key: cord-0803291-s1ca3c6n authors: Hasani, Mahdiyeh; Campbell, Tracey; Wu, Fan; Warriner, Keith title: Decontamination of N95 and surgical masks using a treatment based on a continuous gas phase-Advanced Oxidation Process date: 2021-03-18 journal: PLoS One DOI: 10.1371/journal.pone.0248487 sha: 1d4511e9c6aa3c3237c1c1b9ca21d76260287269 doc_id: 803291 cord_uid: s1ca3c6n A gas-phase Advanced Oxidation Process (gAOP) was evaluated for decontaminating N95 and surgical masks. The continuous process was based on the generation of hydroxyl-radicals via the UV-C (254 nm) photo-degradation of hydrogen peroxide and ozone. The decontamination efficacy of the gAOP was dependent on the orientation of the N95 mask passing through the gAOP unit with those positioned horizontally enabling greater exposure to hydroxyl-radicals compared to when arranged vertically. The lethality of gAOP was independent of the applied hydrogen peroxide concentration (2–6% v/v) but was significantly (P<0.05) higher when H(2)O(2) was introduced into the unit at 40 ml/min compared to 20 ml/min. A suitable treatment for N95 masks was identified as 3% v/v hydrogen peroxide delivered into the gAOP reactor at 40 ml/min with continuous introduction of ozone gas and a UV-C dose of 113 mJ/cm(2) (30 s processing time). The treatment supported >6 log CFU decrease in Geobacillus stearothermophilus endospores, > 8 log reduction of human coronavirus 229E, and no detection of Escherichia coli K12 on the interior and exterior of masks. There was no negative effect on the N95 mask fitting or particulate efficacy after 20 passes through the gAOP system. No visual changes or hydrogen peroxide residues were detected (<1 ppm) in gAOP treated masks. The optimized gAOP treatment could also support >6 log CFU reduction of endospores inoculated on the interior or exterior of surgical masks. G. stearothermophilus Apex spore strips could be applied as a biological indicator to verify the performance of gAOP treatment. Also, a chemical indicator based on the oxidative polymerization of pyrrole was found suitable for reporting the generation of hydroxyl-radicals. In conclusion, gAOP is a verifiable treatment that can be applied to decontaminate N95 and surgical masks without any negative effects on functionality. In the CoVid-19 pandemic, caused by the spread of SARS-Cov-2, there was a shortage of personal protective equipment (PPE) such as N95 masks and surgical (procedure) masks [1] [2] [3] . The N95 mask consists of multiple filter layers that can filter 95% of particulates and provides satisfactory respiratory protection against the virus [1, 4] . Although surgical masks have a looser fit it has been reported that the filter layers also provide adequate protection against the virus within moderate-risk environments [4] . Given the shortage of N95 masks, the general use of surgical masks has been widely adopted for general use. Although surgical masks are intended for single use in moderate-risk environments there has been little attention on how to extend working life or recycle. This becomes relevant considering the demand for surgical masks coupled with the need to avoid noncompostable waste reaching landfills [5] . Consequently, there is a demand for mask decontamination methods that can be applied to enable recycling or extended use without an increased risk of contracting or disseminating infectious agents [6, 7] . To this end, criteria have been developed to define the requirements of a suitable mask decontamination treatment that ensure adequate microbial inactivation while not negatively affecting functionality [8] . Specifically, there is a requirement that the process should support 6 log CFU reduction of Geobacillus stearothermophilus endospores and 4 log reduction of coronavirus or suitable surrogate. No chemical residues should be present on masks or change in functionality in fitting and particulate filtration. There have been several successful N95 mask decontamination methods described based on microwave heating, ethylene oxide, hydrogen peroxide vapor, and Ultraviolet Germicidal irradiation (UVGI) [9] [10] [11] . Each method have advantages and disadvantages with regards to application, the effect on mask integrity, and antimicrobial efficacy [12] . For example, thermal treatments exhibit anti-viral activity although mask integrity can be compromised [13] . Similarly, UVGI has been shown to inactivate viruses such as H1N1 without compromising mask functionality [14] . However, the