key: cord-0817613-kqqwmo39 authors: Kutter, Jasmin S.; de Meulder, Dennis; Bestebroer, Theo M.; Mulders, Ard; Fouchier, Ron A.M.; Herfst, Sander title: Comparison of three air samplers for the collection of four nebulized respiratory viruses ‐ Collection of respiratory viruses from air – date: 2021-06-14 journal: Indoor Air DOI: 10.1111/ina.12875 sha: aef56b1a15f5a465a7ed92b1e079b739f8b613f5 doc_id: 817613 cord_uid: kqqwmo39 Viral respiratory tract infections are a leading cause of morbidity and mortality worldwide. Unfortunately, the transmission routes and shedding kinetics of respiratory viruses remain poorly understood. Air sampling techniques to quantify infectious viruses in the air are indispensable to improve intervention strategies to control and prevent spreading of respiratory viruses. Here, the collection of infectious virus with the six‐stage Andersen cascade impactor was optimized with semi‐solid gelatin as collection surface. Subsequently, the collection efficiency of the cascade impactor, the SKC BioSampler, and an in‐house developed electrostatic precipitator was compared. In an in vitro set‐up, influenza A virus, human metapneumovirus, parainfluenza virus type 3, and respiratory syncytial virus were nebulized and the amount of collected infectious virus and viral RNA was quantified with each air sampler. Whereas only low amounts of virus were collected using the electrostatic precipitator, high amounts were collected with the BioSampler and cascade impactor. The BioSampler allowed straight‐forward sampling in liquid medium, whereas the more laborious cascade impactor allowed size fractionation of virus‐containing particles. Depending on the research question, either the BioSampler or the cascade impactor can be applied in laboratory and field settings, such as hospitals to gain more insight into the transmission routes of respiratory viruses. animal transmission experiments, and virus stability studies) and observational epidemiological studies during outbreaks. 5 Respiratory viruses can spread via different routes: direct contact (e.g., via handshaking), indirect contact (via contaminated surfaces), or through the air via droplets and/or aerosols. We define droplets as particles that can only travel short distances through the air before they settle onto the mucosa of individuals or nearby surfaces, and aerosols as particles that are small enough to remain suspended in the air for prolonged periods of time and cover large distances. To understand the relative importance of various transmission routes, knowledge on the viral load and infectivity of viruses in the air, as well as the size of virus-containing droplets and aerosols is warranted. To meet this need, the collection of viruses from the air with air samplers has gained increasing attention. Such air samplers can be applied in environments such as hospital settings, animal experiments, or livestock farms, [6] [7] [8] [9] [10] [11] however, collecting infectious virus from the air remains challenging. 12, 13 Apart from air samplers that employ a water-condensation growth method to capture particles, a major limitation of air samplers that use physical forces to separate aerosols and droplets from the air stream, is their high cut-off size, which prevents the efficient collection of small aerosols (<1 µm). 11, 14 To overcome this limitation, filters are sometimes installed after air samplers to capture aerosols and droplets that are too small to be collected by the air sampler itself. 6, 15 In addition, the high collection forces and sampling velocities applied inside these air samplers can damage virus particles, resulting in the collection of mainly non-infectious virus. 7, 16 In only a few studies, infectious respiratory viruses were collected from the air as demonstrated by virus isolation in cell cultures, 8, 9, 11, 17, 18 whereas in most studies the presence of virus in the air was solely determined by the detection of viral RNA by (quantitative) RT-PCR. [19] [20] [21] Here, existing and newly developed air samplers with different collection methods were improved and compared: the six-stage Andersen cascade impactor, the SKC BioSampler, and an in-house developed electrostatic precipitator. The BioSampler and cascade impactor are commercially available air samplers that employ inertial forces to remove aerosols and droplets from the airflow. The cascade impactor was originally developed to collect airborne bacteria and fungi onto petri dishes filled with bacteriological agar. However, agar is less suitable as collection medium for viruses because of the high impaction forces and desiccation effects on aerosols and droplets. [22] [23] [24] For the cascade impactor and BioSampler, the collection