key: cord-0830432-t6s6fi0z authors: Suri, Jasjit S.; Agarwal, Sushant; Elavarthi, Pranav; Pathak, Rajesh; Ketireddy, Vedmanvitha; Columbu, Marta; Saba, Luca; Gupta, Suneet K.; Faa, Gavino; Singh, Inder M.; Turk, Monika; Chadha, Paramjit S.; Johri, Amer M.; Khanna, Narendra N.; Viskovic, Klaudija; Mavrogeni, Sophie; Laird, John R.; Pareek, Gyan; Miner, Martin; Sobel, David W.; Balestrieri, Antonella; Sfikakis, Petros P.; Tsoulfas, George; Protogerou, Athanasios; Misra, Durga Prasanna; Agarwal, Vikas; Kitas, George D.; Teji, Jagjit S.; Al-Maini, Mustafa; Dhanjil, Surinder K.; Nicolaides, Andrew; Sharma, Aditya; Rathore, Vijay; Fatemi, Mostafa; Alizad, Azra; Krishnan, Pudukode R.; Ferenc, Nagy; Ruzsa, Zoltan; Gupta, Archna; Naidu, Subbaram; Kalra, Mannudeep K. title: Inter-Variability Study of COVLIAS 1.0: Hybrid Deep Learning Models for COVID-19 Lung Segmentation in Computed Tomography date: 2021-11-01 journal: Diagnostics (Basel) DOI: 10.3390/diagnostics11112025 sha: 207d932daafd9f12e3237bbad5df01f1b3c122e7 doc_id: 830432 cord_uid: t6s6fi0z Background: For COVID-19 lung severity, segmentation of lungs on computed tomography (CT) is the first crucial step. Current deep learning (DL)-based Artificial Intelligence (AI) models have a bias in the training stage of segmentation because only one set of ground truth (GT) annotations are evaluated. We propose a robust and stable inter-variability analysis of CT lung segmentation in COVID-19 to avoid the effect of bias. Methodology: The proposed inter-variability study consists of two GT tracers for lung segmentation on chest CT. Three AI models, PSP Net, VGG-SegNet, and ResNet-SegNet, were trained using GT annotations. We hypothesized that if AI models are trained on the GT tracings from multiple experience levels, and if the AI performance on the test data between these AI models is within the 5% range, one can consider such an AI model robust and unbiased. The K5 protocol (training to testing: 80%:20%) was adapted. Ten kinds of metrics were used for performance evaluation. Results: The database consisted of 5000 CT chest images from 72 COVID-19-infected patients. By computing the coefficient of correlations (CC) between the output of the two AI models trained corresponding to the two GT tracers, computing their differences in their CC, and repeating the process for all three AI-models, we show the differences as 0%, 0.51%, and 2.04% (all < 5%), thereby validating the hypothesis. The performance was comparable; however, it had the following order: ResNet-SegNet > PSP Net > VGG-SegNet. Conclusions: The AI models were clinically robust and stable during the inter-variability analysis on the CT lung segmentation on COVID-19 patients. The WHO's International Health Regulations and Emergency Committee (IHREC) proclaimed COVID-19 a "public health emergency of international significance" or "pandemic" on 30 January 2020. More than 231 million people have been infected worldwide, and nearly 4.7 million people have died due to COVID-19 [1] . Although this "severe acute respiratory syndrome coronavirus 2" (SARS-CoV-2) virus specifically targets the pulmonary and vascular system, it has the potential to travel through the body and lead to complications such as pulmonary embolism [2] , myocardial infarction, stroke, or mesenteric ischemia [3] [4] [5] . Comorbidities such as diabetes mellitus, hypertension, and obesity substantially increase the severity and mortality of COVID-19 [6, 7] . A real-time reverse transcription-polymerase chain reaction (RT-PCR) is the recommended method for diagnosis [8] . Chest radiographs and computed tomography (CT) [9] [10] [11] are used to determine disease severity in patients with moderate to severe disease or underlying comorbidities based on the extent of pulmonary opacities such as ground-glass (GGO), consolidation, and mixed opacities in CT scans [7, [12] [13] [14] . Most radiologists provide a semantic description of the extent and type of opacities to describe the severity of COVID-19 pneumonia. The semiquantitative evaluation of pulmonary opacities is time-consuming, subjective, and tedious [15] [16] [17] [18] . Thus, there is a need for a fast and error-free early COVID-19 disease diagnosis and real-time prognosis solutions. Machine learning (ML) offers a solution to this problem by providing a rich set of algorithms [19] . Previously, ML has been used for detection of cancers in breast [20] , liver [21, 22] , thyroid [23] [24] [25] , skin [26, 27] , prostate [28, 29] , ovary [30] , and lung [31] . There are two main components in disease detection, i.e., segmentation [32] [33] [34] [35] and classification [36, 37] , where segmentation plays a crucial step. An extension of ML called deep learning (DL) employs dense layers to automatically extract and classify all relevant imaging features [38] [39] [40] [41] [42] [43] . Hybrid DL (HDL), a method that combines two DL systems, helps address some of the challenges in solo DL models [44, 45] . This includes overfitting and optimization of hyperparameters, thereby removing the bias [45] . During the AI model training, the most crucial stage is the ground truth (GT) annotation of organs that need to be segmented. It is a time-consuming operation with monetary constraints since skilled personnel such as radiologists are expensive to recruit and difficult to find. These annotations, if conducted by one tracer, make the AI system biased. A plurality of tracers being used to produce the GT annotated dataset makes the system more resilient and lowers the AI bias [46] [47] [48] [49] . This is because the AI model can grasp and adjust to the sensitivity of the difference in the tracings of the tracers. Thus, to avoid AI bias, one needs to have an automated AI-based system with multiple tracers. To establish the validity of such automated AI systems, one must undergo inter-variability analysis with two or more observers. To validate the AI systems, we hypothesize that two conditions must be met: (a) the two observers should perform within 5% range of each other and (b) the performance of the AI system using the ground truth tracings from these two observers should also be within the 5% threshold [48] . The AI