with skin or wounds was widely excluded (Ph. Eur. 3.1.3 & 3.1.6, USP 661.1). A fused filament fabrication (FFF) printer (Ultimaker S5, Ultimaker B.V., Netherlands) was equipped with a steel nozzle (0.8 mm diameter) and fiber-reinforced PP adhesive tape (Scotch extreme packaging tape, 3 M, USA) as the print bed. Printing was performed at a nozzle temperature of 210 °C, build plate temperature of 35 °C, printing speed of 35 mm/s, wall thickness of 1.5 mm, layer height of 0.2 mm, and 100% infill. The application Figure 6 . For demonstration purposes, the original slice has been cut into a 2 mm proximal (yellow) and a 6 mm ventral (orange) part; the ventral part has been stretched ventrally (red) and aligned with the unstretched proximal part (yellow) prior to grouping. www.nature.com/scientificreports/ of support structures was not necessary. Up to four mask frames were printed simultaneously. After printing, the brim was removed from the frame. The horizontal retainer was prebent to an end-to-end distance of 80 mm (Fig. 8) . A cutting pattern was designed for each framework size, including the hole position needed for filter retention. After fixing the melt-blown filter cloth with its holes over the attachment feet and adding the tensioner, rubber bands were added, additionally securing the filter cloth. Seal performance evaluation and comparison with surgical and FFP2 masks. To measure the amount of leaked air, a known breathing-like airflow cycle was generated with a sine wave-like motion at 0.33 Hz and a stroke volume of 1500 ml. This was achieved via a purpose-built, computer-controlled linear motor actuated piston pump. The flow rate was measured via a Fleisch pneumotachograph (Type 2, Dr. Feyves & Gut, Hechingen, Germany), and the pressure was measured using a piezoresistive pressure sensor (Type 2, SI special instruments, Nördlingen, Germany). The sensors were calibrated on a daily basis using a portable calibrator (OM-DM 921, Onneken Mess-und Prüftechnik, Friedrichsdorf, Germany). The in-mask pressure was measured using a tube routed through the back of the 3D-printed test face and into the nose (Fig. 9) . The SSM and the PM were compared to surgical masks (SM, 3 M, Saint Paul, Minnesota, USA) and FFP2 respirator (FFP2, 3 M). Both types of masks are commercially available and in widespread use (Fig. 10) . In addition to the measurements obtained by placing the different masks on the printed faces, as shown in Fig. 10 , perfect seal conditions were measured by using a clay like material (Erkogum, Erkodent, Pfalzgrafenweiler, Germany) to eliminate leak flow past the mask, and the resulting pressure readings were recorded while running the same conditions used for the nonsealed tests. The pressure-flow relationships in the perfect seal measurements were used to determine the fluid mechanical conductance of the respective mask within the flow range of ± 800 ml/s. Assuming a simple model of two conductances in parallel, one representing the conductance of the mask filter cloth and one representing the conductance of the air leakage around the mask, we determined the percentage of leak flow from the pressure-flow relationship determined in the nonsealed measurements. Informed consent statement. Informed consent was obtained from all subjects involved in the study. Written consent was obtained for publication of the patient's picture shown in Fig. 12 . Average face based on sex-neutral shape models. The statistical shape modeling calculation resulted in a sex-neutral form with two sizes, S (height: 205 mm; width: 155 mm) and L (height: 216 mm; width: 163 mm). Table 1 ). The SM showed the statistically highest amount of leakage (p < 0.01). Mask-related leakage differences were comparable for inspiration and expiration. Mask design and 3D printability. The components of the mask were designed following a material-and time-efficient 3D printing process. It allowed for rapidly printing one mask and its retainer in under 60 min using the following printer settings: for PP: 0.8 mm nozzle, 0.2 mm layer height, 35 mm/s; for PLA: 0.4 mm nozzle, 0.3 mm layer height, 50 mm/s. The design is universally combinable with various 3D printing materials. www.nature.com/scientificreports/ Using PLA filaments allowed for support-free printing without reducing the success rate or print quality. For printing using PP filament, 5% of the used material was needed for improved bed adhesion in the form of a 4 mm brim surrounding the first layer of the print. In this study, personalized masks based on the individual face scans of 12 practitioners were created. Additionally, an average face contour was calculated using statistical shape modeling with 190 proband face scans. On these bases, a 3D printable statistically shaped mask was developed. To compare fitting