key: cord-0819726-6y9tn2sy authors: Fernández Fernández, Elena; Joyce, Mary; O’Sullivan, Andrew; MacLoughlin, Ronan title: Evaluation of Aerosol Therapy during the Escalation of Care in a Model of Adult Cystic Fibrosis date: 2021-04-21 journal: Antibiotics (Basel) DOI: 10.3390/antibiotics10050472 sha: 7f9c1c44766f89a9d78f281d7d881734ae9977d5 doc_id: 819726 cord_uid: 6y9tn2sy Lung disease is the main cause of morbidity and mortality in cystic fibrosis (CF). CF patients inhale antibiotics regularly as treatment against persistent bacterial infections. The goal of this study was to investigate the effect of clinical intervention on aerosol therapy during the escalation of care using a bench model of adult CF. Droplet size analysis of selected antibiotics was completed in tandem with the delivered aerosol dose (% of total dose) assessments in simulations of various interventions providing oxygen supplementation or ventilatory support. Results highlight the variability of aerosolised dose delivery. In the homecare setting, the vibrating mesh nebuliser (VMN) delivered significantly more than the jet nebuliser (JN) (16.15 ± 0.86% versus 6.51 ± 2.15%). In the hospital setting, using VMN only, significant variability was seen across clinical interventions. In the emergency department, VMN plus mouthpiece (no supplemental oxygen) was seen to deliver (29.02 ± 1.41%) versus low flow nasal therapy (10 L per minute (LPM) oxygen) (1.81 ± 0.47%) and high flow nasal therapy (50 LPM oxygen) (3.36 ± 0.34%). In the ward/intensive care unit, non-invasive ventilation recorded 19.02 ± 0.28%, versus 22.64 ± 1.88% of the dose delivered during invasive mechanical ventilation. These results will have application in the design of intervention-appropriate aerosol therapy strategies and will be of use to researchers developing new therapeutics for application in cystic fibrosis and beyond. Cystic fibrosis (CF) is one of the most commonly diagnosed genetic diseases worldwide. It is caused by mutations in the cystic fibrosis transmembrane conductance regulator (CFTR) gene [1] . This encodes a cyclic AMP dependent chloride channel expressed in the epithelia of many exocrine tissues including the airways, lung, pancreas, liver, intestine, vas deferens and sweat gland/duct [2] . Mutations in CFTR cause abnormal ion transport in the epithelium of several tissues, which results in the production of abnormally thick and sticky mucus that blocks the organ and is responsible for CF pathology [3] . However, the main cause of morbidity and mortality in CF is lung disease with lung malfunction and pulmonary failure [4] . The production of sticky mucus in the lumen of the lungs impedes mucociliary clearance [5] . Clinically, this manifests as chronic inflammation and recurrent bacterial infections by pathogens such as Pseudomonas aeruginosa and Staphylococus aureus [6] . Mucus and inflammatory cells cause bronchiectasis, and the continuous cycle of infection and inflammation lead to the progressive destruction of the lung tissue [7, 8] . Survival of patients with CF has continued to improve to a median age of 46.2 years. This has resulted in a rapid increase in the adult CF population and The result of droplet size characterisation for a selection of commonly nebulised antibiotic formulations using the Aerogen Solo vibrating mesh nebuliser are presented in Table 1 below. All formulations characterised were within the respirable range with relatively high fine respirable fractions. Of the two main types of nebulisers, compressed-air-driven jet nebulisers (JN) are the most common in the home. Vibrating mesh nebuliser-type (VMN) is also used in the home, but to a lesser degree. Whilst JN is available in the hospital setting, the VMN is the predominant type. VMN is the only one considered suitable for concurrent use across the various patient interventions encountered throughout the escalation of care from home to intensive care [15, 16] . Additionally, current global COVID−19 guidance on the selection of nebuliser recommends the use of devices that maintain a closed system for the prevention of release of both medical and patient-derived bioaerosols into the local environment. VMN is the only nebuliser type that allow this [17] [18] [19] [20] [21] . The aim of this study was to investigate the effect of the