poor penetration of UV-C photons into the filter layers of the N95 masks requires high doses (>270 kJ/ m 2 ) to be applied that can result in 20-51% loss in strap strength [15, 16] . Hydrogen peroxide vapor is the most widely used N95 mask decontamination method and has been demonstrated to support endospore reduction with negligible effects on mask functionality [17] . However, the process is a batch system that requires the application of >480 ppm hydrogen peroxide over an extended period resulting in a process that takes around 3 h. A surgical mask decontamination method based on applying dry heat (60-70˚C) for 1 has been reported [18] . Here, the researchers inoculated surgical masks with different bacteria (Escherichia coli, Staphylococcus aureus, Pseudomonas aeruginosa, Klebsiella pneumonia, Acintobacter spp, Corybacterium spp) yeast (Candida albicans) or virus (H1N1). The inoculation level per mask was 3 log CFU with the virus loading being estimated at 320 virons. No survivors were recovered from masks heated for 1 h at either 60 or 70˚C although filtration efficacy was lost after 2-3 cycles [18] . However, the inactivation of bacterial endospores was not considered in the study. A further study compared the decontamination efficacy of dry heat (90˚C for 70 mins) and microwave generated steam to inactivate S. aureus on surgical masks [19] . Although both processes supported a >6 log reduction of S. aureus the microwave treatment resulted in a loss of surgical mask integrity after a single cycle. By using the dry heat method, the masks could tolerate 3 cycles without loss of function [19] . Again, the researchers did not determine efficacy against endospores given the thermal treatment was below that to support inactivation. In an alternative method, Quaternary ammonium salt surfactant has been applied to masks that supported a 2-3 log reduction of S. aureus although the health implication of chemical residues was not determined [20] . In the following, a process based on gas-phase Advanced Oxidation Process (gAOP) for N95 and surgical mask disinfection was evaluated. The gAOP process is derived from Fentonlike reactions that generate hydroxyl-radicals from the degradation of hydrogen peroxide and/ or ozone [21] . The Advanced Oxidation process has previously been applied for degrading pollutants in water treatment [22] with the gas-phase treatment (gAOP) receiving less attention. Yet, a gAOP process has a history of use in carton packaging decontamination whereby UV-C (254 nm) mediated degradation of 1% v/v hydrogen peroxide can support >5 log CFU inactivation of Bacillus endospores [23] [24] [25] . More recently, gAOP has been applied to inactivate pathogens such as Listeria monocytogenes on fresh fruit and vegetables [21, 26] . Here, the fresh produce to be decontaminated is passed through a hydrogen peroxide spray into a reactor where ozone gas is continuously applied under the illumination of UV-C lamps. The key advantages of the gAOP process is the high anti-microbial efficacy with the ability to inactivate microbes on the surface and sub-surface without negatively affecting the item being sanitized [21, 26] . In this respect, the gAOP process can be envisioned to combine the advantages of hydrogen peroxide vapor and UVGI in delivering a rapid method for decontaminating N95 masks without compromising functionality. A further requirement of an N95 mask decontamination method is the need to verify treatment performance via a biological and/or chemical indicator. The former is satisfied with the inactivation of endospores such as Bacillus pumilus (in the case of irradiation-based treatments) or G. stearothermophilus (for thermal treatments). It has been reported that B. pumilus and G. stearothermophilus have comparable sensitivity to hydroxyl-radicals when applied in solution thereby suggesting that either would be suitable index organisms [23] . With regards to chemical indicators, the AOP process for water disinfection is based on actinometry such as methylene blue [27] . The main advantage of methylene blue is that it exhibits semi-selectivity towards hydroxyl-radicals but relatively insensitive to hydrogen peroxide or UV-C alone [27] . Yet as a chemical indicator, methylene blue is degraded to colorless via the AOP reaction when a positive color change would make it easier to identify decontaminated items. In the following study, pyrrole was evaluated as a chemical indicator to report on the gAOP treatment. Pyrrole