key: cord-0834804-ye9km4dm authors: Wang, Tianqi; Li, Zeyu; Jia, Yuqing; Zhu, Baoli; Cao, Zhipeng title: Interference of hemolysis on the postmortem biochemical analysis of IgE by ECLIA date: 2021-03-30 journal: Int J Legal Med DOI: 10.1007/s00414-021-02578-z sha: 986c7ad29ba368af3636f446e44681aac82de73c doc_id: 834804 cord_uid: ye9km4dm Forensic diagnosis of anaphylactic shock is a challenging task in forensic practice due to the lack of characteristic morphological changes. Postmortem analysis of serum IgE can provide helpful information for determining anaphylaxis. However, postmortem serum always suffers from hemolysis. To investigate the interference of hemolysis on postmortem analysis of total IgE by electrochemiluminescent immunoassay (ECLIA) and verify the suitability of the commercially available ECLIA kit for postmortem hemolyzed blood with the dilution-correction method, different levels of hemolyzed serum were prepared to evaluate the interference of hemolysis. A linear regression analysis was then performed on the concentration of total IgE in the completely hemolyzed blood and the corresponding serum. Our results indicated that hemolysis negatively interfered with the total IgE analysis by ECLIA and the interference (|Bias%|) increased with increasing levels of hemolysis. After controlling for |Bias%| by dilution, the test concentration of total IgE in the completely hemolyzed blood was still significantly lower than that in the serum (P < 0.05) and resulted in eight false-negative cases. A strong correlation was observed between the test concentration of total IgE in the completely hemolyzed blood and that in the serum (r = 0.983). After correction by the regression formula, the corrected concentration revealed no significant differences and exhibited the same diagnostic ability, compared with the serum total IgE concentration. These results indicate that the completely hemolyzed blood is not recommended for postmortem analysis of total IgE directly. The dilution-correction method might have potential utility in forensic practice for evaluating serum total IgE concentrations. Anaphylactic shock is a rapid and fatal multisystem syndrome, and its forensic diagnosis is a challenging task for forensic pathologists in practical work due to the lack of characteristic morphological changes [1] [2] [3] . Previous reports have indicated that serum biomarkers, such as specific and total immunoglobulin E (IgE), tryptase, chymase, and histamine, are useful for postmortem diagnosis of anaphylaxis [4] [5] [6] [7] [8] [9] [10] . As a key mediator of classic IgE-mediated anaphylaxis, IgE is often used as one of the indicators for the diagnosis of anaphylactic diseases in both clinical and forensic medicine [11] . Clinical studies have demonstrated that hemolysis interferes with the measurement of multiple indicators and that hemolyzed samples are not suitable for biochemical analysis in clinical laboratories [12] [13] [14] [15] . However, in forensic practice, blood samples always suffer from irreversible hemolysis due to postmortem changes, even within a short postmortem interval (PMI) [16] , which induces unavailable or unsuitable serum for postmortem biochemical analysis of IgE. As a result, it is necessary to investigate the interference of hemolysis on the postmortem analysis of serum IgE, and effective methods of reducing the interference in hemolyzed blood samples should be identified. The commercial electrochemiluminescent immunoassay (ECLIA) kit has been applied as a clinical measurement method for the biochemical analysis of IgE [17, 18] . Compared with the traditional enzyme-linked immunosorbent assay (ELISA) method, ECLIA has lower background noise and detection limits for analytes [14] . Clinical studies have shown that ECLIA exhibits good anti-interference ability against hemolysis in the measurement of various indicators and drugs, including the current new coronavirus pneumonia virus (COVID-19) antibody test [19] [20] [21] [22] [23] . However, no forensic reports are available on the postmortem analysis of IgE in hemolyzed samples with ECLIA. Therefore, the present study first aims to investigate the interference of hemolysis on the analysis of total IgE with ECLIA by using different levels of hemolyzed serum, and subsequently further verify the suitability of the commercially available ECLIA kit for the postmortem hemolyzed blood samples with the dilution-correction method. Forty-nine autopsy cases (30 