key: cord-0817570-2tdvsslg authors: Dayya, David; O’Neill, Owen J.; Feiertag, Tracy; Tuazon-Boer, Ritzel; Sullivan, John; Perez, Lubiha; Gurash, Sandrah; Eaton, Matthew; Bodley, Taylor; Marker, Joanne; Smykowski, Elizabeth; Hall, T. title: The Use of Oxygen Hoods in Patients Failing on Conventional High-flow Oxygen Delivery Systems, the Effects on Oxygenation, Mechanical Ventilation and Mortality Rates in Hypoxic Patients with COVID-19. A Prospective Controlled Cohort Study date: 2021-02-12 journal: Respir Med DOI: 10.1016/j.rmed.2021.106312 sha: 8ae6df2cebb1275a891fa648afa4d761b8f688db doc_id: 817570 cord_uid: 2tdvsslg INTRODUCTION: Efforts to meet increased oxygen demands in COVID-19 patients are a priority in averting mechanical ventilation (MV), associated with high mortality approaching 76.4 – 97.2%. Novel methods of oxygen delivery could mitigate that risk. Oxygen hoods/helmets may improve: O(2)-saturation (SaO2), reduce in-hospital mechanical ventilation and mortality rates, and reduce length of hospitalization in hypoxic Covid-19 patients failing on conventional high-flow oxygen delivery systems. METHODS: . DESIGN: Prospective Controlled Cohort Study. SETTING: Single Center. PARTICIPANTS: All patients admitted with a diagnosis of COVID-19 were reviewed and 136/347 patients met inclusion criteria. STUDY PERIOD: 3/6/2020 to 5/1/2020. 136 participants completed the study with known status for all outcome measures. INTERVENTION OR EXPOSURE: Oxygen hoods/helmets as compared to conventional high-flow oxygen delivery systems. MAIN OUTCOME(S) AND MEASURE(S): 1) Pre and post change in oxygen saturation (SaO(2)). 2) In-hospital Intubation. 3) In-hospital Mortality. 4) Length of hospitalization. RESULTS: 136 patients including 58-intervention and 78-control patients were studied. Age, gender, and other demographics/prognostic indicators were comparable between cohorts. Oxygen hoods averted imminent intubation/MV in all 58 COVID-19 patients failing on conventional high-flow oxygen delivery systems with a mean improvement in SaO(2) of 8.8%, p < 0.001. MV rates were observed to be higher in the control 37/78 (47.4%) as compared to the intervention cohort 23/58 (39.7%), a difference of 7.7%, a 27% risk reduction, not statistically significant, OR 95%CI 0.73 (0.37 to 1.5). Mortality rates were observed higher in the control 54/78 (69.2%) as compared to the intervention cohort 36/58 (62.1%), a difference of 7.1%, a 27% risk reduction, not statistically significant OR 95%CI 0.73 (0.36 to 1.5). CONCLUSION: Oxygen hoods demonstrate improvement in SaO(2) for patients failing on conventional high-flow oxygen-delivery systems and prevented imminent mechanical ventilation. In-hospital mechanical ventilation and mortality rates were reduced with the use of oxygen hoods but not found to be statistically significant. The oxygen hood is a safe, effective oxygen-delivery system which may reduce intubation/MV and mortality rates. Their use should be considered in treating hypoxic COVID-19 patients. Further research is warranted. TRIAL REGISTRATION: ClinicalTrials.gov Identifier: NCT 04407260. Efforts to meet increased oxygen demands in COVID-19 patients are a priority in averting mechanical ventilation (MV), associated with high mortality approaching 76.4 -97.2%. Novel methods of oxygen delivery could mitigate that risk. Oxygen hoods/helmets may improve: O 2saturation (SaO2), reduce in-hospital mechanical ventilation and mortality rates, and reduce length of hospitalization in hypoxic Covid-19 patients failing on conventional high-flow oxygen delivery systems. Setting: Single Center. Participants: All patients admitted with a diagnosis of COVID-19 were reviewed and 136/347 patients met inclusion criteria. Study period: 3/6/2020 to 5/1/2020. 136 participants completed the study with known status for all outcome measures. Oxygen hoods/helmets as compared to conventional high-flow oxygen delivery systems. 1) Pre and post change in oxygen saturation (SaO 2 ) 2) In-hospital Intubation 3) In-hospital Mortality 4) Length of hospitalization. Results 136 patients including 58-intervention and 78-control patients were studied. Age, gender, and other demographics/prognostic indicators were comparable between cohorts. Oxygen hoods averted imminent intubation/MV in all 58 COVID-19 patients failing on conventional high-flow oxygen delivery systems with a mean improvement in SaO 2 of 8.8%, p < 0.001. MV rates were observed to be higher in the control 37/78 (47.4%) as compared to the intervention cohort 23/58 (39.7%), a difference of 7.7%, a 27% risk reduction, not statistically significant, OR 95%CI 0.73 (0.37 to 1.5). Mortality rates were observed higher in the control 54/78 (69.2%) as compared to the intervention cohort 36/58 (62.1%), a difference of 7.1%, a 27% risk reduction, not statistically significant OR 95%CI 0.73 (0.36 to 1.5). Oxygen hoods