key: cord-0825092-lpjzz169 authors: Wormald, Benjamin W.; Moser, Nicolas; deSouza, Nandita M.; Mantikas, Katerina-Theresa; Malpartida-Cardenas, Kenny; Pennisi, Ivana; Ind, Thomas E. J.; Vroobel, Katherine; Kalofonou, Melpomeni; Rodriguez-Manzano, Jesus; Georgiou, Pantelis title: Lab-on-chip assay of tumour markers and human papilloma virus for cervical cancer detection at the point-of-care date: 2022-05-24 journal: Sci Rep DOI: 10.1038/s41598-022-12557-y sha: 7de747492354a517cf3c1f84ebbfca78d64756c5 doc_id: 825092 cord_uid: lpjzz169 Cervical cancer affects over half a million people worldwide each year, the majority of whom are in resource-limited settings where cytology screening is not available. As persistent human papilloma virus (HPV) infections are a key causative factor, detection of HPV strains now complements cytology where screening services exist. This work demonstrates the efficacy of a handheld Lab-on-Chip (LoC) device, with an external sample extraction process, in detecting cervical cancer from biopsy samples. The device is based on Ion-Sensitive Field-Effect Transistor (ISFET) sensors used in combination with loop-mediated isothermal amplification (LAMP) assays, to amplify HPV DNA and human telomerase reverse transcriptase (hTERT) mRNA. These markers were selected because of their high levels of expression in cervical cancer cells, but low to nil expression in normal cervical tissue. The achieved analytical sensitivity for the molecular targets resolved down to a single copy per reaction for the mRNA markers, achieving a limit of detection of 10(2) for hTERT. In the tissue samples, HPV-16 DNA was present in 4/5 malignant and 2/5 benign tissues, with HPV-18 DNA being present in 1/5 malignant and 1/5 benign tissues. hTERT mRNA was detected in all malignant and no benign tissues, with the demonstrated pilot data to indicate the potential for using the LoC in cervical cancer screening in resource-limited settings on a large scale. | (2022) 12:8750 | https://doi.org/10.1038/s41598-022-12557-y www.nature.com/scientificreports/ varies by region 11, 12 . Furthermore, proliferative capacity in cancer cells is partly attained through telomere maintenance, and telomerase is universally expressed in HPV-positive cancers 13 . hTERT (human telomerase reverse transcriptase) is a telomerase component that is significantly overexpressed in cervical lesions, being detectable in at least 90% of cervical cancers 14, 15 , but presents low to nil expression in normal tissue. Therefore, it is a potentially useful tumour-specific marker. Loop-mediated isothermal amplification (LAMP) methods have been described for point-of-care detection of DNA and RNA from infectious agents, including SARS-CoV-2 16, 17 and breast cancer markers 18, 19 . LAMP was originally developed by Notomi et al. 20 , to allow rapid amplification of nucleic acids at a single temperature, typically between 63 and 65 • C. The reaction occurs without the need for thermocycling, the technique used in PCR that requires access to specialised equipment. Instead, LAMP relies on auto-cycling strand displacement DNA synthesis conducted by a DNA polymerase with high strand displacement activity, thus making the technique ideal for point-of-care testing. LAMP was originally designed with four primers 20 , however this was extended to six by Nagamine et al. 21 as this accelerated the reaction. The six primers are: Forward-Inner (FIP), Backward-Inner (BIP), Forward Outer (F3), Backward Outer (B3), Loop Forward (LF) and Loop Backward (LB). A stemloop structure is formed via the binding of the FIP to the target region and the hybridisation of the F3 primer to its target region, forming the stem-loop and allowing strand displacement DNA synthesis to begin. Similarly, the BIP and B3 primers bind on the other end of the target sequence, forming a 'dumbbell like' structure 20 . Subsequently, an exponential generation of inverted repeats is constructed as the inner primers anneal and cause amplification from the loops in the original structure. The loop primers, LF and LB, allow hybridisation to the available stem-loops which are not hybridised by the inner primers (FIP/BIP). This significantly accelerates the reaction from 1 h to 10-15 min depending upon the concentration of the starting material. LAMP exhibits high specificity due to the use of four to six primers recognizing six to eight distinct regions, with FIP and BIP playing crucial roles in the specificity of the assay. LAMP