escalation of care related to the aerosol therapy delivered to the patient under a simulated adult CF breathing profile, and to provide insight into the variability of nebulised dose delivery across clinical interventions either in the homecare or hospital setting. The result of droplet size characterisation for a selection of commonly nebulised antibiotic formulations using the Aerogen Solo vibrating mesh nebuliser are presented in Table 1 below. All formulations characterised were within the respirable range with relatively high fine respirable fractions. Following tests, the JN was seen to deliver 6.51 ± 2.15% of the nominal dose to the end of the trachea, i.e., the expected lung dose. However, the portable VMN was seen to deliver significantly more, with 16.15 ± 0.86% of the dose delivered to the end of the trachea (p value < 0.001). See Figure 2 . The consequence of device selection is apparent here with choice of nebuliser having a significant bearing on the amount of drug delivered to the patient. 1.41% of the nominal dose delivered to the level of the trachea (p value < 0.001 versus low flow and high flow systems, respectively). When the same VMN device was included in a low flow nasal therapy system at 10 L per minute, 1.81 ± 0.47% of the dose was delivered, compared with 3.36 ± 0.34% delivered during high flow nasal therapy at 50 L per minute oxygen flow rate. See Figure 3 . In this instance, whilst the same nebuliser type was used, it is obvious that clinical intervention can have a significant bearing on the amount of drug delivered to the patient. During both non-invasive and invasive mechanical ventilation, typical of the intensive care and ward settings, similar levels of drug were seen delivered under ventilation strategies appropriate for the adult cystic fibrosis patient. NIV in combination with VMN recorded 19.02 ± 0.28%, whereas IMV recorded 22.64 ± 1.88% of the dose delivered (p value = 0.012). See Figure 3 . In simulated aerosol drug delivery interventions and oxygen therapy interventions that facilitate concurrent aerosol drug delivery commonly prescribed for patients presenting to the hospital emergency department, mouthpiece-mediated VMN aerosol delivery with no supplemental gas flow rate was seen to deliver the highest dose, with 29.02 ± 1.41% of the nominal dose delivered to the level of the trachea (p value < 0.001 versus low flow and high flow systems, respectively). When the same VMN device was included in a low flow nasal therapy system at 10 L per minute, 1.81 ± 0.47% of the dose was delivered, compared with 3.36 ± 0.34% delivered during high flow nasal therapy at 50 L per minute oxygen flow rate. See Figure 3 . In this instance, whilst the same nebuliser type was used, it is obvious that clinical intervention can have a significant bearing on the amount of drug delivered to the patient. Aerosol therapy is a mainstay in the routine treatment of CF wherein many therapeutics, including antibiotics, are delivered via nebuliser. Here, we have assessed aerosol drug delivery across a range of CF-appropriate interventions in both the homecare and hospital settings. Using a tracer drug, salbutamol, in combination with adult CF breath patterns, we describe the variability in dose delivery that could be expected to be delivered to the CF patient. This study, for the first time, investigates the effect of device and intervention on the amount of drug delivered to the patient throughout the escalation of care of a simulated exacerbated adult CF patient. Droplet size is known to affect aerosol drug delivery in the patient, and here we characterised the respirable fraction and droplet size of a selection of commonly nebulised antibiotics [22] . Using a single nebuliser type as the control, indicative droplet size was seen to be relatively consistent with all formulations under test, recording a consistent respirable fraction and droplet size. This would indicate that whilst some differences are noted between formulations, they all remain within the respirable range. Factors affecting droplet size include the physicochemical characteristics, drug concentration, ionic content and temperature of the formulation being aerosolised [23] [24] [25] [26] . Further, significant varia- During both non-invasive and invasive mechanical ventilation, typical of