key: cord-0834995-biqgiikc authors: Dudley, Dawn M.; Newman, Christina M.; Weiler, Andrea M.; Ramuta, Mitchell D.; Shortreed, Cecilia G.; Heffron, Anna S.; Accola, Molly A.; Rehrauer, William M.; Friedrich, Thomas C.; O’Connor, David H. title: Optimizing direct RT-LAMP to detect transmissible SARS-CoV-2 from primary nasopharyngeal swab samples date: 2020-12-31 journal: PLoS One DOI: 10.1371/journal.pone.0244882 sha: a24ea902006c65102c4bb6a1a6318b98c05dca25 doc_id: 834995 cord_uid: biqgiikc SARS-CoV-2 testing is crucial to controlling the spread of this virus, yet shortages of nucleic acid extraction supplies and other key reagents have hindered the response to COVID-19 in the US. Several groups have described loop-mediated isothermal amplification (LAMP) assays for SARS-CoV-2, including testing directly from nasopharyngeal swabs and eliminating the need for reagents in short supply. Frequent surveillance of individuals attending work or school is currently unavailable to most people but will likely be necessary to reduce the ~50% of transmission that occurs when individuals are nonsymptomatic. Here we describe a fluorescence-based RT-LAMP test using direct nasopharyngeal swab samples and show consistent detection in clinically confirmed primary samples with a limit of detection (LOD) of ~625 copies/μl, approximately 100-fold lower sensitivity than qRT-PCR. While less sensitive than extraction-based molecular methods, RT-LAMP without RNA extraction is fast and inexpensive. Here we also demonstrate that adding a lysis buffer directly into the RT-LAMP reaction improves the sensitivity of some samples by approximately 10-fold. Furthermore, purified RNA in this assay achieves a similar LOD to qRT-PCR. These results indicate that high-throughput RT-LAMP testing could augment qRT-PCR in SARS-CoV-2 surveillance programs, especially while the availability of qRT-PCR testing and RNA extraction reagents is constrained. There are more than 13.8 million reported severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) infections in the United States as of December 3rd, 2020 (https://www.cdc.gov/ coronavirus/2019-ncov/cases-updates/cases-in-us.html). The actual number of infections is likely far greater since testing remains limited. Asymptomatic individuals contain similar levels Residual, completely de-identified NP swab samples were provided by the University of Wisconsin-Madison Hospitals and Clinics (UWHC) and the Wisconsin State Laboratory of Hygiene (WSLH) under biosafety protocol B00000117 (IRB 2016-0605), and their use was not considered human subjects research by the University of Wisconsin-Madison School of Medicine and Public Health's Institutional Review Board. Samples were not specifically collected for this study. Samples were collected into a variety of transport media including universal transport media (UTM), viral transport media (VTM), and phosphate-buffered saline (PBS), stored at 4˚C for up to 7 days, and transported to the laboratory at room temperature. Upon arrival at the laboratory, samples were stored at either 4˚C (for immediate same-day use) or -80˚C until use in RT-PCR or RT-LAMP assays. Viral load analysis was performed after samples arrived in our laboratory. RNA was isolated using the Viral Total Nucleic Acid kit for the Maxwell RSC instrument (Promega, Madison, WI) following the manufacturer's instructions. Viral load quantification was performed using a qRT-PCR assay developed by the CDC to detect SARS-CoV-2 (specifically the N1 assay) and commercially available from IDT (IDT cat # 10006770) (Coralville, IA) [28] . The assay was run on a LightCycler 96 or LC480 instrument (Roche, Indianapolis, IN) using the Taqman Fast Virus 1-step Master Mix enzyme (Thermo Fisher, Waltham, MA). The LOD of this assay is estimated to be 200 genome equivalents/ml in swab media. To determine the viral load, samples were interpolated onto a standard curve consisting of serial 10-fold dilutions of in vitro transcribed SARS-CoV-2 N gene RNA kindly provided by Nathan Grubaugh (Yale University). The experiments we describe here were modified from the SARS-CoV-2 RT-LAMP assay developed by Zhang et al. [13] . We used fluorescent-based detection with Warmstart LAMP reagents and the included fluorescent dye (New England