key: cord-0826101-b64zxez7 authors: Rangaiah, Ambica; Shankar, Sathyanarayan Muthur; Basawarajappa, Shantala Gowdara; Shah, Pritik A; Chandrashekar, Aditya; Munegowda, Asha; Padukone, Shashiraja title: Detection of SARS-CoV2 in Clinical Samples: Target-specific Analysis of Qualitative Reverse Transcription–Polymerase Chain Reaction(RT-PCR) Diagnostic Kits date: 2021-11-18 journal: IJID Regions DOI: 10.1016/j.ijregi.2021.11.004 sha: 94f48ba4da7e529448673b1b1068ce89ee0db4a2 doc_id: 826101 cord_uid: b64zxez7 INTRODUCTION The Coronavirus disease 2019 pandemic caused by Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) led to the rise of many available modalities for diagnosis. One such modality is the Reverse Transcription–Polymerase Chain Reaction (RT-PCR) kits which require evaluation amongst the many available commercial kits in the market. METHODS We conducted a performance evaluation of twelve RT‐PCR SARS‐CoV‐2 commercial kits. A total of 75 nasopharyngeal and oropharyngeal clinical samples were selected with their cycling threshold (Ct) values. Inclusion of 5 gene targets: E gene, N gene, S gene, RdRp and ORF1ab. Data was analyzed using R software version 4.1.1 and Microsoft Excel RESULTS We observe that, the positive sample's Ct values differs significantly across the 12 diagnostic kits. However, for gene-specific analysis, we observe that, positive sample's Ct values does not differ significantly across gene targets. There is significant difference in Ct values in Commercial kits targeting all genes except S-gene. All the commercial kits Altona (E and S genes), Thermo (ORF1ab and N genes), Multiplex (E, ORF1ab, RdRdp genes), Meril (N and ORF1ab genes), S D Biosensor (E and ORF1ab genes), Lab Gun (RdRp and N genes) and Lab systems (ORF1ab and E genes) scored a sensitivity of 100%. All other kits scored sensitivity above 95% and lowest sensitivity with the Genes2me (E gene) and Genes2me (RdRp) at 95.08% each. All kits were 100% specific. CONCLUSION This study provides an accurate comprehensive assessment of the different kits in the detection of SARS-CoV-2 which may promote standardization of testing across laboratories. The Coronavirus disease 2019 , caused by Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) which originated from Wuhan, China became a global pandemic which enforced the World Health Organisation(WHO) to label it as a global health concern. It has led to an estimated 179 million confirmed infections and 3 million deaths worldwide and in India an estimated 30 million cases and 3 hundred thousand plus deaths as of July 2021(Zhou P et.al, 2020 and Babu GR et.al,2021) .As the pandemic progressed, the diagnosis of this disease continuously evolved creating the need of many currently available modalities for diagnosis . There are a number of laboratory tests like antigen detection kits, Reverse Transcriptase Polymerase Chain Reactions (RT-PCR), cartridge based nuclear amplification tests etc., however the most widely used and current gold standard for the diagnosis of COVID-19 is Reverse Transcriptase -Polymerase Chain Reaction. SARS-CoV-2 virus belong to the family Coronaviridae and is a positive sense RNA virus which has 14 open reading frames(ORF) that code for the numerous structural and non-structural proteins. The RNA dependent RNA polymerase (RdRp) which is essential for the replication and transcription is encoded by an open reading frame known as ORF1ab. The ORF1ab as well as the genes which encode for the structural proteins like the spike protein(S), envelope protein(E) and nucleocapsid(N) proteins are the major target sites used for the detection of covid19 by RTPCR (Kirtipal N.et.al,2020 , Lu X.et.al,2020 and Barreto HG.et.al,2020) The RT-PCR tests offer several benefits than its other diagnostic counterparts in being very useful for early-stage diagnosis of the infection when antibody titers aren't significantly high. Its added advantages being that it can also be performed on a wide range of specimens including nasal swabs, pharyngeal swabs, sputum, bronchoalveolar lavage fluid, fiber bronchoscope brush biopsies, blood and faces. Furthermore, the RT-PCR has high sensitivity and specificity providing the test results within a few hours making it