key: cord-0806395-55yltan1 authors: Pham, Jimmykim; Meyer, Sarah; Nguyen, Catherine; Williams, Analee; Hunsicker, Melissa; McHardy, Ian; Gendlina, Inessa; Goldstein, D. Yitzchak; Fox, Amy S.; Hudson, Angela; Darby, Paul; Hovey, Paul; Morales, Jose; Mitchell, James; Harrington, Karen; Majlessi, Mehrdad; Moberly, Joshua; Shah, Ankur; Worlock, Andrew; Walcher, Marion; Eaton, Barbara; Getman, Damon; Clark, Craig title: Performance Characteristics of a High-Throughput Automated Transcription-Mediated Amplification Test for SARS-CoV-2 Detection date: 2020-09-22 journal: J Clin Microbiol DOI: 10.1128/jcm.01669-20 sha: 172464d068ac3bcc06fdf25035bfa0de313e9f03 doc_id: 806395 cord_uid: 55yltan1 The COVID-19 pandemic caused by the new SARS-CoV-2 coronavirus has imposed severe challenges on laboratories in their effort to achieve sufficient diagnostic testing capability for identifying infected individuals. In this study, we report the analytical and clinical performance characteristics of a new, high-throughput, fully automated nucleic acid amplification test system for the detection of SARS-CoV-2. The assay utilizes target capture, transcription-mediated amplification, and acridinium ester-labeled probe chemistry on the automated Panther system to directly amplify and detect two separate target sequences in the open reading frame 1ab (ORF1ab) region of the SARS-CoV-2 RNA genome. The probit 95% limit of detection of the assay was determined to be 0.004 50% tissue culture infective dose (TCID(50))/ml using inactivated virus and 25 copies/ml (c/ml) using synthetic in vitro transcript RNA targets. Analytical sensitivity (100% detection) was confirmed to be 83 to 194 c/ml using three commercially available SARS-CoV-2 nucleic acid controls. No cross-reactivity or interference was observed with testing of six related human coronaviruses, as well as 24 other viral, fungal, and bacterial pathogens, at high titers. Clinical nasopharyngeal swab specimen testing (n = 140) showed 100%, 98.7%, and 99.3% positive, negative, and overall agreement, respectively, with a validated reverse transcription-PCR nucleic acid amplification test (NAAT) for SARS-CoV-2 RNA. These results provide validation evidence for a sensitive and specific method for pandemic-scale automated molecular diagnostic testing for SARS-CoV-2. To meet the need for pandemic-scale diagnostic testing, we developed and validated a high-throughput, fully automated nucleic acid amplification test (NAAT) for direct amplification and detection of SARS-CoV-2 RNA from specimens of infected individuals. The assay employs target capture, transcription-mediated amplification (TMA), and acridinium ester-labeled probe chemistries to enable a sample-to-result solution for detection of two different conserved target regions within the open reading frame 1ab (ORF1ab) region of the SARS-CoV-2 genome. Here, we describe the analytical and clinical performance characteristics of the assay. Transcription-mediated amplification test for SARS-CoV-2. The Aptima SARS-CoV-2 assay utilizes magnetic bead-based target capture, isothermal TMA of RNA, and dual kinetic acridinium ester-labeled probe hybridization for the isolation, amplification, and detection of an internal process control RNA and two unique sequences within the ORF1ab region of the SARS-CoV-2 viral genome. The assay is performed on the automated Panther and Panther Fusion instruments (both from Hologic, Inc., San Diego, CA) and received FDA emergency use authorization (EUA) on 14 May 2020. It is intended for the qualitative detection of SARS-CoV-2 RNA isolated and purified from nasopharyngeal (NP) swab, nasal swab (NS), midturbinate and oropharyngeal (OP) swab, NP wash/aspirate, or nasal aspirate specimens obtained from individuals meeting COVID-19 clinical and/or epidemiological criteria. The sample input volume is 0.5 ml with continuous sample and reagent loading access and automated RNA extraction, amplification, detection, and result reporting. The time to first result is 3 h and 30 min, with a capacity of approximately 1,025 results per 24 h per instrument system. Comparison assay. The validated EUA Panther Fusion SARS-CoV-2 reverse transcription-PCR (RT-PCR) assay (Hologic, Inc.) was used as a comparator assay for clinical performance studies. This assay was performed as previously described (6) . Analytical performance. (i) Limit of detection. The