Biolabs, NEB cat # E1700L). We tested primer sets developed in previous studies targeting several SARS-CoV-2 genes as shown in S1 Table [13, 14, 22, [29] [30] [31] [32] [33] . Of note, the Color-Orf1a primers and Lamb-Orf1a primers are identical but were used at different concentrations per the protocols developed by each originating lab. The final 1X primer concentrations are listed in S1 Table. For each reaction, a 10X stock of all 6 primers was combined with Warmstart mastermix and water in 25μl reactions following the manufacturer's recommendations. Unless otherwise stated, 1μl RNA transcript of the SARS-CoV-2 N-gene obtained by Dr. Nathan Grubaugh (S2 Table) volume of direct swab sample addition for highest RT-LAMP efficiency, a 60-cycle program with data collection every 20 seconds was used. The specificity of the primers tested in this manuscript were established in previous publications [13, 14, 22, 29, [31] [32] [33] . Briefly, the Gene-N-A primers, used in most of the experiments in this manuscript, did not cross-react with SARS-CoV-1 N-gene by RT-LAMP and were not anticipated to cross-react with other common respiratory pathogens based on sequence comparisons (communication with Nathan Tanner). The Gene-N2, Gene-E1, and As1e primers were aligned against other coronaviruses and respiratory pathogens and the only pathogen with >80% identity to SARS-CoV-2 in more than one of the 6 primers required for RT-LAMP was SARS-CoV-1 (communication with Nathan Tanner and Brad Langhorst), which hasn't circulated in humans since 2004. The Color Genomics primers were tested for cross-reactivity against 51 organisms for EUA approval and were not cross-reactive to any organism tested [33] . A subset of samples were treated with LucigenQE RNA Extraction Solution (Lucigen, Middleton, WI) in a 1:1 ratio as described in Ladha et al. [34] . Briefly, NP swab media was combined with an equivalent volume of LucigenQE and briefly vortexed. Samples were then incubated for 5 minutes at 95˚C, cooled on ice, and maintained until the addition to the RT-LAMP reaction. To assess improvement in quantification cycle (Cq) values using sample lysis with LucigenQE or RNA isolation, mean Cq values were calculated for each sample. Mean Cq values were not normally distributed for either dataset so we used a nonparametric equivalent to a paired ttest, the Wilcoxon signed rank test with continuity correction, for each set of paired samples. To examine whether sample vRNA load and/or treatment were significantly associated with a positive RT-LAMP result, we used logistic regression with the RT-LAMP result as the dependent variable. For our analysis, an equivocal result in which one replicate was positive while the other was negative, was conservatively treated as negative. We coded RT-LAMP results for each sample tested by each method as a dichotomous outcome with positive samples coded as "1" and negative or equivocal samples coded as "0". For explanatory variables, we chose qRT-PCR vRNA load, with samples greater than 10 3 copies/μl coded as "1" and samples less than 10 3 copies/μl coded as "0", and group, designated as either 5μl RNA, 1μl lysed, or 1μl direct addition. All statistical analyses were performed in RStudio (v. 1.2.1335) using R (v. 3.6.0) [35] . To determine a limit of detection for the RT-LAMP assay, serial 10-fold dilutions of RNA transcripts containing the N-gene were tested in RT-LAMP reactions with Gene-N-A primers in duplicate in 3 independent assays. RNA transcripts were diluted in RNase-free water. Consistent detection of RNA was achieved when 1x10 3 copies or greater of RNA/μl was added into the reactions (Fig 1A) . To obtain a more precise LOD, the transcript was diluted 1:2 starting at 5x10 3 copies/μl down to 78 copies/μl and each concentration was run in 10 replicates. Nine of ten replicates at 625 copies/μl were positive, while 6/10 were positive at 312 copies/μl and 5/10 were positive at 156 copies/μl ( Fig 1A) . Thus, we can consistently detect 625 copies of input into the reaction but can detect down to 156 copies of input in half of the reactions. Zero reactions were detected as positive at 100 copies/μl or below. Residual NP swab samples from 106 patients with diagnosed SARS-CoV-2 and 31 negative samples were tested directly