performance is computed between the GT-area and the AI model-estimated area. The focus of the proposed research is to design a reliable AI system based on the inter-observer paradigm. Figure 1 depicts a COVID-19 CT lung segmentation system in which the CT machine is used to acquire CT volumes. This volume is then annotated by multiple observers (Figure 1 , n denotes the number of observers), and multiple AI models are generated, which is then used for lung segmentation. The segmentation output is the binary mask of the lung, its boundary, and the corresponding boundary overlays. This output can be used for evaluating the performance, analysis, and quantification of the results. The layout of this inter-variability study is as follows: Section 2 presents the methodology with the demographics, COVLIAS 1.0 pipeline, AI architectures, and loss functions. The experimental protocol is shown in Section 3, while results and performance evaluation are presented in Section 4. The discussions and conclusions are presented in Sections 5 and 6, respectively. The dataset consists of 72 adult Italian patients with 46 being male and the remaining being female. The mean height and weight were 173 cm and 79 kg, respectively. A total of 60 patients tested positive for RT-PCR, while 12 patients were confirmed using broncho- The layout of this inter-variability study is as follows: Section 2 presents the methodology with the demographics, COVLIAS 1.0 pipeline, AI architectures, and loss functions. The experimental protocol is shown in Section 3, while results and performance evaluation are presented in Section 4. The discussions and conclusions are presented in Sections 5 and 6, respectively. The dataset consists of 72 adult Italian patients with 46 being male and the remaining being female. The mean height and weight were 173 cm and 79 kg, respectively. A total of 60 patients tested positive for RT-PCR, while 12 patients were confirmed using broncho-alveolar lavage [50] . Overall, the cohort had an average of 4.1 GGO, which was considered low. All chest CT scans were performed in a supine posture during a single full inspiratory breath-hold using a 128-slice multidetector-row Philips Healthcare's "Philips Ingenuity Core" CT scanner. There were no intravenous or oral contrast media administrations. The CT exams were performed using a 120 kV, 226 mAs/slice (utilizing an automatic tube current modulation-Z-DOM by Philips), a 1.08 spiral pitch factor, 0.5-s gantry rotation time, and 64 × 0.625 detector setup. Soft tissue kernel with 512 × 512 matrix (mediastinal window) and lung kernel with 768 × 768 matrix (lung window) was used to reconstruct 1 mm-thick images. The Picture Archiving and Communication System (PACS) workstation that was utilized to review the CT images was outfitted with two Eizo 35 × 43 cm displays with a 2048 × 1536 matrix. Figure 2 shows the raw sample CT scans of COVID-19 patients with varying lung sizes and variable intensity patterns, posing a challenge. The proposed study makes use of the CT data of 72 COVID-positive individuals. Each patient had 200 slices, out of which the radiologist [LS] chose 65-70 slices from the visible lung region, resulting in 5000 images in total. The AI-based segmentation models were trained and tested using these 5000 images. To prepare the data for segmentation, a binary mask was created manually in a selected slice with the help of ImgTracer™ under the supervision of a qualified radiologist [LS] (Global Biomedical Technologies, Inc., Roseville, CA, USA) [47, 48, 51] . Figure 3 shows the white binary mask of the lung region computed using ImgTracer™ during manual tracings by Observer 1 and 2 (both were postgraduate researchers trained by our radiological team). COVLIAS 1.0 system incorporates three models: one solo DL (SDL) and two hybrid DL (HDL). The proposed study incorporates three AI models: (a) PSP Net, (b) VGG-SegNet, and (c) ResNet-SegNet. The Pyramid Scene Parsing Network (PSP Net) [52] is a semantic segmentation network with the ability to consider the global context of the image. The architecture of PSP Net ( Figure 4 ) has four parts: (i) input, (ii) feature map, (iii) pyramid pooling module, and (iv) output. The input to the network is the image to be segmented, which undergoes extraction of the feature map using a set of dilated convolution and pooling blocks. The dilated convolution layer is added at the last two blocks of the network to keep more prominent features at the end. The next stage is the pyramid pooling module; it is the heart of the network, as it helps capture the global context of the image/feature map generated in the previous step. This section consists of four parts, each with a different scaling ability. The scaling of this module includes 1, 2, 3, and 6, where 1 × 1 scaling helps capture the spatial features and thereby increases the resolution of the features captured. The 6 × 6 scaling captures the higher-resolution features. At the end of this module, all the output from these four parts is pooled using global average pooling. For the last part, the global average pooling output is fed to a set of convolutional layers. Finally, the set of prediction classes are generated as the output binary mask. erated in the previous step. This section consists of four parts, each with a different scaling ability. The scaling of this module includes 1, 2, 3, and 6, where 1 × 1 scaling helps capture the spatial features and thereby increases the resolution of the features captured. The 6 × 6 scaling captures the higher-resolution features. At the end of this module, all the output from these four parts is pooled using global average pooling. For the last part, the global average pooling output is fed to a set of convolutional layers. Finally, the set of prediction classes are generated as the output binary mask. The VGGNet architecture ( Figure 5 ) was designed to reduce the training time by replacing the kernel filter in the initial layer with an 11 and 5 sized filter, thereby reducing the # of parameters in the two-dimension convolution (Conv) layers [53] . The VGG-Se-gNet architecture used in this study is composed of three parts (i) encoder, (ii) decoder part, and (iii) a pixel-wise SoftMax classifier at the end. It consists of 16 Conv layers compared to the SegNet architecture, where only 13 Conv layers are used [54] in the encoder part. This increase in #layers helps the model extract more features from the image. The final output of the model is a binary mask with the lung region annotated as 1 (white) and the rest of the image as 0 (black). The VGGNet architecture ( Figure 5 ) was designed to reduce the training time by replacing the kernel filter in the initial layer with an 11 and 5 sized filter, thereby reducing the # of parameters in the two-dimension convolution (Conv) layers [53] . The VGG-SegNet architecture used in this study is composed of three parts (i) encoder, (ii) decoder part, and (iii) a pixel-wise SoftMax classifier at the end. It consists of 16 Conv layers compared to the SegNet architecture, where only 13 Conv layers are used [54] in the encoder part. This increase in #layers helps the model extract more features from the image. The final output of the model is a binary mask with the lung region annotated as 1 (white) and the rest of the image as 0 (black). Although VGGNet was very efficient and fast, it suffered from the problem of vanishing gradients. It results in significantly less or no weight training during backpropagation; at each epoch, it keeps getting multiplied with the gradient, and the update to the initial layers is very small. To overcome this problem, Residual Network or ResNet [55] came into existence ( Figure 6 ). In this architecture, a new connection was introduced known as skip connection which allowed the gradients to bypass a certain number of layers, solving the vanishing gradient problem. Moreover, with the help of one more additions to the network, i.e., an identity function, the local gradient value was kept to one during the backpropagation step. Although VGGNet was very efficient and fast, it suffered from the problem of vanishing gradients. It results in significantly less or no weight training during backpropagation; at each epoch, it keeps getting multiplied with the gradient, and the update to the initial layers is very small. To overcome this problem, Residual Network or ResNet [55] came into existence ( Figure 6 ). In this architecture, a new connection was introduced known as skip connection which allowed the gradients to bypass a certain number of layers, solving the vanishing gradient problem. Moreover, with the help of one more additions to the network, i.e., an identity function, the local gradient value was kept to one during the backpropagation step. The proposed system uses cross-entropy (CE)-loss during the training of the AI models. Equation (1) below represents the CE-loss, symbolized as l CE , for the three AI models: where x i represents the input GT label 1, (1 − x i ) represents the GT label 0, p i represents the probability of the classifier (SoftMax) used at the last layer of the AI model, and × represents the product of the two terms. Figures 4-6 presents the three AI architectures that have been trained using the CE-loss function. ishing gradients. It results in significantly less or no weight training during backpropagation; at each epoch, it keeps getting multiplied with the gradient, and the update to the initial layers is very small. To overcome this problem, Residual Network or ResNet [55] came into existence ( Figure 6 ). In this architecture, a new connection was introduced known as skip connection which allowed the gradients to bypass a certain number of layers, solving the vanishing gradient problem. Moreover, with the help of one more additions to the network, i.e., an identity function, the local gradient value was kept to one during the backpropagation step. The proposed system uses cross-entropy (CE)-loss during the training of the AI models. Equation (1) below represents the CE-loss, symbolized as , for the three AI models: where i represents the input GT label 1, (1i) represents the GT label 0, p i represents the probability of the classifier (SoftMax) used at the last layer of the AI model, and × represents the product of the two terms. Figures 4-6 presents the three AI architectures that have been trained using the CE-loss function. A standardized cross-validation (CV) protocol was adapted for determining the accuracy of the AI models. Our group has published several CV-based protocols of different kinds using AI framework [27, 30, 37, 56, 57] . Since the data were moderate, the K5 protocol was used, which consisted of 80% training data (4000 CT images) and 20% testing (1000 CT images). Five folds were designed in such a way that each fold got a chance to have a A standardized cross-validation (CV) protocol was adapted for determining the accuracy of the AI models. Our group has published several CV-based protocols of different kinds using AI framework [27, 30, 37, 56, 57] . Since the data were moderate, the K5 protocol was used, which consisted of 80% training data (4000 CT images) and 20% testing (1000 CT images). Five folds were designed in such a way that each fold got a chance to have a unique test set. An internal validation mechanism was part of the K5 protocol where 10% data was considered for validation. There were two methods used for quantification of the segmented lungs using AI models. The spirit of these two methods originates from the shape analysis concept. In the first method, lung area (LA) is computed since the region is balloon-shaped, thus the area parameter is well suited for the measurement [58, 59] . In the second method, we compute the long-axis of the lung (LLA) since the shape of the lung is more longitudinal than circular. A similar approach was taken for the long-axis view in heart computation [60] . The lung area (LA) was calculated by counting the number of white pixels in the binary mask segmented lungs, and the lung long axis (LLA) was calculated by the most distant distance segment joining anterior to posterior of the lungs. A resolution factor of 0.52 was used to convert (i) pixel to mm 2 for the LA and (ii) pixel to mm for the LLA computation and quantification. If the total number of the image is represented by N in the database, A ai (m, n) represents