analytical sensitivity of the SARS-CoV-2 TMA assay was assessed using 2 lots of reagents to test 60 replicates each of dilution panels containing cultured SARS-CoV-2 virus strain USA-WA1/2020 (BEI Resources, Manassas, VA) and diluted in Aptima specimen transport medium (STM) matrix to a range of 0.03 to 0.0003 50% tissue culture infectious dose (TCID 50 )/ml. Also tested were replicates (n ϭ 43 to 60) of panels consisting of two in vitro transcribed (IVT) RNA targets, corresponding to two unique target sequences within the ORF1ab region of the SARS-CoV-2 RNA genome, diluted in STM. The assay positivity for both studies was determined using a cutoff value of 560 kilo relative light units (kRLU) according to the manufacturer's instructions for use. Results were analyzed by probit analysis (normal model) to determine the 95% limit of detection (LOD). Analytical sensitivity was confirmed by diluting stock SARS-CoV-2 virus in STM into four specimen matrices (pooled NP swab specimens, STM, saline, and liquid Amies transport medium (Copan, Murrieta, CA) at 0.003 TCID 50 /ml for NP swab, STM, and saline and 0.003 and 0.01 TCID 50 /ml for liquid Amies (n ϭ 20 replicates for each determination). (ii) Analytical specificity/interference. Analytical specificity of the SARS-CoV-2 TMA assay was determined by evaluating assay cross-reactivity and interference using 30 nontarget microorganisms (17 viral species, including 6 non-SARS-CoV-2 coronaviruses, 11 bacterial species, and 2 fungal species; n ϭ 3 replicates each) at the highest titer achievable. Thirty NP swab specimens obtained from asymptomatic donors (with consent) were also tested to represent diverse microbial flora in the human respiratory tract. Interference by high-titer nontarget organisms was assessed by testing 3 replicates of each organism in the presence of low-titer (0.03 TCID 50 /ml) SARS-CoV-2 virus. Cross-reactivity by high-titer nontarget organisms was assessed in the absence of SARS-CoV-2 virus. -ERC0831042) were diluted in stock STM to 6 concentrations ranging from 8 to 833 copies per ml (c/ml), and multiple replicates (n ϭ 20 to 40) were tested with the SARS-CoV-2 TMA assay. The control panels were also tested with the SARS-CoV-2 RT-PCR assay on the automated Panther Fusion platform. Specimen collection and clinical performance. Residual deidentified NP swab samples were collected using standard methods from 140 symptomatic patients at two U.S. clinical sites (Salt Lake City, UT, and Albuquerque, NM). Specimens were transported to the laboratory and tested with the SARS-CoV-2 TMA assay and the Fusion SARS CoV-2 RT-PCR assay. An additional clinical sample set consisting of paired NP swab, OP swab, and NS specimens were collected from 38 patients; complete sets (containing all three specimen types) were obtained from 35 patients. NP swabs were collected using either BD universal viral transport nasopharyngeal swab and universal viral transport medium (VTM) (Becton, Dickinson, San Diego, CA) or Copan Minitip flocked swab and VTM. NS samples were collected first by inserting the swab into the subject's nostril past the inferior turbinate, approximately 3 cm, twisting the swab in the midturbinate area for 3 to 5 s, and placing the swab into a tube of STM. OP swab samples were collected immediately following NS samples by swabbing the posterior pharynx for 3 to 5 s and placing the swab into a specimen tube containing STM. Samples were frozen and shipped to Hologic for testing. Data for the analytical sensitivity determination of the SARS-CoV-2 TMA assay are shown in Fig. 1 . Using a predetermined cutoff value of 560 kRLU, the assay yielded 100% positivity at a concentration of 0.01 TCID 50 /ml of SARS-CoV-2 virus and at 100 c/ml of SARS-CoV-2 IVT RNA targets. Using probit analysis, the 95% limit of detection was determined to be 0.004 TCID 50 /ml (95% confidence interval [CI], 0.003 to 0.007) for SARS-CoV-2 virus, and 25.4 c/ml (95% CI, 16.9 to 50.5) for ORF1ab IVT RNA targets. Analytical sensitivity for the assay was confirmed by testing SARS-CoV-2 virus in 4 specimen matrices (pooled NP swab specimens, STM, saline, and liquid Amies transport medium) at 0.003 TCID 50 /ml. All specimen matrices yielded 95% positivity (19/20 replicates detected) at this concentration, except the liquid Amies transport medium, which was 85% (17/20) positive at 0.003 TCID 50 /ml and 100% (20/20) positive