key: cord-0809283-96f6d8i6 authors: Andreani, Julien; Lupo, Julien; Germi, Raphaële; Laugier, Christel; Roccon, Marie; Larrat, Sylvie; Morand, Patrice; Nemoz, Benjamin title: Evaluation of six commercial SARS-CoV-2 rapid antigen tests in nasopharyngeal swabs: better knowledge for better patient management? date: 2021-08-20 journal: J Clin Virol DOI: 10.1016/j.jcv.2021.104947 sha: 8aa9cd6d8e2d2307879141c707adde6956a73f05 doc_id: 809283 cord_uid: 96f6d8i6 Robust antigen point-of-care SARS-CoV-2 tests have been proposed as an efficient tool to address the COVID-19 pandemic. This requirement was raised after acknowledging the constraints that are brought by molecular biology. However, worldwide markets have been flooded with cheap and potentially underperforming lateral flow assays. Herein we retrospectively compared the overall performance of five qualitative rapid antigen SARS-CoV-2 assays and one quantitative automated test on 239 clinical swabs. While the overall sensitivity and specificity are relatively similar for all tests, concordance with molecular based methods varies, ranging from 75,7% to 83,3% among evaluated tests. Sensitivity is greatly improved when considering patients with higher viral excretion (Ct≤33), proving that antigen tests accurately distinguish infectious patients from viral shedding. These results should be taken into consideration by clinicians involved in patient triage and management, as well as by national authorities in public health strategies and for mass campaign approaches. Testing for SARS-CoV-2 infection is a critical tool to control COVID-19 spread, as it enables infected individuals to (self-)isolate and their contacts to quarantine [1] . The gold standard of COVID-19 diagnosis are molecular biology based techniques such as RT-PCR [2] . This method has several limitations, such as delayed results availability, the need for specialized laboratory equipment, as well as experienced technicians. As a result, appropriate management of positive cases can be delayed. Rapid antigen tests have been proposed as a complementary strategy to improve the capacities of SARS-CoV-2 infection diagnosis. However rapid antigen tests have lower sensitivity in comparison with RT-PCR and the potential benefit of their implementation can easily be jeopardized if they give false negative results [3] [4] [5] . Their clinical performances depend on different factors such as clinical setting use, local epidemiology and testing organizations. Several studies have shown a correlation between high cycle threshold (Ct) values or low viral load obtained by RT-PCR and a lower risk of contagiousness, mitigating the importance of solely working with highly sensitive tests. Rapid antigen tests are a diagnostic tool to perform widespread, repetitive and frequent testing that is crucial to harness COVID-19 [6] . Improved knowledge of the analytical performance of rapid antigen tests is critical to better define their use strategy in a public health emergency setting. This retrospective study assessed the performance of six antigen tests compared to RT-PCR in a cohort of outpatients. We conducted a retrospective study to determine the diagnostic performance of five rapid antigen tests and one automated quantitative antigen assay compared to RT-PCR. All nasopharyngeal swabs were collected by trained healthcare professionals from outpatients attending a specific SARS-CoV-2 diagnostic facility. Swabs were transferred into a Phosphate buffered saline (PBS) medium (Vacuette® Greiner bio-one) and were divided in 2 aliquots of 1.5 mL immediately after reception at the laboratory. One was used for the routine RT-PCR assay and the second made available for antigen tests evaluation and stored for a few hours refrigerated (+4°C). According to the RT-PCR result, positive and negative swabs for SARS-CoV-2 RNA detection were selected to obtain a ratio close to 1 negative :1 positive, and then analysed by all the antigen tests. Delays between nasopharyngeal sample collection and rapid antigen tests never exceeded 12 hours. Twenty-six positive swabs for respiratory pathogens other than SARS-CoV-2 (Supplementary table S1) were also retrospectively investigated to assess potential false positives. These samples have been stored at -80°C for a maximum period of 12 months. All patients were informed and did not oppose to participate in this protocol. Clinical data (age, sex, symptoms and onset of symptoms) were collected and stored according to local ethic procedures. Study protocol has been reviewed and approved by local ethic committee. Two different methods were used to detect SARS-CoV-2 RNA genome in nasopharyngeal swabs according to routine lab organization. For one, nucleic acid was extracted from nasopharyngeal samples using the eMag automated system (BioMérieux, Marcy-l'Etoile, France). Amplification of SARS-CoV-2 RNA was performed using the RT-PCR assay developed by the French National Reference Centre [7] on a LightCycler480 II platform (Roche Diagnostics, Mannheim, Germany). The second method was the Alinity m SARS-CoV-2 assay (Abbott Illinois, Chicago, USA). The molecular diagnosis of respiratory viruses other than SARS-CoV-2 has been performed using multiplexed RT-PCR assays (BioFire TM FilmArray, BioMérieux; RespiFinder® 2Smart, PathoFinder; RealStar® Flu & VRS, Altona Diagnostics). Estimation of SARS-CoV-2 viral loads were obtained using the results of external quality controls acquired from Quality Controls for Molecular Diagnostics (QCMD, Glasgow, Scotland) [8] . Ct values were normalized against the French national reference RT-PCR. A Ct value of 32 corresponds to a viral load of 2.99 log 10 RNA copies/mL (S.D ± 0.0707). Establishing a linear regression between Ct values (y-axis) and RNA viral load expressed in log copies/mL (x-axis), we found the following equation y = -2.66 x + 39.94 (r 2 > 0.98) (data not shown). Samples were analysed using five different rapid antigen tests and one automated quantitative antigen test, according to the manufacturers' recommendations. The characteristics of each test are shown in Table 1 . Rapid antigen test results were assessed by trained staff after 10 to 30 min, according to manufacturer's instructions. Samples that yielded uninterpretable results were immediately retested in order to obtain either a positive or negative result. Only four Lumipulse assays failed due to an insufficient volume used as input in the automate. Contrarily to others tests based on immunochromatographic methods, Lumipulse ® G SARS-CoV-2 Ag kit (Fujirebio ® , Japan) is a semi-quantitative and automated assay. For this assay, the positive threshold was initially set at 1.34 pg/mL, according to the manufacturer's recommendations. Sensitivity, specificity and concordance of rapid antigen tests were calculated using RT-PCR as reference method and following: Based on the recommendations of the French society of microbiology and previous reports, the Ct of 33 was considered of interest to evaluate the contagiousness of SARS-CoV-2 infected patients [9] [10] [11] [12] [13] [14] [15] [16] . Modelling analyses were performed on GraphPad V5 (Prism, La Jolla, USA) using second order polynomial representation. The sensitivities of the different tests were compared using a Chi-square test or measuring the difference between two proportions (Z-test), when appropriate. A total of 239 samples previously tested by RT-PCR were analysed with the 6 antigen tests. Among these patients, 40.2 % (96 on 239) were healthcare professionals. The study population included 51.9% (124 out of 239) of patients with a positive result according to RT-PCR assay and a vast majority of symptomatic patients (64.4%) ( Table 2 ). Among the 35 asymptomatic patients, 30 patients had a negative result for RT-PCR testing and five had positive results. Among the results of 1195 qualitative rapid antigenic tests, 388 were positive (32.5%) and 807 were negative (67.5%). Among them, five hundred seventy negative tests with rapid antigen assays were also negative for RT-PCR. The distribution of antigen test results according to the Ct results of RT-PCR is shown at Figure 1A and 1B. Only one assay yielded a false positive result (SD biosensor® one). The incriminated sample has been controlled a second time by RT-PCR and was confirmed to be a true negative. Interestingly, this sample was positive for human metapneumovirus. Taken together, the five qualitative antigen tests results showed a homogeneity in performance (p = 0.12) (Table 3) . However, using a pair-wise comparison analysis, we observed a lower sensitivity for Certest (53.2%) in comparison with Orient gene (67.7%) and Espline (66.9%) tests (p<0.03 and p<0.03). When considering samples with Ct ≤ 33, the sensitivities of the five qualitative assays were higher, ranging from 61.9% to 78.1%. The quantitative Lumipulse assay showed higher sensitivity than all other qualitative assays (p< 0.001). Finally, the overall concordance with molecular biology results were at 81.3%, 75.7%, 79.5%, 80.8%, 82.8%, and 83.3% for the Lumipulse®, Certest®, Panbio™, SD biosensor, Espline® and the Orient gene tests respectively. We analysed the sensitivity according to the presence of symptoms, and the delay between sample collection and the onset of symptoms. The sensitivities of each test were higher in symptomatic patients group (n=154, median Ct values 21.5) when compared to the group combining asymptomatic patients and patients with unknown status (n=85, median Ct values 29.5) (p<0.01) ( Figure 2A ). Analysing all of the 1434 results, the sensitivities appeared higher for the specimens collected closely from the onset of symptoms. However, this result was only significant for the Certest (p<0.05, when comparing groups D0/D1 versus D5/D6/D7) and this could be attributed to its lower sensitivity as described above ( Figure 2B , Table 2 ). For the PanBio assay, this trend did