the intensive care and ward settings, similar levels of drug were seen delivered under ventilation strategies appropriate for the adult cystic fibrosis patient. NIV in combination with VMN recorded 19.02 ± 0.28%, whereas IMV recorded 22.64 ± 1.88% of the dose delivered (p value = 0.012). See Figure 3 . Aerosol therapy is a mainstay in the routine treatment of CF wherein many therapeutics, including antibiotics, are delivered via nebuliser. Here, we have assessed aerosol drug delivery across a range of CF-appropriate interventions in both the homecare and hospital settings. Using a tracer drug, salbutamol, in combination with adult CF breath patterns, we describe the variability in dose delivery that could be expected to be delivered to the CF patient. This study, for the first time, investigates the effect of device and intervention on the amount of drug delivered to the patient throughout the escalation of care of a simulated exacerbated adult CF patient. Droplet size is known to affect aerosol drug delivery in the patient, and here we characterised the respirable fraction and droplet size of a selection of commonly nebulised antibiotics [22] . Using a single nebuliser type as the control, indicative droplet size was seen to be relatively consistent with all formulations under test, recording a consistent respirable fraction and droplet size. This would indicate that whilst some differences are noted between formulations, they all remain within the respirable range. Factors affecting droplet size include the physicochemical characteristics, drug concentration, ionic content and temperature of the formulation being aerosolised [23] [24] [25] [26] . Further, significant variation in aerosol characteristics can be expected between devices [27, 28] . As such, there will never be a single droplet size distribution that can be relied on, and so these results should be considered as only a feasibility for generation of inhalable antibiotic. In the home, the predominant nebuliser type used is the JN [29] . In more recent times, however, the vibrating mesh type devices have seen increasing adoption [30] [31] [32] . These VMN devices claim higher drug delivery rates, with less drug wastage. Here, the results demonstrate that, in this case at least, the portable VMN delivered approximately 2.5 times more to the simulated patient than the JN. The drug/mass amount delivered, whilst not described in the literature previously using the breathing pattern used herein, are in line with results published in imaging studies assessing performance of the JN in a spontaneously breathing adult patient (for example, 6.51% versus 5.2%) [31] . Whilst only one JN type was assessed here, variability in JN performance can be significant, and so these results should be considered representative of the nebuliser type [32] . The dramatic difference in tracheal dose delivered is due to a combination of factors including the large residual volume recorded in JN of all types [33, 34] , as well as the fact that the drug mass tends to concentrate in the residual volume remaining, as buffer is preferentially aerosolised [23, 35] . Should the patient require medical attention and present at the emergency department (ED) in the hospital, a variety of patient interventions may be prescribed in an effort to have the patient return to stable-state lung function. On confirmation of infection, or infection-mediated bronchospasm, inhaled antibiotics, mucolytics or bronchodilators may be prescribed [36, 37] . Whereas JN are found and used in the ED, VMN is also, with a mounting body of published clinical literature supporting its adoption [38] [39] [40] . As mentioned, VMN is also considered the only nebuliser type suitable for concurrent use with a variety of oxygenation supplementation or ventilatory support strategies. In fact, in some cases, JN is contra-indicated against, as the compressed air required for its normal function interferes with the finely tittered oxygen flow prescribed for the patient [41] . In the ED setting, and depending on how severe the state of the patient, and their supplemental oxygen requirement, e.g., if they become hypoxic, they may be prescribed further aerosol therapy. This may be concurrent with low flow nasal therapy (LFNT) or high flow nasal therapy (HFNT). Across these