by RT-LAMP in duplicate. Additionally, RNA was isolated from the positive samples and tested by qRT-PCR with a transcript standard for quantitation. A total of 63/106 (59%) samples tested positive by RT-LAMP and 106/106 by qRT-PCR (S3 Table) . Another 13 samples were equivocal by RT-LAMP, with one of two replicates positive. As shown in Fig 1B, the LOD of primary samples was similar to that seen with the RNA transcript. For the rest of the analysis, we focus on the ability of this assay to detect samples with a viral load of > 1x10 3 copies/μl as a conservative LOD based on our transcript LOD of 625 copies/μl. 63/78 (81%) samples with viral RNA copy numbers greater than 1x10 3 copies/μl were detected by RT-LAMP, whereas 0/28 samples with concentrations <1x10 3 copies/μl were detected positive and 3/28 were equivocal. All 31 samples that tested negative for SARS-CoV-2 by clinical laboratories also tested negative by RT-LAMP (S3 Table samples To determine whether adding larger volumes of primary NP swab samples could improve the sensitivity of the assay, 1, 3, and 5μl of swab samples from 5 primary SARS-CoV-2-positive samples were tested side-by-side. All replicates were detected as positive when 1μl was added directly into the RT-LAMP reaction ( Fig 1C) . However, one of the two replicates from two samples tested negative when 3μl of the sample was added and both replicates from one sample tested negative when 5μl of the sample was added into the RT-LAMP reaction. Furthermore, Cq thresholds were higher with the addition of higher volumes of sample. Therefore, we chose to use 1μl of straight swab sample in subsequent experiments. To determine whether treatment with lysis buffer improves the sensitivity of RT-LAMP to detect SARS-CoV-2 in clinical samples without the use of traditional nucleic acid isolation methods, we treated 72 clinical samples with a range of SARS-CoV-2 vRNA loads in a 1:1 ratio with LucigenQE as described by Ladha et al. [34] . We then compared the fluorescent RT-LAMP Cq values between 1μl of lysed sample and 1μl of the same samples added directly. The addition of 1μl of NP swab eluate directly into the RT-LAMP reaction resulted in positive detection in both replicates of 46/72 (64%) known SARS-CoV-2-positive samples and in 1 of 2 replicates in 6 additional samples (Fig 2A, S3 Table) . Treatment of the same 72 samples with LucigenQE resulted in detection of both replicates for 56/72 (78%) samples, an additional 10 samples that were undetectable before. An additional 4 samples that were negative when tested directly were equivocal when treated with LucigenQE. Focusing on samples above 1x10 3 copies/μl, 45/53 (85%) samples were positive without extraction while 53/53 (100%) were positive with LucigenQE extraction. Without RNA extraction an additional 5 samples were equivocal indicating that extraction resulted in more consistent detection of positive samples in both replicates above the 1x10 3 copies/μl threshold. While 0/19 samples with viral loads below 1x10 3 copies/μl were positive with 1μl straight, 4 of 19 samples were detectable after lysis in both replicates and 4 additional samples were detected in 1 of 2 replicates. Mean Cq for Luci-genQE-treated samples were significantly lower (Cq = 38.62) than those for directly added samples (Cq = 49.08) (Wilcoxon signed rank test, V = 1830, p = 1.67E -11 ), suggesting that lysis treatment improves the efficiency of amplification in the LAMP reaction (Fig 2A) . We also examined whether sample vRNA load and/or treatment were significantly associated with RT-LAMP detection results. We found that direct addition of 1μl of untreated NP swab was associated with a decreased odds of detecting a positive RT-LAMP result (OR = 0.20, 95% CI = 0.04-0.65, p = 0.015) while the most important factor associated with a positive result was a sample vRNA load of greater than 1x10 3 copies/μl (OR = 88.5, 95%CI = 25.74-434.87, p = 1.88E -10 ). One of the primary reasons the direct RT-LAMP assay is less sensitive than the diagnostic qRT-PCR assay is because the qRT-PCR assay uses 5μl of concentrated and purified RNA as input. To determine whether the RT-LAMP assay would perform to a similar level of detection if the same input was used, 5μl of purified RNA was used in RT-LAMP