primers for HPV-16/18 DNA used alongside hTERT and GAPDH (used as a positive control) mRNA primers are transferable to the LoC device that reports results via Bluetooth to a smart phone. The device is based on Complementary Metal-Oxide-Semiconductor (CMOS) technology coupled with Ion-Sensitive Field-Effect Transistor (ISFET) sensors. This allows the LoC to detect changes in pH that occur during nucleic acid amplification, in which hydrogen ions are released during nucleotide incorporation. LAMP technology on a LoC was first described by Hu and Georgiou et al. 22 and has been applied to the detection of breast cancer biomarkers 18, 19, 23 and infectious diseases diagnostics including COVID-19 17 . The purpose of this study, therefore, was to translate a LAMP technique using optimally designed primers for HPV DNA and hTERT mRNA to a LoC device and test its ability to detect the presence of cervical cancer, in a pilot feasibility study. The LAMP assays for hTERT and GAPDH mRNA were tested with ten-fold serial dilutions of synthetic RNA ranging from 10 7 to 10 0 copies per reaction. For quantification, the time to positive (TTP) values were obtained using the cycle-threshold metric, Ct, set at 0.2 normalized fluorescence units by the LC96 instrument. Standard curves from these results are shown in Fig. 1 with high linearity (R 2 = 0.99) and a limit of detection of 100 and 1000 copies/reaction and maximum TTP of less than 20 and 10 min, respectively. Analytical sensitivity of human markers in pH-LAMP. Validation of HPV DNA and hTERT mRNA using the LOC was done in 10 of 45 available samples from a local tissue collection in which DNA and RNA quantity and quality indicated excellent extraction due to the high number of cells. Of these 10 cervical biopsy samples, there were 5 benign, 4 squamous cell carcinoma (SCC) and 1 adenocarcinoma (ADC) classified on pathology (by K.V. with > 5 years experience as a specialist gynaecological histopathologist). Table 1 . Two cases (H2 and H12) were negative for HPV-16 DNA on the LC96 device using pH-LAMP but were positive when measured on the LoC platform. The presence of HPV-16 DNA in these samples was confirmed using PCR primers. To explain this discrepancy, it may be the case that the copy number of HPV-16 DNA in both of these samples was at the limit of detection or using the same mix/primer regime on both devices yields a more sensitive performance using the LoC platform or that the larger sample volume on the LoC enabled the detection of lower copy numbers at the edge of the detection limit. The presence of hTERT mRNA was detected in 5/5 of cancer samples and 0/5 of benign samples, perfectly delineating the benign from the tumour samples, with results confirmed using the LoC platform. The principle of using a highly sensitive initial test, such as HPV DNA, as well as a secondary triage test with high specificity is appropriately demonstrated here and the near perfect agreement between systems indicates the potential of this platform for further testing. The time to positive (TTP) of less than 25 min on the LoC platform for all tests demonstrates the true point-of-care potential of this system to deliver rapid, accurate results on a portable platform. This pilot feasibility study has demonstrated the capability of a LoC device to accurately detect the presence of cervical cancer from cervical tissue biopsies using LAMP, specific to HPV DNA and hTERT mRNA as a triage marker, but also nucleic acid markers of both HPV infection and the host genomic consequences of persistent HPV effect resulting in malignancy. This allows both high sensitivity and specificity, bypassing the issue of tests that detect solely HPV DNA which can have high rates of false positives, if not performed in conjunction with other screening tests 24 . Furthermore, HPV DNA screening methods are dependent on expensive equipment which can be inaccessible at locations worldwide, particularly LMICs, most affected by cervical cancer. Other screening methods, such as HPV "self-test" kits, are based on detection of antibodies to HPV strains and primarily detect high-risk strains, which are dependent on the host immune reaction to the presence of the HPV and may be falsely negative, especially in the acute phase of the infection. Electrochemical systems with chip based biosensors have also been reported for detection of HPV DNA. One such system has been shown to use 16 gold working electrodes arranged in a four by four