not reach statistical significance. Focusing on the quantitative Lumipulse assay, and considering the recommended threshold of 1.34 pg/mL, the overall sensitivity, the specificity, and the concordance were 87.6% CI95%: 81.7-93.5, 72.7% CI95%: 63.4-82.0, and 81.3%, respectively. When considering positive samples with Ct ≤ 33, the sensitivity of Lumipulse increased to 98.0% CI95%: 95.3-100, specificity to 72.7% CI95%: 63.4-82.0, and concordance to 86.3%. Using a ROC curve analysis and a dot plot representation ( figure 3A , B), we showed that a threshold of 6.55 pg/mL could improve the specificity of this assay yielding values at 95.5% CI95%: 91.1-99.8 but the sensitivity decreased at 81% CI95%:74.0-88.0. However, the concordance was higher at 87.1%. Considering samples with Ct ≤ 33 with this new threshold, the concordance with RT-PCR results also improved (94.2%). It is to be noted that the time-to-result for the Lumipulse assay is longer than for rapid antigen tests, but similar to those of POC RT-PCR tests (≈ 30 minutes). The specificity and sensitivity were 99.0% and 69.4% respectively when choosing a threshold of 87.7 pg/mL. Antigen tests are an easy and rapid tool to perform diagnosis of SARS-CoV-2 infection. In the present study, we evaluated six commercially available antigen tests and compared these to RT-PCR results obtained for nasopharyngeal specimens from 239 outpatients. Our main finding yielded sensitivities ranging from 53.2% to 67.7% for qualitative tests and up to 87.6% for quantitative test. Our work deliberately included similar proportions of positive and negative nasopharyngeal samples tested by RT-PCR. We found values of sensitivity ranging from 53.2% to 87.6% comparable to previous studies including samples with the same heterogeneity of RNA viral loads. [17] [18] [19] [20] [21] . As expected, when considering stronger viral excretions (samples with Ct ≤ 33 in the RT-PCR result), the sensitivity improved for all assays with values ranging from 70% to 98%. In a recent study, the FujireBio® LumiPulse assay showed a sensitivity of 92.0% and a concordance of 99.7%, slightly higher than the results obtained here. However, the frequency of included positive cases was lower (4%) than ours and the results between 1 and 10 pg/mL were considered as inconclusive [22] . Concerns about false positive results of the Lumipulse assay have been reported [23] , also we did not unveil such cases. Recently, the Panbio™ test from Abbott® was used in more than 1300 patients in The Netherlands and showed [18] a very high specificity (100%), thus confirming the specificity obtained in our study. In another study, the comparison of the Espline® test with other antigen tests, molecular biology or viral culture isolation concluded to misdiagnosis in about half positive specimens [3] . Nevertheless, this study was carried out on only 80 clinical specimens of various origin. A recent study compared 7 commercial tests in 238 clinical samples (2 of them were also used herein, the Panbio® and the SD biosensor assays), and the authors described a high specificity and good sensitivity [5] , in accordance with our findings. The question of how important the sensitivity of the assays is, is still debated [24, 25] . The relatively poor sensitivity of antigen tests compared to RT-PCR can be overcome with a widespread and frequent testing strategy [26, 27] . For highly contagious patients, i.e. those with high RNA viral load (higher than 3.8 log 10 copies/mL, Ct number below 25), the sensitivity of antigen tests exceeded 93% limiting the risk of misdiagnosis and mismanagement in the SARS-CoV-2 pandemic. The Lumipulse assay delivered a quantification of antigen amount present in samples. This quantification showed a fairly good agreement with RT-PCR Ct numbers (R 2 =0.767, data not shown), thus offering a viable alternative to RT-PCR in settings where the latter is not available. However the cut off should be set higher to improve the specificity and the positive predictive value in a context of random massive screening tests where the prevalence of SARS-CoV-2 infection may be low. When setting a higher threshold to obtain a specificity of 99%, the sensitivity of this assay was comparable to others lateral flow assays. In our evaluation, we observed only one false positive result with the SD biosensor assay. However, we could not determine the cause of this false positive test (cross reaction with human metapneumovirus versus other non-specific reaction). In our investigation, which included 17 samples with seasonal human coronaviruses, we did not to find any misdetection, in particular with the closely-related betacoronavirus HKU1. This result was similar to the study of Corman et al. [5] , besides they only tested two samples for human coronaviruses. One of the advantages of qualitative antigen tests is their ease of use and we closely monitored the practical aspects of these assays. For Panbio™ test, we