can undergo oxidative polymerization to form a conducting polymer with the degree of conjugation being reflective by darkening [28] . In the course of the polymerization process an anionic dopant is incorporated into the film to counter the positive charge on the polypyrrole backbone. Nafion was used as the doping ion in the current study as the surfactant has been reported to support a homogenous polymerization process and adherent polypyrrole film [29] . It has been reported that hydroxyl-radicals, along with other strong oxidants such as persulfate, can support the oxidative polymerization of pyrrole thereby leading to an irreversible transition from colorless to brown-black [29] . Hence, it is conceivable that the oxidative polymerization of pyrrole can be applied to report on the gAOP decontamination process. Escherichia coli K12 was obtained from ATCC (Atlanta, US) and was stored at -80˚C in tryptic soy broth (TSB; Thermo Fisher, Whitby, Canada) containing 20% w/v glycerol. The bacteria were recovered by streaking out onto a tryptic soy agar (TSA; Thermo Fisher) plate that was incubated at 37˚C for up to 48 h. A colony was transferred to TSB that was incubated overnight at 37˚C and used to inoculate TSB that was incubated overnight. The cells were harvested by centrifugation (5000g for 10 min) and the cell pellet resuspended in saline to give a final cell density of 9 log CFU/ml (OD 600 2.0 measured using a spectrophotometer; BioRad, Mississauga, Canada). The suspension of E. coli K12 was stored at 4˚C until required but used within a week. Geobacillus stearothermophilus ATCC 7963 was cultivated on a TSA slant that was incubated at 55˚C for 72 h. The growth was scraped off the agar layer and transferred to 10 ml sterilized distilled water and used to inoculate tissue culture flasks (100 cm 2 area) containing a layer of TSA. The flasks were closed and incubated at 55˚C for 10 days. The subsequent spores were recovered by flooding the plate with sterilized distilled water and scraping with a spreader. The spore suspension was transferred to a centrifuge tube and the pellet resuspended in sterile water then re-centrifuged (5000g for 10 min). The spores were washed in sterile water three times and the final spore pellet resuspended in 15 ml of sterile water then heattreated at 70˚C for 15 min to inactivate vegetative cells. A dilution series was prepared in sterile water and aliquots (0.1 ml) plated onto TSA that was incubated at 55˚C for up to 3 days to determine the spore density of the preparation. The spore suspension was then adjusted to a density of 7 log CFU/ml and stored at 4˚C until required. Human coronavirus 229E (HCoV-229E) was obtained from the culture collection of the Institute of Infectious Research within McMaster University (Ontario, Canada) and maintained in Opti-PRO SFM media (Thermo-Fisher). MRC-5 cells were seeded at 2 x 10 6 cells/ml into T75 flasks in DMEM supplemented with 10% w/v FBS, 1% w/v L-glutamine, and 1% w/v pen/strep. When the cells were 90-95% confluent, the cells were washed twice with prewarmed Opti-PRO SFM media and then infected with 5 ml of HCoV-229E in Opti-PRO SFM media at a multiplicity of infection of 0.1. Cells were incubated with virus for 80 min at 34˚C with 5% CO 2 . The cells were then washed twice with Opti-PRO SFM media and incubated with 15 ml of 3% w/v FBS DMEM containing 1% w/v sodium pyruvate, 1% w/v L-glutamine, and 1% w/v pen/strep for 5 days. On the fifth day, the supernatant was removed and clarified by centrifugation, and the virus preparation frozen at -80˚C. The TCIS 50 infectivity assay was performed by seeding Huh7.5 cells into a 96-well plate at 1.5 x 10 4 cells/well. When the cells were 80-90% confluent they were washed twice with Opti-PRO SFM and 50 μl of a serially diluted virus was added to the cells. The cells were incubated at 34˚C with 5% CO 2 for 5 days. On the fifth day, cell viability was assessed using Cell Titer Glo 2.0 reagent (Promega; Madison, United States). Aliquots (100 μl) Cell Titer Glo was added directly to the media, the plates were shaken for 2 minutes and then incubated for 10 min at room temperature. The luminescence was read on a Neo2 plate reader (Biotek; Winooski, United States) using luminescence fiber. The TCID 50 was calculated using log 50% endpoint dilution = log dilution showing mortality above 50%-(difference of logarithms x logarithm of dilution factor) [30] . The gAOP unit was provided by Clean Works Inc (Beamsville, Canada) (S1 