males and 19 females) with a mean age of 46.8 years (range 19-92 years) in the School of Forensic Medicine, China Medical University from 2017 to 2019 were included. The causes of death, which were diagnosed by three blinded forensic pathologists, included hemorrhage shock, hanging, mechanical asphyxia, traumatic shock, poisoning (alcohol and carbon monoxide), pneumonia, and cardiovascular and cerebrovascular accidents. No individuals died from anaphylactic shock. Comprehensive medico-legal autopsies were undertaken for each case within a PMI of 48 h, and nonhemolyzed blood samples were collected from the left ventricle via incision at the aortic root using a Pasteur pipette. For each of the 49 included cases, the serum sample was obtained from 1 ml of nonhemolyzed blood following centrifugation at 3000 g for 20 min, and another 2 ml of the nonhemolyzed blood was frozen at − 20 °C and slowly thawed at room temperature (approximately 24 °C) for five cycles to prepare the completely hemolyzed blood sample. The serum and completely hemolyzed blood samples were stored at − 80 °C for biochemical analysis of total IgE. For six of the 49 included cases, 20 ml of nonhemolyzed blood from each case was collected to prepare the concentrated hemolyzed blood for use in the different levels of hemolyzed serum samples according to the previous methods [24] . The flow diagram of the study design is shown in Fig. 1 . Twenty milliliters of the nonhemolyzed blood from each of six cases was centrifuged at 3000 g for 10 min. All the upper serum was collected as the serum for total IgE analysis and preparation for different levels of hemolyzed serum samples. The lower blood cells were mixed with normal saline solution to a total volume of 50 ml, and then, the mixture was centrifuged at 3000 g for 10 min. The lower blood cells were collected, and the supernatant was discarded. After five cycles of mixing-centrifuging-discarding, the total IgE concentration in the final discarded supernatant could not be measured by ECLIA (< 0.1 IU/mL). The blood cells collected in the last round (approximately 10 ml) were vortexed and hemolyzed artificially by ultrasonication (500 Watts, 20 kHz, 5 min), and the concentrated hemolyzed blood sample was successfully prepared (Fig. 2) . The concentrations of hemoglobin in these six concentrated hemolyzed blood samples were determined. Hemoglobin concentration was employed to evaluate the degree of hemolysis. A total volume of 200 μl mixture composed of different ratios of concentrated hemolyzed blood and the corresponding serum was prepared, which contained seven hemolyzed levels with hemoglobin concentrations from 10 to 200 g/L. The volume of the concentrated hemolyzed blood in the mixture is shown in Table 1 . The total IgE concentrations were determined in different levels of hemolyzed serum samples. The total IgE concentration was determined with a commercial ECLIA kit (electrochemiluminescent method, Roche Diagnostic GmbH, Mannheim Germany) using a Roche automatic electrochemiluminescent immunoassay analyzer (Cobas e411, Roche Diagnostic GmbH, Mannheim Germany). The hemoglobin concentration was analyzed with Van Kampen-Zijlstra's reagent (cyanmethemoglobin method, Zhongshan Institution of Tianjin Modern High-Tech Research Institute, Tianjin China) using a UV spectrophotometer (UV-1800, Shimadzu Corporation, Kyoto, Japan). With reference to the previous study, Bias% was used as an indicator to assess the interference of hemolysis on the analysis of total IgE by ECLIA. The formula for Bias% is as follows: where C serum represents the concentration of total IgE in the corresponding serum, which was subsequently added to the mixture of different levels of hemolyzed serum samples; V serum represents the volume of the added serum, which can be calculated with the data presented in Table 1 ; and C test represents the test concentration of total IgE in samples with different levels of hemolyzed serum [24, 25] . A linear regression analysis was conducted to analyze the correlation between the test concentrations of total IgE in the completely hemolyzed blood and that in the corresponding serum. The Wilcoxon signed-rank test was used to evaluate the difference in total IgE concentration among the paired samples. SPSS 20.0 (SPSS Inc., Chicago, IL) was used to perform the statistical analysis, and P < 0.05 was considered statistically significant. To evaluate the interference of hemolysis on the biochemical analysis of total