assays from a subset (n = 44) of COVID-19-positive NP swab samples. The starting sample viral loads ranged from 1.01x10 1 to 1.14x10 7 copies/μl. 25 of 44 had concentrations of virus above the RT-LAMP LOD of 1x10 3 vRNA copies/μl. Of the 44 samples tested with 1μl of the direct swab, 18 tested positive (43%) and 7 were equivocal between replicates (Fig 2B and S3 Table) . When 5μl of purified RNA was used in the reactions instead, 42 of 44 samples (95%) tested positive (Fig 2B, S3 Table) . Note that the process of RNA isolation used in this experiment concentrated the RNA approximately 3-fold from the concentration presented in Fig 2B and that 5μl of that concentrated RNA was used, while 1 μl of the starting sample concentration was used in the 1μl straight reactions. Samples with 100-fold lower vRNA copies/μl were detected after RNA isolation and concentration by RT-LAMP. Adding purified RNA brings the possible LOD of detection down to 150 copies of input of a 10 copies/μl sample, which was detected in 4 of 6 primary samples tested within this viral load range. This is similar to the detection of 156 copies of transcript as shown in Fig 1A. Of the 25 samples with viral loads over the 1x10 3 copies/μl threshold, 18 (72%) were positive when adding straight while 25/25 (100%) were positive with RNA isolation. Overall, the direct addition of 1μl NP swab was associated with reduced odds of a positive RT-LAMP result (OR = 0.0054, 95%CI = 0.00023-0.041, p = 2.56E -05 ). Similar to the results for lysis buffer treatment, samples with qRT-PCR vRNA loads greater than 1x10 3 copies/μl had significantly increased odds of a positive RT-LAMP result (OR = 49.35, 95%CI = 8.22-966.29, p = 0.00044). Lastly, the Cq values were lower for all detected samples when 5μl of RNA was added (mean Cq = 31.42) instead of 1μl straight sample (mean Cq = 58.72), indicating faster and more robust detection with concentrated and purified RNA (Wilcoxon signed rank test, V = 903, p = 1.71E -08 ) (Fig 2B) . Multiple SARS-CoV-2 RT-LAMP primer sets have been published or included in manuscripts on preprint servers since we began our experiments. We compared the efficiency of several alternative primer sets either alone or in combination with the Gene-N-A primers used in our initial studies (S1 Table) . Two primary NP swab samples with high concentrations of SARS-CoV-2 (NP1:1.09x10 9 copies/μl, NP2:4.28x10 7 copies/μl) as well as gamma-irradiated SARS-CoV-2 (BEI) were used to screen different primer combinations using the same fluorescent RT-LAMP conditions. First, a set of primers previously established [14, 30] and used to obtain an FDA EUA by Color Genomics was tested with each primer alone and in different combinations (Fig 3A) . Several primers and primer combinations resulted in a lower Cq value across the board than the Gene-N-A gene primers suggesting improved reaction efficiency. The primer combinations including Color-N/Color-E and Color-N/Color-ORF1a yielded the lowest Cq values in all samples. Next, we compared the Gene-N-A, Color-N, Color-E, and Color-N/E combination to second-generation primers described in Zhang et al. [22] (Gene-N2, Gene-E1 and As1e). We found that As1e, originally published by Rabe et al. [29] , yielded the lowest Cq value followed closely by a combination of As1e with the two primers targeting the Gene-N2 and Gene-E1 genes designed by Zhang et al. (Fig 3B) . The Color-N primer yielded similar Cq values to the triple combination primer set from Zhang et al. We also tested the Color-N, Color-E1, and Gene-N-A gene primer sets against additional published primers that target ORF1a (Lamb, Yu, El-Tholoth, and Zhang primers) either alone or in combination and compared them to the Gene-N-A gene primer set (S1 Table) . The Color-N primer produced the lowest Cq value in this set (S1 Fig). We then tested As1e, Color-N, As1e/Color-N/Gene-E1 primer set with and without guanidine hydrochloride, as recommended by Zhang et al., with Twist RNA SARS-CoV-2 template [22] . Under these conditions, guanidine hydrochloride improved detection with most primer sets with the exception of the Gene-N-A primer set (Fig 3C) . Using the Twist RNA, the primer set that detected samples at the lowest dilutions was the Color-N primer or combination of