distribution on a borosilicate glass chip measuring 21 mm × 23 mm 25 . Each working electrode is placed between a silver pseudo reference and a gold counter electrode of the same size in order to create 16 planar electrochemical cells. The assay is based on co-immobilisation of HPV E7 and E6 thiolated probes with backfiller, a hybridisation process and electrochemical detection. The limits of detection are described in the pM range and are sufficient for most real RNA/DNA samples, but the system is bulky and not easily adaptable to population screening. In another system, target HPV DNA is captured via magnetic bead-modified DNA probes, followed by an antidigoxigenin-peroxidase detection system at screen-printed carbon electrode chips, with eight samples measurable simultaneously 26 . In comparison, the system described in this paper uses a LoC cartridge that performs on-chip real-time DNA amplification and is able to detect viral DNA by LAMP. This simplifies readout of DNA/RNA amplification and is also potentially adaptable to high throughput, as required for screening applications. DNA validation was performed on HPV and similarly RNA validation was performed on hTERT. hTERT was selected as previous literature has suggested its role as a promising tumour marker in cervical cancer due to the fact that it is not expressed in normal tissue 14, 15, 27, 28 . The analytical sensitivity of the selected markers was more than adequate for our purposes. The mRNA markers had a poorer sensitivity compared to the DNA markers, with a limit of detection of 10 2 for hTERT, and are comparable to other LAMP-based markers. Given that the average quantity of extracted total RNA in archived cervical samples is 27.5 ng/µL 29 , the limit of detection of 10 3 copies per reaction for mRNA is several orders of magnitude lower than the one found in normal clinical samples. Further, GAPDH is present in abundance thus a limit of detection of 10 3 per reaction is adequate in order for it to function as a positive control. Nevertheless, www.nature.com/scientificreports/ the limitations of pursuing goals of ultra-sensitivity with these assays comes at the price of loss of specificity and can lead to patients classified as positive, who in truth have no clinically relevant disease. Therefore, a balance between optimal sensitivity without increasing the number of false positive classifications, which would require further (unnecessary) investigations, needs to be made. Our study confirms that the assay was able to differentiate benign from tumor on a standard benchtop PCR machine with equivalent results on the LOC device for solid tumor tissue. We acknowledge however, that in a screening setting, the samples used would be cytological rather than solid tissue biopsy. Although the amount of tumor tissue available from a biopsy sample is greater than from a cytological swab sample, the latter method yields upwards of 50,000 cells, which is more than adequate for extracting detectable amounts of mRNA. Moreover, cell lysis and DNA/ mRNA extraction is much easier from a cytological swab as homogenization of the cellular material is not required. Therefore, the performance of the designed markers seems to sit squarely in the clinically relevant zone andshould be applicable when using brush cervical cytology samples. Validation of the HPV and hTERT LAMP reactions against conventional PCR showed largely equivalent results for both the DNA and RNA. Validation of the clinical testing of hTERT and HPV-16 against standard PCR was also equivalent. The LoC system was in close agreement with the traditional benchtop LC96 system, which indicates the potential of this platform as a diagnostic tool.The sensitivity and specificity of the LOC for detecting CIN needs to be explored in future work. Detection of HPV DNA using this assay would potentially identify all women with CIN and cancer, whilst the addition of hTERT mRNA could be employed as a cancerspecific marker. A large prospective study utilizing the assay on a chip at point of care in a screening population is essential to establish its role in this setting. Primer design. hTERT primers were designed to avoid amplification of the two main splice variants, alpha and beta, which are inhibitory to telomerase function. Primers for hTERT and GAPDH mRNA were designed and selected based on their analytical sensitivity, which was then further documented using a LAMP technique. Together with HPV-16/18 DNA primers designed by