IgE by ECLIA, gradient hemolyzed serum samples were prepared. The color of the different levels of hemolyzed serum gradually became redder and darker as the hemoglobin concentration increased, and the transparency gradually decreased. The test concentration of total IgE in the mixture (C test ) was lower than the corresponding calculation concentration (C calculation ), and |Bias%| increased Table 2 . Given the fact that hemolysis negatively interfered with the analysis of total IgE performed with a commercially available ECLIA kit, we further investigated whether dilution of the completely hemolyzed blood can be used to estimate the level of total IgE in the serum. Due to the lack of quality control standards for postmortem biochemical analysis, a |Bias%| value within 10% was considered acceptable, according to previous relevant clinical research [26] . Based on the Bias% shown in Table 2 , all the 49 completely hemolyzed blood samples (HB: 136.18 ± 44.28 g/L) were diluted automatically on a Roche Cobas e411 with a ninefold volume of Diluent Universal (Roche Diagnostic GmbH, Mannheim Germany) to keep the final hemoglobin concentration less than 20 g/L (HB: 13.62 ± 4.43 g/L). Then, the concentrations of total IgE in the serum and diluted completely hemolyzed blood were tested. After obtaining the final test concentration of the completely hemolyzed blood by multiplying the total IgE concentration in the diluted blood by 10, the results showed that the final test concentration of total IgE in the completely hemolyzed blood (range: 1.65-532.00 IU/mL, median: 20.24 IU/mL) was still significantly lower than that in the serum (range: 3.53-955.70 IU/mL, median: 41.04 IU/ mL) (P < 0.05), which indicate that only dilution of the completely hemolyzed blood cannot be used to estimate serum IgE levels. A linear regression analysis was subsequently carried out for all the 49 cases. The results revealed that the test concentrations of total IgE in the completely hemolyzed blood (x) exhibited a strong correlation with those in the corresponding serum (y), and the regression formula was y = 1.7099x + 15.69 (r = 0.983) (Fig. 4) . Then, we used the regression formula to correct the test concentration of total IgE in the completely hemolyzed blood and obtained the corrected concentration (range: 19.44-924.48 IU/mL, median: 51.16 IU/mL), which showed no significant difference with the concentration of total IgE in serum (P = 0.435) (Fig. 5) . These observations preliminarily confirmed the feasibility of using the corrected concentrations obtained by the dilution-correction method to estimate serum IgE levels. To further verify the effectiveness of the corrected concentration, the forensic diagnostic abilities of total IgE in the serum, completely hemolyzed blood, and the corrected concentration were compared. According to Lara Tran's study, which regarded a postmortem level of 60 IU/mL as the cutoff value of serum total IgE, 20 of 49 cases showed a risk for atopic disposition or sensitivity to an allergen [10] . As a result, positive and negative serum total IgE was observed in 20 and 29 cases, respectively. Eight false-negative cases (concentration of total IgE > 60 IU/ml in the serum but < 60 IU/ml in the completely hemolyzed blood) occurred with the direct use of the test concentration of total IgE in the completely hemolyzed blood. However, the corrected concentration of total IgE showed the same diagnosis result as that of serum, indicating that the corrected concentration obtained by the dilution-correction method had the same forensic diagnostic ability as serum. In the clinical laboratory, hemolysis interferes with the measurement of multiple biomarkers [12] [13] [14] [15] for the following main reasons. First, the released intracellular substances from blood cells impact the progress of the assay. In severely hemolyzed postmortem samples, the main intracellular substance is hemoglobin, which has a strong absorbance peak in the spectrum, and a high concentration of hemoglobin also interferes with the binding of antigens and antibodies [27] . Second, in severe hemolysis, the intracellular substances released from blood cells may affect the concentration of serum biomarkers, such as increased serum potassium and decreased serum sodium after hemolysis [13, 28, 29] . In forensic practice, hemolysis of blood samples is more serious due to the effects of postmortem changes. Clinical research indicates that the ECLIA technique uses