at 0.01 TCID 50 /ml ( Table 1) . The analytical specificity of the SARS-CoV-2 TMA assay was determined by evaluating assay cross-reactivity and interference using 30 nontarget viral, bacterial, and fungal microorganisms at the highest titer achievable, as well as in 30 NP swab specimens obtained with consent from asymptomatic donors at low risk of SARS-CoV-2 infection. As shown in Table S1 in the supplemental material, none of the microorganisms or NP swab specimens tested caused cross-reactivity in the absence of SARS-CoV-2 target or interfered with TMA detection in the presence of SARS-CoV-2 spiked at 0.03 TCID 50 /ml. The clinical accuracy of the SARS-CoV-2 TMA assay was compared to the SARS-CoV-2 RT-PCR assay using 140 patient NP swab specimens ( Table 2, Table S2 ). This analysis resulted in positive, negative, and overall agreements of 100% (95% CI, 94.3% to 100%), Clinical performance of the SARS-CoV-2 TMA assay was also assessed by testing sets of NP swab, OP swab, and nasal swab specimens co-collected from 35 symptomatic patients suspected of being infected with SARS-CoV-2. Figure 2 shows that the SARS-CoV-2 TMA assay had 100% positive and negative agreements of the NP swab specimens with the co-collected OP swab and nasal swab specimens. Similar results were obtained for the paired specimen sets using the SARS-CoV-2 RT-PCR assay (Fig. S1 ). SARS-CoV-2 control panel materials from three commercial suppliers (Exact Diagnostics, SeraCare, and ZeptoMetrix) were evaluated by building dilution panels of each control material and testing multiple replicates with both the SARS-CoV-2 TMA assay and the SARS-CoV-2 RT-PCR assay (Table 3) . Both assays yielded similar results. For the Exact Diagnostics SARS-CoV-2 control, the TMA and RT-PCR assays each had 100% detection down to 83 c/ml (n ϭ 40 each). For the SeraCare SARS-CoV-2 control, the TMA assay was 100% at 83 c/ml (n ϭ 20), while the RT-PCR assay was 90% at 83 c/ml and 100% at 194 c/ml (n ϭ 20 for both). The SARS-CoV-2 TMA assay (n ϭ 37) and the SARS-CoV-2 RT-PCR assay (n ϭ 40) were both 100% reactive at 194 c/ml using the ZeptoMetrix control material. The availability of large-scale diagnostic testing for SARS-CoV-2 addresses a current critical need for identifying the prevalence and spread of the virus in populations and for guiding public health policies and interventions to minimize incident infections. This study describes the performance characteristics of a new high-throughput isothermal TMA NAAT that employs a complete sample-to-result automation system to maximize sample testing throughput in clinical laboratories. The analytical performance data demonstrate that the assay is highly sensitive and specific for detection of viral SARS-CoV-2 RNA and has high clinical agreement (100% positive agreement, 98.7% negative agreement) with an EUA-validated RT-PCR assay for SARS-CoV-2 RNA. The unprecedented need for SARS-CoV-2 testing has resulted in national shortages in sample collection materials, including VTM, prompting the CDC and FDA to recommend optional specimen collection media, such as saline and liquid Amies (7, 8) . To ensure comparable performance across various collection media, we evaluated the analytical sensitivity of the SARS-CoV-2 TMA assay in NP swab matrix (NP swabs collected in VTM), STM, saline, and liquid Amies transport medium using inactivated SARS-CoV-2 virus. NP swab matrix, Aptima STM, and saline demonstrated 95% detection at 0.003 TCID 50 /ml, whereas liquid Amies showed slightly lower sensitivity at 0.01 TCID 50 /ml. Despite this small difference in analytical sensitivity, these results suggest that all of these media are acceptable for use for clinical sample collection. NP swab specimens have been considered the gold standard sampling method for respiratory virus infection (9) , as it has been reported that nasal swab or OP swab samples may have a slightly lower sensitivity than NP swab samples (10, 11) . However, given the challenges of collection device shortages for SARS-CoV-2 diagnosis, particularly NP swabs and VTM, the use of nasal swab and OP swab as alternate samples for diagnosis of SARS-CoV-2 has been evaluated. The CDC recently removed NP swab as the "preferred" sample type from their sample collection guidelines (7) , and others have reported comparable performance of NS and OP swabs for the diagnosis of SARS-CoV-2 (12) (13) (14) (15) . Our data show strong agreement between SARS-CoV-2 detection from NS and OP swab samples compared to paired NP swab samples for the SARS-CoV-2 TMA and SARS-CoV-2 RT-PCR assays. For one patient, we did observe positive NP and nasal swab results but negative results in the paired OP swab; however, overall, the data indicate that both NS and OP swab specimens are adequate sample types for diagnosis of SARS-CoV-2. The performance of other SARS-CoV-2 NAATs that have received emergency use authorization has been characterized using commercially available inactivated virus preparations (e.g., BEI Resources, Manassas, VA) or synthetic RNA quality control materials. Our evaluation of three different external RNA control materials demonstrated comparable analytical sensitivity with both the SARS-CoV-2 TMA assay and the SARS-CoV-2 RT-PCR assay, with LOD values of between 83 c/ml of 194 c/ml. These analytical sensitivity values correlate with previously reported LOD values by Smith et al. (16) . However, using LOD values as determined by external control material to assess or compare assay performance warrants some caution, as different control materials may give considerably different results for the absolute LOD value (16) , and the reported RNA or DNA stock concentrations of these materials may differ from true concentrations (17) . The limitations of this study include the small number of clinical specimens available for testing and the absence of discordant result resolution by testing with a third assay. The paired specimen testing (NS and OP) included only 35 patients (14 of which were positive). The specimens that were included did span a typical clinical titer range for the virus (SARS-CoV-2 RT-PCR cycle threshold [C T ] values ranged from 14.5 to 37.1). Additional clinical data should be collected to more fully assess and compare performance between these sample types. We also only tested control panel material in STM, not in patient swab specimen matrix. In summary, the SARS-CoV-2 TMA assay is highly sensitive and specific and provides an automated high-throughput testing solution for large-scale diagnostic testing for the virus. The assay system is able to process and generate results for Ͼ1,000 samples per day, enabling medical centers, reference laboratories, and public health laboratories to efficiently process and analyze very high volumes of specimens for the detection of SARS-CoV-2 RNA (18). Supplemental material is available online only. SUPPLEMENTAL FILE 1, PDF file, 0.2 MB. WHO. 2020. Q&A on coronaviruses (COVID-19 COVID-19): symptoms of coronavirus Estimation of the asymptomatic ratio of novel coronavirus infections (COVID-19) Coronavirus disease (COVID-19) dashboard Daily state-by-state testing trends Comparison of four molecular in vitro diagnostic assays for the detection of SARS-CoV-2 in nasopharyngeal specimens Interim guidelines for collecting, handling, and testing clinical specimens for COVID-19 FDA. FAQs on testing for SARS-CoV-2 Current best practices for respiratory virus testing Comparing nose-throat swabs and nasopharyngeal aspirates collected from children with symptoms for respiratory virus identification using real-time polymerase chain reaction Detection of multiple respiratory pathogens during primary respiratory infection: nasal swab versus nasopharyngeal aspirate using real-time polymerase chain reaction Virological assessment of hospitalized patients with COVID-2019 Sensitivity of nasopharyngeal, nasal and throat swab for the detection of SARS-CoV-2 Nasal swab sampling for SARS-CoV-2: a convenient alternative in times of nasopharyngeal swab shortage Self-collection: an appropriate alternative during the SARS-CoV-2 pandemic Analytical and clinical comparison of three nucleic acid amplification tests for SARS-CoV-2 detection Commercial stocks of SARS-CoV-2 RNA may report low concentration values, leading to artificially increased apparent sensitivity of diagnostic assays Comparison of two high-throughput reverse transcription-polymerase chain reaction systems for the detection of severe acute respiratory syndrome coronavirus 2 All of the authors except I.M., I.G., D.Y.G., and A.S.F. are scientists employed by Hologic, Inc., the manufacturer of the diagnostic test systems used in this study. I.M., I.G., D.Y.G., and A.S.F. declare no conflicts of interests.This study was funded by Hologic, Inc.