key: cord-0792582-u9qynafz authors: Ishida, Yusuke; Okada, Toshio; Kobayashi, Takayuki; Uchino, Hiroyuki title: ORi™: a new indicator of oxygenation date: 2021-04-26 journal: J Anesth DOI: 10.1007/s00540-021-02938-4 sha: 60078c268710029827dd1e89fa143492266ccfd4 doc_id: 792582 cord_uid: u9qynafz In the perioperative period, hypoxemia and hyperoxia are crucial factors that require attention, because they greatly affect patient prognoses. The pulse oximeter has been the only noninvasive monitor that can be used as a reference of oxygenation in current anesthetic management; however, in recent years, a new monitoring method that uses the oxygen reserve index (ORi™) has been developed by Masimo Corp. ORi is an index that reflects the state of moderate hyperoxia (partial pressure of arterial oxygen [PaO(2)] between 100 and 200 mmHg) using a non-unit scale between 0.00 and 1.00. ORi monitoring performed together with percutaneous oxygen saturation (SpO(2)) measurements may become an important technique in the field of anesthetic management, for measuring oxygenation reserve capacity. By measuring ORi, it is possible to predict hypoxemia and to detect hyperoxia at an early stage. In this review, we summarize the method of ORi, cautions for its use, and suitable cases for its use. In the near future, the monitoring of oxygen concentrations using ORi may become increasingly common for the management of respiratory function before, after, and during surgery. Over the years, intraoperative anesthetic monitoring techniques have been improved similarly to the improvement of surgical instruments. Blood pressure, heart rate, electrocardiography, and blood oxygen saturation level are mainly monitored during intraoperative anesthesia. In the present article, we review the oxygen reserve index (ORi™), which is a measurement that enables the assessment of oxygen reserve capacity, and discuss its various applications and future outlooks. During anesthetic management, predicting the partial pressure of arterial oxygen (PaO 2 ) is difficult when percutaneous oxygen saturation (SpO 2 ) is 100%. In usual cases, in which arterial lines of the patients are secured, arterial blood gas analysis for the appropriate management of PaO 2 is possible. However, as this is an invasive method, it cannot be performed on all patients. In most patients undergoing general anesthesia, tracheal intubation and artificial ventilation are carried out to manage positive-pressure ventilation. However, during surgery, atelectasis and pulmonary injury may occur owing to multiple factors [1] [2] [3] , which may cause sudden hypoxemia during or after surgery, making it difficult to manage the patient [1, 4, 5] . Therefore, PaO 2 during surgery is a useful indicator for anesthesiologists, particularly in patients with low respiratory function (e.g., obstructive/ restrictive ventilatory disorder) and low oxygen reserve (e.g., pregnant women and obese patients) [6, 7] . In addition, controlling oxygen concentrations during surgery in patients with diseases in which hyperoxia should be avoided, such as interstitial pneumonia and retinopathy of prematurity [8] , is a difficult problem. Although hypoxia can lead to adverse effects, hyperoxia during surgery can also induce pulmonary injury and absorption atelectasis, resulting in postsurgical respiratory complications [1] [2] [3] . Furthermore, hyperoxia can also increase the risk of other adverse events [9] (Table 1) . Unfavorable clinical outcomes owing to a high fraction of inspired oxygen (FiO 2 ) have been recently reported in critically ill adults, including patients with chronic obstructive pulmonary disease, myocardial infarction, cardiac arrest, stroke, and traumatic brain injury [10] [11] [12] [13] [14] [15] . Until recently, the pulse oximeter was the only noninvasive instrument to measure arterial blood oxygen saturation (SaO 2 ), which is an index of blood oxygenation. During the 1970s, the world's first pulse oximeter that was able to measure SaO 2 from the fingertip was commercialized. Since then, this device has been used in a wide range of situations for the noninvasive monitoring of blood oxygenation [16] . However, this device has its flaws. Although SpO 2 and PaO 2 are correlated at most points, when SpO 2 reaches 100%, estimating PaO 2 becomes difficult [17] (Fig. 1) . Thus, other methods, such as arterial blood gas analysis, are often performed to measure PaO 2 for the evaluation of oxygenation. The oxygen reserve index (ORi) is a noninvasive and continuous parameter that can be measured by connecting a multi-wavelength sensor to Root ® with Radical-7 ® , Pulse CO-Oximeter ® (Masimo