streptavidin-coated microparticles that are mechanically separated from the supernatant and the potentially interfering substances, so most of the interferences are eliminated in the final buffer [21] . Thus, the ECLIA technique is robust against hemolysis, and the present study applied ECLIA to analyze total IgE in postmortem hemolyzed blood. Our results indicate that hemolysis negatively interfered with the analysis of total IgE by ECLIA. As the level of hemolysis increased, the |Bias%| of total IgE in the hemolyzed blood also increased. Generally, the normal hemoglobin concentration in human blood is 130-160 g/L for males and 120-150 g/L for females [30] ; thus, the total IgE in completely hemolyzed blood samples of a normal human will be decreased by approximate 20% when analyzed through ECLIA due to interference, according to our results. A series of evaluation methods and indicators reflect the interference on different biomarkers and test methods in the clinical laboratories, such as the hemolytic index (HI), ± 10% change of concentrations from baseline (± 10%Δ) [14, 31, 32] . However, for postmortem biochemical analyses, more serious interfering factors are observed relative to clinical situations. Therefore, a set of evaluation methods and indicators should be proposed for different methods and bodily fluid samples used in postmortem biochemical analyses according to the various interfering factors. The present study used Bias% as an indicator to evaluate the interference of hemolysis. After a tenfold dilution of the completely hemolyzed blood to a final hemoglobin concentration less than 20 g/L, |Bias%| was controlled to less than 10%. The test concentration of total IgE in the completely hemolyzed blood was still approximately twofold lower than that in the corresponding serum (P < 0.05). In addition, eight falsenegative values appeared in the completely hemolyzed blood. Regarding the results of the present study and the interference of hemolysis, we speculate that the release of intracellular substances from blood cells may contribute to the dilution of total IgE in the serum, which is similar to the decrease of sodium in postmortem samples [28] . Therefore, the results can be considered positive if the test concentration of total IgE in the completely hemolyzed blood obtained through ECLIA is higher than the cutoff value. However, false-negative values cannot be excluded if the test concentration in the completely hemolyzed blood is lower than the cutoff value. Consequently, the test concentration of total IgE in completely hemolyzed blood is not recommended for direct use in forensic diagnostics. The concentration of total IgE corrected by the regression formula showed no significant difference with the total IgE in serum (P > 0.05) and exhibited the same diagnostic ability as serum. Therefore, we believe that the corrected concentration can be used to evaluate the concentration of total IgE in serum for diagnostics when hemolysis is severe and serum is not available. Certain limitations were observed in the present study. First, the number of included cases was relatively small. More samples will lead to a better regression analysis, and the corrected concentration will be more accurate. However, we believe that the method in the present study for the evaluation of serum total IgE concentrations from completely hemolyzed blood samples of these 49 cases is effective. Second, as we discussed above, in addition to the influence of hemoglobin and other interfering substances, the dilution of serum caused by the release of intracellular substances also leads to hemolysis-induced interference on the analysis of total IgE [13, 27, 29] . We believe that hematocrit (HCT), an indicator that reflects the proportion of red blood cells in the volume of whole blood, may act as another useful correction factor [33, 34] . Regrettably, because of the equipment limitations in our laboratory, the influence of HCT was not investigated in the present study. Therefore, the regression formula should be established according to the different situations in each laboratory. In conclusion, the present study demonstrated that hemolysis negatively interferes with the biochemical analysis of total IgE by ECLIA and that more severe hemolysis corresponds to greater interference. Therefore, completely hemolyzed blood is not recommended for direct postmortem analysis of total IgE, and the corrected concentrations obtained by the dilution-correction