frontline interventions, our studies indicate that aerosol delivery performance can vary significantly. As might be expected, an interface such as the mouthpiece-mediated delivery using the VMN and chamber was seen to deliver the greatest amount of tracer aerosol during simulated breathing, with LFNT and HFNT delivering significantly less. The results recorded here for VMN-mediated aerosol delivery are in line with previous reports for the VMN in combination with the chamber (for Antibiotics 2021, 10, 472 6 of 13 example, 29.02% versus 34.1%) [31, 42] , and HFNT (for example, 3.36% versus 3.46%) [42, 43] . No experimentally comparable published data for LFNT could be found. The larger dose delivered by the combination of VMN and chamber, via mouthpiece, may be explained through a combination of bolus aerosol inhalation, whereby aerosol builds up in the chamber during the exhalation phase of the breath, and the relatively large bore oropharynx, which minimises losses before the lung. The HFNT result is explained through significant aerosol losses throughout the tubing between the nebuliser and the patient but, additionally, the impaction losses seen in the nasal passages after the aerosol exits the nasal cannula [44] . One would expect that, at the lower oxygen flow rates used during LFNT, the ballistic fraction would be smaller and, thus, impaction losses would be minimised; however, a significant fraction of the aerosol is lost in the nasal passages [45] . Further, a major contributor to the reduced aerosol dose here is the very narrow bore oxygen cannula that are typically used [44] . In this instance, the inner diameter of the cannula was 1.8 mm. This is low, compared with the typical internal diameter of adult HFNT cannula, measuring up to as wide as 10 mm [44] . The ratio of nasal cannula outer diameter to nostril inner diameter determines the leak of oxygen, and consequently aerosol from the system also. This ratio also plays a part in the delivered aerosol dose, as well as the levels of fugitive emissions to the local environment, which is a key, but often forgotten consideration for healthcare professionals or bystanders during aerosol therapy [46] [47] [48] [49] . A recent review of aerosol delivery during HFNT by Li and colleagues provides additional insight into the factors at play [50] . Clinically, should these oxygenation supplementation strategies not be successful, and the patient continues to deteriorate, non-invasive ventilation (NIV), or as a last resort, invasive mechanical ventilation (IMV), may be prescribed. These ventilatory support strategies provide pre-set minimum pressures and/or volumes of air and aid the patient in returning to normal pulmonary arterial oxygen levels. NIV is often deployed as pressure support in a conscious spontaneously breathing patient, with IMV having a selection of ventilatory modes available for ventilation of the anaesthetised, unconscious patient via an endotracheal tube. In this study, there were only small differences between NIV and IMV with respect to the delivered aerosol dose (19.02% versus 22.64% respectively). Again, these figures are in line with previous published reports [51, 52] . Whilst little difference was noted between NIV and IMV, these interventions deliver significantly more aerosol compared to LFNT and HFNT. The likely reasons for the poor delivery during those interventions has already been explained; however, NIV benefits from large bore tubing (22 mm inner diameter) and a tight-fitting facemask (no leak). Additionally, with the addition of a fixed minimum positive end expiratory pressure (PEEP), aerosol distribution and deposition is likely to be increased over the lower PEEP levels delivered by LFNT and HFNT [53] . IMV also benefits from larger bore tubing, and depending on the ventilator mode, the bias flow-mediated build-up of an aerosol bolus between inhalations [54] . Furthermore, and to ensure safe volumes and pressures are delivered to the patient's lung on each breath, air leak is minimal, with cuffed endotracheal tubes creating a complete seal within the trachea. This again not only facilitates increased aerosol delivery, but also reduces fugitive medical aerosol emissions to the local environment, and, cognizant of COVID−19, the risk of patient derived transmission of infectious