demonstrate improvement in SaO 2 for patients failing on conventional high-flow oxygendelivery systems and prevented imminent mechanical ventilation. In-hospital mechanical ventilation and mortality rates were reduced with the use of oxygen hoods but not found to be statistically significant. The oxygen hood is a safe, effective oxygen-delivery system which may reduce intubation/MV and mortality rates. Their use should be considered in treating hypoxic COVID-19 patients. Further research is warranted. Trial Registration: ClinicalTrials.gov Identifier: NCT 04407260 J o u r n a l P r e -p r o o f Introduction The shortage of ventilators worldwide during the COVID-19 crisis is an ongoing front-line challenge facing Healthcare practitioners 1 . Large numbers of patients require ventilator management with a limited supply to meet these high demands during the Coronavirus pandemic 2 . Meeting the increased oxygen demands of patient's is a critical priority in delaying or avoiding intubation and mechanical ventilation (MV), associated with a high mortality approaching 76.4 -97.2% 3 4 . Novel methods of oxygen delivery to patients may help mitigate this risk while more resources become available [5] [6] [7] [8] . This could provide needed time for emergency supportive measures to catch up with the critical oxygenation demands of patients worldwide 5 . Oxygen hoods routinely used in undersea & hyperbaric medicine centers were implemented once available at our institution in patients whose oxygen saturation (SaO 2 ) was below 90% and were failing on conventional high-flow oxygen delivery systems. These conventional oxygen delivery systems routinely used include non-rebreather masks (NRB), venturi masks (VM), high-flow nasal cannula (HFNC), continuous positive airway pressure (CPAP), or bilevel positive airway pressure (BiPAP) [7] [8] [9] . The total cost of the hood apparatus varies, but is approximately $250.00 (US) 10 . The oxygen hood system is cost-effective and could help address the demand for effective supplementary alternative oxygen delivery systems in the US and developing nations when compared to ventilator costs which are expensive and where supply is critical 11 . The oxygen hoods can be cleaned and reused according to the position paper published by the European Committee on Hyperbaric Medicine (ECHM) during the Covid-19 pandemic 12 . See Figure 1 below. To compare the differences in oxygenation before and after the use of oxygen hoods on COVID-19 patients failing on conventional high flow oxygen systems. 2) To compare mechanical ventilation rates, mortality rates, and length of hospitalization in COVID-19 patients failing on conventional oxygen delivery systems who are subsequently placed on oxygen hoods as compared to COVID-19 patients failing on conventional high flow oxygen delivery systems prior to the availability of oxygen hoods. 3) To determine which risk factors, and whether the use of the intervention oxygen hoods can predict the outcomes mechanical ventilation and mortality rates. The first cases of confirmed COVID-19 presented to our institution in Westchester, NY on 3/6/2020. The oxygen hoods were available for patient use on 4/3/2020. The study period includes 3/6/2020 to 5/1/2020. Once the oxygen hoods became available ALL COVID-19 patients failing on conventional high flow oxygen delivery systems (non-rebreather masks, high-flow nasal cannula, CPAP, BiPAP) with an SaO 2 below 90%, declining clinical status, and imminent need for intubation were referred for evaluation and use of the oxygen hood to potentially prevent the need for intubation and MV 13 . These patients represented severe refractory cases of respiratory failure requiring immediate or imminent mechanical ventilation prior to the availability of oxygen hoods. Patients using or failing on conventional high-flow oxygen delivery presenting prior to oxygen hood availability served as controls. Controls were selected if they required conventional high-flow oxygen delivery systems to maintain SaO 2 above 90%; or if they required mechanical ventilation due to already failing on conventional high-flow oxygen delivery systems. The intent was to balance intervention and control cohorts for respiratory disease severity. All patients included tested positive for COVID-19 using PCR swabs and/or diagnosis based on clinical/laboratory standard diagnostic criteria 14 . Medical management included evolving standard treatment regimens such as Hydroxychloroquine/Azithromycin, proning and other regimens widely used at the time of this study 15 . Patients were not involved in the decision process for this study as our alternative high-flow oxygen delivery method through the oxygen hood was already an accepted intervention for delivering high-flow oxygen in the Undersea and Hyperbaric Medicine department. In prior studies it was also found to be an alternative to