Fig). In brief, the unit is constructed from stainless steel with the reactor housing ten UV-C lamps (23 W; 254 nm) positioned over a conveyor. The hydrogen peroxide solution was pumped to a spray head positioned at the entrance of the reactor. Ozone gas was generated by UV-184 nm lamps (12 W) positioned at either side of the unit with the gas being introduced at the conveyer level (S1 Fig). The residence time within the reactor was 30 s and controlled by the speed of the conveyor. The temperature within the reactor was maintained at 27-29˚C by heated air with the UV-C, ozone and hydrogen peroxide flow-rate being monitored via sensors (S1 Fig). A 3 x 3 cm area was drawn on either the outer or inner surface of the N95 masks. Aliquots (0.1 ml of ca. 7 log CFU/ml) of the test microbe was deposited on the inner or outer surface of the mask and allowed to dry for at least 1 h prior to treatment. The masks were loaded onto holding trays vertically or horizontally then passed through the gAOP unit that was operating with a hydrogen peroxide solution (2-6% v/v) delivered at a flow rate of 20-40 ml/min (S2 Fig). The UV-C lamps and ozone was continuously applied as the masks transitioned through the unit with a total treatment time of 30 s. The flatbed orientation was when the masks were passed through the gAOP unit twice and inverted between passes. The inoculated masks were left for 15 min before the inoculated area was excised using a sterile blade then transferred to sterile tube containing 20 ml saline. The cells/endospores were released from the mask sections by vortexing for 60 s and a dilution series prepared in saline or sterilized water. E. coli was enumerated on E. coli/coliform Petri Films (3M, London, Canada) that were incubated at 37˚C for 24 h. For G. stearothermphilus enumeration, the rinse solution was heated at 70˚C for 10 min to activate the endospores. A dilution series was prepared in sterile water and plated onto TSA that was then incubated at 55˚C for up to 7 days. The remaining rinse solution and excised mask section were added to an equal volume of TSB that was subsequently incubated at 55˚C for 7 days with daily checks for growth. HCoV-229E was inoculated onto 1 cm x 1 cm square sections drawn on masks and left to attach at room temperature for 5-30 min. The mask sections were transferred to 1 ml extraction buffer and vortexed every 5 min for 20 min at room temperature to extract the virus. The extracted virus was quantified using the TCID 50 assay pre-or post-gAOP treatment. Toxicity testing was performed by taking 1 cm x 1 cm sections of N95 masks with one set being used directly and the other passed through the gAOP process. The mask sections were transferred to extraction buffer at a ratio of 1 ml per 1 cm 2 mask section then incubated for 20 min at room temperature with intermittent vortexing. Samples were then serially diluted using media and incubated with cells for 5 days before assessing viability using Cell Titer Glo 2.0. E. coli K12 was cultivated in TSB overnight at 37˚C and cells harvested by centrifugation with the pellet being resuspended in saline to a final OD 600 0.2 (8 log CFU/ml). A dilution series was prepared in saline and plated onto MacConkey agar that was incubated for 24 h at 37˚C. The cell suspension was held at 4˚C then adjusted to 7 log CFU/ml once the plate count had been determined. Coupons (2 cm x 2 cm) were cut from 8210 N95 masks and used directly. The coupons were sub-divided into two groups. In the first group of coupons, the external surface was inoculated, while in the second group, the inner surface was inoculated. Each coupon was inoculated with 0.1 ml of inoculum and samples (N = 3) withdrawn at time 0, 15 min, 30 min, 45 min and 60 min. The individual coupons were placed in 10 ml of saline then vortexed for 30 s. A dilution series was prepared in saline then plated onto MacConkey agar that was incubated at 37˚C for 24 h. Sections (3 cm x 3 cm) were excised from N95 masks and inoculated with 0.1 ml of 7 log CFU/ ml G. stearothermophilus spore suspension. The sections (N = 3 per treatment) were dried at room temperature for 1 h before passing through the gAOP reactor with the hydrogen peroxide spray (3% v/v delivered at 40 ml/min), ozone (30 s), or UV-C (113 mJ/cm 2 ) switched on. Endospores were recovered from the treated samples and non-treated controls by suspending the sections in 20 ml saline and vortexing for 60 s. A dilution series was prepared and plated