key: cord-0817978-mdg4zuez authors: Kim, Hyeong Rae; An, Sanggwon; Hwang, Jungho title: High Air Flow-rate Electrostatic Sampler for the Rapid Monitoring of Airborne Coronavirus and Influenza Viruses date: 2021-01-23 journal: J Hazard Mater DOI: 10.1016/j.jhazmat.2021.125219 sha: d6ac5b581186c6aa6d68aa21b11e2540fb4ee17b doc_id: 817978 cord_uid: mdg4zuez Capturing virus aerosols in a small volume of liquid is essential when monitoring airborne viruses. As such, aerosol-to-hydrosol enrichment is required to produce a detectable viral sample for real-time quantitative reverse transcription polymerase chain reaction (qRT-PCR) assays. To meet this requirement, the efficient and non-destructive collection of airborne virus particles is needed, while the incoming air flow rate should be sufficiently high to quickly collect a large number of virus particles. To achieve this, we introduced a high air flow-rate electrostatic sampler (HAFES) that collected virus aerosols (human coronavirus 229E, influenza A virus subtypes H1N1 and H3N2, and bacteriophage MS2) in a continuously flowing liquid. Viral collection efficiency was evaluated using aerosol particle counts, while viral recovery rates were assessed using real-time qRT-PCR and plaque assays. An air sampling period of 20 min was sufficient to produce a sample suitable for use in real-time qRT-PCR in a viral epidemic scenario. The monitoring of biological particles suspended in the air has been used for decades to prevent large-scale outbreaks of infection. Real-time quantitative reverse transcription polymerase chain reaction (qRT-PCR), an assay designed for microbes that works by amplifying a target gene, is the most widely used method for monitoring airborne viruses due to its reliability. Before the assay can be conducted, however, virus aerosols need to be captured in a liquid because liquid samples are required for most bioanalytical approaches (e.g., nucleic acid detection, immunoassays, and cell culturing). For the rapid monitoring of airborne viruses, it is essential to develop an air sampler that can collect airborne viruses dispersed in the air at very low concentrations and transfer them into a small volume of liquid above the limit of detection (LOD) for real-time qRT-PCR. The enrichment capacity (EC) of aerosol-to-hydrosol (ATH) sampling is an indicator of how many airborne virus particles can be amassed in this small volume of liquid. Thus, a high EC can reduce the time required to produce a detectable virus sample for use in real-time qRT-PCR. In Eq. (1), denotes the EC of ATH sampling : where is the flow rate of the air, is the flow rate of the carrier liquid, and is the viral collection efficiency of the air sampler. Yang et al. reported that there are relatively low concentrations of virus particles in indoor air (10 3 -10 4 viral genome copies per 1 m 3 of air) even during an influenza epidemic (Yang et al., 2011) . Even during the COVID-19 pandemic, low concentrations of the virus (10 3 -10 4 viral genome copies per 1 m 3 of air) were detected in a negative-pressure room housing a confirmed patient in an isolation facility (Chia et al., 2020; Santarpia et al., 2020) . The LOD of commercial real-time qRT-PCR J o u r n a l P r e -p r o o f devices is typically 10 3 viral genome copies per 1 mL of liquid (Drosten et al., 2020; Corman et al., 2020) . Based on this, an of 10 5 to 10 6 -fold (i.e., from 10 3 -10 4 viral genome copies per 1 m 3 of air in a viral epidemic scenario to 10 3 viral genome copies per 1 mL of liquid) is required. During the air sampling process, airborne virus particles can be damaged by mechanical or chemical stress, rendering them undetectable and leading to a longer sampling time in order to produce a detectable sample, which hinders the rapid monitoring of airborne viruses (Piri et al., 2020; Kim et al., 2018) . Therefore, the sampling of airborne virus particles for biosensors and bioassays, including real-time qRT-PCR, needs to be non-destructive, which can be quantified using the recovery rate (R): where is the total concentration of virus particles collected via air sampling, and is the concentration of virus particles that are not damaged during the air sampling process. The SKC BioSampler (SKC Inc., USA) is a widely used ATH sampler that captures airborne biological particles in a liquid (20 mL) using mechanical impaction. However, it has been experimentally proven that the collection efficiency of the SKC BioSampler for nanosized viruses is less than 50% at a relatively low air sampling flow rate of 12.5 L/min (Li et al., 2018) . For example, the of the SKC BioSampler for a 20-min sampling period is 6,250. Furthermore, Hong et al. reported that the SKC BioSampler has low virus recovery rate (1% for T3 bacteriophages) which is a result of the desiccation and degradation of the collected virus particles