properties, test faces were designed on the basis of the 12 practitioner face scans and 3D printed. The 3D printable mask designs were compared in terms of air seal performance with surgical mask and FFP2 respirator. The investigated masks and respirators showed significant differences in terms of air seal performance. Therefore, the null hypothesis had to be rejected. The personalized masks showed significantly better seal performance than all tested masks and respirators. The statistically shaped masks reached a seal performance comparable to that of the FFP2 respirator, while the surgical mask showed the lowest seal performance. The calculation of the statistical shape models of the midface was performed with the intention of overcoming the disadvantages of commercially available masks, which are often characterized by inadequate fitting, a certain degree of leakage and inconvenience. A uniform product design, which is predicted to fit onto 95% (Gaussian distribution) of the analyzed shapes, can be acquired by using statistical shape modeling. In medical image analysis, statistical shape modeling is used as a tool for modeling ubiquitous physical anatomical surface structures 21, 22 , e.g., for the reconstruction of missing anatomical structures or for the design of preformed implants 23, 24 . Geometric morphometrics is used in biological sciences to capture and quantify geometric information of biological structures, such as the shape variability of human bones or faces 25, 26 . This allowed for the calculation of the scaling factor needed for generating the two sizes of SSM. However, the difference between faces of different sizes and shapes (Fig. 11 ) cannot be completely overcome with one design. Therefore, personal mask designs, with the intention of maximizing seal performance, were developed. For optimal seal performance, the mask size and design were defined by the position of the menton and glabella of the calculated sex-neutral averaged face. While better seal performance can improve the effectiveness of a mask's filtering ability, it can also lead to pronounced discomfort by increasing the breathing resistance. This increase in discomfort could lead to less compliance with wearing a mask, as Ferng et al. concluded that it was one of two major factors for low adherence to mask wearing 27 . Therefore, differential pressure to sustain an Table 1 . Leakage during inspiration and expiration given for each mask. Groups with the same superscript letters (A, B, C) did not differ significantly. SD standard deviation, SM surgical mask, SSMS statistically shaped mask S, SSML statistically shaped mask L, PM personalized mask. www.nature.com/scientificreports/ acceptable level of breathing comfort should be carefully set by application of a suitable filter material with a low follow resistance while still increasing seal performance. To maximize the possibility for consumers to print mask frames even without the need for skin-compliant filaments, the design was created accordingly. This was archived by only allowing the filter cloth to have contact with the skin as it wraps around the mask frame, held by the horizontal tensioner on the inside and by the retention feet on the outside of the mask. Support structures, which are commonly used to prevent overhangs from deformation by gravity during printing, are not needed with the proposed design. This reduces the amount of wasted material and the postprocessing time, which is commonly not the case for publicly available printable mask designs. For optimal printing speed, easily printable shapes were used for the retention feet and the tensioner. Only a brim of 8 lines surrounding the first layer of the object was used to optimize bed adhesion and therefore increase printing success. Compared with another 3D-printed mask optimization design 28 that is meant to enhance wearing comfort, the goal of our design was focused on better seal performance. Bellus3D (Campbell, CA, USA), with their Mask Fitter 29 , followed the same approach. The suggested device is worn on top of a regular mask, which can also be done with the SSM design, although the use of a cut-to-size filter cloth is recommended. Due to the limited supply of medical-grade filter material, other materials could also be used with the PM and SSM designs. It has been shown that household materials are also capable of filtering relevant particle sizes 30 . Konda et al. showed that the filter performance of cotton, natural silk, and chiffon can provide protection, typically above 50% for particle sizes between 10 nm and 6.0 μm, while four-layer silk showed an average efficiency of > 85% for the same particle size range 31 . A PP copolymer with certified biocompatibility according to USP class VI and ISO 10,993 was used as a material for