the most widely used test. Due to the escalating pandemic and the increased demand of easier , more accurate and rapid diagnostic methods, the development of commercial RT-PCR kits has increased , creating different manufactured kits in the market . However the there is still insufficient literature on the independent performance assessment among these kits. Increased accuracy of these kits and identifying them would help us to reduce the burden on the health care system and aid in accurately identifying the required population suffering in this pandemic (Li T.et.al, 2020) . In the State-Level Virus Research and Diagnostics Laboratory (VRDL), Bangalore Medical College and Research Institute, Bangalore, we performed an evaluation of twelve RT-PCR SARS-CoV-2 commercial kits to evaluate the performance characteristics and assess their accuracy. This study targets to provide a comprehensive assessment of the different kits used for the detection of SARS-CoV-2 via RT-PCR which may promote standardization of testing across laboratories, leading to the development of even better diagnostic kits in the future in India. A total of 75 nasopharyngeal and oropharyngeal clinical samples were selected in continuum along with their cycling threshold (Ct) values. The ethical approval was obtained from the Institution Ethics Committee (IEC) as per Certificate Number: BMCRI/PS/298/2020-21 Dt.06.02.2021/IEC. This study encompasses 12 commercial RT-PCR SARS-CoV-2 detection kits available in the Indian market and approved by the Indian Council of Medical Research (ICMR). Kits were obtained from the Karnataka State Mineral Corporation Limited for Covid-19 for analysis as a part of the Karnataka State Government Initiative. None of the manufacturers were involved in the assessment or data analysis. The ICMR listed diagnostic kits were reviewed and analyzed based on availability and need for study. The details of regulation status, country of manufacturer, reaction volume, cycling time, thermocycler used, RNA template volume(microliter), storage conditions, internal control and target genes for all 12 commercial kits have been elaborated in Table1.0. Inclusion of 5 gene targets over these 12 kits are as follows : E gene, N gene , S gene, RdRp and ORF1ab. All the clinical samples used for kit validations were collected from patients attending the Screening Room at the Out Patient Department along with informed consent. 75 samples were included for the study. Clinical samples collected in Viral Transport Media(VTM) were subjected to thorough vortexing, and 150 µl of the specimen was further subjected To QIAamp Viral RNA Mini Kit in accordance to the QIAGEN manual RNA extraction protocol as per manufacturer's instruction. Realtime RT-PCR for all the kits were carried out in a span of less than 3 days. The real-time RT-PCR assays were performed on Thermofisher Quant Studio 5 thermal cycler. The assay procedures and interpretation of the results were done as per the manufacturer's instructions. Data was analysed using R software version 4. negative samples. Hence, the same is considered as gold standard. One positive sample was taken as a positive control. We observe that, the positive sample's Ct values differs significantly across the 12 diagnostic kits (gene The commercial kits had 5 different gene targets (E, N, S, RdRP and ORF1ab). We observe that, positive sample's Ct values does not differ significantly across gene targets. The graph in We observe that, there is significant difference in Ct values in commercial kits targeting E gene. We observe that, there is no significant difference in Ct values in Commercial kits targeting S gene (Altona and Thermo , p value= 0.6595). Figure 4 depicts Mean plot of Ct values in Commercial kits targeting S gene. We observe that, there is significant difference in Ct values in Commercial kits targeting N gene. From post hoc analysis (Tukey's HSD), we observe that, Ct values from Allplex differs significantly from Lab Gun (p-value<0.001), Meril (p-value<0.01) and Thermo (p-value = 0.0014), Lab Gun differs significantly from Meril (p-value <0.001) and Thermo (p-value<0.001). We observe that, there is significant difference in Ct values in Commercial kits targeting RdRp