Corp., Irvine, CA, USA). ORi is an index measured using a non-unit scale between 0.00 and 1.00, which reflects the range of moderate hyperoxia (PaO 2 between 100 and 200 mmHg), and is defined as oxygen reserve reflecting the level of venous blood oxygen saturation (SvO 2 ). ORi remains at 0.00 in a room air environment. With supplemental oxygen, after SpO 2 reaches 100%, SvO 2 also increases as PaO 2 does, and usually reaches a plateau at about a PaO 2 of 200 mmHg [18, 19] (Fig. 2) . In general, a pulse oximeter displays the SpO 2 value measured from light absorption of the pulsatile blood at the site where the sensor is placed. Pulsatile changes are observed in the arteries, capillaries, and venules. Masimo developed rainbow SET ® which can analyze changes in the absorption of incident light of both arterial and venous blood using a multi-wavelength sensor, and display them as ORi, which is a unitless scale between 0.00 and 1.00. Thus, ORi begins to increase from 0.00 at about a PaO 2 of 100 mmHg, and reaches a plateau of 1.00 at approximately 200 mmHg [18] [19] [20] [21] . However, there are reports stating that in cases where FiO 2 is 1.0, even when (Fig. 3) . However, if PaO 2 is higher than 240 mmHg, there was no correlation (r 2 = 0.0016). Furthermore, when ORi was more than 0.24, PaO 2 was more than 100 mmHg. As the report showed that 96.6% of the samples had a PaO 2 of ≥ 150 mmHg when ORi was above 0.55, the authors concluded that ORi can be used as an indicator of PaO 2 . Vos et al. investigated the correlation between PaO 2 and ORi during the administration of oxygen in 20 healthy volunteers, and they also reported that the two values have a strong positive correlation [24] . In addition, regarding the correlation between PaO 2 and ORi during anesthesia under oxygenation, when PaO 2 was less than 240 mmHg (n = 69), linear regression analysis showed a relatively strong positive correlation (r 2 = 0.706) [25] . The following is a list of cases for which the use of ORi is suitable. • Rapid sequence induction Normally in a room air environment, ORi is 0.00. After the administration of oxygen, even after SpO 2 reaches 100%, ORi will continue to increase. Rapid sequence induction (RSI) is appropriate for patients in whom the aspiration of gastric contents may occur, including those with a full stomach, ileus, and expectant mothers [26] [27] [28] . In RSI, tracheal intubation is performed without ventilation after the occurrence of apnea. In such cases, ORi may be an indicator of the time it will take for SpO 2 to decrease during tracheal intubation. Yoshida et al. investigated the usefulness of ORi during RSI [29] . They reported that in 10 out of 17 patients (77%) in whom preoxygenation was performed, ORi started to decrease 32.5 s (IQR: 18.1-51.3) before a decrease in SpO 2 post-apnea. By monitoring the ORi, it is possible to predict a decrease in oxygenation about 30 s before SpO 2 starts to decrease. From this finding, Yoshida concluded that it may be possible to lower the incidence rate of hypoxemia during RSI. Thus, ORi is expected to be useful for RSI. In addition, owing to the recent outbreak of coronavirus disease 2019 (COVID-19), RSI has been recommended to prevent the infection of anesthesiologists during intubation [30, 31] . Furthermore, the use of aerosol boxes for the prevention of COVID-19 infection makes intubation difficult [32] [33] [34] . Thus, aerosol boxes may increase intubation times and therefore increase the risk of hypoxia in patients [32] . Using ORi may hence reduce the risk of hypoxemia and enable safe tracheal intubation. In pediatric patients, oxygen consumption is higher, and the closing volume is also relatively larger than the functional residual capacity compared with adult patients. Owing to this fact, the development of hypoxemia in pediatric patients is more rapid than in adult patients during apnea. Previous studies investigating ORi in pediatric patients during anesthetic induction showed that ORi monitoring enabled detection of a reduction in oxygenation approximately 30 s prior to a rapid change in SpO 2 [21] . These studies suggest that prediction of the onset of a rapid decrease in oxygen saturation enables the prompt implementation of appropriate treatments, which is an advantage. In fact, the use of ORi for tracheoplasty in children was found to enable the earlier prediction of hypoxemia [35] . In addition, as the trachea is shorter in children than in adults, when performing tracheal intubation and head bending during surgery, there is a higher possibility of unilateral intubation in children [36, 37] . Therefore, in the former case, comparison