disease [55, 56] . The implications of these findings may be far-reaching. For example, patients need to be aware of the potential differences between nebulisers when purchasing from a community pharmacy; however, in some instances this choice may be limited if a particular nebuliser is indicated for use on the drug label. Further, clinicians now have insight into potential variability in aerosol delivery across common clinical interventions, and may, where appropriate, titrate the dose accordingly. Finally, developers of novel therapeutics for the treatment of CF are now provided insight into the significant variance in delivery between devices and interventions, and should consider the intended patient and treatment carefully in the selection of devices and interventions for delivery of their therapeutic. Experiments were conducted using a selection of appropriate and commonly used nebulisers in the treatment of cystic fibrosis, and the escalating treatment of lung infection. In the homecare setting, both a compressed air-driven jet nebuliser (Up-draft II jet nebuliser, Hudson, Teleflex Medical, Wake County, NC, USA) as well as a portable vibrating mesh nebuliser (InnoSpire Go, Philips, Farnborough, UK) were evaluated. In the hospital setting, a second VMN was used (Aerogen Solo, Aerogen Ltd., Galway, Ireland). The respirable fraction and aerosol droplet size for a selection of antibiotic formulations were characterised using laser diffraction as previously described [57] . Briefly, the devices were loaded with sample and connected to the inlet of the droplet sizer (Spraytec, Malvern Instruments, Worcestershire, UK). The nebuliser was turned on and run until the entire dose was delivered. Testing was carried out in triplicate. The aerosol characteristics of the nebulisers, with respect to volumetric median diameter (VMD) and fine particle fraction (FPF), were measured using laser diffraction (Spraytec, Malvern Instruments, Worcestershire, UK). For all test combinations, 2.5 mL of 2.5 mg/2.5 mL of Albuterol Sulphate (Ventolin, GlaxoSmithKline, Brentford, UK) was used as a tracer aerosol drug to determine drug dose delivered. This is in line with previous studies [44, 54, 58, 59] and international nebuliser test standard ISO27427. A 3D printed adult head model [60, 61] was attached to a breathing simulator (BRS2100, Copley Scientific Ltd., Nottingham, UK) via a capture filter, at the level of the trachea bifurcation (Respirgard 303, Vyaire, Basingstoke, UK) and breathing pattern applied. See Table 2 for summary of simulated aerosol delivery test scenarios with associated nebuliser types, interfaces and breath patterns utilised for each test. At the end of each dose, the drug captured on the filter was eluted using 10 mL of deionised water. The mass of drug was quantified by means of UV spectrophotometry at a wavelength of 276 nm and interpolation on a standard curve of albuterol sulphate concentrations (100-3.125 µg/mL). Results for tracheal dose were expressed as the percentage of the nominal dose placed in the nebuliser's medication cup. All testing was carried out in independent quintuplicate. To replicate aerosol therapy in a spontaneously breathing adult CF patient undergoing standard care in the homecare setting, both the JN plus open facemask (Up-draft II jet nebuliser, Hudson, Teleflex Medical, Wake County, NC, USA) and portable VMN in combination with a mouthpiece (InnoSpire Go, Philips, Farnborough, UK) were used. The JN was operated at 7 L per minute compressed air flow rate, in line with the accompanying instruction manual. To replicate aerosol therapy in a spontaneously breathing adult CF patient post-exacerbation in the hospital setting a VMN in combination with an aerosol chamber and filtered mouthpiece (Aerogen Ultra, Aerogen Limited, Galway, Ireland) (Respirgard 303 filter, Baxter, Dublin, Ireland) with no supplemental oxygen was used to deliver drug to the simulated adult CF patient. See Figure 4A ,B. Low flow oxygen therapy (LFNT): A nasal oxygen cannula (Hudson RCI comfort flo ® nasal cannula, Teleflex Medical, Delaware County PA, USA) placed in the nostrils of the head model was used to simultaneously deliver low flow oxygen therapy at 10 LPM while aerosol from the VMN (Aerogen Solo, Aerogen Limited, Galway, Ireland) was delivered in combination