method might be of potential utility in forensic practice for evaluating serum total IgE concentrations. Although further studies may be needed, this pilot study provides a possible method for total IgE analysis in postmortem hemolyzed blood. Author contribution The study was designed by Baoli Zhu and Zhipeng Cao. The acquisition and analysis of data and interpretation of the results were carried out by Tianqi Wang, Zeyu Li, and Yuqing Jia. Concentration of total IgE in the serum and completely hemolyzed blood, and the corrected concentration (n = 49 in each group). Data in the boxes are presented in a range (median). *There was a statistically significant difference between the test concentration of total IgE in completely hemolyzed blood and that in serum (Wilcoxon signed-rank paired test, P < 0.05). N.S., no significance Ethics approval This study was approved by the Ethics Committee of China Medical University and was performed in accordance with the 1964 Declaration of Helsinki. All biological samples were anonymized prior to analysis and analyzed in the same laboratory. No ethical approval was necessary to perform biochemical analyses in the collected cases. The authors declare no competing interests. Postmortem diagnosis of acute anaphylaxis by serum tryptase analysis. A case report Postmortem findings after fatal anaphylactic reactions Postmortem serum tryptase levels in anaphylactic and non-anaphylactic deaths Forensic biochemistry for functional investigation of death: concept and practical application Usefulness of serum mast cell-specific chymase levels for postmortem diagnosis of anaphylaxis Diagnostic role of serum tryptase in anaphylactic deaths in forensic medicine: a systematic review and meta-analysis Post mortem tryptase cut-off level for anaphylactic death Laboratory investigation of deaths due to anaphylaxis Utilization of serum tryptase and immunoglobulin e assay in the postmortem diagnosis of anaphylaxis Determination of specific IgE in pericardial and cerebrospinal fluids in forensic casework Pathophysiology of anaphylaxis Influence of hemolysis on routine clinical chemistry testing Effects of hemolysis interferences on routine biochemistry parameters Effect of interference from hemolysis, icterus and lipemia on routine pediatric clinical chemistry assays Influence of in vitro hemolysis on 80 different laboratory tests BNP and NT-proBNP as diagnostic biomarkers for cardiac dysfunction in both clinical and forensic medicine Pediatric reference intervals for 28 chemistries and immunoassays on the Roche cobas 6000 analyzer-a CALIPER pilot study Clinical relevance of plasma concentrations of MBL in accordance with IgE levels in children diagnosed with bronchial asthma Effects of hemolysis and lipemia on the COBAS salicylate and acetaminophen assays compared to GDS assays Clinical performance of the Roche cobas e411 automated assay system for thyrotropin-receptor antibodies for the diagnosis of Graves' disease Spurious hemolysis does not influence the reliability of digoxin testing on Siemens RXL MAX and Roche Cobas e601 Detection of toxoplasma-specific immunoglobulin G in human sera: performance comparison of in house Dot-ELISA with ECLIA and ELISA Evaluation of interference effects from hemolysis, icterus and lipemia on the Roche Elecsys(R) Anti-SARS-CoV-2 assay Experimental study on the interference of hemolytic pericardial fluid on the detection of NT-proBNP The estimation of postmortem serum urea via the ultrafiltration of hemolyzed blood Preliminary study on allowable total error and allowable imprecision in tests for specific proteins in China Pandey AN (2002) Linearization of the relationship between serum sodium, potassium concentration, their ratio and time since death in Chandigarh zone of North-West India Interference testing Importance of hemoglobin concentration and its modification for the outcome of head and neck cancer patients treated with radiotherapy Precision, accuracy, cross reactivity and comparability of serum indices measurement on Abbott Architect c8000, Beckman Coulter AU5800 and Roche Cobas 6000 c501 clinical chemistry analyzers Hemolysis indexes for biochemical tests and immunoassays on Roche analyzers: determination of allowable interference limits according to different calculation methods Anemia screening in naval aviation: is hemoglobin a better indicator than hematocrit as the primary index Publisher's Note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations