key: cord-0830780-uskpfxcj authors: Zhou, Cong; Chase, J. Geoffrey; Sun, Qianhui; Knopp, Jennifer title: A Nonlinear Hysteretic Model for Automated Prediction of Lung Mechanics during Mechanical Ventilation date: 2020-12-31 journal: IFAC-PapersOnLine DOI: 10.1016/j.ifacol.2021.04.177 sha: fc9424b4f1ab4ddfb088939227a6cab66672db05 doc_id: 830780 cord_uid: uskpfxcj Mechanical ventilation (MV) is core intensive care unit (ICU) therapy during the Covid-19 pandemic. Optimising MV care to a specific patient with respiratory failure is difficult due to inter- and intra- patient variability in lung mechanics and condition. The ability to accurately predict patient-specific lung response to a change in MV settings would enable semi-automated care and significantly improve the efficiency of MV monitoring and care. It has particular emphasis when considering MV care required to treat Covid-19 patients, who require longer MV care, where patient-specific care can reduce the time on MV required. This study develops a nonlinear smooth hysteresis loop model (HLM) able to capture the essential lung dynamics in a patient-specific fashion from measured ventilator data, particularly for changes of compliance and infection points of the pressure-volume loop. The automated (no human input) hysteresis loop analysis (HLA) method is applied to identify HLM model parameters, enabling automated digital cloning to create a virtual patient model to accurately predict lung response at a specified positive end expiratory pressure (PEEP) level, as well as in response to the changes of PEEP. The performance of this automated digital cloning approach is assessed using clinical data from 4 patients and 8 recruitment maneuver (RM) arms. Validation results show the HLM-based hysteresis loops identified using HLA match clinical pressure-volume loops very well with root-mean-square (RMS) errors less than 2% for all 8 data sets over 4 patients, validating the accuracy of the developed HLM in capturing the essential lung physiology and respiratory behaviours at different patient conditions. More importantly, the patient-specific digital clones at lower PEEP levels accurately predict lung response at higher PEEP levels with predicted peak inspiratory pressure (PIP) errors less than 2% in average. In addition, the resulted additional lung volume Vfrc obtained with PEEP changes are predicted with average absolute difference of 0.025L. The overall results validate the versatility and potential of the developed HLM for delineating changes of nonlinear lung dynamics, and its capability to create a predictive virtual patient with use of HLA for future treatment personalization and optimisation in MV therapy. Mechanical ventilation (MV) is core therapy for patient suffering respiratory failure and acute respiratory distress syndrome (ARDS) in intensive care unit (ICU), currently one of the major way to support lives of Covid-19 patients (Mahase, 2020) . However, MV is invasive and may cause a further lung injury and higher mortality if MV settings are poorly suited to the patients (Major et al., 2018) . Accurate prediction of patient-specific lung response to the changes of MV settings, such as peak positive end expiratory pressure (PEEP) would enable a more confident and efficient MV design, minimizing the risk of barotrauma and volutrauma (Langdon et al., 2017) . In addition, an optimised MV care would also reduce the treatment time of patient on a ventilator, which is critical for improving the capacity of health systems to provide adequate MV during the Covid-19 pandemic. Current studies of basis functions have shown good modelling and predicting outcome of lung mechanics and airway pressure based on a single compartment linear lung model (Langdon et al., 2017 , Morton et al., 2018 , Morton et al., 2019 . These methods are built on the assumptions of data mode, changes of specific elastance and resistance shapes, lacking of ability to fully represent the lung mechanics, such as the additional lung volume or dynamic functional residual capacity (Vfrc). Hence, this work focuses on the development of a nonlinear hysteresis loop model (HLM) from the perspective of mechanical-physiologica relevance for the dynamic respiratory system. The new HLM is able to capture the essential lung mechanics for creating a patient-specific virtual patient in an automated fashion. Identification and prediction methods are also developed with clinical measurements at a low PEEP level. The performance of the proposed model and methods for predicting lung dynamics at higher PEEP levels Mechanical ventilation (MV) is core therapy for patient suffering respiratory failure and acute respiratory distress syndrome (ARDS) in intensive care unit (ICU), currently one of the major way to support lives of Covid-19 patients (Mahase, 2020) . However, MV is invasive and may cause a further lung injury and higher mortality if MV settings are poorly suited to the patients (Major et al., 2018) . Accurate prediction of patient-specific lung response to the changes of MV settings, such as peak positive end expiratory pressure (PEEP) would enable a more confident and efficient MV design, minimizing the risk of barotrauma and volutrauma (Langdon et al., 2017) . In addition, an optimised MV care would also reduce the treatment time of patient on a ventilator, which is critical for improving the capacity of health systems to provide adequate MV during the Covid-19 pandemic. Current studies of basis functions have shown good modelling and predicting outcome of lung mechanics and airway pressure based on a single compartment linear lung model (Langdon et al., 2017 , Morton et al., 2018 , Morton et al., 2019 . These methods are built on the assumptions of data mode, changes of specific elastance and resistance shapes, lacking of ability to fully represent the lung mechanics, such as the additional lung volume or dynamic functional residual capacity (Vfrc). Hence, this work focuses on the development of a nonlinear hysteresis loop model (HLM) from the perspective of mechanical-physiologica relevance for the dynamic respiratory system. The new HLM is able to capture the essential lung mechanics for creating a patient-specific virtual patient in an automated fashion. Identification and prediction methods are also developed with clinical measurements at a low PEEP level. The performance of the proposed model and methods for predicting lung dynamics at higher PEEP levels This study develops a nonlinear smooth hysteresis loop model (HLM) able to capture the essential lung dynamics in a patient-specific fashion from measured ventilator data, particularly for changes of compliance and infection points of the pressure-volume loop. The automated (no human input) hysteresis loop analysis (HLA) method is applied to identify HLM model parameters, enabling automated digital cloning to create a virtual patient model to accurately predict lung response at a specified positive end expiratory pressure (PEEP) level, as well as in response to the changes of PEEP. The performance of this automated digital cloning approach is assessed using clinical data from 4 patients and 8 recruitment maneuver (RM) arms. Validation results show the HLM-based hysteresis loops identified using HLA match clinical pressurevolume loops very well with root-mean-square (RMS) errors less than 2% for all 8 data sets over 4 patients, validating the accuracy of the developed HLM in capturing the essential lung physiology and respiratory behaviours at different patient conditions. More importantly, the patient-specific digital clones at lower PEEP levels accurately predict lung response at higher PEEP levels with predicted peak inspiratory pressure (PIP) errors less than 2% in average. In addition, the resulted additional lung volume Vfrc obtained with PEEP changes are predicted with average absolute difference of 0.025L. The overall results validate the versatility and potential of the developed HLM for delineating changes of nonlinear lung dynamics, and its capability to create a predictive virtual patient with use of HLA for future treatment personalization and optimisation in MV therapy. Mechanical ventilation (MV) is core therapy for patient suffering respiratory failure and acute respiratory distress syndrome (ARDS) in intensive care unit (ICU), currently one of the major way to support lives of Covid-19 patients (Mahase, 2020) . However, MV is invasive and may cause a further lung injury and higher mortality if MV settings are poorly suited to the patients (Major et al., 2018) . Accurate prediction of patient-specific lung response to the changes of MV settings, such as peak positive end expiratory pressure (PEEP) would enable a more confident and efficient MV design, minimizing the risk of barotrauma and volutrauma (Langdon et al., 2017) . In addition, an optimised MV care would also reduce the treatment time of patient on a ventilator, which is critical for improving the capacity of health systems to provide adequate MV during the Covid-19 pandemic. Current studies of basis functions have shown good modelling and predicting outcome of lung mechanics and airway pressure based on a single compartment linear lung model (Langdon et al., 2017 , Morton et al., 2018 , Morton et al., 2019 . These methods are built on the assumptions of data mode, changes of specific elastance and resistance shapes, lacking of ability to fully represent the lung mechanics, such as the additional lung volume or dynamic functional residual capacity (Vfrc). Hence, this work focuses on the development of a nonlinear hysteresis loop model (HLM) from the perspective of mechanical-physiologica relevance for the dynamic respiratory system. The new HLM is able to capture the essential lung mechanics for creating a patient-specific virtual patient in an automated fashion. Identification and prediction methods are also developed with clinical measurements at a low PEEP level. The performance of the proposed model and methods for predicting lung dynamics at higher PEEP levels This study develops a nonlinear smooth hysteresis loop model (HLM) able to capture the essential lung dynamics in a patient-specific fashion from measured ventilator data, particularly for changes of compliance and infection points of the pressure-volume loop. The automated (no human input) hysteresis loop analysis (HLA) method is applied to identify HLM model parameters, enabling automated digital cloning to create a virtual patient model to accurately predict lung response at a specified positive end expiratory pressure (PEEP) level, as well as in response to the changes of PEEP. The performance of this automated digital cloning approach is assessed using clinical data from 4 patients and 8 recruitment maneuver (RM) arms. Validation results show the HLM-based hysteresis loops identified using HLA match clinical pressurevolume loops very well with root-mean-square (RMS) errors less than 2% for all 8 data sets over 4 patients, validating the accuracy of the developed HLM in capturing the essential lung physiology and respiratory behaviours at different patient conditions. More importantly, the patient-specific digital clones at lower PEEP levels accurately predict lung response at higher PEEP levels with predicted peak inspiratory pressure (PIP) errors less than 2% in average. In addition, the resulted additional lung volume Vfrc obtained with PEEP changes are predicted with average absolute difference of 0.025L. The overall results validate the versatility and potential of the developed HLM for delineating changes of nonlinear lung dynamics, and its capability to create a predictive virtual patient with use of HLA for future treatment personalization and optimisation in MV therapy. Keywords: Nonlinear Modelling; Hysteresis Model; Hysteresis loop analysis; Virtual Patient; Mechanical Ventilation; Lung mechanics. Mechanical ventilation (MV) is core therapy for patient suffering respiratory failure and acute respiratory distress syndrome (ARDS) in intensive care unit (ICU), currently one of the major way to support lives of Covid-19 patients (Mahase, 2020) . However, MV is invasive and may cause a further lung injury and higher mortality if MV settings are poorly suited to the patients (Major et al., 2018) . Accurate prediction of patient-specific lung response to the changes of MV settings, such as peak positive end expiratory pressure (PEEP) would enable a more confident and efficient MV design, minimizing the risk of barotrauma and volutrauma (Langdon et al., 2017) . In addition, an optimised MV care would also reduce the treatment time of patient on a ventilator, which is critical for improving the capacity of health systems to provide adequate MV during the Covid-19 pandemic. Current studies of basis functions have shown good modelling and predicting outcome of lung mechanics and airway pressure based on a single compartment linear lung model (Langdon et al., 2017 , Morton et al., 2018 , Morton et al., 2019 . These methods are built on the assumptions of data mode, changes of specific elastance and resistance shapes, lacking of ability to fully represent the lung mechanics, such as the additional lung volume or dynamic functional residual capacity (Vfrc). Hence, this work focuses on the development of a nonlinear hysteresis loop model (HLM) from the perspective of mechanical-physiologica relevance for the dynamic respiratory system. The new HLM is able to capture the essential lung mechanics for creating a patient-specific virtual patient in an automated fashion. Identification and prediction methods are also developed with clinical measurements at a low PEEP level. The performance of the proposed model and methods for predicting lung dynamics at higher PEEP levels Cong Zhou* , **, J. Geoffrey Chase**, Qianhui Sun**, Jennifer Knopp** *School of Civil Aviation, Northwestern Polytechnical University, China(e-mail: cong.zhou@nwpu.edu.cn) **Department of Mechanical Engineering; Dept of Mechanical Eng, Centre for Bio-Engineering, University of Canterbury Christchurch, New Zealand, (e-mail: geoff.chase@canterbury.ac.nz; qianhui.sun@pg.canterbury.ac.nz; jennifer.knopp@canterbury.ac.nz) Abstract: Mechanical ventilation (MV) is core intensive care unit (ICU) therapy during the Covid-19 pandemic. Optimising MV care to a specific patient with respiratory failure is difficult due to inter-and intra-patient variability in lung mechanics and condition. The ability to accurately predict patient-specific lung response to a change in MV settings would enable semi-automated care and significantly improve the efficiency of MV monitoring and care. It has particular emphasis when considering MV care required to treat Covid-19 patients, who require longer MV care, where patient-specific care can reduce the time on MV required. This study develops a nonlinear smooth hysteresis loop model (HLM) able to capture the essential lung dynamics in a patient-specific fashion from measured ventilator data, particularly for changes of compliance and infection points of the pressure-volume loop. The automated (no human input) hysteresis loop analysis (HLA) method is applied to identify HLM model parameters, enabling automated digital cloning to create a virtual patient model to accurately predict lung response at a specified positive end expiratory pressure (PEEP) level, as well as in response to the changes of PEEP. The performance of this automated digital cloning approach is assessed using clinical data from 4 patients and 8 recruitment maneuver (RM) arms. Validation results show the HLM-based hysteresis loops identified using HLA match clinical pressurevolume loops very well with root-mean-square (RMS) errors less than 2% for all 8 data sets over 4 patients, validating the accuracy of the developed HLM in capturing the essential lung physiology and respiratory behaviours at different patient conditions. More importantly, the patient-specific digital clones at lower PEEP levels accurately predict lung response at higher PEEP levels with predicted peak inspiratory pressure (PIP) errors less than 2% in average. In addition, the resulted additional lung volume Vfrc obtained with PEEP changes are predicted with average absolute difference of 0.025L. The overall results validate the versatility and potential of the developed HLM for delineating changes of nonlinear lung dynamics, and its capability to create a predictive virtual patient with use of HLA for future treatment personalization and optimisation in MV therapy. Keywords: Nonlinear Modelling; Hysteresis Model; Hysteresis loop analysis; Virtual Patient; Mechanical Ventilation; Lung mechanics. Mechanical ventilation (MV) is core therapy for patient suffering respiratory failure and acute respiratory distress syndrome (ARDS) in intensive care unit (ICU), currently one of the major way to support lives of Covid-19 patients (Mahase, 2020) . However, MV is invasive and may cause a further lung injury and higher mortality if MV settings are poorly suited to the patients (Major et al., 2018) . Accurate prediction of patient-specific lung response to the changes of MV settings, such as peak positive end expiratory pressure (PEEP) would enable a more confident and efficient MV design, minimizing the risk of barotrauma and volutrauma (Langdon et al., 2017) . In addition, an optimised MV care would also reduce the treatment time of patient on a ventilator, which is critical for improving the capacity of health systems to provide adequate MV during the Covid-19 pandemic. Current studies of basis functions have shown good modelling and predicting outcome of lung mechanics and airway pressure based on a single compartment linear lung model (Langdon et al., 2017 , Morton et al., 2018 , Morton et al., 2019 . These methods are built on the assumptions of data mode, changes of specific elastance and resistance shapes, lacking of ability to fully represent the lung mechanics, such as the additional lung volume or dynamic functional residual capacity (Vfrc). Hence, this work focuses on the development of a nonlinear hysteresis loop model (HLM) from the perspective of mechanical-physiologica relevance for the dynamic respiratory system. The new HLM is able to capture the essential lung mechanics for creating a patient-specific virtual patient in an automated fashion. Identification and prediction methods are also developed with clinical measurements at a low PEEP level. The performance of the proposed model and methods for predicting lung dynamics at higher PEEP levels This study develops a nonlinear smooth hysteresis loop model (HLM) able to capture the essential lung dynamics in a patient-specific fashion from measured ventilator data, particularly for changes of compliance and infection points of the pressure-volume loop. The automated (no human input) hysteresis loop analysis (HLA) method is applied to identify HLM model parameters, enabling automated digital cloning to create a virtual patient model to accurately predict lung response at a specified positive end expiratory pressure (PEEP) level, as well as in response to the changes of PEEP. The performance of this automated digital cloning approach is assessed using clinical data from 4 patients and 8 recruitment maneuver (RM) arms. Validation results show the HLM-based hysteresis loops identified using HLA match clinical pressurevolume loops very well with root-mean-square (RMS) errors less than 2% for all 8 data sets over 4 patients, validating the accuracy of the developed HLM in capturing the essential lung physiology and respiratory behaviours at different patient conditions. More importantly, the patient-specific digital clones at lower PEEP levels accurately predict lung response at higher PEEP levels with predicted peak inspiratory pressure (PIP) errors less than 2% in average. In addition, the resulted additional lung volume Vfrc obtained with PEEP changes are predicted with average absolute difference of 0.025L. The overall results validate the versatility and potential of the developed HLM for delineating changes of nonlinear lung dynamics, and its capability to create a predictive virtual patient with use of HLA for future treatment personalization and optimisation in MV therapy. Keywords: Nonlinear Modelling; Hysteresis Model; Hysteresis loop analysis; Virtual Patient; Mechanical Ventilation; Lung mechanics. Mechanical ventilation (MV) is core therapy for patient suffering respiratory failure and acute respiratory distress syndrome (ARDS) in intensive care unit (ICU), currently one of the major way to support lives of Covid-19 patients (Mahase, 2020) . However, MV is invasive and may cause a further lung injury and higher mortality if MV settings are poorly suited to the patients (Major et al., 2018) . Accurate prediction of patient-specific lung response to the changes of MV settings, such as peak positive end expiratory pressure (PEEP) would enable a more confident and efficient MV design, minimizing the risk of barotrauma and volutrauma (Langdon et al., 2017) . In addition, an optimised MV care would also reduce the treatment time of patient on a ventilator, which is critical for improving the capacity of health systems to provide adequate MV during the Covid-19 pandemic. Current studies of basis functions have shown good modelling and predicting outcome of lung mechanics and airway pressure based on a single compartment linear lung model (Langdon et al., 2017 , Morton et al., 2018 , Morton et al., 2019 . These methods are built on the assumptions of data mode, changes of specific elastance and resistance shapes, lacking of ability to fully represent the lung mechanics, such as the additional lung volume or dynamic functional residual capacity (Vfrc). Hence, this work focuses on the development of a nonlinear hysteresis loop model (HLM) from the perspective of mechanical-physiologica relevance for the dynamic respiratory system. The new HLM is able to capture the essential lung mechanics for creating a patient-specific virtual patient in an automated fashion. Identification and prediction methods are also developed with clinical measurements at a low PEEP level. The performance of the proposed model and methods for predicting lung dynamics at higher PEEP levels This study develops a nonlinear smooth hysteresis loop model (HLM) able to capture the essential lung dynamics in a patient-specific fashion from measured ventilator data, particularly for changes of compliance and infection points of the pressure-volume loop. The automated (no human input) hysteresis loop analysis (HLA) method is applied to identify HLM model parameters, enabling automated digital cloning to create a virtual patient model to accurately predict lung response at a specified positive end expiratory pressure (PEEP) level, as well as in response to the changes of PEEP. The performance of this automated digital cloning approach is assessed using clinical data from 4 patients and 8 recruitment maneuver (RM) arms. Validation results show the HLM-based hysteresis loops identified using HLA match clinical pressurevolume loops very well with root-mean-square (RMS) errors less than 2% for all 8 data sets over 4 patients, validating the accuracy of the developed HLM in capturing the essential lung physiology and respiratory behaviours at different patient conditions. More importantly, the patient-specific digital clones at lower PEEP levels accurately predict lung response at higher PEEP levels with predicted peak inspiratory pressure (PIP) errors less than 2% in average. In addition, the resulted additional lung volume Vfrc obtained with PEEP changes are predicted with average absolute difference of 0.025L. The overall results validate the versatility and potential of the developed HLM for delineating changes of nonlinear lung dynamics, and its capability to create a predictive virtual patient with use of HLA for future treatment personalization and optimisation in MV therapy. Ventilation; Lung mechanics. Mechanical ventilation (MV) is core therapy for patient suffering respiratory failure and acute respiratory distress syndrome (ARDS) in intensive care unit (ICU), currently one of the major way to support lives of Covid-19 patients (Mahase, 2020) . However, MV is invasive and may cause a further lung injury and higher mortality if MV settings are poorly suited to the patients (Major et al., 2018) . Accurate prediction of patient-specific lung response to the changes of MV settings, such as peak positive end expiratory pressure (PEEP) would enable a more confident and efficient MV design, minimizing the risk of barotrauma and volutrauma (Langdon et al., 2017) . In addition, an optimised MV care would also reduce the treatment time of patient on a ventilator, which is critical for improving the capacity of health systems to provide adequate MV during the Covid-19 pandemic. Current studies of basis functions have shown good modelling and predicting outcome of lung mechanics and airway pressure based on a single compartment linear lung model (Langdon et al., 2017 , Morton et al., 2018 , Morton et al., 2019 . These methods are built on the assumptions of data mode, changes of specific elastance and resistance shapes, lacking of ability to fully represent the lung mechanics, such as the additional lung volume or dynamic functional residual capacity (Vfrc). Hence, this work focuses on the development of a nonlinear hysteresis loop model (HLM) from the perspective of mechanical-physiologica relevance for the dynamic respiratory system. The new HLM is able to capture the essential lung mechanics for creating a patient-specific virtual patient in an automated fashion. Identification and prediction methods are also developed with clinical measurements at a low PEEP level. The performance of the proposed model and methods for predicting lung dynamics at higher PEEP levels This study develops a nonlinear smooth hysteresis loop model (HLM) able to capture the essential lung dynamics in a patient-specific fashion from measured ventilator data, particularly for changes of compliance and infection points of the pressure-volume loop. The automated (no human input) hysteresis loop analysis (HLA) method is applied to identify HLM model parameters, enabling automated digital cloning to create a virtual patient model to accurately predict lung response at a specified positive end expiratory pressure (PEEP) level, as well as in response to the changes of PEEP. The performance of this automated digital cloning approach is assessed using clinical data from 4 patients and 8 recruitment maneuver (RM) arms. Validation results show the HLM-based hysteresis loops identified using HLA match clinical pressurevolume loops very well with root-mean-square (RMS) errors less than 2% for all 8 data sets over 4 patients, validating the accuracy of the developed HLM in capturing the essential lung physiology and respiratory behaviours at different patient conditions. More importantly, the patient-specific digital clones at lower PEEP levels accurately predict lung response at higher PEEP levels with predicted peak inspiratory pressure (PIP) errors less than 2% in average. In addition, the resulted additional lung volume Vfrc obtained with PEEP changes are predicted with average absolute difference of 0.025L. The overall results validate the versatility and potential of the developed HLM for delineating changes of nonlinear lung dynamics, and its capability to create a predictive virtual patient with use of HLA for future treatment personalization and optimisation in MV therapy. Ventilation; Lung mechanics. Mechanical ventilation (MV) is core therapy for patient suffering respiratory failure and acute respiratory distress syndrome (ARDS) in intensive care unit (ICU), currently one of the major way to support lives of Covid-19 patients (Mahase, 2020) . However, MV is invasive and may cause a further lung injury and higher mortality if MV settings are poorly suited to the patients (Major et al., 2018) . Accurate prediction of patient-specific lung response to the changes of MV settings, such as peak positive end expiratory pressure (PEEP) would enable a more confident and efficient MV design, minimizing the risk of barotrauma and volutrauma (Langdon et al., 2017) . In addition, an optimised MV care would also reduce the treatment time of patient on a ventilator, which is critical for improving the capacity of health systems to provide adequate MV during the Covid-19 pandemic. Current studies of basis functions have shown good modelling and predicting outcome of lung mechanics and airway pressure based on a single compartment linear lung model (Langdon et al., 2017 , Morton et al., 2018 , Morton et al., 2019 . These methods are built on the assumptions of data mode, changes of specific elastance and resistance shapes, lacking of ability to fully represent the lung mechanics, such as the additional lung volume or dynamic functional residual capacity (Vfrc). Hence, this work focuses on the development of a nonlinear hysteresis loop model (HLM) from the perspective of mechanical-physiologica relevance for the dynamic respiratory system. The new HLM is able to capture the essential lung mechanics for creating a patient-specific virtual patient in an automated fashion. Identification and prediction methods are also developed with clinical measurements at a low PEEP level. The performance of the proposed model and methods for predicting lung dynamics at higher PEEP levels Optimising MV care to a specific patient with respiratory failure is difficult due to inter-and intra-patient variability in lung mechanics and condition. The ability to accurately predict patient-specific lung response to a change in MV settings would enable semi-automated care and significantly improve the efficiency of MV monitoring and care. It has particular emphasis when considering MV care required to treat Covid-19 patients, who require longer MV care, where patient-specific care can reduce the time on MV required. This study develops a nonlinear smooth hysteresis loop model (HLM) able to capture the essential lung dynamics in a patient-specific fashion from measured ventilator data, particularly for changes of compliance and infection points of the pressure-volume loop. The automated (no human input) hysteresis loop analysis (HLA) method is applied to identify HLM model parameters, enabling automated digital cloning to create a virtual patient model to accurately predict lung response at a specified positive end expiratory pressure (PEEP) level, as well as in response to the changes of PEEP. The performance of this automated digital cloning approach is assessed using clinical data from 4 patients and 8 recruitment maneuver (RM) arms. Validation results show the HLM-based hysteresis loops identified using HLA match clinical pressurevolume loops very well with root-mean-square (RMS) errors less than 2% for all 8 data sets over 4 patients, validating the accuracy of the developed HLM in capturing the essential lung physiology and respiratory behaviours at different patient conditions. More importantly, the patient-specific digital clones at lower PEEP levels accurately predict lung response at higher PEEP levels with predicted peak inspiratory pressure (PIP) errors less than 2% in average. In addition, the resulted additional lung volume Vfrc obtained with PEEP changes are predicted with average absolute difference of 0.025L. The overall results validate the versatility and potential of the developed HLM for delineating changes of nonlinear lung dynamics, and its capability to create a predictive virtual patient with use of HLA for future treatment personalization and optimisation in MV therapy. Keywords: Nonlinear Modelling; Hysteresis Model; Hysteresis loop analysis; Virtual Patient; Mechanical Ventilation; Lung mechanics. Mechanical ventilation (MV) is core therapy for patient suffering respiratory failure and acute respiratory distress syndrome (ARDS) in intensive care unit (ICU), currently one of the major way to support lives of Covid-19 patients (Mahase, 2020) . However, MV is invasive and may cause a further lung injury and higher mortality if MV settings are poorly suited to the patients (Major et al., 2018) . Accurate prediction of patient-specific lung response to the changes of MV settings, such as peak positive end expiratory pressure (PEEP) would enable a more confident and efficient MV design, minimizing the risk of barotrauma and volutrauma (Langdon et al., 2017) . In addition, an optimised MV care would also reduce the treatment time of patient on a ventilator, which is critical for improving the capacity of health systems to provide adequate MV during the Covid-19 pandemic. Current studies of basis functions have shown good modelling and predicting outcome of lung mechanics and airway pressure based on a single compartment linear lung model (Langdon et al., 2017 , Morton et al., 2018 , Morton et al., 2019 . These methods are built on the assumptions of data mode, changes of specific elastance and resistance shapes, lacking of ability to fully represent the lung mechanics, such as the additional lung volume or dynamic functional residual capacity (Vfrc). Hence, this work focuses on the development of a nonlinear hysteresis loop model (HLM) from the perspective of mechanical-physiologica relevance for the dynamic respiratory system. The new HLM is able to capture the essential lung mechanics for creating a patient-specific virtual patient in an automated fashion. Identification and prediction methods are also developed with clinical measurements at a low PEEP level. The performance of the proposed model and methods for predicting lung dynamics at higher PEEP levels are investigated. In addition, the additional lung volume during the change of PEEP is also predicted using the created virtual patient model. The key features for lung respiratory dynamics include the recruitment of alveoli and alveolar distension in inspiratory, as well as the recoil of lung elastance in expiratory. The identifiable HLM is thus proposed with a linear spring, Ke, to represent the lung tissue elastance, and two nonlinear hysteretic spring, Kh1 and Kh2, for alveolar expanding elastance during inspiration and expiration, respectively. To account for the end-inspiratory plateau pressure, a nonlinear element controlled by energy absorption is also added to the inspiration hysteretic spring. Thus, the dynamic equation of motion for the proposed model is defined: (1) where V is the volume of air delivered to lungs, Vh1 and Vh2 are hysteretic volume response during inspiration and expiration, respectively. R is the airway resistance, PEEP is the positive end-expiratory pressure, and ( ) is the steady-state input force. The nonlinear stiffness, K(t), for a breath can thus be defined in a differential form (Zhou and Chase, 2020) : Thus, the change of nonlinear stiffness is determined by the two hysteretic springs ℎ1̇ℎ 1 and ℎ2̇ℎ 2 . Particularly, the inspiratory hysteretic spring is defined: where Vm1 is the lower deflection point, δ controls the endinspiratory stiffness for plateau pressure, Eh1 is the dissipated energy due to inspiratory hysteresis, q controls the smoothness of plateau stiffness and Em1 is the maximum energy without inspiratory pause at the peak alveolar pressure. Kc controls stiffness changes from inspiration to expiration with defined signum functions + and − : The expiratory hysteretic spring is defined with a similar curve shape to inspiration for alveolar derecruitment: where Vm2 is the upper deflection point. Therefore, ten (10) parameters, including Ke, Kh1, Kh2, R, Vm1, Vm2, Em1, Kc, q and δ are defined to model the lung hysteresis mechanics using the proposed HLM. The goal of section 2.2 is to create a virtual patient at a low PEEP level based on the identification of the proposed HLM with the ability to predict patient lung dynamics at higher PEEP levels. The identification process can be implemented in an automated fashion without requiring human input, including: • HLA identification of Ke, Kh1, Kh2, Vm1, Vm2 and Kc; • Calculation of R and ; • Forward simulation to identify Em1 and q. In particular, HLA is an automated algorithm of identifying the change of slopes and breakpoints for a hysteresis loop (Zhou et al., 2015 , Zhou et al., 2017a , Zhou et al., 2017b . the measured PV loop is divided into an inspiratory and expiratory half cycles using the turning point Tmax at the maximum volume point, shown in Fig 1. The inspiration half cycle is then approximated using two segments with identified stiffness k1, k2 and the breakpoint Vm1 for inspiration. Similarly, the expiration half cycle is segmented with stiffness k3, k4 and the breakpoint Vm2 in HLA. The end-inspiratory parameter δ can also be calculated: where k2end is the end-inspiratory stiffness identified in HLA shown in Fig 1. A steady-state harmonic input force is assumed as the input force, defined: where Vmax and Vmin are the maximum and minimum volume in the PV loop, respectively. ξ is the damping ratio. The resistance parameter R can then be calculated: Finally, both the maximum energy parameter Em1 without considering plateau pressure and the smoothness parameter q can be obtained using forward simulation with the identified Vm1, Ke, Kh1 and Kh2 for inspiratory half cycle. In the identified HLM, the parameters Kh1, Kh2, R, Vm2, Em1, Kc, q and δ are fixed as constants, while the lung tissue elastance Ke is updated as PEEP changes. Particularly, a pair of linear and parabolic shaped basis functions (Morton et al., 2018 , Morton et al., 2019 are defined to the "post-yielding" ratio α representing the change of stiffness from alveolar recruitment to distention in the HLM: where E1 is the coefficient for distention basis function and E2 is the coefficient for recruitment basis function. To calculate E1 and E2 in Equation (14), the identified Ke1 and k1 at the first low PEEP level, PEEP1, can provide In addition, the maximum value for "post-yielding" ratio α is equal to 1 at the upper limit volume, yielding: Combining Equations (15) and (16) In addition, the prediction of Vm1 can also be determined based on changes of the transition radius from alveolar recruitment to distention. Clinical data of 4 patients and 8 staircase recruitment maneuver (RM) from the pilot CURE trial conducted at Christchurch Hospital ICU in 2016 are used to validate the modelling accuracy of the developed HLM, as well as the ability of the created virtual patient to predict lung response forward as PEEP changes (Davidson et al., 2014) . The New Zealand Southern Regional Ethics Committee grated ethics approval for this pilot trial. Pressure and flow data were recorded at a sampling rate of 50Hz from a Puritan Bennett 840 ventilator (Covidien Boulder, CO, USA). Patient demographics are presented in Table 1 . Each patient was ventilated with two pairs of RM, and each pair of RM includes a staircase increase in PEEP changes followed by a staircase decrease changes of PEEP. The method validation focuses on the two increasing staircase (Arm1 and Arm3) of RM for each patient, thus giving 8 data sets. Table 2 shows the details of PEEP changes for each data set of the 4 patient. Note the set value for PEEP in ventilator are not necessary matched by the real PEEP level performed in patient response. The median values across all breaths samples for each PEEP level are given in Table 2 . The PIP errors and RMS errors for all 8 data sets are listed in Table 3 , showing all absolute PIP error are less than 1.4% and RMS errors less than 1.7%, all of which validate the accuracy of the identification and modelling accuracy. The overall results validates the accuracy of identification and HLM modelling for capturing the essential lung physiology and respiratory behaviours at different patient conditions. The identified HLM at PEEP1 is combined with the prediction functions to predict the higher PEEP levels (PEEP2-5). More specifically, this virtual patient model created at PEEP1 provides not only one PEEP step higher (PEEP2) response prediction but also two (PEEP3), three (PEEP4) and four (PEEP5) steps ahead response with only using PEEP1 measurements. Fig. 4 shows a prediction example for comparing the predicted hysteresis to the clinical measurements at PEEP2-5 for Patient1, Arm1. The predicted PIP errors for all 8 data sets of the 4 patients are listed in Table 4 . It shows 57 out of 64 error values are less than 3%. The maximum absolute PIP error is 5.3% from Patient 2 who shows a negative compliance at PEEP1 and PEEP2 while volume is increasing known as spontaneously breathing (SB) (Damanhuri et al., 2016) . This SB results in a relatively larger error in predicting accurate values of elastance and response. However, the accuracies for Patient 2 at the highest PEEP (PEEP5) are very good, while the errors for other 3 patients are also very low with the maximum errors of 2.9% for PIP. These results show the unique capability of the created virtual patient for predicting both small and very large PEEP intervals using only one PEEP measurement. Fig. 4 . Response prediction to PEEP2-5 for Patient 1, Arm1. Finally, the change of Vfrc due to change of PEEP can be readily simulated using the predictive virtual patient. The predicted absolute Vfrc errors for all data sets are listed in Table 5 . The maximum errors is 0.096L again from Patient 2 as should be expected due to the maximum PIP errors. However, 20 out 32 predicted errors are less than 0.025L, which is 5% of a normal ventilated tidal volume (0.5L) (Salyer, 2007) , and 30 out of 32 are less than 0.05L. This work proposes an automated modelling approach based on a new developed HLM to predict the change of lung response to the change PEEP levels in a patient-specific fashion. Identification and modelling results at the low PEEP level show a very good match of the clinical and simulation hysteresis, which thus validates the identification accuracy and model efficiency. More importantly, the created virtual patient model with the proposed prediction functions is able to provide a very accurate prediction of both pressure and Vfrc at higher PEEP levels for the 4 patients. The developed algorithm requires no human-intervention, eliminating the potential subjective uncertainty due to the variations of clinician skills. Assessing respiratory mechanics using pressure reconstruction method in mechanically ventilated spontaneous breathing patient. Computer methods and programs in biomedicine Clinical Utilisation of Respiratory Elastance (CURE): Pilot trials for the optimisation of mechanical ventilation settings for the critically Ill Extrapolation of a non-linear autoregressive model of pulmonary mechanics Covid-19: most patients require mechanical ventilation in first 24 hours of critical care Biomedical engineer's guide to the clinical aspects of intensive care mechanical ventilation A virtual patient model for mechanical ventilation Predictive virtual patient modelling of mechanical ventilation: impact of recruitment function Essential emergency medicine: for the healthcare practitioner Physical Parameter Identification of Structural Systems with Hysteretic Pinching Damage assessment by stiffness identification for a fullscale three-story steel moment resisting frame building subjected to a sequence of earthquake excitations Comparing model-based adaptive LMS filters and a model-free hysteresis loop analysis method for structural health monitoring A new pinched nonlinear hysteretic structural model for automated creation of digital clones in structural health monitoring. Structural Health Monitoring