efficiency is low for aerosols in the submicron range. [25] [26] [27] To overcome the poor collection efficiency of small aerosols, we developed an electrostatic precipitator. Electrostatic precipitators are widely used to remove small particles such as dust from the air, and they are being increasingly explored for air sampling of airborne microorganisms. [28] [29] [30] [31] In electrostatic precipitators, the air around a conductor is ionized through the application of high voltage. Incoming droplets and aerosols get charged and attracted into a neutral or oppositely charged collection medium. These air samplers have a low flow rate and subject aerosols and droplets to less physical stress, thereby yielding higher recovery rates of infectious microorganisms. 30, 31 In this study, the collection of infectious virus with the cascade impactor was first improved by optimizing the collection medium. Subsequently, the efficiency to collect infectious virus and viral RNA of the BioSampler, cascade impactor and electrostatic precipitator for nebulized pandemic H1N1 influenza A virus (pH1N1), human metapneumovirus (HMPV), human parainfluenza virus type 3 (PIV3), and respiratory syncytial virus (RSV) was compared in an in vitro set-up. Finally, the sensitivity of the BioSampler, and cascade impactor for low virus concentrations was evaluated. • Respiratory viruses pose a continuous health threat, and knowledge on their transmission routes will help to prevent their spread in the population. • Knowledge on the quantity of an infectious respiratory virus in the air will help to predict the importance of transmission of this virus through the air. • Most currently available air sampling devices have not been designed to collect infectious viruses from the air, hence the optimization and comparison of their ability to collect different respiratory viruses from the air is of utmost importance. medium. Vero-118 cells were inoculated with either HMPV-GFP or PIV3-GFP at an moi of 0.01. Cells and supernatant were harvested 7 (HMPV-GFP) or 5 (PIV3) dpi and stored at −80℃. After one freezethaw cycle, cell-free supernatants were concentrated and purified on a 30/60% sucrose cushion and subsequently stored in 25% sucrose at −80℃. Vero-118 cells were maintained in IMDM supplemented with 10% FBS and 100 IU/ml penicillin-100 µg/ml streptomycin mixture (Lonza). (Lonza), 10 mM Hepes (Lonza). All cells were cultured at 37℃ and 5% CO 2 . Virus stock concentrations were calculated as described below and were 6.8, 6.6, 6.9, and 7.5 log 10 TCID 50 /ml for pH1N1, HMPV, PIV3, and RSV, respectively. The viable six-stage Andersen cascade impactor (Thermo Scientific) operates at 28.3 liters per minute (LPM) and consists of six stages, with 400 orifices each, and 6 petri dishes 36 ( Figure 1A ). With increasing stage number, the size of the orifices decreases, and hence, impaction velocity increases, enabling the collection of sizefractionated droplets and aerosols over the different petri dishes. (1) to generate aerosols and droplets containing virus particles in an air-tight chamber (3), which was connected via a tube (4) to an air sampler (5) . A second tube (6) was placed between the air sampler and a vacuum pump (8) that was placed outside the BSL II cabinet. High-efficiency particulate air (HEPA) filters (2, 7) were installed on both sides of the air sampling set-up to guarantee that clean air entered the box and to prevent contamination of the environment. The direction of airflow is indicated with orange arrows. The flow rate through the system equals the recommended flow rate of the different samplers. For each experiment, nebulized viruses were collected from the chamber with air samplers for 5 min minimizes the chances of particle re-nebulization and maintains the infectiousness of collected particles. When aerosols and droplets exit the nozzles they get impinged into the liquid medium, while the remaining air exits the air sampler through the outlet. As collection medium, 15 ml virus transport medium (VTM) was used. Twenty µl antifoam B emulsion (Sigma Aldrich) was added to prevent the generation of bubbles and foam due to the swirling motion of the collection medium during air sampling. The in-house developed electrostatic precipitator is made of a glass chamber consisting of an upper and bottom part ( Figure 1C) . A voltage of 13 kV is applied to a 80 mm long corona wire that is attached to the upper part of the glass chamber with a distance of 20 mm between the wire and the bottom part ( Figure S1 ). Application of high voltage produces an ion discharge which ionizes the air in the chamber. Upon collision of incoming aerosols and droplets with ionized air molecules, aerosols and droplets become charged and attracted by the neutral bottom part of the glass chamber which is filled with 20 ml VTM. As a side effect, corona discharges also generate ozone, which is known to inactive viruses. 