key: cord-0838096-rde9jmyg authors: Shatz, Andrew C; Potvin, Richard title: Spectacle Independence and Quality of Vision After Bilateral Implantation of a Trifocal Intraocular Lens date: 2021-06-17 journal: Clin Ophthalmol DOI: 10.2147/opth.s318203 sha: 01183eefc908be38d9c30d98ac1ec362d7e06a04 doc_id: 838096 cord_uid: rde9jmyg PURPOSE: To evaluate the patient-reported spectacle independence and quality of vision for patients who were bilaterally implanted with a trifocal intraocular lens (IOL). DESIGN: Single site, prospective, single-arm study. METHODS: Subjects presenting for cataract surgery who had an interest in reducing their dependence on glasses at multiple distances were enrolled. Subjects were bilaterally implanted with a trifocal IOL and followed for 3 months post-surgery. A spectacle independence questionnaire and a quality of vision questionnaire were administered at the 3-month visit. The binocular visual acuity at distance (4 m), intermediate (60 cm) and near (40 cm) was also measured in the uncorrected and distance-corrected state. RESULTS: All but one subject reported no need for glasses for distance and intermediate work, and most (25/29) reported no need for glasses for near work. Almost all subjects reported never wearing glasses or wearing them only a little, and being able to function without glasses at any distance all or most of the time. Ninety percent of subjects (26/29) reported being completely or mostly satisfied with their vision overall, without the use of glasses or contact lenses. The mean binocular visual acuity 1 and 3 months after surgery was better than 0.1 logMAR (20/25) at all test distances in both the uncorrected and distance-corrected states. The reported quality of vision appeared similar to previously reported data for the same IOL. CONCLUSION: The trifocal IOL provided nearly all subjects with functional vision and spectacle independence for distance, intermediate and near work. It is a viable alternative for patients looking to reduce their overall dependence on spectacles after cataract surgery. Patients presenting for cataract surgery are often interested in options to reduce their dependence on spectacles for a variety of tasks, from distance vision to computer work (intermediate vision) and/or reading (near vision). Intraocular lens (IOL) options are available for a variety of patient needs. The major considerations for the patient are usually the distances at which they desire spectacle independence and whether the slightly increased potential for visual disturbances is likely to be a bother to them. Currently, the best option for those who desire good vision at distance, intermediate and near is likely to be a trifocal IOL, which provides 3 distinct foci at those distances. Trifocal IOLs have been shown to provide better intermediate vision than is obtained with a bifocal IOL design 1,2 and to provide better near vision than can be obtained with an extended depth of focus IOL. 3, 4 Early trifocal IOL designs were constrained by their optics to provide an intermediate focus that was half the dioptric distance of the near focus. For instance, the FineVision trifocal (PhysIOL, Liège, Belgium) design included a 1.75 D and 3.50 D intermediate and near focus respectively, corresponding to roughly 80cm for intermediate viewing and 40 cm for near viewing. 5 These values are similar to those reported for the AT Lisa tri IOL (Carl Zeiss Meditec, Jena, Germany), with a 1.66 D intermediate add and a 3.33 D near add. 6 The best intermediate focus for both of these trifocals is thus located further than where most patients place a computer or hold a cellphone or tablet; in the latter two cases "arm's length" is often a limiting factor. 7 With traditional trifocal designs, the only way the intermediate add could be increased was if there was a corresponding increase in the near add -an undesirable side effect as the optimal reading distance would be moved closer than 40 cm. The PanOptix ® Trifocal IOL (Alcon, Fort Worth, USA) is currently the only trifocal IOL approved by the FDA for use in the USA. It is fundamentally based on an aspheric diffractive quadrifocal IOL design, but the light at the first focal point is redirected to improve distance vision, making it a trifocal IOL in function. The benefit of this design is that it redistributes light energy in a unique manner to provide focal points at distance, intermediate (60 cm) and near (40 cm). 8 Studies of the PanOptix IOL have shown improved intermediate vision at 60 cm relative to the traditional trifocal IOLs described above. [9] [10] [11] Prior reported clinical results indicate that the lens provides a good range of vision (~ 3 D) with contrast sensitivity values in the normal range. [12] [13] [14] The purpose of the current study was to provide normative spectacle independence, quality of vision and visual acuity data related to bilateral implantation of the PanOptix trifocal IOL. This study was a single-site prospective single-arm study of clinical and subjective visual outcomes after bilateral implantation of a trifocal IOL. The study was approved by an institutional review board (Salus IRB, Austin, TX, USA). All subjects signed an appropriate informed consent document. The study was conducted in compliance with Good Clinical Practice (GCP), the tenets of the Declaration of Helsinki and International Harmonization (ICH) guidelines. The study was registered with clinicaltrials.gov (NCT04124952). Data are not available for sharing. Subjects over 40 years of age who were eligible for cataract surgery and interested in a reduced dependence on spectacles for near, intermediate and distance vision were considered for enrollment. They had to be considered appropriate candidates for trifocal IOL implantation (toric or non-toric) with no preoperative ocular pathology (eg corneal dystrophy, diabetic retinopathy), prior corneal or intraocular surgery, or any other pre-existing condition likely to confound the results of the study. All potential multifocal patients in the practice complete a lifestyle questionnaire and are individually counseled by the surgeon regarding the need for good lighting when reading, the need to manage dry eye and the potential to see halos around lights when driving at night for at least several months post-operatively. Candidates for any multifocal IOL must have 20/30 or better potential acuity in both eyes. Subjects were bilaterally implanted with the trifocal IOL on the same day and followed for 3 months after surgery, with refractive and VA data collected at both 1 and 3 months postoperative. Quality of vision and spectacle independence questionnaires were completed at the 3-month visit. All subjects were monitored for adverse https://doi.org/10.2147/OPTH.S318203 Clinical Ophthalmology 2021:15 events over the course of the study. The primary measure of interest was the reported level of spectacle wear for distance, intermediate and near vision, and overall, 3 months postoperative; this was based on responses to the Patient Reported Spectacle Independence Questionnaire, or PRSIQ. 15 Secondary measures included binocular visual acuities (uncorrected and best distance-corrected) at distance (4m), intermediate (60 cm) and near (40cm) at the 1-month and 3-month postoperative visits. The Quality of Vision questionnaire, a Rasch-scored questionnaire related to the frequency, severity, and degree of bother of various potential visual disturbances, was used to evaluate subjectively reported quality of vision. 16 All clinical data and questionnaire responses were tabulated in Microsoft Access (Microsoft, Redmond, WA, USA). Statistical analyses were performed using Statistica 12 (TIBCO Software Inc., Palo Alto, CA, USA). As a single-arm study, results were generally limited to descriptive statistics. Where appropriate, statistical testing of parametric variables was based on an analysis of variance (ANOVA), with a level of statistical significance of p ≤ 0.05. While a specific sample size for a single-arm study could not be calculated, a sample of 30 subjects (60 eyes) was considered reasonable to characterize the performance of this trifocal IOL. A total of 30 subjects were enrolled in the study, with all subjects bilaterally implanted with a non-toric or toric trifocal IOL between 10/2019 and 10/2020. One subject was lost to follow-up after successful bilateral IOL implantation (no 1-month or 3-month data available), leaving 29 subjects completing the study. One subject missed their 1 month visit due to a COVID-19 closure. Table 1 summarizes the demographic, preoperative and operative characteristics of the group. While not shown, mean IOL power was not statistically significantly different between the toric and non-toric eyes (p = 0.20). The PRSIQ results are summarized in Table 2 . As can be seen, only one subject reported needing glasses for distance work and one needed glasses for intermediate work. A significant majority (25/29, 86%) reported no need for glasses for near work. Almost all subjects reported never wearing glasses or wearing them only a little, and being able to function without glasses at any distance all or most of the time. Ninety percent of subjects (26/29) reported being completely satisfied or mostly satisfied with their vision without glasses or contact lenses overall. The two least-satisfied subjects were both observed to have dry eye postoperatively, with variable visual acuity and refractive results. The Quality of Vision questionnaire measures the reported frequency, severity and degree of bother associated with 10 different