filament-based 3D printing to meet the highest medical standards 32 . The Purell PP used here features high durability, balanced flexibility (tensile modulus 1.0 GPa), allowing the frame to adapt to face contours, and chemical resistance to, for example, common disinfectants. Moreover, its high Vicat softening temperature of 142 °C (VST/A/50, ISO 306) allows it to be steam sterilizable while retaining its shape to enable reuse of the 3D-printed mask frames. Like all polyolefins, PP exhibits unrivaled low raw material and manufacturing costs, making it ideal for global use in medical applications that are accessible to all converters and end-users. The combination of a sustainable solvent-free polymerization process and simple recycling due to its hydrocarbon nature have led to excellent values in life cycle analysis 33 . The certified PP copolymer can be processed with all common FFF printers. Thus, it offers an excellent eco-friendly cost-benefit ratio, both for small decentralized applications and for high-volume production of protective masks or other personal protective equipment. The leak performance results of the tested masks revealed that the PM showed the lowest leak flow (p < 0.01) of inspired or expired unfiltered air of approximately 10.4 ± 16.4%. While the FFP2 classification implies that for filter cloth, > 94% of passing particles become trapped, we showed in our test that when placed on our face model, 34.9 ± 18.5% of air bypassed the cloth entirely. A comparable leakage was achieved with the SSM (p > 0.68). The SM showed the highest (p < 0.01) leakage, with 84.9 ± 7.7%. In most studies 16, 34 , seal performance tests are conducted by creating aerosol particles, which can then be quantified both in the surrounding air and inside of the mask. The goal of this study was to evaluate seal performance. Therefore, we relied on pressure and flow rate measurements. Furthermore, the use of 3D-printed faces allowed for repeatable results, eliminating inconsistencies by face movements. On the other hand, the lack of compressible skin might have resulted in a loss of seal performance. Nevertheless, from the authors' point of view, the repeatability outweighs this effect considering that the results are not meant to reflect real-world values but to allow for comparability. The mentioned effects could explain the higher leakage compared to Grinshpun et al., who, however, showed in accordance with our results that the FFP2 mask outperformed the surgical mask 16 . While Cai et al. showed that a personalized mask fitter can improve the contact pressure 28 , to our knowledge, no research has addressed the impact of personalized mask frames based on individual face scans and the resulting seal performance. The 3D printability of the mask offers three main advantages. First, fast and inexpensive production on demand is possible without a large stock capacity and complexity (e.g., expiring of stored materials), and the risk of uncontrolled material drain (e.g., waste, robbery) can be minimized. The use of filament-based 3D printing is particularly noteworthy, as finished parts can be printed in a one-step production process without the need for posttreatment as required for other additive manufacturing techniques. Second, in-house (quarantined private setting), in-hospital or on-campus production can be easily established if production and supply chains are disrupted, as recognized temporally around the world 35 . Third, the developed PM workflow can be utilized by every company around the world to create personalized masks for their employees at companies with 3D printing facilities. However, the designs cannot yet be recommended for clinical application since mask validation is mandatory according to defined standards (DIN EN 14683:2019-10, DIN EN 149:2009-08, ASTM F2100, NIOSH 42 CFR 84). Furthermore, the proposed approach of personalizing medical devices could be adopted to respirators for patients with severe trauma-induced swelling or those who require long-term ventilation. This could lead to improved seal performance and prevent uneven pressure that would result in damage to the patient's skin (Fig. 12) . Because of the present public health emergency, the material supply for healthcare workers and the general public needs to be realized quickly. 