of ORi before and after intubation may be useful, and in the latter case, a sudden drop in ORi during the operation may be a useful indicator of subsequent hypoxemia. Therefore, the use of ORi may enable the earlier implementation of appropriate treatments. All of the above demonstrate that ORi is a useful indicator in pediatric anesthesia during the perioperative period. • Other suitable cases [23] . PaO 2 partial pressure of arterial oxygen, ORi oxygen reserve index ORi is also useful for patients in whom respiration management is required during one-lung ventilation (OLV). Hypoxemia may still occur in about 10% of such patients [38] . When switching from two-lung to OLV, the shunt fraction increases, oxygenation is impaired, and hypoxemia may occur [39] . Studies monitoring ORi during the anesthetic management of OLV showed that decreases in PaO 2 could be detected much earlier than when only SpO 2 was monitored. These studies suggest that ORi can reduce the risk of complications during OLV [40] . Another study reported that the monitoring of ORi during OLV could predict hypoxemia [41] . These findings may be helpful to adjust FiO 2 individually in patients undergoing OLV, and to avoid unnecessarily high concentrations of oxygen. In addition, the use of ORi is suitable for surgeries that have a high probability of leading to atelectasis, such as surgeries on obese patients [42, 43] , surgeries in the Trendelenburg position, and laparoscopic surgeries [43] [44] [45] . Other studies also suggested that ORi is useful in situations in which awake intubation is performed [22] . ORi monitoring may be able to avoid hypoxemia in cases in which bronchial stenting using a rigid endoscope is performed while the patient continues to breathe spontaneously [46] . For surgeries in which patients' respiratory statuses can vary, ORi monitoring can be an indicator to decide whether or not early intervention is required to avoid hypoxemia [47] . ORi monitoring during the induction of anesthesia can be an indicator of whether preoxygenation is being carried out appropriately. In addition, hyperoxia can be detected by monitoring ORi in patients under general anesthesia, and thus the administration of unnecessarily high concentrations of oxygen can possibly be avoided. In fact, there are several reports demonstrating that the use of ORi in managing oxygen levels during and after surgery, and for patients in the intensive care unit (ICU) can prevent exposure to excess oxygen [25, 48, 49] . In general, high oxygen concentrations are more harmful to infants, children, and older patients than to adults [50] . Oxidative stress that occurs as a consequence of treatment with high oxygen concentrations during the perinatal period notably affects the lung and brain development of premature and newborn infants. For instance, bronchopulmonary dysplasia is caused by hyperoxic lung injury [51] [52] [53] . In older patients, the production and accumulation of reactive oxygen species (ROS) are largely increased, and antioxidative functions are decreased, resulting in various aging-associated diseases [54, 55] . Hyperoxia promotes ROS production, suggesting that perioperative hyperoxia might enhance the adverse effects of aging-associated diseases, particularly in lung and brain diseases. Therefore, adequate oxygenation using ORi may lead to a reduction in complications. ORi is measured by placing an RD rainbow Lite SET ® sensor, which is a multiwavelength sensor, on a patient's fingertip and connecting it to Root ® with Radical-7 ® . In addition to ORi, this sensor can measure a patient's SpO 2 , pulse rate (PR), Perfusion Index (Pi), and Pleth Variability Index (PVi ® ) (Fig. 4) . In addition, ORi can be measured by a RD rainbow SET ® -2 sensor, which can measure Noninvasive and Continuous Hemoglobin (SpHb ® ). However, it is necessary to block any outside light when using the SpHb sensor, as this sensor is easily affected by light owing to its high sensitivity to long wavelengths (Fig. 4) . Furthermore, special attention is needed to make sure that the SpO 2 is greater than 98% because if this criterion is not established, ORi does not show a positive number. One caution during monitoring is that measurement is difficult when the patient's body is moving or when the blood perfusion rate is very low. Furthermore, ORi values may fluctuate if a pigmented drug is administered intravenously. There is a report showing that ORi rapidly decreased after indigo carmine was Fig. 4 ORi function built into Root ® with Radical-7 ® . a ORi, b SpO 2 , c blockage of light is necessary to measure ORi using a SpHb sensor. ORi oxygen reserve