noted practical difficulties (leaks during liquid handling) raising a threat of cross contamination between tests, especially when performed in batches. A limitation of our study stems from the retrospective selection of our population with an artificial overrepresentation of positive cases in RT-PCR. However, this approach enabled us to obtain a diversity of samples in term of viral RNA loads. As reported in other studies , the delay between the collection and the analysis of the swab stored refrigerated in PBS for a maximum of 12 hours could affect the results of sensitivity [5, 18, 21] . In the future, with the need for point-of-care tests to quickly control epidemics due to emergent pathogens [28] , clinicians should be aware of the performance of novel commercialized assays to guide them in their practices, and for the management of infected patients. Our result highlight that rapid antigen tests have a better sensitivity for the detection of symptomatic cases than asymptomatic cases. However, considering the kinetic of RNA viral load and the diagnosis window, tests should be repeated on different days. Finally, mass rapid antigen test campaigns are one of the available epidemic control measures and they have shown incredible impact (i.e. Slovakia [27] ), proving this strategy is efficient. Using better tests among those available will undoubtedly yield better outcomes for populations. % is expressed on the total population of the cohort (n = 239). In figure 2A , 85 swabs from asymptomatic patients or for which status was unknown were included for each assay. One hundred fifty-four symptomatic cases were included for all tests. For the figure 1B , details for each subgroup are indicated in table 2. Declaration of interests ☒ The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper. ☐The authors declare the following financial interests/personal relationships which may be considered as potential competing interests: Transmission of COVID-19 in 282 clusters in Catalonia, Spain: a cohort study To Interpret the SARS-CoV-2 Test, Consider the Cycle Threshold Value Comparison of Rapid Antigen Tests for COVID-19 Analytical Sensitivity of the Abbott BinaxNOW COVID-19 Ag CARD Comparison of seven commercial SARS-CoV-2 rapid point-of-care antigen tests: a singlecentre laboratory evaluation study Test sensitivity is secondary to frequency and turnaround time for COVID-19 screening Clinical and virological data of the first cases of COVID-19 in Europe: a case series Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR Viral RNA load as determined by cell culture as a management tool for discharge of SARS-CoV-2 patients from infectious disease wards SARS-CoV-2 Viral Load in Upper Respiratory Specimens of Infected Patients Viral load dynamics and disease severity in patients infected with SARS-CoV-2 in Zhejiang province, China Quantitative Detection and Viral Load Analysis of SARS-CoV-2 in Infected Patients Interpreting Diagnostic Tests for SARS-CoV-2 Viral load of SARS-CoV-2 in clinical samples Clinical and laboratory characteristics of symptomatic healthcare workers with suspected COVID-19: a prospective cohort study Virological assessment of hospitalized patients with COVID-2019 Performance of a Point-of-Care Test for the Rapid Detection of SARS-CoV-2 Antigen, Microorganisms Real-life validation of the Panbio™ COVID-19 antigen rapid test (Abbott) in communitydwelling subjects with symptoms of potential SARS-CoV-2 infection Field performance and public health response using the BinaxNOW TM Rapid SARS-CoV-2 antigen detection assay during community-based testing Clinical Evaluation of Self-Collected Saliva by Quantitative Reverse Transcription-PCR (RT-qPCR), Direct RT-qPCR, Reverse Transcription-Loop-Mediated Isothermal Amplification, and a Rapid Antigen Test To Diagnose COVID-19 Evaluation of a rapid antigen test to detect SARS-CoV-2 infection and identify potentially infectious individuals Prospective study of 1308 nasopharyngeal swabs from 1033 patients using the LUMIPULSE SARS-CoV-2 antigen test: Comparison with RT-qPCR Another falsepositive problem for a SARS-CoV-2 antigen test in Japan Rethinking Covid-19 Test Sensitivity -A Strategy for Containment Fast coronavirus tests: what they can and can't do Impact of mass testing during an epidemic rebound of SARS-CoV-2: a modelling study using the example of France The impact of population-wide rapid antigen testing on SARS-CoV-2 prevalence in Slovakia Current and Future Point-of-Care Tests for Emerging and New Respiratory Viruses and Future Perspectives We thank Eric Maniez for his kind help and proofreading regarding with English corrections.Certest, Espline and Lumipulse tests have been kindly provided by manufacturers. Companies had no role in study design, data collection and interpretation or in the decision to submit this work. The authors have not received specific funding for this study. The authors have no conflict of interest to declare.