As1e/Color-N/Gene-E1 with guanidine hydrochloride, though only in one of two replicates. The As1e primer with and without GuHCl as well as the combination As1e/Color-N/Gene-E1 with GuHCl often yielded the lowest Cq value. Lastly, to determine which primer set or combination worked best with primary NP swab samples, fourteen samples with viral loads ranging from 7.33 x10 4 -1.52 x10 8 vRNA copies/μl were tested using 1μl of straight swab sample with the most promising primer combinations with and without GuHCl. Both the Color-N primer set alone and the As1e/Color-N/Gene-E1 combination yielded the lowest Cq value across the different levels of virus (Fig 4A) . Guanidine hydrochloride did not improve detection in primary samples as seen with the Twist RNA. Twelve additional primary NP swab samples with high Ct values ranging from 25 to 35 were tested with the Color-N and As1e/Color-N/Gene-E1 combination. Detection of these samples was very similar between the two primer conditions (Fig 4B) . Note that several samples were equivocal at Ct 25 for both primer sets equivalently and that we might expect better detection at that threshold. Since we do not eliminate inhibitors in the samples and since these samples were stored for several days and freeze-thawed between the time the hospital ran the sample to acquire the Ct value and the time the sample was run in our assay, some of the replicates became undetectable. We would expect to more consistently detect virus at a Ct of 25 in both replicates in fresh samples. Frequent, widespread testing is considered the best mitigation strategy to control SARS-CoV-2 before a vaccine or effective therapy becomes widely available in 2021. Traditional qRT-PCR is sensitive but is time-consuming and reliant on very specific reagents that are in short supply. RT-LAMP has become a promising alternative to qRT-PCR, but the sensitivity of this assay has been poorly characterized from primary NP swab samples when added directly into the reaction. Many published studies establish a LOD based on free RNA transcript or isolated RNA from cultured virus of around 100 copies/rxn. These LODs apply to RT-LAMP only when purified RNA is used as input. They do not apply to direct RT-LAMP methods containing whole virions in primary samples, which also contain host enzymes and other host components. In this study, we established that 625 copies/reaction was necessary in order to detect RNA transcripts consistently in 9/10 reactions in our assay. In primary samples we rarely detected virus in samples with a vRNA load of less than 5x10 2 -1x10 3 vRNA copies/μl, establishing this threshold as a conservative LOD. While not as sensitive as methods with RNA extraction first, this LOD range is sufficient for SARS-CoV-2 surveillance to detect virus in individuals with the minimal amount of virus necessary to isolate the virus and therefore most likely to transmit the virus [1, [36] [37] [38] . The LOD of 1x10 3 copies/μl is likely due to inefficiencies associated with virus lysis at 65˚C during the LAMP reaction and possible degradation of liberated RNA by enzymes, including RNases, present in primary samples. Indeed, adding more sample, including more host enzymes and media with potential inhibitors, reduced detection. On the other hand, RNA extraction eliminates these inhibitors and allows 100% consistent detection above 10 copies/μl (150 copies/reaction). Lysis with LucigenQE was compatible with RT-LAMP and improved our sensitivity to detect SARS-CoV-2 in primary samples with vRNA loads less than 10 3 copies/μl. When compared with direct addition, we were able to increase our detection of true positives for both replicates from 61% to 78% after lysis treatment. In samples with viral loads above 1x10 3 copies/μl we improved detection to 100%. Guanidine hydrochloride has also been shown to improve sensitivity in other studies and while our results showed better detection with synthetic Twist RNA, bringing the LOD down to 62.5 copies/μl, the same improvement was not observed in primary samples in our hands. The largest increase in sensitivity occurred with RNA isolation prior to RT-LAMP using an alternative RNA isolation method to those approved for the CDC qRT-PCR assay. With RNA isolation we detected 95% of the samples