intubation in patients suffering from ARDS prior to COVID-19. The oxygen hoods were attached to the oxygen flowmeters in negative-pressure ventilation rooms via a modified adapter (Sully-adaptor), developed in the department for this purpose. The oxygen hoods required a high-flow rate minimally set at 30 Liters/minute. Flowmeters, in patient rooms, can supply rates exceeding 30 L/min when turned to the full-flush position (40-60 Liters). Full-flush is when the knob on the flowmeter was turned to the fully open position or until it stopped turning 16 . This is done irrespective of the standard ball indicator which usually reads to a maximum of 15 liters/minute. The higher flow rates (all sources of oxygen delivery) required the main oxygen supply pressure to be increased accordingly to support the high-flow demands throughout the hospital. HEPA filters (Hudson RCI Bacterial/Viral Filter, or Ultipor 100 Filter) were placed on the oxygen hood exhaust and had a reported Bacterial/Viral Filtration Efficiency (VFE) equivalent to 99.999+% 17 . These filters helped reduce the exposure risk to staff whereas conventional high-flow delivery systems lack this safeguard and flow freely as open systems to the surrounding environments unfiltered. A Positive-End-Expiratory-Pressure (PEEP) adaptor is an attachment option available for use with the oxygen hoods, but controversy exists on using PEEP in COVID-19 patients 18 . The excess oxygen released from the filtered exhaust manifold of the hood was subsequently vented from the negative-pressure ventilation rooms. This reduced the risks associated with increased oxygen levels in confined spaces including the leaks inherent in all conventional high-flow oxygen delivery systems. The transparent oxygen hood also allows for visual assessment of the patient's head. Oxy-hemoglobin saturation is continuously measured by pulse-oximetry including immediately before and after oxygen hood placement. Serial arterial blood gas (ABG) measurements are not widely used due to the invasiveness of recurrent arterial punctures, bleeding, thrombosis, and over-extended staff burdens during the pandemic. Clinical decisions were frequently made based on pulse-oximetry, clinical, and laboratory assessments of respiratory decompensation and failure. Patients were followed to discharge/survival, or in-hospital mortality as endpoints in both control and intervention cohorts. 14 . Routine follow-up evaluation was maintained until final in-hospital outcomes were known for all the following: • Hospital census was reviewed for ALL patients seen in the ED and admitted with COVID -19 diagnosis experiencing hypoxia requiring supplemental high-flow oxygen delivery or mechanical ventilation, along with those failing on these high-flow oxygen systems. • Consent (native language services provided). • No limitations/restrictions based on age, gender, race/ethnicity, comorbidities, pregnancy status, DNR/DNI status. • SaO 2 > 90% and controlled on room air (RA) or low-flow O 2 -delivery system (nasal cannula, simple mask, or venturi-mask). • Confinement anxiety post-oxygen hood placement with request for removal. Randomization was not possible as it is unethical to deny a hypoxic patient an alternative approved means of oxygenation for patients failing on conventional oxygen delivery systems. There was no blinding of participants or investigators. The control cohort included hypoxic COVID-19 patients presenting for treatment prior to 4/3/20 when hyperbaric oxygen hoods were not available. The patients maintained on conventional O 2 -delivery systems or failing on these systems and requiring mechanical ventilation prior to the availability of oxygen hoods were studied. Using an alpha cut off = 0.05, power = 0.90, with minimum effect size differences of 10% and 20% in mechanical ventilation or mortality, would require a minimal sample size of 526 and 158 patients, respectively. Sensitivity analysis was not required as there was no missing outcome data from either cohort. Between 3/6/2020 and 5/1/2020 a total of 347 patients were evaluated in the emergency department and admitted to the hospital with a diagnosis of COVID-19 and respiratory symptoms. 63 consultations were conducted, 58 patients met inclusion criteria and were placed on oxygen hoods, see Figure 2 Flowchart. A total of 78/347 patients who did not receive oxygen hoods and were maintained on conventional high-flow oxygen delivery systems during hospitalization or who had failed conventional high-flow oxygen delivery and required mechanical ventilation were used as controls. The total sample size included known final status outcomes on all 136 patients in both cohorts by 5/29 including: home-discharge status, mechanical ventilation status, in-hospital mortality, and length of stay. There was no missing outcome data to report in either cohort. A total of 136 patients were studied 19 . 