potential visual disturbances, then provides an aggregate Rasch-scored (1-100 scale) result for each. Figure 1 shows the distribution of aggregate results for these three patient-reported characteristics. Table 3 summarizes the reported frequency, severity, and degree of bother for haloes, glare, and starbursts. Haloes were the most frequently reported visual disturbance, with 41% of subjects (12/29) rating them as "moderate" or "severe". Haloes were also rated as the most bothersome; 20% (6/29) of subjects reported halos as "quite" or "very" bothersome. Glare and starbursts were the next most frequent disturbances noted but were reported less frequently and as less bothersome than haloes. The higher outliers were associated with the two subjects with dry eye, identified above. The mean monocular spherical equivalent refraction at 3 months was 0.06 ± 0.53 D, while the mean residual refractive cylinder was 0.49 ± 0.36 D. Mean binocular visual acuity results at 1 month and 3 months are shown in Figure 2 (uncorrected) and Figure 3 (distance-corrected) . The mean visual acuity was better than 0.1 logMAR (20/ 25) at all distances in both the uncorrected and distancecorrected states at both time points. There was no statistically significant difference in any VA measure between 1 and 3 months postoperative. At the 3-month visit, binocular distance-corrected acuity was slightly better (less than half a line) than uncorrected acuity at 4 m (p < 0.01) but not statistically significantly different at 60 cm (p = 0. 16) or 40 cm (p = 0.33). There was one unrelated non-serious adverse event reported for one subject during the study. They had a bicycle accident with minor injuries, none of which involved the eye. There was no apparent effect on their The current study was designed to collect spectacle independence, subjective quality of vision and visual acuity data for the only trifocal IOL currently approved for use in the USA, with an emphasis on the degree of spectacle independence achieved. Previous studies have used different tests of patient functional vision and spectacle independence, or different scoring, so it is difficult to directly compare results. However, overall spectacle independence reported by Donmez et al for the same trifocal IOL was 94%, 17 while Cochener reported 90%; 3 both results appear consistent with findings in the current study. The highest reported need for spectacles at any distance in the current study (14%) was slightly lower than reported in the study performed to obtain FDA approval for the lens (19.5%), 18 indicating somewhat better results in the current study. The distance at which spectacles were needed most often was 40 cm (near vision) in both studies. In the current study, 90% of subjects reported that they were completely or mostly satisfied with their vision overall, which appears consistent with the findings of Garcia-Perez et al 19 and Rementeira-Capelo et al 4 , though the questionnaires used in all 3 studies were different. At present, there is no standardized questionnaire related to spectacle Abbreviation: PRSIQ, patient reported spectacle independence questionnaire. independence and satisfaction. We used the PRSIQ because it is very short, with relevant questions and is easy for patients to understand. In general, overall satisfaction with vision appears higher with trifocal IOLs than with extended depth of focus IOLs, primarily due to the better near vision. 4 Quality of vision scores reported here were slightly higher (worse) than results from a recent study in Norway, 10 but very similar to recently reported results in the USA 20 and somewhat lower than reported results from a study in Portugal, 21 all of which included the same questionnaire and IOL. It appears that responses may reflect different patient attitudes and expectations in different regions. This suggests that comparisons of questionnaire data from different regions may be less reliable. While relative values varied, in all four studies haloes were the most frequently reported visual disturbance and reported as the most bothersome. The quality of vision scores here are also similar to those reported for other trifocal IOLs, and generally equivalent to or better than those reported with extended depth of focus (EDOF) IOLs. 9, 10, 21 In all studies patient satisfaction with the trifocal IOL is high, indicating that there is an appreciated tradeoff between visual disturbances and good near vision. As noted in the results, the outliers in the quality of vision data collected in the current study were associated with subjects who had postoperative dry eye; their reported satisfaction was also the lowest. This reiterates the need to identify and aggressively treat dry eye before and after cataract surgery with advanced technology IOLs. One recent study indicated that the cause of dissatisfaction DovePress with presbyopia correcting IOLs could be traced to dry eye in 35% of subjects. 22 Appropriate management of dry eye postoperatively can improve results in many patients. 23 The binocular visual acuities reported here appear consistent with previous findings for the same IOL. 4, 5, 24 Several studies have found, as in the current study, that intermediate VA was slightly worse than for both distance and near VA, but the relative differences reported were generally small (less than one line). 17, 22, 25 There are limitations to the current study. It was conducted at a single site with a relatively small sample size. Objective quality of vision data such as contrast sensitivity or low contrast acuity were not collected. Finally, the study was a single-arm study, so no comparative data relative to other IOLs were collected. In summary, the trifocal IOL evaluated in this study appears to be a good alternative for appropriate patients interested in reducing their overall dependence on spectacles for distance, intermediate and near work. As with all multifocal IOLs, potential patients should be advised of the increased likelihood of visual disturbances such as halos, relative to implantation of a monofocal IOL. The management of dry eye is also important. Trifocal versus bifocal diffractive intraocular lens implantation after cataract surgery or refractive lens exchange: a meta-analysis Comparison of visual quality and subjective outcomes among 3 trifocal intraocular lenses and 1 bifocal intraocular lens A comparative evaluation of a new generation of diffractive trifocal and extended depth of focus intraocular lenses Real-world evaluation of visual results and patient satisfaction for extended range of focus intraocular lenses compared to trifocal lenses Visual outcomes and subjective experience after bilateral implantation of a new diffractive trifocal intraocular lens Evaluation of visual outcomes and patient satisfaction after implantation of a diffractive trifocal intraocular lens Viewing distance and eyestrain symptoms with prolonged viewing of smartphones First implantation of a diffractive quadrafocal (trifocal) intraocular lens Comparison of visual and refractive outcomes of 2 trifocal intraocular lenses Trifocal intraocular lenses: a comparison of the visual performance and quality of vision provided by two different lens designs Visual and refractive outcomes, spectacle independence, and visual disturbances after cataract or refractive lens exchange surgery: comparison of 2 trifocal intraocular lenses Visual function after implantation of a presbyopia-correcting trifocal intraocular lens Clinical outcomes with a diffractive trifocal intraocular lens Results of a clinical evaluation of a trifocal intraocular lens in Japan Patient-Reported Spectacle Independence Questionnaire (PRSIQ): development and validation The development of an instrument to measure quality of vision: the Quality of Vision (QoV) questionnaire Patients satisfaction and clinical outcomes of binocular implantation of a new trifocal intraocular lens Visual and patient-reported outcomes of a diffractive trifocal intraocular lens compared with those of a monofocal intraocular lens Short term visual outcomes of a new trifocal intraocular lens Visual acuity, quality of vision, and patient-reported outcomes after bilateral implantation with a trifocal or extended depth of focus intraocular lens Through-focus vision performance and light disturbances of 3 new intraocular lenses for presbyopia correction Causes and correction of dissatisfaction after implantation of presbyopia-correcting intraocular lenses Evaluation of quality of life after implantation of a new trifocal intraocular lens Refractive and visual outcomes after bilateral implantation of a trifocal intraocular lens in a large population A comparative study of the visual outcomes between a new trifocal and an extended depth of focus intraocular lens Key topics include: Optometry; Visual science; Pharmacology and drug therapy in eye diseases; Basic Sciences; Primary and Secondary eye care; Patient Safety and Quality of Care Improvements. This journal is indexed on PubMed Central and CAS, and is the official journal of The Society of Clinical Ophthalmology (SCO). The manuscript management system is completely online and includes a very quick and fair peer-review system Victoria Wiltshire, RN, MBA, of SightTrust Eye Institute provided data collection and study administration support for this study. This research was conducted with support from Alcon, in the form of an investigator-initiated study grant (IIT # 51718703). Richard Potvin is a consultant to Alcon and Carl Zeiss Meditec. The authors report no other conflicts of interest in this work.