lung area for in the image "n" using the AI model "m", A ai (m) represents the mean lung area corresponding to the AI model "m," and mean area of the GT binary mask is represented by A gt , then mathematically A ai (m) and A gt can be computed as shown in Equation (2). Similarly, LA ai (m, n) represents LLA for in the image "n" using the AI model "m", LA ai (m) represents the mean LLA corresponding to the AI model "m,"LA gt represents the corresponding mean LLA of the GT binary lung mask, then mathematically LA ai (m) and LA gt can be computed as shown in Equation (3). The accuracy of the AI system was measured by comparing the predicted output and the ground truth pixel values. These values were interpreted as binary (0 or 1) numbers as the output lung mask was only black and white, respectively. Finally, these binary numbers were summed up and divided by the total number of pixels in the image. If TP, TN, FN, and FP represent true positive, true negative, false negative, and false positive, then the accuracy of the AI system can be computed as shown in Equation (4) [61] . Previously, COVLIAS 1.0 [54] was designed to run on a training: testing ratio of 2:3 dataset from 5000 images. However, this study proposes an inter-observer variability study with K5 in a CV framework. The training was performed on two sets of annotations, i.e., Observer 1 and Observer 2. The output results are similar to the previously published study, i.e., a binary mask of the segmented lungs. Figures 7-9 show the AI-generated binary mask, segmented lung, and color segmented lung with grayscale background as an overlay for the three AI models. Diagnostics 2021, 11, x FOR PEER REVIEW 11 of 40 study, i.e., a binary mask of the segmented lungs. Figure 7 , Figure 8 and Figure 9 show the AI-generated binary mask, segmented lung, and color segmented lung with grayscale background as an overlay for the three AI models. This section deals with the performance evaluation (PE) of the three AI models for Observer 1 vs. Observer 2. Section 4.2.1 presents the visual comparison of the results, which includes (i) boundary overlays against the ground truth boundary and (ii) lung long axis against the ground truth axis. Section 4.2.2 shows the PE for lung area error, which consists of (i) cumulative frequency (CF) plot, (ii) Bland-Altman plot, (iii) Jaccard Index (JI) and Dice Similarity (DS), and (iv) ROC and AUC curves for the three AI-based models' performance for Observer 1 vs. Observer 2. Similarly, lung long axis error (LLAE) presents PE using (i) cumulative plot, (ii) correlation coefficient (CC), and (iii) Bland-Altman plot. Finally, statistical analyses of the LA and LLA are presented using paired t-test, Wilcoxon, Mann-Whitney, and CC values for all 12 possible combinations for three AI models between Observer 1 and Observer 2. This section deals with the performance evaluation (PE) of the three AI models for Observer 1 vs. Observer 2. Section 4.2.1 presents the visual comparison of the results, which includes (i) boundary overlays against the ground truth boundary and (ii) lung long axis against the ground truth axis. Section 4.2.2 shows the PE for lung area error, which consists of (i) cumulative frequency (CF) plot, (ii) Bland-Altman plot, (iii) Jaccard Index (JI) and Dice Similarity (DS), and (iv) ROC and AUC curves for the three AI-based models' performance for Observer 1 vs. Observer 2. Similarly, lung long axis error (LLAE) presents PE using (i) cumulative plot, (ii) correlation coefficient (CC), and (iii) Bland-Altman plot. Finally, statistical analyses of the LA and LLA are presented using paired t-test, Wilcoxon, Mann-Whitney, and CC values for all 12 possible combinations for three AI models between Observer 1 and Observer 2. The overlay for the three AI model boundaries (green) and GT-boundary (red) corresponding to Observer 1 (left) and Observer 2 (right) with a grayscale COVID-19 CT slice in the background is shown in Figure 10 , while Figure 11 shows the AI-long axis (green) and GT-long axis (red) between Observer 1 and Observer 2 for three AI models. It shows the reach of anterior to posterior of the left and right lungs, with the GT boundary (white) corresponding to Observer 1 (left) and Observer 2 (right) of the lungs by the tracer using ImgTracer™. The three AI models follow the order: PSP Net, VGG-SegNet, and ResNet-SegNet. Figure 9 . Results from ResNet-SegNet while using Observers 1 and 2. Columns are the raw, binary mask output, segmented lung region, and overlay of the estimated lung region vs. ground truth region. The overlay for the three AI model boundaries (green) and GT-boundary (red) corresponding to Observer 1 (left) and Observer 2 (right) with a grayscale COVID-19 CT slice in the background is shown in Figure 10 , while Figure 11 shows the AI-long axis (green) and GT-long axis (red) between Observer 1 and Observer 2 for three AI models. It shows the reach of anterior to posterior of the left and right lungs, with the GT boundary (white) corresponding to Observer 1 (left) and Observer 2 (right) of the lungs by the tracer using ImgTracer™. The three AI models follow the order: PSP Net, VGG-SegNet, and ResNet-SegNet. The frequency of occurrence of the LAE is compared to a reference value in the cumulative frequency analysis and shown in Figure 12 (left lung) and Figure 13 (right lung) for three AI models between Observer 1 and Observer 2. A cutoff-score of 80% was chosen to show the difference between the three AI models. The LAE with the selected cutoff for the left lung was 1123.36 mm 2 , 725.90 mm 2 , and 571.65 mm 2 for the three AI models using Observer 1, and 834.08 mm 2 , 1730.58 mm 2 , and 683.42 mm 2 , respectively, for the three AI models using Observer 2. A similar trend was followed by the right lung with 1158.93 mm 2 , 612.47 mm 2 , and 532.44 mm 2 for the three AI models using Observer 1, and 809.77 mm 2 , 1610.15 mm 2 , and 572.56 mm 2 , respectively, for the three AI models using Observer 2. The three AI models follow the order: PSP Net, VGG-SegNet, and ResNet-SegNet. Correlation Plot for Lung Area Error Coefficient of correlations (CC) plots for the three AI models' LA vs. GT, area corresponding to the left