Luo et al. 30 , they were subsequently transferred to a specially designed LOC platform, where they were used to detect hTERT mRNA and HPV DNA extracted from tissue samples taken from women with a diagnosis of cervical cancer. The genomic sequences for hTERT (GenBank Accession number NG_009265.1, NM_198253.2, NM_001193376.1, NR_149162.1, NR_149163.1) were retrieved from the National Institute of Health GenBank system and visualised with Geneious software. Within Geneious the sequences were aligned with the MUSCLE method (MUltiple Sequence Comparison by Log-Expectation) 31 . The mRNA transcript variant sequences were also downloaded. LAMP primers were designed using Primer Explorer V5 (http:// prime rexpl orer. jp/ lampv 5e/ index. html). The splice variant locations for TERT 32 were added to the sequences prior to primer design. Within hTERT, there are multiple transcript variants but only the full length mRNA transcript conserves telomerase activity 33 . Therefore, exclusion of the two main variants, alpha and beta, that are inhibitory to telomerase function, was necessary. The alpha splice variant removes the first 36 bp of exon 6, while the beta splice variant removes 182 bp from exons 7 and 8 32 . As the F2 primer is located within the 36 bp alpha splice sequence and the B2 and B3 primers are located in the portion of exon 7 which is part of the beta splice, a primer set specific to full length mRNA, that was unable to amplify the main inhibitory splice variants and was exon-exon spanning was designed; this is depicted in Fig. 2. Primer selection. Primer selection based on analytical sensitivity used synthetic DNA samples for validation. Synthetic double-stranded DNA containing the sequence of the target DNA or RNA (Integrated DNA Technologies, Leuven, Belgium) was quantified using a Qubit 3.0 fluorometer and high-sensitivity doublestranded DNA assay kits (Life Technologies, Carlsbad, CA) on a LightCycler 96 System (Roche Molecular Systems, Pleasanton, CA). The hTERT primers were tested against a synthetic sequence specific to alpha and beta splice variants. The best primer sets were selected as follows: Step 1 Primer sets for RNA based markers were tested against human genomic DNA (Thermo Fisher Scientific) to ensure the primers were specific to RNA, and not DNA. Step 2 All primer sets were tested against a synthetic double-stranded DNA or RNA containing the sequence of interest, including appropriate non-template controls. In this initial primer screen the synthetic sequence was diluted to a 10 6 copy/reaction. Step 3 The two primer set candidates that achieved the fastest TTP were then tested in a dilution screen. Analytical sensitivity of the assays was evaluated in serial 10-fold dilutions of synthetic DNA or RNA target. PCR reaction conditions. PCR was used as the 'gold standard' method against which the efficacy of the newly developed LAMP assay was validated. The PCR primers used are displayed in Table 2 . The reaction conditions were based on the conditions laid out in Wang et al. 27 . Determining the analytical sensitivity of selected primers using a LAMP technique. For the RNA markers, synthetic double-stranded DNA containing a T7 promoter for RNA transcription synthesised by Integrated DNA Technologies (Leuven, Belgium) was used. The HiScribe T7 High-yield RNA synthesis kit (New England Biolabs, Hitchin, UK) was used to produce target specific RNA to test the designed primers. The synthetic DNA and RNA was quantified using a Qubit 3.0 fluorometer and high-sensitivity double-stranded DNA and single stranded RNA assay kits (Life Technologies, Carlsbad, CA). Sensitivity of the assay was evaluated in serial 10-fold dilutions of synthetic DNA/RNA target, ranging from 10 8 to 10 0 copies per reaction with each condition run in triplicate. pH-LAMP assays were performed on the LightCycler 96 System and data analysed using the system software version SW1.1. Transferring the LAMP technique to the LoC platform. The LAMP assay was translated to a portable LoC platform. This utilises a cartridge which integrates standard and low cost components including a state-of-the-art microchip, a Printed Circuit Board (PCB) and a microfluidic manifold (Fig. 3) . The microchip is implemented in a standard CMOS process which is ideal for scalability and low cost. It includes a large array of 78 × 56 ISFETs for large-scale potentiometric sensing to monitor nucleic acid amplification during LAMP through the release of protons during