due to the evaporation of the collection media (Cao et al., 2011) . Hong et al. thus introduced a batch-type electrostatic particle concentrator (EPC) that collects aerosolized virus particles in 0.5 mL of liquid . The collection efficiency of the EPC was found to be 35% with an applied voltage of -10 kV and an air sampling flow rate of 12.5 L/min for 100-nm virus particles. The of the EPC with a 20-min sampling period was 96,250, much higher than that of the SKC BioSampler. However, the recovery rate of the collected virus particles remained low (10% for the T3 bacteriophage). Hong et al. believed that the high-intensity electric field employed during the sampling process damaged the collected virus particles . Amin et al. also reported that long-term exposure to the ozone or reactive oxygen species (ROS) generated by corona discharge during the electrostatic sampling process could damage collected microorganisms (Piri et al., 2020) . Therefore, the air flow rate, viral collection efficiency, and recovery rate of the collected viruses need to be high. This study proposes a high air flow-rate electrostatic air sampler (HAFES) than can achieve an higher than 10 5 . This study also investigates the use of a continuously flowing carrier liquid because it is essential that the virus sample collected in the liquid is evacuated immediately to prevent exposure to ROS and ozone and thus ensure a high recovery rate. (Russell et al., 2008) . The MS2 bacteriophage has been widely used to evaluate air sampler performance because its morphology is similar to that of pathogenic viruses, including foot-and-mouth disease (FMD) virus, rhinovirus, and poliovirus Verreault et al., 2008; Pan et al., 2016) . The HAFES was designed to have a high collection efficiency and recovery rate with a high air flow rate (100 L/min). To achieve this, lab-scale tests were carried out to evaluate the collection efficiency of the HAFES for the test viruses using aerosol counts. The recovery rates of the viruses collected using the HAFES were compared with those collected using the SKC BioSampler with real-time qRT-PCR and plaque assays. Finally, the virus collection performance of the HAFES and the SKC BioSampler was compared using real-time qRT-PCR assays in a simulated viral epidemic scenario. to test the efficiency of size-dependent particle collection. NaCl has been used as a standard material for evaluating particle collection performance of an electrostatic air sampler (Miller et al., 2010; Han et al., 2009; Mainelis et al., 2002) . The prepared virus stocks were mixed J o u r n a l P r e -p r o o f with deionized water (total volume of 50 mL), and the mixture was atomized using a A/H1N1, A/H3N2, and HCoV-229E. These were similar to those of TEM images (see Section 3.1). The selectively generated virus particles entered the HAFES after 2 L/min of particleladen air flow had been diluted with clean air. The total air flow rate (40 to 100 L/min) entering the HAFES was controlled by two fans. In the section 3 of Supplementary Information, a method for generating the NaCl particles is described in detail. A 3D model of the proposed HAFES is presented in Fig. 2 . The HAFES contained a cap with 12 tungsten-wire electrodes and a body with an air flow inlet and an outlet. The Corona currents were measured under various air flow rates and applied voltages using a multimeter (8845a, Fluke, USA). I-V measurements were repeated three times. The electrostatic module of COMSOL Multiphysics (Version 5.4) was employed to simulate the electrical potential in the HAFES. The collection efficiency of the HAFES (Equation (1)) was calculated using where and are the concentrations of airborne particles measured at a location downstream the HAFES when the power for corona discharge is on and off, respectively. The size and concentration of the aerosolized virus and NaCl particles were measured using a scanning mobility particle sizer (SMPS, TSI, USA). The SMPS consisted of a classifier (3080, TSI Inc., USA), a DMA (3081, TSI Inc., USA), an aerosol charge neutralizer (Soft X-ray Charger 4530, HCT, Republic of Korea), and a condensation particle counter (CPC, 3775, TSI Inc., USA). The aerosol particles entering the CPC were condensed with butanol to spur growth; these particles then passed through a laser, and the number of particles was counted. To calculate R (Equation (2)), was defined as follows: . After 30 sec, materials other than the virus particles were wicked away using filter paper (01531055, ADVANTEC, USA). After this, 4 μL of 1% uranyl acetate (E22400-1, Science Services, Germany) was placed on the grid for 15 sec, and filter paper was used to remove the reagent. Another 4 μL of ultra-pure water was used to wash away the residue. The negatively stained virus on the grid was observed after drying for 30 min at room temperature. J o u r