58/58 patients demonstrated immediate improvement in their oxygen saturation to above 90%. The overall mean improvement in SaO 2 was 8.9%, 95%CI (7.8 to 9.8). See table 1. The intervention and control cohorts were compared and not found to have statistically significant differential levels of comorbidities, or potential confounding factors with the exception of Diabetes (DM), Chronic Kidney Disease (CKD), as these were higher in the control as compared to the intervention group. Convalescent plasma antibody administration was higher in the intervention than the control group. The CKD and Plasma had few events in at least one cell with events < 5. Differences in CKD between the cohorts were observed to be significant but differences in creatinine levels upon admission between the cohorts was not found to be statistically significant. The comorbidity chronic lung disease was observed to be higher in the control group 13/78 (17%) as compared to the intervention group 7/58 (12%); whereas smoking was lower in the control 20/66 (26%) as compared to the intervention group 19/55 (35%). Neither chronic lung disease nor smoking reached the threshold for significance (p=0.45); (p=0.62). Cardiovascular disease was higher in the control 28/78 (36%) as compared to the intervention group 14/58 (24%) but was not found to be statistically significant. These findings were similar whether equal variances were assumed or not. There were differences in lab markers of disease between both groups however none of these rose to the threshold of significance, see Table 2 . *Independent t-test used for analysis ** Subsample of hooded patients used for portion of study which excluded 10 existing hospitalized patients whose final outcomes on intubation and mortality were available at the end of the study period. *** number of subjects with the comorbidity over the total number of subjects in the sample where information on the comorbidity was available in the record. ****All percentages are given as percent of the total column sample irrespective of missing data. ***** Events < 5 in at least one cells. The total sample included 60/136 (44.1%) patients that required and 76/136 (55.9%) patients that did not require mechanical ventilation (MV). Mortality in the total sample was observed to be 45/60 (75%) in the patients requiring mechanical ventilation and 42/71 (59.2%) in the patients not requiring mechanical ventilation, or a difference of 15.8% and a 107% increase in mortality in the ventilated patients. These observed differences were not found to be statistically significant OR 95% CI 2.07 (0.98 to 4.34). See Table 3a . Mechanical ventilation rates were observed to be higher in the control 37/78 (47.4%) as compared to the intervention cohort 23/58 (39.7%), a difference of 7.7% or a 27% reduction in the MV rate in the group receiving the oxygen hoods, this reduction was not found to be statistically significant, OR 95%CI 0.73 (0.37 to 1.5). Mortality rates were observed to be higher in the control 54/78 (69.2%) as compared to the intervention cohort 36/58 (62.1%), a difference of 7.1% or a 27% reduction in mortality in the group receiving oxygen hoods, this reduction was not found to be statistically significant, OR 95%CI 0.73 (0.36 to 1.5). The mean length of hospitalization was observed to be 12.62 days in the intervention and 10.13 days in the control cohort. The intervention cohort receiving oxygen hoods experienced a longer LOS than the control cohort, this reduction was not found to be statistically significant, Mean 95%CI, 2.49 days (-0.37 to 5.35). See table 3b. Ho: There is no difference in the MV rates, and/or mortality rates, and/or hospital length of stay in COVID-19 patient receiving oxygen hoods who failed on conventional high-flow oxygen delivery systems; as compared to those maintained on high-flow oxygen delivery systems OR failing on high-flow oxygen delivery systems AND requiring mechanical ventilation, under normobaric conditions. Ha: There is a difference in the intubation rates, and/or mortality rates, and/or hospital length of stay in COVID-19 patient receiving oxygen hoods who failed on conventional high-flow oxygen delivery systems; as compared to those maintained on highflow oxygen delivery systems OR failing on high-flow oxygen delivery systems AND requiring mechanical ventilation, under normobaric conditions. Mechanical ventilation rates were observed to be higher in the control 36/76 (47.4%) as compared to the intervention cohort 23/37 (40.5%), a difference of 6.9% or a 24% reduction in MV in the group receiving oxygen hoods, this reduction was not found to be statistically significant, OR 95%CI 0.76 (0.34 -1.7). Mortality rates were observed to be higher in the control 52/76 (68.4%) as compared to the intervention cohort 23/37 (62.2%), a difference of 6.2% or a 24% reduction in mortality in the group