nucleotide incorporation. The increase in proton concentration is proportional to the total number of nucleotides inserted and inversely proportional to the buffering capacity of the solution. Therefore, DNA amplification modifies the overall pH of the solution causing a change in voltage detected by the ISFETs. When no amplification occurs, the pH of the solution remains unchanged therefore giving a constant voltage signal. The reaction mix of a standard LAMP therefore was modified to pH-LAMP 34 by adjusting the buffering capabilities to allow for changes in pH to be measured during DNA amplification. The CMOS microchip (Fig. 3A ) was attached to a printed circuit board (PCB) and a laser cut acrylic manifold was bolted on top to form a sealed reaction chamber for the reaction. The manifold was designed for a reaction volume of 20 µ L and sealing between the manifold and chip surface was accomplished through laser cut doublestick Tessa adhesive gaskets. An Ag/AgCl reference electrode (chloridised 0.03 mm Ag wire) was placed between the chip surface and the tape gasket. The cartridge is shown in Fig. 3B . Reactions were performed for 30 min, and data was recorded in situ in real-time. The LoC device (Lacewing) was designed as a handheld platform to perform heating and temperature regulation for the LAMP assay, data acquisition from the microchip, and Bluetooth transmission to a smartphone app developed on AndroidOS (Fig. 3C) . Temperature regulation was performed with a PID controller on the device microcontroller using on-chip temperature sensors and a Peltier Table 2 . LAMP and PCR primer sequences for hTERT and GAPDH mRNA, in 5 ′ -3 ′ direction. F3 GCC TGA GCT GTA CTT TGT CA GAT GCT GGC GCT GAG TAC B3 GGT GAG CCA CGA ACT GTC GCT AAG CAG TTG GTG GTG C FIP TGG GGT TTG ATG ATG CTG GCG AGG GCG CGT ACG ACA CCA TCC CTT TTG GCT CCC CCC TGC AAA TGG AGTCC ACT GGC GTCTT BIP GGT CCA GAA GGC CGC CCA TGC TGG AGG TCT GTC AAG GTA TCT GCT GAT GCC CCC ATG TTC GGA GGCAT TGC TGA TGA TCT LF ACC TCC GTG AGC CTG TCC TG AGC CTT CTC CAT GGT GGT www.nature.com/scientificreports/ module placed in contact with the bottom side of the cartridge. The device is powered using a battery integrated at the bottom of the 3D printed case. Spatial monitoring of the reaction pH ( Fig. 3D ) was averaged over all active sensors and processed in MatLab to interpolate and cancel sensor drift (Fig. 3E) 35, 36 . The output was linearised and then normalised in order to provide results proportional to the number of nucleotides in solution corresponding to the amplification curve (Fig. 3F) . The TTP was then extracted based on cross-thresholding at 0.2 normalized fluorescence units. Validation of LoC technology using cervical tissue samples. Tissue samples from cervical cancer patients, from which adequate DNA and mRNA was extracted, were tested using the LAMP technique on the LoC in order to test its ability to detect HPV DNA/RNA and hTERT mRNA as biomarkers for the presence of cervical cancer. Furthermore, to confirm the presence of high quality mRNA obtained from the extractions, GAPDH (glyceraldehyde-3-phosphate dehydrogenase, a common 'housekeeping' gene {hlwhich is used as a positive control) mRNA primers were designed and LAMP assays were performed, prior to using the mRNA for hTERT assays. An existing holding of 45 cervical tissue biopsies were used, collected in 2001 from patients treated at Hammersmith Hospital, London, with their informed consent, and stored at −80 • C. The use of these samples for our project was reviewed and approved by London-Surrey Research Ethics Committee Ref: REC18/LO/0865 and performed in accordance with the Declaration of Helsinki. Pathological verification of cancer was undertaken by splitting the sample in half and using half for DNA/RNA extraction and the other half for histological evaluation. For DNA/RNA extraction, the tissue was homogenized in 2 mL tubes containing RLT+/BME lysis buffer solution from the AllPrep DNA/RNA Kit (Qiagen, Hilden, Germany) and zirconium oxide beads in a Precellys 24 Homogenizer (Bertin, Villeurbanne, France). DNA and RNA were extracted using the AllPrep kit, according to the manufacturer's instructions. Total DNA and RNA yield was determined using a NanoDrop ND-2000 spectrophotometer (ThermoFisher Scientific, Waltham, USA). Only those samples which yielded both DNA and RNA were selected for testing. Testing was undertaken on the LOC platform and a comparison made with a pH-LAMP www.nature.com/scientificreports/ experiment on the LC96 system. For histological evaluation, paraffin embedded sections were mounted on glass slides and stained with haemotoxylin and eosin. Slides were reviewed by a specialist consultant gynaecological histopathologist, who classified the sample as benign or malignant tissue ("Supplementary information"). Received: 2 November 2021; Accepted: 22 April 2022 Global Cancer Statistics 2020: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries Human papillomavirus E7 induces p63 expression to modulate DNA damage response Chromosomal copy number alterations and HPV integration in cervical precancer and invasive cancer Understanding the HPV integration and its progression to cervical cancer The human papillomavirus E7 oncoprotein as a regulator of transcription Papillomavirus E6 oncoproteins Cytology versus HPV testing for cervical cancer screening in the general population Accuracy of combinations of visual inspection using acetic acid or lugol iodine to detect cervical precancer: A meta-analysis Performance of careHPV, hybrid capture 2 and visual inspection with acetic acid for detection of high-grade cervical lesion in Tanzania: A cross-sectional study Human papillomavirus genotype attribution in invasive cervical cancer: A retrospective cross-sectional worldwide study Prophylactic vaccination against human papillomaviruses to prevent cervical cancer and its precursors Activation of telomerase by HPVs Telomerase activity and human papillomavirus in malignant, premalignant and benign cervical lesions Telomerase activity exclusively in cervical carcinomas and a subset of cervical intraepithelial neoplasia grade III lesions: Strong association with elevated messenger RNA levels of its catalytic subunit and high-risk human papillomavirus DNA Rapid detection of mobilized colistin resistance using a nucleic acid based lab-on-a-chip diagnostic system Handheld point-of-care system for rapid detection of SARS-CoV-2 extracted RNA in under 20 min Detection of multiple breast cancer ESR1 mutations on an ISFET based lab-on-chip platform A novel hotspot specific isothermal amplification method for detection of the common PIK3CA p.H1047R breast cancer mutation Loop-mediated isothermal amplification of DNA Accelerated reaction by loop-mediated isothermal amplification using loop primers A robust ISFET pH-measuring front-end for chemical reaction monitoring Detection of breast cancer ESR1 p. E380Q mutation on an ISFET lab-on-chip platform The feasibility of new HPV/DNA test as a primary cervical cancer screening method among 35-years-old ever-married women in Kalutara district; a cross-sectional study Electrochemical biosensor for the multiplexed detection of human papillomavirus genes Electrochemical chip-based genomagnetic assay for detection of high-risk human papillomavirus DNA Use of hTERT and HPV E6/E7 mRNA RT-qPCR TaqMan assays in combination for diagnosing high-grade cervical lesions and malignant tumors Diagnostic performance of HPV E6/E7, hTERT, and Ki67 mRNA RT-qPCR assays on formalin-fixed paraffinembedded cervical tissue specimens from women with cervical cancer Analytic and diagnostic performances of human papillomavirus E6/E7 mRNA test on up-to 11-year-old liquid-based cervical samples A biobank-based longitudinal study Visual detection of high-risk human papillomavirus genotypes 16, 18, 45, 52, and 58 by loop-mediated isothermal amplification with hydroxynaphthol blue dye Multiple sequence alignment with high accuracy and high throughput Telomerase activity in human development is regulated by human telomerase reverse transcriptase (hTERT) transcription and by alternate splicing of hTERT transcripts Alternative splicing regulation of telomerase: A new paradigm Quantitative and rapid Plasmodium falciparum malaria diagnosis and artemisinin-resistance detection using a CMOS Lab-on-Chip platform Isfets in cmos and emergent trends in instrumentation: A review Rapid detection of azole-resistant aspergillus fumigatus in clinical and environmental isolates by use of a lab-on-achip diagnostic system The authors declare no competing interests. The online version contains supplementary material available at https:// doi. org/ 10. 1038/ s41598-022-12557-y.Correspondence and requests for materials should be addressed to P.G.Reprints and permissions information is available at www.nature.com/reprints. Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http:// creat iveco mmons. org/ licen ses/ by/4. 0/.