and right between Observers 1 and 2, are shown in Figure 14 and Figure 15 . The CC values are summarized in Table 1 with a percentage difference between Observers 1 and 2. The percentage difference for the CC value (p < 0.001) ranges from 0% Figure 13 . Cumulative frequency plot of right LAE using three AI models: Observer 1 vs. Observer 2. Coefficient of correlations (CC) plots for the three AI models' LA vs. GT, area corresponding to the left and right between Observers 1 and 2, are shown in Figures 14 and 15 . The CC values are summarized in Table 1 with a percentage difference between Observers 1 and 2. The percentage difference for the CC value (p < 0.001) ranges from 0% to 2.04%, which is <5% as part of the error threshold chosen as the hypothesis. This clearly shows that the AI models are clinically valid for the proposed setting of the inter-observer variability study. to 2.04%, which is <5% as part of the error threshold chosen as the hypothesis. This clearly shows that the AI models are clinically valid for the proposed setting of the inter-observer variability study. Figure 14 . CC of left lung area using three AI models: Observer 1 vs. Observer 2. Figure 14 . CC of left lung area using three AI models: Observer 1 vs. Observer 2. Diagnostics 2021, 11, x FOR PEER REVIEW 19 of 40 Figure 15 . CC of right lung area using three AI models: Observer 1 vs. Observer 2. Figure 15 . CC of right lung area using three AI models: Observer 1 vs. Observer 2. Jaccard Index and Dice Similarity Figure 16 depicts a cumulative frequency plot for dice similarity (DS) for three AI models between Observers 1 and Observer 2. It shows that 80% of the CT images had a DS > 0.95. A cumulative frequency plot for the Jaccard Index (JI) is presented in Figure 17 and shows that 80% of the CT scans had a JI > 0.90 between Observer 1 and Observer 2. The three AI models follow the order: PSP Net, VGG-SegNet, and ResNet-SegNet. A Bland-Altman plot is used to demonstrate the consistency of two methods that employ the same variable. Based on our prior paradigms [48, 62] , we follow the Bland-Altman computing procedure. Figures 18 and 19 show the (i) mean and (ii) standard A Bland-Altman plot is used to demonstrate the consistency of two methods that employ the same variable. Based on our prior paradigms [48, 62] , we follow the Bland-Altman An ROC curve represents how an AI system's diagnostic performance changes as the discrimination threshold changes. Figure 20 shows the ROC curve and corresponding AUC value for the three AI models between Observer 1 and Observer 2. The three AI models follow the order: PSP Net, VGG-SegNet, and ResNet-SegNet. Figure 19 . BA for right LA using three AI models: Observer 1 vs. Observer 2. An ROC curve represents how an AI system's diagnostic performance changes as the discrimination threshold changes. Figure 20 shows the ROC curve and corresponding AUC value for the three AI models between Observer 1 and Observer 2. The three AI models follow the order: PSP Net, VGG-SegNet, and ResNet-SegNet. Cumulative Frequency Plot for Lung Long Axis Error Figures 21 and 22 show the cumulative frequency plot LLAE for left and right lung, respectively, corresponding to Observer 1 and Observer 2 for the three AI models. Based on the 80% threshold, the LLAE for the left lung ( Figure 21 ) using the three AI models for Observer 1 and Observer 2 were 6.12 mm (for PSP Net), 4.77 mm (for VGG-SegNet), and 5.01 mm (for ResNet-SegNet) and 10.88 mm (for PSP Net), 13.30 mm (for VGG-SegNet), and 9.18 mm (for ResNet-SegNet), respectively. Similarly, for the right lung (Figure 22 ), the error was 7.81 mm (for PSP Net), 5.47 mm (for VGG-SegNet), and 3.10 mm (for Res-Net-SegNet) and 9.14 mm (for PSP Net), 11.33 mm (for VGG-SegNet), and 6.88 mm (for ResNet-SegNet), respectively, for Observer 1 and Observer 2. The three AI models follow the order: PSP Net, VGG-SegNet, and ResNet-SegNet. Figures 21 and 22 show the cumulative frequency plot LLAE for left and right lung, respectively, corresponding to Observer 1 and Observer 2 for the three AI models. Based on the 80% threshold, the LLAE for the left lung ( Figure 21 ) using the three AI models for Observer 1 and Observer 2 were 6.12 mm (for PSP Net), 4.77 mm (for VGG-SegNet), and 5.01 mm (for ResNet-SegNet) and 10.88 mm (for PSP Net), 13.30 mm (for VGG-SegNet), and 9.18 mm (for ResNet-SegNet), respectively. Similarly, for the right lung (Figure 22 ), the error was 7.81 mm (for PSP Net), 5.47 mm (for VGG-SegNet), and 3.10 mm (for ResNet-SegNet) and 9.14 mm (for PSP Net), 11.33 mm (for VGG-SegNet), and 6.88 mm (for ResNet-SegNet), respectively, for Observer 1 and Observer 2. The three AI models follow the order: PSP Net, VGG-SegNet, and ResNet-SegNet. Correlation Plot for Lung Long Axis Error Figure 23 and Figure 24 show the CC plot for the three AI models considered in the proposed inter-observer variability study for Observers 1 and 2. Table 2 summarizes the CC values for the left, right, and mean errors of the LLA. It proves the hypothesis that the percentage difference between the results using the two observers has a difference of < Correlation Plot for Lung Long Axis Error Figures 23 and 24 show the CC plot for the three AI models considered in the proposed inter-observer variability study for Observers 1 and 2. Table 2 summarizes the CC values for the left, right, and mean errors of the LLA. It proves the hypothesis that the percentage difference between the results using the two observers has a difference of <5%. This demon-strates that the proposed system is clinically valid in the suggested inter-observer variability study context. 