receiving oxygen hoods, this reduction was not found to be statistically significant, OR 95%CI 0.76 (0.33 -1.7). The mean length of hospitalization was observed to be 10.57 days in the intervention and 10.36 days in the control group. The controls demonstrated a slightly shorter LOS, than the group receiving oxygen hoods, this reduction was not found to be statistically significant, 0.21 days 95%CI (-2.6 to 3.0). Intubation rates were observed to be higher in the control 49/76 (69.7%) as compared to the intervention cohort 24/37 (63.9%), a difference of 5.8% or a 23% reduction in MV in the group receiving oxygen hoods, these reductions were not found to be statistically significant, OR (95%CI) 0.77 (0.33 -1.8). A logistic regression analysis was conducted to investigate the effects of prognosticators reported by the CDC and other sources that may be predictive of outcomes in COVID-19 patients 3 14 19 . The likelihood that predictor covariates effect rates of mechanical Ventilation and/or mortality rates were investigated. The intervention oxygen hoods, along with the covariates age, BMI, gender, D-Dimer level, lymphocyte count, chronic lung disease, cardiovascular disease, chronic kidney disease, immunosuppression, and DM were investigated 3 14 . Patients who expired but whose status was DNI/DNR were reassigned as having been mechanically ventilated when investigating predictors of mechanical ventilation for both cohorts. This was done since the patients failing conventional oxygen delivery systems would have been placed on mechanical ventilation had their status not been DNI/DNR. None of the covariates were identified as predictor variables for mechanical ventilation or mortality in the model except age. Age was identified when using logistic regression with Forward stepwise Likelihood Ratio. Backward stepwise regression failed to identify any of the variables including age as predictors of mortality or mechanical ventilation. Age remained a predictor of mechanical ventilation and mortality in subgroup analysis when patients receiving convalescent plasma antibodies were excluded. The logistic regression model (Model 1) predicting mechanical ventilation was found to be statistically significant χ 2 = 6.618, p = 0.010. The model explained 8.8% (Nagelkerke R 2 ) of the variance in mechanical ventilation and correctly classified 60/99 (60.6%) of cases. 37 cases included missing data due to nonreporting of one or more of the covariates other than the predictor variable hood status or the outcome variable of interest mechanical ventilation. The model predicted an OR = 1.036 or a 3.6% increase in risk of mechanical ventilation for every one-year of age within the studied age range. See Table 4c . The logistic regression model (Model 2) predicting mortality was found to be statistically significant χ 2 = 10.623, p = 0.001. The model explained 13.8% (Nagelkerke R 2 ) of the variance in mortality and correctly classified 62/99 (62.6%) of cases. 37 cases were missing data due to nonreporting of one or more of the covariates other than the predictor variable hood status or the outcome variable of interest mortality. The model predicts an OR based on age of 1.047 or a 4.7% increase in risk of mortality for every year within the studied age range, see Table 4d . The use of hyperbaric oxygen hoods prevented immediate/imminent intubation and mechanical ventilation (MV) in the short-term in 58 out of the 58 COVID-19 patients receiving oxygen hoods with a mean improvement in oxygen saturation of 8.8%, 95%CI (7.8 to 9.8). This intervention affords healthcare practitioners an effective alternative high-flow oxygen delivery system in these hypoxic COVID-19 patients who are failing conventional high-flow oxygen delivery systems. Intubation and mechanical ventilation of hypoxic COVID-19 patients has been associated with high mortality, approaching 76.4 -97.2% 3 . Mortality associated with intubation/MV in the total sample was consistent with reported estimates and was observed to be 75%, OR 95% CI 2.07 (0.98 to 4.34), which was not found to be statistically significant. Though the confidence interval nearly reached the threshold of significance. Oxygen hoods may afford additional time for existing life-saving measures and evolving medical interventions to work and for more resources to become available, potentially delaying or foregoing the need for intubation and mechanical ventilation. Effect size reductions for in-hospital mechanical ventilation and mortality rates with oxygen hoods were consistently observed. The effects were relatively small 7.7% and 7.1% respectively but were not found to be statistically significant. These reductions were also observed in subgroup analyses. Subgroup analysis included reassignment of DNI/DNR patients who expired to the intubation/MV outcome and controlling for patients receiving convalescent plasma