with an aerosol chamber and filtered mouthpiece (Aerogen Ultra, Aerogen Limited., Galway, Ireland) (Respirgard 303 filter, Vyaire, Basingstoke, UK), with no supplemental oxygen. See Figure 5 . Low flow oxygen therapy (LFNT): A nasal oxygen cannula (Hudson RCI comfort flo ® nasal cannula, Teleflex Medical, Delaware County PA, USA) placed in the nostrils of the head model was used to simultaneously deliver low flow oxygen therapy at 10 LPM while aerosol from the VMN (Aerogen Solo, Aerogen Limited, Galway, Ireland) was delivered in combination with an aerosol chamber and filtered mouthpiece (Aerogen Ultra, Aerogen Limited, Galway, Ireland) (Respirgard 303 filter, Vyaire, Basingstoke, UK), with no supplemental oxygen. See Figure 5 . High flow oxygen therapy (HFNT): A nasal cannula (OPT944, Optiflow + Medium, Fisher & Paykel, New Zealand) placed on the head model was used to deliver oxygen at 50 LPM with the turbine driven HFNT device (AIRVO2, Fisher & Paykel, Auckland New Zealand). Aerosol from the VMN (Aerogen Solo, Aerogen Limited, Ireland) was delivered in combination with an aerosol chamber and filtered mouthpiece (Aerogen Ultra, Aerogen Limited, Galway, Ireland) (Respirgard 303 filter, Vyaire, Basingstoke UK), with no supplemental oxygen. See Figure 6 . Low flow oxygen therapy (LFNT): A nasal oxygen cannula (Hudson RCI comfort flo ® nasal cannula, Teleflex Medical, Delaware County PA, USA) placed in the nostrils of the head model was used to simultaneously deliver low flow oxygen therapy at 10 LPM while aerosol from the VMN (Aerogen Solo, Aerogen Limited, Galway, Ireland) was delivered in combination with an aerosol chamber and filtered mouthpiece (Aerogen Ultra, Aerogen Limited., Galway, Ireland) (Respirgard 303 filter, Vyaire, Basingstoke, UK), with no supplemental oxygen. See Figure 5 . A critical care mechanical ventilator with non-invasive functionality (Bellavista, IMT Medical, Buchs Switzerland), was set to deliver continuous positive airway pressure (CPAP) of 10 cmH20. It was attached to the breathing simulator via a single limb, vented circuit with a non-vented NIV mask (RT045, Nivairo, Fisher and Paykel, Auckland New Zealand). The breathing simulator was set to deliver the same breathing pattern as the spontaneously breathing patient (see Table 2 ). In line with published in vivo imaging studies of the optimal nebuliser placement, the VMN was placed between the exhalation port of the single limb circuit and the mask [65] . The mask was attached to the breathing simulator via a capture filter. See Figure 7 . A critical care mechanical ventilator with non-invasive functionality (Bellavista, IMT Medical, Buchs Switzerland), was set to deliver continuous positive airway pressure (CPAP) of 10 cmH20. It was attached to the breathing simulator via a single limb, vented circuit with a non-vented NIV mask (RT045, Nivairo, Fisher and Paykel, Auckland New Zealand). The breathing simulator was set to deliver the same breathing pattern as the spontaneously breathing patient (see Table 2 ). In line with published in vivo imaging studies of the optimal nebuliser placement, the VMN was placed between the exhalation port of the single limb circuit and the mask [65] . The mask was attached to the breathing simulator via a capture filter. See Figure 7 . (CPAP) of 10 cmH20. It was attached to the breathing simulator via a single limb, vented circuit with a non-vented NIV mask (RT045, Nivairo, Fisher and Paykel, Auckland New Zealand). The breathing simulator was set to deliver the same breathing pattern as the spontaneously breathing patient (see Table 2 ). In line with published in vivo imaging studies of the optimal nebuliser placement, the VMN was placed between the exhalation port of the single limb circuit and the mask [65] . The mask was attached to the breathing simulator via a capture filter. See Figure 7 . Data are expressed as the arithmetic mean ± standard deviation (SD) tracheal dose (percentage). Statistical analysis was carried out using Minitab Software, version 18 (Coventry, UK). Student's t-tests were conducted to establish if the tracheal dose varied significantly across two different aerosol generators in the homecare setting. A One-Way ANOVA was performed to assess different drug delivery interfaces, across various patient interventions using a VMN