5%. This demonstrates that the proposed system is clinically valid in the suggested interobserver variability study context. Figure 23 . CC of left LLA for three AI models: Observer 1 vs. Observer 2. Figure 23 . CC of left LLA for three AI models: Observer 1 vs. Observer 2. Diagnostics 2021, 11, x FOR PEER REVIEW 29 of 40 Figure 24 . CC of right LLA using three AI models: Observer 1 vs. Observer 2. Figure 24 . CC of right LLA using three AI models: Observer 1 vs. Observer 2. The (i) mean and (ii) standard deviation of the lung long axis corresponding to Observer 1 and Observer 2 for the three AI models is shown in Figure 25 for the left lung and Figure 26 for the right lung. The (i) mean and (ii) standard deviation of the lung long axis corresponding to Observer 1 and Observer 2 for the three AI models is shown in Figure 25 for the left lung and Figure 26 for the right lung. The system's dependability and stability were assessed using a standard paired ttest, ANOVA, and Wilcoxon test. The paired t-test can be used to see if there is enough data to support a hypothesis; the Wilcoxon test is its alternative when the distribution is not normal. ANOVA helps in the analysis of the difference between the means of groups of the input data. MedCalc software (Osteen, Belgium) was used to perform the statistical analysis. To validate the system presented in this study, we have presented all the possible combinations (twelve in total) for the three AI models between Observer 1 and Observer 2. Table 3 shows the paired t-test, ANOVA, and Wilcoxon test results for the 12 combinations. Figure 26 . BA for the right LLA using the three AI models: Observer 1 vs. Observer 2. The system's dependability and stability were assessed using a standard paired t-test, ANOVA, and Wilcoxon test. The paired t-test can be used to see if there is enough data to support a hypothesis; the Wilcoxon test is its alternative when the distribution is not normal. ANOVA helps in the analysis of the difference between the means of groups of the input data. MedCalc software (Osteen, Belgium) was used to perform the statistical analysis. To validate the system presented in this study, we have presented all the possible combinations (twelve in total) for the three AI models between Observer 1 and Observer 2. Table 3 shows the paired t-test, ANOVA, and Wilcoxon test results for the 12 combinations. The likelihood of the error in the system is known as the figure of merit (FoM). We have calculated FoM for (i) lung area and (ii) lung long axis to show the acceptability of the hypothesis if the % difference between the two observers is <5%. Table 4 shows the values for FoM using Equation (5) and the % difference for the three AI models against the two observers. Similarly, Table 5 shows the values for FoM using Equation (6) and the % difference for the three AI models against the two observers. The study presented the inter-observer variability analysis for the COVLIAS 1.0 using three AI models, PSP Net, VGG-SegNet, and ResNet-SegNet. These models have considered tissue characterization approaches since they analyze the tissue data for better feature extraction to evaluate for ground vs. background, thus are more akin to a tissue characterization in classification framework [30, 37] . Our group has strong experience in tissue characterization approaches with different AI models and applications for classification using ML frameworks such as plaque, liver, thyroid, breast [21, 28, 30, [63] [64] [65] [66] [67] [68] , and DL framework [1, 36, 69, 70] . These three AI models were trained using the GT annotated data from the two observers. The percentage difference between the outputs of the two AI model results was less than 5%, and thus the hypothesis was confirmed. During the training, the K5 cross-validation protocol was adapted on a set of 5000 CT images. For the PE of the proposed inter-observer variability system, the following ten metrics were considered: (i) visualization of the lung boundary, (ii) visualization of the lung long axis, cumulative frequency plots for (iii) LAE, (iv) LLAE, CC plots for (v) lung area, (vi) lung long axis, BA plots for (vii) lung area, (viii) lung long axis, (ix) ROC and AUC curve, and (x) JI and DS for estimated AI model lung regions. These matrices showed consistent and stable results. The training, evaluation, and quantification were implemented on the GPU environment (DGX V100) using python. We adapted vectorization provided by python during the implementation of the Numba library. The proposed inter-observer variability study used three AI models for the analysis, where PSP Net was implemented for the first time for COVID-19 lung segmentation. The other models VGG-SegNet and ResNet-SegNet were used for benchmarking. The AUC for the mean lung region for the three AI models was >0.95 for both Observer 1 and Observer 2. Our results, shown below in Table 6 , compared various metrics that included the interobserver variability study for the three AI models. All the models behaved consistently while using the two different observers. Our results showed that ResNet-SegNet was the best performing model for all the PE metrics. The percentage difference between the two observers was 0.4%, 3.7%, and 0.4%, respectively, for the three models PSP Net, VGG-SegNet, and ResNet-SegNet, respectively. This further validated our hypothesis for every AI model, keeping the error threshold less than 5%. Even though all three AI models passed the hypothesis, VGG-SegNet is the least superior. This is because the number of the layers in the VGG-SegNet architecture ( Figure 5 ) is 19, compared to~50 in PSP Net ( Figure 4 ) and 51 (encoder part) in the ResNet-SegNet model ( Figure 6 ). By taking the results from both the observers into account, the order of the performance of the models is ResNet-SegNet > PSP Net > VGG-SegNet. Further, we also conclude that HDL models are superior to SDL (PSP Net). The aggregate score was computed as the mean for all the models for Observer 1, Observer 2, and the mean of the two Observers. Even though the performance of all the models was comparable, when carefully looking at the performance of Observer 1 the order of performance was ResNet-SegNet > VGG-SegNet > PSP Net. For Observer 2, the order of performance was ResNet-SegNet > PSP Net > VGG-SegNet. Further, the performance of the left lung was better than the right lung for the reasons unclear at this point, and more investigations would be needed to evaluate this. There have been several studies in the area of DL for lung segmentation, but only a few in the region of COVID-19 [71] [72] [73] [74] , and even less that involved variability analysis. Table 7 shows the benchmarking table having three variability studies: Saba et