gene. From post hoc analysis (Games-Howell test), we observe that, Ct values from Allplex differs significantly from Lab Gun (p-value<0.001) and QLINE (p-value<0.001), Genes2me differs significantly from Lab Gun (p-value<0.001) and QLINE (p-value = 0.005), Lab Gun differs significantly from NIV Multiplex V 3.1 (p-value < 0.001), Saragene (p-value < 0.001) and TRUPCR (p-value<0.001), Multiplex varies significantly from QLINE (p-value < 0.001), QLINE differs significantly from Saragene (p-value < 0.001) and TRUPCR (p-value < 0.001). Figure 4 depicts the Mean plot of Ct values in Commercial kits targeting RdRp gene. We observe that, there is significant difference in Ct values in Commercial kits targeting ORF1ab gene. The commercial kits Altona (E and S genes), Thermo (ORF1ab and N genes), Multiplex (E, ORF1ab, RdRdp genes), Meril (N and ORF1ab genes), S D Biosensor (E and ORF1ab genes), Lab Gun (RdRp and N genes) and Lab systems (ORF1ab and E genes) scored a sensitivity of 100%. All other kits scored sensitivity above 95%. The lowest sensitivity was observed with the Genes2me (E gene) and Genes2me (RdRp) with 95.08% each. For specificity, all the tested kits scored 100%. For the Kappa agreement tests, the highest score was 100% and was observed with Altona (E and S genes), Thermo (ORF1ab and N genes), Multiplex (E, ORF1ab, RdRdp genes), Meril (N and ORF1ab genes), S D Biosensor (E and ORF1ab genes), Lab Gun (RdRp and N genes) and Lab systems (ORF1ab and E genes) kits, the other kits scored an agreement above 90% except for Genes2me (E gene) and Genes2me (RdRp) which had 88% agreement each. The statistical summaries of all commercial kits on genes, sensitivity, specificity, Having passed two major waves of the coronavirus disease 2019 (COVID-19) pandemic, many testing kits were developed and were made available commercially. The ICMR has approved several diagnostic kits for several gene-specific analysis and data validation. In our study we evaluate and analyse the performance of 12 commercial kits in detecting the various genes of the SARS-CoV-2 virus. We performed target gene specific analysis from which we derived that although the commercial kits had 5 different gene targets (E, N, S, RdRP and ORF1ab), the positive sample's cycling threshold (Ct) values differs significantly across gene targets. From post hoc analysis, we observe that, Ct value of different gene targets did not differ significantly as elaborated in the results above. Among the kits we also notice that Altona (E and S genes), Thermo (ORF1ab and N genes), Multiplex (E, ORF1ab, RdRdp genes), Meril (N and ORF1ab genes), S D Biosensor (E and ORF1ab genes), Lab Gun (RdRp and N genes) and Lab systems (ORF1ab and E genes) scored a sensitivity of 100%. All other kits scored sensitivity above 95%. The lowest sensitivity was observed with the Genes2me (E gene) and Genes2me (RdRp) with 95.08% each. The study conducted by Altamimi A.M et.al (Altamimi AM.et.al, 2021) on the performance of these commercial kits by gene target showed no significant change in Ct values which indicates that kits disparities are mainly linked to the choice of the gene target. Their study showed that RdRP and E gene targets show significant differences by the reported Ct values. However in our study Ct-value for different gene targets did not differ significantly with the others. Comparing commercial kits showed similar results in detection, however, different targets did not show variation in Ct values. Moreover, kits with multiple targets such as in another evaluation study did show higher sensitivity and specificity than with other kits detecting a single target (Iglói Z.et.al, 2020 and Basawarajappa SG.et.al,2021) The escalation of these various mutations in SARS-CoV-2 in multiple geographic locations indicates the need for regular genetic testing in each country to capture these changes as they are used in the selection of the commercial diagnostic kit used in health centres [9] . An issue regarding the current situation in India is regarding the rise of the delta variant. India has experienced multiple outbreaks of SARS-CoV2 in 2020 and has reached a population sero-positivity of more than 50% by 2021. We hope by the results, new data and analysis