onto TSA that was subsequently incubated at 55˚C for 7 days. Hydrogen peroxide residues on N95 masks treated with the optimized gAOP treatment were determined using MQuant Peroxide test kit (Millipore Sigma, Oakville, ON, Canada). Here, N95 masks (N = 3) were passed through the gAOP unit operating at 3% v/v hydrogen peroxide at 40 ml/min with ozone being continuously introduced and a UV-C dose of 113 mJ/cm 2 . A 3 x 3 cm section of the N95 mask was excised and transferred to a tube containing 25 mL of distilled water then left for 1h at room temperature. The residual water was then decanted into a tube and hydrogen peroxide was determined using test strips with a low detection limit of 1.0 mg/l. The fit-related evaluations of each mask were performed by measuring the headband mechanical properties. Both upper and lower headbands from each N95 mask sample were evaluated for mechanical properties on an Instron model 5966 Universal Test System with a 1 kN load cell. Three elongation cycles were applied to each sample in the following order and magnitude: 200%, 50%, and 25%. These cycles simulate the donning and re-donning of a filtering facepiece respirator with non-adjustable elastic straps. The gAOP treated masks were also visually inspected for shrinkage, deformation of the nose-foam, and inner or outer shell. Three inoculation areas (1.5 cm diameter) were drawn on the exterior and interior of N95 masks, ensuring no overlapping between the areas (S3 Fig). Aliquots (0.1 ml) of spore suspension (7 log CFU/ml) were inoculated onto the designated areas then left for 1h at room temperature to attach. The masks were then passed through the gAOP unit and the inoculated sections were transferred to tubes containing TSB (20 ml). The spores were released by vortexing the tubes for 60 s and a dilution series prepared in saline that was subsequently plated onto TSA that was then incubated at 55˚C for up to 7 days to account for the delayed germination of super-dormant endospores. Apex Biological Indicator strips (MesaLabs, Bozeman, MT, USA) were applied in the current study. The stainless-steel strips are inoculated with Geobacillus stearothermophilus endospores (6 log CFU) on one face and have previously been applied for verification of hydrogen peroxide-based decontamination processes. Apex ribbon biological indicator strips were mounted onto N95 masks using two configurations (S4 Fig). In the first configuration, the Apex strips were attached to the outer and inner surface of the masks using double-sided tape. In Configuration 2, a window was cut from the center of the N95 masks with the Apex strip positioned with the inoculated side facing up with the other inverted (S4 Fig). In both configurations, the six masks were placed into the holder face-up and passed through the gAOP unit operating under the optimized working parameters. The Apex strips were removed from the masks after treatment and transferred to sterile tubes into which 10 ml of TSB was added. The tubes were incubated for 48h at 55˚C then checked for growth of residual survivors as per manufacture instructions. The chemical indicator strips were fabricated by depositing 10 μl of Nafion resin 1100W (Sigma-Aldrich, ON, Canada) onto strips (1 cm x 3 cm) or disks (1 cm diameter) of thick blotting (BioRad, ON, Canada). Aliquots (10 μl) of pyrrole (Sigma-Aldrich, ON, Canada) were deposited over the Nafion and the strip/disk attached to the masks via double-sided tape then used within 10 min. The indicator was positioned within the interior and exterior of the mask with a positive reaction being recorded by a transition from pink to black. Trials were also performed with chemical indicator strips passed through the gAOP reactor when the individual components (hydrogen peroxide, UV-C or ozone) were applied alone and then in combination. The color change of the chemical indicator was then recorded. Values represent the average of 3-6 masks per treatment with log counts of bacteria or TCID 50 values compared using one-way ANOVA in combination with the Tukey test. Statistical analysis was performed using IBM SPSS statistics software (Armok, NY, USA). Analysis of variance was used to test P � 0.05 was considered statistically significant. Fit and particulate testing represents results of 10 masks for control, ten or twenty passes through the gAOP reactor. The gAOP treatment was optimized in terms of mask orientation, hydrogen peroxide concentration, and flow rate with the UV-C dose and