3D printing allows for easy and quick production of PPE, which has already been demonstrated by the wide adoption of additively manufactured face shields 36, 37 . To overcome the shortage of face masks, the use of alternative filtering materials and designs for protective face wear is needed 30 . Using 3D scanning software available for smartphones (STL Maker, Scandy LCC, New Orleans, USA) could also allow consumers to follow our workflow and design and print personalized mask frames for their enhanced protection. A workflow for designing 3D printable frameworks for an oronasal mask was created, offering a fast and inexpensive on-demand, in-house, or in-hospital production. The seal performance of personalized masks based on individual face scans was superior to that of all tested masks and the FFP2 respirator. Prior to application in a clinical setting, the mask performance has to be approved according to the Emergency Use Authorization (EUA) under section 564 of the Federal Food, Drug, and Cosmetic Act (FD&C Act) and Commission Recommendation (EU) 2020/403 from March 13, 2020. The datasets generated and/or analyzed during the current study are available from the corresponding author on reasonable request. Critical supply shortages-The need for ventilators and personal protective equipment during the Covid-19 pandemic World Health Organization. Rational Use of Personal Protective Equipment for Coronavirus Disease (COVID-19): Interim Guidance Department of Health and Human Services, Determination of Public Health Emergency Commission Recommendation (EU) 2020/403 of 13 March 2020 on Conformity Assessment and Market Surveillance Procedures Within the Context of the COVID-19 Threat Protecting healthcare staff from severe acute respiratory syndrome: Filtration capacity of multiple surgical masks Effect of fit testing on the protection offered by N95 filtering facepiece respirators against fine particles in a laboratory setting Effectiveness of N95 respirators versus surgical masks against influenza: A systematic review and meta-analysis Particle size-selective assessment of protection of European standard FFP respirators and surgical masks against particles-tested with human subjects Simulated workplace protection factors for half-facepiece respiratory protective devices Specification for Performance of Materials Used in Medical Face Masks DIN EN 14683:2019-10 Medical Face Masks-Requirements and Test Methods NIOSH 42 CFR Part 84 Respiratory Protective Devices Respiratory Protective Devices-Filtering Half Masks to Protect Against Particles-Requirements Escalating infection control response to the rapidly evolving epidemiology of the coronavirus disease 2019 (COVID-19) due to SARS-CoV-2 in Hong Kong Comparison of performance of three different types of respiratory protection devices Performance of an N95 filtering facepiece particulate respirator and a surgical mask during human breathing: Two pathways for particle penetration Do various respirator models fit the workers in the norwegian smelting industry 3D bioprinted bacteriostatic hyperelastic bone scaffold for damage-specific bone regeneration 3D-printing techniques in a medical setting: a systematic literature review Current and emerging applications of 3D printing in medicine Statistical shape models for 3D medical image segmentation: A review Chapter 2-Statistical deformation model: Theory and methods Virtual anthropology: Virtual anthrophology Orbital reconstruction: Prefabricated implants, data transfer, and revision surgery Geometric morphometrics Forensic age estimation in human skeletal remains: Current concepts and future directions Barriers to mask wearing for influenza-like illnesses among urban Hispanic households: Barriers to masks for influenza-like illnesses Customized design and 3D printing of face seal for an N95 filtering facepiece respirator Inc, Bellus3D Face Mask Fitter Simple respiratory protection-Evaluation of the filtration performance of cloth masks and common fabric materials against 20-1000 nm size particles Aerosol filtration efficiency of common fabrics used in respiratory cloth masks Pandemic-driven development of a medical-grade, economic and decentralized applicable polyolefin filament for additive fused filament fabrication Sustainability metrics: Life cycle assessment and green design in polymers Face seal leakage of half masks and surgical masks Responding to Covid-19-A once-in-a-century pandemic 3-D printed protective equipment during COVID-19 pandemic 3D printed face shields: A community response to the COVID-19 global pandemic The filament used in this study was donated by LyondellBasell. The test faces were designed and printed in cooperation with 3D-LABS (St. Georgen, Baden-Württemberg, Germany). The article processing charge was funded by the Baden-Wuerttemberg Ministry of Science, Research and Art and the University of Freiburg in the funding program Open Access Publishing. Open Access funding enabled and organized by Projekt DEAL. The authors declare no competing interests. Correspondence and requests for materials should be addressed to B.C.S. Publisher's note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations. License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http:// creat iveco mmons. org/ licen ses/ by/4. 0/.