only, in the hospital setting. Differences were considered statistically significant when p ≤ 0.05 (*). Aerosol droplet size characterisation work was carried out in triplicate. Tracheal dose experiments were conducted in quintuplicate and with at least five technical replicates per experiment. The experimental approach taken here to describe the gross or total aerosol inhaled and delivered to the level of the trachea does not consider the disease specific features of Figure 8 . IMV setup, the VMN was placed on the dry side of a heated humidifier and the capture filter placed between the ETT and a test lung. Data are expressed as the arithmetic mean ± standard deviation (SD) tracheal dose (percentage). Statistical analysis was carried out using Minitab Software, version 18 (Coventry, UK). Student's t-tests were conducted to establish if the tracheal dose varied significantly across two different aerosol generators in the homecare setting. A One-Way ANOVA was performed to assess different drug delivery interfaces, across various patient interventions using a VMN only, in the hospital setting. Differences were considered statistically significant when p ≤ 0.05 (*). Aerosol droplet size characterisation work was carried out in triplicate. Tracheal dose experiments were conducted in quintuplicate and with at least five technical replicates per experiment. The experimental approach taken here to describe the gross or total aerosol inhaled and delivered to the level of the trachea does not consider the disease specific features of CF such as mucus secretions, inflammation, infection etcetera and the influence they may have on distribution and deposition throughout the airways. This study highlights the potential for sometimes significant variance in aerosol drug delivery between both nebuliser type, and clinical intervention. These results will have application in the design of intervention-appropriate aerosol therapy strategies and will be of use to researchers developing new therapeutics for application in cystic fibrosis and beyond, as well as clinicians in the selection of nebuliser/intervention combinations in the treatment of patients. Identification of the cystic fibrosis gene: Genetic analysis Molecular mechanisms of CFTR chloride channel dysfunction in cystic fibrosis CFTR dysfunction in cystic fibrosis and chronic obstructive pulmonary disease Pathophysiology and Management of Pulmonary Infections in Cystic Fibrosis Cystic fibrosis: A disease of vulnerability to airway surface dehydration Cystic Fibrosis Since 1938 The neutrophil in chronic obstructive pulmonary disease Cystic fibrosis Treatment of pulmonary exacerbations in cystic fibrosis Aerosol therapy in cystic fibrosis: A survey of 54 CF centers Cystic Fibrosis Pulmonary Guidelines Pulmonary drug delivery. Part I: Physiological factors affecting therapeutic effectiveness of aerosolized medications Acute Pulmonary Exacerbations in Cystic Fibrosis The Effects of Heliox on the Output and Particle-Size Distribution of Salbutamol Using Jet and Vibrating Mesh Nebulizers Vibrating Mesh Nebulisers-Can Greater Drug Delivery to the Airways and Lungs Improve Respiratory Outcomes? Antibiotics 2021 Global Initiative for the Diagnosis, Management, and Prevention of Chronic Obstructive Lung Disease. The 2020 GOLD Science Committee Report on COVID-19 and Chronic Obstructive Pulmonary Disease Expert consensus on protective measures related to respiratory therapy in patients with severe coronavirus infection Association for Respiratory Care (AARC) 2020 Airway Management and Related Procedures in Critically Ill COVID-19 Patients: Position Statement of the Indian Society of Critical Care Medicine Clinical Consensus Recommendations Regarding Non-Invasive Respiratory Support in the Adult Patient with Acute Respiratory Failure Secondary to SARS-CoV-2 infection Regional Lung Deposition and Bronchodilator Response as a Function of β2-Agonist Particle Size Factors affecting aerosol performance during nebulization with jet and ultrasonic nebulizers The influence of fluid physicochemical properties on vibrating-mesh nebulization Viscosity effects on nebulisation of aqueous solutions A study of the effects of sodium halides on the performance of air-jet and vibrating-mesh nebulizers Mesh nebulizers have become the first choice for