antibodies. The reductions in mechanical ventilation and mortality were observed favoring the use of the oxygen hood but were not found to be statistically significant. Mechanical ventilation and mortality rate reductions have been observed by other researchers to be significant with the use of similar oxygen hoods/helmets in patients with ARDS 7 . During a pandemic small effect sizes may impact large numbers of patients, if the reductions are true effects. The evaluation for differential levels of comorbidities and potential confounders, including the use of logistic regression analysis to predict intubation/mechanical ventilation and mortality, did not demonstrate statistically significant differences between the intervention and control cohorts with the exception of DM, CKD. Age was observed in the logistic regression model to be predictive of mortality in COVID-19 patients consistent with similar findings by other reseachers 6 . The analysis did not support other covariates as predictors of mechanical ventilation or mortality rates. Reasons that could have explained not reaching the threshold for significance despite consistently observed reductions for in-hospital mechanical ventilation and mortality rates include the following. Despite the observed balance in predictor covariates for both cohorts; Covid-19 respiratory disease severity was not considered balanced between the intervention and control cohorts and is a limitation in this study. Efforts were made to choose a comparable control group, using the requirement for conventional high-flow oxygen delivery systems as a marker of respiratory disease severity. However, except for the control patients who had already failed on these systems and were subsequently intubated, there were control patients in this group who did not fail on these conventional high-flow oxygen delivery systems remaining on them throughout hospitalization and who never required mechanical ventilation. This distinction is important as in contrast every patient receiving the oxygen hoods was failing on these same conventional high-flow oxygen delivery systems. This critical distinction supports the assertion that the intervention cohort patients demonstrated a higher level of critical illness and severity of respiratory failure as compared to the control cohort; yet trends demonstrating observed reductions of intubation and mortality rates were consistently lower in the intervention as compared to the control cohort despite this limitation. The reductions in intubations and mortality may have risen to the threshold level of significance had the control group been balanced in respiratory disease severity. However, this would have decreased the size of the control cohort in the sample. A larger sample size would have improved power and reduced the likelihood of a type-2 error. The national emergency created by the pandemic made it difficult to gauge exactly how many patients would be available to study. The gravity of the COVID-19 pandemic and the positive results in improved patient oxygenation warranted reporting these findings. Lower effect sizes than those used for our sample size calculation could be considered clinically significant in the absence of a pandemic. The oxygen hood is a novel alternative high-flow oxygen delivery system to use with hypoxic COVID-19 patients presenting with impending respiratory failure, however nursing and respiratory staff were initially unfamiliar with its use. Despite oxygen hood training, equipment inexperience could have led to premature discontinuation of the hood in favor of a returning prematurely to conventional high-flow O 2delivery systems, or an incorrect presumption that the oxygen hoods were titratable like conventional high-flow oxygen delivery systems when they are not. These factors could have led to suboptimal use of oxygen hood systems in COVID-19 patients further confounding the study. The implementation of the oxygen hood demonstrated significant improvement in O 2 -saturation above that of conventional high-flow oxygen delivery systems and prevented immediate/imminent intubation and mechanical ventilation in the short-term; potentially allowing additional time for other treatments to work. The observed intubation and mortality effect sizes may have been larger between the intervention and control cohorts if not for the limitations discussed. Larger effect sizes could have reached the level of significance with the sample size used in the study reducing the probability of type 2 error. Longer-term reductions in the mechanical ventilation and mortality rates with use of similar oxygen hoods among ARDS patients has been observed by other researchers and found to be significant, this could be similar for hypoxic COVID-19 patients 7 . The oxygen hood is a safe, reliable, and an effective form of oxygen delivery which may reduce mechanical