can be done for the analysis of RT-PCR versus genomic detection of the new strains and variants (Dhar MS.et.al,2021) . But recent reports and current evidence from scientific literature tell a different story. There was speculation that SARS-CoV-2 escaped RTPCR due to a series of mutations received. Host-dependent RNA editing and high human-to-human transmission may have mutated the virus, thereby allowing the virus to spread faster and even escape RT-PCR ( Jindal H.et.al, 2021) . Limitations in our study include the absence of evaluation of the limit of detection which can play a big role, as with several of the commercial kits approved for the pandemic situation many were approved without evaluation with appropriate numbers. Another limitation is that we were unable to examine the cross-reactivity of these commercial kits with other viruses, which can significantly alter the detection results. As we were not able to include all the RT-PCR diagnostic kits available on the market and the performance of some kits may vary depending on the extraction method, there may be a limitation posed as well. Lastly, in our study our sample size is limited to the number of tests per kit for that particular Lot/Batch number and might vary with manufacturer provisions. In conclusion, we entrust upon the fact that most of the commercially available RT-PCR kits endorsed in this study can be used for routine diagnosis of clinical samples suspicious of SARS-CoV-2 patients. Most of the kits were validated in detecting the virus, however, few distinctions were found with specific kits and targets. In comparison with other studies our study covers the difference in Ct-values with respect to target specific gene-analysis in a detailed comprehensive manner. Multiple targets have been shown to be more sensitive and specific as they counted for target variations. New methods for diagnosis have been recently mentioned in literature, such as, saliva samples and oral dry swabs amongst many others (Teo, A.et.al, 2021) . In addition, we recommend that regardless of the choice of commercial diagnostic kit for the clinical detection of COVID-19 patients in the laboratory, a good validation plan and collaboration with external laboratories and government agencies are required to identify virus changes, procedures, technicians and the various kit services. The author(s) declare that there are no conflicts of interest This work received no specific grant from any funding agency. Karnataka State Mineral Corporation Limited provided the RT-PCR kit required for this study. The ethical approval was obtained from the Institution Ethics Committee (IEC) as per Certificate Number: BMCRI/PS/298/2020-21 Dt.06.02.2021/IEC. Approval has been obtained from Institution Ethics Committee (IEC). Assessment of 12 qualitative RT-PCR commercial kits for the detection of SARS-CoV-2 The burden of active infection and anti-SARS-CoV-2 IgG antibodies in the general population: Results from a statewide sentinel-based population survey in Diagnosing the novel SARS-CoV-2 by quantitative RT-PCR: variations and opportunities Performance evaluation of Truenat™ Beta CoV & Truenat™ SARS-CoV-2 point-of-care assays for coronavirus disease 2019 Genomic characterization and epidemiology of an emerging SARS-CoV-2 variant in Delhi Comparison of commercial realtime reverse transcription PCR assays for the detection of SARS-CoV-2 False-Negative RT-PCR Findings and Double Mutant Variant as Factors of an Overwhelming Second Wave of COVID-19 in India: an Emerging Global Health Disaster From SARS to SARS-CoV-2, in-sights on structure, pathogenicity and immunity aspects of pandemic human coronaviruses Diagnosis and clinical management of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) infection: an operational recommendation of Peking Union Medical College Hospital (V2. 0) working group of 2019 novel coronavirus, Peking union medical college hospital US CDC real-time reverse transcription PCR panel for detection of severe acute respiratory syndrome coronavirus 2 Saliva is more sensitive than nasopharyngeal or nasal swabs for diagnosis of asymptomatic and mild COVID-19 infection Mean plot of CT values in Commercial kits targeting We acknowledge the support by