new nebulized pharmaceutical drug developments Nebulized drug delivery in respiratory medicine: What does the future hold? Home nebulizer use among patients with cystic fibrosis Preference and quality of life of adult chronic obstructive lung disease (COPD) patients when using a novel mesh nebulizer compared to traditional jet nebulizer (TJN). Airway Pharmacol SPECT-CT Comparison of Lung Deposition using a System combining a Vibrating-mesh Nebulizer with a Valved Holding Chamber and a Conventional Jet Nebulizer: A Randomized Cross-over Study Aerosolization Performance of Jet Nebulizers and Biopharmaceutical Aspects Investigation of the Quantity of Exhaled Aerosols Released into the Environment during Nebulisation Assessment of Antibiotic Aerosol Generation Using Commercial Jet Nebulizers Evaluation of jet nebulisers for use with gentamicin solution The future of cystic fibrosis care: A global perspective European Cystic Fibrosis Society Standards of Care: Best Practice guidelines Comparison of bronchodilator administration with vibrating mesh nebulizer and standard jet nebulizer in the emergency department Comparison of Vibrating Mesh Nebulizer Versus Jet Nebulizer in the Pediatric Asthma Patient: A Randomized Controlled Trial Quality Improvement Project to Compare Vibrating Mesh Nebulizer Therapy with Hour Long Jet Nebulizer Therapy for Albuterol Delivery in Asthma and Reactive Airway Disorder Patients in a Pediatric Emergency Department Influence of Jet Nebulization and Oxygen Delivery on the Fraction of Inspired Oxygen: An Experimental Model Vibrating Mesh Nebulisation of Pro-Antimicrobial Peptides for Use in Cystic Fibrosis Impact of Gas Flow and Humidity on Trans-Nasal Aerosol Deposition via Nasal Cannula in Adults: A Randomized Cross-Over Study In Vitro Determination of the Main Effects in the Design of High-Flow Nasal Therapy Systems with Respect to Aerosol Performance Nasal high flow nebulization in infants and toddlers: An in vitro and in vivo scintigraphic study Investigation of Fugitive Aerosols Released into the Environment during High-Flow Therapy Occupational risk factors for asthma among nurses and related healthcare professionals in an international study Exposure of health care workers to pentamidine isethionate Precautionary Practices of Respiratory Therapists and Other Health-Care Practitioners Who Administer Aerosolized Medications A narrative review on trans-nasal pulmonary aerosol delivery A mesh nebulizer is more effective than jet nebulizer to nebulize bronchodilators during non-invasive ventilation of subjects with COPD: A randomized controlled trial with radiolabeled aerosols Aerosol delivery with two ventilation modes during mechanical ventilation: A randomized study Optimization of aerosol deposition pressure support in children with cystic fibrosis: An experimental and clinical study Influence of nebulizer type, position, and bias flow on aerosol drug delivery in simulated pediatric and adult lung models during mechanical ventilation Fugitive Aerosol Therapy Emissions during Mechanical Ventilation: In Vitro Assessment of the Effect of Tidal Volume and Use of Protective Filters Nebuliser Type Influences Both Patient-Derived Bioaerosol Emissions and Ventilation Parameters during Mechanical Ventilation Early-Stage Development of Novel Cyclodextrin-siRNA Nanocomplexes Allows for Successful Postnebulization Transfection of Bronchial Epithelial Cells Effective nebulization of interferon-γ using a novel vibrating mesh Quantifying continuous nebulization via high flow nasal cannula and large volume nebulizer in a pediatric model Absorption and Clearance of Pharmaceutical Aerosols in the Human Nose: Development of a CFD Model Intermittent Aerosol Delivery to the Lungs During High-Flow Nasal Cannula Therapy Changes in Pulmonary Mechanics with Increasing Disease Severity in Children and Young Adults with Cystic Fibrosis Cartilaginous airway wall dimensions and airway resistance in cystic fibrosis lungs Ventilator-associated pneumonia caused by multidrug-resistant Gram-negative bacteria: Understanding nebulization of aminoglycosides and colistin Radioaerosol Pulmonary Deposition Using Mesh and Jet Nebulizers During Noninvasive Ventilation in Healthy Subjects This research was funded by Aerogen Limited. All authors are employees of Aerogen Limited.