ventilation and mortality rates. Their use should be considered in treating COVID-19 patients and possibly other patients with hypoxia failing on conventional high-flow oxygen delivery systems to prevent mechanical ventilation. *Independent t-test used for analysis ** Subsample of hooded patients used for portion of study which excluded 10 existing hospitalized patients whose final outcomes on intubation and mortality were available at the end of the study period. *** number of subjects with the comorbidity over the total number of subjects in the sample where information on the comorbidity was available in the record. ****All percentages are given as percent of the total column sample irrespective of missing data. ***** Events < 5 in at least one cells. Ho: There is no difference in the MV rates, and/or mortality rates, and/or hospital length of stay in COVID-19 patient receiving oxygen hoods who failed on conventional high-flow oxygen delivery systems; as compared to those maintained on high-flow oxygen delivery systems OR failing on high-flow oxygen delivery systems AND requiring mechanical ventilation, under normobaric conditions. Ha: There is a difference in the intubation rates, and/or mortality rates, and/or hospital length of stay in COVID-19 patient receiving oxygen hoods who failed on conventional high-flow oxygen delivery systems; as compared to those maintained on highflow oxygen delivery systems OR failing on high-flow oxygen delivery systems AND requiring mechanical ventilation, under normobaric conditions. J o u r n a l P r e -p r o o f Excluded patients were either stable on room air or on low-flow oxygen delivery systems (Low flow nasal cannula, simple masks, venturi masks) throughout hospitalization and discharge. J o u r n a l P r e -p r o o f Rolling updates on Coronavirus disease (COVID -19) -Coronavirus disease (COVID-19) Situation Report.: World Health Organization Recommended Preparedness and Response Activities in Healthcare. Supplement C: Preparedness and Response In Healthcare Facilities. Public Health Guidance for Community-Level Preparedness and Response to Severe Acute Respiratory Syndrome (SARS) Version 2/3 Presenting Characteristics, Comorbidities, and Outcomes Among 5700 Patients Hospitalized With COVID-19 in the Severe Covid-19 Amid Ongoing COVID-19 Pandemic, Governor Cuomo Announces State Department of Health Has Approved New Protocol to Allow BiPAP Machines to be Converted Into Ventilators Coronavirus disease 2019 in elderly patients: Characteristics and prognostic factors based on 4-week follow-up Effect of Noninvasive Ventilation Delivered by Helmet vs Face Mask on the Rate of Endotracheal Intubation in Patients With Acute Respiratory Distress Syndrome. A Randomized Clinical Trial One Year Outcomes in Patients with Acute Respiratory Distress Syndrome Enrolled in a Randomized Clinical Trial of Helmet versus Facemask Noninvasive Ventilation Oxygen Administration: What Is the Best Choice Department of Health and Human Services -Food and drug Administration -Concwence of CDRH OoDEO. Device Evaluation and Approval Local production could solve shortages of essential pandemic-fighting equipment ECHM position on Hyperbaric Oxygen Therapy (HBOT) in multiplace hyperbaric chambers during coronavirus disease (COVID-19) outbreak COVID-19: Oxygen Escalation Therapy and Noninvasive Ventilation Evaluating and Testing Persons for Coronavirus Disease 2019 (COVID-19) Interim Clinical Guidance for Management of Patients with Confirmed Coronavirus Disease (COVID-19) High-flow nasal oxygen therapy A comparison of facemask and respirator filtration test methods Management of COVID-19 Respiratory Distress IBM SPSS Statistics for Windows All authors have completed and submitted Disclosure of Potential Conflicts of Interest and none were reported.Funding/Support: The oxygen hoods were purchased using funds available through a generous donation to the Division of Undersea and Hyperbaric Medicine from the Melvin family. No other grants or support were received.Role of the Funder/Sponsor: The family had no role in the design and conduct of the study; collection, management, analysis, and interpretation of the data; preparation, review, or approval of the manuscript; and decision to submit the manuscript for publication.Additional Acknowledgements: We thank the Phelps Hospital Northwell Health Departments and staff of: Respiratory Therapy, Nursing, Family Medicine, Hospitalist Service, Emergency Medicine, Critical Care/Medicine, Anesthesia, and Hospital Administration for their support during this study. Institutional administrative approval was requested and supported in lieu of formal IRB approval under the emergency conditions that existed during the COVID-19 pandemic.Trial Registration: ClinicalTrials.gov Identifier: NCT 04407260 https://clinicaltrials.gov/ct2/results?cond=&term=04407260&cntry=&state=&city=&dist=