index, SpO 2 percutaneous oxygen saturation administered intravenously [56] . Another report showed that when indocyanine green (ICG) was administered to 10 patients, ORi increased in all of the patients [57] . The authors of this study suggested that it is hence important to take into account an increase in ORi when using ICG during surgery. ORi values can also fluctuate for other unknown reasons, and hence it is also important to make a comprehensive assessment of the patient's condition and possible effects. In addition, the Radical-7 software version for the calculation of ORi and the RD rainbow Lite SET sensor were updated in January 2018, and although the peak is different for each individual, a plateau is reached at about 1.00 in most cases, and the artifact elimination function has been improved compared with the previous version. It is hence important to note that the obtained ORi values may differ depending on the version of the device used, so care must be taken when interpreting the results of previous studies. As mentioned above, in recent years, ORi has been used as an index of oxygenation during surgery. However, in the near future, ORi may also be used in situations in which changes in oxygenation occur frequently, such as in patients in the ICU and general wards. Yoshida et al. suggested the usefulness of ORi for patients in the ICU, when suctioning through an endotracheal tube is performed during ventilation [58] . Furthermore, the fact that the ORi device is noninvasive and easy-to-wear will enable its use in various clinical situations. Recently, COVID-19 has become a worldwide epidemic. ORi might be useful to prevent infections, such as COVID-19, by reducing the frequency of contact of clinical workers with patient blood when performing blood gas analysis or securing an arterial line. The use of ORi will not only enable the early prediction of hypoxemia, but may also be useful for avoiding hyperoxia. In recent years, there have been reports that the mortality rate of patients managed for hyperoxia is high [59] . Thus, in the near future, ORi monitoring may become an effective method to lower the adverse effects caused by high oxygen concentrations. ORi not only determines the optimal inhaled oxygen concentration during the perioperative period, but also acts as an index of proper oxygenation, and has the potential to help avoid complications caused by hypoxemia and hyperoxia. Studies on how the use of ORi can change patient prognoses should be performed in the near future. Postoperative pulmonary complications Optimal oxygen concentration during induction of general anesthesia Oxygen concentration and characteristics of progressive atelectasis formation during anaesthesia Does intraoperative lungprotective ventilation reduce postoperative pulmonary complications? Intraoperative ventilation strategies to prevent postoperative pulmonary complications: systematic review, meta-analysis, and trial sequential analysis Preoxygenation: physiologic basis, benefits, and potential risks How to preoxygenate in operative room: healthy subjects and situations "At Risk Retinopathy of prematurity The New World Health Organization recommendations on perioperative administration of oxygen to prevent surgical site infections: a dangerous reductionist approach? Arterial hyperoxia and mortality in critically ill patients: a systematic review and meta-analysis Air versus oxygen in ST-segmentelevation myocardial infarction Effect of high flow oxygen on mortality in chronic obstructive pulmonary disease patients in prehospital setting: randomised controlled trial Oxygen therapy in the pre-hospital setting for acute exacerbations of chronic obstructive pulmonary disease Emergency Medicine Shock Research Network (EMShockNet) Investigators. Relationship between supranormal oxygen tension and outcome after resuscitation from cardiac arrest Relationship between arterial partial oxygen pressure after resuscitation from cardiac arrest and mortality in children Takuo Aoyagi: discovery of pulse oximetry Mastering ABGs. The art of arterial blood gas measurement The oxygen reserve index (ORI): a new tool to monitor oxygen therapy Correction to: the oxygen reserve index (ORI): a new tool to monitor oxygen therapy Oxygen reserve index, a new method of monitoring oxygenation status: what do we need to know Oxygen reserve index: a novel noninvasive measure of oxygen reserve-a pilot study Oxygen reserve index: a novel and noninvasive method for monitoring oxygenation The relationship between oxygen reserve index and arterial partial pressure of oxygen during surgery Oxygen reserve index: validation of a new variable