detected by qRT-PCR, including those with the lowest viral loads. There are now many primers available targeting different regions of SARS-CoV-2. In this work, we tested several sets to iteratively choose which primer set worked best with primary NP swab samples. We found that several primer sets performed more efficiently than the Gene-N-A primers we used in most of our experiments. The two best performing primer sets that were nearly indistinguishable in performance with both high and low viral load primary samples were a combination of As1e/Color-N/Gene E1 and the Color-N primer set alone. For this study, we chose to use fluorescent RT-LAMP for detection rather than colorimetric detection. Using fluorescence enabled analysis of the differences in the Cq values providing a quantitative evaluation of how each condition changed the efficiency of the assay. However, when considering how to deploy RT-LAMP in the field, our group has developed a mobile RT-LAMP workflow that uses saliva and colorimetric readouts for low cost and portability [39] . Additional work needs to be done to determine whether the benefits of fluorescent detection can be inexpensively migrated to decentralized point-of-need testing. Many studies are comparing RT-LAMP to RT-PCR results presented as Ct value, rather than vRNA copies/μl. We chose to focus on comparing methods and determining LODs based on a standard qRT-PCR assay performed in our laboratory with a quantitative standard, rather than Ct values generated by the varying sources of our primary samples. Diagnostic laboratories have transitioned between different methods as reagents were available and as new assays became available, which means that the Ct values can vary. By comparing all our results to the vRNA copies/ml that we generated in our lab using a consistent primer set and protocol (CDC qRT-PCR assay) that targeted the same gene (and primers) as our SARS-CoV-2 RT-LAMP assay, we were able to ensure our comparisons were consistent across all samples. Overall, we have shown that direct RT-LAMP using fluorescent detection can detect SARS-CoV-2 in primary NP swab samples with viral loads greater than 1x10 3 vRNA copies/μl. We were able to improve this slightly with the quick and low-cost addition of Lucigen lysis buffer to the reaction and could detect 100% of samples above the 1x10 3 copies/μl threshold. We also saw improvement in efficiency with several alternative primer sets. While direct RT-LAMP is not as sensitive as qRT-PCR, it is sufficient to detect the levels of virus that are necessary to culture virus from a sample. This means that this assay detects people who are likely to transmit the virus for a significant reduction in cost, time, and reagents relative to qRT-PCR. Direct RT-LAMP costs approximately $7 per sample while qRT-PCR costs $25. RT-LAMP from setup to results can be performed in 60-90 minutes depending on the scale of the assay (more samples take longer to set up on the front end, but the reaction time is 30 minutes), while qRT-PCR requires 60 minutes for RNA extraction, 30 minutes to set-up qRT-PCR, and an additional 90 minutes for the qRT-PCR run to finish. One proposed way to utilize this test could be to surveil large numbers of individuals who are then directed toward diagnostic testing by qRT-PCR if they test positive by this test, significantly reducing the burden on diagnostic labs and their resources. This test, if transitioned to a colorimetric version, could be set up at point-of-need sites with minimal equipment and could provide same-day test results to individuals in a work or school environment. For this test to be used in a clinical or diagnostic capacity, additional clinical validation would be required. Overall, direct RT-LAMP is an important addition to the repertoire of currently available tests to identify samples containing SARS-CoV-2 nucleic acids. Samples that were not detectable were plotted on the ND line set at Cq 80, the hißghest cycle number in our assay. (TIF) S1 Table. qRT-PCR N-gene viral loads and RT-LAMP Cq values from 106 primary NP swab samples run in duplicate with 1ul of swab sample or a subset of NP swab samples run in duplicate with either 1μl of primary samples treated with Lucigen QuickExtract or 5μl of extracted