n a l P r e -p r o o f The virus sampling performance of the HAFES was compared with that of the SKC BioSampler for A/H1N1 and HCoV-229E, which were aerosolized and collected using the air samplers. According to Yang et al. (Yang et al., 2011) , the airborne virus concentration in indoor environments (a day-care center, health center, and airplane) during an IAV outbreak varies from 5.8×10 3 to 3.7×10 4 RNA genome copies/m 3 . Therefore, for A/H1N1 and HCoV- It was found that 99.0% and 99.7% of the RNA of A/H1N1 and HCoV-229E, respectively, were damaged during aerosolization with the atomizer. Therefore, a concentration of 3.57×10 6 RNA copies/m 3 (100-fold higher than 3.57×10 4 RNA copies/m 3 ) for A/H1N1 and 1.19×10 7 RNA copies/m 3 (333-fold higher than 3.57×10 4 RNA copies/m 3 ) for HCoV-229E were aerosolized with the atomizer. The measurement of the concentration Before testing with the virus aerosols, preliminary tests were conducted using NaCl. Section subsequent virus sampling. for 30-nm particles is higher than that for 100-nm particles. HCoV-229E, the recovery rate using the HAFES (81%) was 2.7 times higher than that using the SKC BioSampler (32%). For A/H1N1 and A/H3N2, the recovery rates with the HAFES were 4.9 times (88%) and 3.2 times (87%) higher, respectively, than those with the SKC BioSampler (18% and 27%, respectively). For the MS2 bacteriophage, the HAFES produced a 1.25-times higher recovery rate (94%) than did the SKC BioSampler (75%). More detailed results are available in Section 4 of Supplementary Information (Table S1 ). The higher viral recovery rate for the HAFES could be because the velocity of the airborne particles in the collection medium using electrostatic sampling is about two to four orders of magnitude J o u r n a l P r e -p r o o f lower than when using inertial impaction sampling at comparable air flow rates Mainelis, 1999) , meaning that the HAFES was likely to be less destructive to the virus particles than was the SCK BioSampler. The ozone concentrations were measured using an ozone counter (OZ 2000 G, SERES, France). The results were14±3 ppb, 10±2 ppb, and nondetectable at 0 cm, 3 cm, and 5 cm from the outlet of HAFES, respectively. However, the SKC BioSampler exhibited much lower recovery rates for HCoV-229E (32%) and the influenza viruses (18% and 27% for H1N1 and H3N2, respectively) than for the MS2 bacteriophage (75%). It has been reported in a previous study that influenza viruses generally have a lower recovery rate than the MS2 bacteriophage when an inertia impaction air sampling method is used (Ge et al., 2014) . When a large virus particle is captured by the collection media, it experiences a strong inertia force resulting in significant mechanical stress. Therefore, it is more likely that the coronavirus and influenza viruses will be damaged than the MS2 bacteriophage when the same air sampling speed is employed. The recovery rate for the MS2 bacteriophage collected using the HAFES was compared with the recovery rates obtained using the batch-type EPC and a continuous-type electrostatic ATH sampler (ATHS) . The recovery rate for the MS2 bacteriophage with the HAFES (94%) was higher than that with the ATHS (59%) and the EPC (55%). Further details are presented in Section 5 of Supplementary Information (Fig. S3(A) ). A/H1N1 and HCoV-229E particles at a concentration of 3.57×10 4 RNA copies/m 3 were separately suspended in the air to simulate a viral epidemic scenario. On the other hand, Fig. 5 confirms that a sampling time of 20 min was sufficient to produce detectable samples for real-time qRT-PCR when using the HAFES in a viral epidemic scenario. The HAFES was able to rapidly capture airborne viruses with a high collection efficiency even at a high air flow rate (71%, 79%, 76%, and 76% for HCoV-229E, the MS2 The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper. (Jaschhof, 1992) . The volume of the virus mixture used for the 1-min atomization was 0.5 mL. The gelatin filter loaded with virus particles was dissolved in 1 mL of deionized water. Finally, the viral RNA was extracted and analyzed. The concentrations of the A/H1N1 and HCoV-229E virus solutions in the atomizer were 3.03×10 7 RNA copy number/mL (Ct = 19.0) and 3.54×10 8 RNA copy number /mL (Ct = J o u r n a l P r e -p r o o f 18.3), respectively. Theoretically, it was expected that the concentrations of the A/H1N1 and HCoV-229E samples collected for 1 min would be 1.49×10 7 RNA copy number/mL (3.03×10 7 RNA copy number/mL × 0.5 mL / 1 mL, Ct = 19.9) and 1.74×10 8 RNA copy number/mL (3.54×10 8 RNA copy number /mL × 0.5 mL / 1 mL, Ct = 19.3). However, the actual concentrations of the detected virus samples were 1.49×10 5 RNA copy number/mL (Ct = 26.5) and 5.21×10 5 RNA copy number/mL (Ct = 27.7) for A/H1N1 and HCoV-229E, respectively. The unexpected difference in the Ct (26.5 -19.9 = 6.6 for A/H1N1 and 27.7 -19.3 = 8.4 for HCoV-229E) was caused by the damage that occurred during the aerosolization process with the atomizer. The damage to the virus ( ) during the aerosolization process can be calculated using Eq. S1: where is the concentration of the detected virus sample, and ℎ is the calculated concentration of the collected virus sample. The viral damage during the aerosolization process was 99.0% and 99.7% for A/H1N1 and HCoV-229E, respectively. The effect of the shear rate and oxygen injury during atomization and the combination of these factors have been reported as the reasons for the lower survival rate for microorganisms (Ghandi et al., 2012) . Therefore, 3.57×10 6 RNA copies/m 3 (100-fold higher than 3.57×10 4 RNA copies/m 3 ) of A/H1N1 and 1.19×10 7 RNA copies/m 3 (333-fold higher than 3. Fig. S3 (B) shows that the corona starting voltage was ± 3kV for all air flow rates. Large and Pierce reported that, in winds of less than a few meters per second, the point-discharge current is not a function of the wind speed (Large et al., 1957) . In this study, the air velocity in the HAFES was between 0.4 and 1.0 m/s. Beyond the corona starting voltage, the current gradually increased with the applied voltage. The current for a negative voltage was higher than that for a positive voltage of the same magnitude, which has been reported by previous studies . and 100 L/min, respectively. The collection efficiency increased as the applied voltage increased for the same polarity. According to classical aerosol-charging theory, particle collection efficiency increases with a higher particle charge, while the particle charge J o u r n a l P r e -p r o o f increases with a higher ion concentration (Hinds, 1999) . In Fig. S3 (D), S3(E), and S3(F), the collection efficiency was higher under an applied negative voltage than that under an applied positive voltage of the same magnitude because the negative discharge generated more air ions than did the positive discharge ( Fig. S3(B) ). Similarly, the collection efficiency with an applied voltage of -8 kV was higher than that with an applied voltage of +10 kV because the concentration of air ions was higher with an applied voltage of -8 kV (Fig. S3(B) ). The collection efficiency decreased as the air flow rate increased at all voltages. According to particle-charging theory, the charge number of a particle is proportional to the residence time of the particle exposed to the electric field (Hinds, 1999) . As the air flow rate increases, the time during which the particles are exposed to the electric field shortens, thus they are not sufficiently charged, resulting in lower aerosol collection efficiency. When the air flow rates were 60, 80, and 100 L/min, the residence times were 0.18, 0.14, and 0.11 sec, respectively. All the collection efficiency curves exhibited a U-shape, as in previous studies (McLean, 1988; Yoo et al., 1997; Ylätalo and Hautanen, 1998; Huang and Chen, 2002; Park et al., 2018) . The collection efficiency was lowest for particle sizes of 200-300 nm. This range represents the most penetrated particle size (MPPS) and does not change with differences in the applied voltage or the air flow rate. J o u r n a l P r e -p r o o f was 6×10 9 RNA copies/mL. The stock was diluted to a concentration of 6,462,000 RNA copies/mL and was assayed using real-time qRT-PCR. The Ct obtained for the diluted sample was 27.5. HCoV-229E particles collected using the HAFES were also assayed with real-time qRT-PCR. If the obtained Ct was 27.5, then the concentration of the HCoV-229E sample must be 6,462,000 RNA copies/mL. The concentration of HCoV-229E at the inlet of the HAFES was 7,980,000 RNA copies/mL, which was obtained using the method described in the main text. According to Equation (2) in the main text, the recovery rate for HCoV-229E, for example, was 81% (6,462,000/7,980,000) and 32% (2,556,000/7,980,000) using the HAFES and the SKC BioSampler, respectively. The recovery rates were very similar (81-90%) for HCoV-229E, A/H1N1, A/H3N2, and the MS2 bacteriophage when the HAFES was used for J o u r n a l P r e -p r o o f air sampling. However, the SKC BioSampler demonstrated much lower recovery rates for HCoV-229E (32%) and the influenza viruses (18% and 27% for H1N1 and H3N2, respectively) than for the MS2 bacteriophage (75%). The recovery rate of the MS2 bacteriophage using the HAFES was compared with that for the batch-type EPC and the continuous-type ATHS (Fig. S4(A) ) . Hong et al. reported that the recovery rate with the EPC was 55% for an applied voltage of -10 kV (which enabled the largest number of virus particles to be collected over a specific period of time) (Hong aet al., 2016) . In Park et al., the recovery rate using the ATHS was measured using the same experimental protocol described in this study . The recovery rate was 