Adjustment of oxygen reserve index (ORi™) to avoid excessive hyperoxia during general anesthesia Rapid sequence induction Rapid sequence induction for cesarean section Rapid sequence induction and intubation in patients with risk of aspiration: recommendations for action for practical management of anesthesia Usefulness of oxygen reserve index (ORi™), a new parameter of oxygenation reserve potential, for rapid sequence induction of general anesthesia Guiding airway management and personal protective equipment for COVID-19 intubation teams A practical guide for anesthesia providers on the management of coronavirus disease 2019 patients in the acute care hospital The aerosol box for intubation in coronavirus disease 2019 patients: an insitu simulation crossover study A comparison of droplet and contact contamination using 3 simulated barrier techniques for COVID-19 intubation: a quality assurance study COVID-19 aerosol box as protection from droplet and aerosol contaminations in healthcare workers performing airway intubation: a randomised cross-over simulation study Oxygen reserve index (ORi™) as an alarm for oxygenation deterioration in pediatric tracheostomaplasty: a case report Detection and correction of endotracheal-tube position in premature neonates A comparative study of endotracheal tube positioning methods in children: safety from neck movement One-lung ventilation and arterial oxygenation Hypoxemia during one-lung ventilation: prediction, prevention, and treatment Monitoring the oxygen reserve index can contribute to the early detection of deterioration in blood oxygenation during one-lung ventilation Oxygen reserve index predicts hypoxemia during one-lung ventilation: an observational diagnostic study Study of early warning for desaturation provided by oxygen reserve index in obese patients Anesthetic management of a morbidly obese patient with endometrial cancer during robot-assisted laparoscopic surgery Atelectasis in general anesthesia and alveolar recruitment strategies Intraoperative ventilatory strategies for prevention of pulmonary atelectasis in obese patients undergoing laparoscopic bariatric surgery Oxygen reserve index (ORi™) contributes to prediction of hypoxemia and patient safety during tracheal stent insertion using rigid bronchoscopy: a case report Oxygen reserve index: utility as an early warning for desaturation in high-risk surgical patients ORI monitoring allows a reduction of time with hyperoxia in critically ill patients: the randomized control ORI 2 study The oxygen reserve index as a determinant of the necessary amount of postoperative supplemental oxygen Perioperative use of oxygen: variabilities across age Encouraging pulmonary outcome for surviving, neurologically intact, extremely premature infants in the post surfactant era Pulmonary complications of oxygen therapy Aging might augment reactive oxygen species (ROS) formation and affect reactive nitrogen species (RNS) level after myocardial ischemia/reperfusion in both humans and rats Aging aggravates nitrate mediated ROS/RNS changes Effects of indigo carmine intravenous injection on oxygen reserve index (ORi™) measurement Effects of intravenous injection of indocyanine green on the oxygen reserve index (ORi™) Making use of the oxygen reserve index (ORi™): a new parameter of oxygenation reserve potential The impact of hyperoxia on outcome of patients treated with noninvasive respiratory support We would like to thank T. W. L. Scheeren, M.D., PhD, Professor of Anesthesiology, University Medical Center Groningen, and Springer Nature, for giving permission for using Fig. 2 included in this article. We would like to thank Richard L. Applegate II, M.D., FASA, Professor and Chair, Anesthesiology and Pain Medicine, University of California Davis, and Wolters Kluwer Health, for giving permission for using Fig. 3 included in this article. We would like to thank Helena Akiko Popiel and Rikako Tomita for English language editing. We are deeply grateful to Koichi Nakazawa, Hiroshi Hamada, Kenji Kurita, Aya Kawachi, Kei Tojimbara, Rieko Yamato and Yusuke Chiba for helpful discussions and comments on the manuscript. Author contributions YI conducted the literature search and analyzed the data and wrote the draft. TO and TK helped review all the literature included. HU was the overall supervisor of this review. All authors read and approved the final manuscript.Funding None.Availability of data and materials Not applicable. The authors declare that they have no competing interests associated with this manuscript.Ethics approval and consent to participate Not applicable. Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.