vRNA. (DOCX) Presymptomatic SARS-CoV-2 Infections and Transmission in a Skilled Nursing Facility Transmission of 2019-nCoV Infection from an Asymptomatic Contact in Germany CT imaging and clinical course of asymptomatic cases with COVID-19 pneumonia at admission in Wuhan Suppression of a SARS-CoV-2 outbreak in the Italian municipality of Vo'. Nature SARS-CoV-2 Viral Load in Upper Respiratory Specimens of Infected Patients Clinical Course and Molecular Viral Shedding Among Asymptomatic and Symptomatic Patients With SARS-CoV-2 Infection in a Community Treatment Center in the Republic of Korea It's Like Groundhog Day': Coronavirus Testing Labs Again Lack Key Supplies. The New York Times Quest Diagnostics Media Statement about COVID-19 Testing Long lines and delayed results again plague coronavirus tests as Thanksgiving approaches. The New York Times How to Get My COVID-19 Test Result Simple and sensitive molecular diagnostic test for SARS-CoV-2 surveillance Development of Reverse Transcription Loop-Mediated Isothermal Amplification Assays Targeting Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) Rapid Molecular Detection of SARS-CoV-2 (COVID-19) Virus RNA Using Colorimetric LAMP Rapid detection of novel coronavirus/Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) by reverse transcription-loop-mediated isothermal amplification Multiplex loop-mediated isothermal amplification (M-LAMP) assay for the detection of influenza A/H1, A/H3 and influenza B can provide a specimen-to-result diagnosis in 40 min with single genome copy sensitivity Development of a reverse transcription-loop-mediated isothermal amplification as a rapid early-detection method for novel SARS-CoV-2. Emerg Microbes Infect RT-LAMP for rapid diagnosis of coronavirus SARS-CoV-2 Evaluation of rapid diagnosis of novel coronavirus disease (COVID-19) using loop-mediated isothermal amplification A Novel Reverse Transcription Loop-Mediated Isothermal Amplification Method for Rapid Detection of SARS-CoV-2 Development of a Novel Reverse Transcription Loop-Mediated Isothermal Amplification Method for Rapid Detection of SARS-CoV-2 Rapid and visual detection of 2019 novel coronavirus (SARS-CoV-2) by a reverse transcription loop-mediated isothermal amplification assay Enhancing colorimetric loop-mediated isothermal amplification speed and sensitivity with guanidine chloride Interim Guidance for Rapid Antigen Testing for SARS-CoV-2 Test sensitivity is secondary to frequency and turnaround time for COVID-19 surveillance Rethinking Covid-19 Test Sensitivity-A Strategy for Containment Evaluation of Testing Frequency and Sampling for Severe Acute Respiratory Syndrome Coronavirus 2 Surveillance Strategies in Long-Term Care Facilities Assessment of SARS-CoV-2 Screening Strategies to Permit the Safe Reopening of College Campuses in the United States US Centers for Disease Control. CDC 2019-Novel Coronavirus (2019-nCoV) Real-Time RT-PCR Diagnostic Panel. 2020 SARS-CoV-2 detection using isothermal amplification and a rapid, inexpensive protocol for sample inactivation and purification CRISPR-Cas12-based detection of SARS-CoV-2 A single and two-stage, closed tube, molecular test for the 2019 novel coronavirus (COVID-19) at home, clinic, and points of entry Rapid Detection of COVID-19 Coronavirus Using a Reverse Transcriptional Loop-Mediated Isothermal Amplification (RT-LAMP) Diagnostic Platform SARS-CoV-2 LAMP Diagnostic Assay A 5-min RNA preparation method for COVID-19 detection with RT-qPCR R: A language and environment for statistical computing Virological assessment of hospitalized patients with COVID-2019 Clinical and virological data of the first cases of COVID-19 in Europe: a case series Respiratory disease and virus shedding in rhesus macaques inoculated with SARS-CoV-2. bioRxiv Initial evaluation of a mobile SARS-CoV-2 RT-LAMP testing strategy We thank Nathan Tanner for discussions about primers and optimizations of the RT-LAMP assay. We thank Brad Langhorst for specificity data for NEB primer sets. We thank Dr. Nathan Grubaugh for the Sars-CoV-2 N-gene transcript. The following reagent was deposited by the Centers for Disease Control and Prevention and obtained through BEI Resources, NIAID, NIH: SARS-Related Coronavirus 2, Isolate USA-WA1/2020, Gamma-Irradiated, NR-52287.