59% for an applied voltage of -7 kV (which enabled the largest number of virus particles to be collected over a specific period of time). Fig. S4 (A) illustrates that the recovery rate using the HAFES was 94%, which was much higher than that using the ATHS. It should be noted that the distance between the corona discharge electrode and the ground electrode in the HAFES (20 mm) was larger than that of the ATHS (10 mm). Therefore, the electric field in the HAFES was lower than that in the ATHS, resulting in the lower generation of ozone and ROS in the HAFES (Park and Hwang, 2013) . This may be the reason why the MS2 bacteriophage incurred less damage when using the HAFES. where V is the volume of the collected medium. The × R of the EPC was 4812.5 with an air flow rate, collection efficiency, recovery rate, and liquid volume of 12.5 L/min, 35%, 55%, and 0.5 mL, respectively. Physical collection efficiency of filter materials for bacteria and viruses Development of an improved methodology to detect infectious airborne influenza virus using the NIOSH bioaerosol sampler Accessible on Sep Detection of air and surface contamination by SARS-CoV-2 in hospital rooms of infected patients Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR Rapid Detection and Quantification of RNA of Ebola and Marburg Viruses, Lassa Virus, Crimean-Congo Hemorrhagic Fever Virus, Rift Valley Fever Virus, Dengue Virus, and Yellow Fever Virus by Real-Time Reverse Transcription-PCR Airborne virus survivability during long-term sampling using a non-viable Andersen cascade impactor in an environmental chamber Efficient collection of atmospheric aerosols with a particle concentrator-electrostatic precipitator sampler Gentle sampling of submicrometer airborne virus particles using a personal electrostatic particle concentrator Aerosol-to-hydrosol sampling and simultaneous enrichment of airborne bacteria for rapid biosensing Comparison of lab-made electrostatic rod-type sampler with single stage viable impactor for identification of indoor airborne bacteria Numerical investigation of the separation mechanism in an electrostatic aerosol-to-hydrosol separator by glow corona discharge: a quantitative comparison of the effects of ionic wind and Coulomb force Comparing the performance of 3 bioaerosol samplers for influenza virus Design and collection efficiency of a new electrostatic precipitator for bioaerosol collection Collection of airborne microorganisms by electrostatic precipitation Interaction between functionalized multiwalled carbon nanotubes and MS2 bacteriophages in water A handheld electrostatic precipitator for sampling airborne particles and nanoparticles Efficient collection of viable virus aerosol through laminar-flow, water-based condensational particle growth Continuous and real-time bioaerosol monitoring by combined aerosol-to-hydrosol sampling and ATP bioluminescence assay Prevention of damage caused by corona dischargegenerated reactive oxygen species under electrostatic aerosol-to-hydrosol sampling The global circulation of seasonal influenza A (H3N2) viruses Characterization of the 1918 influenza virus polymerase genes Methods for sampling of airborne viruses Concentrations and size distributions of airborne influenza A viruses measured indoors at a health centre, a day-care centre and on aeroplanes A novel coronavirus from patients with pneumonia in China Effect of shear rate and oxygen stresses on the survival of Lactococcus lactis during the atomization and drying stages of spray drying: a laboratory and pilot scale study Aerosol technology: Properties, behavior, and measurement of airborne particles Gentle sampling of submicrometer airborne virus particles using a personal electrostatic particle concentrator Ultrafine aerosol penetration through electrostatic precipitators Sampling virus aerosols using the gelatin membrane filter Aerosol-to-hydrosol sampling and simultaneous enrichment airborne bacteria for rapid biosensing The dependence of point-discharge currents on wind as examined by a new experimental approach Electrostatic precipitators Susceptibility constants of airborne bacteria to dielectric barrier discharge for antibacterial performance evaluation Computational fluid dynamic modelling of particle charging and collection in a wire-to-plate type single-stage electrostatic precipitator Continuous and real-time bioaerosol monitoring by combined aerosol-to-hydrosol sampling and ATP bioluminescence assay Electrostatic precipitator penetration function for pulverized coal combustion Charging and collection of submicron particles in two-stage parallelplate electrostatic precipitators Supplementary material related to this article can be found, in the online version, at doi:https://doi.org/10.1016/j.jhazmat.2020.xxxxxx.J o u r n a l P r e -p r o o f High Air Flow-rate Electrostatic Sampler for the Rapid