37, 38 A positive charge was used in the electrostatic precipitator because it produces less ozone than a negative charge. 39, 40 The electrostatic precipitator is operated at 4 LPM. Air samplers were connected to a chamber (Nalgene BioTransport nebulizer that uses electricity to produce aerosols and droplets between 1 and 5 µm, with a reported mass median aerodynamic diameter of 3.05 µm, when nebulized in a 12 L chamber. 41, 42 The vacuum pump was switched on just before nebulization and air was drawn through the air samplers for a total of 5 min ( Figure 2B ). Subsequently, air samples were retrieved from the samplers, agar and semi-solid gelatin samples were processed as described above and all samples were subjected to further analysis. All experiments were performed in a class 2 biosafety cabinet. Prior to the air sampling experiments, the performance of the nebulizer was assessed, as the mechanical nebulization process may af- A direct comparison of the titers with and without nebulization demonstrated that pH1N1 virus infectivity was barely affected by this process, as the virus titer after nebulization was only 0.03 log 10 T-CID 50 lower than without nebulization. The virus titers of HMPV, PIV3, and RSV were reduced by 0.58, 0.55, and 0.54 log 10 TCID 50 , respectively ( Table 1 ). The loss of virus infectivity during nebulization may be due to incomplete nebulization of virus suspensions or due to viruses | 1879 KUTTER ET al. Virus titer before nebulization (log 10 TCID 50 +/−SD) Virus titer after nebulization (log 10 TCID 50 +/−SD) For pH1N1 virus, collection of infectious virus was equally efficient when agar or semi-solid gelatin was used, and only 0.6 and 0.8 log 10 TCID 50 were lost, respectively, as compared to the positive control. Collection of infectious pH1N1 virus in VTM was much less efficient and resulted in a reduction of 2.4 log 10 TCID 50 ( Figure 3A ). The total collection efficiency of the cascade impactor for pH1N1 virus RNA as compared to the positive control was 16.4% and 6.4% for agar and semi-solid gelatin, but interestingly 39.4% for VTM, despite the substantial loss of virus infectivity ( Figure 3B ). Also, for HMPV, the amount of infectious virus collected with each medium varied. The collection efficiency was highest with semi-solid gelatin and VTM, where 1.4 and 1.6 log 10 TCID 50 less infectious virus was recovered after air sampling as compared to the positive control ( Figure 3C ). When agar was used as collection medium, considerably less infectious HMPV was collected with a reduction of 2.4 log 10 T-CID 50 as compared to the positive control ( Figure 3C ). The total collection efficiency of the cascade impactor for HMPV RNA varied from 1.4%, 5.4%, to 12.3% for semi-solid gelatin, agar, and VTM, respectively. Thus, also for HMPV, the highest physical collection efficiency was obtained with VTM. (Figure 3D ). numbers as compared to the lower stage numbers. (Figure 3E , F). The distribution of infectious HMPV and HMPV RNA over the six stages was slightly more variable than that of pH1N1 virus. Interestingly, substantially lower amounts of infectious HMPV and viral RNA were captured in stage 6 as compared to pH1N1 ( Figure 3G, H) . Overall, the infectivity of both viruses was well conserved with semi-solid gelatin, which was therefore used in subsequent experiments. The highest collection efficiency of infectious virus was obtained with the cascade impactor for pH1N1 virus and the BioSampler for HMPV and RSV, whereas similar amounts of PIV3 were collected with the cascade impactor and BioSampler ( Figure 4A ). For all four viruses, only low amounts of infectious virus were collected with the in-house developed electrostatic precipitator ( Figure 4A ). The collection efficiency of the electrostatic precipitator for virus RNA was also very low demonstrating that the overall collection of viruses with this sampler was poor ( Figure 4B ). When the distribution patterns of the four viruses over the six stages of the cascade impactor were investigated, the amounts of collected infectious pH1N1 virus and pH1N1 RNA were found to be similar in all stages. In contrast, considerably lower amounts of infectious virus and virus RNA were collected in stage 6 for HMPV, PIV3, and RSV as compared to the other stages ( Figure 4C, D) . In cascade impactors, originally agar was used to collect bacteria from the air. However, agar is generally considered less suitable as a collection surface for viruses, given the possibility of desiccation and increased