concentration of ozone remaining constant (Table 1; S2 Fig) . The inactivation of E. coli was used as a metric to assess the different treatments given that the bacterium exhibited higher tolerance to hydroxyl-radicals compared to enveloped viruses [31] . E. coli was inoculated onto the exterior or interior of masks then orientated in different positions during passage through the gAOP unit operating at different hydrogen peroxide flow rates (Table 1) . With non-treated controls, the recovery of E. coli from the interior of the N95 masks were significantly (P<0.05) lower compared to the exterior even though the same cell density was inoculated (Table 1) . To determine the underlying reasons for the lower cell density of E. coli recovered from the internal surfaces of N95 masks trials were performed to determine the fate of the bacterium when inoculated inner or outer surface then held for 60 min (Fig 1; S2 Table) . It was found that E. coli inoculated onto the outer surface of masks retained viability with no significant change in counts over the 60 min holding period. However, E. coli inoculated into the internal surface of N95 masks progressively declined over the 60 min holding period and resulted in a 2.02 log CFU reduction (Fig 1) . It has previously been reported that the viability of E. coli and viruses decreases over time when introduced onto N95 masks presumably due to the electrostatic filter layers [32] . It was found that the inactivation of E. coli on N95 masks by gAOP treatment was independent of the hydrogen peroxide concentration (2-6% v/v) applied with no E. coli being recovered treated masks. It has previously been reported that a combination of hydrogen peroxide at 1-2% v/v combined with UV-C can support a 5 log CFU reduction of Bacillus endospores illustrating the high antimicrobial activity of hydroxyl-radicals [23] . In subsequent studies, a 3% v/v hydrogen peroxide solution was applied for the gAOP treatment given that the concentration is commercially widely available without the need for dilution. The flow rate of hydrogen peroxide within the gAOP unit and orientation of the N95 masks during treatment affected the decontamination efficacy (Table 1) . Specifically, residual E. coli survivors were recovered from N95 masks treated with gAOP operating at 20 ml/min (Table 1) . However, no E. coli was recovered from N95 masks treated with gAOP when the hydrogen peroxide was introduced at 40 ml/min ( Table 1 ). The results indicate that the efficacy of the gAOP was dependent on the degree of misting by hydrogen peroxide to cover the N95 masks and to generate sufficient hydroxyl radicals to inactivate E. coli. The importance of misting was also highlighted by the effect on the orientation of the mask. N95 masks passed through the gAOP unit horizontally (Treatment 4 and 5; Table 1 ) resulted in a higher decontamination efficacy compared with those positioned vertically (Treatment 2 and 3; Table 1 ). No E. coli were recovered from masks that had been double passed (flatbed) through the gAOP being inverted after the first pass (Treatment 1; Table 1 ). The effect of position again reflects the exposure to hydroxyl radicals to the surfaces on the outside and the internal surface of masks. The orientation of masks is critical in UVGI treatments given that the surfaces are required to be directly exposed to the UV-C photons [12] . With hydrogen peroxide vapor the orientation is less critical but there is a need to facilitate free-flow of the antimicrobial gas around the masks [6] . In a similar manner, gAOP treatment requires the circulation of hydroxyl-radicals to the surfaces of the mask but do not need to be directly exposed to UV as with gas plasma treatment [33] . Unlike E. coli, the levels of endospores recovered from the internal surface of the N95 masks were not significantly different (P>0.05) compared to the inner layers ( Table 2) . No surviving endospores were recovered from N95 masks passed through the gAOP unit thereby verifying the antimicrobial efficacy of the process ( Table 2 ). The results are in agreement with studies performed on the sporicidal activity of hydroxyl-radical activity in carton sterilization [25] . To verify that endospore inactivation was via hydroxyl-radicals trials were performed to assess the lethality of the individual components of the gAOP process. Here, mask sections were inoculated with Geobacillus endospores were passed through the reactor with only the hydrogen peroxide spray (3% v/v at 40 ml/min), ozone or UV-C lamps (dose 113 mJ/cm 2 ) turned on. The log count reductions obtained were 0.97±0.07 log CFU for hydrogen peroxide, 0.10±0 log CFU for ozone and 0.93±0.09 log CFU for UV-C alone. The results confirm that the generation of hydroxyl-radicals exhibited a synergistic lethality compared to the individual component parts. The initial TCID 50 of CoV-229E inoculated onto N95 masks was determined to be >2 x 10 8 with no survivors being detected on samples treated with gAOP (S3 Table) . The results confirm the susceptibility of coronavirus to hydroxyl-radicals [31] . There was no significant difference (P>0.05) in the viability of Huh7.5 cells exposed to the extract taken from mask sections that had been treated with gAOP (0.939± 0.017 RLU) relative to non-treated controls (0.867 ±0.012 RLU) (S4 Table) . The result verifies that no toxic products were generated as a result of the gAOP process. The penetration test illustrated that N95 masks treated with the gAOP process were not significantly different (P>0.05) from non-treated controls and within the 0.1-2.0% tolerance limits ( Table 3 ). The insignificant difference in the pressure drop across gAOP treated masks compared to controls also provided evidence that the N95 masks retained functionality. The headband properties of the gAOP treated N95 masks were not significantly (P>0.05) differ compared to the controls and no visual deformation, shrinkage, or change in texture was found. Collectively the results confirmed the N95 masks treated up to 20 cycles in the gAOP unit did not alter the performance or fit of masks. Surgical masks were inoculated at different positions on the interior and exterior of surgical masks then passed through the gAOP reactor operating under optimized conditions (Table 4 ; The biological indicator strips were affixed to the internal or external surface of N95 masks or bridging a cut-out section (S4 Fig). The masks were then passed through the gAOP reactor operating under the developed working parameters (3% v/v hydrogen peroxide delivered at 40 ml/min with a UV-C dose of 113 mJ/cm 2 with ozone being applied throughout the 30 s process). None of the Apex biological indicator strips tested negative when passed through the gAOP reactor irrespective of being position in Configuration 1 or 2 (S4 Fig). In contrast, the non-treated Apex biological strips tested positive thereby confirming that the Geobacillus endospores were viable. The results confirm that the Biological Indicator Geobacillus endospore strips could be applied for routine verification of the gAOP process. The chemical indicator serves to differentiate between those N95 masks that have been treated using gAOP from those awaiting treatment. Besides, indicators also serve to verify that the treatment has been adequately applied so can be viewed as a complement to biological indicators. In the current study, the Nafion-Pyrrole indicator transitioned from pink to blueblack when passed through the gAOP reactor operating under the aforementioned working parameters. However, when each of the inputs to the gAOP process (i.e UV-C, hydrogen peroxide, or ozone) were applied alone the color change was less distinct and incomplete (Fig 2) . The results are in agreement with other reports demonstrating that the generation of hydroxyl-radicals can support the oxidative polymerization of pyrrole [29, 34] . The study was directed towards validating and verifying the efficacy of a gAOP to decontaminate N95 and surgical masks without resulting in detrimental changes in functionality. In addition, the utility of a biological and chemical indicator to verify decontamination performance was also performed. It was found that the gAOP treatment could support the inactivation of human coronavirus, vegetative bacterial cells, and endospores without causing negative effects on mask functionality. Given the continuous, rapid, nature of the process, it could be envisaged that the gAOP unit could be applied on-site (for example, hospital ward) for recycling or extending the use of masks. The different operating parameters (hydrogen peroxide, ozone, and UV-C dose) can be continuously monitored in real-time thereby resulting in a controlled process. Yet, a biological and chemical indicator has been identified in the current study that can be applied for process verification. The current study tested a limited number of masks due to availability issues and future studies will be looking at a broader