al. [48] , Jeremy et al. [75] , and Joskowicz et al. [76] , that are compared against Suri et al. in this proposed study. Saba et al. has used a dataset of 96 patients with three observers for tracings, and ROC curves were also not presented in the study. Jeremy et al. [60] have demonstrated the variability analysis using five different observers that used the area error as the metric. The boundary error, ROC, JI, and DS were not discussed. Finally, Joskowicz et al. [76] used 480 images and 11 observers to annotate the dataset, but no area and boundary errors were present. Moreover, they did not present the ROC curves, JI, and DS for the tracings. All three studies [48, 75, 76] , only performed manual annotation of the non-COVID dataset, and there was no involvement of the AI techniques to generate the boundaries automatically. Comparatively, the proposed study provides a first-of-its-kind for inter-observer variability analysis alongside HDL and SDL solutions, supporting our hypothesis that the error between the AI models trained using the two observers involved is less than 5%. The proposed study successfully validated the hypothesis for the inter-observer variability settings, demonstrating that the difference between the two AI models when trained by the two observers was less than 5%. It was the first-time inter-observer variability was presented for COVID-19 lung segmentation using HDL and SDL models. In spite of encouraging results, the study could not include more than two observers due to reasons such as cost, time, and availability of the radiologists. The imaging analysis component could be extended to handle more dense pulmonary opacities such as consolidation or mixed opacities during lung segmentation. As part of the extension, the HDL models can be extended, which combines DL with ML or two solo DL models for lung segmentation. Conventional methods [77, 78] can be used for lung segmentation embedded with denoising methods [79] and benchmarked against the AI models. The system can be extended to unseen data where the training data is taken from one clinical site and testing data can be from the other clinical site. It would also be interesting to explore the segmentation of lungs in the healthy patients using the AI model trained on COVID-19 patients. Other neural network techniques such as generative adversarial networks (GANs) [80] or transfer learning and loss schemes [38, 44, 81] can also be adapted. A big data framework can be used to integrate comorbidity factors [82] in the AI models. The proposed study is the first of its kind to evaluate the effect of ground-truth tracings on the AI models for COVID-19 CT lung segmentation. Three kinds of AI models, PSP Net, VGG-SegNet, and ResNet-SegNet, were adapted for lung segmentation. Two different Observers were used to annotate 5000 CT lung slices taken from 72 COVID-19 patients. Thus, six AI training models (three AI models times two Observers) were generated and evaluated using the K5 cross-validation protocol. Ten different kinds of metrics were used for the evaluation of the six AI models. The two Observers' error metrics were compared to validate the hypothesis for every AI model, keeping below the error threshold of 5%. Our results showed that the difference in these errors were 0%, 0.51%, and 2.04% (all < 5%), respectively, for the three AI models, validating the hypothesis. Statistical analysis was conducted using a standard paired t-test, ANOVA, and Wilcoxon test to prove the system's hypothesis. The inter-variability COVLIAS 1.0 showed clinically robust and statistically stable outcomes for this pilot study and, thus, can be adapted in clinical settings. The authors declare no conflict of interest. GBTI deals in lung image analysis and Jasjit S. Suri is affiliated with GBTI. LA ai (m, n) Estimated lung long axis using AI model 'm' and image 'n' 10 LA gt (n) GT lung long axis for image 'n' 11 LA gt Mean ground truth long axis for all images N in the database 12 FoM A (m) Figure- of-Merit for segmentation model 'm ' 14 FoM LA (m) Figure- Wilson disease tissue classification and characterization using seven artificial intelligence models embedded with 3D optimization paradigm on a weak training brain magnetic resonance imaging datasets: A supercomputer application Complications in COVID-19 patients: Characteristics of pulmonary embolism Molecular pathways triggered by COVID-19 in different organs: ACE2 receptor-expressing cells under attack? A review Imaging in COVID-19-related myocardial injury Bidirectional link between diabetes mellitus and coronavirus disease 2019 leading to cardiovascular disease: A narrative review A narrative review on characterization of acute respiratory distress syndrome in COVID-19-infected lungs using artificial intelligence Computed tomography findings of COVID-19 pneumonia in Intensive Care Unit-patients Real-time reverse transcription-polymerase chain reaction assay for SARS-associated coronavirus Deep learning-based multi-view fusion model for screening 2019 novel coronavirus pneumonia: A multicentre study Deep transfer learning based classification model for COVID-19 disease Multi-Detector CT Imaging: Principles, Head, Neck, and Vascular Systems Rapid ai development cycle for the coronavirus (COVID-19) pandemic: Initial results for automated detection & patient monitoring using deep learning ct image analysis Automated detection of COVID-19 using ensemble of transfer learning with deep convolutional neural network based on CT scans A CT scan dataset about COVID-19. arXiv 2020 Detection Using X-ray Images and Artificial Intelligence Hybrid Systems CNN-based transfer learning-BiLSTM network: A novel approach for COVID-19 infection detection An explainable COVID-19 diagnosis system by joint classification and segmentation Recent Advances in Breast Imaging, Mammography, and Computer-Aided Diagnosis of Breast Cancer Symtosis: A liver ultrasound tissue characterization and risk stratification in optimized deep learning paradigm Data mining framework for fatty liver disease classification in ultrasound: A hybrid feature extraction paradigm Non-invasive automated 3D thyroid lesion classification in ultrasound: A class of ThyroScan™ systems A review on ultrasoundbased thyroid cancer tissue characterization and automated classification Characterization of single thyroid nodules by contrast-enhanced 3-D ultrasound Computer-aided diagnosis of psoriasis skin images with HOS, texture and color features: A first comparative study of its kind Reliable and accurate psoriasis disease classification in dermatology images using comprehensive feature space in machine learning paradigm Prostate tissue characterization/classification in 144 patient population using wavelet and higher order spectra features from transrectal ultrasound images El-Baz, A. In-vitro and in-vivo diagnostic techniques for prostate cancer: A review Evolutionary algorithm based classifier parameter tuning for automatic diabetic retinopathy grading: A hybrid feature extraction approach Lung disease stratification using amalgamation of Riesz and Gabor transforms in machine learning framework Biomedical Image Segmentation: Advances and Trends Shape recovery algorithms using level sets in 2-D/3-D medical imagery: A state-of-the-art review Multi Modality State-of-the-Art Medical Image Segmentation and Registration Methodologies Level Set Method in Medical Imaging Segmentation Multimodality carotid plaque tissue characterization and classification in the artificial intelligence paradigm: A narrative review for stroke application Automated classification of patients with coronary artery disease using grayscale features from left ventricle echocardiographic images A novel block imaging technique using nine artificial intelligence models for COVID-19 disease classification, characterization and severity measurement in lung computed tomography scans on an Italian cohort Six artificial intelligence paradigms for tissue characterisation and classification of non-COVID-19 pneumonia against COVID-19 pneumonia in computed tomography lungs 3-D optimized classification and characterization artificial intelligence paradigm for cardiovascular/stroke risk stratification using carotid ultrasound-based delineated plaque: Atheromatic™ 2.0 Multiclass magnetic resonance imaging brain tumor classification using artificial intelligence paradigm Web-based efficient dual attention networks to detect COVID-19 from X-ray images A Means of Assessing Deep Learning-Based Detection of ICOS Protein Expression in Colon Cancer Hybrid deep learning segmentation models for atherosclerotic plaque in internal carotid artery B-mode ultrasound Artificial Intelligence-based Hybrid Deep Learning Models for Image Classification: The First Narrative Review Systematic Review of Artificial Intelligence in Acute Respiratory Distress Syndrome for COVID-19 Lung Patients: A Biomedical Imaging Perspective Intra-and inter-operator reproducibility of automated cloud-based carotid lumen diameter ultrasound measurement Inter-observer Variability Analysis of Automatic Lung Delineation in Normal and Disease Patients Inter-and Intra-Observer Variability Assessment of in Vivo Carotid Plaque Burden Quantification Using Multi-Contrast Dark Blood MR Images Factors limiting the utility of bronchoalveolar lavage in the diagnosis of COVID-19 Accurate cloud-based smart IMT measurement, its validation and stroke risk stratification in carotid ultrasound: A web-based point-of-care tool for multicenter clinical trial Pyramid scene parsing network Very deep convolutional networks for large-scale image recognition COVLIAS 1.0: Lung Segmentation in COVID-19 Computed Tomography Scans Using Hybrid Deep Learning Artificial Intelligence Models Deep residual learning for image recognition An accurate and generalized approach to plaque characterization in 346 carotid ultrasound scans Ovarian tumor characterization and classification: A class of GyneScan™ systems Shape-based approach for coronary calcium lesion volume measurement on intravascular ultrasound imaging and its association with carotid intima-media thickness Three different strategies for real-time prostate capsule volume computation from 3-D end-fire transrectal ultrasound Left ventricle longitudinal axis fitting and its apex estimation using a robust algorithm and its performance: A parametric apex model Risk stratification of 2D ultrasound-based breast lesions using hybrid feature selection in machine learning paradigm Statistics in Medicine Symptomatic vs. asymptomatic plaque classification in carotid ultrasound Diagnosis of Hashimoto's thyroiditis in ultrasound using tissue characterization and pixel classification Understanding symptomatology of atherosclerotic plaque by image-based tissue characterization Symptomatic vs. asymptomatic classification of carotid ultrasound plaque using a combination of HOS, DWT & texture Atherosclerotic plaque tissue characterization in 2D ultrasound longitudinal carotid scans for automated classification: A paradigm for stroke risk assessment Accurate and automatic carotid plaque characterization in contrast enhanced 2-D ultrasound images Ultrasound-based carotid stenosis measurement and risk stratification in diabetic cohort: A deep learning paradigm Deep learning fully convolution network for lumen characterization in diabetic patients using carotid ultrasound: A tool for stroke risk Deep CNN models for predicting COVID-19 in CT and x-ray images COVIDNet-CT: A Tailored Deep Convolutional Neural Network Design for Detection of COVID-19 Cases From Chest CT Images Performance analysis of lightweight CNN models to segment infectious lung tissues of COVID-19 cases from tomographic images Lung Infection Segmentation for COVID-19 Pneumonia Based on a Cascade Convolutional Network from CT Images Interobserver and intraobserver variability in measurement of non-small-cell carcinoma lung lesions: Implications for assessment of tumor response Inter-observer variability of manual contour delineation of structures in CT Lung Imaging and CADx Lung Imaging and Computer Aided Diagnosis Speckle reduction in medical ultrasound images using an unbiased non-local means method Skin lesion segmentation using a lightweight generative adversarial network Brain MRI-based Wilson disease tissue classification: An optimised deep transfer learning approach Big Data in Multimodal Medical Imaging