particle bounce. [43] [44] [45] In addition, it was demonstrated for infectious bursal disease virus, that virus recovery from agar is reduced significantly when petri dishes are processed at later time points, and immediate processing is not always possible in field Therefore, to collect smaller aerosols more efficiently, an electrostatic precipitator was developed in-house and also tested with all four respiratory viruses. Although the collection efficiency of the electrostatic precipitator, as determined by using an aerodynamic particle sizer, ranged from 65% to 43% for particles between 0.2 and 0.5 µm (Table S1) should be noted that the nebulizer creates a higher relative humidity compared to a jet nebulizer and that in the current study aerosols and droplets were directly nebulized into a small chamber (11.5 L). 42 In other studies, a dryer is often added to the experimental set-up that further dries aerosols and droplets before entering a collection chamber and thereby further decreasing particle sizes. 49 Despite in the current set-up particles were not dried and therefore particle sizes potentially larger as compared to other studies, Bowling et al. showed that the average particle size produced by the nebulizer was 3.1 µm in a 12 L chamber with 80.89% and 59.03% of the generated particles being ≤5 µm and ≤3 µm, and that the average humidity was not significantly higher in that chamber than when particles were produced with the Collison nebulizer. 42 Also, in the current experimental set-up high amounts of virus were nebulized, whereas, under field conditions, it is more likely that lower amounts of virus are dispersed over larger volumes of air. For the collection of representative samples under field conditions therefore air samplers likely need to run for longer time periods, whereas in the experimental set-up presented here, air samplers only run for five minutes. Particularly in the case of electrostatic precipitation, no device is commercially available yet that is specifically designed to collect infectious respiratory viruses from the air, and hence, air samplers are custom made designs. 25, 28, 57, 58 This makes it very difficult to directly compare the performance of the air samplers evaluated here, with other studies. In conclusion, in the present study, the commercially available BioSampler and cascade impactor are both capable of collecting artificially generated droplets and aerosols containing respiratory viruses while maintaining their infectivity during the sampling process. With the cascade impactor, quantitative data on the sizes of virus-containing particles can be obtained, and in combination with a semi-solid gelatin layer as collection surface, the cascade impactor is also easy to use in various field settings such as hospitals. With the BioSampler size fractionation of the collected aerosols and droplets is unfortunately not possible. However, collection is more facile, since only one air sample is obtained per collection moment and no post-air sampling processing is needed. The choice for either of the two air samplers therefore also depends on the environment in which it is to be used and on the research questions to be addressed. Overall, implementation of these air samplers in field studies will help to obtain more quantitative data on the amount of infectious respiratory virus that is present in the air, thereby generating a better understanding of respiratory virus transmission. This work was financed through an NWO VIDI grant (contract number 91715372), NIH/NIAID contract HHSN272201400008C and European Union's Horizon 2020 research and innovation program VetBioNet (grant agreement No 731014). The authors declare no competing interests. The peer review history for this article is available at https://publo ns.com/publo n/10.1111/ina.12875. World Health Organization. The top 10 causes of death Global, regional, and national disease burden estimates of acute lower respiratory infections due to respiratory syncytial virus in young children in 2015: a systematic review and modelling study The etiological role of common respiratory viruses in acute respiratory infections in older adults: a systematic review and meta-analysis Estimates of the global, regional, and national morbidity, mortality, and aetiologies of lower respiratory infections in 195 countries, 1990-2016: a systematic analysis for the Global Burden of Disease Study Transmission routes of respiratory viruses among humans Quantification of influenza virus RNA in aerosols in patient rooms Detection of measles virus RNA in air and surface specimens in a hospital setting Assessment of air sampling methods and size distribution of virus-laden aerosols in outbreaks in swine and poultry