range of N95 mask types. The COVID-19 pandemic, personal protective equipment and respirator: A narrative review. International Journal of Clinical Practice Challenges and solutions for addressing critical shortage of supply chain for personal and protective equipment (PPE) arising from Coronavirus disease (COVID19) pandemic-Case study from the Republic of Ireland Availability of Personal Protective Equipment (PPE) Among US and Pakistani Doctors in COVID-19 Pandemic A rapid systematic review of the efficacy of face masks and respirators against coronaviruses and other respiratory transmissible viruses for the community, healthcare workers and sick patients Disposable masks: Disinfection and sterilization for reuse, and non-certified manufacturing, in the face of shortages during the COVID-19 pandemic Development of a highly effective low-cost vaporized hydrogen peroxide-based method for disinfection of personal protective equipment for their selective reuse during pandemics Important regulatory consideration for the reprocessing of single use N95 respirators during the CoVid-19 response Effect of various decontamination procedures on disposable N95 mask integrity and SARS-CoV-2 infectivity Evaluation of five decontamination methods for filtering facepiece respirators Filter quality of electret masks in filtering 14.6-594 nm aerosol particles: Effects of five decontamination methods Effects of Ultraviolet Germicidal Irradiation (UVGI) on N95 respirator filtration performance and structural integrity Ultraviolet-C and other methods of decontamination of filtering facepiece N-95 respirators during the COVID-19 pandemic A pandemic influenza preparedness study: use of energetic methods to decontaminate filtering facepiece respirators contaminated with H1N1 aerosols and droplets The importance of the minimum dosage necessary for UVC decontamination of N95 respirators during the COVID-19 pandemic. Photodermol Photoimmunol Photomed Invited editorial from the social media editor of JPRAS; Leadership in the time of COVID-19 Addressing decontaminated respirators: Some methods appear to damage mask integrity and protective function Decontamination and reuse of N95 respirators with hydrogen peroxide vapor to address worldwide personal protective equipment shortages during the SARS-CoV-2 (COVID-19) pandemic Decontamination of surgical face masks and N95 respirators by dry heat pasteurization for one hour at 70 degrees C Dry heat and microwavegenerated steam protocols for the rapid decontamination of respiratory personal protective equipment in response to COVID-19-related shortages Application of a quaternary ammonium agent on surgical face masks before use for pre-decontamination of nosocomial infection-related bioaerosols Inactivation of Salmonella, Listeria monocytogenes, Aspergillus and Penicillium on lemons using advanced oxidation process optimized through response surface methodology Advanced oxidation processes (AOP) for water purification and recovery The combined effect of hydrogen peroxide and ultraviolet irradiation on bacterial spores Carton sterilization by uv-C excimer laser light: recovery of Bacillus subtilis spores on vegetable extracts and food simulation matrices Combined treatment with hydrogen peroxide and UV irradiation to reduce microbial contamination levels in performed food packaging cartons Inactivation of Listeria monocytogenes on and within apples destined for caramel apple production by using sequential forced air ozone gas followed by a continuous advanced oxidative process treatment Advanced Oxidation Process with peracetic Acid and Fe(II) for contaminant degradation Organic conducting polymer electrode based sensors for detection of Salmonella infecting bacteriophages A short report on the polymerization of pyrrole and its copolymers by sonochemical synthesis of fluorescent carbon dots A simple method of estimating fifty percent endpoints Broad spectrum microbicidal activity of photocatalysis by TiO 2 . Catalysts Evaluation of the survivability of microorganisms deposited on filtering respiratory protective devices under varying conditions of humidity Cold plasma-activated hydrogen peroxide aerosol inactivates Escherichia coli O157:H7, Salmonella Typhimurium, and Listeria innocua and maintains quality of grape tomato, spinach and cantaloupe The sonochemical synthesis of Ga C-dots particles We thank Clean Works Inc for providing access to a Clean Flow unit and 3M for undertaking the mask evaluation testing.