farms Detection and isolation of airborne influenza a H3N2 virus using a sioutas personal cascade impactor sampler Measurement of airborne influenza virus in a hospital emergency department Viable SARS-CoV-2 in the air of a hospital room with COVID-19 patients Near-patient sampling to assist infection control-a case report and discussion Methods for sampling of airborne viruses Efficient collection of viable virus aerosol through laminar-flow, water-based condensational particle growth The environmental deposition of influenza virus from patients infected with influenza A(H1N1)pdm09: implications for infection prevention and control Detection of airborne severe acute respiratory syndrome (SARS) coronavirus and environmental contamination in SARS outbreak units Viable influenza A virus in airborne particles expelled during coughs versus exhalations Influenza virus aerosols in human exhaled breath: particle size, culturability, and effect of surgical masks Influenza virus in human exhaled breath: an observational study Exposure to influenza virus aerosols during routine patient care Aerosol sampling in a hospital emergency room setting: a complementary surveillance method for the detection of respiratory viruses Effect of impact stress on microbial recovery on an agar surface Improved aerosol collection by combined impaction and centrifugal motion Enhancing bioaerosol sampling by andersen impactors using mineral-oil-spread agar plate Gentle sampling of submicrometer airborne virus particles using a personal electrostatic particle concentrator Sampling methodologies and dosage assessment techniques for submicrometre and ultrafine virus aerosol particles Investigation of cut-off sizes and collection efficiencies of portable microbial samplers Sampling and detection of airborne influenza virus towards point-of-care applications Electrostatic precipitation of airborne bio-aerosols Collection of airborne microorganisms by electrostatic precipitation Assessment of electrical charge on airborne microorganisms by a new bioaerosol sampling method Experimental human metapneumovirus infection of cynomolgus macaques (Macaca fascicularis) results in virus replication in ciliated epithelial cells and pneumocytes with associated lesions throughout the respiratory tract An improved plaque reduction virus neutralization assay for human metapneumovirus Real-time reverse transcriptase pcr assay for detection of human metapneumoviruses from all known genetic lineages Incidence of viral respiratory pathogens causing exacerbations in adult cystic fibrosis patients New sampler for the collection, sizing and enumeration of viable airborne particles Inactivation of surface viruses by gaseous ozone In vitro inactivation of herpes virus by ozone Ozone generation in positive corona electrostatic precipitators Discharge electrodes and electrostatic precipitators A vibrating mesh nebulizer as an alternative to the collison 3-jet nebulizer for infectious disease aerobiology Experimental studies on particle impaction and bounce: effects of substrate design and material Particle bounce in cascade impactors Performance evaluation of commonly used impaction substrates under various loading conditions Airborne virus sampling: efficiencies of samplers and their detection limits for infectious bursal disease virus (IBDV) Evidence of respiratory syncytial virus spread by aerosol time to revisit infection control strategies? Evaluation of bioaerosol samplers for the detection and quantification of influenza virus from artificial aerosols and influenza virusinfected ferrets Collection and measurement of aerosols of viable influenza virus in liquid media in an Andersen cascade impactor Sampling efficiencies of two modified viable cascade impactors Airborne influenza virus detection with four aerosol samplers using molecular and infectivity assays: considerations for a new infectious virus aerosol sampler A newly discovered human pneumovirus isolated from young children with respiratory tract disease The morphology and assembly of respiratory syncytial virus revealed by cryo-electron tomography Spherical influenza viruses have a fitness advantage in embryonated eggs, while filament-producing strains are selected in vivo The M segment of the 2009 pandemic influenza virus confers increased neuraminidase activity, filamentous morphology, and efficient contact transmissibility to A/Puerto Rico/8/1934-based reassortant viruses Influenza virus survival in aerosols and estimates of viable virus loss resulting from aerosolization and air-sampling Design and collection efficiency of a new electrostatic precipitator for bioaerosol collection Ionizing air affects influenza virus infectivity and prevents airborne-transmission